Cuff electrodes for chronic stimulation and recording of peripheral nerve activity
Abstract
A comparative study of 5 different designs of nerve cuff electrodes was undertaken to determine their relative merits for stimulating and recording whole-nerve activity over extended periods of chronic implantation on large and small peripheral nerves in 8 cats. Four of the designs represent novel fabrication strategies, including 2 based on flexible, thin-film substrates and 2 based on dip-coating silicone elastomer on a cylindrical mandrel. Various advantages and shortcomings of these materials and designs are discussed in the context of the biophysical factors that influence these electrophysiological interfaces, particularly the problem of recording microvolt-level neurograms in the presence of millivolt-level electromyograms from adjacent muscles in freely behaving subjects. The most effective design was one in which a thin sheath of silicone rubber was wrapped around and intra-operatively sealed to a longitudinally slit, tripolar cuff made by dip-coating silicone over stranded stainless steel leads that were prepositioned on a mandrel using polyvinyl alcohol as a temporary adhesive. When properly installed, these electrodes had stable impedances, recruitment thresholds and relatively interference-free recording properties for the duration of this study (up to 9 weeks).
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Nerve electrodes implanted adjacent to, around, or within a peripheral nerve trunk or spinal root reduce the stimulus intensity required for activation compared with muscular or surface electrodes, thereby decreasing hazardous electrochemical processes and power consumption.
“Due to the proximity to the nerve, the stimulus intensity required for activation is reduced in comparison with muscular or surface electrodes and, consequently, hazardous electrochemical processes and power consumption of the stimulator system can be decreased”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Loeb and Peck, 1996
Neuromuscular stimulating electrodes should provide stimulation below the charge-carrying capacity and density that induce reversible electrochemical processes and axonal damage.
“Neuromuscular stimulating electrodes should provide stimulation below the charge-carrying capacity and density that induce reversible electrochemical processes and axonal damage”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Loeb and Peck, 1996
Cuff electrodes confine the stimulating current to the inner space of the electrode, avoiding stimulation of neighboring nerves and tissues.
“Cuff electrodes placed around the nerves have several advantages compared with surface, intramuscular, and epimysial electrodes: they allow for correct positioning of electrode leads to minimize mechanical distortion and lead failure, and the stimulating current is confined to the inner space of the electrode, thus avoiding the stimulation of other neighboring nerves and tissues (Loeb and Peck, 19”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Loeb and Peck, 1996
Cuff electrodes allow correct positioning of electrode leads to minimize mechanical distortion and lead failure.
“Cuff electrodes placed around the nerves have several advantages compared with surface, intramuscular, and epimysial electrodes: they allow for correct positioning of electrode leads to minimize mechanical distortion and lead failure, and the stimulating current is confined to the inner space of the electrode, thus avoiding the stimulation of other neighboring nerves and tissues (Loeb and Peck, 19”
From A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems (2005) · cites this paper as Loeb and Peck, 1996
Extrafascicular electrodes such as nerve cuff electrodes transduce signals with amplitudes below 20 μVpp.
“In addition, the signals transduced by extrafascicular electrodes, such as nerve cuff electrodes, have amplitudes less than 20 µVpp [41, 75, 127, 138].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [75]
Because of noise constraints, PNS electrophysiological recordings are mostly done with commercial bench-top low-noise amplifier setups that require tethering the subject.
“Due to the noise constraints placed on recording instrumentation, electrophysiological recordings from the PNS are largely performed using commercial bench-top low-noise amplifier setups [75, 82, 109] that require subjects to be tethered to the equipment.”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [75]
Extraneural interfaces leave the nerve intact but usually target several bundles of axons or fascicles within it, hence offer lower selectivity.
“Extraneural interfaces leave the nerve intact but usually target several bundles of axons or fascicles within it, hence offer lower selectivity.”
From Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations (2018) · cites this paper as Loeb and Peck 1996
Nerve electrodes implanted in previous studies have induced inflammation around the nerve causing device degradation and failure, limiting functionality in small animals due to mechanical damage and nerve injury.
“In addition, nerve electrodes implanted in previous studies have induced inflammation around the nerve causing device degradation and failure, limiting the functionality in small animals due to mechanical damage and nerve injury [14].”
From A novel flexible cuff-like microelectrode for dual purpose, acute and chronic electrical interfacing with the mouse cervical vagus nerve (2017) · cites this paper as [14]
Cites
A spiral nerve cuff electrode for peripheral nerve stimulation
Naples, G. G., Mortimer, J. T., Scheiner, A., & Sweeney, J. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.8670 · Cited by 415 · see citing works
Selective control of muscle activation with a multipolar nerve cuff electrode
Veraart, C., Grill, W. M., & Mortimer, J. T. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.237694 · Cited by 320 · see citing works
Brummer, S. B., Robblee, L. S., & Hambrecht, F. T. (1983) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1983.tb31628.x · Cited by 205 · see citing works
Functionally complex muscles of the cat hindlimb
Pratt, C. A., Chanaud, C. M., & Loeb, G. E. (1991) · Experimental Brain Research · doi:10.1007/bf00229407 · Cited by 179 · see citing works
Self-citationStable long-term recordings from cat peripheral nerves
Stein, R. B., Nichols, T. R., Jhamandas, J., Davis, L., & Charles, D. (1977) · Brain Research · doi:10.1016/0006-8993(77)90233-5 · Cited by 160 · see citing works
A method to effect physiological recruitment order in electrically activated muscle
Fang, Z.-P., & Mortimer, J. T. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.76384 · Cited by 125 · see citing works
Skin contact force information in sensory nerve signals recorded by implanted cuff electrodes
Haugland, M. K., Hoffer, J. A., & Sinkjær, T. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.296346 · Cited by 122 · see citing works
Loeb, G. E. (1993) · Experimental Brain Research · doi:10.1007/bf00230446 · Cited by 98 · see citing works
Self-citationMcCreery, D. B., Agnew, W. F., Yuen, T. G. H., & Bullara, L. A. (1992) · Medical & Biological Engineering & Computing · doi:10.1007/bf02446202 · Cited by 97 · see citing works
Loeb, G. E., Walker, A. E., Uematsu, S., & Konigsmark, B. W. (1977) · Journal of Biomedical Materials Research · doi:10.1002/jbm.820110206 · Cited by 95 · see citing works
Self-citationA Technique for Collision Block of Peripheral Nerve: Single Stimulus Analysis
van den Honert, C., & Mortimer, J. T. (1981) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1981.324718 · Cited by 82 · see citing works
Single unit conduction velocities from averaged nerve cuff electrode records in freely moving cats
Hoffer, J. A., Loeb, G. E., & Pratt, C. A. (1981) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(81)90033-9 · Cited by 73 · see citing works
Self-citationHoffer, J. A., Loeb, G. E., Marks, W. B., O'Donovan, M. J., Pratt, C. A., & Sugano, N. (1987) · Journal of Neurophysiology · doi:10.1152/jn.1987.57.2.510 · Cited by 71 · see citing works
Self-citationMarks, W. B., & Loeb, G. E. (1976) · Biophysical Journal · doi:10.1016/s0006-3495(76)85719-0 · Cited by 61 · see citing works
Self-citationElectroneurographic recordings with polymer cuff electrodes in paralyzed cats
Julien, C., & Rossignol, S. (1982) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(82)90078-4 · Cited by 54 · see citing works
Haugland, M. K., & Hoffer, J. A. (1994) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.296344 · Cited by 50 · see citing works
Impedance Properties of Metal Electrodes for Chronic Recording from Mammalian Nerves
Stein, R. B., Charles, D., Gordon, T., Hoffer, J. A., & Jhamandas, J. H. (1978) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1978.326287 · Cited by 48 · see citing works
Krarup, C., & Loeb, G. E. (1988) · Muscle & Nerve · doi:10.1002/mus.880110905 · Cited by 46 · see citing works
Self-citationA simple electrode for intact nerve stimulation and/or recording in semi-chronic rats
Sauter, J. F., Berthoud, H. R., & Jeanrenaud, B. (1983) · Pflügers Archiv - European Journal of Physiology · doi:10.1007/bf00585171 · Cited by 28 · see citing works
A bipolar electrode for peripheral nerve stimulation
Barone, F. C., Wayner, M. J., Aguilar-Baturoni, H. U., & Guevara-Aguilar, R. (1979) · Brain Research Bulletin · doi:10.1016/s0361-9230(79)80019-2 · Cited by 20 · see citing works
Physical Sensors for Biomedical Applications
&NA; (1980) · Journal of Clinical Engineering · doi:10.1097/00004669-198001000-00023 · Cited by 19 · see citing works
Selective activation of distant nerve fibers
Grill, W. M., & Mortimer, J. T. (2005) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1993.979119 · Cited by 13 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845 · see citing works
Selective Stimulation of Cat Sciatic Nerve Using an Array of Varying-Length Microelectrodes
Branner, A., Stein, R. B., & Normann, R. A. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.85.4.1585 · Cited by 358 · see citing works
Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces
Stieglitz, T., Beutel, H., Schuettler, M., & Meyer, J.-U. (2000) · Biomedical Microdevices · doi:10.1023/a:1009955222114 · Cited by 343 · see citing works
A multielectrode array for intrafascicular recording and stimulation in sciatic nerve of cats
Branner, A., & Normann, R. A. (2000) · Brain Research Bulletin · doi:10.1016/s0361-9230(99)00231-2 · Cited by 234 · see citing works
Safety of long-term electrical peripheral nerve stimulation: review of the state of the art
Günter, C., Delbeke, J., & Ortiz-Catalan, M. (2019) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-018-0474-8 · Cited by 198 · see citing works
Long-term usability and bio-integration of polyimide-based intra-neural stimulating electrodes
Wurth, S., Capogrosso, M., Raspopovic, S., Gandar, J., Federici, G., Kinany, N., Cutrone, A., Piersigilli, A., Pavlova, N., Guiet, R., et al. (2017) · Biomaterials · doi:10.1016/j.biomaterials.2017.01.014 · Cited by 176 · see citing works
Ortiz-Catalan, M., Brånemark, R., Håkansson, B., & Delbeke, J. (2012) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-11-33 · Cited by 144 · see citing works
A review for the peripheral nerve interface designer
Larson, C. E., & Meng, E. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2019.108523 · Cited by 133 · see citing works
A Microchannel Neuroprosthesis for Bladder Control After Spinal Cord Injury in Rat
Chew, D. J., Zhu, L., Delivopoulos, E., Minev, I. R., Musick, K. M., Mosse, C. A., Craggs, M., Donaldson, N., Lacour, S. P., McMahon, S. B., et al. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3007186 · Cited by 123 · see citing works
Thin films and microelectrode arrays for neuroprosthetics
Ordonez, J., Schuettler, M., Boehler, C., Boretius, T., & Stieglitz, T. (2012) · MRS Bulletin · doi:10.1557/mrs.2012.117 · Cited by 118 · see citing works
Chronic interfacing with the autonomic nervous system using carbon nanotube (CNT) yarn electrodes
McCallum, G. A., Sui, X., Qiu, C., Marmerstein, J., Zheng, Y., Eggers, T. E., Hu, C., Dai, L., & Durand, D. M. (2017) · Scientific Reports · doi:10.1038/s41598-017-10639-w · Cited by 102 · see citing works
Interfaces with the Peripheral Nerve for the Control of Neuroprostheses
del Valle, J., & Navarro, X. (2013) · International Review of Neurobiology · doi:10.1016/b978-0-12-420045-6.00002-x · Cited by 105 · see citing works
Thin film multi-electrode softening cuffs for selective neuromodulation
González-González, M. A., Kanneganti, A., Joshi-Imre, A., Hernandez-Reynoso, A. G., Bendale, G., Modi, R., Ecker, M., Khurram, A., Cogan, S. F., Voit, W. E., et al. (2018) · Scientific Reports · doi:10.1038/s41598-018-34566-6 · Cited by 97 · see citing works
Cuff electrodes for long-term recording of natural sensory information
Struijk, J. J., Thomsen, M., Larsen, J. O., & Sinkjaer, T. (1999) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.765194 · Cited by 97 · see citing works
Tissue-engineered peripheral nerve interfaces
Spearman, B. S., Desai, V. H., Mobini, S., McDermott, M. D., Graham, J. B., Otto, K. J., Judy, J. W., & Schmidt, C. E. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701713 · Cited by 96 · see citing works
Implantable neurotechnologies: a review of integrated circuit neural amplifiers
Ng, K. A., Greenwald, E., Xu, Y. P., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1431-3 · Cited by 83 · see citing works
Long Micro-Channel Electrode Arrays: A Novel Type of Regenerative Peripheral Nerve Interface
Lacour, S. P., Fitzgerald, J. J., Lago, N., Tarte, E., McMahon, S., & Fawcett, J. (2009) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2031241 · Cited by 75 · see citing works
Innovations in Prosthetic Interfaces for the Upper Extremity
Kung, T. A., Bueno, R. A., Alkhalefah, G. K., Langhals, N. B., Urbanchek, M. G., & Cederna, P. S. (2013) · Plastic and Reconstructive Surgery · doi:10.1097/prs.0b013e3182a97e5f · Cited by 71 · see citing works
Chapter 2 Bidirectional Interfaces with the Peripheral Nervous System
Micera, S., & Navarro, X. (2009) · International Review of Neurobiology · doi:10.1016/s0074-7742(09)86002-9 · Cited by 69 · see citing works
Delivopoulos, E., Chew, D. J., Minev, I. R., Fawcett, J. W., & Lacour, S. P. (2012) · Lab on a Chip · doi:10.1039/c2lc21277c · Cited by 68 · see citing works
Klinge, P. M., Vafa, M. A., Brinker, T., Brandis, A., Walter, G. F., Stieglitz, T., Samii, M., & Wewetzer, K. (2001) · Biomaterials · doi:10.1016/s0142-9612(00)00420-8 · Cited by 64 · see citing works
Compliant peripheral nerve interfaces.
Paggi, V., Akouissi, O., Micera, S., & Lacour, S. P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abcdbe · Cited by 60 · see citing works
“Different designs and materials therefore influence the implantation protocols and the long-term performance of the device, affecting parameters such as SNR and scarring [94].”
BackgroundFoldes, E. L., Ackermann, D. M., Bhadra, N., Kilgore, K. L., & Bhadra, N. (2011) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2010.12.020 · Cited by 61 · see citing works
A Parylene Self-Locking Cuff Electrode for Peripheral Nerve Stimulation and Recording
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2014.2333733 · Cited by 56 · see citing works
Giagka, V., & Serdijn, W. A. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0010-y · Cited by 58 · see citing works
Caravaca, A. S., Tsaava, T., Goldman, L., Silverman, H., Riggott, G., Chavan, S. S., Bouton, C., Tracey, K. J., Desimone, R., Boyden, E. S., et al. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa7a42 · Cited by 57 · see citing works
Polyimide micro-channel arrays for peripheral nerve regenerative implants
Lacour, S. P., Atta, R., FitzGerald, J. J., Blamire, M., Tarte, E., & Fawcett, J. (2008) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2008.05.031 · Cited by 56 · see citing works
Lee, Y. J., Kim, H.-J., Do, S. H., Kang, J. Y., & Lee, S. H. (2016) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2016.06.169 · Cited by 54 · see citing works
The design of and chronic tissue response to a composite nerve electrode with patterned stiffness
Freeberg, M. J., Stone, M. A., Triolo, R. J., & Tyler, D. J. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6632 · Cited by 54 · see citing works
Polymeric C-shaped cuff electrode for recording of peripheral nerve signal
Xue, N., Sun, T., Tsang, W. M., Delgado-Martinez, I., Lee, S.-H., Sheshadri, S., Xiang, Z., Merugu, S., Gu, Y., Yen, S.-C., et al. (2015) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2015.01.006 · Cited by 48 · see citing works
Regenerated sciatic nerve axons stimulated through a chronically implanted macro-sieve electrode
MacEwan, M. R., Zellmer, E. R., Wheeler, J. J., Burton, H., & Moran, D. W. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00557 · Cited by 48 · see citing works
Stable Detection of Movement Intent From Peripheral Nerves: Chronic Study in Dogs
Dweiri, Y. M., Eggers, T. E., Gonzalez-Reyes, L. E., Drain, J., McCallum, G. A., & Durand, D. M. (2017) · Proceedings of the IEEE · doi:10.1109/jproc.2016.2607520 · Cited by 44 · see citing works
Lee, S., & Lee, C. (2018) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2018.05.004 · Cited by 48 · see citing works
Biomechanical and functional variation in rat sciatic nerve following cuff electrode implantation
Restaino, S. M., Abliz, E., Wachrathit, K., Krauthamer, V., & Shah, S. B. (2014) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-11-73 · Cited by 44 · see citing works
Printable microscale interfaces for long-term peripheral nerve mapping and precision control
Otchy, T. M., Michas, C., Lee, B., Gopalan, K., Nerurkar, V., Gleick, J., Semu, D., Darkwa, L., Holinski, B. J., Chew, D. J., et al. (2020) · Nature Communications · doi:10.1038/s41467-020-18032-4 · Cited by 35 · see citing works
From softening polymers to multimaterial based bioelectronic devices
Ecker, M., Joshi-Imre, A., Modi, R., Frewin, C. L., Garcia-Sandoval, A., Maeng, J., Gutierrez-Heredia, G., Pancrazio, J. J., & Voit, W. E. (2019) · Multifunctional Materials · doi:10.1088/2399-7532/aaed58 · Cited by 34 · see citing works
Dweiri, Y. M., Stone, M. A., Tyler, D. J., McCallum, G. A., & Durand, D. M. (2016) · Journal of Visualized Experiments · doi:10.3791/54388 · Cited by 23 · see citing works
ICFEM 2000. Third IEEE International Conference on Formal Engineering Methods
(2000) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.867869 · Cited by 20 · see citing works
Čvančara, P., Valle, G., Müller, M., Guiho, T., Hiairrassary, A., Petrini, F., Raspopovic, S., Strauss, I., Granata, G., Fernandez, E., et al. (2019) · doi:10.1101/653964 · Cited by 12 · see citing works
A cable-tie-type parylene cuff electrode for peripheral nerve interfaces
Yu, H., Xiong, W., Zhang, H., Wang, W., & Li, Z. (2014) · 2014 IEEE 27th International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2014.6765560 · Cited by 10 · see citing works
Rabischong, B., Larraín, D., Rabischong, P., Botchorishvili, R., Fraisse, G., Gallego, S., Gaydier, P., Chardigny, J. M., & Avan, P. (2011) · Surgical Endoscopy · doi:10.1007/s00464-011-1750-x · Cited by 7 · see citing works
- Authors
- Loeb, G. E., & Peck, R. A.
- Affiliations
- Queen's University (CA)
- Corresponding author
- Loeb, G.E.
- ORCID
- Loeb, G.E.
- Venue
- Journal of Neuroscience Methods
- Publisher
- Elsevier BV
- Published
- 1996-01
- Volume
- 64
- Issue
- 1
- Pages
- 95-103
- ISSN
- 0165-0270 (ISSN-L)
- OpenAlex
- W2163710888
- Cited by
- 215
- References
- 22
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 7 · 2023: 6 · 2022: 4 · 2021: 13 · 2020: 12 · 2019: 14 · 2018: 13 · 2017: 15 · 2016: 19 · 2015: 8 · 2014: 9 · 2013: 10 · 2012: 9
- FWCI
- 0.81
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Journal of Neuroscience Methods (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institute of Neurological Disorders and Stroke (N01-NS-3-2380)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Keywords
- cuff electrodes, peripheral nerve stimulation, peripheral nerve recording, chronic implantation, thin film substrates, electromyographic interference, electrode impedance, recruitment threshold, polyvinyl alcohol adhesive
- MeSH terms
- Animals, Cats, Electric Stimulation, Electrodes, Electromyography, Electrophysiology, Equipment Design, Evoked Potentials, Muscles, Peripheral Nerves, Silicone Elastomers, Time Factors
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review