Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during Exercise
Abstract
PURPOSE: The purpose of this study was to evaluate the accuracy of the Polar M600 optical heart rate (OHR) sensor compared with ECG heart rate (HR) measurement during various physical activities. METHODS: Thirty-six subjects participated in a continuous 76-min testing session, which included rest, cycling warm-up, cycling intervals, circuit weight training, treadmill intervals, and recovery. HR was measured using a three-lead ECG configuration and a Polar M600 Sport Watch on the left wrist. Statistical analyses included OHR percent accuracy, mean difference, mean absolute error, Bland-Altman plots, and a repeated-measures generalized estimating equation design. OHR percent accuracy was calculated as the percentage of occurrences where OHR measurement was within and including ±5 bpm from the ECG HR value. RESULTS: Of the four exercise phases performed, the highest OHR percent accuracy was found during cycle intervals (91.8%), and the lowest OHR percent accuracy occurred during circuit weight training (34.5%). OHR percent accuracy improved steadily within exercise transitions during cycle intervals to a maximum of 98.5% and during treadmill intervals to a maximum of 89.0%. Lags in HR calculated by the Polar M600 OHR sensor existed in comparison to ECG HR, when exercise intensity changed until steady state occurred. There was a tendency for OHR underestimation during intensity increases and overestimation during intensity decreases. No statistically significant interaction effect with device was found in this sample on the basis of sex, body mass index, V˙O2max, skin type, or wrist size. CONCLUSIONS: The Polar M600 was accurate during periods of steady-state cycling, walking, jogging, and running, but less accurate during some exercise intensity changes, which may be attributed to factors related to total peripheral resistance changes and pulse pressure.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Horton et al assessed the validity of heart rate only using the Polar M600 when compared with a 3-lead electrocardiogram during both aerobic and resistance exercises, and the accuracy of the wearable device was reported to be better during aerobic exercise (92%) than during resistance exercises (35%).
“Horton et al [22] assessed the validity of HR only using the Polar M600 when compared with a 3-lead electrocardiogram (ECG) during both aerobic and resistance exercises. The accuracy of the wearable device was reported to be better during aerobic exercise (92%) as compared with only 35% accurate during the resistance exercises.”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [22]
Presently, there are at least four studies that have investigated the accuracy of wearable devices during resistance exercises and none during high-intensity interval training.
“Presently, there are at least 4 studies [21-24] that have investigated the accuracy of wearable devices during resistance exercises and none during HIIT training.”
From Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study (2018) · cites this paper as [22]
The PPG signal recorded in a real environment often contains movement artifacts and is affected by changing pressure of sensor on skin.
“The PPG signal, when recorded in a real environment, often contains movement artifacts and is also affected by the changing pressure of the sensor on the skin [80].”
From The Current State of Optical Sensors in Medical Wearables (2022) · cites this paper as [80]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661 · see citing works
Gil, E., Orini, M., Bailón, R., Vergara, J. M., Mainardi, L., & Laguna, P. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/9/015 · Cited by 461 · see citing works
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 455 · see citing works
Accuracy of wrist-worn heart rate monitors
Wang, R., Blackburn, G., Desai, M., Phelan, D., Gillinov, L., Houghtaling, P., & Gillinov, M. (2017) · JAMA Cardiology · doi:10.1001/jamacardio.2016.3340 · Cited by 386 · see citing works
Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise
GILLINOV, S., ETIWY, M., WANG, R., BLACKBURN, G., PHELAN, D., GILLINOV, A. M., HOUGHTALING, P., JAVADIKASGARI, H., & DESAI, M. Y. (2017) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001284 · Cited by 345 · see citing works
Stahl, S. E., An, H.-S., Dinkel, D. M., Noble, J. M., & Lee, J.-M. (2016) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2015-000106 · Cited by 276 · see citing works
The advantages of wearable green reflected photoplethysmography
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9506-z · Cited by 209 · see citing works
Evaluation of wearable consumer heart rate monitors based on photopletysmography
Parak, J., & Korhonen, I. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6944419 · Cited by 148 · see citing works
Validation of photoplethysmography as a method to detect heart rate during rest and exercise
Spierer, D. K., Rosen, Z., Litman, L. L., & Fujii, K. (2015) · Journal of Medical Engineering & Technology · doi:10.3109/03091902.2015.1047536 · Cited by 139 · see citing works
Evaluation of the beat-to-beat detection accuracy of PulseOn wearable optical heart rate monitor
Parák, J., Tarniceriu, A., Renevey, P., Bertschi, M., Delgado-Gonzalo, R., & Korhonen, I. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7320273 · Cited by 100 · see citing works
Evaluation of accuracy and reliability of PulseOn optical heart rate monitoring device
Delgado-Gonzalo, R., Parák, J., Tarniceriu, A., Renevey, P., Bertschi, M., & Korhonen, I. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318391 · Cited by 55 · see citing works
Lee, S. M., Shin, H., & Hahm, C. (2016) · International Conference on Advanced Communication Technology · doi:10.1109/icact.2016.7423470 · Cited by 38 · see citing works
Lin, Z., Zhang, J., Chen, Y., & Zhang, Q. (2015) · International Conference on Communications · doi:10.1109/icc.2015.7248369 · Cited by 17 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438 · see citing works
Validity of Wearable Activity Monitors during Cycling and Resistance Exercise
BOUDREAUX, B. D., HEBERT, E. P., HOLLANDER, D. B., WILLIAMS, B. M., CORMIER, C. L., NAQUIN, M. R., GILLAN, W. W., GUSEW, E. E., & KRAEMER, R. R. (2018) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000001471 · Cited by 205 · see citing works
Reddy, R. K., Pooni, R., Zaharieva, D. P., Senf, B., El Youssef, J., Dassau, E., Doyle III, F. J., Clements, M. A., Rickels, M. R., Patton, S. R., et al. (2018) · JMIR mHealth and uHealth · doi:10.2196/10338 · Cited by 157 · see citing works
The Current State of Optical Sensors in Medical Wearables
Vavrinsky, E., Esfahani, N. E., Hausner, M., Kuzma, A., Rezo, V., Donoval, M., & Kosnacova, H. (2022) · Biosensors · doi:10.3390/bios12040217 · Cited by 103 · see citing works
“The PPG signal, when recorded in a real environment, often contains movement artifacts and is also affected by the changing pressure of the sensor on the skin [80].”
BackgroundZhang, Y., Weaver, R. G., Armstrong, B., Burkart, S., Zhang, S., & Beets, M. W. (2020) · Journal of Sports Sciences · doi:10.1080/02640414.2020.1767348 · Cited by 88 · see citing works
Validation of a Photoplethysmographic Heart Rate Monitor: Polar OH1
Hermand, E., Cassirame, J., Ennequin, G., & Hue, O. (2019) · International Journal of Sports Medicine · doi:10.1055/a-0875-4033 · Cited by 59 · see citing works
Skin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches
Ray, I., Liaqat, D., Gabel, M., & de Lara, E. (2021) · 2021 IEEE International Conference on Pervasive Computing and Communications Workshops and other Affiliated Events (PerCom Workshops) · doi:10.1109/percomworkshops51409.2021.9431120 · Cited by 23 · see citing works
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
- Authors
- HORTON, J. F., STERGIOU, P., FUNG, T. S., & KATZ, L.
- Affiliations
- University of Calgary (CA)
- Venue
- Medicine & Science in Sports & Exercise
- Published
- 2017-12
- Volume
- 49
- Issue
- 12
- Pages
- 2600-2607
- ISSN
- 0195-9131 (ISSN-L)
- OpenAlex
- W2740969659
- Cited by
- 79
- References
- 15
- Citations by year
- 2026: 4 · 2025: 6 · 2024: 9 · 2023: 6 · 2022: 11 · 2021: 12 · 2020: 11 · 2019: 15 · 2018: 3 · 2017: 2
- FWCI
- 3.81
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Medicine & Science in Sports & Exercise (publishedVersion)
- PubMed (publishedVersion)
- Funders
- University of Calgary
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Complementary and alternative medicine › Cardiovascular and exercise physiology
- Health Sciences › Medicine › Physiology › Physical Activity and Health
- Keywords
- heart rate accuracy, cycling, treadmill running, circuit weight training, systemic vascular resistance, pulse pressure, wearable heart rate sensors
- MeSH terms
- Adult, Bicycling, Electrocardiography, Exercise Test, Female, Heart Rate, Humans, Male, Middle Aged, Oxygen Consumption, Running, Exercise, Walking, Resistance Training
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review