Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management
Open accessAbstract
source of the sample, and specimen matrix (i.e., plasma versus whole blood).This study will review assay principles, patient variables, and systematic variables and then encourage clinicians to carefully consider how standard recommendations regarding glycemic targets, particularly in the ICU, should be implemented in their individual health care facilities.ASSAY PRINCIPLES -In this review, we will signify reference laboratory methods with the term "central laboratory method.""POC" refers to hand-held devices or portable ward-based analyzers.We recognize that some of these devices are also used in the ambulatory setting."Plasma correlated" refers to glucose concentrations measured in samples of whole blood but are converted to values that would be expected of plasma measurements.•
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Intensive insulin therapy in critically ill patients.
Van den Berghe, G., Wouters, P., Weekers, F., Verwaest, C., Bruyninckx, F., Schetz, M., Vlasselaers, D., Ferdinande, P., Lauwers, P., & Bouillon, R. (2001) · New England Journal of Medicine · doi:10.1056/nejmoa011300 · Cited by 9,977
Intensive insulin therapy in the medical ICU
Van den Berghe, G., Wilmer, A., Hermans, G., Meersseman, W., Wouters, P. J., Milants, I., Van Wijngaerden, E., Bobbaers, H., & Bouillon, R. (2006) · New England Journal of Medicine · doi:10.1056/nejmoa052521 · Cited by 3,639
Standards of Medical Care in Diabetes–2006
American Diabetes Association (2006) · Diabetes Care · doi:10.2337/diacare.29.s1.06.s4 · Cited by 1,885
Malmberg, K., Rydén, L., Efendic, S., Herlitz, J., Nicol, P., Waldenstrom, A., Wedel, H., & Welin, L. (1995) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(95)00126-k · Cited by 1,560
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Sacks, D. B., Bruns, D. E., Goldstein, D. E., Maclaren, N. K., McDonald, J. M., & Parrott, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.436 · Cited by 1,130
American College of Endocrinology Position Statement on Inpatient Diabetes and Metabolic Control
Garber, A. J., Moghissi, E. S., Bransome, E. D., Clark, N. G., Clement, S., Cobin, R. H., Furnary, A. P., Hirsch, I. B., Levy, P., Roberts, R., et al. (2004) · Endocrine Practice · doi:10.4158/ep.10.s2.4 · Cited by 409
Reliability of point-of-care testing for glucose measurement in critically ill adults*
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000189939.10881.60 · Cited by 395
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
American College of Endocrinology Position Statement on Inpatient Diabetes and Metabolic Control
Garber, A. J., Moghissi, E. S., Bransome, E. D., Clark, N. G., Clement, S., Cobin, R. H., Furnary, A. P., Hirsch, I. B., Levy, P., Robert, R., et al. (2004) · Endocrine Practice · doi:10.4158/ep.10.1.77 · Cited by 284
The Original Clarke Error Grid Analysis (EGA)
Clarke, W. L. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.776 · Cited by 280
Boyd, J. C., & Bruns, D. E. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.2.209 · Cited by 227
Approved IFCC recommendation on reporting results for blood glucose (abbreviated)
D'Orazio, P. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.051979 · Cited by 208
Tang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · American Journal of Clinical Pathology · doi:10.1309/qaw1-x5xw-bvrq-5lkq · Cited by 207
Fingerstick Glucose Determination in Shock
Atkin, S. H., Dasmahapatra, A., Jaker, M. A., Chorost, M. I., & Reddy, S. (1991) · Annals of Internal Medicine · doi:10.7326/0003-4819-114-12-1020 · Cited by 189
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Meijering, S., Corstjens, A. M., Tulleken, J. E., Meertens, J. H., Zijlstra, J. G., & Ligtenberg, J. J. (2006) · Critical Care · doi:10.1186/cc3981 · Cited by 172
Dermal Neurovascular Dysfunction in Type 2 Diabetes
Vinik, A. I., Erbas, T. A., Park, T. S., Stansberry, K. B., Scanelli, J. A., & Pittenger, G. L. (2001) · Diabetes Care · doi:10.2337/diacare.24.8.1468 · Cited by 159
Kuwa, K., Nakayama, T., Hoshino, T., & Tominaga, M. (2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00426-0 · Cited by 142
Agreement Between Bedside Blood and Plasma Glucose Measurement in the ICU Setting
Finkielman, J. D., Oyen, L. J., & Afessa, B. (2005) · CHEST Journal · doi:10.1378/chest.127.5.1749 · Cited by 112
Capillary versus venous bedside blood glucose estimations
Boyd, R. (2005) · Emergency Medicine Journal · doi:10.1136/emj.2003.011619 · Cited by 112
Bergenstal, R. M., Pearson, J., Cembrowski, G. S., Bina, D. M., Davidson, J. L., & List, S. (2000) · The Science of Diabetes Self-Management and Care · doi:10.1177/014572170002600610 · Cited by 107
Comparability of venous and capillary glucose measurements in blood
Colagiuri, S., Sandbaek, A., Carstensen, B., Christensen, J. O., Glümer, C., Lauritzen, T., & Borch‐Johnsen, K. (2003) · Diabetic Medicine · doi:10.1046/j.1464-5491.2003.01048.x · Cited by 101
Tang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0577-eopogm · Cited by 92
Berth, M., & Delanghe, J. (2004) · Acta Clinica Belgica · doi:10.1179/acb.2004.039 · Cited by 91
Sell-Monitoring of Blood Glucose
(1996) · Diabetes Care · doi:10.2337/diacare.19.1.s62 · Cited by 89
Accuracy of fingerstick glucose values in shock patients
Sylvain, H., Pokorny, M., English, S. M., BENSON, N., Whitley, T., Ferenczy, C. J., & Harrison, J. G. (1995) · American Journal of Critical Care · doi:10.4037/ajcc1995.4.1.44 · Cited by 89
Kulkarni, A. P., Saxena, M. K., Price, G., O’Leary, M. J., Jacques, T. C., & Myburgh, J. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2500-5 · Cited by 88
Newer portable glucose meters—analytical improvement compared with previous generation devices?
Weitgasser, R., Gappmayer, B., & Pichler, M. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.10.1821 · Cited by 85
A Previously Undescribed Side Effect of Icodextrin: Overestimation of Glycemi a by Glucose Analyzer
Wens, R., Taminne, M., Devriendt, J., Collart, F., Broeders, N., Mestrez, F., Germanos, H., & Dratwa, M. (1998) · Peritoneal Dialysis International · doi:10.1177/089686089801800608 · Cited by 82
Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction
Rao, L. V., Jakubiak, F., Sidwell, J. S., Winkelman, J. W., & Snyder, M. (2005) · Clinica Chimica Acta · doi:10.1016/j.cccn.2005.01.027 · Cited by 81
Positive Interference of Icodextrin Metabolites in Some Enzymatic Glucose Methods
Janssen, W., Harff, G. A., Caers, M., & SCHELLEKENS, A. P. M. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.11.2379 · Cited by 71
Öberg, D., & Östenson, C. (2005) · Diabetes Care · doi:10.2337/diacare.28.5.1261 · Cited by 71
Kristensen, G. B. B., Christensen, N. G., Thue, G., & Sandberg, S. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.049080 · Cited by 70
Maser, R. E., BUTLER, M. A., & DeCherney, G. S. (1994) · Critical Care Medicine · doi:10.1097/00003246-199404000-00014 · Cited by 69
Heinemann, L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322250578 · Cited by 68
HemoCue: Evaluation of a Portable Photometric System for Determining Glucose in Whole Blood
Ashworth, L., Gibb, I., & Alberti, K. G. M. M. (1992) · Clinical Chemistry · doi:10.1093/clinchem/38.8.1479 · Cited by 67
Differences between capillary and venous blood glucose during oral glucose tolerance tests
Larsson-Cohn, U. (1976) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365517609081941 · Cited by 65
Pilot study of the accuracy of bedside glucometry in the intensive care unit
Ray, J. G., Hamielec, C. M., & Mastracci, T. M. (2001) · Critical Care Medicine · doi:10.1097/00003246-200111000-00025 · Cited by 64
Oxygen Effects on Glucose Measurements with a Reference Analyzer and Three Handheld Meters
Tang, Z., Louie, R. F., Payes, M., Chang, K.-C. J., & Kost, G. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194215 · Cited by 63
Oyibo, S. O., Pritchard, G., Mclay, L., James, E., Laing, I., Gokal, R., & Boulton, A. J. M. (2002) · Diabetic Medicine · doi:10.1046/j.1464-5491.2002.00753.x · Cited by 56
Lewandrowski, K. B., Cheek, R., Nathan, D. M., Godine, J. E., Hurxthal, K., Eschenbach, K., & Laposata, M. (1992) · The American Journal of Medicine · doi:10.1016/0002-9343(92)90172-8 · Cited by 54
Point-of-Care Testing in Diabetes Mellitus
Price, C. P. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.185 · Cited by 54
Kost, G. J., Nguyen, T. H., & Tang, Z. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1128-wbgal · Cited by 45
The difference between the glucose concentrations in plasma and whole blood
Holtkamp, H. C., Verhoef, N. J., & Leijnse, B. (1975) · Clinica Chimica Acta · doi:10.1016/0009-8981(75)90216-8 · Cited by 45
Clinical Evaluation of Alternative-Site Glucose Measurements in Patients After Major Cardiac Surgery
Ellmerer, M., Haluzı́k, M., Bláha, J., Kremen, J., Svacina, S., Toller, W. G., Mader, J. K., Schaupp, L. A., Plank, J., & Pieber, T. R. (2006) · Diabetes Care · doi:10.2337/dc05-2377 · Cited by 44
Evaluation of Roche Accu-Chek Go and Medisense Optium blood glucose meters
Hawkins, R. C. (2004) · Clinica Chimica Acta · doi:10.1016/j.cccn.2004.10.013 · Cited by 43
Soller, B. R., Idwasi, P. O., Balaguer, J., Levin, S., Simsir, S. A., Salm, T. J. V., Collette, H., & Heard, S. O. (2003) · Critical Care Medicine · doi:10.1097/01.ccm.0000086999.21673.6a · Cited by 42
Variations in Sample pH and pO2 Affect ExacTech Meter Glucose Measurements
Kilpatrick, E. S., Rumley, A. G., & Smith, E. A. (1994) · Diabetic Medicine · doi:10.1111/j.1464-5491.1994.tb00315.x · Cited by 41
An Assessment of the Effect of Haematocrit on the Hemocue Blood Glucose Analyser
Wiener, K. (1993) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329303000116 · Cited by 39
Kurahashi, K., Maruta, H., Usuda, Y., & Ohtsuka, M. (1997) · Critical Care Medicine · doi:10.1097/00003246-199702000-00006 · Cited by 37
Haupt, A., Berg, B., Paschen, P., Dreyer, M., Häring, H.-U., Smedegaard, J., & Matthaei, S. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.597 · Cited by 36
Point-of-Care Testing for Diabetes
Blake, D. R., & Nathan, D. M. (2004) · Critical Care Nursing Quarterly · doi:10.1097/00002727-200404000-00007 · Cited by 29
Effect of packed cell volume on blood glucose estimations.
Dacombe, C. M., Dalton, R. G., Goldie, D. J., & Osborne, J. P. (1981) · Archives of Disease in Childhood · doi:10.1136/adc.56.10.789 · Cited by 28
Intra‐operative blood glucose measurements
Smith, E. A., & Kilpatrick, E. S. (1994) · Anaesthesia · doi:10.1111/j.1365-2044.1994.tb03369.x · Cited by 26
Kaufmann-Raab, I., Jonen, H. G., Jähnchen, E., Kahl, G. F., & Groth, U. (1976) · Clinical Chemistry · doi:10.1093/clinchem/22.10.1729 · Cited by 26
Monitoring Blood Glucose Meters
Rainey, P. M., & Jatlow, P. I. (1995) · American Journal of Clinical Pathology · doi:10.1093/ajcp/103.2.125 · Cited by 26
Cartier, L.-J., Leclerc, P., Pouliot, M., Nadeau, L., Turcotte, G., & de Laclos, B. F. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.4.893 · Cited by 24
Kavsak, P. A., Zielinski, N., Li, D., McNamara, P. J., & Adeli, K. (2004) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2004.06.003 · Cited by 22
Farrer, M., Albers, C. J., Neil, H. A. W., Adams, P. C., Laker, M. F., & Alberti, K. G. M. M. (1995) · Diabetic Medicine · doi:10.1111/j.1464-5491.1995.tb00486.x · Cited by 20
Evaluation of a glucose meter against analytical quality specifications for hospital use
Dhatt, G. S., Agarwal, M., & Bishawi, B. (2004) · Clinica Chimica Acta · doi:10.1016/j.cccn.2004.01.021 · Cited by 18
Nobels, F. R. E., Beckers, F., Bailleul, E., Schrijver, P. D., Sierens, L., & Crombrugge, P. V. (2004) · Diabetic Medicine · doi:10.1111/j.1464-5491.2004.01333.x · Cited by 17
Yuoh, C., Elghetany, M. T., Petersen, J. R., Mohammad, A. A., & Okorodudu, A. O. (2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00442-9 · Cited by 17
Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus
Stahl, M., & Brandslund, I. (2003) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510310002590 · Cited by 16
Kempe, K. C., Czeschin, L. I., Yates, K. H., et al. (1997) · Clinical Chemistry · doi:10.1093/clinchem/43.9.1803 · Cited by 13
Cross, M. H., & Brown, D. G. (1996) · The Journal of Clinical Monitoring · doi:10.1007/bf02025307 · Cited by 13
Sánchez‐Margalet, V., Oliva, M. S. R., Sánchez-Pozo, M. C., Fernández-Gallardo, M. F., & Goberna, R. (2005) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2005.147 · Cited by 12
Investigation of the Potential for Interference with Whole Blood Glucose Strips
Sylvester, E., Price, C. P., & Burrin, J. M. (1994) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456329403100118 · Cited by 11
Chun, T.-Y., Hirose, M., Sawa, T., Harada, M., Hosokawa, T., Tanaka, Y., & Miyazaki, M. (1994) · Anesthesia & Analgesia · doi:10.1213/00000539-199411000-00030 · Cited by 10
Interference of Accel® wipes with LifeScan SureStep® Flexx glucose meters
Larsen, C. L., Jackson, C., & Lyon, M. E. (2005) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2005.11.003 · Cited by 8
Cited by
Standards of medical care in diabetes—2014
American Diabetes Association (2014) · Diabetes Care · doi:10.2337/dc14-s014 · Cited by 4,572
Standards of medical care in diabetes—2013
American Diabetes Association (2013) · Diabetes Care · doi:10.2337/dc13-s011 · Cited by 4,369
Standards of medical care in diabetes 2010
American Diabetes Association (2010) · Diabetes Care · doi:10.2337/dc10-s011 · Cited by 3,470
Standards of medical care in diabetes—2011
American Diabetes Association (2011) · Diabetes Care · doi:10.2337/dc11-s011 · Cited by 2,880
Standards of Medical Care in Diabetes—2012
(2012) · Diabetes Care · doi:10.2337/dc12-s011 · Cited by 2,240
Metadata
- Authors
- Dungan, K., Chapman, J., Braithwaite, S. S., & Buse, J.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2007-02-01
- Volume
- 30
- Issue
- 2
- Pages
- 403-409
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2034084425
- Cited by
- 292
- References
- 70
- FWCI
- 18.55
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.