Rapid Changes in Postprandial Blood Glucose Produce Concentration Differences at Finger, Forearm, and Thigh Sampling Sites
Open accessAbstract
OBJECTIVE—To compare pre- and postmeal capillary blood glucose concentrations measured at the finger, forearm, and thigh in adults with diabetes. RESEARCH DESIGN AND METHODS—For phase 1, capillary blood glucose concentrations were measured at six time points (premeal and at ∼60, 90, 120, 150, and 180 min postmeal) using a blood glucose monitoring system and technician-obtained samples collected from finger, forearm, and thigh sites of 42 adults with diabetes. The finger samples were also tested with a laboratory instrument. For phase 2, ∼14 weeks later, the testing procedures were repeated with 38 subjects from the original study population. RESULTS—Meter finger results were accurate at all time points. Alternate sites tended to produce lower glucose readings compared to finger readings at times when glucose was increasing rapidly (60 and 90 min postmeal). Forearm-to-finger differences correlated with rates of glucose change (r = 0.56, P < 0.001), as did the thigh-to-finger differences (r = 0.52, P < 0.001). Other factors, such as subject age, BMI, diabetes type, and insulin dependence did not have a significant impact on site differences. When the testing procedures were repeated with the same subjects, the pattern of site differences was consistent, although individual results were variable. CONCLUSIONS—Changes in blood glucose immediately after a meal may be identified at finger sites before detection at forearm or thigh sites. Alternate site testing appears to be a useful option for routine self-monitoring before meals; however, patients and clinicians should recognize that results may be different from fingertip results when glucose levels are changing rapidly.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“When glucose concentrations are rising or have just risen sharply, fingertip capillary glucose concentrations are greater than those in capillary blood from other sites including forearm, thigh and abdomen (Ellison et al. 2002;Jungheim & Koschinsky, 2002;van der Valk et al. 2002), with variable differences in repeated tests on the same subject (Ellison et al. 2002);”
From Glycaemic index methodology (2005) · cites this paper as Ellison et al. 2002
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“The anatomical site from which skin puncture samples are obtained influences results: Use of blood from so-called alternate sites (such as forearm or thigh rather than fingertip) for testing may exhibit a temporal lag between the circulating and measured concentrations of glucose when blood glucose is changing in vivo (117).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [117]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 178
Kuwa, K., Nakayama, T., Hoshino, T., & Tominaga, M. (2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00426-0 · Cited by 142
Tanizawa, Y., Matsuda, K., Matsuo, M., Ohta, Y., Ochi, N., Adachi, M., Koga, M., Mizuno, S., Kajita, M., Tanaka, Y., et al. (2000) · Diabetes · doi:10.2337/diabetes.49.1.114 · Cited by 66
Differences between capillary and venous blood glucose during oral glucose tolerance tests
Larsson-Cohn, U. (1976) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365517609081941 · Cited by 65
Risky Delay of Hypoglycemia Detection by Glucose Monitoring at the Arm
Jungheim, K., & Koschinsky, T. (2001) · Diabetes Care · doi:10.2337/diacare.24.7.1303-a · Cited by 62
Use of an Automated Device for Alternative Site Blood Glucose Monitoring
Fineberg, S. E., Bergenstal, R. M., Bernstein, R. M., Laffel, L. M., & Schwartz, S. L. (2001) · Diabetes Care · doi:10.2337/diacare.24.7.1217 · Cited by 59
High Incidence of Maternal Transmission of Diabetes in Obese Uruguayan Children
Pisabarro, R., Recalde, A., & Chaftare, Y. (2001) · Diabetes Care · doi:10.2337/diacare.24.7.1303 · Cited by 38
Capillary-venous differences in blood glucose values during the oral glucose tolerance test.
Eriksson, K. F., Fex, G., & Trell, E. (1983) · Clinical Chemistry · doi:10.1093/clinchem/29.5.993a · Cited by 23
Response to Jungheim and Koschinsky
McGarraugh, G. (2001) · Diabetes Care · doi:10.2337/diacare.24.7.1304 · Cited by 20
Cited by
Sorted by citations
Brouns, F., Bjorck, I., Frayn, K. N., Gibbs, A. L., Lang, V., Slama, G., & Wolever, T. M. S. (2005) · Nutrition Research Reviews · doi:10.1079/nrr2005100 · Cited by 1,093
Goldstein, D. E., Little, R. R., Lorenz, R. A., Malone, J. I., Nathan, D., Peterson, C. M., & Sacks, D. B. (2004) · Diabetes Care · doi:10.2337/diacare.27.7.1761 · Cited by 625
Assessing Glycemia in Diabetes Using Self-monitoring Blood Glucose and Hemoglobin A 1c
Saudek, C. D., Derr, R. L., & Kalyani, R. R. (2006) · JAMA · doi:10.1001/jama.295.14.1688 · Cited by 373
Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy
Yadav, J., Rani, A., Singh, V., & Murari, B. M. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.01.005 · Cited by 394
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Brand-Miller, J. C., Stockmann, K. S., Atkinson, F. S., Petocz, P., & Denyer, G. S. (2008) · American Journal of Clinical Nutrition · doi:10.3945/ajcn.2008.26354 · Cited by 265
Xie, X., Doloff, J. C., Yesilyurt, V., Sadraei, A., McGarrigle, J. J., Omami, M., Veiseh, O., Farah, S., Isa, D., Ghani, S., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0273-3 · Cited by 235
Bao, J., Atkinson, F., Petocz, P., Willett, W. C., & Brand-Miller, J. C. (2011) · The American Journal of Clinical Nutrition · doi:10.3945/ajcn.110.005033 · Cited by 187
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 164
Overview on self-monitoring of blood glucose
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.01.002 · Cited by 116
Monitoring glycemic control: the importance of self-monitoring of blood glucose
Renard, E. (2005) · The American Journal of Medicine · doi:10.1016/j.amjmed.2005.07.052 · Cited by 99
Berard, L., Blumer, I. R., Houlden, R. L., Miller, D. B., & Woo, V. C. (2013) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2013.01.017 · Cited by 98
Gebhart, S. S. P., Faupel, M. L., Fowler, R. T., Kapsner, C., Lincoln, D., McGee, V., Pasqua, J. J., Steed, L., Wangsness, M., Xu, F., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503321827812 · Cited by 69
Valeri, C., Pozzilli, P., & Leslie, D. (2004) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.512 · Cited by 69
Brand-Miller, J. C., Fatima, K., Middlemiss, C., Bare, M., Liu, V., Atkinson, F., & Petocz, P. (2007) · The American Journal of Clinical Nutrition · doi:10.1093/ajcn/85.6.1545 · Cited by 69
Minimally invasive capillary blood sampling methods
Hoffman, M. S. F., McKeage, J. W., Xu, J., Ruddy, B. P., Nielsen, P. M. F., & Taberner, A. J. (2023) · Expert Review of Medical Devices · doi:10.1080/17434440.2023.2170783 · Cited by 67
Bina, D. M., Anderson, R., Johnson, M. L., Bergenstal, R. M., & Kendall, D. M. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.981 · Cited by 66
Methodology for adding glycemic index values to 24-hour recalls
Louie, J. C. Y., Flood, V. M., Turner, N. A., Everingham, C., & Gwynn, J. (2010) · Nutrition · doi:10.1016/j.nut.2009.12.006 · Cited by 64
Koschinsky, T., Jungheim, K., & Heinemann, L. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322527030 · Cited by 64
Glycemic response and glycemic index of semolina spaghetti enriched with barley β-glucan
Chillo, S., Ranawana, D. V., Pratt, M., & Henry, C. J. (2010) · Nutrition · doi:10.1016/j.nut.2010.07.003 · Cited by 61
Stout, P. J., Racchini, J. R., & Hilgers, M. E. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2004.6.635 · Cited by 59
Gerber, K. L., & Freeman, K. P. (2016) · Veterinary Clinical Pathology · doi:10.1111/vcp.12310 · Cited by 56
Strauss, S. M., Wheeler, A. J., Russell, S. L., Brodsky, A., Davidson, R. M., Gluzman, R., Lin, L., Malo, R. G., Salis, B., Schoor, R., et al. (2009) · Journal of Periodontology · doi:10.1902/jop.2009.080542 · Cited by 51
Dunn, T. C., Eastman, R. C., & Tamada, J. A. (2004) · Diabetes Care · doi:10.2337/diacare.27.9.2161 · Cited by 50
Assessing the quality of glucose monitor studies: a critical evaluation of published reports
Mahoney, J., & Ellison, J. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.083493 · Cited by 48
Self-citationBerard, L. D., Siemens, R., & Woo, V. (2018) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2017.10.007 · Cited by 69
Forst, T., Caduff, A., Talary, M., Weder, M., Brändle, M., Kann, P., Flacke, F., Friedrich, C., & Pfützner, A. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.94 · Cited by 51
The Diabetes Research in Children Network (Direcnet) Study Group (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.675 · Cited by 48
Prevalência de fatores de risco cardiovascular em trabalhadores de uma indústria brasileira
Schmidt, A. (2009) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas) · Cited by 46
The Diabetes Research In Children, N. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322640971 · Cited by 44
Prevalência de fatores de risco cardiovascular em trabalhadores de uma indústria brasileira
Cassani, R. S. L., Nobre, F. D., Pazin‐Filho, A., & Schmidt, A. (2009) · Arquivos Brasileiros de Cardiologia · doi:10.1590/s0066-782x2009000100004 · Cited by 41
Successful Pregnancy in Women with Type 1 Diabetes: From Preconception Through Postpartum Care
Jovanovič, L., & Nakai, Y. (2006) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2005.09.008 · Cited by 39
Howard, R., Guo, J., & Hall, K. D. (2020) · American Journal of Clinical Nutrition · doi:10.1093/ajcn/nqaa198 · Cited by 37
Semi-Implantable Bioelectronics
Fang, J., Huang, S., Liu, F., He, G., Li, X., Huang, X., Chen, H., & Xie, X. (2022) · Nano-Micro Letters · doi:10.1007/s40820-022-00818-4 · Cited by 37
Self-monitoring of blood glucose in children and teens with diabetes
Bui, H., Perlman, K., & Daneman, D. (2005) · Pediatric Diabetes · doi:10.1111/j.1399-543x.2005.00095.x · Cited by 37
Haupt, A., Berg, B., Paschen, P., Dreyer, M., Häring, H.-U., Smedegaard, J., & Matthaei, S. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.597 · Cited by 36
Sakaguchi, K., Hirota, Y., Hashimoto, N., Ogawa, W., Sato, T., Okada, S., Hagino, K., Asakura, Y., Kikkawa, Y., Kojima, J., et al. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0255 · Cited by 35
Delayed effects of coffee, tea and sucrose on postprandial glycemia in lean, young, healthy adults.
Louie, J. C. Y., Atkinson, F. S., Petocz, P., & Brand-Miller, J. C. (2008) · PubMed · Cited by 31
Wang, G., Poscente, M. D., Park, S. S., Andrews, C. N., Yadid-Pecht, O., & Mintchev, M. P. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2669440 · Cited by 33
Self-monitoring Blood Glucose (SMBG): Now and the Future
Briggs, A. L., & Cornell, S. A. (2004) · Journal of Pharmacy Practice · doi:10.1177/0897190003261306 · Cited by 31
Glucose recovery in a microfluidic microdialysis biochip
Hsieh, Y., & Zahn, J. D. (2005) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2004.11.039 · Cited by 30
Evaluation of a glucose meter for determining the glycemic responses of foods
Velangi, A., Fernandes, G., & Wolever, T. M. S. (2005) · Clinica Chimica Acta · doi:10.1016/j.cccn.2005.01.025 · Cited by 29
Use of the Noninvasive GlucoWatch ® Biographer ® During Exercise of Varying Intensity
Nunnold, T., Colberg, S. R., Herriott, M. T., & Somma, C. T. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/1520915041705848 · Cited by 29
Schmid, C., Haug, C., Heinemann, L., & Freckmann, G. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0047 · Cited by 30
Evaluation of CGMS ® During Rapid Blood Glucose Changes in Patients with Type 1 Diabetes
Wilhelm, B., Forst, S., Weber, M. M., Larbig, M., Pfützner, A., & Forst, T. M. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.146 · Cited by 28
Ito, K., Okuda‐Shimazaki, J., Kojima, K., Mori, K., Tsugawa, W., Asano, R., Ikebukuro, K., & Sode, K. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112911 · Cited by 25
Gingival crevicular blood: As a non-invasive screening tool for diabetes mellitus in dental clinics
Shetty, N., Shankarapillai, R., Mathur, L. K., Manohar, B., Mathur, A., & Jain, M. (2013) · Journal of Indian Society of Periodontology · doi:10.4103/0972-124x.118319 · Cited by 25
Louie, J. C. Y., Markovic, T., Ross, G. P., Foote, D., & Brand-Miller, J. C. (2012) · Nutrients · doi:10.3390/nu5010001 · Cited by 24
Alternate site blood glucose testing: do patients prefer it?
Tieszen, K. L., & New, J. P. (2003) · Diabetic Medicine · doi:10.1046/j.1464-5491.2003.00927.x · Cited by 24
Zahn, J. D., Hsieh, Y., & Yang, M. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.536 · Cited by 24
Ridder, T. D., Steeg, B. J. V., & Laaksonen, B. D. (2009) · Journal of Biomedical Optics · doi:10.1117/1.3253353 · Cited by 23
Alternative Site Blood Glucose Testing: A Multicenter Study
Fedele, D., Corsi, A. M., Noacco, C., Prisco, F., Squatrito, S., Torre, E. L. M., Iafusco, D., Errico, M. K., Toniato, R., Nicolucci, A., et al. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322641033 · Cited by 22
Precision and accuracy of two blood glucose meters: FreeStyle Flash versus One Touch Ultra
Rivers, S. M., Kane, M. P., Bakst, G., Busch, R. S., & Hamilton, R. A. (2006) · American Journal of Health-System Pharmacy · doi:10.2146/ajhp050473 · Cited by 22
Aptamer and Oligonucleotide-Based Biosensors for Health Applications
Mayol, B., Qubbaj, I. Z., Nava-Granados, J., Vasquez, K., Keene, S. T., & Sempionatto, J. R. (2025) · Biosensors · doi:10.3390/bios15050277 · Cited by 29
Lucidarme, N., Alberti, C., Zaccaria, I., Claude, E., & Tubiana‐Rufi, N. (2005) · Diabetes Care · doi:10.2337/diacare.28.3.710 · Cited by 20
Gingival Crevicular Blood as a Potential Screening Tool: A Cross Sectional Comparative Study
Rapone, B., Ferrara, E., Santacroce, L., Topi, S., Converti, I., Gnoni, A., Scarano, A., & Scacco, S. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17207356 · Cited by 19
Assessment of three frequently used blood glucose monitoring devices in clinical routine
Züger, T., Schuler, V., Stettler, C., Diem, P., & Christ, E. (2012) · Swiss Medical Weekly · doi:10.4414/smw.2012.13631 · Cited by 20
Kempe, K. C., Budd, D., Stern, M., Ellison, J. M., Saari, L. A., Adiletto, C. A., Olin, B., Price, D. A., & Horwitz, D. L. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.916 · Cited by 19
Self-citationEvron, S., Tress, V., Ezri, T., Szmuk, P., Landau, O., Hendel, D., Schechter, P., & Medalion, B. (2007) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2006.04.005 · Cited by 19
Continuous glucose monitoring: a stepping stone in the journey towards a cure for diabetes
Thorsell, A., Gordon, M., & Jovanovic, L. G. (2004) · The Journal of Maternal-Fetal & Neonatal Medicine · doi:10.1080/14767050310001650671 · Cited by 18
Self-monitoring of blood glucose with a focus on analytical quality: an overview
Kristensen, G. B. B., & Sandberg, S. (2010) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2010.186 · Cited by 16
Alternative Sampling Site for Blood Glucose Testing in Cats: Giving the Ears a Rest
Zeugswetter, F. K., Rebuzzi, L., & Karlovits, S. (2010) · Journal of Feline Medicine and Surgery · doi:10.1016/j.jfms.2010.07.014 · Cited by 16
Ito, T., Kamoi, K., Minagawa, S., Kimura, K., & Kobayashi, A. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400420 · Cited by 16
Boutati, E. I., & Raptis, S. (2004) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.528 · Cited by 17
Forearm blood glucose testing in diabetes mellitus
Greenhalgh, S. (2004) · Archives of Disease in Childhood · doi:10.1136/adc.2002.019307 · Cited by 16
The use of glycaemic index tables to predict glycaemic index of breakfast meals
Brand-Miller, J. C., & Wolever, T. M. S. (2005) · British Journal Of Nutrition · doi:10.1079/bjn20041423 · Cited by 14
Zurbau, A., Jenkins, A. L., Jovanovski, E., Au‐Yeung, F., Bateman, E. A., Brissette, C., Wolever, T. M. S., Hanna, A., & Vuksan, V. V. V. (2018) · European Journal of Clinical Nutrition · doi:10.1038/s41430-018-0182-2 · Cited by 15
Alternative-Site Blood Glucose Measurement at the Abdomen
van der Valk, P. R., van der Schatte Olivier-Steding, I., Wientjes, K.-J. C., Schoonen, A. J., & Hoogenberg, K. (2002) · Diabetes Care · doi:10.2337/diacare.25.11.2114 · Cited by 15
Wainstein, J., Chimin, G., Landau, Z., Boaz, M., Jakubowicz, D. J., GODDARD, G. F. J., & Bar‐Dayan, Y. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0278 · Cited by 14
Improving the Safety of Blood Glucose Monitoring
Klonoff, D. C. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500601 · Cited by 13
Introduction to the Glucose Sensing Problem
Wilson, G. S., & Zhang, Y. (2009) · doi:10.1002/9780470567319.ch1 · Cited by 13
Steudle, J., Schön, C., Wargenau, M., Pauly, L., Schwejda-Guettes, S., Gaigg, B., Kuchinka-Koch, A., & Stover, J. (2018) · World Journal of Gastrointestinal Pharmacology and Therapeutics · doi:10.4292/wjgpt.v9.i3.22 · Cited by 12
Knapp, P. E., Showers, K. M., Phipps, J. C., Speckman, J. L., Sternthal, E., Freund, K. M., Ash, A. S., & Apovian, C. M. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0060 · Cited by 12
Self-Monitoring of Blood Glucose—Epidemiological and Practical Aspects
Nielsen, J. K., & Christiansen, J. S. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0007 · Cited by 12
Ugi, S., Maegawa, H., Morino, K., Nishio, Y., Sato, T., Okada, S., Kikkawa, Y., Watanabe, T., Nakajima, H., & Kashiwagi, A. (2016) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2016.40.4.326 · Cited by 11
Special Considerations for Older Adults With Diabetes Residing in Skilled Nursing Facilities
Haas, L. B. (2014) · Diabetes Spectrum · doi:10.2337/diaspect.27.1.37 · Cited by 10
Equivalence of Earlobe Site Blood Glucose Testing With Finger Stick
Anzalone, P. (2008) · Clinical Nursing Research · doi:10.1177/1054773808325050 · Cited by 10
Ridder, T. D., Steeg, B. J. V., Laaksonen, B. D., & Hull, E. L. (2011) · Journal of Biomedical Optics · doi:10.1117/1.3535594 · Cited by 10
Owiredu, W. K. B. A., Amegatcher, G., & Amidu, N. (2009) · Journal of Medical Sciences(Faisalabad) · doi:10.3923/jms.2009.185.193 · Cited by 10
Park, K., Park, M.-S., Cha, Y.-J., Kim, W., Choi, S.-S., Kim, K. O., Cha, E., & Kim, K.-A. (2010) · Annals of Laboratory Medicine · doi:10.3343/kjlm.2010.30.3.264 · Cited by 9
Continuous glucose monitoring: key challenges to replacing episodic SMBG
Stout, P. J., Racchini, J. R., Hilgers, M. E., & Noujaim, S. E. (2006) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(06)70007-9 · Cited by 9
Stavrianos, C., & Anastasiou, E. (2004) · Diabetes Care · doi:10.2337/diacare.27.2.627 · Cited by 9
Pavithran, A. A., Ramamoorthy, L., Suryanarayana, B. S., Murugesan, R., & Kumari, M. J. (2020) · Europe PMC (PubMed Central) · doi:10.34172/jcs.2020.028 · Cited by 8
Structured Blood Glucose Monitoring in Primary Care: A Practical, Evidence-Based Approach
Logan, A. D., Jones, J., & Kuritzky, L. (2020) · Clinical Diabetes · doi:10.2337/cd20-0045 · Cited by 8
Alternative Site Self Blood Glucose Testing Is Preferred by Women with Gestational Diabetes
Clarke, P., Coleman, M. A. G., & Holt, R. I. G. (2005) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2005.7.604 · Cited by 7
Sephel, G. C., & Laposata, M. (2013) · American Journal of Clinical Pathology · doi:10.1309/ajcpe2z0evfetjqt · Cited by 7
ACROCYANOSIS AS A CAUSE OF PSEUDOHYPOGLYCEMIA
Crevel, E., Ardigo, S., Perrenoud, L., & Vischer, U. M. (2009) · Journal of the American Geriatrics Society · doi:10.1111/j.1532-5415.2009.02357.x · Cited by 7
Dahan, R., Sutton, G. A., Oreff, G. L., & Kelmer, G. (2015) · Australian Veterinary Journal · doi:10.1111/avj.12303 · Cited by 6
The accuracy and acceptability of performing capillary blood glucose measurements at the earlobe
Chan, H. Y. L., Lau, T. S. L., Ho, S. Y., Leung, D. Y. P., & Lee, D. (2016) · Journal of Advanced Nursing · doi:10.1111/jan.12944 · Cited by 6
Jacoby, J. M. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400421 · Cited by 6
Lascola, K. M., Werf, K. A. V., Freese, S., Morgera, A. E., Schaeffer, D. J., & Wilkins, P. A. (2017) · Journal of Veterinary Emergency and Critical Care · doi:10.1111/vec.12588 · Cited by 6
New Developments in Home Glucose Monitoring: Minimizing the Pain
Gabbay, R. A., Phd, M., & Gabbay, R. A. (2015) · Cited by 6
Preventive Systems for the Late Complications of Diabetes
Ładyżyński, P., Wojcicki, J. M., & Foltyński, P. (2011) · Journal of Applied Biomedicine · doi:10.1016/s0208-5216(11)70007-9 · Cited by 6
Wang, H., Yao, C., Liu, Z., Wang, X., Liu, Z., Zhang, T., Huang, X., Wang, L., Wang, Y., Xiao, G., et al. (2024) · Advanced Intelligent Systems · doi:10.1002/aisy.202400547 · Cited by 5
Extending the Overnight Fast: Sex Differences in Acute Metabolic Responses to Breakfast
Atkinson, F. S., Heruc, G. A., Tan, V. M. H., Petocz, P., & Brand-Miller, J. C. (2020) · Nutrients · doi:10.3390/nu12082173 · Cited by 5
How and when to use an alternative site in self-monitoring of blood glucose
Dufaitre-Patouraux, L., Djemli, K., & Vague, P. (2004) · Diabetes & Metabolism · doi:10.1016/s1262-3636(07)70145-3 · Cited by 5
Dooley, G., Godbole, S. V., Wrobel, J., Henthorn, T., Brooks‐Russell, A., Limbacher, S. A., & Kosnett, M. J. (2025) · Journal of Analytical Toxicology · doi:10.1093/jat/bkaf043 · Cited by 4
Kesavadev, J., Chandran, G. B., Basanth, A., Krishnan, G., Shankar, A., & Jothydev, S. (2023) · Diabetes & Metabolic Syndrome Clinical Research & Reviews · doi:10.1016/j.dsx.2023.102731 · Cited by 4
Glycemic indices of three commonly consumed foods: a clinical trial in Iranian healthy adults
Hajifaraji, M., Rezvani, V., Yaghoobi, A., Morteza, H., & Maddah, M. (2012) · Mediterranean Journal of Nutrition and Metabolism · doi:10.1007/s12349-012-0094-y · Cited by 4
Pfützner, A., Hermanns, N., Schröder, S., Wegenstein, M., Larbig, M., Mondok, Á., Forst, T. M., & Haak, T. (2002) · Swiss Medical Weekly · doi:10.4414/smw.2002.09991 · Cited by 4
Metadata
- Authors
- Ellison, J. M., Stegmann, J. M., Colner, S. L., Michael, R. H., Sharma, M. K., Ervin, K. R., & Horwitz, D. L.
- Affiliations
- LifeScan, Milpitas, California
- ORCID
- Horwitz, David L.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2002-06-01
- Volume
- 25
- Issue
- 6
- Pages
- 961-964
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2154078343
- Semantic Scholar
- e98ef81443d8253dc4e47cf37f2bfa46dd424c26
- Cited by
- 133
- References
- 12
- Citations by year
- 2026: 2 · 2025: 4 · 2024: 1 · 2023: 4 · 2022: 2 · 2021: 2 · 2020: 6 · 2018: 4 · 2017: 3 · 2016: 3 · 2015: 5 · 2014: 1 · 2013: 9 · 2012: 4
- FWCI
- 8.08
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Diabetes Care (publishedVersion)
- PubMed (publishedVersion)
- Funders
- American Diabetes Association
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Keywords
- postprandial blood glucose, capillary blood glucose, forearm, thigh, self-monitoring of blood glucose, meter accuracy, diabetes mellitus
- MeSH terms
- Adult, Blood Glucose, Blood Specimen Collection, Capillaries, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Female, Fingers, Forearm, Humans, Leg, Male, Middle Aged, Regression Analysis, Time Factors, Blood Glucose Self-Monitoring, Reproducibility of Results, Postprandial Period
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review