Assessment of rat sciatic nerve function following acute implantation of high density utah slanted electrode array (25 electrodes/mm2) based on neural recordings and evoked muscle activity
Abstract
Introduction: High density Utah slanted electrode arrays (HD‐USEAs) have been developed recently for intrafascicular access to submillimeter neural structures. Insertion of such high electrode density devices may cause nerve crush injury, counteracting the intended improved selective nerve fiber access. Methods: HD‐USEAs were implanted into sciatic nerves of anesthetized rats. Nerve function was assessed before and after HD‐USEA implantation by measuring changes in evoked muscle and nerve compound action potentials and single unit neuronal recordings. Results: Neural activity was recorded with over half of all implanted electrodes. Average decreases of 38%, 36%, and 13% in nerve, medial gastrocnemius, and tibialis anterior compound action potential amplitudes, respectively, were observed following array implantation. Only 1 of 8 implantations resulted in loss of all signals. Conclusions: These studies demonstrate that HD‐USEAs provide a useful neural interface without causing a nerve crush injury that would otherwise negate their use in acute preparations (<12 h). Muscle Nerve 50: 417–424, 2014
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The observed increase in CNAP peak potentials with increasing applied current amplitudes is consistent with the literature.
“which are consistent with the literature (10, 12, 54)”
From Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording (2019) · cites this paper as [54]
Cites
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · cited 1× · see citing works
“In the field of neural interfaces, the use of intrafascicular electrodes has allowed for increasingly selective access to peripheral nerve fibers, resulting in lower stimulus intensities and higher signal-tonoise ratios for recording of neural activity.(1) One such device, the Utah slanted electrode array (USEA) conta…”
BackgroundFiber composition of the rat sciatic nerve
Schmalbruch, H. (1986) · The Anatomical Record · doi:10.1002/ar.1092150111 · Cited by 402 · see citing works
Selective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 358 · cited 3× · see citing works
“This long exposure length, combined with the small nerve diameter, leaves the nerve more vulnerable to dehydration than larger nerves, such as the feline sciatic, where such challenges have not been reported.(2) These factors, along with possible crush injury, may also account for the loss of all signals in 1 experime…”
BackgroundSelf-citationA method for pneumatically inserting an array of penetrating electrodes into cortical tissue
Rousche, P. J., & Normann, R. A. (1992) · Annals of Biomedical Engineering · doi:10.1007/bf02368133 · Cited by 305 · see citing works
Self-citationLong-term stimulation and recording with a penetrating microelectrode array in cat sciatic nerve
Branner, A., Stein, R. B., Fernandez, E., Aoyagi, Y., & Normann, R. A. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.820321 · Cited by 301 · see citing works
Self-citationA multielectrode array for intrafascicular recording and stimulation in sciatic nerve of cats
Branner, A., & Normann, R. A. (2000) · Brain Research Bulletin · doi:10.1016/s0361-9230(99)00231-2 · Cited by 234 · see citing works
Self-citationWark, H. A. C., Sharma, R., Mathews, K. S., Fernandez, E., Yoo, J., Christensen, B., Tresco, P., Rieth, L., Solzbacher, F., Normann, R. A., et al. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/4/045003 · Cited by 148 · see citing works
Self-citationThe distribution of conduction velocities (DCV) in peripheral nerves: A review
Dorfman, L. J. (1984) · Muscle & Nerve · doi:10.1002/mus.880070103 · Cited by 99 · see citing works
Topographical distribution of motor fascicles in the sciatic‐tibial nerve of the rat
Badia, J., Pascual‐Font, A., Vivó, M., Udina, E., & Navarro, X. (2010) · Muscle & Nerve · doi:10.1002/mus.21652 · Cited by 98 · see citing works
Effects of terpineol on the compound action potential of the rat sciatic nerve
Moreira, M. R. C., da Cruz, G. M. P., Lopes, M. S., Albuquerque, A. A. C., & Leal‐Cardoso, J. H. (2001) · Brazilian Journal of Medical and Biological Research · doi:10.1590/s0100-879x2001001000015 · Cited by 61 · see citing works
Variability of repeated nerve conduction studies
Bleasel, A., & Tuck, R. R. (1991) · Electroencephalography and Clinical Neurophysiology/Evoked Potentials Section · doi:10.1016/0168-5597(91)90049-4 · Cited by 60 · see citing works
Models and analysis of compound nerve action potentials.
Schoonhoven, R., & Stegeman, D. F. (1991) · PubMed · Cited by 55 · see citing works
Effects of estragole on the compound action potential of the rat sciatic nerve
Leal‐Cardoso, J. H., Matos-Brito, B. G., Lopes-Junior, J. E. G., et al. (2004) · Brazilian Journal of Medical and Biological Research · doi:10.1590/s0100-879x2004000800009 · Cited by 36 · see citing works
Dalkılıç, N., Tuncer, S., Barışkaner, H., et al. (2009) · YAKUGAKU ZASSHI · doi:10.1248/yakushi.129.485 · Cited by 25 · see citing works
Current-Distance Relationships for Peripheral Nerve Stimulation Localization
Li, J., Kong, X., Gozani, S. N., Shi, R., & Borgens, R. B. (2010) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181fca16b · Cited by 24 · see citing works
An early diagnostic tool for diabetic neuropathy: Conduction velocity distribution
Tuncer, S., Dalkılıç, N., Esen, H. H., et al. (2011) · Muscle & Nerve · doi:10.1002/mus.21837 · Cited by 24 · see citing works
Local anesthetic effects of cocaethylene and isopropylcocaine on rat peripheral nerves
Tokuno, H. A., Bradberry, C. W., Everill, B., et al. (2003) · Brain Research · doi:10.1016/j.brainres.2003.10.024 · Cited by 14 · see citing works
Does the conduction velocity distribution change along the nerve?
Pehlivan, F., Dalkılıç, N., & Kızıltan, E. (2004) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2004.02.009 · Cited by 14 · see citing works
Murphy, B., Krieger, C., & Hoffer, J.-A. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2003.08.013 · Cited by 13 · see citing works
Carrington, A. L., Ettlinger, C. B., & Tomlinson, D. R. (1994) · Journal of Diabetes and its Complications · doi:10.1016/1056-8727(94)90008-6 · Cited by 7 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Zhang, Y., Zheng, N., Cao, Y., Wang, F., Wang, P., Ma, Y., Lu, B., Hou, G., Fang, Z., Liang, Z., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw1066 · Cited by 314 · see citing works
Implantable neurotechnologies: a review of micro- and nanoelectrodes for neural recording
Patil, A. C., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1430-4 · Cited by 175 · see citing works
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
Xiang, Z., Yen, S., Sheshadri, S., Wang, J., Lee, S., Liu, Y., Liao, L., Thakor, N. V., & Lee, C. (2016) · Advanced Materials · doi:10.1002/adma.201503423 · Cited by 116 · see citing works
Wark, H. A. C., Mathews, K. S., Normann, R. A., & Fernandez, E. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/4/046027 · Cited by 54 · see citing works
Self-citationFlexible epineural strip electrode for recording in fine nerves
Lee, S., Yen, S.-C., Sheshadri, S., Martinez, I. D., Xue, N., Xiang, Z., Thakor, N. V., & Lee, C. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2466442 · Cited by 33 · see citing works
Frederick, R. A., Troyk, P. R., & Cogan, S. F. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac2bb8 · Cited by 7 · see citing works
- Authors
- Mathews, K. S., Wark, H. A. C., & Normann, R. A.
- Affiliations
- University of Utah (US)
- Corresponding author
- Normann, Richard A.
- Venue
- Muscle & Nerve
- Publisher
- Wiley
- Published
- 2014-05-14
- Volume
- 50
- Issue
- 3
- Pages
- 417-424
- ISSN
- 0148-639X (ISSN-L)
- 1097-4598
- OpenAlex
- W2109286850
- Cited by
- 26
- References
- 20
- Citations by year
- 2024: 1 · 2023: 2 · 2022: 2 · 2021: 3 · 2020: 2 · 2019: 3 · 2018: 2 · 2016: 2 · 2015: 6 · 2014: 3
- FWCI
- 1.67
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Muscle & Nerve (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Science Foundation (CBET-1134545)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Life Sciences › Neuroscience › Cognitive Neuroscience › Undergraduate Neuroscience Education and Research
- Keywords
- high-density electrode arrays, sciatic nerve, nerve crush injury, compound action potential, single-unit recording, neural recording, neural interfaces, rats
- MeSH terms
- Action Potentials, Animals, Electrodes, Implanted, Evoked Potentials, Foot, Microelectrodes, Sciatic Nerve, Rats, Sprague-Dawley, Muscle, Skeletal, Rats, Electrophysiological Phenomena
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review