Superparamagnetic iron oxides as MPI tracers: A primer and review of early applications
Open accessThe DOI matches its Crossref record; 68 of the 68 references it cites were found in a registry or an index.
Abstract
Magnetic particle imaging (MPI) has recently emerged as a non-invasive, whole body imaging technique that detects superparamagnetic iron oxide (SPIO) nanoparticles similar as those used in magnetic resonance imaging (MRI). Based on tracer “hot spot” detection instead of providing contrast on MRI scans, MPI has already proven to be truly quantitative. Without the presence of endogenous background signal, MPI can also be used in certain tissues where the endogenous MRI signal is too low to provide contrast. After an introduction to the history and simplified principles of MPI, this review focuses on early MPI applications including MPI cell tracking, multiplexed MPI, perfusion and tumor MPI, lung MPI, functional MPI, and MPI-guided hyperthermia. While it is too early to tell if MPI will become a mainstay imaging technique with the (theoretical) sensitivity that it promises, and if it can successfully compete with SPIO-based 1H MRI and perfluorocarbon-based 19F MRI, it provides unprecedented opportunities for exploring new nanoparticle-based imaging applications.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Tomographic imaging using the nonlinear response of magnetic particles
Gleich, B., & Weizenecker, J. (2005) · Nature · doi:10.1038/nature03808 · Cited by 2,003 · cited 1×
“years later, in 2005, the scientific world first heard about MPI [2].”
BackgroundIron oxide MR contrast agents for molecular and cellular imaging
Bulte, J. W. M., & Kraitchman, D. L. (2004) · NMR in Biomedicine · doi:10.1002/nbm.924 · Cited by 1,500 · cited 1×
“It was realized that these nanosized magnetic entities change the microenvironment of the strong external magnetic fields, leading to a loss of phase coherence of resonating protons with as result a loss of MRI signal [8].”
BackgroundSelf-citationThree-dimensional real-time in vivo magnetic particle imaging
Weizenecker, J., Gleich, B., Rahmer, J., Dahnke, H., & Borgert, J. (2009) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/5/l01 · Cited by 651 · cited 1×
“With the high temporal resolution of 40 frames per second for the FFP method, flowing blood can be observed in real-time in the heart [40] and the brain [38], and it is possible to obtain blood flow velocity measurements [41].”
MethodDynamic Imaging of Allogeneic Mesenchymal Stem Cells Trafficking to Myocardial Infarction
Kraitchman, D. L., Tatsumi, M., Gilson, W. D., Ishimori, T., Kedziorek, D., Walczak, P., Segars, W. P., H. Chen, H., Fritzges, D., Izbudak, I., et al. (2005) · Circulation · doi:10.1161/circulationaha.105.537480 · Cited by 602 · cited 1×
“Systemically injected SPIO-labeled MSCs initially became entrapped within the lung vasculature, and then redistributed within one day to the liver and spleen (Figure 4)[21], which is a typical distribution pattern previously observed with single photon emission computed tomography (SPECT) of intravenously injected 111…”
BackgroundSelf-citationCheng, Y., Zhao, L., Li, Y., & Xu, T. (2011) · Chemical Society Reviews · doi:10.1039/c0cs00097c · Cited by 515 · cited 1×
“This biological feature has previously been exploited to non-selectively target tumors with SPIO nanoparticles [45] and other nanosized contrast agents [46, 47] by means of the so-called enhanced permeability and retention (EPR) effect, leading to enhanced tumor MR images.”
BackgroundTracking immune cells in vivo using magnetic resonance imaging
Ahrens, E. T., & Bulte, J. W. M. (2013) · Nature Reviews Immunology · doi:10.1038/nri3531 · Cited by 491 · cited 1×
“Another example is imaging of inflammation, following intravenous injection of nanoparticles that accumulate in macrophages at the site of inflammation [11].”
BackgroundSelf-citationFluorine (19F) MRS and MRI in biomedicine
Ruiz‐Cabello, J., Barnett, B. P., Bottomley, P. A., & Bulte, J. W. M. (2011) · NMR in Biomedicine · doi:10.1002/nbm.1570 · Cited by 489 · cited 1×
“While 19F MRI avoids the problems encountered with 1H MRI and has been successfully used for lung ventilation imaging using fluorous gases, it suffers in from low sensitivity [53].”
Self-citationTay, Z. W., Chandrasekharan, P., Chiu-Lam, A., Hensley, D. W., Dhavalikar, R., Zhou, X. Y., Yu, E. Y., Goodwill, P. W., Zheng, B., Rinaldi, C., et al. (2018) · ACS Nano · doi:10.1021/acsnano.8b00893 · Cited by 459 · cited 1×
“To proof this, localization of thermal damage and therapy was validated with luciferase activity and histological assessment [55].”
MethodIn vivo detection of single cells by MRI
Shapiro, E. M., Sharer, K., Skrtic, S., & Koretsky, A. P. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20718 · Cited by 378
Magnetic Particle Imaging: A Novel in Vivo Imaging Platform for Cancer Detection
Yu, E. Y., Bishop, M., Zheng, B., Ferguson, R. M., Khandhar, A. P., Kemp, S. J., Krishnan, K. M., Goodwill, P. W., & Conolly, S. M. (2017) · Nano Letters · doi:10.1021/acs.nanolett.6b04865 · Cited by 366 · cited 1×
“In another study, the tumor could be highlighted with a tumor-to-background ratio of up to 50 at the peak uptake of 6 hours after tracer injection [49].”
BackgroundZheng, B., von See, M. P., Yu, E., Gunel, B., Lu, K., Vazin, T., Schaffer, D. V., Goodwill, P. W., & Conolly, S. M. (2016) · Theranostics · doi:10.7150/thno.13728 · Cited by 322 · cited 1×
“Systemically injected SPIO-labeled MSCs initially became entrapped within the lung vasculature, and then redistributed within one day to the liver and spleen (Figure 4)[21], which is a typical distribution pattern previously observed with single photon emission computed tomography (SPECT) of intravenously injected 111…”
BackgroundMultidimensional X-Space Magnetic Particle Imaging
Goodwill, P. W., & Conolly, S. M. (2011) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2011.2125982 · Cited by 319 · cited 3×
“An alternative method is to apply a “field-free line” (FFL) [3] instead of multiple FFPs to obtain increased sensitivity, but at the cost of temporal resolution.”
MethodInfluentialHeyn, C., Ronald, J. A., Ramadan, S. S., Snir, J. A., Barry, A. M., MacKenzie, L. T., Mikulis, D. J., Palmieri, D., Bronder, J. L., Steeg, P. S., et al. (2006) · Magnetic Resonance in Medicine · doi:10.1002/mrm.21029 · Cited by 291
Zheng, B., Vazin, T., Goodwill, P. W., Conway, A., Verma, A., Ulku Saritas, E., Schaffer, D., & Conolly, S. M. (2015) · Scientific Reports · doi:10.1038/srep14055 · Cited by 274 · cited 1×
“Using MSCs, it was demonstrated that SPIO-labeled NSCs could be detected by MPI in rat brain, and that the particles persisted for several months [20].”
BackgroundSong, G., Chen, M., Zhang, Y., Cui, L., Qu, H., Zheng, X., Wintermark, M., Liu, Z., & Rao, J. (2018) · Nano Letters · doi:10.1021/acs.nanolett.7b03829 · Cited by 256 · cited 3×
“While the early field of MPI cell tracking has not shown this feasibility yet, an in vivo sensitivity of 250 cells within the FOV of an entire mouse has been reported using these MPI-optimized Janus iron oxide nanoparticles [25].”
MethodInfluentialSchleich, N., Po, C., Jacobs, D., Ucakar, B., Gallez, B., Danhier, F., & Préat, V. (2014) · Journal of Controlled Release · doi:10.1016/j.jconrel.2014.07.059 · Cited by 238 · cited 1×
“This biological feature has previously been exploited to non-selectively target tumors with SPIO nanoparticles [45] and other nanosized contrast agents [46, 47] by means of the so-called enhanced permeability and retention (EPR) effect, leading to enhanced tumor MR images.”
BackgroundMagnetic Particle Imaging for Real-Time Perfusion Imaging in Acute Stroke
Ludewig, P., Gdaniec, N., Sedlacik, J., Forkert, N. D., Szwargulski, P., Graeser, M., Adam, G., Kaul, M. G., Krishnan, K. M., Ferguson, R. M., et al. (2017) · ACS Nano · doi:10.1021/acsnano.7b05784 · Cited by 224 · cited 2×
“MPI cerebral perfusion scans have been performed in stroke [38] and traumatic brain injury [39], and has shown its utility for unambiguous detection of neuropathological changes (Figure 5).”
BackgroundMethodMagnetic particle imaging with tailored iron oxide nanoparticle tracers
Ferguson, R. M., Khandhar, A. P., Kemp, S. J., Arami, H., Saritas, E. U., Croft, L. R., Konkle, J., Goodwill, P. W., Halkola, A., Rahmer, J., et al. (2015) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2014.2375065 · Cited by 222
CHIN, B. B., NAKAMOTO, Y., BULTE, J. W., PITTENGER, M. F., WAHL, R., & KRAITCHMAN, D. L. (2003) · Nuclear Medicine Communications · doi:10.1097/00006231-200311000-00005 · Cited by 210 · cited 1×
“Systemically injected SPIO-labeled MSCs initially became entrapped within the lung vasculature, and then redistributed within one day to the liver and spleen (Figure 4)[21], which is a typical distribution pattern previously observed with single photon emission computed tomography (SPECT) of intravenously injected 111…”
BackgroundSelf-citationIntracellular labeling of T‐cells with superparamagnetic contrast agents
Yeh, T., Zhang, W., Ildstad, S. T., & Ho, C. (1993) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910300513 · Cited by 198
Model-Based Reconstruction for Magnetic Particle Imaging
Knopp, T., Sattel, T. F., Biederer, S., Rahmer, J., Weizenecker, J., Gleich, B., Borgert, J., & Buzug, T. M. (2010) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2009.2021612 · Cited by 184
Ferromagnetic particles as contrast agents for magnetic resonance imaging of liver and spleen
Dias, M. H. M., & Lauterbur, P. C. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030218 · Cited by 180
Graeser, M., Knopp, T., Szwargulski, P., Friedrich, T., von Gladiss, A., Kaul, M. G., Krishnan, K. M., Ittrich, H., Adam, G., & Buzug, T. M. (2017) · Scientific Reports · doi:10.1038/s41598-017-06992-5 · Cited by 176
Bulte, J. W. M., Walczak, P., Janowski, M., Krishnan, K. M., Arami, H., Halkola, A., Gleich, B., & Rahmer, J. (2015) · Tomography · doi:10.18383/j.tom.2015.00172 · Cited by 173 · cited 3×
“Using the initial Philips pre-clinical MPI prototype scanner at Philips Research, Hamburg, Germany, cells could be readily detected by MPI at a detection threshold of about 5×104 cells, with MPI/MRI overlays showing a good agreement between the hypointense MRI areas and MPI hot spots (Figure 3) [19].”
InfluentialSelf-citationFirst experimental evidence of the feasibility of multi-color magnetic particle imaging
Rahmer, J., Halkola, A., Gleich, B., Schmale, I., & Borgert, J. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/5/1775 · Cited by 170
Ferromagnetic contrast agents: A new approach
Renshaw, P. F., Owen, C. S., McLaughlin, A. C., Frey, T. G., & Leigh, J. S. (1986) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910030205 · Cited by 162
Yu, E. Y., Chandrasekharan, P., Berzon, R., Tay, Z. W., Zhou, X. Y., Khandhar, A. P., Ferguson, R. M., Kemp, S. J., Zheng, B., Goodwill, P. W., et al. (2017) · ACS Nano · doi:10.1021/acsnano.7b04844 · Cited by 161 · cited 1×
“Similarly, MPI can be properly used in abdominal studies [54], where the colon and other parts of the GI system often contain air-filled pockets or food packages that show up hypointense on MRI.”
BackgroundTomographic magnetic particle imaging of cancer targeted nanoparticles
Arami, H., Teeman, E., Troksa, A., Bradshaw, H., Saatchi, K., Tomitaka, A., Gambhir, S. S., Häfeli, U. O., Liggitt, D., & Krishnan, K. M. (2017) · Nanoscale · doi:10.1039/c7nr05502a · Cited by 158 · cited 1×
“This phenomenon has also been shown to apply MPI, where subcutaneous tumors were selectively highlighted [48].”
BackgroundMonodisperse magnetite nanoparticle tracers for in vivo magnetic particle imaging
Khandhar, A. P., Ferguson, R. M., Arami, H., & Krishnan, K. M. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.01.087 · Cited by 152 · cited 1×
“e, the MRI contrast agents Resovist and Feridex, as well as MPI-optimized particles [18].”
BackgroundCombining magnetic particle imaging and magnetic fluid hyperthermia in a theranostic platform
Hensley, D., Tay, Z. W., Dhavalikar, R., Zheng, B., Goodwill, P., Rinaldi, C., & Conolly, S. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa5601 · Cited by 147
Multicore Magnetic Nanoparticles for Magnetic Particle Imaging
Eberbeck, D., Dennis, C. L., Huls, N. A. F., Krycka, K. L., Grüttner, C., & Westphal, F. (2012) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2012.2226438 · Cited by 144
First in vivo magnetic particle imaging of lung perfusion in rats
Zhou, X. Y., Jeffris, K. E., Yu, E. Y., Zheng, B., Goodwill, P. W., Nahid, P., & Conolly, S. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa616c · Cited by 139 · cited 1×
“This enables new MPI applications for the lung, including the study of (impaired) lung perfusion [51], or inhalation of aerosolized SPIO formulations [52], which was suggested to possibly serve as an alternative mucociliary clearance test.”
BackgroundClinical Tracking of Cell Transfer and Cell Transplantation: Trials and Tribulations
Bulte, J. W. M., & Daldrup-Link, H. E. (2018) · Radiology · doi:10.1148/radiol.2018180449 · Cited by 137
Self-citationBulte, J. W. M., Ma, L. D., Magin, R. L., Kamman, R. L., Hulstaert, C. E., Go, K. G., THE, T. H., & de Leij, L. (1993) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910290108 · Cited by 135
Self-citationTay, Z. W., Chandrasekharan, P., Zhou, X. Y., Yu, E. Y., Zheng, B., & Conolly, S. (2018) · Theranostics · doi:10.7150/thno.26608 · Cited by 131 · cited 1×
“This enables new MPI applications for the lung, including the study of (impaired) lung perfusion [51], or inhalation of aerosolized SPIO formulations [52], which was suggested to possibly serve as an alternative mucociliary clearance test.”
BackgroundLiu, G., Moake, M. M., Har‐el, Y., Long, C. M., Chan, K. W. Y., Cardona, A., Jamil, M., Walczak, P., Gilad, A. A., Sgouros, G., et al. (2011) · Magnetic Resonance in Medicine · doi:10.1002/mrm.23100 · Cited by 120
Self-citationEdge Preserving and Noise Reducing Reconstruction for Magnetic Particle Imaging
Storath, M., Brandt, C., Hofmann, M., Knopp, T., Salamon, J., Weber, A., & Weinmann, A. (2016) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2016.2593954 · Cited by 119
System Characterization of a Highly Integrated Preclinical Hybrid MPI-MRI Scanner
Franke, J., Heinen, U., Lehr, H., Weber, A., Jaspard, F., Ruhm, W., Heidenreich, M., & Schulz, V. (2016) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2016.2542041 · Cited by 117 · cited 1×
“Indeed, industrial efforts are underway to build an MPI/MRI scanner [58].”
BackgroundFirst in vivo traumatic brain injury imaging via magnetic particle imaging
Orendorff, R. D., Peck, A. J., Zheng, B., Shirazi, S. N., Ferguson, R. M., Khandhar, A. P., Kemp, S. J., Goodwill, P. W., Krishnan, K. M., Brooks, G. A., et al. (2017) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aa52ad · Cited by 116 · cited 1×
“MPI cerebral perfusion scans have been performed in stroke [38] and traumatic brain injury [39], and has shown its utility for unambiguous detection of neuropathological changes (Figure 5).”
BackgroundFrequency dependence of MR relaxation times II. Iron oxides
Bulte, J. W. M., Vymazal, J., Brooks, R. A., Pierpaoli, C., & Frank, J. A. (1993) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1880030414 · Cited by 113 · cited 1×
“Although SPIO MR contrast agents are the most effective MRI contrast agents known today, with field-dependent T2 relaxivities well exceeding >100 mM−1s−1 [12], they have several drawbacks.”
BackgroundSelf-citationRahmer, J., Antonelli, A., Sfara, C., Tiemann, B., Gleich, B., Magnani, M., Weizenecker, J., & Borgert, J. (2013) · Physics in Medicine and Biology · doi:10.1088/0031-9155/58/12/3965 · Cited by 111
Ultra‐small superparamagnetic iron oxides for metastatic lymph node detection: back on the block
Fortuin, A. S., Brüggemann, R. J. M., van der Linden, J., Panfilov, I., Israël, B., Scheenen, T. W. J., & Barentsz, J. O. (2017) · Wiley Interdisciplinary Reviews Nanomedicine and Nanobiotechnology · doi:10.1002/wnan.1471 · Cited by 110 · cited 1×
“Since then, their primary clinical use has been as negative contrast agents for liver and lymph node imaging [9], where normal tissue contains an abundance of phagocytic cell types that take up Bulte Page 2”
BackgroundDextran-magnetite: A new relaxation reagent and its application to T2 measurements in gel systems
Ohgushi, M., Nagayama, K., & Wada, A. (1978) · Journal of Magnetic Resonance (1969) · doi:10.1016/0022-2364(78)90018-5 · Cited by 102
Rahmer, J., Wirtz, D., Bontus, C., Borgert, J., & Gleich, B. (2017) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2679099 · Cited by 101
Development of a Trimodal Contrast Agent for Acoustic and Magnetic Particle Imaging of Stem Cells
Lemaster, J. E., Chen, F., Kim, T., Hariri, A., & Jokerst, J. V. (2018) · ACS Applied Nano Materials · doi:10.1021/acsanm.8b00063 · Cited by 90 · cited 1×
“Multi-modal MPI cell tracking agents are now also being developed, although at the cost of a lower sensitivity in the order of 1 million cells [29].”
BackgroundDesign analysis of an MPI human functional brain scanner
Mason, E. E., Cooley, C. Z., Cauley, S., Griswold, M. A., Conolly, S., & Wald, L. L. (2016) · International journal on magnetic particle imaging · doi:10.18416/ijmpi.2017.1703008 · Cited by 80 · cited 2×
“are developing a human functional MPI scanner [59], and Magnetic Insight has also launched an initiative to build one.”
BackgroundMulti-color magnetic particle imaging for cardiovascular interventions
Haegele, J., Vaalma, S., Panagiotopoulos, N., Barkhausen, J., VOGT, F. M., Borgert, J., & Rahmer, J. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/16/n415 · Cited by 73
Online reconstruction of 3D magnetic particle imaging data
Knopp, T., & Hofmann, M. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/11/n257 · Cited by 72
In vivo MRI visualization of different cell populations labeled with PARACEST agents
Ferrauto, G., Castelli, D. D., Terreno, E., & Aime, S. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24411 · Cited by 72 · cited 1×
“This approach can be compared to multi-color paraCEST (chemical exchange saturation transfer) MRI using two different paramagnetic agents [35] or diaCEST MRI using two different diaCEST agents [36], but there are no potentially toxic metals involved (as in the case of paraCEST MRI) and there is no background contrast…”
MethodCooley, C. Z., Mandeville, J. B., Mason, E. E., Mandeville, E. T., & Wald, L. L. (2018) · NeuroImage · doi:10.1016/j.neuroimage.2018.05.004 · Cited by 71
Wang, P., Goodwill, P. W., Pandit, P., Gaudet, J. M., Ross, A., Wang, J., Yu, E. Y., Hensley, D. W., Doyle, T. C., Contag, C. H., et al. (2018) · Quantitative Imaging in Medicine and Surgery · doi:10.21037/qims.2018.02.06 · Cited by 65
Perspectives on clinical magnetic particle imaging
Borgert, J., Schmidt, J. D., Schmale, I., Bontus, C., Gleich, B., David, B. R., Weizenecker, J., Jockram, J., Lauruschkat, C., Mende, O., et al. (2013) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmt-2012-0064 · Cited by 60
Weber, A., Werner, F., Weizenecker, J., Buzug, T. M., & Knopp, T. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/2/475 · Cited by 60
Magnetic particle imaging for in vivo blood flow velocity measurements in mice
Kaul, M. G., Salamon, J., Knopp, T., Ittrich, H., Adam, G., Weller, H., & Jung, C. (2018) · Physics in Medicine and Biology · doi:10.1088/1361-6560/aab136 · Cited by 57 · cited 1×
“With the high temporal resolution of 40 frames per second for the FFP method, flowing blood can be observed in real-time in the heart [40] and the brain [38], and it is possible to obtain blood flow velocity measurements [41].”
MethodFerromagnetic particles as contrast agent in T2 NMR imaging
Olsson, M. B. E., Persson, B. R. B., Salford, L. G., & Schröder, U. (1986) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(86)90054-8 · Cited by 54
Khandhar, A. P., Ferguson, R. M., Arami, H., Kemp, S. J., & Krishnan, K. M. (2015) · IEEE Transactions on Magnetics · doi:10.1109/tmag.2014.2321096 · Cited by 52
Joint Reconstruction of Tracer Distribution and Background in Magnetic Particle Imaging
Straub, M., & Schulz, V. (2018) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2777878 · Cited by 49
Dextran‐magnetite particles: Contrast‐enhanced MRI of blood–brain barrier disruption in a rat model
Bulte, J. W. M., Jonge, M. W. A., Kamman, R. L., Go, K. G., Zuiderveen, F., Blaauw, B., Oosterbaan, J. A., Hauw, T., & de Leij, L. (1992) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910230203 · Cited by 48
Self-citationMPI cell tracking: what can we learn from MRI?
Bulte, J. W. M., Walczak, P., Gleich, B., Weizenecker, J., Markov, D. E., Aerts, H. C. J., Boeve, H., Borgert, J., & Kuhn, M. H. (2011) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.879844 · Cited by 43 · cited 1×
“MPI cell tracking—One of the earliest SPIO MRI applications has been in cell tracking [13–15], and the same can be said for MPI [16, 17].”
BackgroundSelf-citationRed blood cells as carriers in magnetic particle imaging
Antonelli, A., Sfara, C., Rahmer, J., Gleich, B., Borgert, J., & Magnani, M. (2013) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmt-2012-0065 · Cited by 43
Multi-Channel Acquisition for Isotropic Resolution in Magnetic Particle Imaging
Lu, K., Goodwill, P. W., Zheng, B., & Conolly, S. (2017) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2787500 · Cited by 38
Top, C. B., İlbey, S., & Güven, H. E. (2017) · Medical Physics · doi:10.1002/mp.12604 · Cited by 37 · cited 1×
“Engineers are thrilled to build a clinical machine [59, 60].”
BackgroundTrillaud, H., Degrèze, P., Mesplède, Y., et al. (1995) · Academic Radiology · doi:10.1016/s1076-6332(05)80187-4 · Cited by 21 · cited 1×
“Perfusion and tumor MPI—(U)SPIO particles have been used with MRI early on for perfusion imaging [37] and the detection of leaky endothelium in the case of bloodbrain barrier disruption [13].”
MethodFast multiresolution data acquisition for magnetic particle imaging using adaptive feature detection
Gdaniec, N., Szwargulski, P., & Knopp, T. (2017) · Medical Physics · doi:10.1002/mp.12628 · Cited by 20
Kuboyabu, T. (2016) · International Journal of Nanomedicine and Nanosurgery · doi:10.16966/2470-3206.114 · Cited by 19
DEVELOPING CELLULAR MPI: INITIAL EXPERIENCE
Bulte, J. W. M., Walczak, P., Bernard, S., Gleich, B., Weizenecker, J., Borgert, J., AERTS, H., & Boeve, H. (2010) · doi:10.1142/9789814324687_0028 · Cited by 17 · cited 1×
“MPI cell tracking—One of the earliest SPIO MRI applications has been in cell tracking [13–15], and the same can be said for MPI [16, 17].”
BackgroundSelf-citationTAKEUCHI, Y., Suzuki, H., Sasahara, H., Ueda, J., Yabata, I., ITAGAKI, K., Saito, S., & Murase, K. (2014) · Advanced Biomedical Engineering · doi:10.14326/abe.3.37 · Cited by 14
Simulational validation of color magnetic particle imaging (cMPI)
Han, S., Cho, E., Lee, D.-K., Cho, G., Kim, Y. R., & Cho, H. (2014) · Physics in Medicine and Biology · doi:10.1088/0031-9155/59/21/6521 · Cited by 10
Cited by
Recent progress of magnetic nanoparticles in biomedical applications: A review
Anik, M. I., Hossain, M. K., Hossain, I., Mahfuz, A. M. U. B., Rahman, M. T., & Ahmed, I. (2021) · Nano Select · doi:10.1002/nano.202000162 · Cited by 300
Magnetic Nanostructures: Rational Design and Fabrication Strategies toward Diverse Applications
Wang, S., Xu, J., Li, W., Sun, S., Gao, S., & Hou, Y. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00370 · Cited by 129
Metadata
- Authors
- Bulte, J. W. M.
- Venue
- Advanced Drug Delivery Reviews
- Publisher
- Elsevier BV
- Published
- 2019-01
- Volume
- 138
- Pages
- 293-301
- ISSN
- 0169-409X (ISSN-L)
- OpenAlex
- W2903992024
- Cited by
- 202
- References
- 68
- FWCI
- 1.62
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.