Comparison of different algorithms for lupus anticoagulant detection: a single-center experience
Open accessAbstract
Background: The laboratory tests for lupus anticoagulant (LA) detection comprise complex and multistep coagulation testing procedures. There is no established gold standard assay or direct comparison of algorithms as recommended by different guidelines. Objectives: This study aimed to evaluate and compare the LA detection performance of different laboratory algorithms suggested by the existing guidelines. Methods: The routine LA test data of 1801 plasma samples, including 188 LA-positive and 1613 LA-negative samples, were re-evaluated by applying the algorithms recommended by existing guidelines and were interpreted using various methods. Diagnostic performance indices for each LA detection algorithm were compared with those of the other algorithms. The efficacies of the different interpretation methods were analyzed to determine a suitable interpretation methodology for each assay. Results: The diagnostic performance for detecting LA varied by the algorithm and method of interpretation used. All laboratory algorithms displayed exceptional diagnostic performance with all diagnostic parameters of >90.0%. Nearly perfect agreement was observed in all algorithms when compared to the Clinical and Laboratory Standards Institute 2014 guideline interpreted by normalized screen-to-confirm ratio (NSCR) and mixing test-specific cutoff (MTC), as a reference assay (Cohen's kappa coefficient, >0.90 [range, 0.94-1.00]). A combination of the index of circulating anticoagulant and NSCR was optimal for interpreting the activated partial thromboplastin time-based test, whereas a combination of the MTC and NSCR was suitable for the diluted Russell's viper venom time-based test. Conclusion: All laboratory algorithms showed equivalent diagnostic performance. Establishing the best method of interpretation for each assay is recommended to improve LA detection performance.
AI snapshot
Written from the paper’s full text.
TL;DR
This single-center retrospective study re-evaluated routine lupus anticoagulant (LA) testing data from 1801 nonanticoagulated patients against algorithms from ISTH SSC 2009/2020, BCSH 2012, and CLSI 2014 guidelines. All algorithms showed high diagnostic performance (>90% for most parameters) and near-perfect agreement with the CLSI 2014 reference, but the optimal interpretation method differed by assay.
Key findings
- Among 1801 plasma samples (188 LA-positive, 1613 LA-negative by the reference algorithm), all laboratory algorithms had >90.0% PPA/sensitivity, NPA/specificity, PPV, NPV, and ORA/accuracy; agreement with the CLSI 2014 NSCR+MTC reference was nearly perfect (Cohen’s kappa >0.90; range 0.94–1.00).
- For aPTT-based assays, the best interpretation combination was ICA plus NSCR (ORA 57.0%; PPA 59.3%; NPA 52.9%), whereas for dRVVT-based assays it was MTC plus NSCR (ORA 91.3%; PPA 86.3%; NPA 95.7%).
- ISTH SSC 2020 algorithms improved PPA versus ISTH SSC 2009 (median PPA 100.0% vs 93.6%) and identified 4 additional persistently weakly LA-positive patients with a “negative mixing but positive confirmatory tests” pattern.
- Using 97.5th vs 99.0th percentile cutoffs increased median PPA (97.3% vs 93.6%) but lowered median PPV (95.6% vs 98.9%); dRVVT-based assays had AUCs >0.90, while aPTT-based assays averaged an AUC of 0.83.
Methods
- Retrospective single-center study at Phramongkutklao Hospital, Bangkok, Thailand (January 2020–December 2022); 1938 samples were screened, 137 excluded, leaving 1801 nonanticoagulant patients clinically investigated for APS, plus 120 healthy subjects for local reference intervals and cutoffs.
- Routine LA data were re-evaluated using algorithms recommended by ISTH SSC 2009, BCSH 2012, CLSI 2014, and ISTH SSC 2020, with CLSI 2014 interpreted by NSCR and MTC as the reference algorithm.
- Diagnostic performance was assessed with PPA, NPA, PPV, NPV, ORA, McNemar’s test, Cohen’s kappa, and AUC analysis; optimal interpretation methods combined mixing-test parameters (ICA or MTC) with confirmatory parameters (%Correct, SCR, or NSCR).
Limitations
- The authors state that only one aPTT and one dRVVT reagent were used, that the ellagic acid–based aPTT reagents may be less sensitive to LA, and that anticoagulated patients were excluded, so drug interference was not evaluated.
- The authors also state that costs and labor associated with each algorithm were not evaluated, and larger studies are needed to assess clinical outcomes.
- Evident from the text: this was a single-center retrospective study without a true gold standard assay, and the reference algorithm itself was a guideline-based comparator rather than an independent diagnostic standard.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
27 of 58 cited sources checked · 22 supported · 0 flagged · accuracy not yet measured
- Premisecitation
Using mixing tests interpreted with ICA demonstrates superior specificity as there are few false positives of inhibition by non-LA causes in cases of positive LA screening test results.
“In contrast, using mixing tests interpreted with ICA demonstrates superior specificity as there are few false positives of inhibition by non-LA causes in cases of positive LA screening test results [16].”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [16] Kumano, 2019, Ruling out lupus anticoagulants with mixing test-specific cutoff assessment and the index of circulating anticoagulant, Res Pract Thromb Haemost source
- Findingown data
When the ISTH SSC 2020 guideline was employed, a maximum of an additional 8 patients classified LA-negative by the reference algorithm were considered LA-positive with a "negative mixing but positive confirmatory tests" pattern.
“Interestingly, when the ISTH SSC 2020 guideline [20] was employed, a maximum of an additional 8 patients who were classified LA-negative by the reference algorithm were considered LA-positive with a “negative mixing but positive confirmatory tests” pattern.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [20] Devreese, 2020, Guidance from the Scientific and Standardization Committee for lupus anticoagulant/antiphospholipid antibodies of the International Society on Thrombosis and Haemostasis: update of the guidelines for lupus anticoagulant detection and interpretation, J Thromb Haemost source
- Premisecitation
Technical and practical issues, including test algorithms, interpretation parameters, cutoff values, and methods of interpretation, are problematic in many laboratories.
“Technical and practical issues, including test algorithms, interpretation parameters, cutoff values, and methods of interpretation, are problematic in many laboratories [6,8,9,21].”
Quote in the paperSubstantial claim4 of 4 sources checked- cites [6] Moore, 2014, Recent guidelines and recommendations for laboratory detection of lupus anticoagulants, Semin Thromb Hemost source
- cites [8] Moore, 2014, Commonalities and contrasts in recent guidelines for lupus anticoagulant detection, Int J Lab Hematol source
- cites [9] Moore, 2016, Current controversies in lupus anticoagulant detection, Antibodies (Basel) source
- cites [21] Cohen, 2019, J Thromb Haemost source
- Findingown data
Applying algorithms following the updated ISTH SSC 2020 guideline improved PPA compared with the previous 2009 guideline (median PPA, 93.6% vs 100.0%).
“It was revealed that the PPA was clearly improved by applying algorithms following the updated ISTH SSC 2020 guideline compared with the previous guideline in 2009 (median PPA, 93.6% vs 100.0%) (Table 1).”
- Also stated in: § Discussion · hedged
“Our results indicated that employing the algorithms based on the ISTH SSC 2020 recommendations resulted in significantly improved PPA when compared to their previous version.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [20] Devreese, 2020, Guidance from the Scientific and Standardization Committee for lupus anticoagulant/antiphospholipid antibodies of the International Society on Thrombosis and Haemostasis: update of the guidelines for lupus anticoagulant detection and interpretation, J Thromb Haemost source
- Also stated in: § Discussion · hedged
- Findingcitation
Patients were classified into LA-positive (n = 188) and LA-negative (n = 1613) based on LA test results interpreted by NSCR and MTC under the CLSI 2014 reference algorithm.
“The patients were classified into 2 groups based on their LA test results interpreted by using NSCR and MTC under the CLSI 2014 guidelines [13] (the reference algorithm): LA-positive (n = 188) and LA-negative (n = 1613).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] 2014
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingown data
Overall, the laboratory algorithms showed satisfactory diagnostic performance, as shown by their high PPA, NPA, PPV, NPV, and ORA.
“Although the LA-positive case number varied by the algorithm and method of interpretation used, overall, the laboratory algorithms showed satisfactory diagnostic performance, as shown by their high PPA, NPA, PPV, NPV, and ORA.”
Quote in the paperSubstantial claim - Findingown data
Almost all algorithms using aPTT-based assays showed moderate to good diagnostic performance, with an average AUC of 0.83.
“In contrast, almost all of the algorithms using the aPTT-based assays showed moderate to good diagnostic performance, with an average AUC of 0.83.”
Quote in the paperSubstantial claim - Findingown data
Using different cutoffs led to inconsistent LA-positive results with all algorithms (McNemar's test P value for each algorithm was less than .05).
“It was revealed that using different cutoffs led to inconsistent LA-positive results with all algorithms (McNemar’s test P value for each algorithm was less than .05).”
- Also stated in: § Discussion · plain
“The findings showed inconsistencies in the LA results for each approach.”
Quote in the paperSubstantial claim - Also stated in: § Discussion · plain
- Findingown data
The positive results of LA varied based on the algorithms and methods of interpretation used.
“It was observed that the positive results of LA varied based on algorithms and methods of interpretation used, as shown in Figure 2, Figure 3, Figure 4, Figure 5.”
Quote in the paperSubstantial claim - Findingown data
Among those, 4 patients were persistently positive with the same pattern when re-examined after 12 weeks and were classified as weakly LA-positive.
“Among these, 4 patients were persistently positive with the same pattern when the tests were re-examined after 12 weeks, and they were classified as weakly LA-positive.”
- Also stated in: § Discussion · boosted
“we identified 4 additional patients as weak LA-positive, who were classified as LA-negative according to our reference algorithm. In addition, the findings were validated by repeating the tests 12 weeks later, and the results demonstrated a persistently positive with “negative mixing but positive confirmatory tests” pattern.”
Quote in the paperSubstantial claim - Also stated in: § Discussion · boosted
- Findingown data
Excellent performance was observed in only ICA-containing algorithms applied to the aPTT-based assays.
“Notably, excellent performance was observed in only ICA-containing algorithms applied to the aPTT-based assays.”
- Also stated in: § Discussion · boosted
“in our receiver operating characteristic analysis, only ICA-containing algorithms demonstrated excellent performance for the aPTT-based assays.”
Quote in the paperSubstantial claim - Also stated in: § Discussion · boosted
- Findingown data
Of the 188 LA-positive cases, 72.3% were diagnosed with APS.
“Of these, 72.3% (136 of 188) were diagnosed with APS.”
Quote in the paperNot a substantial claim - Findingown data
For aPTT-based assays, using ICA and NSCR for mixing and LA confirmatory test interpretations showed a superior ability to identify both LA-positive and LA-negative cases, with an ORA of 57.0% (PPA = 59.3% and NPA = 52.9%).
“For aPTT-based assays, using ICA and NSCR for mixing and LA confirmatory test interpretations showed a superior ability to identify both LA-positive and LA-negative cases, with an ORA of 57.0% (PPA = 59.3% and NPA = 52.9%) (Figure 6C).”
Quote in the paperSubstantial claim - Findingown data
A combination of the MTC and NSCR was suitable for the diluted Russell's viper venom time-based test.
“A combination of the index of circulating anticoagulant and NSCR was optimal for interpreting the activated partial thromboplastin time–based test, whereas a combination of the MTC and NSCR was suitable for the diluted Russell’s viper venom time–based test.”
Quote in the paperSubstantial claim - Findingown data
Using the 97.5th percentile cutoff yielded a lower PPV than the 99.0th percentile cutoff (median PPV, 95.6% vs 98.9%).
“Considering each particular algorithm, using the 97.5th percentile cutoff yielded a higher PPA (median PPA, 97.3% vs 93.6%) but lower PPV than the 99.0th percentile cutoff (median PPV, 95.6% vs 98.9%).”
Quote in the paperSubstantial claim - Findingown data
All laboratory algorithms displayed exceptional diagnostic performance, with >90% of PPA, NPA, PPV, NPV, and ORA for LA detection.
“Even so, all laboratory algorithms displayed exceptional diagnostic performance, with >90% of PPA, NPA, PPV, NPV, and ORA for LA detection.”
- Also stated in: § Results · plain
“All laboratory algorithms displayed exceptional diagnostic performance with all diagnostic parameters of >90.0%.”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Findingown data
Considering all algorithms recommended by each guideline, only 172 samples were mutually identified as LA-positive.
“Considering all algorithms recommended by each guideline, we found that only 172 samples were mutually identified as LA-positive.”
Quote in the paperSubstantial claim - Findingown data
For dRVVT-based assays, combining the mixing test interpreted by MTC with LA confirmation interpreted by NSCR was the ideal interpretation method, with an ORA of 91.3% (PPA = 86.3% and NPA = 95.7%).
“In contrast, the ideal interpretation method for dRVVT-based assays was combining the mixing test interpreted by MTC with LA confirmation interpreted by NSCR, which resulted in an ORA of 91.3% (PPA = 86.3% and NPA = 95.7%) (Figure 6L).”
Quote in the paperSubstantial claim - Findingown data
All approaches displayed excellent agreement when compared with the reference algorithm.
“Furthermore, all approaches displayed excellent agreement when compared with the reference algorithm.”
Quote in the paperSubstantial claim - Findingown data
A nearly perfect agreement was observed between each algorithm and the CLSI 2014 guidelines interpreted by NSCR and MTC as the reference assays, with Cohen's kappa coefficients greater than 0.90 (95% CI, 0.94-1.00) for each.
“Additionally, a nearly perfect agreement was observed between each algorithm and the CLSI 2014 [13] guidelines interpreted by NSCR and MTC as the reference assays, with Cohen’s kappa coefficients greater than 0.90 (95% CI, 0.94-1.00) for each.”
- Also stated in: § Results · plain
“Nearly perfect agreement was observed in all algorithms when compared to the Clinical and Laboratory Standards Institute 2014 guideline interpreted by normalized screen-to-confirm ratio (NSCR) and mixing test–specific cutoff (MTC), as a reference assay (Cohen’s kappa coefficient, >0.90 [range, 0.94-1.00]).”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] 2014
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Also stated in: § Results · plain
- Findingown data
Based on the reference algorithm, 8 patients had negative mixing test results but positive LA confirmatory test results.
“Based on this reference algorithm, there were 8 patients who had negative mixing test results but positive LA confirmatory test results.”
Quote in the paperNot a substantial claim - Findingown data
All algorithms employing dRVVT-based assays demonstrated excellent performance in LA detection, with AUCs greater than 0.90 (P < .001).
“All algorithms employing dRVVT-based assays demonstrated excellent performance in LA detection, with AUCs greater than 0.90 (P < .001).”
Quote in the paperSubstantial claim - Findingown data
Using the 97.5th percentile cutoff yielded a higher PPA than the 99.0th percentile cutoff (median PPA, 97.3% vs 93.6%).
“Considering each particular algorithm, using the 97.5th percentile cutoff yielded a higher PPA (median PPA, 97.3% vs 93.6%) but lower PPV than the 99.0th percentile cutoff (median PPV, 95.6% vs 98.9%).”
Quote in the paperSubstantial claim - Findingown data
The diagnostic performance for detecting LA varied with the algorithm and method of interpretation used.
“It was observed that the diagnostic performance for detecting LA varied with the algorithm and method of interpretation used.”
- Also stated in: § Results · plain
“The diagnostic performance for detecting LA varied by the algorithm and method of interpretation used.”
Quote in the paperSubstantial claim - Also stated in: § Results · plain
- Findingown data
A combination of the index of circulating anticoagulant and NSCR was optimal for interpreting the activated partial thromboplastin time-based test.
“A combination of the index of circulating anticoagulant and NSCR was optimal for interpreting the activated partial thromboplastin time–based test, whereas a combination of the MTC and NSCR was suitable for the diluted Russell’s viper venom time–based test.”
Quote in the paperSubstantial claim - Study design and conductown data
Of 1938 patient samples, 137 were excluded because of anticoagulant treatment (123), FVIII inhibitors (3), or other coagulopathies (11).
“Among the 1938 patient samples, 137 samples, including 123 samples obtained from patients receiving anticoagulant treatment, 3 samples from patients with FVIII inhibitors, and 11 samples from patients with other coagulopathies, were excluded.”
Quote in the paperNot a substantial claim - Study design and conductown data
The data distributions for the 120 normal samples were confirmed as Gaussian using the Anderson‒Darling test (P > .05), with the exception of %Correct.
“The data distributions for the 120 normal samples were confirmed as Gaussian using the Anderson‒Darling test (P > .05), with the exception of %Correct.”
Quote in the paperSubstantial claim - Premisecitation
A recent study revealed that using ICA for interpreting the mixing tests of LA screening and confirming reagents greatly improved the capacity to distinguish between samples containing LA and direct FXa inhibitors with high specificity.
“A recent study revealed that using ICA for interpreting the mixing tests of LA screening and confirming reagents greatly improved the capacity to distinguish between samples containing LA and direct FXa inhibitors with high specificity [28].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [28] Kumano, 2021, Applying index of circulating anticoagulant to mixing tests with lupus anticoagulant screen and confirm reagents can distinguish with high specificity between lupus anticoagulants and direct factor Xa inhibitors, Int J Lab Hematol source
- Interpretationown data
Using ICA could improve the aPTT-based mixing test accuracy regardless of the phospholipid-dependent interpretation indices used.
“These findings suggest that using the ICA could improve the aPTT-based mixing test accuracy regardless of the phospholipid-dependent interpretation indices used.”
Quote in the paperSubstantial claim - Interpretationown data
All laboratory algorithms showed equivalent diagnostic performance.
“All laboratory algorithms showed equivalent diagnostic performance.”
Quote in the paperSubstantial claim - Premisecitation
The commercial reagents employed in this study are potentially less sensitive to LA.
“Although the sensitivity of ellagic acid-based reagents to LA is comparable with that of other silica-based reagents, the commercial reagents employed in this study are potentially less sensitive to LA [31].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [31] Kumano, 2012, APTT reagent with ellagic acid as activator shows adequate lupus anticoagulant sensitivity in comparison to silica-based reagent, J Thromb Haemost source
- Study design and conductown data
All 188 LA-positive cases were repeated and confirmed positive after at least 12 weeks to evidence antibody persistence.
“All 188 LA-positive cases were repeated and confirmed positive after at least 12 weeks to evidence the antibody persistence.”
Quote in the paperSubstantial claim
- Classified as probably not a substantial claim. 24
- May combine two claims. 18
Classified as probably not a substantial claim.
PremisecitationLA detection laboratory tests comprise complex and multistep coagulation testing procedures, including screening, confirmatory, and mixing tests.
“Laboratory tests for LA detection comprise complex and multistep coagulation testing procedures, including screening, confirmatory, and mixing tests [[11], [12], [13]], and heterogeneity in these is an important cause of interlaboratory variation.”
Quote in the paperSubstantial claim2 of 3 sources checkedMay combine two claims.
PremisereasoningThere is currently no established gold standard assay or reference plasma for LA detection.
“As a result, there are currently no established gold standard assays or reference plasmas for LA detection [[6], [7], [8], [9], [10]].”
Quote in the paperSubstantial claim4 of 5 sources checked- cites [6] Moore, 2014, Recent guidelines and recommendations for laboratory detection of lupus anticoagulants, Semin Thromb Hemost source
The Abstract says that an absence of gold standards complicates LA detection, and “Which Screening Tests to Use and How Many to Use?” explicitly notes the lack of a gold-standard LA detection method. The paper does not establish the more specific claim that no reference plasma exists; its “Ratios” section discusses variation among normal pooled plasmas instead. (LLM, conf=0.84)
Experimental — accuracy not yet measured - cites [7] Moore, 2017, Application of different lupus anticoagulant diagnostic algorithms to the same assay data leads to interpretive discrepancies in some samples, Res Pract Thromb Haemost source
- cites [8] Moore, 2014, Commonalities and contrasts in recent guidelines for lupus anticoagulant detection, Int J Lab Hematol source
- cites [9] Moore, 2016, Current controversies in lupus anticoagulant detection, Antibodies (Basel) source
- cites [10] Devreese, 2012, Standardization of antiphospholipid antibody assays. Where do we stand?, Lupus source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationMixing tests are performed to improve the specificity of LA detection and rule out other causes of prolonged clotting time, including coagulation factor deficiency, anticoagulant use, and coagulation factor inhibitors.
“Mixing tests are performed to improve the specificity of LA detection and rule out other causes of prolonged clotting time (CT), including coagulation factor deficiency, anticoagulant use, and coagulation factor inhibitors [15,16].”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [15] Moore, 2020, Mixing studies for lupus anticoagulant: mostly no, sometimes yes, Clin Chem Lab Med source
- cites [16] Kumano, 2019, Ruling out lupus anticoagulants with mixing test-specific cutoff assessment and the index of circulating anticoagulant, Res Pract Thromb Haemost source
The Introduction says mixing tests help distinguish inhibitors from factor deficiencies, and the Discussion reports that ICA improves specificity when investigating non-LA causes of prolonged clotting time. But the Discussion also says mixing tests are not specific for LA-induced inhibition: factor inhibitors and direct anticoagulants can affect their results, and confirmatory tests are needed. Thus the source does not support saying mixing tests rule out all the listed causes. (LLM, conf=0.88)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationIn tests with absence of a gold standard assay, PPA, NPA, and ORA are recommended in place of sensitivity, specificity, and %accuracy.
“However, in tests with absence of a gold standard assay, the terms positive percent agreement (PPA), negative percent agreement (NPA), and overall rate of agreement (ORA) were recommended in place of sensitivity, specificity, and %accuracy, respectively.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [22] Obermeier, 2016, Enabling precision medicine with digital case classification at the point-of-care, EBioMedicine source
Probably not a substantial claim: may not pass S2 (checkable).
PremisecitationBecause no single assay is sensitive to all LA antibodies, at least 2 coagulation tests based on different assay principles are recommended for LA detection.
“Because there is no single assay sensitive to all LA antibodies, at least 2 coagulation tests based on different assay principles are recommended for LA detection [[11], [12], [13]].”
Quote in the paperNot a substantial claim2 of 3 sources checkedClassified as probably not a substantial claim.
PremisecitationVarious interpretation indices, including %Correct, SCR, and NSCR, are suggested to determine the presence of LA in phospholipid-dependent assays.
“Various interpretation indices, including percentage of correction (%Correct), screen-to-confirm ratio (SCR), and normalized SCR (NSCR), are suggested to determine the presence of LA in these phospholipid-dependent assays [[11], [12], [13]].”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [11] Pengo, 2009, J Thromb Haemost source
- cites [12] Keeling, 2012, Guidelines on the investigation and management of antiphospholipid syndrome, Br J Haematol source
The ‘Lupus anticoagulant testing’ section recommends percentage correction or a normalized screen-to-confirm ratio for interpreting LA tests. It does not recommend an unnormalized screen-to-confirm ratio (SCR) as a separate index; it specifies that screen and confirm results should first be expressed as ratios against a normal plasma pool. No page number is available in our copy. (LLM, conf=0.87)
Experimental — accuracy not yet measured - cites [13] 2014
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationISTH SSC 2020 guideline recommends performing the mixing and confirmatory tests simultaneously if the screening test shows the presence of LA.
“In contrast, the ISTH SSC 2020 guideline recommends performing the mixing and confirmatory tests simultaneously if the screening test shows the presence of LA [20].”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [20] Devreese, 2020, Guidance from the Scientific and Standardization Committee for lupus anticoagulant/antiphospholipid antibodies of the International Society on Thrombosis and Haemostasis: update of the guidelines for lupus anticoagulant detection and interpretation, J Thromb Haemost source
Section 3.4, “Three step procedure” (page not provided), says that performing the mixing and confirmatory steps at the same time after a suggestive screening test can be helpful. It does not explicitly recommend simultaneous testing, so the claim states the guidance more strongly than the source does. (LLM, conf=0.86)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationMTC and ICA can be calculated and employed as interpretation indices for mixing tests.
“In addition, the mixing test–specific cutoff (MTC) and index of circulating anticoagulant (ICA) can be calculated and employed as interpretation indices for the mixing tests [[15], [16], [17], [18], [19]].”
Quote in the paperNot a substantial claim2 of 5 sources checked- cites [15] Moore, 2020, Mixing studies for lupus anticoagulant: mostly no, sometimes yes, Clin Chem Lab Med source
- cites [16] Kumano, 2019, Ruling out lupus anticoagulants with mixing test-specific cutoff assessment and the index of circulating anticoagulant, Res Pract Thromb Haemost source
- cites [17] Depreter, 2016, Differences in lupus anticoagulant final conclusion through clotting time or Rosner index for mixing test interpretation, Clin Chem Lab Med source
- cites [18] Kumano, 2018, Lupus anticoagulant mixing tests for multiple reagents are more sensitive if interpreted with a mixing test-specific cut-off than index of circulating anticoagulant, Res Pract Thromb Haemost source
- cites [19] Moore, 2016, Mixing test specific cut-off is more sensitive at detecting lupus anticoagulants than index of circulating anticoagulant, Thromb Res source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe LA-sensitive aPTT assay can be used as a secondary test for LA detection.
“The diluted Russell’s viper venom time (dRVVT) assay is the first recommended test for detecting LA due to its high sensitivity and specificity [14], while the LA-sensitive activated partial thromboplastin time (aPTT) assay can be used as a secondary test [[11], [12], [13], [14]].”
Quote in the paperNot a substantial claim2 of 4 sources checked- cites [11] Pengo, 2009, J Thromb Haemost source
- cites [12] Keeling, 2012, Guidelines on the investigation and management of antiphospholipid syndrome, Br J Haematol source
- cites [13] 2014
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [14] Favaloro, 2019, The Russell viper venom time (RVVT) test for investigation of lupus anticoagulant (LA), Am J Hematol source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSeveral studies have recommended the use of MTC over ICA as it offers greater sensitivity in detecting in vitro inhibition of LA.
“For the interpretation of the LA mixing test result, several studies have recommended the use of MTC over ICA as it offers greater sensitivity in detecting in vitro inhibition of LA [[17], [18], [19]].”
Quote in the paperNot a substantial claim1 of 3 sources checked- cites [17] Depreter, 2016, Differences in lupus anticoagulant final conclusion through clotting time or Rosner index for mixing test interpretation, Clin Chem Lab Med source
- cites [18] Kumano, 2018, Lupus anticoagulant mixing tests for multiple reagents are more sensitive if interpreted with a mixing test-specific cut-off than index of circulating anticoagulant, Res Pract Thromb Haemost source
- cites [19] Moore, 2016, Mixing test specific cut-off is more sensitive at detecting lupus anticoagulants than index of circulating anticoagulant, Thromb Res source
May combine two claims.
PremisecitationAlthough MTC provides greater sensitivity for the detection of LA, especially when weak LA is present, it has a greater tendency to produce false-positive results than ICA.
“Although MTC provides greater sensitivity for the detection of LA, especially when weak LA is present, it has a greater tendency to produce false-positive results than ICA [16,18,19].”
Quote in the paperSubstantial claim2 of 3 sources checked- cites [16] Kumano, 2019, Ruling out lupus anticoagulants with mixing test-specific cutoff assessment and the index of circulating anticoagulant, Res Pract Thromb Haemost source
- cites [18] Kumano, 2018, Lupus anticoagulant mixing tests for multiple reagents are more sensitive if interpreted with a mixing test-specific cut-off than index of circulating anticoagulant, Res Pract Thromb Haemost source
The Results and Discussion report higher MTC sensitivity than ICA for detecting LA, including cases that mixing-test dilution could otherwise cause to appear negative. But the reported comparison uses samples positive for LA in undiluted plasma; it does not establish that MTC produces more false-positive results than ICA. No page numbers are available in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measured - cites [19] Moore, 2016, Mixing test specific cut-off is more sensitive at detecting lupus anticoagulants than index of circulating anticoagulant, Thromb Res source
May combine two claims.
Findingown dataAmong the 188 LA-positive cases, 67.0% were LA-positive by dRVVT alone, 7.5% by aPTT alone, and 25.5% by both dRVVT and aPTT.
“Among the 188 LA-positive cases, 67.0% (126 of 188), 7.5% (14 of 188), and 25.5% (48 of 188) were considered LA-positive by dRVVT alone, aPTT alone, and both dRVVT and aPTT, respectively.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn the present study, only 20.7% of APS patients were triple-positive APS, whereas 79.3% were non-triple-positive APS.
“In the present study, we found that only 20.7% (57 of 275) of these patients were triple-positive APS, whereas the remaining 79.3% (218 of 275) were considered non–triple-positive APS.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe number of cases requiring complete 3-step testing was less when using the 99.0th percentile cutoff than when using the 97.5th percentile cutoff, though not statistically significant (P = .448; average number of cases requiring complete 3-step testing, 185 vs 199 cases).
“Although it did not reach statistical significance, the number of cases requiring complete 3-step testing was less when using the 99.0th percentile cutoff than when using the 97.5th percentile cutoff (P = .448; average number of cases requiring complete 3-step testing, 185 vs 199 cases) (Table 1).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataApproximately 56.4% of LA-positive patients experienced arterial or venous thrombosis, and 16.0% presented with obstetrical and gynecological comorbidities.
“Approximately 56.4% of LA-positive patients (106 of 188) experienced arterial or venous thrombosis, whereas 16.0% (30 of 188) of the patients presented with obstetrical and gynecological comorbidities.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataOnly 3 patients had positive LA mixing test results but negative confirmatory test results; they demonstrated clearly negative results when the tests were repeated after 12 weeks.
“Only 3 patients who had positive LA mixing test results but negative confirmatory test results were observed; however, they demonstrated clearly negative results when the tests were repeated after 12 weeks.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataAlthough different positive patterns of aPTT and dRVVT tests were observed in some samples, all cases demonstrated consistent positive results whether the 97.5th or 99.0th percentile cutoffs were used.
“Although the different positive pattern of aPTT and dRVVT tests was observed in some samples, all cases demonstrated consistent positive results whether the 97.5th or 99.0th percentile cutoffs were used.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataAmong LA-positive patients, 52.7% had positive aβ2-GPI and 31.4% had positive aCL results.
“As expected, among the LA-positive patients, 52.7% (99 of 188) and 31.4% (59 of 188) had positive aβ2-GPI and aCL results, respectively.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationCLSI 2014 guideline reprioritizes the test order to screen-confirm-mix to reduce false negatives by mixing tests.
“While the ISTH SSC 2009 and the BCSH 2012 guidelines suggest that the screening test should be followed by the mixing study and then the confirmatory test (screen-mix-confirm), the CLSI 2014 guideline reprioritizes the test order to screen-confirm-mix to reduce false negatives by mixing tests [[11], [12], [13],20].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [11], [12], [13],20
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Classified as probably not a substantial claim.
Study design and conductreasoningThe remaining 1801 plasma samples from nonanticoagulant patients clinically suspected of APS were analyzed.
“The remaining 1801 plasma samples obtained from nonanticoagulant patients clinically suspected of being investigated for APS were analyzed (Figure 1).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationISTH SSC 2009 and BCSH 2012 guidelines suggest that the screening test should be followed by the mixing study and then the confirmatory test (screen-mix-confirm).
“While the ISTH SSC 2009 and the BCSH 2012 guidelines suggest that the screening test should be followed by the mixing study and then the confirmatory test (screen-mix-confirm), the CLSI 2014 guideline reprioritizes the test order to screen-confirm-mix to reduce false negatives by mixing tests [[11], [12], [13],20].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [11], [12], [13],20
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe dRVVT assay is the first recommended test for detecting LA due to its high sensitivity and specificity.
“The diluted Russell’s viper venom time (dRVVT) assay is the first recommended test for detecting LA due to its high sensitivity and specificity [14], while the LA-sensitive activated partial thromboplastin time (aPTT) assay can be used as a secondary test [[11], [12], [13], [14]].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] Favaloro, 2019, The Russell viper venom time (RVVT) test for investigation of lupus anticoagulant (LA), Am J Hematol source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationISTH SSC 2009 and 2020 guidelines advise using 99.0th percentile cutoffs, whereas BCSH 2012 and CLSI 2014 guidelines advise using 97.5th percentile cutoffs.
“The ISTH SSC 2009 and 2020 guidelines advise using 99.0th percentile cutoffs, whereas the BCSH 2012 and the CLSI 2014 guidelines advise using 97.5th percentile cutoffs [[11], [12], [13],20].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [11], [12], [13],20
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe sensitivity of ellagic acid-based reagents to LA is comparable with that of other silica-based reagents.
“Although the sensitivity of ellagic acid-based reagents to LA is comparable with that of other silica-based reagents, the commercial reagents employed in this study are potentially less sensitive to LA [31].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [31] Kumano, 2012, APTT reagent with ellagic acid as activator shows adequate lupus anticoagulant sensitivity in comparison to silica-based reagent, J Thromb Haemost source
May combine two claims.
Interpretationown dataApplying the updated ISTH SSC 2020 guideline not only improves sensitivity but also increases the chance of detecting patients with weak LA presence by reducing the chance of a false negative result involving the mixing step.
“These findings suggest that applying this updated guideline not only improves sensitivity but also increases the chance of detecting patients with weak LA presence by reducing the chance of a false negative result involving the mixing step.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisecitationDifferences in reagent composition and concentration of phospholipids between various types of reagents are considered to be the cause of interreagent variation in LA sensitivity.
“Differences in reagent composition and concentration of phospholipids between various types of reagents are considered to be the cause of interreagent variation in LA sensitivity [30,31].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [30] Brandt, 1987, The sensitivity of different coagulation reagents to the presence of lupus anticoagulants, Arch Pathol Lab Med
Not checked in the free run.
- cites [31] Kumano, 2012, APTT reagent with ellagic acid as activator shows adequate lupus anticoagulant sensitivity in comparison to silica-based reagent, J Thromb Haemost source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationA previous study recommended the use of MTC for the dRVVT-based mixing test and ICA for the aPTT-based mixing test to achieve the highest sensitivity.
“a previous study recommended the use of MTC for the dRVVT-based mixing test and ICA for the aPTT-based mixing test to achieve the highest sensitivity [27].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [27] Martinuzzo, 2012, New guidelines for lupus anticoagulant: sensitivity and specificity of cut-off values calculated with plasmas from healthy controls in mixing and confirmatory tests, Int J Lab Hematol source
Classified as probably not a substantial claim.
Study design and conductown dataThe study did not evaluate the costs and labor associated with each algorithm and guideline.
“Nevertheless, we did not evaluate the costs and labor associated with each algorithm and guideline in our study, so further research comparing the cost-effectiveness of these guidelines is needed to draw a firm conclusion.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
PremisecitationThe NSCR is recommended by the existing guidelines as a parameter for interpreting phospholipid-dependent assays because it can reduce both interlaboratory and interreagent variations.
“The NSCR is recommended by the existing guidelines as a parameter for interpreting phospholipid-dependent assays [[11], [12], [13],20] because it can reduce both interlaboratory and interreagent variations [25].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [11], [12], [13],20
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [25] Ling, 2022, Interpretation of clot-based lupus anticoagulant assays—normalizing clotting time against different denominators, Int J Lab Hematol source
Classified as probably not a substantial claim.
Study design and conductnone givenThe analyzed data were only applied to the results of clinical samples with 1 aPTT and 1 dRVVT reagent.
“First, the analyzed data were only applied to the results of clinical samples with 1 aPTT and 1 dRVVT reagent.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe aPTT reagent pair used in this study (Actin FSL for screening and Actin FS for confirmation) is one of only a few paired aPTT reagents offered by commercial manufacturers, with most manufacturers of silica-based aPTT assays offering an LA-sensitive reagent, but not providing an LA insensitive silica-based paired reagent.
“On the other hand, the aPTT reagent pair used in this study (Actin FSL for screening and Actin FS for confirmation) is one of only a few paired aPTT reagents offered by commercial manufacturers, with most manufacturers of silica-based aPTT assays offering an LA-sensitive reagent, but not providing an LA insensitive silica-based paired reagent [32].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [32] Favaloro EJ, Dean E, Arunachalam S. Variable performance of lupus anticoagulant testing: the Australasian/Asia-Pacific experience. Semin Thromb Hemost 2023; in press. https://doi.org/10.1055/s-0043-1776406 source
May combine two claims.
Study design and conductreasoningThe samples of patients who received anticoagulants were excluded, so the interfering effect of drugs on the diagnostic performance of LA testing was not evaluated.
“Second, as the samples of patients who received anticoagulants were excluded, the interfering effect of the drugs on the diagnostic performance of LA testing was not evaluated in our study.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataMTC was optimal for interpreting the dRVVT-based test results, while using ICA for interpreting the aPTT-based mixing tests yielded the most precise results as it offered better positive agreement rates with the highest rate of overall agreements when compared with the reference assay.
“Similarly, our results demonstrated that MTC was optimal for interpreting the dRVVT-based test results, while using ICA for interpreting the aPTT-based mixing tests yielded the most precise results as it offered better positive agreement rates with the highest rate of overall agreements when compared with the reference assay.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisecitationThe ICA was considered the most robust approach to express the LA mixing test results.
“The ICA was considered the most robust approach to express the LA mixing test results [26]”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [26] Tripodi, 2007, Laboratory testing for lupus anticoagulants: a review of issues affecting results, Clin Chem source
May combine two claims.
PremisecitationCompared with dRVVT, aPTT is a more complex assay involving multiple enzymatic steps and may be influenced by physiologic and pathologic factors unrelated to the inhibition of LA.
“Compared with dRVVT, aPTT is a more complex assay involving multiple enzymatic steps and may be influenced by physiologic and pathologic factors unrelated to the inhibition of LA [14,29].”
Quote in the paperSubstantial claim0 of 2 sources checkedClassified as probably not a substantial claim.
Study design and conductnone given120 normal plasma samples were used to establish and validate the specific local RIs and cutoff values.
“Additionally, 120 normal plasma samples were used to establish and validate the specific local RIs and cutoff values.”
Quote in the paperSubstantial claim
The paper relies on this without support.
No study has compared the testing algorithms recommended by the ISTH SSC 2009, BCSH 2012, CLSI 2014, and ISTH SSC 2020 guidelines in terms of LA diagnostic performance.
“Furthermore, no study has compared the testing algorithms recommended by each of the abovementioned guidelines in terms of their LA diagnostic performance.”
The paper relies on this without support.
There is no gold standard assay available for LA detection.
“Although the current guidelines have helped standardize the LA testing process and interpretation, there is no gold standard assay available for LA detection.”
The paper relies on this without support.
There is a lack of empirical data to support any strong evidence-based recommendations.
“Furthermore, there is a lack of empirical data to support any strong evidence-based recommendations.”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
The stated ORA formula does not produce the reported ORA
MinorThe stated formula gives (95.7 + 99.9) / 1801 × 100 = 10.9%, not the 99.4% ORA reported for the first algorithm in Table 1. The table appears to use overall agreement based on case counts instead.
“ORA = ([PPA + NPA] / total cases) × 100”
“| Screen→mix (ICA)→confirm (%Correct) | 189 | 182 | 180/2 | 95.7 | 99.9 | 98.9 | 99.5 | 99.4 | 0.97 (0.95-0.99) |”
Identical classifications have different kappa coefficients
MinorThe two CLSI 2014 rows each report 188 positive cases, comprising the same 184 reference-positive and four reference-negative cases. Against the same reference results, their Cohen's kappa coefficients must be identical, but the table reports 0.98 and 0.95.
“| Screen→confirm (%Correct)→mix (MTC) | 235 | 188 | 184/4 | 97.8 | 99.8 | 97.9 | 99.8 | 99.6 | 0.98 (0.96-0.99) |”
“| Screen→confirm (SCR)→mix (MTC) | 362 | 188 | 184/4 | 97.8 | 99.8 | 97.9 | 99.8 | 99.6 | 0.95 (0.93-0.98) |”
Specificity increases despite an additional false positive
MinorBoth ISTH SSC 2009 rows have 172 reference-positive cases among their positive results. With the same 1613 reference-negative samples, the row with three reference-negative positives cannot have higher specificity than the row with two; the reported values are reversed.
“| Screen→mix (MTC)→confirm (SCR) | 181 | 174 | 172/2 | 91.5 | 99.8 | 98.9 | 99.0 | 99.0 | 0.95 (0.92-0.97) |”
“| Screen→mix (MTC)→confirm (NSCR) | 181 | 175 | 172/3 | 91.5 | 99.9 | 98.3 | 99.0 | 98.9 | 0.94 (0.92-0.97) |”
Cites
MIYAKIS, S., LOCKSHIN, M. D., ATSUMI, T., BRANCH, D. W., BREY, R. L., CERVERA, R., DERKSEN, R. H. W. M., DE GROOT, P. G., KOIKE, T., MERONI, P. L., et al. (2006) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2006.01753.x · Cited by 7,203 · see citing works
Update of the guidelines for lupus anticoagulant detection
PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M., & DE GROOT, P. G. (2009) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03555.x · Cited by 1,197 · see citing works
Guidelines on the investigation and management of antiphospholipid syndrome
Keeling, D., Mackie, I., Moore, G. W., Greer, I. A., Greaves, M., & British Committee for Standards in Haematology (2012) · British Journal of Haematology · doi:10.1111/j.1365-2141.2012.09037.x · Cited by 740 · see citing works
Clinical course of high‐risk patients diagnosed with antiphospholipid syndrome
PENGO, V., RUFFATTI, A., LEGNANI, C., GRESELE, P., BARCELLONA, D., ERBA, N., TESTA, S., MARONGIU, F., BISON, E., DENAS, G., et al. (2010) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03674.x · Cited by 633 · cited 1× · see citing works
“In addition, the presence of LA is a well-known risk factor for clinical manifestations of APS patients[3 – 5].”
BackgroundDevreese, K. M. J., de Groot, P. G., de Laat, B., Erkan, D., Favaloro, E. J., Mackie, I., Martinuzzo, M., Ortel, T. L., Pengo, V., Rand, J. H., et al. (2020) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15047 · Cited by 439 · see citing works
Lockshin, M. D., Kim, M., Laskin, C. A., Guerra, M., Branch, D. W., Merrill, J., Petri, M., Porter, T. F., Sammaritano, L., Stephenson, M. D., et al. (2012) · Arthritis & Rheumatism · doi:10.1002/art.34402 · Cited by 388 · see citing works
Laboratory criteria for antiphospholipid syndrome: communication from the SSC of the ISTH
Devreese, K. M. J., Ortel, T. L., Pengo, V., & de Laat, B. (2018) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.13976 · Cited by 269 · see citing works
Coagulation Testing in the Core Laboratory
Winter, W. E., Flax, S. D., & Harris, N. S. (2017) · Laboratory Medicine · doi:10.1093/labmed/lmx050 · Cited by 238 · see citing works
Recent Guidelines and Recommendations for Laboratory Detection of Lupus Anticoagulants
Moore, G. (2014) · Seminars in Thrombosis and Hemostasis · doi:10.1055/s-0033-1364185 · Cited by 148 · cited 6× · see citing works
“Under the ISTH SSC 2009 recommendations [11], using the 99.0th percentile as the cutoff can increase speci fi city while decreasing sensitivity [6,8,20].”
BackgroundMethodInfluentialLaboratory Testing for Lupus Anticoagulants: A Review of Issues Affecting Results
Tripodi, A. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2007.089524 · Cited by 81 · see citing works
Standardization of antiphospholipid antibody assays. Where do we stand?
Devreese, K. (2012) · Lupus · doi:10.1177/0961203312439335 · Cited by 76 · see citing works
The sensitivity of different coagulation reagents to the presence of lupus anticoagulants
(1986) · Thrombosis Research · doi:10.1016/0049-3848(86)91499-4 · Cited by 74 · see citing works
Gebhart, J., Posch, F., Koder, S., Perkmann, T., Quehenberger, P., Zoghlami, C., Ay, C., & Pabinger, I. (2015) · Blood · doi:10.1182/blood-2014-11-611129 · Cited by 72 · cited 1× · see citing works
“In addition, the presence of LA is a well-known risk factor for clinical manifestations of APS patients[3 – 5].”
BackgroundKUMANO, O., IEKO, M., NAITO, S., YOSHIDA, M., & TAKAHASHI, N. (2012) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2012.04906.x · Cited by 59 · see citing works
Cohen, H., Mackie, I. J., & Devreese, K. M. J. (2019) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.14560 · Cited by 48 · see citing works
Moore, G. W., Culhane, A. P., Daw, C. R., Noronha, C. P., & Kumano, O. (2016) · Thrombosis Research · doi:10.1016/j.thromres.2016.01.019 · Cited by 42 · see citing works
Commonalities and contrasts in recent guidelines for lupus anticoagulant detection
Moore, G. W. (2014) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12227 · Cited by 41 · see citing works
Current Controversies in Lupus Anticoagulant Detection
Moore, G. (2016) · Antibodies · doi:10.3390/antib5040022 · Cited by 36 · see citing works
Depreter, B., & Devreese, K. M. J. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0978 · Cited by 34 · see citing works
Uludağ, G., Onghanseng, N., Tran, A. N. T., Hassan, M., Halim, M. S., Sepah, Y. J., Diana, V., & Nguyen, Q. D. (2021) · Journal of Ophthalmic Inflammation and Infection · doi:10.1186/s12348-021-00240-8 · Cited by 32 · see citing works
Enabling Precision Medicine With Digital Case Classification at the Point-of-Care
Obermeier, P., Muehlhans, S., Hoppe, C., Karsch, K., Tief, F., Seeber, L., Chen, X., Conrad, T., Boettcher, S., Diedrich, S., et al. (2016) · EBioMedicine · doi:10.1016/j.ebiom.2016.01.008 · Cited by 29 · see citing works
Laurent, C., Ricard, L., Nguyen, Y., Boffa, J. J., Rondeau, É., Gerotziafas, G., Elalamy, I., Deriaz, S., MOREUIL, C. D., Planche, V., et al. (2023) · RMD Open · doi:10.1136/rmdopen-2022-002534 · Cited by 28 · see citing works
Kumano, O., & Moore, G. W. (2018) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12069 · Cited by 26 · see citing works
The Russell viper venom time (RVVT) test for investigation of lupus anticoagulant (LA)
Favaloro, E. J. (2019) · American Journal of Hematology · doi:10.1002/ajh.25606 · Cited by 25 · see citing works
Kumano, O., & Moore, G. W. (2019) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12245 · Cited by 24 · see citing works
Moore, G. W., Maloney, J. C., de Jager, N., Dunsmore, C. L., Gorman, D. K., Polgrean, R. F., & Bertolaccini, M. L. (2017) · Research and Practice in Thrombosis and Haemostasis · doi:10.1002/rth2.12006 · Cited by 20 · see citing works
Mixing studies for lupus anticoagulant: mostly no, sometimes yes
Moore, G. W. (2020) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2019-1248 · Cited by 19 · see citing works
MARTINUZZO, M. E., CERRATO, G. S., VARELA, M. L. I., ADAMCZUK, Y. P., & FORASTIERO, R. R. (2012) · International Journal of Laboratory Hematology · doi:10.1111/j.1751-553x.2011.01376.x · Cited by 16 · see citing works
Variable Performance of Lupus Anticoagulant Testing: The Australasian/Asia-Pacific Experience
Favaloro, E. J., Dean, E., & Arunachalam, S. (2024) · Seminars in Thrombosis and Hemostasis · doi:10.1055/s-0043-1776406 · Cited by 11 · see citing works
Ling, L., Liu, C., Huang, X., Liao, J., Jia, J., & Zhou, J. (2022) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.13830 · Cited by 5 · see citing works
Kumano, O., Amiral, J., Dunois, C., Peyrafitte, M., & Moore, G. W. (2021) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.13571 · Cited by 2 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Barion, B. G., Stefanello, B., de Paula, M. L. S. A., Saraiva, T. S., Villaça, P. R., Rocha, V., Orsi, F. A., & da Rocha, T. R. F. (2025) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.14549 · Cited by 2 · see citing works
Ibrahim‐Kosta, M., Alessi, M., & Frère, C. (2025) · Encyclopédie médico-chirurgicale. Hématologie · doi:10.1016/s1155-1984(24)47828-x
Novelli, C., Gatti, A., Ierna, F., Riccardi, E., & Cuppari, I. (2026) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.70153
Wongkrajang, P., Pientong, T., Hanyongyuth, R., et al. (2026) · PLoS ONE · doi:10.1371/journal.pone.0352430
- Authors
- Apipongrat, D., Lamool, R., Arnutti, P., Ruangpratheep, C., & Chantkran, W.
- Affiliations
- Phramongkutklao Hospital (TH)
- Phramongkutklao College of Medicine
- Corresponding author
- Chantkran, Wittawat
- Venue
- Research and Practice in Thrombosis and Haemostasis
- Publisher
- Elsevier BV
- Published
- 2024-01
- Volume
- 8
- Issue
- 1
- Pages
- 102333
- ISSN
- 2475-0379 (ISSN-L)
- PMC
- PMC10885790
- OpenAlex
- W4391344566
- Semantic Scholar
- 55e52d27bf50b1791b68c473413a5b0602ffd544
- Cited by
- 4
- References
- 31
- Citations by year
- 2026: 2 · 2025: 2
- FWCI
- 1.09
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Research and Practice in Thrombosis and Haemostasis (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion, cc-by-nc-nd)
- Topics
- Health Sciences › Medicine › Rheumatology › Systemic Lupus Erythematosus Research
- Health Sciences › Medicine › Hematology › Blood Coagulation and Thrombosis Mechanisms
- Health Sciences › Medicine › Surgery › Heparin-Induced Thrombocytopenia and Thrombosis
- Keywords
- lupus anticoagulant, diagnostic performance, activated partial thromboplastin time, Russell's viper venom time, assay interpretation, antiphospholipid syndrome
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-19.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-19.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review