Characterization of nerve-cuff electrode interface for biocompatible and chronic stimulating application
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A lower impedance magnitude is more suitable for both electrical recording and stimulation, increasing the recorded SNR and the stimulation efficiency.
“A lower impedance magnitude is more suitable for both electrical recording and stimulation, i.e., increasing the recorded SNR and the stimulation efficiency.”
From Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording (2019) · cites this paper as [40]
A larger CDC can decrease the current amplitudes that are required to activate the nerves during stimulation, which can also decrease the electrical damage to the nerve.
“A larger CDC can decrease the current amplitudes that are required to activate the nerves during the stimulation, which can also decrease the electrical damage to the nerve (10, 40, 41).”
From Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording (2019) · cites this paper as [40]
The average CDC (~9.7 mC/cm2) and the impedance magnitude (~156 ohms) at 1 kHz are comparable to those of previously reported extraneural electrodes.
“In addition, the average CDC (~9.7 mC/cm2) and the impedance magnitude (~156 ohms) at 1 kHz are comparable to those of previously reported extraneural electrodes (12, 40).”
From Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording (2019) · cites this paper as [40]
Cites
Neural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · see citing works
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · see citing works
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · see citing works
Kozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721 · see citing works
Oxygen, Reactive Oxygen Species and Tissue Damage
Bergamini, C. M., Gambetti, S., Dondi, A., & Cervellati, C. (2004) · Current Pharmaceutical Design · doi:10.2174/1381612043384664 · Cited by 661 · see citing works
Abidian, M. R., Corey, J. M., Kipke, D. R., & Martin, D. C. (2010) · Small · doi:10.1002/smll.200901868 · Cited by 405 · see citing works
Conducting Polymers for Neural Prosthetic and Neural Interface Applications
Green, R. A., & Abidian, M. R. (2015) · Advanced Materials · doi:10.1002/adma.201501810 · Cited by 397 · see citing works
In vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes
Weiland, J. D., Anderson, D. J., & Humayun, M. S. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.805487 · Cited by 382 · see citing works
In Vitro and In Vivo Evaluation of PEDOT Microelectrodes for Neural Stimulation and Recording
Venkatraman, S., Hendricks, J., King, Z. A., Sereno, A. J., Richardson-Burns, S., Martin, D., & Carmena, J. M. (2011) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2109399 · Cited by 319 · see citing works
Cuff electrodes for chronic stimulation and recording of peripheral nerve activity
Loeb, G. E., & Peck, R. A. (1996) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(95)00123-9 · Cited by 215 · see citing works
Negi, S., Bhandari, R., Rieth, L. W., & Solzbacher, F. (2010) · Biomedical Materials · doi:10.1088/1748-6041/5/1/015007 · Cited by 202 · see citing works
Sputtered iridium oxide films for neural stimulation electrodes
Cogan, S. F., Ehrlich, J., Plante, T. D., Smirnov, A., Shire, D. B., Gingerich, M. D., & Rizzo, J. F. (2008) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.31223 · Cited by 188 · see citing works
Yeo, M., Lee, J., Chun, W., & Kim, G. (2016) · Biomacromolecules · doi:10.1021/acs.biomac.5b01764 · Cited by 159 · see citing works
Hybrid Conducting Polymer–Hydrogel Conduits for Axonal Growth and Neural Tissue Engineering
Abidian, M. R., Daneshvar, E. D., Egeland, B. M., Kipke, D. R., Cederna, P. S., & Urbanchek, M. G. (2012) · Advanced Healthcare Materials · doi:10.1002/adhm.201200182 · Cited by 138 · see citing works
Performance of conducting polymer electrodes for stimulating neuroprosthetics
Green, R. A., Matteucci, P. B., Hassarati, R. T., Giraud, B., Dodds, C. W. D., Chen, S., Byrnes-Preston, P., Suaning, G. J., Poole-Warren, L. A., & Lovell, N. H. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/1/016009 · Cited by 127 · see citing works
Lee, H. J., Son, Y., Kim, J., Lee, C. J., Yoon, E.-S., & Cho, I. (2015) · Lab on a Chip · doi:10.1039/c4lc01321b · Cited by 106 · see citing works
Getting signals into the brain: visual prosthetics through thalamic microstimulation
Pezaris, J. S., & Eskandar, E. N. (2009) · Neurosurgical FOCUS · doi:10.3171/2009.4.focus0986 · Cited by 98 · see citing works
Cogan, S. F., Troyk, P. R., Ehrlich, J., Gasbarro, C. M., & Plante, T. D. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/2/008 · Cited by 88 · see citing works
Advances in auditory prostheses
Shannon, R. V. (2011) · Current Opinion in Neurology · doi:10.1097/wco.0b013e32834ef878 · Cited by 81 · see citing works
Nanowire-Based Electrode for Acute In Vivo Neural Recordings in the Brain
Suyatin, D. B., Wallman, L., Thelin, J., Prinz, C. N., Jörntell, H., Samuelson, L., Montelius, L., & Schouenborg, J. (2013) · PLoS ONE · doi:10.1371/journal.pone.0056673 · Cited by 81 · see citing works
Blau, A. W., Ziegler, C., Heyer, M., Endres, F., Schwitzgebel, G., Matthies, T., Stieglitz, T., Meyer, J.-U., & Göpel†, W. (1997) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(97)00017-1 · Cited by 79 · see citing works
In Vivo and In Vitro Comparison of the Charge Injection Capacity of Platinum Macroelectrodes
Leung, R. T., Shivdasani, M. N., Nayagam, D. A. X., & Shepherd, R. K. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2366514 · Cited by 72 · see citing works
Foldes, E. L., Ackermann, D. M., Bhadra, N., Kilgore, K. L., & Bhadra, N. (2011) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2010.12.020 · Cited by 61 · see citing works
Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer
Wang, K., Liu, C.-C., & Durand, D. M. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2008.926691 · Cited by 57 · see citing works
Cytotoxic analysis of the conducting polymer PEDOT using myocytes
Miriani, R. M., Abidian, M. R., & Kipke, D. R. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4649538 · Cited by 55 · see citing works
Conducting Polymers in Neural Stimulation Applications
Zhou, D. D., Cui, X. T., Hines, A., & Greenberg, R. J. (2009) · Biological and medical physics series · doi:10.1007/978-0-387-98120-8_8 · Cited by 53 · see citing works
Chu, J., Song, K., Han, S., Lee, S. H., Kim, J., Kang, J. Y., Hwang, D., Suh, J. F., Choi, K., & Youn, I. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/6/943 · Cited by 46 · see citing works
Self-citationCharacterization of iridium film as a stimulating neural electrode
Lee, I.-S. (2002) · Biomaterials · doi:10.1016/s0142-9612(01)00373-8 · Cited by 45 · see citing works
Lee, S. H., Jung, J. H., Chae, Y. M., Suh, J. F., & Kang, J. Y. (2010) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/20/3/035015 · Cited by 44 · see citing works
Self-citationFabrication of strongly adherent platinum black coatings on microelectrodes array
Tang, R., Pei, W., Chen, S., Zhao, H., Chen, Y., Han, Y. F., Wang, C., & Chen, H. (2013) · Science China Information Sciences · doi:10.1007/s11432-013-4825-6 · Cited by 44 · see citing works
Mikulewicz, M., & Chojnacka, K. (2010) · Biological Trace Element Research · doi:10.1007/s12011-010-8798-7 · Cited by 41 · see citing works
Evaluation of high-perimeter electrode designs for deep brain stimulation
Howell, B. L., & Grill, W. M. (2014) · Journal of Neural Engineering · doi:10.1088/1741-2560/11/4/046026 · Cited by 35 · see citing works
Vrabec, T. L., Bhadra, N., Wainright, J. S., Bhadra, N., Franke, M., & Kilgore, K. L. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1385-5 · Cited by 34 · see citing works
Hydrogel‐Mediated Direct Patterning of Conducting Polymer Films with Multiple Surface Chemistries
Park, S.-H., Yang, G. C., Madduri, N., Abidian, M. R., & Majd, S. (2014) · Advanced Materials · doi:10.1002/adma.201306093 · Cited by 34 · see citing works
Implantable electrode array with platinum black coating for brain stimulation in fish
Zhang, C., Liu, J., Tian, H., Kang, X., Du, J., Rui, Y.-F., Yang, B. T., & Yang, C. (2013) · Microsystem Technologies · doi:10.1007/s00542-013-2017-3 · Cited by 26 · see citing works
Sputtered platinum–iridium layers as electrode material for functional electrostimulation
Ganske, G., Slavcheva, E. P., van Ooyen, A., Mokwa, W., & Schnakenberg, U. (2011) · Thin Solid Films · doi:10.1016/j.tsf.2011.01.344 · Cited by 24 · see citing works
In vitro study of titanium nitride electrodes for neural stimulation
Aryan, N. P., Asad, M. I. H. B., Brendler, C., et al. (2011) · doi:10.1109/iembs.2011.6090791 · Cited by 20 · see citing works
Márton, G., Bakos, I., Fekete, Z., Ulbert, I., & Pongracz, A. (2013) · Journal of Materials Science Materials in Medicine · doi:10.1007/s10856-013-5114-z · Cited by 18 · see citing works
Successful removal and reimplant of vagal nerve stimulator device after 10 years
Giulioni, M., Martinoni, M., Naldi, I., Bisulli, F., Pozzati, E., & Tinuper, P. (2012) · Annals of Indian Academy of Neurology · doi:10.4103/0972-2327.94997 · Cited by 7 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Zhang, Y., Zheng, N., Cao, Y., Wang, F., Wang, P., Ma, Y., Lu, B., Hou, G., Fang, Z., Liang, Z., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw1066 · Cited by 314 · see citing works
Long-term usability and bio-integration of polyimide-based intra-neural stimulating electrodes
Wurth, S., Capogrosso, M., Raspopovic, S., Gandar, J., Federici, G., Kinany, N., Cutrone, A., Piersigilli, A., Pavlova, N., Guiet, R., et al. (2017) · Biomaterials · doi:10.1016/j.biomaterials.2017.01.014 · Cited by 176 · see citing works
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
Compliant peripheral nerve interfaces.
Paggi, V., Akouissi, O., Micera, S., & Lacour, S. P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abcdbe · Cited by 60 · see citing works
“Functional Electrical Stimulation Society pp 265–8 [217] Lee Y J, Kim H-J, Do S H, Kang J Y and Lee S H 2016 Characterization of nerve-cuff electrode interface for biocompatible and chronic stimulating application Sensors Actuators B 237 924–34 [218] Jiman A A, Ratze D C, Welle E J, Patel P R, Richie J M, Bottorff E C…”
BackgroundStretchable gold nanowire-based cuff electrodes for low-voltage peripheral nerve stimulation
Lienemann, S., Zötterman, J., Farnebo, S., & Tybrandt, K. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abfebb · Cited by 48 · see citing works
Printable microscale interfaces for long-term peripheral nerve mapping and precision control
Otchy, T. M., Michas, C., Lee, B., Gopalan, K., Nerurkar, V., Gleick, J., Semu, D., Darkwa, L., Holinski, B. J., Chew, D. J., et al. (2020) · Nature Communications · doi:10.1038/s41467-020-18032-4 · Cited by 35 · see citing works
Wu, Y., & Guo, L. (2020) · Neural Interface Engineering · doi:10.1007/978-3-030-41854-0_5 · Cited by 3 · see citing works
- Authors
- Lee, Y. J., Kim, H.-J., Do, S. H., Kang, J. Y., & Lee, S. H.
- Affiliations
- Korea Institute of Science and Technology (KR)
- Konkuk University (KR)
- Corresponding author
- Lee, Soo Hyun
- Venue
- Sensors and Actuators B: Chemical
- Publisher
- Elsevier BV
- Published
- 2016-12
- Volume
- 237
- Pages
- 924-934
- ISSN
- 0925-4005 (ISSN-L)
- OpenAlex
- W2471527680
- Cited by
- 54
- References
- 39
- Citations by year
- 2026: 3 · 2025: 12 · 2024: 1 · 2023: 4 · 2022: 5 · 2021: 8 · 2020: 7 · 2019: 5 · 2018: 5 · 2017: 4
- FWCI
- 2.62
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Sensors and Actuators B Chemical (publishedVersion)
- Funders
- Korea Institute of Science and Technology (2E26180)
- National Research Foundation of Korea
- Ministry of Education, Science and Technology (2009-0091918, 2010-0020786)
- Ministry of Science, ICT and Future Planning (NRF-2014R1A1A3052557)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- cuff electrodes, peripheral nerve stimulation, platinum black, charge injection capacity, interfacial impedance, chronic implantation, biocompatibility, polyimide electrodes, PEDOT, iridium oxide, platinum electrode, nerve injury, biphasic pulses
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review