Point-of-Care Glucose Testing
The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Objective.—To assess the clinical performance of glucose meter systems when used with critically ill patients. Design.—Two glucose meter systems (SureStepPro and Precision G) and a modular adaptation (Immediate Response Mobile Analysis-SureStepPro) were assessed clinically using arterial samples from critically ill patients. A biosensor-based analyzer (YSI 2700) and a hospital chemistry analyzer (Synchron CX-7) were the primary and secondary reference instruments, respectively. Patients and Setting.—Two hundred forty-seven critical care patients at the University of California, Davis, Medical Center participated in this study. Outcome Measures.—Error tolerances of ±15 mg/dL for glucose levels ≤100 mg/dL and ±15% for glucose levels >100 mg/dL were used to evaluate glucose meter performance; 95% of glucose meter measurements should fall within these tolerances. Results.—Compared to the primary reference method, 98% to 100% of SureStepPro and 91% to 95% of Precision G measurements fell within the error tolerances. Paired differences of glucose measurements versus critical care variables (Po2, pH, Pco2, and hematocrit) were analyzed to determine the effects of these variables on meter measurements. Po2 and Pco2 decreased Precision G and SureStepPro measurements, respectively, but not enough to be clinically significant based on the error tolerance criteria. Hematocrit levels affected glucose measurements on both meter systems. Modular adaptation did not affect test strip performance. Conclusions.—Glucose meter measurements correlated best with primary reference instrument measurements. Overall, both glucose meter systems showed acceptable performance for point-of-care testing. However, the effects of some critical care variables, especially low and high hematocrit values, could cause overestimated or underestimated glucose measurements.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Ferrocene-mediated enzyme electrode for amperometric determination of glucose
Cass, A. E. G., Davis, G., Francis, G. D., Hill, H. A. O., Aston, W. J., Higgins, I. J., Plotkin, E. V., Scott, L. D. L., & Turner, A. P. F. (1984) · Analytical Chemistry · doi:10.1021/ac00268a018 · Cited by 1,646
A Text-Book of Physical Chemistry
(1936) · Nature · doi:10.1038/138864b0 · Cited by 1,146
A textbook of physical chemistry
Adamson, A. W. (1973) · Cited by 343
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Self-citationConsensus Statement on Self-Monitor ing of Blood Glucose
(1987) · Diabetes Care · doi:10.2337/diacare.10.1.95 · Cited by 248
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
Self-citationCommercial biosensors : applications to clinical, bioprocess, and environmental samples
Ramsay, G. (1998) · Wiley eBooks · Cited by 109
Electrochemical measurement of glucose using various electron acceptors
Schläpfer, P., Mindt, W., & Racine, P. (1974) · Clinica Chimica Acta · doi:10.1016/0009-8981(74)90408-2 · Cited by 103
Sell-Monitoring of Blood Glucose
(1996) · Diabetes Care · doi:10.2337/diacare.19.1.s62 · Cited by 89
Cited by
Jacobi, J., Bircher, N., Krinsley, J., Agus, M., Braithwaite, S. S., Deutschman, C., Freire, A. X., Geehan, D., Kohl, B., Nasraway, S. A., et al. (2012) · Critical Care Medicine · doi:10.1097/ccm.0b013e3182653269 · Cited by 571
Reliability of point-of-care testing for glucose measurement in critically ill adults*
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000189939.10881.60 · Cited by 395
Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Self-citationGlucose Measurement: Confounding Issues in Setting Targets for Inpatient Management
Dungan, K., Chapman, J., Braithwaite, S. S., & Buse, J. (2007) · Diabetes Care · doi:10.2337/dc06-1679 · Cited by 292
Continuous glucose monitoring: an Endocrine Society Clinical Practice Guideline
Klonoff, D. C., Buckingham, B., Christiansen, J. S., Montori, V. M., Tamborlane, W. V., Vigersky, R. A., & Wolpert, H. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-2756 · Cited by 211
Tang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00038 · Cited by 188
Self-citationCorstjens, A. M., Ligtenberg, J. J., van der Horst, I. C., Spanjersberg, R., Lind, J. S., Tulleken, J. E., Meertens, J. H., & Zijlstra, J. G. (2006) · Critical Care · doi:10.1186/cc5048 · Cited by 163
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
Ellison, J. M., Stegmann, J. M., Colner, S. L., Michael, R. H., Sharma, M. K., Ervin, K. R., & Horwitz, D. L. (2002) · Diabetes Care · doi:10.2337/diacare.25.6.961 · Cited by 133
Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816641433 · Cited by 115
Glucose Metabolism in Pregnancy at High Altitude
Krampl, E., Kametas, N. A., Nowotny, P., Roden, M., & Nicolaides, K. H. (2001) · Diabetes Care · doi:10.2337/diacare.24.5.817 · Cited by 41
Performance of a new interference-resistant glucose meter
Vanavanan, S., Santanirand, P., Chaichanajarernkul, U., Chittamma, A., DuBois, J. A., Shirey, T., & Heinz, M. (2010) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.09.010 · Cited by 39
Evaluation of hematocrit interference with MyStar extra and seven competitive devices
Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814565790 · Cited by 22
Warner, J. V., Wu, J. Y., Buckingham, N., McLeod, D. S. A., Mottram, B., & Carter, A. C. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0129 · Cited by 20
Factors interfering with the accuracy of five blood glucose meters used in Chinese hospitals
Lv, H., Zhang, G., Kang, X., Yuan, H., Lv, Y., Wang, W., & Randall, R. (2013) · Journal of Clinical Laboratory Analysis · doi:10.1002/jcla.21611 · Cited by 20
MacRury, S., Srinivasan, A., & Mahoney, J. J. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700215 · Cited by 7
Metadata
- Authors
- Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J.
- Venue
- Archives of Pathology & Laboratory Medicine
- Published
- 2000-02-01
- Volume
- 124
- Issue
- 2
- Pages
- 257-266
- ISSN
- 0003-9985 (ISSN-L)
- 1543-2165
- OpenAlex
- W2138905008
- Cited by
- 180
- References
- 23
- FWCI
- 4.72
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.