Clinical evaluation of the GlucoWatch® biographer: a continual, non-invasive glucose monitor for patients with diabetes
Abstract
A device providing frequent, automatic, and non-invasive glucose measurements for persons with diabetes has been developed: the GlucoWatch biographer. This device extracts glucose through intact skin via reverse iontophoresis where it is detected by an amperometric biosensor. The biographer can provide glucose readings every 20 min for 12 h. The performance of this device was evaluated in two large clinical studies in a controlled clinical environment (n=231), and the home environment (n=124). Accuracy of the biographer was evaluated by comparing the automatic biographer readings to serial finger-stick blood glucose (BG) measurements. Biographer performance was comparable in both environments. Mean difference between biographer and finger-stick measurements was -0.01 and 0.26 mmol l(-1) for the clinical and home environments, respectively. The mean absolute value of the relative difference was 1.06 and 1.18 mmol l(-1) for the same studies. Correlation coefficient (r) between biographer and finger-stick measurements was 0.85 and 0.80 for the two studies. In both studies, over 94% of the biographer readings were in the clinically acceptable A+B region of the Clarke Error Grid. A slight positive bias is observed for the biographer readings at low BG levels. Biographer accuracy is relatively constant over all rates of BG changes, except when BG decreases more than 10 mmol l(-1) h(-1), which occurred for only 0.2% of points in the home environment study. Biographer precision, as measured by CV%, is approx. 10%. Skin irritation, characterized by erythema and edema, was either non-existent or mild in >90% of subjects and resolved in virtually all subjects without treatment in several days.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The GlucoWatch Biographer was an FDA-approved device capable of non-invasive glucose monitoring over a 12 hour period with 6 measurements per hour.
“This US Food and Drug Administration (FDA)-approved device is capable of non-invasive glucose monitoring over a 12 hour period with 6 measurements per hour.”
From Wearable biosensors for healthcare monitoring (2019) · cites this paper as [30]
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667 · see citing works
Turner, R. C., Fox, C., Matthews, McElroy, H. H., Cull, C. A., Holman, R. R., Neil, P., Hadden, D. R., Wright, D., Manley, E., et al. (1998) · The Lancet · Cited by 8,419 · see citing works
Draize, J. H., Woodard, G. E., & Calvery, H. O. (1944) · Journal of Pharmacology and Experimental Therapeutics · doi:10.1016/s0022-3565(25)08751-8 · Cited by 2,607 · see citing works
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435 · see citing works
Statistical Adjustment of Data
Anderson, R. L., & Deming, W. E. (1944) · Southern Economic Journal · doi:10.2307/1052578 · Cited by 775 · see citing works
Noninvasive Glucose Monitoring
Tamada, J. A., Garg, S. K., Jovanovič, L., Pitzer, K. R., Fermi, S. J., & Potts, R. O. (1999) · JAMA · doi:10.1001/jama.282.19.1839 · Cited by 281 · see citing works
Self-citationGarg, S. K., Potts, R. O., Ackerman, N. R., Fermi, S. J., Tamada, J. A., & Chase, H. P. (1999) · Diabetes Care · doi:10.2337/diacare.22.10.1708 · Cited by 200 · see citing works
Self-citationElimination of the Acetaminophen Interference in an Implantable Glucose Sensor
Zhang, Y., Hu, Y., Wilson, G. S., Moatti-Sirat, D., Poitout, V., & Reach, G. (1994) · Analytical Chemistry · doi:10.1021/ac00079a038 · Cited by 184 · see citing works
Measurement of glucose in diabetic subjects using noninvasive transdermal extraction
Tamada, J. A., Bohannon, N. J. V., & Potts, R. O. (1995) · Nature Medicine · doi:10.1038/nm1195-1198 · Cited by 182 · see citing works
Self-citationLouie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 178 · see citing works
The GlucoWatch®biographer: a frequent, automatic and noninvasive glucose monitor
Tierney, M. J., Tamada, J. A., Potts, R. O., Eastman, R. C., Pitzer, K. R., Ackerman, N. R., & Fermi, S. J. (2000) · Annals of Medicine · doi:10.3109/07853890009002034 · Cited by 154 · see citing works
Self-citationReverse Iontophoresis: Development of a Noninvasive Approach for Glucose Monitoring
Rao, G., Glikfeld, P., & Guy, R. H. (1993) · Pharmaceutical Research · doi:10.1023/a:1018926215306 · Cited by 126 · see citing works
Validation of home blood glucose meters with respect to clinical and analytical approaches
Brunner, G. A., Ellmerer, M., Sendlhofer, G., Wutte, A., Trajanoski, Z., Schaupp, L., Quehenberger, F., Wach, P., Krejs, G. J., & Pieber, T. R. (1998) · Diabetes Care · doi:10.2337/diacare.21.4.585 · Cited by 119 · see citing works
Statistical adjustment of data
Oppermann, R. H. (1944) · Journal of the Franklin Institute · doi:10.1016/s0016-0032(44)90975-0 · Cited by 100 · see citing works
An Overview of Minimally Invasive Technologies
Ginsberg, B. H. (1992) · Clinical Chemistry · doi:10.1093/clinchem/38.9.1596 · Cited by 71 · see citing works
Design of a Biosensor for Continual, Transdermal Glucose Monitoring
Tierney, M. J., Jayalakshmi, Y., Parris, N. A., Reidy, M. P., Uhegbu, C. E., & Vijayakumar, P. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.9.1681 · Cited by 59 · see citing works
Self-citationKurnik, R. T. (1999) · Sensors and Actuators B Chemical · doi:10.1016/s0925-4005(99)00239-7 · Cited by 41 · see citing works
Modelling and simulation of a diffusion limited glucose biosensor
Cambiaso, A., Delfino, L., Grattarola, M., Verreschi, G., Ashworth, D. C., Maines, A., & Vadgama, P. M. (1996) · Sensors and Actuators B Chemical · doi:10.1016/0925-4005(96)80099-2 · Cited by 36 · see citing works
A Ward Comparison Between the One Touch II and Glucometer II Blood Glucose Meters
Kilpatrick, E. S., McLeod, M. J., Rumley, A. G., & SMALL, M. P. (1994) · Diabetic Medicine · doi:10.1111/j.1464-5491.1994.tb02023.x · Cited by 13 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373 · see citing works
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707 · see citing works
Yang, J. C., Mun, J., Kwon, S. Y., Park, S., Bao, Z., & Park, S. (2019) · Advanced Materials · doi:10.1002/adma.201904765 · Cited by 1,641 · see citing works
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486 · see citing works
Non-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247 · see citing works
A review of smart homes—Present state and future challenges
Chan, M., Estève, D., Escriba, C., & Campo, E. (2008) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2008.02.001 · Cited by 1,007 · see citing works
Wearable health devices-vital sign monitoring, systems and technologies
Dias, D., & Paulo Silva Cunha, J. (2018) · Sensors · doi:10.3390/s18082414 · Cited by 919 · see citing works
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660 · see citing works
Glucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584 · see citing works
Non-invasive glucose monitoring technology in diabetes management: A review
Vashist, S. K. (2012) · Analytica Chimica Acta · doi:10.1016/j.aca.2012.03.043 · Cited by 575 · see citing works
Non-Invasive Blood Glucose Monitoring Technology: A Review
Tang, L., Chang, S. J., Chen, C.-J., & Liu, J.-T. (2020) · Sensors · doi:10.3390/s20236925 · Cited by 481 · see citing works
Wearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Technologies for Continuous Glucose Monitoring: Current Problems and Future Promises
Vaddiraju, S., Burgess, D. J., Tomazos, I., Jain, F. C., & Papadimitrakopoulos, F. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400632 · Cited by 285 · see citing works
Tura, A., Maran, A., & Pacini, G. (2007) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2006.10.027 · Cited by 283 · see citing works
Current development in non-invasive glucose monitoring
Ferrante do Amaral, C. E., & Wolf, B. (2008) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2007.06.003 · Cited by 269 · see citing works
Achievements and Challenges for Real-Time Sensing of Analytes in Sweat within Wearable Platforms
Brothers, M. C., DeBrosse, M., Grigsby, C. C., Naik, R. R., Hussain, S. M., Heikenfeld, J., & Kim, S. S. (2019) · Accounts of Chemical Research · doi:10.1021/acs.accounts.8b00555 · Cited by 207 · see citing works
Serum 1,5-Anhydroglucitol (GlycoMark™): A Short-Term Glycemic Marker
Buse, J. B., Freeman, J. L. R., Edelman, S. V., Jovanovic, L., & McGill, J. B. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503765691839 · Cited by 187 · see citing works
Self-citationMunje, R. D., Muthukumar, S., & Prasad, S. (2017) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2016.07.088 · Cited by 190 · see citing works
Li, G., & Wen, D. (2020) · Journal of Materials Chemistry B · doi:10.1039/c9tb02474c · Cited by 119 · see citing works
Venugopal, M., Feuvrel, K. E., Mongin, D., Bambot, S., Faupel, M., Panangadan, A., Talukder, A., & Pidva, R. (2008) · IEEE Sensors Journal · doi:10.1109/jsen.2007.912544 · Cited by 76 · see citing works
Kühner, L., Semenyshyn, R., Hentschel, M., Neubrech, F., Tarín, C., & Giessen, H. (2019) · ACS Sensors · doi:10.1021/acssensors.9b00488 · Cited by 68 · see citing works
Modeling Glucose Transport From Systemic Circulation to Sweat
La Count, T. D., Jajack, A., Heikenfeld, J., & Kasting, G. B. (2019) · Journal of Pharmaceutical Sciences · doi:10.1016/j.xphs.2018.09.026 · Cited by 60 · see citing works
Wang, Z., Dong, S., Gui, M., Asif, M., Wang, W., Wang, F., & Liu, H. (2018) · Analytical Biochemistry · doi:10.1016/j.ab.2017.12.010 · Cited by 59 · see citing works
Dudde, R., Vering, T., Piechotta, G., & Hintsche, R. (2006) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2006.864477 · Cited by 47 · see citing works
Enhancing glucose flux into sweat by increasing paracellular permeability of the sweat gland
Jajack, A., Brothers, M., Kasting, G., & Heikenfeld, J. (2018) · PLOS ONE · doi:10.1371/journal.pone.0200009 · Cited by 35 · see citing works
Souza, S. L., Graça, G., & Oliva, A. (2018) · Skin Research and Technology · doi:10.1111/srt.12412 · Cited by 34 · see citing works
The Role of Wearable Technologies and Telemonitoring in Managing Vascular Disease
Chan, C., Sounderajah, V., Acharya, A., Normahani, P., Bicknell, C., & Riga, C. (2020) · Vascular and Endovascular Review · doi:10.15420/ver.2019.11 · Cited by 6 · see citing works
- Authors
- Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S.
- Affiliations
- Cygnus (United States) (US)
- Sansum Medical Clinic (US)
- University of Colorado Health (US)
- Corresponding author
- Tierney, M.J.
- Venue
- Biosensors and Bioelectronics
- Publisher
- Elsevier BV
- Published
- 2001-12
- Volume
- 16
- Issue
- 9-12
- Pages
- 621-629
- ISSN
- 0956-5663 (ISSN-L)
- OpenAlex
- W2066037251
- Cited by
- 390
- References
- 20
- Citations by year
- 2026: 20 · 2025: 22 · 2024: 19 · 2023: 28 · 2022: 23 · 2021: 21 · 2020: 22 · 2019: 30 · 2018: 24 · 2017: 12 · 2016: 12 · 2015: 9 · 2014: 6 · 2013: 7 · 2012: 7
- FWCI
- 5.77
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Biosensors and Bioelectronics (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Keywords
- non-invasive glucose monitoring, diabetes mellitus, reverse iontophoresis, amperometric biosensor, Clarke error grid, skin irritation, home monitoring, continuous glucose monitoring
- MeSH terms
- Diabetes Mellitus, Equipment Design, Glucose, Humans, Iontophoresis, Skin, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review