The effect of unfractionated heparin, enoxaparin and danaparoid on lupus anticoagulant testing. Can activated carbon eliminate false positive results?
Open accessAbstract
BACKGROUND: Heparins and heparinoids interfere with functional clotting assays used for lupus anticoagulant (LAC) detection. However, current guidelines for LAC testing do not provide clear guidance on this matter. OBJECTIVES: We aimed to assess to effect of unfractionated heparin (UFH), enoxaparin, and danaparoid on LAC assays over broad anti-Xa activity ranges and to evaluate whether activated carbon (AC) is able to neutralize these effects. METHODS: UFH (0.1-3.0 IU/mL), enoxaparin (0.2-2.9 IU/mL), and danaparoid (0.6-2.2 IU/mL) were spiked to normal pooled plasma. AC was added at multiple activity levels. Anti-Xa assays and LAC tests were performed on all samples using Stago analyzers and reagents. RESULTS: Abnormal activated partial thromboplastin time (APTT) screening and mixing tests were obtained at the lowest levels for all compounds. Abnormal APTT confirmation tests were seen from 2.5 and 1.9 anti-Xa IU/mL for enoxaparin and danaparoid, respectively. Abnormal dilute Russell's viper venom test (dRVVT) screening tests were obtained from 1.6, 1.4, and 1.1 anti-Xa IU/mL for UFH, enoxaparin, and danaparoid, respectively. Mixing tests were abnormal from 2.5 and 1.3 anti-Xa IU/mL for enoxaparin and danaparoid, respectively. Abnormal dRVVT confirmation results were seen for danaparoid only from 1.9 anti-Xa IU/mL. AC was unable to neutralize anti-Xa activity in plasma and overcome the effect of the tested anticoagulants on LAC assays but may cause prolongation of APTT clotting times. CONCLUSIONS: UFH, enoxaparin, and danaparoid clearly affected LA tests; however, false-positive LAC conclusions were obtained at supratherapeutic enoxaparin and danaparoid levels only. AC may prolong APTT screen clotting times, requiring 3-step testing to avoid potential misdiagnosis of LAC.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Pretreatment with adsorbents may alter clotting times, producing false-positive or false-negative LA results or incomplete removal of direct oral anticoagulants.
“Several studies indicated that pretreatment with adsorbents may affect clotting times resulting in false-positive or false-negative LA results, or incomplete removal of DOACs.33,[41][42][43]”
From Guidance from the Scientific and Standardization Committee for lupus anticoagulant/antiphospholipid antibodies of the International Society on Thrombosis and Haemostasis (2020) · cites this paper as [33]
The Results and Discussion report that activated-carbon pretreatment prolonged APTT screening clotting times, and the Conclusions warn that three-step testing is needed to avoid potential misdiagnosis. But the study found that pretreatment did not alter its final LA interpretations; it does not report false-negative LA results or incomplete DOAC removal. Its false-positive LA findings concerned high levels of enoxaparin and danaparoid, not adsorbent pretreatment. (LLM, conf=0.86)
Experimental — accuracy not yet measuredEnoxaparin caused false-positive APTT-based LA detection only at supra-therapeutic anti-Xa activity levels.
“In contrast, enoxaparin caused false-positive aPTT-based LA detection only at supra-therapeutic anti-Xa activity levels.”
Unfractionated heparin and enoxaparin affect dRVVT at supra-therapeutic anti-Xa levels but did not cause false-positive LA in the three-step procedure.
“a recent study 33 showed unfractionated heparin (UFH) and enoxaparin affect the dRVVT at supra-therapeutic anti-Xa levels but did not lead to false-positive LA in the three step procedure.”
False positives have been described for aPTT and dRVVT even when utilizing heparin neutralizers.
“False positives have been described for aPTT and dRVVT even when utilizing neutralizers [68,69,70].”
From Lupus Anticoagulant Detection under the Magnifying Glass (2023) · cites this paper as [70]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
MIYAKIS, S., LOCKSHIN, M. D., ATSUMI, T., BRANCH, D. W., BREY, R. L., CERVERA, R., DERKSEN, R. H. W. M., DE GROOT, P. G., KOIKE, T., MERONI, P. L., et al. (2006) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2006.01753.x · Cited by 7,203 · see citing works
EULAR recommendations for the management of antiphospholipid syndrome in adults
Tektonidou, M. G., Andreoli, L., Limper, M., Amoura, Z., Cervera, R., Costedoat-Chalumeau, N., Cuadrado, M. J., Dörner, T., Ferrer-Oliveras, R., Hambly, K., et al. (2019) · Annals of the Rheumatic Diseases · doi:10.1136/annrheumdis-2019-215213 · Cited by 1,261 · see citing works
Update of the guidelines for lupus anticoagulant detection
PENGO, V., TRIPODI, A., REBER, G., RAND, J. H., ORTEL, T. L., GALLI, M., & DE GROOT, P. G. (2009) · Journal of Thrombosis and Haemostasis · doi:10.1111/j.1538-7836.2009.03555.x · Cited by 1,197 · see citing works
Garcia, D. A., Baglin, T. P., Weitz, J. I., & Samama, M. M. (2012) · CHEST Journal · doi:10.1378/chest.11-2291 · Cited by 1,098 · see citing works
Guidelines on the investigation and management of antiphospholipid syndrome
Keeling, D., Mackie, I., Moore, G. W., Greer, I. A., Greaves, M., & British Committee for Standards in Haematology (2012) · British Journal of Haematology · doi:10.1111/j.1365-2141.2012.09037.x · Cited by 740 · see citing works
Heparin-Induced Thrombocytopenia (HIT): An Overview of 230 Patients Treated with Orgaran (Org 10172)
Magnani, H. N. (1993) · Thrombosis and Haemostasis · doi:10.1055/s-0038-1649626 · Cited by 445 · see citing works
Laboratory criteria for antiphospholipid syndrome: communication from the SSC of the ISTH
Devreese, K. M. J., Ortel, T. L., Pengo, V., & de Laat, B. (2018) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.13976 · Cited by 269 · see citing works
Self-citationIampietro, R., Woolley, A. M., Preston, F. E., & Kitchen, S. (1999) · Thrombosis and Haemostasis · doi:10.1055/s-0037-1614377 · Cited by 112 · see citing works
Martinuzzo, M. E., Barrera, L. H., D 'Adamo, M. A., Otaso, J. C., Gimenez, M. I., & Oyhamburu, J. (2014) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12138 · Cited by 91 · see citing works
Simple method for removing DOACs from plasma samples
Exner, T., Michalopoulos, N., Pearce, J., et al. (2018) · Thrombosis Research · doi:10.1016/j.thromres.2018.01.047 · Cited by 81 · see citing works
Detection of lupus anticoagulant in the era of direct oral anticoagulants
Hoxha, A., Banzato, A., Ruffatti, A., & Pengo, V. (2016) · Autoimmunity Reviews · doi:10.1016/j.autrev.2016.12.010 · Cited by 74 · see citing works
Platton, S., & Hunt, C. (2019) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12950 · Cited by 74 · see citing works
Favresse, J., Lardinois, B., Sabor, L., Devalet, B., Vandepapeliere, J., Braibant, M., Lessire, S., Chatelain, B., Jacqmin, H., Douxfils, J., et al. (2018) · TH Open · doi:10.1055/s-0038-1657785 · Cited by 73 · see citing works
Favaloro, E. J., Mohammed, S., Curnow, J., & Pasalic, L. (2019) · Pathology · doi:10.1016/j.pathol.2018.11.008 · Cited by 67 · see citing works
Mixing studies in lupus anticoagulant testing are required at least in some type of samples
Devreese, K. M. J., & de Laat, B. (2015) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.13020 · Cited by 63 · see citing works
Self-citationFavaloro, E. J., Gilmore, G., Arunachalam, S., Mohammed, S., & Baker, R. I. (2019) · Thrombosis Research · doi:10.1016/j.thromres.2019.05.013 · Cited by 53 · see citing works
Cohen, H., Mackie, I. J., & Devreese, K. M. J. (2019) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.14560 · Cited by 48 · see citing works
Self-citationZąbczyk, M., Kopytek, M., Natorska, J., & Undas, A. (2019) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2018-1197 · Cited by 47 · see citing works
Frans, G., Meeus, P., & Bailleul, E. (2019) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.14488 · Cited by 44 · see citing works
Moore, G. W., Culhane, A. P., Daw, C. R., Noronha, C. P., & Kumano, O. (2016) · Thrombosis Research · doi:10.1016/j.thromres.2016.01.019 · Cited by 42 · see citing works
Exner, T., Ahuja, M., & Ellwood, L. (2018) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2018-0967 · Cited by 42 · see citing works
Use of DOAC Stop for elimination of anticoagulants in the thrombin generation assay
Kopatz, W. F., Brinkman, H. J. M., & Meijers, J. C. M. (2018) · Thrombosis Research · doi:10.1016/j.thromres.2018.08.014 · Cited by 39 · see citing works
Depreter, B., & Devreese, K. M. J. (2016) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2015-0978 · Cited by 34 · see citing works
Self-citationJacquemin, M. G., Toelen, J., Feyen, L., Schoeters, J., Horenbeeck, I. V., Vanlinthout, I., Debasse, M., Vanassche, T., Peerlinck, K., & Verhamme, P. B. (2018) · International Journal of Laboratory Hematology · doi:10.1111/ijlh.12807 · Cited by 33 · see citing works
The Effects of Indirect- and Direct-Acting Anticoagulants on Lupus Anticoagulant Assays
Seheult, J. N., Meyer, M. P., Bontempo, F. A., & Chibisov, I. (2017) · American Journal of Clinical Pathology · doi:10.1093/ajcp/aqx035 · Cited by 31 · see citing works
Prevalence of Heparin in Samples Submitted for Lupus Anticoagulant Testing
Sangle, N. A., Rodgers, G. M., & Smock, K. J. (2011) · Laboratory Hematology · doi:10.1532/lh96.10024 · Cited by 5 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Devreese, K. M. J., de Groot, P. G., de Laat, B., Erkan, D., Favaloro, E. J., Mackie, I., Martinuzzo, M., Ortel, T. L., Pengo, V., Rand, J. H., et al. (2020) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15047 · Cited by 439 · see citing works
Self-citationTripodi, A., Cohen, H., & Devreese, K. M. J. (2020) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.14846 · Cited by 143 · see citing works
Self-citationDevreese, K. M. J., Bertolaccini, M. L., Branch, D. W., de Laat, B., Erkan, D., Favaloro, E. J., Pengo, V., Ortel, T. L., Wahl, D., & Cohen, H. (2025) · Journal of Thrombosis and Haemostasis · doi:10.1016/j.jtha.2024.10.022 · Cited by 92 · see citing works
Self-citationMoore, G. W., Jones, P. O., Platton, S., Hussain, N., White, D., Thomas, W., Rigano, J., Pouplard, C., Gray, E., & Devreese, K. M. J. (2021) · Journal of Thrombosis and Haemostasis · doi:10.1111/jth.15438 · Cited by 47 · see citing works
Self-citationLupus Anticoagulant Detection under the Magnifying Glass
Molinari, A. C., Martini, T., Banov, L., Ierardi, A., Leotta, M., Strangio, A., & Santoro, R. C. (2023) · Journal of Clinical Medicine · doi:10.3390/jcm12206654 · Cited by 9 · see citing works
Analytical dilemmas in lupus anticoagulant detection
Moore, G. W. (2023) · Exploration of Immunology · doi:10.37349/ei.2023.00104 · Cited by 5 · see citing works
Lupus Anticoagulant Testing: Taipan Snake Venom Time with Ecarin Time as Confirmatory Test
Moore, G. W. (2023) · Methods in Molecular Biology · doi:10.1007/978-1-0716-3175-1_16 · Cited by 5 · see citing works
Viewpoint: Lupus anticoagulant detection and interpretation in antiphospholipid syndrome
Efthymiou, M., Bertolaccini, M. L., & Cohen, H. (2024) · Rheumatology · doi:10.1093/rheumatology/kead623 · Cited by 5 · see citing works
- Authors
- De Kesel, P. M. M., & Devreese, K. M. J.
- Affiliations
- Ghent University Hospital (BE)
- Corresponding author
- Devreese, Katrien M.J.
- Venue
- Research and Practice in Thrombosis and Haemostasis
- Publisher
- Elsevier BV
- Published
- 2020-01
- Volume
- 4
- Issue
- 1
- Pages
- 161-168
- ISSN
- 2475-0379 (ISSN-L)
- PMC
- PMC6971310
- OpenAlex
- W2995572514
- Cited by
- 47
- References
- 26
- Citations by year
- 2026: 4 · 2025: 2 · 2024: 8 · 2023: 9 · 2022: 8 · 2021: 7 · 2020: 9
- FWCI
- 4.54
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Research and Practice in Thrombosis and Haemostasis (publishedVersion, cc-by-nc-nd)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- Ghent University Academic Bibliography (Ghent University) (submittedVersion, cc-by-nc-nd)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Rheumatology › Systemic Lupus Erythematosus Research
- Health Sciences › Medicine › Hematology › Blood Coagulation and Thrombosis Mechanisms
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Keywords
- lupus anticoagulant, unfractionated heparin, enoxaparin, danaparoid, heparinoids, activated carbon, false positives, anti-Xa activity, activated partial thromboplastin time, mixing test, anticoagulant interference
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review