Mixing studies for lupus anticoagulant: mostly no, sometimes yes
Abstract
Mixing tests have long been a mainstay in the lupus anticoagulant (LA) testing armoury of screen, mix and confirm assays. If a sample with an elevated screening test does not evidence inhibition in the mixing test, the search for an LA is halted and a different diagnostic pathway embarked upon. Recent years have seen studies evidencing sometimes high frequencies of false-negative mixing tests with perhaps sinister implications for missed diagnoses and skewed patient management. Issues such as the dilution effect, between-reagent sensitivity and specificity differences, variability of normal pooled plasma (NPP) quality and suitability and interpretive inconsistencies all contribute to questioning the reliability of mixing tests and their pivotal place in the LA assay hierarchy. The advent of integrated testing, where phospholipid-dependence is demonstrated or excluded prior to any attempt to evidence inhibitory properties with a fallible analytical principle, provides an alternative path to LA detection. In the absence of other causes of elevated clotting times, LA assay screen and confirm discordance is sufficient to secure a laboratory diagnosis of the presence of an LA, leaving the mixing test in a supplementary yet valuable role when further diagnostic discrimination is required.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A majority of ISTH-SSC survey respondents, 84.1%, affirmed the need for mixing tests.
“The need for mixing tests is reflected in the affirmative response of majority of the ISTH-SSC survey respondents (84.1%) and recent literature.15,61,62”
Mixing tests are performed to improve the specificity of LA detection and rule out other causes of prolonged clotting time, including coagulation factor deficiency, anticoagulant use, and coagulation factor inhibitors.
“Mixing tests are performed to improve the specificity of LA detection and rule out other causes of prolonged clotting time (CT), including coagulation factor deficiency, anticoagulant use, and coagulation factor inhibitors [15,16].”
From Comparison of different algorithms for lupus anticoagulant detection: a single-center experience (2024) · cites this paper as [15]
MTC and ICA can be calculated and employed as interpretation indices for mixing tests.
“In addition, the mixing test–specific cutoff (MTC) and index of circulating anticoagulant (ICA) can be calculated and employed as interpretation indices for the mixing tests [[15], [16], [17], [18], [19]].”
From Comparison of different algorithms for lupus anticoagulant detection: a single-center experience (2024) · cites this paper as [15]
The unavoidable dilution effect of mixing tests on antibody titer can make LA undetectable, leading to false-negative results.
“The principal limitation regarding mixing is the unavoidable dilution effect on the antibody titer, which in some cases can make the LA undetectable, leading to false-negative results [104].”
From Lupus Anticoagulant Detection under the Magnifying Glass (2023) · cites this paper as [104]
Cites
Update of the guidelines for lupus anticoagulant detection
PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M., & DE GROOT, P. G. (2009) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03555.x · Cited by 1,197 · see citing works
Guidelines on the investigation and management of antiphospholipid syndrome
Keeling, D., Mackie, I., Moore, G. W., Greer, I. A., Greaves, M., & British Committee for Standards in Haematology (2012) · British Journal of Haematology · doi:10.1111/j.1365-2141.2012.09037.x · Cited by 740 · see citing works
Self-citationLupus Anticoagulant Testing: Performance and Practices by North American Clinical Laboratories
Dembitzer, F. R., Ledford Kraemer, M. R., Meijer, P., & Peerschke, E. I. B. (2010) · American Journal of Clinical Pathology · doi:10.1309/ajcp4spplg5xvixf · Cited by 53 · see citing works
Interpretation of normal plasma mixing studies in the laboratory diagnosis of lupus anticoagulants
Devreese, K. M. J. (2007) · Thrombosis Research · doi:10.1016/j.thromres.2006.03.012 · Cited by 46 · see citing works
Moore, G. W., & Savidge, G. F. (2006) · Thrombosis Research · doi:10.1016/j.thromres.2005.09.010 · Cited by 46 · see citing works
Self-citationDEVREESE, K. M. J. (2010) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2010.03807.x · Cited by 42 · see citing works
Normal plasma mixing studies in the laboratory diagnosis of lupus anticoagulant
Thom, J., Ivey, L., & Eikelboom, J. (2003) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2003.0543h.x · Cited by 39 · see citing works
Current Controversies in Lupus Anticoagulant Detection
Moore, G. (2016) · Antibodies · doi:10.3390/antib5040022 · Cited by 36 · see citing works
Self-citationQuerrec, A. L., Arnout, J., Arnoux, D., Borg, J. Y., Caron, C., Darnige, L., Delahousse, B., Reber, G., & Siè, P. (2001) · Thrombosis and Haemostasis · doi:10.1055/s-0037-1616090 · Cited by 35 · see citing works
Depreter, B., & Devreese, K. M. J. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0978 · Cited by 34 · see citing works
Transient Lupus Anticoagulants in Children: Stepwise Disappearance of Diagnostic Features
Lechner, K., Speiser, W., Pabinger, I., & Male, C. (2000) · Thrombosis and Haemostasis · doi:10.1055/s-0037-1613777 · Cited by 33 · see citing works
Kumano, O., & Moore, G. W. (2018) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12069 · Cited by 26 · cited 5× · see citing works
“A variety of formulae have been proposed and employed for interpreting mixing tests and current guidelines recommend either the index of circulating anticoagulant (ICA) calculation, or a mixing test-specific cut-off (MTC) derived from the upper limit of a population distribution [11, 17].”
BackgroundInfluentialSelf-citationKumano, O., & Moore, G. W. (2019) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12245 · Cited by 24 · see citing works
Self-citation<i>In</i> ECAT <i>veritas?</i>
Reber, G., & Meijer, P. (2012) · Lupus · doi:10.1177/0961203312446389 · Cited by 21 · see citing works
Biguzzi, E., Asti, D., Palmucci, C., Tripodi, A., & Chantarangkul, V. (2013) · Thrombosis and Haemostasis · doi:10.1160/th12-11-0850 · Cited by 21 · see citing works
Moore, G. W., Maloney, J. C., de Jager, N., Dunsmore, C. L., Gorman, D. K., Polgrean, R. F., & Bertolaccini, M. L. (2017) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12006 · Cited by 20 · see citing works
Self-citationClyne, L. P., Yen, Y., Kriz, N. J., & Breitenstein, M. G. (1993) · PubMed · Cited by 16 · see citing works
Favaloro, E. J., Bonar, R., & Marsden, K. (2013) · Blood Coagulation & Fibrinolysis · doi:10.1097/mbc.0b013e328360c71b · Cited by 14 · see citing works
Chandrashekar, V. (2015) · Blood Coagulation & Fibrinolysis · doi:10.1097/mbc.0000000000000463 · Cited by 11 · see citing works
Moore, G. W. (2017) · Thrombosis Research · doi:10.1016/j.thromres.2017.09.017 · Cited by 8 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Devreese, K. M. J., de Groot, P. G., de Laat, B., Erkan, D., Favaloro, E. J., Mackie, I., Martinuzzo, M., Ortel, T. L., Pengo, V., Rand, J. H., et al. (2020) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15047 · Cited by 439 · see citing works
Moore, G. W., Jones, P. O., Platton, S., Hussain, N., White, D., Thomas, W., Rigano, J., Pouplard, C., Gray, E., & Devreese, K. M. J. (2021) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15438 · Cited by 47 · see citing works
Self-citationLupus Anticoagulant Detection under the Magnifying Glass
Molinari, A. C., Martini, T., Banov, L., Ierardi, A., Leotta, M., Strangio, A., & Santoro, R. C. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12206654 · Cited by 9 · see citing works
“The principal limitation regarding mixing is the unavoidable dilution effect on the antibody titer, which in some cases can make the LA undetectable, leading to false-negative results [104].”
BackgroundAnalytical dilemmas in lupus anticoagulant detection
Moore, G. W. (2023) · Exploration of Immunology · doi:10.37349/ei.2023.00104 · Cited by 5 · see citing works
Self-citationLupus Anticoagulant Testing: Taipan Snake Venom Time with Ecarin Time as Confirmatory Test
Moore, G. W. (2023) · Methods in Molecular Biology · doi:10.1007/978-1-0716-3175-1_16 · Cited by 5 · see citing works
Self-citationComparison of different algorithms for lupus anticoagulant detection: a single-center experience
Apipongrat, D., Lamool, R., Arnutti, P., Ruangpratheep, C., & Chantkran, W. (2024) · Research and Practice in Thrombosis and Haemostasis · doi:10.1016/j.rpth.2024.102333 · Cited by 4 · see citing works
- Authors
- Moore, G. W.
- Affiliations
- Addenbrooke's Hospital (GB)
- Corresponding author
- Moore, Gary W.
- ORCID
- Moore, Gary W.
- Venue
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Publisher
- Walter de Gruyter GmbH
- Published
- 2019-12-24
- Volume
- 58
- Issue
- 4
- Pages
- 492-495
- ISSN
- 1434-6621 (ISSN-L)
- 1437-4331
- OpenAlex
- W2997725712
- Cited by
- 19
- References
- 20
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 3 · 2023: 7 · 2022: 5 · 2021: 1 · 2020: 1
- FWCI
- 1.33
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Chemistry and Laboratory Medicine (CCLM) (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Rheumatology › Systemic Lupus Erythematosus Research
- Health Sciences › Medicine › Hematology › Blood Coagulation and Thrombosis Mechanisms
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Keywords
- lupus anticoagulant, mixing test, false-negative results, dilution effect, phospholipid dependence, integrated testing, anticoagulant screening, assay sensitivity and specificity
- MeSH terms
- Anticoagulants, False Negative Reactions, Humans, Partial Thromboplastin Time, Plasma, Prothrombin Time, Reproducibility of Results, Lupus Coagulation Inhibitor, Antiphospholipid Syndrome
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review