The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
Open accessAbstract
To compare pulse oximetry measurement bias between infants with hypoxemia with either dark skin or light skin with Masimo Radical 7 and Nellcor Oximax. There was no significant difference in systematic bias based on skin pigment for either oximeter.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Among the 11/35 low risk studies, 3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation.
“3/11 (27.27%) reported no demonstrable loss of accuracy caused by skin pigmentation [46,48,70]”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [48]
The skin pigmentation effect on pulse oximetry accuracy has long been studied, reporting some contradictory conclusions.
“The skin pigmentation effect on pulse oximetry accuracy has long been studied reporting some contradictory conclusions.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [48]
Objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) and 4/37 studies (10.81%), respectively.
“We find that objective scales such as the Fitzpatrick Skin Pigmentation (FSP) scale or the Munsell scale were used by 6/37 studies (16.21%) [61,69,70,71,76,77] and 4/37 studies (10.81%) [47,48,74,89], respectively.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [48]
One study asserted that there is no significant difference in systematic bias based on skin pigment for pulse oximetry, but this study was not statistically significant since it had a small sample size and a stringent inclusion criterion.
“Although one study asserted that there is no significant difference in systematic bias based on skin pigment for pulse oximetry [48], this study was not statistically significant since it had a small sample size and a stringent inclusion criterion that was not typical for premature infants.”
From Skin pigmentation influence on pulse oximetry accuracy: a systematic review and bibliometric analysis (2022) · cites this paper as [48]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Callaway, C. W., Soar, J., Aibiki, M., Böttiger, B. W., Brooks, S. C., Deakin, C. D., Donnino, M. W., Drajer, S., Kloeck, W., Morley, P. T., et al. (2015) · Circulation · doi:10.1161/cir.0000000000000273 · Cited by 1,390 · see citing works
Part 7: Neonatal Resuscitation
Perlman, J. M., Wyllie, J. P., Kattwinkel, J., et al. (2015) · Circulation · doi:10.1161/cir.0000000000000276 · Cited by 696 · see citing works
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427 · see citing works
Feiner, J. R., Severinghaus, J. W., & Bickler, P. E. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000285988.35174.d9 · Cited by 400 · see citing works
Pulse Oximetry in Pediatric Practice
Fouzas, S., Priftis, K. Ν., & Anthracopoulos, M. Β. (2011) · PEDIATRICS · doi:10.1542/peds.2011-0271 · Cited by 218 · see citing works
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145 · see citing works
Accuracy of Pulse Oximetry in Children
Ross, P. A., Newth, C. J. L., & Khemani, R. G. (2014) · Pediatrics · doi:10.1542/peds.2013-1760 · Cited by 129 · see citing works
Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
Adler, J. N., Hughes, L. A., Vtvilecchia, R., & Jr., C. A. C. (1998) · Academic Emergency Medicine · doi:10.1111/j.1553-2712.1998.tb02772.x · Cited by 102 · see citing works
Association of Interstage Home Monitoring With Mortality, Readmissions, and Weight Gain
Oster, M. E., Ehrlich, A. C., King, E. C., et al. (2015) · Circulation · doi:10.1161/circulationaha.114.014107 · Cited by 73 · see citing works
Batchelder, P. B., & Raley, D. M. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000268495.35207.ab · Cited by 62 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Shi, C., Goodall, M., Dumville, J., Hill, J., Norman, G., Hamer, O., Clegg, A., Watkins, C. L., Georgiou, G., Hodkinson, A., et al. (2022) · doi:10.1101/2022.02.16.22271062 · Cited by 158 · see citing works
A Review of Wearable Multi-Wavelength Photoplethysmography
Ray, D., Collins, T., Woolley, S., & Ponnapalli, P. (2023) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2021.3121476 · Cited by 155 · see citing works
Cabanas, A. M., Fuentes-Guajardo, M., Latorre, K., León, D., & Martín-Escudero, P. (2022) · Sensors · doi:10.3390/s22093402 · Cited by 124 · see citing works
Racial discrepancy in pulse oximeter accuracy in preterm infants
Vesoulis, Z., Tims, A., Lodhi, H., Lalos, N., & Whitehead, H. (2022) · Journal of Perinatology · doi:10.1038/s41372-021-01230-3 · Cited by 120 · see citing works
Pulse Oximeter Performance, Racial Inequity, and the Work Ahead
Okunlola, O. E., Lipnick, M. S., Batchelder, P. B., Bernstein, M., Feiner, J. R., & Bickler, P. E. (2022) · Respiratory Care · doi:10.4187/respcare.09795 · Cited by 96 · see citing works
Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024) · Journal of Medical Internet Research · doi:10.2196/62769 · Cited by 36 · see citing works
Noninvasive Monitoring and Assessment of Oxygenation in Infants
Poets, C. F. (2019) · Clinics in Perinatology · doi:10.1016/j.clp.2019.05.010 · Cited by 24 · see citing works
- Authors
- Foglia, E. E., Whyte, R. K., Chaudhary, A., Mott, A., Chen, J., Propert, K. J., & Schmidt, B.
- Affiliations
- Children's Hospital of Philadelphia (US)
- Dalhousie University (CA)
- Baylor College of Medicine (US)
- University of Pennsylvania (US)
- Corresponding author
- Foglia, Elizabeth E.
- Venue
- The Journal of Pediatrics
- Publisher
- Elsevier BV
- Published
- 2017-03
- Volume
- 182
- Pages
- 375-377.e2
- ISSN
- 0022-3476 (ISSN-L)
- PMC
- PMC5328979
- OpenAlex
- W2560394825
- Cited by
- 96
- References
- 10
- Citations by year
- 2026: 10 · 2025: 21 · 2024: 15 · 2023: 7 · 2022: 31 · 2021: 7 · 2020: 3 · 2019: 1 · 2017: 1
- FWCI
- 0.32
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- The Journal of Pediatrics (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- Eunice Kennedy Shriver National Institute of Child Health and Human Development (L40 HD077522, T32HD060550)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Keywords
- pulse oximetry, skin pigmentation, hypoxemia, infants, measurement bias, dark skin, oxygen saturation monitoring
- MeSH terms
- Hypoxia, Cross-Sectional Studies, Female, Heart Defects, Congenital, Hospitals, Pediatric, Humans, Infant, Newborn, Infant, Premature, Intensive Care Units, Neonatal, Male, Oximetry, Prospective Studies, Sensitivity and Specificity, Skin Pigmentation, Critical Illness, Statistics, Nonparametric
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review