A nonlinear QCT-based finite element model validation study for the human femur tested in two configurations in vitro
The DOI matches its Crossref record; 72 of the 72 references it cites were found in a registry or an index.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
An estimate of the worldwide prevalence and disability associated with osteoporotic fractures
Johnell, O., & Kanis, J. A. (2006) · Osteoporosis International · doi:10.1007/s00198-006-0172-4 · Cited by 4,704 · cited 1×
“Osteoporosis is a common skeletal disease, which increases lifetime fracture risk [1].”
BackgroundTests for comparing elements of a correlation matrix
Steiger, J. H. (1980) · Psychological Bulletin · doi:10.1037/0033-2909.87.2.245 · Cited by 4,621
Epidemiology and outcomes of osteoporotic fractures
Cummings, S. R., & Melton, L. J. (2002) · The Lancet · doi:10.1016/s0140-6736(02)08657-9 · Cited by 3,821 · cited 1×
“In osteoporotic individuals, femoral fracture is a major clinical problemwith high morbidity andmortality [2,3].”
BackgroundASSESSMENT OF FRACTURE RISK AND ITS APPLICATION TO SCREENING FOR POSTMENOPAUSAL OSTEOPOROSIS
Alexeeva, L. P., Burkhardt, P., Christiansen, C., Cooper, C. C., Delmas, P. D., Johnell, O., Johnston, C. C., Kanis, J. A., Lips, P. T. A. M., Melton, L. J., et al. (1994) · Cited by 2,655
Picture Thresholding Using an Iterative Selection Method
(1978) · IEEE Transactions on Systems, Man, and Cybernetics · doi:10.1109/tsmc.1978.4310039 · Cited by 2,262 · cited 2×
“A Gaussian filter (σ=1.2, radius equal to two voxels) was applied to the μCT images to remove high frequency noise and an optimal single level threshold was computed using the iterative selection method of Ridler and Calvard [51] to segment the images.”
MethodInfluentialComparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue
Bayraktar, H. H., Morgan, E. F., Niebur, G. L., Morris, G. E., Wong, E. K., & Keaveny, T. M. (2004) · Journal of Biomechanics · doi:10.1016/s0021-9290(03)00257-4 · Cited by 1,142 · cited 3×
“Therefore, a nonlinear scalar function (tissue function TF, Appendix A) for both compression and tension was defined to provide, once extrapolated for poreless bone (BV/TV=ρ=1), an elastic modulus equal to 24 GPa (Emax), a compression ultimate stress equal to 266 MPa and a tension ultimate stress equal to 200 MPa.”
BackgroundMethodBone strength and its determinants
Ammann, P., & Rizzoli, R. (2003) · Osteoporosis International · doi:10.1007/s00198-002-1345-4 · Cited by 761 · cited 1×
“the whole variance in the mechanical properties [23] and in particular its predictive ability for femoral strength has shown a wide range of variability (R(2)=0.”
BackgroundPrediction of femoral fracture load using automated finite element modeling
Keyak, J. H., Rossi, S. A., Jones, K. A., & Skinner, H. B. (1997) · Journal of Biomechanics · doi:10.1016/s0021-9290(97)00123-1 · Cited by 682 · cited 2×
“Due to the complexity of the experimental and modelling procedures, to the authors’ knowledge, only a few studies have investigated both loading conditions with the same model (17 pairs in [43] and one pair in [44,45]) and no one has investigated all above mentioned structural and local predictions of the same model.”
BackgroundResultParkkari, J., Kannus, P., Palvanen, M., Natri, A., Vainio, J., Aho, H., Vuori, I., & Järvinen, M. (1999) · Calcified Tissue International · doi:10.1007/s002239900679 · Cited by 563 · cited 2×
“In the first case, a fall scenario is important because femoral fractures are often associated with an impact load due to a sideways fall [5-8].”
BackgroundQuantitative computed tomography
Adams, J. E. (2009) · European Journal of Radiology · doi:10.1016/j.ejrad.2009.04.074 · Cited by 458 · cited 3×
“For its higher precision in measuring BMD compared to DXA [16,17], QCT has been recently used in clinical trials to estimate the effect of drug treatments on the femoral ultimate load [18–20] and to discriminate between patients who did or did not undergo a hip fracture [21,22].”
MethodFemoral strength is better predicted by finite element models than QCT and DXA
Cody, D. D., Gross, G. J., J. Hou, F., Spencer, H. J., Goldstein, S. A., & P. Fyhrie, D. (1999) · Journal of Biomechanics · doi:10.1016/s0021-9290(99)00099-8 · Cited by 440 · cited 1×
“These results are in line with previous validation of nonlinear FE for either STANCE [41,43] or SIDE [42,43].”
ResultSchileo, E., Dall’Ara, E., Taddei, F., Malandrino, A., Schotkamp, T., Baleani, M., & Viceconti, M. (2008) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2008.05.017 · Cited by 372 · cited 2×
“99, [36-39]), to qualitatively predict the failure location [40] and to predict the femoral ultimate load better than BMD [41-43].”
BackgroundSelf-citationGreenspan, S. L., Myers, E. R., Kiel, D. P., Parker, R. A., Hayes, W. C., & Resnick, N. M. (1998) · The American Journal of Medicine · doi:10.1016/s0002-9343(98)00115-6 · Cited by 349 · cited 2×
“In the first case, a fall scenario is important because femoral fractures are often associated with an impact load due to a sideways fall [5-8].”
BackgroundFabric and elastic principal directions of cancellous bone are closely related
Odgaard, A., Kabel, J., van Rietbergen, B., Dalstra, M., & Huiskes, R. (1997) · Journal of Biomechanics · doi:10.1016/s0021-9290(96)00177-7 · Cited by 338 · cited 1×
“In fact, it is well accepted that bone shows orthotropic material properties with plane of mechanical symmetry being coincident with the planes of morphological symmetry [69].”
BackgroundHip Fractures and the Contribution of Cortical Versus Trabecular Bone to Femoral Neck Strength
Holzer, G., von Skrbensky, G., Holzer, L. A., & Pichl, W. (2009) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.081108 · Cited by 319 · cited 1×
“mechanics in one-legged stance is necessary to study the incidence of spontaneous fractures [9,10], the behavior of the hip during daily activities [11], the stress redistribution after hip surgery [12], the load sharing between cortical and trabecular bone [13], and the effect of pathologies like bone metastasis on b…”
BackgroundMechanics of Biological Tissue
(2006) · Springer eBooks · doi:10.1007/3-540-31184-x · Cited by 318
Alffram, P.-A. (1964) · Acta Orthopaedica Scandinavica · doi:10.3109/ort.1964.35.suppl-65.01 · Cited by 317 · cited 1×
“The setup was able to generate fractures that are usually recorded during clinical practice (sub-capital, neck and trochanteric fractures) [65,66].”
BackgroundImproved prediction of proximal femoral fracture load using nonlinear finite element models
Keyak, J. H. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00045-5 · Cited by 314 · cited 2×
“If applied to the proximal femur [26-28], QCT-based finite element (FE) models can improve the prediction of mechanical properties by combining information about BMD distribution, femoral geometry, femoral size, bone material properties and loading configuration.”
MethodFracture Prediction for the Proximal Femur Using Finite Element Models: Part I—Linear Analysis
Lotz, J. C., Cheal, E. J., & Hayes, W. C. (1991) · Journal of Biomechanical Engineering · doi:10.1115/1.2895412 · Cited by 309 · cited 1×
“Due to the complexity of the experimental and modelling procedures, to the authors’ knowledge, only a few studies have investigated both loading conditions with the same model (17 pairs in [43] and one pair in [44,45]) and no one has investigated all above mentioned structural and local predictions of the same model.”
BackgroundFinite Element Analysis of the Proximal Femur and Hip Fracture Risk in Older Men
Orwoll, E. S., Marshall, L. M., Nielson, C. M., Cummings, S. R., Lapidus, J., Cauley, J. A., Ensrud, K., Lane, N., Hoffmann, P. R., Kopperdahl, D. L., et al. (2009) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.081201 · Cited by 265 · cited 1×
“Homogenized continuum level voxel based FE models (hvFE, where coarsened voxels are directly converted to hexahedron elements) have been recently used in clinical studies to evaluate the effect of drug treatments compared to baseline femoral ultimate load [29,30], to investigate the effect of gender and age on femoral…”
MethodKeyak, J. H., & Falkinstein, Y. (2003) · Medical Engineering & Physics · doi:10.1016/s1350-4533(03)00081-x · Cited by 259 · cited 1×
“) and hard tissues (disarticulation from the pelvis) which were shown to affect (5–13%) the FE prediction of failure load in an in vitro study based on two subjects [68].”
BackgroundReduction in proximal femoral strength due to long-duration spaceflight
Keyak, J. H., Koyama, A. K., LeBlanc, A., Lu, Y., & Lang, T. F. (2009) · Bone · doi:10.1016/j.bone.2008.11.014 · Cited by 236 · cited 1×
“Homogenized continuum level voxel based FE models (hvFE, where coarsened voxels are directly converted to hexahedron elements) have been recently used in clinical studies to evaluate the effect of drug treatments compared to baseline femoral ultimate load [29,30], to investigate the effect of gender and age on femoral…”
MethodDragomir-Daescu, D., Op Den Buijs, J., McEligot, S., Dai, Y., Entwistle, R. C., Salas, C., Melton, L. J., Bennet, K. E., Khosla, S., & Amin, S. (2011) · Annals of Biomedical Engineering · doi:10.1007/s10439-010-0196-y · Cited by 232 · cited 2×
“parameters like yield strain [42] and Young'smodulus-ash density relationship [46] were defined from a training set to obtain more realistic results.”
MethodResultKeaveny, T. M., Hoffmann, P. F., Singh, M., Palermo, L., Bilezikian, J. P., Greenspan, S. L., & Black, D. M. (2008) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.080805 · Cited by 218 · cited 1×
“Homogenized continuum level voxel based FE models (hvFE, where coarsened voxels are directly converted to hexahedron elements) have been recently used in clinical studies to evaluate the effect of drug treatments compared to baseline femoral ultimate load [29,30], to investigate the effect of gender and age on femoral…”
MethodA comparative study on different methods of automatic mesh generation of human femurs
Viceconti, M., Bellingeri, L., Cristofolini, L., & Toni, A. (1998) · Medical Engineering & Physics · doi:10.1016/s1350-4533(97)00049-0 · Cited by 217 · cited 1×
“Inaccuracies in failure location prediction are probably enhanced by the large hexahedron elements used in this study, which do not represent accurately the external smooth surface of the femur and introduce therefore geometrical artifacts [67].”
MethodBlack, D. M., Bouxsein, M. L., Marshall, L. M., Cummings, S. R., Lang, T. F., Cauley, J. A., Ensrud, K. E., Nielson, C. M., & Orwoll, E. S. (2008) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.080316 · Cited by 209 · cited 1×
“For its higher precision in measuring BMD compared to DXA [16,17], QCT has been recently used in clinical trials to estimate the effect of drug treatments on the femoral ultimate load [18–20] and to discriminate between patients who did or did not undergo a hip fracture [21,22].”
MethodPrediction of fracture location in the proximal femur using finite element models
Keyak, J. H., Rossi, S. A., Jones, K. A., Les, C. M., & Skinner, H. B. (2001) · Medical Engineering & Physics · doi:10.1016/s1350-4533(01)00094-7 · Cited by 199 · cited 2×
“99, [36–39]), to qualitatively predict the failure location [40] and to predict the femoral ultimate load better than BMD [41–43].”
BackgroundResult(2006) · Technical Report Series 936 · doi:10.2471/trs.05.931 · Cited by 184
Compressive behaviour of child and adult cortical bone
Öhman, C., Baleani, M., Pani, C., Taddei, F., Alberghini, M., Viceconti, M., & Manfrini, M. (2011) · Bone · doi:10.1016/j.bone.2011.06.035 · Cited by 181 · cited 3×
“Therefore, a nonlinear scalar function (tissue function TF, Appendix A) for both compression and tension was defined to provide, once extrapolated for poreless bone (BV/TV=ρ=1), an elastic modulus equal to 24 GPa (Emax), a compression ultimate stress equal to 266 MPa and a tension ultimate stress equal to 200 MPa.”
BackgroundMethodDall’Ara, E., Pahr, D., Varga, P., Kainberger, F., & Zysset, P. (2012) · Osteoporosis International · doi:10.1007/s00198-011-1568-3 · Cited by 181 · cited 1×
“The external bone contour was found for each image with a filling out algorithm [52] as already described in a previous study for the human vertebra [53].”
MethodSelf-citationMulti-axial mechanical properties of human trabecular bone
Rincón-Kohli, L., & Zysset, P. K. (2009) · Biomechanics and Modeling in Mechanobiology · doi:10.1007/s10237-008-0128-z · Cited by 177 · cited 6×
“Contents lists available at SciVerse ScienceDirect Bone j ourna l homepage: www.e lsev ie r .com/ locate /bone Original Full Length Article A nonlinear QCT-based finite element model validation study for the human femur tested in two configurations in vitro E. Dall'Ara a,⁎, B. Luisier a, R. Schmidt b, F. Kainberger c,…”
MethodInfluentialSelf-citationAge-dependence of femoral strength in white women and men
Keaveny, T. M., Kopperdahl, D. L., Melton, L. J., Hoffmann, P. F., Amin, S., Riggs, B. L., & Khosla, S. (2010) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.091033 · Cited by 171 · cited 1×
“Homogenized continuum level voxel based FE models (hvFE, where coarsened voxels are directly converted to hexahedron elements) have been recently used in clinical studies to evaluate the effect of drug treatments compared to baseline femoral ultimate load [29,30], to investigate the effect of gender and age on femoral…”
MethodCharacteristics of Falls and Risk of Hip Fracture in Elderly Men
Schwartz, A. V., Kelsey, J. L., Sidney, S., & Grisso, J. A. (1998) · Osteoporosis International · doi:10.1007/s001980050060 · Cited by 171 · cited 2×
“[8] A.”
BackgroundYosibash, Z., Trabelsi, N., & Milgrom, C. (2007) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2007.06.017 · Cited by 170 · cited 2×
“99, [36-39]), to qualitatively predict the failure location [40] and to predict the femoral ultimate load better than BMD [41-43].”
BackgroundKeyak, J. H., Sigurdsson, S., Karlsdottir, G., Oskarsdottir, D., Sigmarsdottir, A., Zhao, S., Kornak, J., Harris, T. B., Sigurdsson, G., Jonsson, B. Y., et al. (2011) · Bone · doi:10.1016/j.bone.2011.03.682 · Cited by 165 · cited 1×
“Homogenized continuum level voxel based FE models (hvFE, where coarsened voxels are directly converted to hexahedron elements) have been recently used in clinical studies to evaluate the effect of drug treatments compared to baseline femoral ultimate load [29,30], to investigate the effect of gender and age on femoral…”
MethodThe distribution of cartilage thickness within the joints of the lower limb of elderly individuals
ADAM, C., Eckstein, F., Milz, S., & Putz, R. V. (1998) · Journal of Anatomy · doi:10.1046/j.1469-7580.1998.19320203.x · Cited by 153 · cited 1×
“In fact, the cartilage thickness and its intra-subject variation are higher in the cranial side of the femoral head, which was loaded in STANCE configuration, compared to the medial-caudal region of the femoral head, which was loaded in the SIDE configuration [63].”
MethodViceconti, M., Davinelli, M., Taddei, F., & Cappello, A. (2004) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2003.12.030 · Cited by 151
Bouxsein, M., Coan, B. S., & Lee, S. C. (1999) · Bone · doi:10.1016/s8756-3282(99)00093-9 · Cited by 149
Effect of force direction on femoral fracture load for two types of loading conditions
Keyak, J. H., Skinner, H. B., & Fleming, J. A. (2001) · Journal of Orthopaedic Research · doi:10.1016/s0736-0266(00)00046-2 · Cited by 144 · cited 1×
“This SIDE configuration was chosen because it was found to be most critical for the femoral strength [49].”
MethodA three-dimensional elastic plastic damage constitutive law for bone tissue
Garcia, D., Zysset, P. K., Charlebois, M., & Curnier, A. (2008) · Biomechanics and Modeling in Mechanobiology · doi:10.1007/s10237-008-0125-2 · Cited by 137 · cited 10×
“[54] (for a more detailed formulation please see Appendix A).”
BackgroundMethodInfluentialSelf-citationKoivumäki, J., Thevenot, J., Pulkkinen, P., Kühn, V., Link, T. M., Eckstein, F., & Jämsä, T. J. (2012) · Bone · doi:10.1016/j.bone.2012.01.012 · Cited by 136 · cited 1×
“The range of Exp_Fu and Exp_S for the two loading configurations were in line with the values reported in the literature for similar in vitro test on human femora [9,61,62].”
ResultA comparison of enhanced continuum FE with micro FE models of human vertebral bodies
Pahr, D. H., & Zysset, P. K. (2009) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2008.11.028 · Cited by 133 · cited 1×
“A future study with improved smooth models [64] will evaluate this uncertainty on the model outcome.”
BackgroundSelf-citationPahr, D. H., & Zysset, P. K. (2008) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255840802144105 · Cited by 130 · cited 1×
“The external bone contour was found for each image with a filling out algorithm [52] as already described in a previous study for the human vertebra [53].”
MethodSelf-citationAccuracy of finite element predictions in sideways load configurations for the proximal human femur
Grassi, L., Schileo, E., Taddei, F., Zani, L., Juszczyk, M., Cristofolini, L., & Viceconti, M. (2011) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2011.10.019 · Cited by 127 · cited 2×
“99, [36-39]), to qualitatively predict the failure location [40] and to predict the femoral ultimate load better than BMD [41-43].”
BackgroundA 3D damage model for trabecular bone based on fabric tensors
Zysset, P. K., & Curnier, A. (1996) · Journal of Biomechanics · doi:10.1016/s0021-9290(96)80006-6 · Cited by 117
Self-citationFracture Prediction for the Proximal Femur Using Finite Element Models: Part II—Nonlinear Analysis
Lotz, J. C., Cheal, E. J., & Hayes, W. C. (1991) · Journal of Biomechanical Engineering · doi:10.1115/1.2895413 · Cited by 115 · cited 1×
“Due to the complexity of the experimental and modelling procedures, to the authors’ knowledge, only a few studies have investigated both loading conditions with the same model (17 pairs in [43] and one pair in [44,45]) and no one has investigated all above mentioned structural and local predictions of the same model.”
BackgroundBorggrefe, J., Graeff, C., Nickelsen, T. N., Marin, F., & Glüer, C. C. (2010) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.090820 · Cited by 112 · cited 2×
“[18] J.”
MethodAdvanced Imaging Assessment of Bone Quality
Genant, H. K., & Jiang, Y. (2006) · Annals of the New York Academy of Sciences · doi:10.1196/annals.1346.038 · Cited by 107 · cited 3×
“For its higher precision in measuring BMD compared to DXA [16,17], QCT has been recently used in clinical trials to estimate the effect of drug treatments on the femoral ultimate load [18–20] and to discriminate between patients who did or did not undergo a hip fracture [21,22].”
MethodGraeff, C., Chevalier, Y., Charlebois, M., Varga, P., Pahr, D. H., Nickélsen, T., Morlock, M. M., Glüer, C., & Zysset, P. K. (2009) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.090416 · Cited by 107 · cited 2×
“[54] which has been already used in a number of studies to model bone tissue [70-73].”
BackgroundSelf-citationLochmüller, E., Müller, R., Kühn, V., Lill, C. A., & Eckstein, F. (2003) · Journal of Bone and Mineral Research · doi:10.1359/jbmr.2003.18.5.906 · Cited by 104 · cited 2×
“[25] E.”
Femoral strength in osteoporotic women treated with teriparatide or alendronate
Keaveny, T. M., McClung, M. R., Wan, X., Kopperdahl, D. L., Mitlak, B., & Krohn, K. (2011) · Bone · doi:10.1016/j.bone.2011.10.002 · Cited by 103 · cited 1×
“Homogenized continuum level voxel based FE models (hvFE, where coarsened voxels are directly converted to hexahedron elements) have been recently used in clinical studies to evaluate the effect of drug treatments compared to baseline femoral ultimate load [29,30], to investigate the effect of gender and age on femoral…”
MethodValidation of an anatomy specific finite element model of Colles’ fracture
Varga, P., Baumbach, S. F., Pahr, D. H., & Zysset, P. K. (2009) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2009.04.017 · Cited by 100
Self-citationLewiecki, E., Keaveny, T., Kopperdahl, D., Genant, H., Engelke, K., Fuerst, T., Kivitz, A., Davies, R., & Fitzpatrick, L. (2009) · Journal of Clinical Endocrinology and Metabolism · doi:10.1210/jc.2008-1807 · Cited by 97 · cited 2×
“For its higher precision in measuring BMD compared to DXA [16, 17], QCT has been recently used in clinical trials to estimate the effect of drug treatments on the femoral ultimate load [18-20] and to discriminate between patients who did or did not undergo a hip fracture [21, 22].”
MethodOptimisation of brace treatment with numerical modelling
Berteau, J., Pastre, J.-P., Mesure, S., Pithioux, M., Bolini, G., & Chabrand, P. R. (2009) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255840903065399 · Cited by 91
Keaveny, T. M. (2010) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.231 · Cited by 86
Bousson, V., Adams, J., Engelke, K., Aout, M., Cohen-Solal, M., Bergot, C., Haguenauer, D., Goldberg, D., Champion, K., Aksouh, R., et al. (2011) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.270 · Cited by 85 · cited 1×
“For its higher precision in measuring BMD compared to DXA [16,17], QCT has been recently used in clinical trials to estimate the effect of drug treatments on the femoral ultimate load [18–20] and to discriminate between patients who did or did not undergo a hip fracture [21,22].”
MethodJuszczyk, M., Cristofolini, L., & Viceconti, M. (2011) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2011.05.038 · Cited by 82 · cited 2×
“Moreover, as the loading rate was found to affect the mechanical properties of the human femur in a STANCE configuration [11], it is likely that a similar effect would occur also for a SIDE configuration.”
BackgroundCong, A. X., den Buijs, J. O., & Dragomir‐Daescu, D. (2010) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2010.09.018 · Cited by 82 · cited 1×
“parameters like yield strain [42] and Young'smodulus-ash density relationship [46] were defined from a training set to obtain more realistic results.”
MethodMechanical testing of bones: the positive synergy of finite–element models and in vitro experiments
Cristofolini, L., Schileo, E., Juszczyk, M., Taddei, F., Martelli, S., & Viceconti, M. (2010) · Philosophical Transactions of the Royal Society A Mathematical Physical and Engineering Sciences · doi:10.1098/rsta.2010.0046 · Cited by 80 · cited 1×
“have to be accurately compared with experimental outcomes in vitro) to understand how reliable they are in predicting specific mechanical properties under certain conditions [35].”
BackgroundChevalier, Y., Pahr, D. H., & Zysset, P. K. (2009) · Journal of Biomechanical Engineering · doi:10.1115/1.3212097 · Cited by 79 · cited 2×
“[54] which has been already used in a number of studies to model bone tissue [70-73].”
BackgroundSelf-citationChevalier, Y., Quek, E., Borah, B., Gross, G. J., Stewart, J. A., Lang, T., & Zysset, P. K. (2009) · Bone · doi:10.1016/j.bone.2009.09.032 · Cited by 73
Self-citationTrabelsi, N., & Yosibash, Z. (2011) · Journal of Biomechanical Engineering · doi:10.1115/1.4004180 · Cited by 73 · cited 1×
“study has investigated the effect of including bone anisotropy (defined by geometrical andmicromechanical considerations) intoQCT-based FE models of the human femur [38].”
BackgroundRelationships between femoral fracture loads for two load configurations
Keyak, J. H. (2000) · Journal of Biomechanics · doi:10.1016/s0021-9290(99)00202-x · Cited by 71 · cited 1×
“The range of Exp_Fu and Exp_S for the two loading configurations were in line with the values reported in the literature for similar in vitro test on human femora [9,61,62].”
ResultMnif, H., Koubaa, M., Zrig, M., Trabelsi, R., & Abid, A. (2009) · Orthopaedics & Traumatology Surgery & Research · doi:10.1016/j.otsr.2009.08.001 · Cited by 70 · cited 1×
“In osteoporotic individuals, femoral fracture is a major clinical problemwith high morbidity andmortality [2,3].”
BackgroundViceconti, M., Taddei, F., Cristofolini, L., Martelli, S., Falcinelli, C., & Schileo, E. (2011) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2011.11.048 · Cited by 66 · cited 1×
“mechanics in one-legged stance is necessary to study the incidence of spontaneous fractures [9,10], the behavior of the hip during daily activities [11], the stress redistribution after hip surgery [12], the load sharing between cortical and trabecular bone [13], and the effect of pathologies like bone metastasis on b…”
BackgroundA sideways fall and hip fracture
Kannus, P., Leiponen, P., Parkkari, J. P., Palvanen, M., & Järvinen, M. (2006) · Bone · doi:10.1016/j.bone.2006.01.148 · Cited by 58
DXA predictions of human femoral mechanical properties depend on the load configuration
Dall’Ara, E., Luisier, B., Schmidt, R., Pretterklieber, M. L., Kainberger, F. M., Zysset, P. K., & Pahr, D. H. (2013) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2013.04.008 · Cited by 50
Self-citationLenaerts, L., & van Lenthe, H. (2009) · Philosophical Transactions of the Royal Society A Mathematical Physical and Engineering Sciences · doi:10.1098/rsta.2008.0302 · Cited by 41
Engelke, K., Fuerst, T., Dašić, G., Davies, R., & Genant, H. (2010) · Bone · doi:10.1016/j.bone.2010.03.003 · Cited by 37
The epidemiology of fractures of the proximal femur
Cordey, J., Schneider, M., & Bühler, M. (2000) · Injury · doi:10.1016/s0020-1383(00)80033-9 · Cited by 37 · cited 1×
“The setup was able to generate fractures that are usually recorded during clinical practice (sub-capital, neck and trochanteric fractures) [65,66].”
BackgroundA calibration methodology of QCT BMD for human vertebral body with registered micro-CT images
Dall’Ara, E., Varga, P., Pahr, D. H., & Zysset, P. K. (2011) · Medical Physics · doi:10.1118/1.3582946 · Cited by 37 · cited 3×
“[50] was found to be similar to the one previously found for the human vertebrae (μCTBV/TV=0.”
MethodResultInfluentialSelf-citationMartelli, S., Taddei, F., Schileo, E., Cristofolini, L., Rushton, N., & Viceconti, M. (2011) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2011.07.006 · Cited by 25 · cited 1×
“mechanics in one-legged stance is necessary to study the incidence of spontaneous fractures [9,10], the behavior of the hip during daily activities [11], the stress redistribution after hip surgery [12], the load sharing between cortical and trabecular bone [13], and the effect of pathologies like bone metastasis on b…”
Background
Cited by
Kopperdahl, D. L., Aspelund, T., Hoffmann, P. F., Sigurdsson, S., Siggeirsdottir, K., Harris, T. B., Gudnason, V., & Keaveny, T. M. (2014) · Journal of Bone and Mineral Research · doi:10.1002/jbmr.2069 · Cited by 286
Finite element analysis for prediction of bone strength
Zysset, P. K., Dall'Ara, E., Varga, P., & Pahr, D. H. (2013) · BoneKEy Reports · doi:10.1038/bonekey.2013.120 · Cited by 188
Self-citation
Metadata
- Authors
- Dall'Ara, E., Luisier, B., Schmidt, R., Kainberger, F., Zysset, P., & Pahr, D.
- Venue
- Bone
- Publisher
- Elsevier BV
- Published
- 2013-01
- Volume
- 52
- Issue
- 1
- Pages
- 27-38
- ISSN
- 1873-2763 (ISSN-L)
- 8756-3282
- OpenAlex
- W1983023120
- Cited by
- 176
- References
- 72
- FWCI
- 11.69
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.