Development and Characterization of a Cryogenic Tuned Amplifier for Penning Trap Applications
Open accessAbstract
A two-stage tuned amplifier has been developed and characterised for operation at cryogenic temperatures for Penning Trap application. Two pHEMT devices were tested at 300 K, 77 K and 4.2 K for their DC and AC characteristics. The developed amplifier has shown an amplification of 40 dB at a quiescent power consumption of ~ 1 mW at liquid helium temperature. Considering the feeble intensity of the image charge signal from Penning trap, the input impedance of the first stage amplifier is kept high whereas the output impedance of the second stage is kept 50 Ω for impedance matching with the transmission line. The bandwidth was ~ 200 kHz with the centre frequency around 40 MHz to match with the axial frequency of the electrons confined in the Penning trap. The amplifier was tested at 5 T magnetic field and it showed similar performance as in no field condition. The signal of trapped electrons, in a Penning trap at 4.2 K, was detected using this amplifier through the resonance absorption technique, confirming its suitability for the system.
Overview
CoChat Summary
## TL;DR
The paper develops a two-stage cryogenic tuned amplifier for Penning trap readout and shows that it can deliver 40 dB gain at 4.2 K with about 1 mW power consumption. The amplifier also works in a 5 T magnetic field and was able to detect trapped-electron signals in a Penning trap.
## Key findings
- Two GaAs pHEMTs were characterized at 300 K, 77 K, and 4.2 K; the SKY65050 showed higher performance, reaching up to 40 dB gain at 4.2 K, while the CE3512 K2 reached 27.8 dB at 4.2 K.
- The developed amplifier operated at about 1 mW quiescent power consumption and was tuned to a center frequency around 40 MHz with a bandwidth of about 200 kHz.
- At 5 T magnetic field and 4.2 K, the SKY65050-based amplifier showed no major deviation in performance; the second-stage response indicated a resonance frequency of 39.1 MHz.
- When integrated with a five-electrode Penning trap at 4.2 K, the combined detection circuit reached a maximum Q-factor of 3500 at less than 1 mW power consumption, and trapped-electron signals were detected using the amplifier.
## Methods
- A two-stage tune-coupled amplifier was designed for Penning trap readout, with a high-input-impedance first stage and a second stage providing amplification and 50 Ω matching; the PCB was built on a PTFE substrate.
- Two GaAs pHEMTs (CE3512 K2 and SKY65050) were tested for DC and AC characteristics at 300 K, 77 K, and 4.2 K; DC I-V data were acquired using a Visual Studio GUI interfaced with two Keithley SMUs.
- AC characterization used both sinusoidal input and pseudorandom white noise; the selected device was then tested at 5 T and integrated with a five-electrode closed-end Penning trap using the resonance absorption technique.
## Limitations
- The paper states that the datasheets do not provide device-level parameters needed for very low-power, low-frequency cryogenic operation, so some design choices relied on experimental tuning rather than complete vendor data.
- It notes that the contradictory temperature dependence of resonance frequency for the two HEMTs “requires further investigation” (stated in the paper).
- Evident from the text: the study evaluates only two specific HEMTs and reports performance in a single Penning trap setup, so generalization to other devices or trap configurations is not established in the paper.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
0 Valid · 1 Not valid · 11 Not checked · 26 Own results and interpretations
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“whereas III-V direct band gap semiconductor devices like Ga-As based pHEMTs are one of the most preferred choice for developing cryogenic amplifiers [12-28].”
What the sources say
- Not checked — no full text of the sourcecites [12] M. Varonen et al., An MMIC low-noise amplifier design technique. IEEE Trans. Microw. Theory Tech. 64, 826 (2016). https://doi.org/10.1109/TMTT.2016.2521650 source
- Not checked — no full text of the sourcecites [13] M. Varonen et al., A 75–116-GHz LNA with 23-K noise temperature at 108 GHz, IEEE MTT-S Int. Microw. Symp. 1 (2013). https://doi.org/10.1109/MWSYM.2013.6697595 source
- Not checked — no full text of the sourcecites [14] E. Cha et al., Two-finger InP HEMT design for stable cryogenic operation of ultra-low-noise Ka- and Q-Band LNAs. IEEE Trans. Microw. Theory Tech. 65, 5171 (2017). https://doi.org/10.1109/TMTT.2017.2765318 source
- Could not verifycites [15] E. Dogmus et al., InAlGaN/GaN HEMTs at cryogenic temperatures. Electronics 5, 31 (2016). https://doi.org/10.3390/electronics5020031 sourceExperimental — accuracy not yet measured
Our copy of Dogmus et al. (2016) is an experimental study of a sub-10-nm quaternary InAlGaN/AlN/GaN HEMT measured from 4-400 K; the Introduction (p. 1), Results/DC-RF section (pp. 3-4) and Conclusions (p. 5) discuss only GaN-based 2DEG mobility, sheet resistance and f_T/f_max improvements at 77 K. No passage mentions GaAs, pHEMTs, direct band gap materials, or which device family is preferred for cryogenic amplifiers (searched: 'GaAs pHEMT cryogenic low noise amplifier', 'III-V direct band gap preferred choice cryogenic amplifiers', 'advantages of GaN over conventional III-V materials'). The paper's reference list contains no GaAs-amplifier entry either, so our copy neither supports nor contradicts the attributed statement. (LLM, conf=0.68)
- Not checked — no full text of the sourcecites [16] X. de la Broïse et al., Cryogenic ultra-low noise HEMT amplifiers board. Nucl. Instrum. Methods Phys. Res. A 787, 51 (2015). https://doi.org/10.1016/j.nima.2014.11.016 source
- Not checked — no full text of the sourcecites [17] Q. Dong et al., Ultra-low noise high electron mobility transistors for high-impedance and low-frequency deep cryogenic readout electronics. Appl. Phys. Lett. 105, 013504 (2014). https://doi.org/10.1063/1.4887368 source
- Not checked — no full text of the sourcecites [18] X. de la Broïse et al., Cryogenic low noise and low dissipation multiplexing electronics, using HEMT+SiGe ASICs, for the readout of high impedance sensors: new version. Nucl. Instrum. Methods A 787, 64 (2015). https://doi.org/10.1016/j.nima.2014.11.019 source
- Not checked — no full text of the sourcecites [19] T. Arakawa et al., Cryogenic amplifier for shot noise measurement at 20 mK. Appl. Phys. Lett. 103, 172104 (2013). https://doi.org/10.1063/1.4826681 source
- Not checked — no full text of the sourcecites [20] J. Schleeh et al., Cryogenic broadband ultra-low-noise MMIC LNAs for radio astronomy applications. IEEE Trans. Microw. Theory Tech. 61, 871 (2013). https://doi.org/10.1109/TMTT.2012.2235856 source
- Not checked — no full text of the sourcecites [21] P. Neilinger et al., Cryogenic low noise 2.2–3GHz amplifier. Cryogenics 52, 362 (2012). https://doi.org/10.1016/j.cryogenics.2012.02.006 source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“A maximum quality factor of 3500 has been obtained at a power consumption of less than 1 mW.”
- Also stated in: § 6 Conclusion · plain
“The quality factor of the helical resonator coupled with Penning trap and amplifier was obtained up to 3500 at 4.2 K while keeping the total drain current below 1mA.”
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
- Also stated in: § 6 Conclusion · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The developed amplifier has shown an amplification of 40dB at a quiescent power consumption of ~ 1 mW at liquid helium temperature.”
- Also stated in: § 6 Conclusion · plain
“The amplifiers with both the devices consumed ~ 1 mW power to provide sufficient gain for the targeted application.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § 6 Conclusion · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The signal of trapped electrons, in a Penning trap at 4.2K, was detected using this amplifier through the resonance absorption technique, confirming its suitability for the system.”
- Also stated in: § 6 Conclusion · plain
“Signal of the trapped electron cloud at liquid helium temperature was observed using the amplifier.”
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Also stated in: § 6 Conclusion · plain
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“and the AC characteristics displayed a minor deviation.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
24 more claims
Create a free account to see all 38 claims and every source verdict.
- It may state two separate claims as one. 6
- The quote only partly states this claim. 1
- Probably not a substantial claim. 2
Citations
Cites
The AME 2020 atomic mass evaluation (II). Tables, graphs and references*
Wang, M., Huang, W. J., Kondev, F. G., Audi, G., & Naimi, S. (2021) · Chinese Physics C · doi:10.1088/1674-1137/abddaf · Cited by 2,915 · cited 1×
“The mass of the trapped particles can be determined with high precision by accurately measuring the axial frequency of their motions [2–5].”
The AME2016 atomic mass evaluation (II). Tables, graphs and references
Wang, M., Audi, G., Kondev, F. G., Huang, W. J., Naimi, S., & Xu, X. (2017) · Chinese Physics C · doi:10.1088/1674-1137/41/3/030003 · Cited by 2,791
High-accuracy mass spectrometry with stored ions
Blaum, K. (2006) · Physics Reports · doi:10.1016/j.physrep.2005.10.011 · Cited by 685 · cited 2×
“The axial frequency ( ω z ) of the stored electron cloud depends on the applied voltages ( U 0 = V R − V E ) between the ring and end-cap electrodes [1, 9, 42]: where q and m are charge and mass of the electron respectively and d refers to the characteristic dimension of Penning trap.”
The AME 2020 atomic mass evaluation (I). Evaluation of input data, and adjustment procedures*
Huang, W. J., Wang, M., Kondev, F. G., Audi, G., & Naimi, S. (2021) · Chinese Physics C · doi:10.1088/1674-1137/abddb0 · Cited by 436
Feedback Cooling of a One-Electron Oscillator
D’Urso, B., Odom, B., & Gabrielse, G. (2003) · Physical Review Letters · doi:10.1103/physrevlett.90.043001 · Cited by 82
Oukhanski, N., Grajcar, M., Il’ichev, E., & Meyer, H.-G. (2003) · Review of Scientific Instruments · doi:10.1063/1.1532539 · Cited by 52 · cited 1×
“Devices used in the earlier developed amplifiers [29] for such applications completed their lifecycle and are obsolete now.”
Dong, Q., Liang, Y. X., Ferry, D., Cavanna, A., Gennser, U., Couraud, L., & Jin, Y. (2014) · Applied Physics Letters · doi:10.1063/1.4887368 · Cited by 52
Cryogenic Broadband Ultra-Low-Noise MMIC LNAs for Radio Astronomy Applications
Schleeh, J., Wadefalk, N., Nilsson, P.-Å., Starski, J. P., & Grahn, J. (2013) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2012.2235856 · Cited by 51
Operation of SiGe heterojunction bipolar transistors in the liquid-helium temperature regime
Joseph, A. J., Cressler, J. D., & Richey, D. M. (1995) · IEEE Electron Device Letters · doi:10.1109/55.790731 · Cited by 48 · cited 1×
“The Si and Ge devices are not suitable due to heavy doping effects and carrier freeze-out at cryogenic temperature [10, 11], whereas III–”
Ulmer, S., Blaum, K., Kracke, H., Mooser, A., Quint, W., Rodegheri, C. C., & Walz, J. (2013) · Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment · doi:10.1016/j.nima.2012.12.071 · Cited by 48 · cited 1×
“The obtained Q factor of the trap assembly at this operating frequency of 39.1 MHz is relatively higher than the values reported previously [41].”
Influential
Research works printed but not linked: 2
- M. A. Wessels, Progress Toward a Single-Electron Qubit in an Optimized Planar Penning Trap, Ph.D. thesis, Harvard (2019).
- F. Teyssandier, D. Prêle, Commercially Available Capacitors at Cryogenic Temperatures, Ninth Int. Workshop Low Temp. Electronics, WOLTE9, Jun 2010, Guaruja, Brazil. hal-00623399
Metadata
- Authors
- Chaddha, N., Sikdar, A. K., Nandi, J., Vyshnav, C. H., Chatterjee, M., Das, P., & Ray, A.
- Affiliations
- Variable Energy Cyclotron Centre (IN)
- Central University of Haryana (IN)
- Jadavpur University (IN)
- Corresponding author
- Sikdar, A. K.
- Venue
- Journal of Low Temperature Physics
- Published
- 2025-06-05
- Volume
- 220
- Issue
- 3-6
- Pages
- 281-293
- ISSN
- 0022-2291 (ISSN-L)
- 1573-7357
- OpenAlex
- W4411068540
- Cited by
- 0
- References
- 38
- FWCI
- 0.00
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- cc-by-nc-nd
- creativecommons.org/licenses/by-nc-nd/4.0 (tdm)
- creativecommons.org/licenses/by-nc-nd/4.0 (vor)
- Locations
- Journal of Low Temperature Physics (publishedVersion, cc-by-nc-nd)
- Funders
- Science and Engineering Research Board (CRG/2020/003237)
- Variable Energy Cyclotron Centre
- Topics
- Physical Sciences › Physics and Astronomy › Nuclear and High Energy Physics › Particle accelerators and beam dynamics
- Physical Sciences › Physics and Astronomy › Nuclear and High Energy Physics › Particle Detector Development and Performance
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Superconducting and THz Device Technology
- Keywords
- cryogenic amplifiers, Penning trap, pHEMT, two-stage amplifier, trapped electrons, resonant absorption, impedance matching, cryogenic electronics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (qwen3.8-omni-flash).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Development and Characterization of a Cryogenic Tuned Amplifier for Penning Trap Applications. https://app.cochat.ai/papers/10.1007/s10909-025-03309-z
- The paper
- Chaddha, N., Sikdar, A. K., Nandi, J., Vyshnav, C. H., Chatterjee, M., Das, P., & Ray, A. (2025). Development and Characterization of a Cryogenic Tuned Amplifier for Penning Trap Applications. Journal of Low Temperature Physics. https://doi.org/10.1007/s10909-025-03309-z
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.