Effectiveness of continuous glucose monitoring systems on glycemic control in adults with type 1 diabetes: A systematic review and meta-analysis
Open accessAbstract
Introduction: Continuous glucose monitoring (CGM) technologies have been advancing rapidly, but evidence on their comparative effectiveness stills limited to date yet. We conducted a systematic review and meta-analysis to evaluate and investigate the impact of CGM systems on glycemic control in adults with type 1 diabetes. Methods: We searched electronic literature databases from inception through April 30, 2025, for comparative studies investigating CGM systems with standard monitoring or different CGM technologies in adults with type 1 diabetes. Primary outcomes included HbA1c reduction, time in range (TIR), and hypoglycemia reduction. We performed random-effects meta-analyses, network meta-analysis, and subgroup analyses by baseline HbA1c and intervention duration. Evidence quality was assessed using GRADE methodology. Results: Twenty-seven studies with total of 2975 participants were included. CGM significantly reduced HbA1c compared to standard monitoring (mean difference: 0.38 %, 95 % CI: 0.49 to -0.27 %). TIR increased by 7.9 % (95 % CI: 5.8-10.0 %), representing 114 additional minutes daily in best range. Real-time CGM showed advantages over intermittently scanned CGM for TIR (+5.63 %, P-value<0.001) and hypoglycemia reduction (-1.28 %, P-value<0.001). Automated closed-loop systems achieved the highest ranking in network meta-analysis (SUCRA = 0.92). Benefits were greater among patients with higher baseline HbA1c (>8.5 %: 0.68 % reduction in HbA1c vs. <7.5 %: 0.24 % reduction in HbA1c, P-value = 0.009). Conclusions: CGM technologies significantly improve glycemic control in adults with type 1 diabetes, with greater benefits for those with higher baseline HbA1c. Advanced systems demonstrate progressively greater improvements, with automated closed-loop systems showing the strongest evidence of effectiveness. These findings support broader implementation of CGM technologies, with selection tailored to individual patient needs.
AI snapshot
Written from the paper’s full text.
TL;DR
This systematic review and meta-analysis of 27 RCTs (2,975 adults with type 1 diabetes) found that continuous glucose monitoring improves glycemic control versus standard monitoring, with HbA1c reduction of 0.38% and time-in-range increase of 7.9% (114 min/day). Network meta-analysis ranked automated closed-loop systems highest (SUCRA 0.92), followed by real-time CGM (0.84), with greater benefits in patients with higher baseline HbA1c.
Key findings
- CGM versus standard monitoring reduced HbA1c (mean difference −0.38%, 95% CI 0.49 to −0.27%, P<0.001) and increased time in range by 7.9% (95% CI 5.8–10.0%, P<0.001), equal to 114 additional minutes daily in range.
- In direct rtCGM versus isCGM comparisons, rtCGM reduced HbA1c by 0.23% (95% CI 0.34 to −0.12%, P<0.001), increased TIR by 5.36% (95% CI 3.41–7.31%, P<0.001; 77 additional minutes/day), and produced a 1.03% greater reduction in time below range (95% CI 1.45 to −0.61%, P<0.001); heterogeneity was low (I² 12.9–23.5%).
- Network meta-analysis ranked automated hybrid closed-loop systems highest overall (SUCRA 0.92), followed by rtCGM (0.84), isCGM (0.57), sensor-augmented pump (0.39), and MDI + SMBG (0.05); closed-loop ranked first for HbA1c (P best 0.71) and TIR (P best 0.82), while rtCGM ranked first for hypoglycemia reduction (P best 0.62).
- Subgroup analyses showed greater HbA1c reduction with higher baseline HbA1c (>8.5%: 0.68% vs <7.5%: 0.24%, P=0.009) and greater TIR improvement (>8.5%: +9.8% vs <7.5%: +5.3%, P=0.015); duration subgroup differences were not significant for HbA1c (P=0.42), while TIR peaked in 3–6-month studies (+8.4%, P=0.002).
Methods
- PRISMA 2020 systematic review and meta-analysis; searched MEDLINE/PubMed, Scopus, CENTRAL, Google Scholar, and Web of Science from inception through April 30, 2025, for RCTs in adults ≥18 years with type 1 diabetes comparing any CGM system (rtCGM, isCGM, or automated insulin delivery/closed-loop) with SMBG or another CGM technology; 27 studies with 2,975 participants were included.
- Random-effects meta-analyses calculated mean differences with 95% CIs; network meta-analysis ranked technologies using SUCRA values; subgroup analyses were by baseline HbA1c (<7.5%, 7.5–8.5%, >8.5%) and intervention duration (<3, 3–6, >6 months); heterogeneity used I², publication bias used funnel plots, Egger’s test, and trim-and-fill, and evidence quality used RoB 2 and GRADE.
Limitations
- The paper states that inability to blind participants to visibly different devices caused performance bias; outcome definitions and measurement tools were heterogeneous, especially for hypoglycemia and patient-reported outcomes; few direct head-to-head comparisons limited precision; most studies lasted 3–6 months with few beyond 1 year; most studies were from high-income countries with predominantly Caucasian populations; exclusion of conference abstracts may have introduced publication bias; subgroup findings may be confounded by other patient characteristics; and adverse events, technical failures, and cost-effectiveness were inconsistently reported.
- The paper also states that no formal multiple-testing correction was applied despite multiple outcomes and comparisons, and that older studies evaluated CGM systems now superseded by newer generations.
- Evident from the text: reported rtCGM-versus-isCGM effects are inconsistent between the abstract (+5.63% TIR, −1.28% hypoglycemia) and results (+5.36% TIR, 1.03% greater hypoglycemia reduction), and the HbA1c confidence interval is printed as “0.49 to −0.27%,” which appears directionally inconsistent.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
10 of 21 cited sources checked · 3 supported · 1 flagged · accuracy not yet measured
- Findingcitation
The DPLG system evaluated by Benhamou et al. (2019) achieved the largest improvement in TIR (+14.5%, P<0.001).
“The DPLG system evaluated by Benhamou et al. (2019) achieved the largest improvement in TIR (+14.5 %, P-value<0.001), followed by similar benefits with other closed-loop systems.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Benhamou et al. (2019) source
The Results report a 9.2% improvement in time in range for DBLG1 versus sensor-assisted pump therapy (p<0.0001), not 14.5%. The Discussion describes this improvement as similar to those in previous trials, rather than the largest among closed-loop systems. (LLM, conf=0.88)
Checked by a reasoning model — may be wrong.
Experimental — accuracy not yet measured
- Findingcitation
Benhamou et al. (2019) reported an 8.3% reduction in CV (P < 0.001).
“The study by Benhamou et al. (2019) reported an 8.3 % reduction in CV (P-value<0.001)”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Benhamou et al. (2019) source
The Methods list glucose coefficient of variation (CV) as a secondary outcome, but the passages available from our copy do not report an 8.3% reduction in CV or its P-value. The Results report a 9.2% increase in time in the target glucose range, which is a different outcome. I searched for CV, coefficient of variation, and the claimed 8.3% figure. (LLM, conf=0.74)
Experimental — accuracy not yet measured
- Findingcitation
Visser et al. (2021) quantified an almost 21% reduction in hypoglycemia fear scores with rtCGM compared to isCGM (P = 0.003).
“The study by Visser et al. (2021) quantified an almost 21 % reduction in hypoglycemia fear scores with rtCGM compared to isCGM (P-value = 0.003).”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Visser et al. (2021) source
The Results and Table 2 (Additional information; page unavailable) support lower hypoglycaemia-worry scores with rtCGM than isCGM at six months. They report scores of 15.4 versus 18.0, an adjusted difference of −2.62 points and p=0.0071—not an almost 21% reduction with p=0.003. (LLM, conf=0.91)
Experimental — accuracy not yet measured
- Findingcitation
Studies evaluating closed-loop systems (Choudhary et al., 2022; Matejko et al., 2022) reported significant improvements in quality of life measures.
“the studies evaluating closed-loop systems (Choudhary et al., 2022, Matejko et al., 2022) reported significant improvements in quality of life measures.”
Quote in the paperNot a substantial claim1 of 2 sources checked - Findingcitation
Garg et al. (2011) found an 8.9% reduction in MAGE (P < 0.01).
“while Garg et al. (2011) found an 8.9 % reduction in MAGE (P-value<0.01).”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Garg et al. (2011) source
The Results (Glycemic variability indexes) report that MAGE fell during the six-month unblinded period. Table 2 (Results, Glycemic excursions; page not available) gives P < 0.001 for the blinded-versus-unblinded comparison, supporting P < 0.01, but it does not report an 8.9% reduction; the displayed MAGE values do not yield that percentage for either group. (LLM, conf=0.88)
Experimental — accuracy not yet measured
- Premisecitation
Glycemic variability is linked to oxidative stress, endothelial dysfunction, and cardiovascular outcomes independent of mean glucose levels.
“new literature evidence links glycemic variability to oxidative stress, endothelial dysfunction, and cardiovascular outcomes independent of mean glucose levels [[51], [52], [53], [54]].”
Quote in the paperNot a substantial claim1 of 3 sources checked- cites [51] Martinez, 2021, Glycemic variability and cardiovascular disease in patients with type 2 diabetes, BMJ Open Diabetes Res Care source
- cites [52] Tateishi, 2022, Impact of glycemic variability on coronary and peripheral endothelial dysfunction in patients with coronary artery disease, J Cardiol source
- cites [53], [54]
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Findingown data
In the rtCGM versus isCGM meta-analysis, heterogeneity was low across all outcomes (I2 ranging from 12.9% to 23.5%).
“Heterogeneity was low across all outcomes (I2 ranging from 12.9 % to 23.5 %), suggesting consistency in the observed advantages of rtCGM over isCGM.”
Quote in the paperSubstantial claim - Findingown data
In studies of automated insulin delivery systems, TIR absolute increases ranged from +4.1% to +14.5%.
“TIR showed the most observable benefits, with absolute increases ranging from +4.1 % to +14.5 % across studies.”
Quote in the paperSubstantial claim - Findingown data
In direct head-to-head comparisons, rtCGM was associated with a greater reduction in time below range than isCGM (mean difference 1.03%, 95% CI 1.45 to −0.61%, P<0.001; abstract reports −1.28%).
“Real-time CGM showed advantages over intermittently scanned CGM for TIR (+5.63 %, P-value<0.001) and hypoglycemia reduction (−1.28 %, P-value<0.001).”
- Also stated in: § Comparison of rtCGM vs. isCGM technologies · boosted
“A significant advantage was observed for hypoglycemia reduction, with rtCGM demonstrating a 1.03 % greater reduction in time below range compared to isCGM (95 % CI: 1.45 to −0.61 %, P-value<0.001).”
Quote in the paperSubstantial claim - Also stated in: § Comparison of rtCGM vs. isCGM technologies · boosted
- Findingown data
Among included studies, HbA1c reduction varied from no significant change to significant reductions up to −1.0%.
“The magnitude of HbA1c reduction varied across studies, ranging from no significant change to significant reductions of up to −1.0 %.”
Quote in the paperNot a substantial claim - Findingown data
Closed-loop and automated insulin delivery systems generally demonstrated larger HbA1c reductions (−0.5% to −1.0%) compared to standalone CGM systems.
“Studies investigating closed-loop and automated insulin delivery systems generally demonstrated larger HbA1c reductions (−0.5 % to −1.0 %) compared to standalone CGM systems, suggesting additional benefits with automated insulin delivery (Fig. 2).”
Quote in the paperSubstantial claim - Findingown data
The largest TIR improvements were observed with closed-loop/automated insulin delivery systems, with increases up to 18.5%.
“The largest TIR improvements were observed with closed-loop/automated insulin delivery systems, with increases of up to 18.5 %.”
Quote in the paperNot a substantial claim - Findingown data
In network meta-analysis pairwise comparisons, transitioning from MDI + SMBG to SAP yielded the largest TIR improvement (+6.8%, 95% CI 4.2-9.4%).
“Pairwise comparisons showed significant benefits for each progressive technology tier, with the largest TIR improvement seen when transitioning from MDI + SMBG to SAP (+6.8 %, 95 % CI: 4.2–9.4 %), and the most significant hypoglycemia reduction observed when moving from isCGM to rtCGM (−1.28 %, 95 % CI: 1.77 to −0.79 %).”
Quote in the paperSubstantial claim - Findingown data
Automated insulin delivery systems consistently demonstrated improvements in glycemic outcomes compared to standard therapy.
“These systems consistently demonstrated improvements in glycemic outcomes.”
Quote in the paperSubstantial claim - Findingown data
In network meta-analysis pairwise comparisons, moving from isCGM to rtCGM yielded the most significant hypoglycemia reduction (−1.28%, 95% CI 1.77 to −0.79%).
“Pairwise comparisons showed significant benefits for each progressive technology tier, with the largest TIR improvement seen when transitioning from MDI + SMBG to SAP (+6.8 %, 95 % CI: 4.2–9.4 %), and the most significant hypoglycemia reduction observed when moving from isCGM to rtCGM (−1.28 %, 95 % CI: 1.77 to −0.79 %).”
Quote in the paperSubstantial claim - Findingown data
Hypoglycemia outcomes in automated insulin delivery studies generally showed either no significant difference or modest reductions.
“Hypoglycemia outcomes generally showed either no significant difference or modest reductions with automated systems.”
Quote in the paperNot a substantial claim - Findingown data
In network meta-analysis, automated hybrid closed-loop systems achieved the highest overall effectiveness ranking among CGM technologies (SUCRA 0.92).
“Automated closed-loop systems achieved the highest ranking in network meta-analysis (SUCRA = 0.92).”
- Also stated in: § Network meta-analysis of CGM technologies · boosted
“Automated hybrid closed-loop systems demonstrated the highest overall effectiveness (SUCRA value: 0.92), ranking first for HbA1c reduction (probability of being best: 0.71) and TIR (probability of being best: 0.82).”
Quote in the paperSubstantial claim - Also stated in: § Network meta-analysis of CGM technologies · boosted
- Findingown data
The systematic review and meta-analysis included 27 studies with a total of 2975 participants.
“Twenty-seven studies with total of 2975 participants were included.”
- Also stated in: § Study characteristics · plain
“Our literature search identified 27 studies including 2975 participants that met our inclusion criteria and were included in the meta-analysis”
Quote in the paperSubstantial claim - Also stated in: § Study characteristics · plain
- Findingown data
Some studies reported reductions in hypoglycemia exposure up to 72% in time <3.0 mmol/L.
“Several studies reported significant reductions in hypoglycemia exposure, with decreases of up to 72 % in time <3.0 mmol/L.”
Quote in the paperSubstantial claim - Findingown data
Four studies provided direct head-to-head comparisons of rtCGM versus isCGM.
“Direct head-to-head comparisons of rtCGM versus isCGM technologies were available from four studies (Table 4).”
Quote in the paperSubstantial claim - Findingown data
Among studies reporting hypoglycemia outcomes, 20 of 24 (83%) reported reductions in time in hypoglycemia.
“For hypoglycemia outcomes, most studies (20/24, 83 %) reported reductions in TIH, despite these findings, the magnitude and statistical significance varied across studies.”
Quote in the paperSubstantial claim - Findingown data
In direct head-to-head comparisons, rtCGM was associated with a higher time in range than isCGM (mean difference +5.36%, 95% CI 3.41-7.31%, P<0.001; abstract reports +5.63%).
“Real-time CGM showed advantages over intermittently scanned CGM for TIR (+5.63 %, P-value<0.001) and hypoglycemia reduction (−1.28 %, P-value<0.001).”
- Also stated in: § Comparison of rtCGM vs. isCGM technologies · plain
“Similarly, TIR was significantly higher with rtCGM (mean difference: +5.36 %, 95 % CI: 3.41–7.31 %, P-value<0.001), equivalent to 77 additional minutes daily in the target range.”
Quote in the paperSubstantial claim - Also stated in: § Comparison of rtCGM vs. isCGM technologies · plain
- Findingown data
In adults with type 1 diabetes, CGM significantly reduced HbA1c compared with standard monitoring (pooled mean difference −0.38%, 95% CI 0.49 to −0.27%, P<0.001).
“CGM significantly reduced HbA1c compared to standard monitoring (mean difference: 0.38 %, 95 % CI: 0.49 to −0.27 %).”
- Also stated in: § Effect of CGM on HbA1c · boosted
“Among the studies comparing CGM to standard care, the pooled mean difference in HbA1c was −0.38 % (95 % CI: 0.49 to −0.27 %, P-value<0.001), indicating a clinically meaningful improvement.”
Quote in the paperSubstantial claim - Also stated in: § Effect of CGM on HbA1c · boosted
- Findingown data
Most included studies showed statistically significant improvements in HbA1c with CGM use.
“Most of the included studies have showed statistically significant improvements in HbA1c with CGM use.”
Quote in the paperSubstantial claim - Findingown data
Eight studies evaluated automated insulin delivery systems compared to standard therapy approaches.
“Eight studies evaluated automated insulin delivery systems compared to standard therapy approaches (Table 4).”
Quote in the paperNot a substantial claim - Findingcitation
The most substantial HbA1c reduction among automated insulin delivery studies was observed with closed-loop systems compared to MDI + CGM (−1.0%, P<0.001) in Choudhary et al. (2022).
“The most substantial HbA1c reduction was observed with closed-loop systems compared to MDI + CGM (−1.0 %, P-value<0.001) in the study by Choudhary et al. (2022).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Choudhary et al. (2022) source
- Findingown data
Our systematic review and meta-analysis provides significant and detailed evidence that CGM technologies significantly improve glycemic control across multiple dimensions.
“Our systematic review and meta-analysis provides significant and detailed evidence that CGM technologies significantly improve glycemic control across multiple dimensions.”
Quote in the paperSubstantial claim - Findingown data
Across the included studies, treatment satisfaction consistently improved.
“Fourteen studies reported patient-related outcomes, with consistent improvements in treatment satisfaction, quality of life, and reduced fear of hypoglycemia.”
Quote in the paperNot a substantial claim - Findingown data
Across the included studies, fear of hypoglycemia was consistently reduced.
“Fourteen studies reported patient-related outcomes, with consistent improvements in treatment satisfaction, quality of life, and reduced fear of hypoglycemia.”
Quote in the paperNot a substantial claim - Findingown data
Across the included studies, quality of life consistently improved.
“Fourteen studies reported patient-related outcomes, with consistent improvements in treatment satisfaction, quality of life, and reduced fear of hypoglycemia.”
Quote in the paperNot a substantial claim - Findingown data
Fourteen studies reported patient-related outcomes.
“Fourteen studies reported patient-related outcomes, with consistent improvements in treatment satisfaction, quality of life, and reduced fear of hypoglycemia.”
Quote in the paperNot a substantial claim - Findingown data
Ten studies reported improvements in glycemic variability across various metrics including CV, MAGE, and SD.
“For glycemic variability, ten studies reported improvements across various metrics including Coefficient of Variation (CV), Mean Amplitude of Glycemic Excursions (MAGE), and Standard Deviation (SD) of glucose values.”
Quote in the paperNot a substantial claim - Findingown data
We found significant improvements in TIR that are equivalent to nearly two additional hours daily in target range.
“We found consistent HbA1c reductions with CGM use, significant improvements in TIR that are equivalent to nearly two additional hours daily in target range, and meaningful reductions in hypoglycemia exposure across various definitions.”
Quote in the paperSubstantial claim - Findingown data
We found meaningful reductions in hypoglycemia exposure across various definitions.
“We found consistent HbA1c reductions with CGM use, significant improvements in TIR that are equivalent to nearly two additional hours daily in target range, and meaningful reductions in hypoglycemia exposure across various definitions.”
Quote in the paperSubstantial claim - Findingown data
We found consistent HbA1c reductions with CGM use.
“We found consistent HbA1c reductions with CGM use, significant improvements in TIR that are equivalent to nearly two additional hours daily in target range, and meaningful reductions in hypoglycemia exposure across various definitions.”
Quote in the paperSubstantial claim - Findingown data
Sensitivity analyses excluding studies before 2010 showed consistent effect directions (HbA1c reduction: 0.35% vs. −0.38%).
“However, sensitivity analyses excluding studies before 2010 showed consistent effect directions as, HbA1c reduction: 0.35 % vs. −0.38 %, suggesting that technological improvements have improved magnitude rather than altered the effectiveness patterns of CGM technologies.”
Quote in the paperSubstantial claim - Findingown data
The network meta-analysis demonstrated a clear hierarchy of effectiveness, with automated closed-loop systems achieving the highest overall ranking, followed by rtCGM, and isCGM.
“The network meta-analysis has demonstrated a clear hierarchy of effectiveness, with automated closed-loop systems achieving the highest overall ranking, followed by rtCGM, and isCGM.”
Quote in the paperSubstantial claim - Findingown data
Automated insulin delivery systems demonstrate superior effectiveness across outcomes compared to other systems.
“Our findings on automated insulin delivery systems support previous studies findings and extend previous meta-analyses by demonstrating their superior effectiveness across outcomes.”
Quote in the paperSubstantial claim - Findingown data
Network meta-analysis demonstrated a clear effectiveness hierarchy: automated closed-loop systems and rtCGM have superior benefits compared to isCGM and standard monitoring.
“Our network meta-analysis has demonstrated a clear effectiveness hierarchy, with automated closed-loop systems and rtCGM demonstrating superior benefits compared to isCGM and standard monitoring.”
Quote in the paperSubstantial claim - Findingown data
CGM use resulted in increases in TIR by 7.9%, equivalent to 114 minutes daily.
“CGM use resulted in meaningful reductions in HbA1c by 0.38 %, significant increases in TIR by 7.9 %, which is equivalent to 114 min daily, and consistent decreases in hypoglycemia exposure.”
Quote in the paperSubstantial claim - Findingown data
CGM improves hypoglycemia fear scores.
“Improvements in hypoglycemia fear scores, treatment satisfaction, and overall quality of life highlight and address the psychosocial impact of CGM technologies.”
Quote in the paperNot a substantial claim - Findingreasoning
CGM benefits were maintained or slightly improved with longer intervention durations.
“Importantly, CGM benefits were maintained or slightly improved with longer intervention durations, suggesting sustained effectiveness over time.”
Quote in the paperSubstantial claim - Findingown data
CGM use resulted in reductions in HbA1c by 0.38%.
“CGM use resulted in meaningful reductions in HbA1c by 0.38 %, significant increases in TIR by 7.9 %, which is equivalent to 114 min daily, and consistent decreases in hypoglycemia exposure.”
Quote in the paperSubstantial claim - Findingown data
Subgroup analysis by duration suggested maintained or improved benefits over time.
“However, our subgroup analysis by duration suggested maintained or improved benefits over time, longer-term effectiveness data would provide greater confidence regarding sustained benefits and adherence.”
Quote in the paperNot a substantial claim - Findingown data
rtCGM has clear advantages for hypoglycemia reduction estimated by −1.28% in time below range.
“The clear advantages of rtCGM for hypoglycemia reduction estimated by −1.28 % in time below range may be better for certain patients with hypoglycemia unawareness or frequent hypoglycemic events.”
Quote in the paperSubstantial claim - Findingown data
Statistically significant differences between subgroups by baseline HbA1c were observed.
“While we observed statistically significant differences between subgroups, individual studies within each stratum may differ in other important characteristics such as diabetes duration, prior technology experience, socioeconomic status, or healthcare system factors that could confound the observed relationships.”
Quote in the paperNot a substantial claim - Findingown data
The HbA1c reduction of −0.38% observed in our results represents a significant improvement.
“The HbA1c reduction of −0.38 % observed in our results represents a significant improvement that could translate to better and optimized reductions in microvascular complication risk.”
Quote in the paperSubstantial claim - Findingown data
The observed difference in HbA1c between rtCGM and isCGM is 0.23% (decreased).
“However, the slighter observed differences in HbA1c estimated by decreased 0.23 %, suggest that both technologies offer better benefits over standard monitoring, with specific technology selection could be guided by individual patient factors including hypoglycemia risk, cost considerations, and personal preferences.”
Quote in the paperSubstantial claim - Findingown data
Benefits were maintained or improved with longer-term use.
“Also, we identified that patients with higher baseline HbA1c derived nearly three-fold greater absolute improvements, while benefits were maintained or improved with longer-term use.”
Quote in the paperSubstantial claim - Findingown data
Patients with higher baseline HbA1c derived nearly three-fold greater absolute improvements.
“Also, we identified that patients with higher baseline HbA1c derived nearly three-fold greater absolute improvements, while benefits were maintained or improved with longer-term use.”
Quote in the paperSubstantial claim - Findingreasoning
The improvement in TIR of 7.9% translated into 114 min daily.
“The improvement in TIR of 7.9 % translated into 114 min daily, exceeds the threshold of 5 % considered clinically significant in recent consensus guidelines.”
Quote in the paperSubstantial claim - Findingown data
CGM reduces glycemic variability.
“Our assessment of glycemic variability, despite its limitation by the heterogeneous measurement strategies has consistently demonstrated reductions.”
Quote in the paperSubstantial claim - Findingown data
CGM use resulted in consistent decreases in hypoglycemia exposure.
“CGM use resulted in meaningful reductions in HbA1c by 0.38 %, significant increases in TIR by 7.9 %, which is equivalent to 114 min daily, and consistent decreases in hypoglycemia exposure.”
Quote in the paperSubstantial claim - Findingcitation
TIR improvements with automated insulin delivery systems were up to 15%.
“The TIR improvements which were up to increased 15 % estimate, with these systems represent an important consideration for us in diabetes management, approaching glucose profiles of individuals without diabetes [[45], [46], [47]].”
Quote in the paperNot a substantial claim - Findingown data
CGM improves treatment satisfaction.
“Improvements in hypoglycemia fear scores, treatment satisfaction, and overall quality of life highlight and address the psychosocial impact of CGM technologies.”
Quote in the paperNot a substantial claim - Findingown data
Subgroup analyses demonstrated a dose-response relationship between baseline HbA1c and treatment effect, with patients having poorer initial control deriving the greatest absolute benefits.
“Subgroup analyses demonstrated a dose-response relationship between baseline HbA1c and treatment effect, with patients having poorer initial control deriving the greatest absolute benefits.”
Quote in the paperNot a substantial claim - Findingown data
Baseline HbA1c is related to CGM treatment effect.
“The identified relationship between baseline HbA1c and treatment effect has important implications for resource allocation and patient selection.”
Quote in the paperSubstantial claim - Findingown data
CGM improves overall quality of life.
“Improvements in hypoglycemia fear scores, treatment satisfaction, and overall quality of life highlight and address the psychosocial impact of CGM technologies.”
Quote in the paperNot a substantial claim - Findingown data
Patients with baseline HbA1c over 8.5% achieved nearly three times the absolute HbA1c reduction (0.68%) compared to those with baseline HbA1c less than 7.5% (0.24%).
“The finding that patients with higher baseline HbA1c values over 8.5 % have achieved nearly three times the absolute HbA1c reduction by 0.68 % compared to those with good baseline control who are less than 7.5 % by estimated 0.24 %”
Quote in the paperSubstantial claim - Premisecitation
Glycemic management of T1DM has increasingly focused beyond HbA1c alone to include metrics such as TIR, hypoglycemia exposure, and glycemic variability.
“The glycemic management of T1DM has increasingly focused beyond HbA1c alone to include metrics such as TIR, hypoglycemia exposure, and glycemic variability, which reflect day-to-day glucose control and its impact on patients' lives.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [42] Maiorino, 2020, Effects of continuous glucose monitoring on metrics of glycemic control in diabetes: a systematic review with meta-analysis of randomized controlled trials source
- cites [43] Rizos, 2025, Continuous glucose monitoring in type 1 diabetes, type 2 diabetes, and diabetes during pregnancy source
- Study design and conductown data
Funnel plot analysis for HbA1c outcomes suggested minimal publication bias.
“Assessment of publication bias for HbA1c outcomes using funnel plot analysis suggested minimal bias (Fig. 3).”
Quote in the paperNot a substantial claim - Study design and conductown data
Egger's test was not statistically significant (P = 0.078).
“The funnel plot was reasonably symmetrical, and Egger's test was not statistically significant resulting with P-value = 0.078.”
Quote in the paperSubstantial claim - Study design and conductreasoning
The most common source of bias was performance bias.
“The most common source of bias was performance bias, that was present in studies due to the difficulty of blinding participants to visibly different glucose monitoring devices.”
Quote in the paperNot a substantial claim - Study design and conductown data
For patient-reported outcomes and glycemic variability, evidence quality was very low, with limitations across multiple GRADE domains including inconsistent measurement tools, indirectness, and imprecision.
“For patient-reported outcomes and glycemic variability, evidence quality was very low (⊕◯◯◯), with limitations across multiple GRADE domains including inconsistent measurement tools, indirectness, and imprecision.”
Quote in the paperNot a substantial claim - Interpretationown data
Technological improvements have improved the magnitude rather than altered the effectiveness patterns of CGM technologies.
“However, sensitivity analyses excluding studies before 2010 showed consistent effect directions as, HbA1c reduction: 0.35 % vs. −0.38 %, suggesting that technological improvements have improved magnitude rather than altered the effectiveness patterns of CGM technologies.”
Quote in the paperNot a substantial claim - Study design and conductreasoning
No formal multiple testing corrections were applied despite multiple outcomes and comparisons.
“Given the multiple outcomes and comparisons analyzed, we acknowledge the possible risk for Type I error inflation. However, we did not apply formal multiple testing corrections as our primary outcomes (HbA1c, TIR, and hypoglycemia) represent clinically distinct and established endpoints in diabetes care.”
Quote in the paperSubstantial claim- Rests on an assumption: HbA1c, TIR, and hypoglycemia represent clinically distinct and established endpoints in diabetes care.
- Study design and conductown data
For hypoglycemia reduction, evidence quality was low due to risk of bias and inconsistency in hypoglycemia definitions across studies.
“For hypoglycemia reduction, evidence quality was low (⊕⊕◯◯) due to both risk of bias and inconsistency in hypoglycemia definitions across studies.”
Quote in the paperSubstantial claim - Interpretationown data
The small adjustment from trim-and-fill suggests that possible publication bias had minimal impact on the overall findings.
“This small adjustment suggests that possible publication bias had minimal impact on our overall findings, providing additional confidence in the significance of the meta-analysis results.”
Quote in the paperNot a substantial claim - Study design and conductown data
GRADE assessment demonstrated variable quality of evidence across outcomes.
“GRADE assessment demonstrated variable quality of evidence across outcomes (Supplementary Table 2).”
Quote in the paperNot a substantial claim - Study design and conductnone given
Evidence for automated systems was rated moderate quality.
“Evidence for automated systems was rated as moderate quality (⊕⊕⊕◯).”
Quote in the paperNot a substantial claim - Premisenone given
Observational studies are susceptible to selection bias, confounding by indication, and unmeasured covariates that could significantly impact glycemic outcomes.
“While observational studies may offer larger sample sizes and real-world applicability, they are susceptible to selection bias, confounding by indication, and unmeasured covariates that could significantly impact glycemic outcomes.”
Quote in the paperSubstantial claim - Premisecitation
Despite several RCTs demonstrating benefits of various CGM technologies, uncertainty has persisted regarding their comparative effectiveness across different outcomes, patient subgroups, and technology types.
“Despite the several RCTs demonstrating benefits of various CGM technologies, uncertainty has persisted regarding their comparative effectiveness across different outcomes, patient subgroups, and technology types [40,41].”
Quote in the paperSubstantial claim0 of 2 sources checked - Study design and conductown data
Most studies had moderate overall risk of bias.
“Risk of bias assessment revealed that most studies had moderate overall risk of bias, (Supplementary Table 1).”
Quote in the paperNot a substantial claim - Study design and conductreasoning
Selection bias was low in most studies.
“Selection bias was low in most studies, indicating appropriate randomization procedures”
Quote in the paperNot a substantial claim - Study design and conductreasoning
Performance bias was present due to difficulty blinding participants to visibly different glucose monitoring devices.
“The most common source of bias was performance bias, that was present in studies due to the difficulty of blinding participants to visibly different glucose monitoring devices.”
Quote in the paperNot a substantial claim - Study design and conductown data
The funnel plot was reasonably symmetrical.
“The funnel plot was reasonably symmetrical, and Egger's test was not statistically significant resulting with P-value = 0.078.”
Quote in the paperSubstantial claim - Study design and conductown data
Attrition bias was predominantly low.
“Attrition bias was also predominantly low, indicating good follow-up and data completeness.”
Quote in the paperNot a substantial claim - Study design and conductreasoning
Detection bias was generally low.
“Detection bias was generally low, suggesting reliable outcome assessment despite difficulties with blinding.”
Quote in the paperNot a substantial claim - Study design and conductown data
Trim-and-fill method identified three potentially missing studies.
“Trim-and-fill method has identified three potentially missing studies, resulting in an adjusted pooled effect of −0.32 % (95 % CI: 0.45 to −0.19 %) compared to the original estimate of −0.38 % (95 % CI: 0.49 to −0.27 %).”
Quote in the paperNot a substantial claim - Study design and conductown data
Evidence for rtCGM versus isCGM comparisons was rated low quality, limited by risk of bias and imprecision due to the small number of direct comparison studies.
“The evidence for rtCGM versus isCGM comparisons was also rated as low quality (⊕⊕◯◯), limited by risk of bias and imprecision due to the small number of direct comparison studies.”
Quote in the paperNot a substantial claim - Study design and conductreasoning
Inability to blind participants to visibly different glucose monitoring devices resulted in performance bias in most included studies.
“First, the inability to blind participants to visibly different glucose monitoring devices resulted in performance bias in most included studies.”
Quote in the paperNot a substantial claim - Interpretationown data
CGM technologies meaningfully improve glycemic control across multiple dimensions in adults with T1D.
“This systematic review and meta-analysis provides significant evidence that CGM technologies meaningfully improve glycemic control across multiple dimensions in adults with T1D.”
Quote in the paperSubstantial claim - Interpretationcitation
The observed HbA1c reduction could translate to better reductions in microvascular complication risk.
“The HbA1c reduction of −0.38 % observed in our results represents a significant improvement that could translate to better and optimized reductions in microvascular complication risk.”
Quote in the paperSubstantial claim - Interpretationown data
CGM effectiveness is sustained over time.
“Importantly, CGM benefits were maintained or slightly improved with longer intervention durations, suggesting sustained effectiveness over time.”
Quote in the paperSubstantial claim - Study design and conductown data
Inspection of funnel plot asymmetry and trim-and-fill method suggested minimal impact of publication bias on primary findings.
“However, the inspection of funnel plot of asymmetry and application of trim-and-fill method suggested minimal impact of publication bias on our primary findings.”
Quote in the paperNot a substantial claim - Interpretationreasoning
Heterogeneity likely contributed to lower quality evidence ratings for these outcomes in the GRADE assessment.
“This heterogeneity likely contributed to the lower quality evidence ratings for these outcomes in our GRADE assessment.”
Quote in the paperSubstantial claim - Interpretationown data
rtCGM may be better for certain patients with hypoglycemia unawareness or frequent hypoglycemic events.
“The clear advantages of rtCGM for hypoglycemia reduction estimated by −1.28 % in time below range may be better for certain patients with hypoglycemia unawareness or frequent hypoglycemic events.”
Quote in the paperSubstantial claim - Study design and conductreasoning
Significant heterogeneity existed in outcome definitions and measurement approaches, especially for hypoglycemia and patient-reported outcomes.
“Second, significant heterogeneity existed in outcome definitions and measurement approaches, especially for hypoglycemia and patient-reported outcomes.”
Quote in the paperSubstantial claim - Interpretationown data
The TIR improvement exceeds the threshold of 5% considered clinically significant in recent consensus guidelines.
“The improvement in TIR of 7.9 % translated into 114 min daily, exceeds the threshold of 5 % considered clinically significant in recent consensus guidelines.”
Quote in the paperSubstantial claim- Rests on an assumption: A TIR improvement of 5% is considered clinically significant in recent consensus guidelines.
- Interpretationcitation
CGM-induced reductions in glycemic variability may have implications beyond acute glycemic control.
“These improvements may have implications beyond acute glycemic control”
Quote in the paperNot a substantial claim
- Classified as probably not a substantial claim. 15
- May combine two claims. 25
Probably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationPatient preference surveys consistently favored rtCGM systems over isCGM.
“Patient preference surveys consistently favored rtCGM systems over isCGM (Hašková et al., 2020)”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites Hašková et al., 2020 source
The Results (Secondary End Points; page unavailable) report no significant between-group differences in quality-of-life measures, but quality of life is not a patient-preference survey. The Procedures and Conclusions passages likewise do not report patient preferences. Searches for preference surveys and whether patients preferred rtCGM over isCGM found no passage addressing the claim. (LLM, conf=0.76)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationThis systematic review and meta-analysis was conducted following PRISMA 2020 guidelines.
“This systematic review and meta-analysis was conducted following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines [11].”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [11] Page, 2021, The PRISMA 2020 statement: an updated guideline for reporting systematic reviews, Rev Esp Cardiol source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationcitationReduced diabetes distress and improved quality of life may contribute to better treatment adherence and long-term engagement with self-management behaviors.
“Reduced diabetes distress and improved quality of life may contribute to better treatment adherence and long-term engagement with self-management behaviors [[48], [49], [50]].”
Quote in the paperNot a substantial claim2 of 3 sources checked- cites [48] Schierloh, 2024, Fear of hypoglycemia and sleep in children with type 1 diabetes and their parents, Front Endocrinol source
The Introduction discusses the effects of diabetes management and fear of hypoglycemia on sleep and quality of life, while the Results report fear-of-hypoglycemia scores. Neither section links reduced diabetes distress or improved quality of life to treatment adherence or long-term self-management engagement. Searches for adherence, long-term engagement, and related terms found no passage making that link. (LLM, conf=0.73)
Experimental — accuracy not yet measured - cites [49] Barnard-Kelly, 2024, Identifying the deficiencies of currently available CGM to improve uptake and benefit, Diabet Med source
- cites [50] Glocker, 2022, Fear of hypoglycemia and quality of life in young people with type 1 diabetes and their parents in the era of sensor glucose monitoring, Front Endocrinol source
The Introduction (page unavailable) says psychosocial coping may affect glycemic outcomes, which supports a possible link between psychosocial well-being and diabetes management. It does not say that reduced distress or improved quality of life improves treatment adherence or long-term self-management engagement. The Discussion (page unavailable) cautions against assuming causality between psychosocial and glycemic outcomes. (LLM, conf=0.77)
Experimental — accuracy not yet measured
May combine two claims.
Findingown dataFor HbA1c outcomes in automated insulin delivery studies, three studies showed significant reductions and five found no significant difference despite TIR improvements.
“For HbA1c outcomes, results were more variable, with three studies showing significant reductions and five finding no significant difference despite TIR improvements.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSeveral automated insulin delivery studies reported additional benefits including improved quality of life, reduced glycemic variability, and improved overnight glucose control, with multiple studies highlighting nocturnal glucose management benefits.
“Several studies reported additional benefits including improved quality of life, reduced glycemic variability, and improved overnight glucose control, with multiple studies highlighting particular benefits in nocturnal glucose management.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn subgroup analysis by intervention duration, HbA1c reduction was 0.21% for <3 months, 0.35% for 3-6 months, and 0.41% for >6 months, with no statistically significant subgroup difference (P=0.42).
“HbA1c reductions progressively increased from short-term studies (less than three-months: 0.21 %) to medium-term (three to six-months: 0.35 %) and long-term studies (over six-months: 0.41 %), however the difference was not statistically significant with P-value = 0.42.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn subgroup analysis, HbA1c reduction with CGM was greater in adults with baseline HbA1c >8.5% (−0.68%) than in those with baseline HbA1c <7.5% (−0.24%), with a statistically significant subgroup difference (P=0.009).
“Benefits were greater among patients with higher baseline HbA1c (>8.5 %: 0.68 % reduction in HbA1c vs. <7.5 %: 0.24 % reduction in HbA1c, P-value = 0.009).”
- Also stated in: § Subgroup analyses by baseline HbA1c and intervention duration · hedged
“When subgrouped by baseline HbA1c, a clear dose-response relationship appeared, with progressively greater HbA1c reductions as baseline values increased: 0.24 % for low HbA1c (<7.5 %), −0.39 % for moderate HbA1c (7.5–8.5 %), and −0.68 % for high HbA1c (>8.5 %). The difference between subgroups was statistically significant with P-value = 0.009.”
Quote in the paperSubstantial claim- Also stated in: § Subgroup analyses by baseline HbA1c and intervention duration · hedged
May combine two claims.
Findingown dataIn network meta-analysis, isCGM consistently ranked third across all outcomes (SUCRA 0.57), followed by sensor-augmented pump therapy (SUCRA 0.39) and standard multiple daily injections with SMBG (SUCRA 0.05).
“Intermittently scanned CGM consistently ranked third across all outcomes (SUCRA: 0.57), followed by sensor-augmented pump therapy (SUCRA: 0.39) and the standard multiple daily injections with SMBG (SUCRA: 0.05).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn subgroup analysis, TIR improvement with CGM was greater in adults with baseline HbA1c >8.5% (+9.8%) than in those with baseline HbA1c <7.5% (+5.3%), with a statistically significant subgroup difference (P=0.015).
“A similar pattern was observed for TIR improvements, with greater absolute increases in the high HbA1c subgroup (+9.8 %) compared to the low HbA1c subgroup (+5.3 %, P-value = 0.015 for subgroup difference).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn subgroup analysis by intervention duration, TIR improvement peaked in medium-duration studies (+8.4%), with a statistically significant subgroup difference (P=0.002).
“TIR improvements peaked in medium-duration studies (+8.4 %, P-value = 0.002 for subgroup difference).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataIn network meta-analysis, rtCGM achieved the second highest overall ranking (SUCRA 0.84) and ranked first for hypoglycemia reduction (probability of being best 0.62).
“Real-time CGM achieved the second highest overall ranking (SUCRA: 0.84) and ranked first for hypoglycemia reduction (probability of being best: 0.62).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataCGM increased time in range by 7.9% (95% CI 5.8-10.0%, P<0.001), equivalent to 114 additional minutes daily in the target range.
“TIR increased by 7.9 % (95 % CI: 5.8–10.0 %), representing 114 additional minutes daily in best range.”
- Also stated in: § Effect of CGM on time in range and hypoglycemia · plain
“The pooled absolute increase in TIR was 7.9 % (95 % CI: 5.8–10.0 %, P-value<0.001), representing 114 additional minutes daily spent in the optimal glucose range (70–180 mg/dL or 3.9–10.0 mmol/L).”
Quote in the paperSubstantial claim- Also stated in: § Effect of CGM on time in range and hypoglycemia · plain
May combine two claims.
Findingown dataDirect head-to-head comparisons revealed significant advantages of real-time CGM over intermittently scanned systems for all key outcomes, especially for hypoglycemia reduction.
“Direct head-to-head comparisons revealed significant advantages of real-time CGM over intermittently scanned systems for all key outcomes, especially regarded for hypoglycemia reduction.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe adjusted pooled effect after trim-and-fill was −0.32% (95% CI: 0.45 to −0.19%) compared to the original estimate of −0.38% (95% CI: 0.49 to −0.27%).
“Trim-and-fill method has identified three potentially missing studies, resulting in an adjusted pooled effect of −0.32 % (95 % CI: 0.45 to −0.19 %) compared to the original estimate of −0.38 % (95 % CI: 0.49 to −0.27 %).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataBeyond glycemic metrics, we documented consistent improvements in patient-reported outcomes and glycemic variability measures.
“Beyond glycemic metrics, we documented consistent improvements in patient-reported outcomes and glycemic variability measures.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductown dataFor HbA1c reduction and TIR, evidence quality was moderate, mostly limited by risk of bias due to performance bias in most studies.
“For HbA1c reduction and TIR, evidence quality was moderate (⊕⊕⊕◯), that was mostly limited by risk of bias due to performance bias in most studies.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Study design and conductnone givenTwo independent reviewers assessed risk of bias for each included study using the Cochrane Risk of Bias tool for RCTs (RoB 2).
“Two independent reviewers assessed the risk of bias for each included study using the Cochrane Risk of Bias tool for RCTs (RoB 2), which evaluates bias across five domains: randomization process, deviations from intended interventions, missing outcome data, outcome measurement, and selective reporting.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisereasoningRCTs provide the highest level of evidence for causal inference by minimizing confounding through randomization.
“We aimed to focus on RCTs only as they provide the highest level of evidence for causal inference by minimizing confounding through randomization.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataReporting bias was consistently low.
“while reporting bias was consistently low, suggesting transparent outcome reporting.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe quality of evidence for each outcome was assessed using GRADE.
“We assessed the quality of evidence for each outcome using the Grading of Recommendations, Assessment, Development, and Evaluations (GRADE) approach.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenSubgroup comparisons by baseline HbA1c have inherent limitations because individual studies within subgroups may differ in unmeasured characteristics beyond baseline HbA1c levels.
“We acknowledge that such comparisons have inherent limitations as individual studies within subgroups may differ in unmeasured characteristics beyond baseline HbA1c levels.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationCGM systems have progressed from retrospective professional devices to sophisticated real-time monitors and automated closed-loop systems that integrate with insulin delivery.
“These systems have progressed from retrospective professional devices to sophisticated real-time monitors and automated closed-loop systems that integrate with insulin delivery [39].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [39] Lin, 2021, Continuous glucose monitoring: a review of the evidence in type 1 and 2 diabetes mellitus source
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningThe exclusion of conference abstracts without full-text publications may introduce publication bias.
“Conference abstracts without full-text publications were excluded to ensure adequate data extraction and quality assessment, however we acknowledge this may introduce publication bias.”
Quote in the paperNot a substantial claim- Rests on an assumption: Conference abstracts without full-text publications do not allow adequate data extraction and quality assessment.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenPublication bias was assessed using funnel plots, Egger's test, and trim-and-fill analysis for outcomes with at least ten studies.
“Publication bias was assessed using funnel plots, Egger's test, and trim-and-fill analysis for outcomes with at least ten studies.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductreasoningRandom-effects models were used to account for anticipated heterogeneity between studies.
“We performed meta-analyses using random-effects models to account for anticipated heterogeneity between studies.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataMost included studies had relatively short durations around three to six months, with few extending beyond one year.
“Fourth, most included studies had relatively short durations that were around three to six months, with few extending beyond one year.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataLimited number of direct head-to-head comparison studies for certain technology types reduced precision of some effect estimates and necessitated reliance on network meta-analysis.
“Third, the limited number of direct head-to-head comparison studies for certain technology types reduced the precision of some effect estimates and necessitated reliance on network meta-analysis for detailed technology comparisons.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningFor a typical patient, the TIR improvement represents nearly two additional hours daily spent in optimal glucose range, which forms a significant improvement in day-to-day glucose management that likely contributes to reduced glycemic variability and improved quality of life.
“For a typical patient, this represents nearly two additional hours daily spent in optimal glucose range, which forms a significant improvement in day-to-day glucose management that likely contributes to reduced glycemic variability and improved quality of life.”
Quote in the paperSubstantial claim- Rests on an assumption: Increased TIR is associated with reduced risk of microvascular complications independent of HbA1c.
May combine two claims.
InterpretationreasoningBoth technologies offer better benefits over standard monitoring.
“However, the slighter observed differences in HbA1c estimated by decreased 0.23 %, suggest that both technologies offer better benefits over standard monitoring, with specific technology selection could be guided by individual patient factors including hypoglycemia risk, cost considerations, and personal preferences.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataNetwork meta-analysis demonstrated good consistency and transitivity.
“While our network meta-analysis demonstrated good consistency and transitivity, direct comparative evidence would strengthen confidence in the relative effectiveness rankings.”
Quote in the paperSubstantial claimMay combine two claims.
Study design and conductown dataMajority of studies were conducted in high-income countries with majority Caucasian populations, limiting generalizability to diverse global settings.
“In addition to that, the majority of studies were conducted in high-income countries with majority of Caucasian populations, which is limiting the generalizability to diverse global settings.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningWhile the observed reduction is smaller than 1%, consistent benefits across studies and the maintained effect over longer durations suggest important clinical implications for long-term outcomes.
“While our observed reduction is smaller than 1 %, the consistent benefits across studies and the maintained effect over longer durations suggest important clinical implications for long-term outcomes.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationUKPDS established that each 1% reduction in HbA1c is associated with a 37% decrease in microvascular complications and a 21% reduction in diabetes-related deaths.
“The UK Prospective Diabetes Study (UKPDS) established that each 1 % reduction in HbA1c is associated with a 37 % decrease in microvascular complications and a 21 % reduction in diabetes-related deaths [44].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [44] Holman, 2022, A brief history of the UK prospective diabetes study: rury R holman, for the UKPDS group source
May combine two claims.
Study design and conductnone givenReporting of adverse events, technical failures, and cost-effectiveness was inconsistent across studies, limiting ability to assess these factors.
“We shall not forget that the reporting of adverse events, technical failures, as well as the cost-effectiveness was inconsistent across studies, limiting our ability to confidently assess these important factors.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningAutomated insulin delivery systems may be more beneficial for patients with challenging glycemic profiles or those unable to achieve targets with standard management strategies.
“These systems may be more beneficial for patients with challenging glycemic profiles or those unable to achieve targets with standard management strategies.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductreasoningExclusion of conference abstracts without full-text publications may have introduced publication bias favoring positive results.
“Also, our exclusion of conference abstracts without full-text publications may have introduced publication bias, possibly favoring studies with positive results that proceeded to full publication.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningObserved improvements in glycemic control may reflect both direct technological impact and associated behavioral adaptations.
“The observed improvements in glycemic control may therefore reflect both the direct technological impact and associated behavioral adaptations.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
InterpretationreasoningDifferential effects by baseline HbA1c should be interpreted as suggestive rather than definitive evidence for treatment effect modification.
“The differential effects by baseline HbA1c should therefore be interpreted as suggestive rather than definitive evidence for treatment effect modification.”
Quote in the paperSubstantial claim
The paper relies on this without support.
High-quality RCTs on CGM in adults with type 1 diabetes are available.
“Given the availability of high-quality RCTs in the studied topic, we prioritized internal validity over external validity to establish clear causal relationships between CGM technologies and glycemic outcomes.”
The paper relies on this without support.
HbA1c, TIR, and hypoglycemia represent clinically distinct and established endpoints in diabetes care.
“However, we did not apply formal multiple testing corrections as our primary outcomes (HbA1c, TIR, and hypoglycemia) represent clinically distinct and established endpoints in diabetes care.”
The paper relies on this without support.
Children and adolescents with type 1 diabetes have different physiological characteristics including different insulin sensitivity patterns, growth-related hormonal fluctuations, and varying adherence behaviors that may differentially impact CGM effectiveness compared to adults.
“We excluded pediatric populations as children and adolescents with T1DM as they have different physiological characteristics including different insulin sensitivity patterns, growth-related hormonal fluctuations, and varying adherence behaviors that may differentially impact CGM effectiveness compared to adults.”
The paper relies on this without support.
Pregnant women have pregnancy-specific glycemic targets and altered glucose metabolism that require specialized management.
“Pregnant women were excluded due to pregnancy-specific glycemic targets and altered glucose metabolism that require specialized management.”
The paper relies on this without support.
Type 2 diabetes has different underlying pathophysiology, management lines, and glycemic targets from type 1 diabetes.
“T2DM participants were excluded as the underlying pathophysiology, management lines, and glycemic targets differ significantly from T1D, possibly confounding the analysis of effectiveness.”
The paper relies on this without support.
Conference abstracts without full-text publications do not allow adequate data extraction and quality assessment.
“Conference abstracts without full-text publications were excluded to ensure adequate data extraction and quality assessment, however we acknowledge this may introduce publication bias.”
The paper relies on this without support.
A TIR improvement of 5% is considered clinically significant in recent consensus guidelines.
“exceeds the threshold of 5 % considered clinically significant in recent consensus guidelines.”
The paper relies on this without support.
Increased TIR is associated with reduced risk of microvascular complications independent of HbA1c.
“This improvement is important for us given that increased TIR has been associated with reduced risk of microvascular complications independent of HbA1c.”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Network hypoglycemia estimate falls outside its confidence interval
MajorThe reported interval runs from −0.79% to +1.77%, so it excludes the stated reduction of −1.28%.
“the most significant hypoglycemia reduction observed when moving from isCGM to rtCGM (−1.28 %”
“95 % CI: 1.77 to −0.79 %”
The largest listed TIR improvement is attributed to the wrong technology
MajorThe text attributes an increase of up to 18.5% to closed-loop or automated systems, but Table 3 assigns the 18.5% increase to a CGM-versus-SMBG study.
“The largest TIR improvements were observed with closed-loop/automated insulin delivery systems, with increases of up to 18.5 %.”
“| Thabit et al. (2020) | CGM | SMBG | Increased | +18.5 % |”
A non-RCT appears among studies said to be RCTs only
MajorThe eligibility criteria restrict inclusion to RCTs, but Table 1 classifies the included Garg study as a Non-RCT.
“We included RCTs conducted in adults (≥18 years) with T1DM”
“| Garg et al. (2011) | United States | Non-RCT | 60 |”
Conflicting rtCGM-versus-isCGM TIR estimates
MinorThe abstract gives a 5.63% TIR advantage, while the results give 5.36% for the same technology comparison.
“Real-time CGM showed advantages over intermittently scanned CGM for TIR (+5.63 %, P-value<0.001)”
“TIR was significantly higher with rtCGM (mean difference: +5.36 %, 95 % CI: 3.41–7.31 %, P-value<0.001)”
Direct rtCGM HbA1c estimate falls outside its confidence interval
MinorThe reported endpoints, −0.12% and +0.34%, do not contain the stated mean difference of −0.23%.
“The pooled mean difference in HbA1c was −0.23 %”
“95 % CI: 0.34 to −0.12 %”
Automated-system HbA1c results count unreported outcomes as nonsignificant
MinorThe text says that, of eight studies, three found significant reductions and five found no significant difference. Table 4 instead lists HbA1c differences as not reported for Benhamou and Tauschmann, so those two cannot be counted as finding no significant difference.
“three studies showing significant reductions and five finding no significant difference despite TIR improvements.”
“| Benhamou et al. (2019) | DPLG | SAP | +14.5 % (p < 0.001) | Not reported |”
Cites
The PRISMA 2020 statement: An updated guideline for reporting systematic reviews
Page, M. J., McKenzie, J. E., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., Shamseer, L., Tetzlaff, J. M., Akl, E. A., Brennan, S. E., et al. (2021) · BMJ · doi:10.1136/bmj.n71 · Cited by 108,283 · see citing works
Declaración PRISMA 2020: una guía actualizada para la publicación de revisiones sistemáticas
Page, M. J., McKenzie, J. E., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., Shamseer, L., Tetzlaff, J. M., Akl, E. A., Brennan, S. E., et al. (2021) · Revista Española de Cardiología · doi:10.1016/j.recesp.2021.06.016 · Cited by 2,095 · cited 1× · see citing works
“This systematic review and meta-analysis was conducted following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines [11].”
DiMeglio, L. A., Evans-Molina, C., & Oram, R. A. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31320-5 · Cited by 1,561 · see citing works
Declaración PRISMA 2020: una guía actualizada para la publicación de revisiones sistemáticas
Page, M. J., McKenzie, J. E., Bossuyt, P. M., Boutron, I., Hoffmann, T. C., Mulrow, C. D., Shamseer, L., Tetzlaff, J. M., Akl, E. A., Brennan, S. E., et al. (2021) · Revista Española de Cardiología (English Edition) · doi:10.1016/j.rec.2021.07.010 · Cited by 1,313 · see citing works
Beck, R. W., Riddlesworth, T., Ruedy, K., Ahmann, A., Bergenstal, R., Haller, S., Kollman, C., Kruger, D., McGill, J. B., Polonsky, W., et al. (2017) · JAMA · doi:10.1001/jama.2016.19975 · Cited by 1,185 · see citing works
Heinemann, L., Freckmann, G., Ehrmann, D., Faber-Heinemann, G., Guerra, S., Waldenmaier, D., & Hermanns, N. (2018) · The Lancet · doi:10.1016/s0140-6736(18)30297-6 · Cited by 522 · see citing works
Tauschmann, M., Thabit, H., Bally, L., Allen, J. M., Hartnell, S., Wilinska, M. E., Ruan, Y., Sibayan, J., Kollman, C., Cheng, P., et al. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31947-0 · Cited by 440 · see citing works
van Beers, C. A. J., DeVries, J. H., Kleijer, S. J., Smits, M. M., Geelhoed-Duijvestijn, P. H., Kramer, M. H. H., Diamant, M., Snoek, F. J., & Serné, E. H. (2016) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(16)30193-0 · Cited by 386 · see citing works
Maiorino, M. I., Signoriello, S., Maio, A., Chiodini, P., Bellastella, G., Scappaticcio, L., Longo, M., Giugliano, D., & Esposito, K. (2020) · Diabetes Care · doi:10.2337/dc19-1459 · Cited by 364 · see citing works
Aleppo, G., Ruedy, K. J., Riddlesworth, T. D., Kruger, D. F., Peters, A. L., Hirsch, I., Bergenstal, R. M., Toschi, E., Ahmann, A. J., Shah, V. N., et al. (2017) · Diabetes Care · doi:10.2337/dc16-2482 · Cited by 329 · see citing works
Continuous glucose monitoring systems for type 1 diabetes mellitus
Langendam, M., Luijf, Y. M., Hooft, L., DeVries, J. H., Mudde, A. H., & Scholten, R. J. (2012) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd008101.pub2 · Cited by 262 · see citing works
Forlenza, G. P., Li, Z., Buckingham, B. A., Pinsker, J. E., Cengiz, E., Wadwa, R. P., Ekhlaspour, L., Church, M. M., Weinzimer, S. A., Jost, E., et al. (2018) · Diabetes Care · doi:10.2337/dc18-0771 · Cited by 246 · see citing works
Raccah, D., Sulmont, V., Reznik, Y., Guerci, B., Renard, É., Hanaire, H. P., Jeandidier, N., & Nicolino, M. P. (2009) · Diabetes Care · doi:10.2337/dc09-0750 · Cited by 236 · see citing works
Benhamou, P.-Y., Franc, S., Reznik, Y., Thivolet, C., Schaepelynck, P., Renard, E., Guerci, B., Chaillous, L., Lukas-Croisier, C., Jeandidier, N., et al. (2019) · The Lancet Digital Health · doi:10.1016/s2589-7500(19)30003-2 · Cited by 217 · see citing works
Choudhary, P., Kolassa, R., Keuthage, W., Kroeger, J., Thivolet, C., Evans, M., Ré, R., de Portu, S., Vorrink, L., Shin, J., et al. (2022) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00212-1 · Cited by 204 · see citing works
Continuous glucose monitoring: A review of the evidence in type 1 and 2 diabetes mellitus
Lin, R., Brown, F., James, S., Jones, J., & Ekinci, E. (2021) · Diabetic Medicine · doi:10.1111/dme.14528 · Cited by 197 · see citing works
Intermittently scanned continuous glucose monitoring for type 1 diabetes
Leelarathna, L., Evans, M. L., Neupane, S., Rayman, G., Lumley, S., Cranston, I., Narendran, P., Barnard-Kelly, K., Sutton, C. J., Elliott, R. A., et al. (2022) · New England Journal of Medicine · doi:10.1056/nejmoa2205650 · Cited by 174 · see citing works
Visser, M. M., Charleer, S., Fieuws, S., De Block, C., Hilbrands, R., Van Huffel, L., Maes, T., Vanhaverbeke, G., Dirinck, E., Myngheer, N., et al. (2021) · The Lancet · doi:10.1016/s0140-6736(21)00789-3 · Cited by 165 · see citing works
Glycemic variability and cardiovascular disease in patients with type 2 diabetes
Martinez, M., Santamarina, J., Pavesi, A., Musso, C., & Umpierrez, G. E. (2021) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2020-002032 · Cited by 155 · see citing works
Thabit, H., Lubina-Solomon, A., Stadler, M., Leelarathna, L., Walkinshaw, E., Pernet, A., Allen, J. M., Iqbal, A., Choudhary, P., Kumareswaran, K., et al. (2014) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(14)70114-7 · Cited by 150 · see citing works
Hásková, A., Radovnická, L., Petruželková, L., Parkin, C. G., Grunberger, G., Horová, E., Navrátilová, V., Kádě, O., Matoulek, M., Prázný, M., et al. (2020) · Diabetes Care · doi:10.2337/dc20-0112 · Cited by 146 · see citing works
Tanenberg, R. J., Bode, B. W., Lane, W. S., Levetan, C. S., Mestman, J. H., Peters, A. L., Tobian, J. A., Gross, T. M., & Mastrototaro, J. J. (2004) · Mayo Clinic Proceedings · doi:10.4065/79.12.1521 · Cited by 135 · see citing works
Dicembrini, I., Cosentino, C., Monami, M., Mannucci, E., & Pala, L. (2021) · Acta Diabetologica · doi:10.1007/s00592-020-01589-3 · Cited by 106 · see citing works
Riveline, J., Schaepelynck, P., Chaillous, L., Renard, É., Sola‐Gazagnes, A., Penfornis, A., Tubiana‐Rufi, N., Sulmont, V., Catargi, B., Lukas, C., et al. (2012) · Diabetes Care · doi:10.2337/dc11-2021 · Cited by 96 · see citing works
Molfetta, S. D., Gioia, L. D., Caruso, I., Cignarelli, A., Green, S. C., Natale, P., Strippoli, G. F., Sorice, G. P., Perrini, S., Natalicchio, A., et al. (2024) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3842 · Cited by 86 · see citing works
Garg, S. K., Voelmle, M. K., Beatson, C. R., Miller, H. A., Crew, L. B., Freson, B. J., & Hazenfield, R. M. (2011) · Diabetes Care · doi:10.2337/dc10-1852 · Cited by 85 · see citing works
The Role of Glycemic Variability in Cardiovascular Disorders
Alfieri, V., Myasoedova, V. A., Vinci, M. C., Rondinelli, M., Songia, P., Massaiu, I., Cosentino, N., Moschetta, D., Valerio, V., Ciccarelli, M., et al. (2021) · International Journal of Molecular Sciences · doi:10.3390/ijms22168393 · Cited by 77 · see citing works
Matejko, B., Juza, A., Kieć-Wilk, B., Cyranka, K., Krzyżowska, S., Chen, X., Cohen, O., Da Silva, J., Malecki, M. T., & Klupa, T. (2022) · Diabetes Care · doi:10.2337/dc22-0470 · Cited by 72 · see citing works
Thabit, H., Prabhu, J. N., Mubita, W., Fullwood, C., Azmi, S., Urwin, A., Doughty, I., & Leelarathna, L. (2020) · Diabetes Care · doi:10.2337/dc20-0736 · Cited by 69 · see citing works
Advances in Continuous Glucose Monitoring: Clinical Applications
Kwon, S. Y., & Moon, J. S. (2025) · Endocrinology and Metabolism · doi:10.3803/enm.2025.2370 · Cited by 67 · see citing works
Randomized Controlled Trial of Mobile Closed-Loop Control
Kovatchev, B., Anderson, S. M., Raghinaru, D., Kudva, Y. C., Laffel, L. M., Levy, C., Pinsker, J. E., Wadwa, R. P., Buckingham, B., Doyle, F. J., et al. (2020) · Diabetes Care · doi:10.2337/dc19-1310 · Cited by 60 · see citing works
Renard, É., Joubert, M., Villard, O., Dreves, B., Reznik, Y., Farret, A., Place, J., Breton, M. D., & Kovatchev, B. P. (2023) · Diabetes Care · doi:10.2337/dc23-0685 · Cited by 52 · see citing works
Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes
Eichenlaub, M., Waldenmaier, D., Wehrstedt, S., Pleus, S., Link, M., Jendrike, N. R. M., Öter, Ş., Haug, C., Schinz, M., Braunack-Mayer, V., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251315459 · Cited by 50 · see citing works
Automated insulin delivery systems: from early research to routine care of type 1 diabetes
Renard, É. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01929-5 · Cited by 47 · see citing works
Aiello, E. M., Deshpande, S., Özaslan, B., Wolkowicz, K. L., Dassau, E., Pinsker, J. E., & Doyle, F. J. (2021) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2021.100312 · Cited by 44 · see citing works
Identifying the deficiencies of currently available <scp>CGM</scp> to improve uptake and benefit
Barnard‐Kelly, K. D., Martínez‐Brocca, M. A., Glatzer, T., & Oliver, N. (2024) · Diabetic Medicine · doi:10.1111/dme.15338 · Cited by 37 · see citing works
Avari, P., Moscardó, V., Jugnee, N., Oliver, N. S., & Reddy, M. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819867688 · Cited by 29 · see citing works
Hybrid Closed Loop Using a Do-It-Yourself Artificial Pancreas System in Adults With Type 1 Diabetes
Nanayakkara, N., Sharifi, A., Burren, D., Elghattis, Y., Jayarathna, D. K., & Cohen, N. D. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231153882 · Cited by 27 · see citing works
Association between Variability of Metabolic Risk Factors and Cardiometabolic Outcomes
Park, M., & Choi, K. M. (2022) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2021.0316 · Cited by 25 · see citing works
Glocker, V., Bachmann, S., Hess, M., Szinnai, G., & Burckhardt, M.-A. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.958671 · Cited by 25 · see citing works
Rizos, E. C., Markozannes, G., Charitakis, N., Filis, P., Stoimeni, A. E., Nørgaard, K., Ntzani, E. E., & Tsilidis, K. K. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0599 · Cited by 18 · cited 1× · see citing works
“This progression and advancement in outcome assessment necessitates a re-evaluation of the evidence supporting different CGM technologies, with attention to these expanded metrics and patient-centered outcomes [42,43].”
Tateishi, K., Saito, Y., Kitahara, H., & Kobayashi, Y. (2022) · Journal of Cardiology · doi:10.1016/j.jjcc.2021.08.009 · Cited by 16 · see citing works
Narrative Review: Continuous Glucose Monitoring (CGM) in Older Adults with Diabetes
Wilson, A., Morrison, D., Sainsbury, C. A. R., & Jones, G. C. (2025) · Diabetes Therapy · doi:10.1007/s13300-025-01720-z · Cited by 16 · see citing works
Kim, J. Y., Kim, J. H., & Kim, J. H. (2025) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2024.0160 · Cited by 14 · see citing works
Langeland, L. B. L., Salvesen, Ø. O., Selle, H., Carlsen, S. M., & Fougner, K. J. (2012) · International Journal of Clinical Practice · doi:10.1111/j.1742-1241.2012.02947.x · Cited by 13 · see citing works
Yan, J., Zhou, Y., Zheng, X., Zheng, M., Lü, J., Luo, S., Yang, D., Deng, H., Xu, W., Bi, Y., et al. (2023) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.3614 · Cited by 11 · see citing works
Azzam, A. Y., Medina, L., Morsy, M. M., Essibayi, M. A., Altschul, D., & Nassar, M. (2025) · ASIDE Internal Medicine · doi:10.71079/aside.im.02202517 · Cited by 11 · see citing works
Self-citationZeng, J., Beck, M., Barouti, A. A., Löfvenborg, J. E., Carlsson, S., & Lampousi, A. (2025) · EClinicalMedicine · doi:10.1016/j.eclinm.2025.103222 · Cited by 11 · see citing works
Gupta, A. V., Mukherjee, S., Bhadada, S. K., Ram, S., Pal, R., Sachdeva, N. K., & Dutta, P. R. (2024) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2024.111610 · Cited by 10 · see citing works
brief history of the UK Prospective Diabetes Study: RURY R HOLMAN, FOR THE UKPDS GROUP
Holman, R. (2022) · British Journal of Diabetes · doi:10.15277/bjd.2022.359 · Cited by 10 · see citing works
Fear of hypoglycemia and sleep in children with type 1 diabetes and their parents
Schierloh, U., Aguayo, G. A., Fichelle, M., De Melo Dias, C., Schritz, A., Vaillant, M., Barnard-Kelly, K., Cohen, O., Gies, I., & de Beaufort, C. (2024) · Frontiers in Endocrinology · doi:10.3389/fendo.2024.1419502 · Cited by 7 · see citing works
Jensen, M. H., Vestergaard, P., Hirsch, I. B., & Hejlesen, O. K. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820957662 · Cited by 5 · see citing works
Uduku, C. N., Pendolino, V., Jugnee, N., Oliver, N. S., Fothergill, R. T., & Reddy, M. (2023) · Diabetic Medicine · doi:10.1111/dme.15100 · Cited by 4 · see citing works
van Baal, L., Heinemann, L., Reinold, J., von Conta, J., Bahnsen, F. H., Kleesiek, J., Führer-Sakel, D., & Tan, S. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251334633 · Cited by 4 · see citing works
Nassar, M., Azzam, A. Y., Morsy, M. M., Patoli, I., Gill, A. S., & Marte, E. (2024) · ASIDE Internal Medicine · doi:10.71079/aside.im.0000012262412 · Cited by 2 · see citing works
Self-citationContinuous and Intermittent Glucose Monitoring in 2024
Dovč, K., Bode, B. W., & Battelino, T. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.8802.kd · Cited by 1 · see citing works
van Baal, L., Heinemann, L., Reinold, J., et al. (2025) · Diabetologie und Stoffwechsel · doi:10.1055/s-0045-1807376 · Cited by 1 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- Dicembrini, 2021, 58, 401
Cited by
Still being checked.
- Authors
- Alfadli, S. F., Alotaibi, Y. S., Aqdi, M. J., Almozan, L. A., Alzubaidi, Z. B., Altemani, H. A., Almutairi, S. D., Alabdullah, H. A., Almehmadi, A. A., Alanzi, A. L., & Azzam, A. Y.
- Affiliations
- Imam Mohammad ibn Saud Islamic University (SA)
- Jazan University (SA)
- Umm al-Qura University (SA)
- Northern Border University (SA)
- King Abdulaziz University (SA)
- Imam Abdulrahman Bin Faisal University (SA)
- Ibn Sina National College for Medical Studies (SA)
- American Society for Inclusion, Diversity, and Equity in Healthcare (ASIDE) (US)
- Corresponding author
- Azzam, Ahmed Y.
- Venue
- Metabolism Open
- Publisher
- Elsevier BV
- Published
- 2025-09
- Volume
- 27
- Pages
- 100382
- ISSN
- 2589-9368 (ISSN-L)
- PMC
- PMC12337206
- OpenAlex
- W4412865072
- Semantic Scholar
- 0a55aba6855407c0a674ed13b8555a571d1cec73
- Cited by
- 11
- References
- 57
- Citations by year
- 2026: 8 · 2025: 3
- FWCI
- 2.33
- Type
- journal article
- Language
- en
- Update notices
- erratum (2025-09-01)
- Open access
- gold
- doi.org
- Licence
- Locations
- Metabolism Open (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- PubMed Central (submittedVersion, cc-by-nc-nd)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- continuous glucose monitoring, type 1 diabetes, glycemic control, HbA1c, time in range, hypoglycemia, flash glucose monitoring, real-time CGM, baseline HbA1c, adults
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review