Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal Polymer
Open accessAbstract
Current electrode designs require flexible substrates that absorb little moisture and provide large charge injection capability. Sputtered iridium oxide films have superior charge injection capabilities versus noble metals and can adhere to various substrates. Liquid crystal polymers (LCPs) have very little water absorption compared to other flexible substrates. Therefore, the combination of sputtered iridium oxide film on LCP substrate was studied using 50 Hz, 100 micros duration, and 10 mA biphasic current waveforms for 700 h at 67 degrees C in bicarbonate buffer saline. Scanning electron micrograph analysis showed no delamination and approximately 1% of electrode material was lost to the bicarbonate buffer. The charge injection limit and the cathodic charge storage capacity within the water window were 4.6 +/- 1.0 and 31.5 +/-6.6 mC/cm2, respectively. Additional electrochemical analysis revealed significant charge imbalance attributed to oxygen reduction within the water window. These results, along with the flexible, chemically inert, and biocompatible substrate, indicate that sputtered iridium oxide films on LCP could become the method of choice for flexible substrate nerve electrodes.
AI snapshot
Written from the paper’s full text.
TL;DR
Reactively sputtered iridium oxide on liquid crystal polymer survived long-term in vitro stimulation without delamination and showed high charge injection, but exhibited significant charge imbalance under oxygen/carbon dioxide conditions attributed to oxygen reduction. The LCP lamination process damaged the iridium oxide films, so non-laminated epoxy-insulated electrodes were used for further testing.
Key findings
- After ~700 hours of stimulation at 67°C with 10 mA, 100 μs, 50 Hz biphasic waveforms, SEM showed no delamination and approximately 1% of electrode material was lost to bicarbonate buffer (0.051 μg iridium per electrode), equivalent to eight months at 37°C.
- Charge injection limit was 4.6 ± 1.0 mC/cm²; cathodic charge storage capacity was 31.5 ± 6.6 mC/cm² under 100% N₂ and 33.7 ± 10.1 mC/cm² under O₂/CO₂.
- Under O₂/CO₂, net average charge loss was −3.0 ± 1.1 mC/cm²·cycle, versus a net charge gain of 0.43 ± 0.38 mC/cm²·cycle under N₂; the water window narrowed to −0.3 to 0.6 V SCE under O₂/CO₂ versus −0.64 to 0.75 V SCE under N₂.
- Laminated electrodes showed cracks in the iridium oxide, sometimes traversing traces and preventing electrical continuity; non-laminated epoxy-insulated electrodes were used thereafter.
Methods
- Reactively sputtered iridium oxide films (2.03 μm) on 50 μm LCP with a 2 nm titanium adhesion layer; four sets of 1 mm² stimulation pads, traces, and bonding pads were fabricated, with half laminated with 50 μm LCP and half epoxy-insulated.
- Electrochemical analysis used cyclic voltammetry at 50 mV/s, impedance spectroscopy, and charge injection limit testing with 5 ms capacitively coupled biphasic waveforms; long-term testing stimulated eight electrodes in bicarbonate buffer at 67°C for four weeks (equivalent to eight months at 37°C) with 10 mA, 100 μs, 50 Hz, plus four unstimulated controls.
- Post-test analysis included SEM imaging and inductively coupled plasma mass spectroscopy (ICP-MS) for iridium in solution.
Limitations
- Stated: the LCP lamination process at around 300°C damaged the thin metal films, interrupting iridium oxide continuity; charge imbalance under O₂/CO₂ suggests oxygen reduction and potentially harmful species, and further studies are needed to confirm oxygen reduction, tissue damage, and whether electrodes must be recharacterized to a −0.3 V SCE charge injection limit.
- Evident: all testing was in vitro with accelerated soak conditions, with no in vivo validation; charge injection limit was measured only under O₂/CO₂; the laminated electrode design was not viable in the tested form; CSCc varied across electrodes.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 55 cited sources checked · 0 supported · 0 flagged
- Findingown data
Additional electrochemical analysis revealed significant charge imbalance within the water window.
“Additional electrochemical analysis revealed significant charge imbalance attributed to oxygen reduction within the water window.”
Quote in the paperNot a substantial claim - Findingown data
The charge injection limit within the water window was 4.6 +/− 1.0 mC/cm2.
“The charge injection limit and the cathodic charge storage capacity within the water window were 4.6 +/− 1.0mC/cm2 and 31.5 +/− 6.6mC/cm2 respectively.”
Quote in the paperSubstantial claim - Findingown data
Approximately 1% of electrode material was lost to the bicarbonate buffer.
“Scanning electron micrograph (SEM) analysis showed no delamination and approximately 1% of electrode material was lost to the bicarbonate buffer.”
Quote in the paperSubstantial claim - Findingown data
The cathodic charge storage capacity within the water window was 31.5 +/− 6.6 mC/cm2.
“The charge injection limit and the cathodic charge storage capacity within the water window were 4.6 +/− 1.0mC/cm2 and 31.5 +/− 6.6mC/cm2 respectively.”
Quote in the paperSubstantial claim - Findingown data
This voltage range was significantly smaller compared to the −0.6V to 0.7V SCE voltage range under 100% nitrogen gas.
“This voltage range was significantly smaller compared to the −0.6V to 0.7V SCE voltage range under 100% nitrogen gas.”
Quote in the paperSubstantial claim
- Findingown data
Visual inspection under a light microscope and SEM analysis of the laminated iridium oxide electrodes revealed many cracks in the iridium oxide that often traversed the whole electrode.
“Visual inspection under a light microscope and SEM analysis of the laminated iridium oxide electrodes revealed many cracks in the iridium oxide that often traversed the whole electrode.”
Quote in the paperSubstantial claim - Findingown data
At high magnification (10,000x), the iridium oxide film morphology was shown to be sponge-like with an amorphous structure.
“At high magnification (10,000x), the iridium oxide film morphology was shown to be sponge-like with an amorphous structure (Fig. 5b).”
Quote in the paperNot a substantial claim - Findingown data
Cyclic voltammogram data obtained under oxygen/carbon dioxide gas conditions was notably different from CVs under 100% nitrogen gas.
“Cyclic voltammogram data obtained under oxygen/carbon dioxide gas conditions was notably different from CVs under 100% nitrogen gas.”
Quote in the paperSubstantial claim - Findingown data
Integration of the cyclic voltammetry profile obtained within the water window of eight different electrodes revealed the cathodic charge storage capacity (CSCc) of the electrodes to be 31.5 +/− 6.6mC/cm2 (N = 8) under 100% nitrogen gas.
“Integration of the cyclic voltammetry profile obtained within the water window of eight different electrodes revealed the cathodic charge storage capacity (CSCc) of the electrodes to be 31.5 +/− 6.6mC/cm2 (N = 8) under 100% nitrogen gas (Fig. 2b).”
Quote in the paperSubstantial claim - Findingown data
The lower voltage limit boundary before significant charge imbalance occurred was between −0.6V and −0.7V SCE (N = 16).
“The lower voltage limit boundary before significant charge imbalance occurred was between −0.6V and −0.7V SCE (N = 16).”
Quote in the paperSubstantial claim - Findingcitation
Compared to activated iridium oxide films, the current density in the −0.1V to −0.64V SCE voltage range was significantly higher compared to other activated and electrodeposited iridium oxide films reported in literature [3].
“Compared to activated iridium oxide films, the current density in the −0.1V to −0.64V SCE voltage range was significantly higher compared to other activated and electrodeposited iridium oxide films reported in literature [3].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3] van Balken MR Vergunst H Bemelmans BL The Use of Electrical Devices for the Treatment of Bladder Dysfunction: a review of methods The Journal of Urology 172 846 851 2004 15310981
Not checked in the free run.
- Findingown data
SEM images of the epoxy-insulated iridium oxide electrodes also showed no cracks in the bulk electrode area.
“SEM images of the epoxy-insulated iridium oxide electrodes also showed no cracks in the bulk electrode area (Fig. 5a).”
Quote in the paperNot a substantial claim - Findingown data
A net average charge loss was determined to be −3.0+/−1.1mC/cm2·cycle for the electrodes under oxygen/carbon dioxide conditions (N = 8) compared to nitrogen-saturated solutions.
“A net average charge loss was determined to be −3.0+/−1.1mC/cm2·cycle for the electrodes under oxygen/carbon dioxide conditions (N = 8) (Fig 3c) compared to nitrogen-saturated solutions (Fig 3b).”
Quote in the paperSubstantial claim - Findingown data
Charge loss under these conditions increased linearly over multiple cycles R2 > 0.99.
“Charge loss under these conditions increased linearly over multiple cycles R2 > 0.99).”
Quote in the paperNot a substantial claim - Findingown data
The CSCc of the iridium oxide electrodes with oxygen/carbon dioxide gas was calculated to be 33.7 +/−10.1mC/cm2 (n = 8).
“Lastly, the CSCc of the iridium oxide electrodes with oxygen/carbon dioxide gas was calculated to be 33.7 +/−10.1mC/cm2 (n = 8).”
Quote in the paperNot a substantial claim - Findingown data
Taking the voltage limit to be equal to the lower voltage limit of the water window, −0.645V SCE, the data interpolation of the post-cathodic phase voltages yielded a charge injection limit of 4.6 +/−1.0mC/cm2 for the given 5ms duration, capacitive biphasic current waveform at 10Hz.
“Taking the voltage limit to be equal to the lower voltage limit of the water window, −0.645V SCE, the data interpolation of the post-cathodic phase voltages yielded a charge injection limit of 4.6 +/−1.0mC/cm2 for the given 5ms duration, capacitive biphasic current waveform at 10Hz.”
Quote in the paperSubstantial claim - Findingown data
After approximately 700 hours of stimulation at 67°C with 10mA, 100μs pulse width, biphasic cathodic waveforms stimulated at 50Hz, visual inspection under a light microscope revealed intact iridium oxide films with no visible cracking.
“After approximately 700 hours of stimulation at 67°C with 10mA, 100μs pulse width, biphasic cathodic waveforms stimulated at 50Hz, visual inspection under a light microscope revealed intact iridium oxide films with no visible cracking.”
Quote in the paperSubstantial claim - Findingown data
Under oxygen and carbon dioxide gas, the −0.3V and 0.6V SCE voltage range did not result in significant charge loss (N = 24).
“Under oxygen and carbon dioxide gas, the −0.3V and 0.6V SCE voltage range did not result in significant charge loss (N = 24).”
Quote in the paperSubstantial claim - Findingown data
Integration of cyclic voltammetry data obtained within the water window for the iridium oxide films under the oxygen and carbon dioxide gas mixture revealed a charge imbalance profile that was much larger than that obtained in 100% nitrogen gas.
“Integration of cyclic voltammetry data obtained within the water window for the iridium oxide films under the oxygen and carbon dioxide gas mixture revealed a charge imbalance profile that was much larger than that obtained in 100% nitrogen gas.”
Quote in the paperNot a substantial claim - Findingown data
The morphology and structure of the iridium oxide on the LCP substrate electrodes were markedly different compared to the cauliflower morphology reported by Cogan et al [1] and Slavcheva et al [2].
“The morphology and structure of the iridium oxide on the LCP substrate electrodes were markedly different compared to the cauliflower morphology reported by Cogan et al [1]and Slavcheva et al [2].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [1] Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot IEEE Transactions on Rehabilitation Engineering 10 260 279 2002 12611364 source
Not checked in the free run.
- cites [2] Degnan GG Wind TC Jones EV Edlich RF Functional Electrical Stimulation in Tetraplegic Patients to Restore Hand Function Journal of Long-term effects of Medical Implants 12 175 188 2002 12545943 source
Not checked in the free run.
- Findingown data
Cyclic voltammetry curves under 100% nitrogen gas revealed peaks at +0.2 V SCE (reduction), −0.4 V SCE (reduction), −0.1 V SCE (oxidation), and +0.45V SCE (oxidation).
“Cyclic voltammetry curves generated under 100% nitrogen gas revealed peaks present at +0.2 V SCE (reduction), −0.4 V SCE (reduction), −0.1 V SCE (oxidation), and +0.45V SCE (oxidation) (arrows, Fig. 2a).”
Quote in the paperSubstantial claim - Findingown data
Under 100% nitrogen gas, there was a net charge gain of 0.43 +/− 0.38mC/cm2·cycle.
“Under 100% nitrogen gas, there was a net charge gain of 0.43 +/− 0.38mC/cm2·cycle.”
Quote in the paperNot a substantial claim - Findingown data
The impedance profile in the Bode plot could not be fitted to a first order series resistor and capacitor model.
“The impedance profile in the Bode plot could not be fitted to a first order series resistor and capacitor model.”
Quote in the paperNot a substantial claim - Findingown data
The upper voltage limit boundary before significant charge imbalance occurred was between 0.8V and 0.7V SCE (N = 16) using a sample t-test (P<0.005).
“The upper voltage limit boundary before significant charge imbalance occurred was found to be between 0.8V and 0.7V SCE (N = 16) using a sample t-test (P<0.005).”
Quote in the paperSubstantial claim - Findingown data
The SEM pictures of the iridium electrode films revealed no cracking or significant loss of bulk iridium oxide material in both the control and the stimulated electrodes.
“The SEM pictures of the iridium electrode films revealed no cracking or significant loss of bulk iridium oxide material in both the control and the stimulated electrodes (Fig. 5c, d).”
Quote in the paperSubstantial claim - Findingown data
The reactively sputtered iridium oxide film did not delaminate from the liquid crystal polymer substrate during long-term stimulation and accelerated soak study.
“Despite these changes in the voltage waveform and the electrochemical response that occurred with the iridium oxide electrodes, the long-term stimulation and accelerated soak study showed that the reactively sputtered iridium oxide film did not delaminate from the liquid crystal polymer substrate.”
- Also stated in: § V. Conclusions · plain
“Long term testing at 100uC/cm2 (10mA at 100μs) showed that the electrode potential was stable, did not delaminate or generate significant iridium oxide dissolution.”
Quote in the paperSubstantial claim - Also stated in: § V. Conclusions · plain
- Findingown data
The charge injection limit of the iridium oxide film was 4.6 ± 1.0 mC/cm2 (around 4.5 mC/cm2).
“The large discrepancy between the cathodic charge storage capacity (33.7 +/−10.1mC/cm2) and the charge injection limit (4.6 +/− 1.0 mC/cm2) is most likely due to the rate of charge injection and withdrawal during the analyses.”
- Also stated in: § V. Conclusions · plain
“Electrochemical characterization of the sputtered iridium oxide electrodes before and after long term testing showed a large charge injection limit around 4.5mC/cm2.”
Quote in the paperSubstantial claim - Also stated in: § V. Conclusions · plain
- Findingown data
During the 700-hour stimulation with accelerated soak procedure, several changes in the iridium oxide film were observed, most notably CV changes with an oxidation peak at around 0.1V SCE.
“During the course of the 700-hour stimulation with accelerated soak procedure, several changes in the iridium oxide film were observed, most notably the CV changes in the film with the oxidation peak at around 0.1V SCE.”
Quote in the paperNot a substantial claim - Findingown data
The CSCc under oxygen/carbon dioxide conditions also varied over time.
“The CSCc under oxygen/carbon dioxide conditions also varied over time.”
Quote in the paperNot a substantial claim - Findingown data
The same iridium analysis performed on the bicarbonate saline samples from the control electrodes and from bicarbonate saline samples taken before the stimulation and soak procedure revealed no detectable concentrations of iridium in solution.
“The same iridium analysis performed on the bicarbonate saline samples from the control electrodes and from bicarbonate saline samples taken before the stimulation and soak procedure revealed no detectable concentrations of iridium in solution.”
Quote in the paperSubstantial claim - Findingown data
Trend analysis over the course of the stimulation procedure revealed an increase of approximately 0.8mC/cm2·day in the CSCc of the iridium oxide films.
“Trend analysis over the course of the stimulation procedure revealed an increase of approximately 0.8mC/cm2·day in the CSCc of the iridium oxide films, indicating further activation during the course of the stimulation and soak protocol.”
Quote in the paperNot a substantial claim - Findingown data
The cathodic charge storage capacity of the iridium oxide film was 33.7 ± 10.1 mC/cm2.
“The large discrepancy between the cathodic charge storage capacity (33.7 +/−10.1mC/cm2) and the charge injection limit (4.6 +/− 1.0 mC/cm2) is most likely due to the rate of charge injection and withdrawal during the analyses.”
Quote in the paperSubstantial claim - Findingown data
During the course of the long-term accelerated soak and stimulation procedure, the impedance of the electrodes did not vary significantly.
“During the course of the long-term accelerated soak and stimulation procedure, the impedance of the electrodes did not vary significantly.”
Quote in the paperSubstantial claim - Findingown data
The LCP lamination process damaged the thin metal films deposited on the substrate.
“The first problem is that the LCP lamination process damaged the thin metals films deposited on the substrate.”
- Also stated in: § V. Conclusions · plain
“Scanning electron microscopy revealed that the lamination process damaged the deposited thin films.”
Quote in the paperNot a substantial claim - Also stated in: § V. Conclusions · plain
- Findingown data
The working electrode voltage waveform baseline varied significantly over time, with −0.174V (reference to SCE) and a standard deviation of 0.153V.
“The working electrode voltage waveform baseline varied significantly over time, with −0.174V (reference to SCE) and a standard deviation of 0.153V.”
Quote in the paperNot a substantial claim - Findingown data
The variability in the CSCc of the iridium oxide film under oxygen/carbon dioxide conditions during the course of stimulation was observed in this study.
“The variability in the CSCc of the iridium oxide film under oxygen/carbon dioxide conditions during the course of the stimulation observed in this study has also been reported in other studies [8].”
Quote in the paperNot a substantial claim - Findingown data
The corners and edges were still well-defined with no signs of delamination or iridium oxide layer thinning after the soak and stimulus protocol when compared to the iridium oxide films before the procedure and to controls.
“The corners and edges were still well-defined with no signs of delamination or iridium oxide layer thinning after the soak and stimulus protocol when compared to the iridium oxide films before the procedure and to controls (Fig. 5e, f).”
Quote in the paperSubstantial claim - Findingreasoning
Charge imbalance was observed in the water window under oxygen/carbon dioxide conditions.
“Another problem is the charge imbalance observed in the water window under oxygen/carbon dioxide conditions.”
- Also stated in: § V. Conclusions · plain
“However, significant charge imbalance occurred within the water window voltage range attributed to faradic transfer reactions associated with oxygen reduction.”
Quote in the paperSubstantial claim - Also stated in: § V. Conclusions · plain
- Findingown data
Long-term testing at 100 uC/cm2 (10 mA at 100 μs) showed that the electrode potential was stable.
“Long term testing at 100uC/cm2 (10mA at 100μs) showed that the electrode potential was stable, did not delaminate or generate significant iridium oxide dissolution.”
Quote in the paperSubstantial claim - Findingown data
The amount of iridium found in the bicarbonate solution by ICP-MS was minimal compared to the amount of iridium oxide electrode material, on the order of 1% over the equivalent time of eight months.
“The amount of iridium found in the bicarbonate solution by ICP-MS was also considered to be minimal in comparison to the amount of iridium oxide electrode material, on the order of 1% over the equivalent time of eight months.”
- Also stated in: § V. Conclusions · plain
“Long term testing at 100uC/cm2 (10mA at 100μs) showed that the electrode potential was stable, did not delaminate or generate significant iridium oxide dissolution.”
Quote in the paperSubstantial claim - Also stated in: § V. Conclusions · plain
- Findingown data
The liquid crystal polymer substrate appeared intact after the accelerated soak protocol with no noticeable changes to size, thickness, or condition.
“The liquid crystal polymer substrate appeared intact after the accelerated soak protocol with no noticeable changes to size, thickness, or condition.”
Quote in the paperSubstantial claim - Findingreasoning
The iridium oxide film's charge storage capacity was significantly greater than its charge injection limit.
“The charge storage capacity of the iridium oxide film reported in this analysis was also significantly greater than the charge injection limit.”
Quote in the paperNot a substantial claim - Findingown data
Analysis of the baseline voltage before the current pulse did not reveal any significant trend in the baseline drift.
“Analysis of the baseline voltage before the current pulse did not reveal any significant trend in the baseline drift.”
Quote in the paperNot a substantial claim - Findingown data
Analysis of the bicarbonate buffer saline after the long-term stimulation and soak procedure revealed the presence of 0.051 μg of iridium in solution per electrode, equivalent to approximately 1% of the estimated mass of the electrodes after the equivalent stimulation and soak time of eight months at human body temperature.
“Analysis of the bicarbonate buffer saline after the long-term stimulation and soak procedure revealed the presence of 0.051 μg of iridium in solution per electrode, equivalent to approximately 1% of the estimated mass of the electrodes after the equivalent stimulation and soak time of eight months at human body temperature.”
Quote in the paperSubstantial claim- Rests on an assumption: The accelerated soak and stimulation protocol at 67°C for approximately 700 hours is equivalent to eight months at human body temperature.
- Premisecitation
Liquid crystal polymers have very little water absorption compared to other flexible substrates, absorbing one hundredth the amount of moisture compared to polyimide.
“Liquid crystal polymers (LCP) have very little water absorption compared to other flexible substrates.”
- Also stated in: § I. Introduction · plain
“However, LCP absorbs one hundredth the amount of moisture compared to polyimide [25] [26].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [25] Wang X Engel J Liu C Liquid Crystal Polymer (LCP) for MEMS: processes and applications Journal of Micromechanics and Microengineering 13 628 633 2003
Not checked in the free run.
- cites [26] Hayden T New Liquid Crystal Polymer Flex Circuits to Meet Demanding Reliability and End-use Applications International Conference on Advanced Packaging and Systems 2002
Not checked in the free run.
- Also stated in: § I. Introduction · plain
- Interpretationown data
Sputtered iridium oxide films on liquid crystal polymer could become the method of choice for flexible substrate nerve electrodes.
“These results, along with the flexible, chemically inert, biocompatible substrate, indicate that sputtered iridium oxide films on liquid crystal polymer could become the method of choice for flexible substrate nerve electrodes.”
Quote in the paperNot a substantial claim- Rests on an assumption: LCP is chemically inert, facilitating long-term biological implantation.
- Premisecitation
Sputtered iridium oxide films have superior charge injection capabilities versus noble metals.
“Sputtered iridium oxide films have superior charge injection capabilities versus noble metals and can adhere to various substrates.”
- Also stated in: § I. Introduction · plain
“Charge injection limits for iridium oxide films made from iridium wire have been reported to be +/− 2 mC/cm2 and −/+ 1 mC/cm2 for anodic-first and cathodic-first current waveforms, respectively and are significantly higher than those reported for stainless steel or platinum [9].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [9] Agnew WF McCreery DB Neural Prostheses: Fundamental Studies Englewood Cliffs, NJ Prentice Hall 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [10] Weiland JD Anderson DJ Humayun MS In Vitro Electrical Properties for Iridium Oxide Versus Titanium Nitride Stimulating Electrodes IEEE Transactions on Neural Systems and Rehabilitation Engineering 40 1574 1579 2002 12549739 source
Not checked in the free run.
- Also stated in: § I. Introduction · plain
- Premisecitation
Sputtered iridium oxide films can adhere to various substrates.
“Sputtered iridium oxide films have superior charge injection capabilities versus noble metals and can adhere to various substrates.”
- Also stated in: § I. Introduction · plain
“Sputtered iridium oxide films have been deposited onto platinum [11] and silicon [16] substrates.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [11] Cogan SF Plante TD Ehrlich J Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes 26th Annual International Conference of the IEEE EMBS 2004 4153 4156 PMC2709748 17271216 source
Not checked in the free run.
- cites [16] Tanghe SJ Najafi K Wise KD A Planar IrO Multichannel Stimulating Electrode for Use in Neural Prostheses Sensors and Actuators, B1 1 464 467 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Also stated in: § I. Introduction · plain
- Interpretationown data
The charge imbalance is attributed to oxygen reduction within the water window.
“Additional electrochemical analysis revealed significant charge imbalance attributed to oxygen reduction within the water window.”
Quote in the paperNot a substantial claim - Interpretationown data
The lamination process damaged the iridium oxide layer.
“These data suggest that the lamination process damaged the iridium oxide layer and further experiments were carried on electrodes without LCP lamination.”
Quote in the paperSubstantial claim - Study design and conductcitation
The time and temperature used in the long-term stimulation and accelerated soak study were equivalent to a period of eight months at 37°C.
“Based on the empirical relationship that an environmental temperature increase of 10°C doubles the effective time of the study [27], the time and temperature used in this study were equivalent to a period of eight months at 37°C.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [27] Masel RI Chemical Kinetics and Catalysis New York Wiley-Interscience 2001
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
The significant negative charge gain with the addition of oxygen strongly suggests the presence of oxygen reduction reactions occurring between −0.3V to −0.6V SCE.
“The significant negative charge gain with the addition of oxygen strongly suggests the presence of oxygen reduction reactions occurring between −0.3V to −0.6V SCE.”
Quote in the paperSubstantial claim - Premisecitation
Some applications in electrical stimulation, such as peripheral nerve cuff electrodes, require flexible substrates.
“However, some applications in electrical stimulation, such as peripheral nerve cuff electrodes, require flexible substrates [12].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [12] Slavcheva E Vitushinsky R Mokwa W Schnakenberg U Sputtered Iridium Oxide Films as Charge Injection Material for Functional Electrostimulation Journal of The Electrochemical Society 151 E226 E237 2004
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Premisecitation
Moisture absorption in polyimide results in stresses associated with delamination between metal/polyimide and polyimide/polyimide interfaces.
“However, polyimide is known to absorb moisture, causing hydrolytic attack of the substrate material, and resulting in stresses associated with delamination between metal/polyimide and polyimide/polyimide interfaces [7] [21] [22].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [7] Weiland JD Liu W Humayun MS Retinal Prosthesis Annual Review of Biomedical Engineering 7 361 401 2005 16004575 source
Not checked in the free run.
- cites [21] Xu T Farris RJ Stresses Associated with Diffusion in Polyimide and Polyacrylics Films Journal of Applied Polymer Science 99 2661 2670 2006
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [22] Edell DJ Farrell B Implantable devices having a liquid crystal polymer substrate U. S. P. a. T. Office United States of America 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationreasoning
The 1 pH unit difference of the solution between the 100% nitrogen gas and the oxygen/carbon dioxide gas mixture does not account for the 0.3V difference between the lower voltage window limit.
“The 1 pH unit different of the solution between the 100% nitrogen gas and the oxygen/carbon dioxide gas mixture does not account for the 0.3V difference between the lower voltage window limit.”
Quote in the paperSubstantial claim - Study design and conductown data
The 10mA current amplitude was chosen because preliminary working electrode voltage measurements indicated that the working electrode was operating within the water window.
“The 10mA current amplitude was chosen since preliminary working electrode voltage measurements indicated that the working electrode was operating within the water window, and 10mA is ten times higher than the normal amplitude used for peripheral nerve electrodes.”
Quote in the paperSubstantial claim - Interpretationown data
The peaks indicated an active Ir(IV) + e− ⇋ Ir(III) charge transfer mechanism.
“However, these peaks indicated an active Ir(IV) + e− ⇋ Ir(III) charge transfer mechanism.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [1] Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot IEEE Transactions on Rehabilitation Engineering 10 260 279 2002 12611364 source
Not checked in the free run.
- cites [2] Degnan GG Wind TC Jones EV Edlich RF Functional Electrical Stimulation in Tetraplegic Patients to Restore Hand Function Journal of Long-term effects of Medical Implants 12 175 188 2002 12545943 source
Not checked in the free run.
- Premisecitation
Polyimide absorbs moisture, causing hydrolytic attack of the substrate material.
“However, polyimide is known to absorb moisture, causing hydrolytic attack of the substrate material, and resulting in stresses associated with delamination between metal/polyimide and polyimide/polyimide interfaces [7] [21] [22].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [7] Weiland JD Liu W Humayun MS Retinal Prosthesis Annual Review of Biomedical Engineering 7 361 401 2005 16004575 source
Not checked in the free run.
- cites [21] Xu T Farris RJ Stresses Associated with Diffusion in Polyimide and Polyacrylics Films Journal of Applied Polymer Science 99 2661 2670 2006
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [22] Edell DJ Farrell B Implantable devices having a liquid crystal polymer substrate U. S. P. a. T. Office United States of America 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- Interpretationown data
Sputtered iridium oxide on LCP substrate is a feasible long-term implantable iridium oxide electrode.
“The absence of delamination and very low percentage of iridium material loss to solution suggest that sputtered iridium oxide on LCP substrate is a feasible long-term implantable iridium oxide electrode.”
Quote in the paperSubstantial claim - Interpretationown data
The large charge injection capacity of the electrodes can be attributed to a large surface area generated by the amorphous and sponge-like morphology.
“The large charge injection capacity of the electrodes can be attributed to a large surface area generated by the amorphous and sponge-like morphology (Fig. 5e, f).”
Quote in the paperSubstantial claim - Premisecitation
The faradic charge transfer mechanism is a markedly slower process.
“The faradic charge transfer mechanism is a markedly slower process, shown by variable CV voltage sweep rates in other studies [8].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [8] Moller AR History of Cochlear Implants and Auditory Brainstem Implants Advances in Oto-rhino-laryngology 64 1 10 2006 16891833 source
Not checked in the free run.
- Interpretationreasoning
The continuity and electrical conduction of the iridium oxide film was interrupted most likely by stresses in the film during the lamination process performed at around 300°C.
“The continuity and electrical conduction of the iridium oxide film was interrupted most likely by the stresses in the film during the lamination process performed at around 300°C.”
Quote in the paperSubstantial claim - Interpretationreasoning
Charge imbalance will probably occur in vivo when the electrodes operate at voltages less than −0.3V SCE.
“It is also probable that charge imbalance will also occur in vivo as well when the electrodes operate at voltages less than −0.3V SCE.”
Quote in the paperSubstantial claim - Interpretationreasoning
Charge imbalance indicates the generation of potentially harmful chemical species that diffuse away from the electrode surface into the surrounding environment.
“This charge imbalance indicates the generation of potentially harmful chemical species that diffuse away from the electrode surface into the surrounding environment.”
Quote in the paperSubstantial claim- Rests on an assumption: Charge imbalance is associated with net charge consumption for or output from electrochemical reactions that are not reversed.
- Premisecitation
The reduction of oxygen is known to produce superoxides, peroxides, and peroxide ions that can initiate free radical reactions and result in tissue damage.
“The reduction of oxygen is known to produce superoxides, peroxides, and peroxide ions that can initiate free radical reactions and result in tissue damage [10] [11].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [10] Weiland JD Anderson DJ Humayun MS In Vitro Electrical Properties for Iridium Oxide Versus Titanium Nitride Stimulating Electrodes IEEE Transactions on Neural Systems and Rehabilitation Engineering 40 1574 1579 2002 12549739 source
Not checked in the free run.
- cites [11] Cogan SF Plante TD Ehrlich J Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes 26th Annual International Conference of the IEEE EMBS 2004 4153 4156 PMC2709748 17271216 source
Not checked in the free run.
- Interpretationreasoning
Oxygen reduction is the most probable reason behind the charge imbalance resulting from the introduction of the oxygen/carbon dioxide gas mixture.
“With the charge imbalance resulting from the introduction of the oxygen/carbon dioxide gas mixture, oxygen reduction is the most probable reason behind the charge imbalance.”
- Also stated in: § V. Conclusions · plain
“However, significant charge imbalance occurred within the water window voltage range attributed to faradic transfer reactions associated with oxygen reduction.”
Quote in the paperSubstantial claim - Also stated in: § V. Conclusions · plain
- Interpretationreasoning
The large discrepancy between cathodic charge storage capacity and charge injection limit is most likely due to the rate of charge injection and withdrawal during the analyses.
“The large discrepancy between the cathodic charge storage capacity (33.7 +/−10.1mC/cm2) and the charge injection limit (4.6 +/− 1.0 mC/cm2) is most likely due to the rate of charge injection and withdrawal during the analyses.”
Quote in the paperNot a substantial claim - Interpretationreasoning
If recharacterization to a −0.3V SCE charge injection limit is required, larger electrodes or larger electrode arrays would be required to pass the same amount of current in tissue.
“If the recharacterization is required, larger electrodes or larger electrode arrays would be required to pass the same amount of current in tissue.”
Quote in the paperSubstantial claim - Interpretationown data
The increase in CSCc indicates further activation during the course of the stimulation and soak protocol.
“indicating further activation during the course of the stimulation and soak protocol.”
Quote in the paperNot a substantial claim
- The quote only partly states this claim. 1
- May combine two claims. 22
- Classified as probably not a substantial claim. 14
The quote only partly states this claim.
Findingown dataSEM analysis showed no delamination of the sputtered iridium oxide on LCP electrodes after long-term stimulation and accelerated soak.
“Scanning electron micrograph (SEM) analysis showed no delamination and approximately 1% of electrode material was lost to the bicarbonate buffer.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAt lower frequencies, the impedance increased and was strongly dependent on frequency.
“At lower frequencies, the impedance increased and was strongly dependent on frequency.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSome laminated electrodes showed cracks traversing the traces prevented electrical continuity.
“Some laminated electrodes showed cracks traversing the traces prevented electrical continuity.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe charge storage capacity of the iridium oxide electrodes here was observed to be larger than the charge storage capacity obtained with activated iridium oxide films reported by Agnew et al [4] and sputtered iridium oxide films reported by Cogan et al. [1].
“The charge storage capacity of the iridium oxide electrodes here was observed to be larger than the charge storage capacity obtained with activated iridium oxide films reported by Agnew et al [4] and sputtered iridium oxide films reported by Cogan et al. [1]”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [4] Wichmann T DeLong MR Deep Brain Stimulation for Neurologic and Neuropsychiatric Disorders Neuron 52 197 204 October 2006 17015236 source
Not checked in the free run.
- cites [1] Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot IEEE Transactions on Rehabilitation Engineering 10 260 279 2002 12611364 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataVisual inspection of the electrode under a light microscope showed an iridescent grey color to the electrode with no visible cracking in the film.
“Visual inspection of the electrode under a light microscope showed an iridescent grey color to the electrode with no visible cracking in the film.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFitting the Bode plot with a constant phase element model with series resistance (Rs + [(j ω)P C]−1) revealed an Rs of 2.72 Ω/cm2, a C of 6.55mF/cm2, and exponent P of 0.8 with all variables having errors of less than 2%.
“However, fitting the Bode plot with a constant phase element model with series resistance (Rs + [(j ω)P C]−1) revealed an Rs of 2.72 Ω/cm2, a C of 6.55mF/cm2, and exponent P of 0.8 with all variables having errors of less than 2%.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe post-cathodic working electrode voltages were plotted with respect to the amount of charge injected into the electrode, and a linear relationship was observed (R2 = 0.96).
“The post-cathodic working electrode voltages were plotted with respect to the amount of charge injected into the electrode, and a linear relationship was observed (R2 = 0.96) (Fig. 4b).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataAdditional voltage ramps sweeping from −0.645V to 0.755V SCE (−0.6V to 0.8V Ag|AgCl) that has been consistently used in literature for electrode characterization also yielded minimal charge loss, indicating that the water window was −0.64V to 0.75V SCE.
“Additional voltage ramps sweeping from −0.645V to 0.755V SCE (−0.6V to 0.8V Ag|AgCl) that has been consistently used in literature for electrode characterization also yielded minimal charge loss, indicating that the water window was −0.64V to 0.75V SCE.”
Quote in the paperSubstantial claimMay combine two claims.
Findingnone givenDespite the large CSCc variation among the electrodes, the variability of the CSCc of each individual electrode was very small.
“Despite the large CSCc variation among the electrodes, the variability of the CSCc of each individual electrode was very small.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataUnder oxygen and carbon dioxide conditions, only two peaks were observed: an oxidation peak at +0.45 V SCE and a reduction peak at −0.4 V SCE.
“Under oxygen and carbon dioxide conditions, only two peaks were observed: an oxidation peak at +0.45 V SCE and a reduction peak at −0.4 V SCE (arrows, Fig. 3a).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe variation of the CSCc for each individual electrode was also observed to be small under oxygen/carbon dioxide conditions.
“The variation of the CSCc for each individual electrode was also observed to be small under oxygen/carbon dioxide conditions.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenThe peaks were not as well defined compared to the CVs of activated iridium oxide in et al [1].
“These peaks were not as well defined compared to the CVs of activated iridium oxide in et al [1].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [1] Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot IEEE Transactions on Rehabilitation Engineering 10 260 279 2002 12611364 source
Not checked in the free run.
May combine two claims.
Findingown dataThe impedances were frequency-independent, between 290Ω and 500Ω (average |Z| = 390 +/− 80Ω/cm2) (N = 7), for frequencies higher than 1kHz.
“The impedances were frequency-independent, between 290Ω and 500Ω (average |Z| = 390 +/− 80Ω/cm2) (N = 7), for frequencies higher than 1kHz.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataCyclic voltammetry analysis showed stable CV traces of the iridium oxide electrodes but with an additional oxidation peak at 0.1V SCE appearing approximately halfway into the stimulation protocol and continuing to grow throughout the remainder of the stimulation and soak protocol.
“Cyclic voltammetry analysis showed stable CV traces of the iridium oxide electrodes but with an additional oxidation peak at 0.1V SCE appearing approximately halfway into the stimulation protocol and continuing to grow throughout the remainder of the stimulation and soak protocol.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
FindingcitationThe frequency dependence of the sputtered iridium oxide film impedance was also similar to other activated and electrodeposited iridium oxide films [3].
“The frequency dependence of the sputtered iridium oxide film impedance was also similar to other activated and electrodeposited iridium oxide films [3].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [3] van Balken MR Vergunst H Bemelmans BL The Use of Electrical Devices for the Treatment of Bladder Dysfunction: a review of methods The Journal of Urology 172 846 851 2004 15310981
Not checked in the free run.
May combine two claims.
Findingown dataThe voltage excursion of the working electrode, as defined in Fig. 4a, seemed to vary greatly during the course of the stimulation procedure, averaging 0.125V with a standard deviation of 0.050V but did not drift with time.
“The voltage excursion of the working electrode, as defined in Fig. 4a, seemed to vary greatly during the course of the stimulation procedure, averaging 0.125V with a standard deviation of 0.050V but did not drift with time.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataElectrochemical testing of the sputtered iridium oxide films on liquid crystal polymer showed that the cathodic charge storage capacity and the injectable charge limit were higher than those obtained for activated and electrodeposited iridium oxide films [4] [5] [6] [7].
“Electrochemical testing of the sputtered iridium oxide films on liquid crystal polymer showed that the cathodic charge storage capacity and the injectable charge limit were higher than those obtained for activated and electrodeposited iridium oxide films [4] [5] [6] [7].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [4] Wichmann T DeLong MR Deep Brain Stimulation for Neurologic and Neuropsychiatric Disorders Neuron 52 197 204 October 2006 17015236 source
Not checked in the free run.
- cites [5] Theodore WH Brain Stimulation for Epilepsy Nature Clinical Practice Neurology 1 64 65 2005 16932495 source
Not checked in the free run.
- cites [6] Wang GJ Yang J Volkow ND Telang F Ma Y Zhu W Wong CT Tomasi D Thanos PK Fowler JS Gastric stimulation in obese subjects activates the hippocampus and other regions involved in brain reward circuitry Proceedings of the National Academy of Sciences of the United States 103 15641 15645 2006 PMC1592230 17023542 source
Not checked in the free run.
May combine two claims.
Findingown dataImpedance testing of the sputtered iridium oxide films also revealed Bode plots with slightly higher impedances compared the sputtered iridium oxide films reported by Cogan et al. [1] and Slavcheva et al [2], but lower compared to electrodeposited and activated iridium oxide films [3].
“Impedance testing of the sputtered iridium oxide films also revealed Bode plots with slightly higher impedances compared the sputtered iridium oxide films reported by Cogan et al. [1] and Slavcheva et al [2], but lower compared to electrodeposited and activated iridium oxide films [3].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [1] Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot IEEE Transactions on Rehabilitation Engineering 10 260 279 2002 12611364 source
Not checked in the free run.
- cites [2] Degnan GG Wind TC Jones EV Edlich RF Functional Electrical Stimulation in Tetraplegic Patients to Restore Hand Function Journal of Long-term effects of Medical Implants 12 175 188 2002 12545943 source
Not checked in the free run.
- cites [3] van Balken MR Vergunst H Bemelmans BL The Use of Electrical Devices for the Treatment of Bladder Dysfunction: a review of methods The Journal of Urology 172 846 851 2004 15310981 source
Not checked in the free run.
May combine two claims.
PremisecitationSputtered iridium oxide films are mechanically stable and can resist corrosion and degradation.
“In contrast, sputtered iridium oxide films have been shown to be mechanically stable and can resist corrosion and degradation [11].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [11] Cogan SF Plante TD Ehrlich J Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes 26th Annual International Conference of the IEEE EMBS 2004 4153 4156 PMC2709748 17271216 source
Not checked in the free run.
Classified as probably not a substantial claim.
PremisecitationIridium oxide has been used to increase the amount of total charge that can be delivered for electrical stimulation.
“Iridium oxide has been used to increase the amount of total charge that can be delivered for electrical stimulation [9].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [9] Agnew WF McCreery DB Neural Prostheses: Fundamental Studies Englewood Cliffs, NJ Prentice Hall 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationFor iridium oxide films made from sputtered iridium metal, charge injection limits have been reported to be 4-6 mC/cm2.
“For iridium oxide films made from sputtered iridium metal, charge injection limits have been reported to be 4–6mC/cm2 [10].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Weiland JD Anderson DJ Humayun MS In Vitro Electrical Properties for Iridium Oxide Versus Titanium Nitride Stimulating Electrodes IEEE Transactions on Neural Systems and Rehabilitation Engineering 40 1574 1579 2002 12549739 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSputtered iridium oxide films have been deposited onto platinum and silicon substrates.
“Sputtered iridium oxide films have been deposited onto platinum [11] and silicon [16] substrates.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [11] Cogan SF Plante TD Ehrlich J Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes 26th Annual International Conference of the IEEE EMBS 2004 4153 4156 PMC2709748 17271216 source
Not checked in the free run.
- cites [16] Tanghe SJ Najafi K Wise KD A Planar IrO Multichannel Stimulating Electrode for Use in Neural Prostheses Sensors and Actuators, B1 1 464 467 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
PremisecitationCharge injection limits for iridium oxide films made from iridium wire are +/− 2 mC/cm2 for anodic-first and −/+ 1 mC/cm2 for cathodic-first current waveforms.
“Charge injection limits for iridium oxide films made from iridium wire have been reported to be +/− 2 mC/cm2 and −/+ 1 mC/cm2 for anodic-first and cathodic-first current waveforms, respectively and are significantly higher than those reported for stainless steel or platinum [9].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Agnew WF McCreery DB Neural Prostheses: Fundamental Studies Englewood Cliffs, NJ Prentice Hall 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationCharge injection limits for iridium oxide films made from iridium wire are significantly higher than those reported for stainless steel or platinum.
“Charge injection limits for iridium oxide films made from iridium wire have been reported to be +/− 2 mC/cm2 and −/+ 1 mC/cm2 for anodic-first and cathodic-first current waveforms, respectively and are significantly higher than those reported for stainless steel or platinum [9].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Agnew WF McCreery DB Neural Prostheses: Fundamental Studies Englewood Cliffs, NJ Prentice Hall 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
PremisecitationActivated iridium oxide films have uneven growth of the oxide layer that introduces strain and can result in a tendency to delaminate from the underlying iridium metal.
“For activated iridium oxide films, uneven growth of the oxide layer introduces strain in the oxide layer and can result in the tendency for the activated iridium oxide film to delaminate from the underlying iridium metal [12] [13].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [12] Slavcheva E Vitushinsky R Mokwa W Schnakenberg U Sputtered Iridium Oxide Films as Charge Injection Material for Functional Electrostimulation Journal of The Electrochemical Society 151 E226 E237 2004
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [13] Cogan SF Guzelian AA Agnew WF Yuen TGH McCreery DB Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation Journal of Neuroscience Methods 137 141 150 2004 15262054 source
Not checked in the free run.
May combine two claims.
PremisecitationElectrodeposited iridium oxide adheres poorly with its underlying substrate, resulting in a tendency to delaminate while under operation, which shortens the electrode lifespan.
“Electrodeposited iridium oxide has been reported to adhere poorly with its underlying substrate, resulting in a tendency, to delaminate while under operation, which shortens the electrode lifespan [14] [15].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [14] Marzouk SAM Improved Electrodeposited Iridium Oxide pH Sensor Fabricated on Etched Titanium Substrates Analytical Chemistry 75 1258 1266 February 19 2003 12659184 source
Not checked in the free run.
- cites [15] Mailley SC Hyland M Mailley P McLaughlin JM McAdams ET Electrochemical and Structural Characterizations of Electrodeposited Iridium Oxide Thin-film Electrodes Applied to Neurostimulating Electrical Signal Materials Science and Engineering 21 167 175 2002 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPolyimide is a biocompatible polymer.
“Polyimide, a biocompatible polymer [17] [18], has been a substrate of choice for many electrodes with electro-deposited iridium metal layers and iridium films [19] [20].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [17] Metz S Bertsch A Bertrand D Renauld P Flexible Polyimide probes with microelectrodes and embedded microfluidic channels for simultaneous drug delivery and multi-channel monitoring of bioelectric activity Biosensors and Bioelectronics 19 1309 1318 2004 15046764 source
Not checked in the free run.
- cites [18] Stieglitz T Beutel H Schuettler M Meyer JU Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces Biomedical Microdevices 2 283 294 2000
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationSimilar multiple peaks had been reported by Cogan et al. [1] and Slavcheva et al. [2].
“Similar multiple peaks had been reported by Cogan et al. [1] and Slavcheva et al. [2].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [1] Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot IEEE Transactions on Rehabilitation Engineering 10 260 279 2002 12611364 source
Not checked in the free run.
- cites [2] Degnan GG Wind TC Jones EV Edlich RF Functional Electrical Stimulation in Tetraplegic Patients to Restore Hand Function Journal of Long-term effects of Medical Implants 12 175 188 2002 12545943 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationPolyimide has been a substrate of choice for many electrodes with electro-deposited iridium metal layers and iridium films.
“Polyimide, a biocompatible polymer [17] [18], has been a substrate of choice for many electrodes with electro-deposited iridium metal layers and iridium films [19] [20].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [19] Meyer RD Cogan SF Nguyen TH Rauh RD Electrodeposited Iridium Oxide for Neural Stimulation and Recording Electrodes IEEE Transactions on Neural Systems and Rehabilitation Engineering 9 2 11 March 2001 11482359 source
Not checked in the free run.
- cites [20] Stieglitz T Beutel H Keller R Schuettler M Meyer J-U Flexible, Polyimide-Based Neural Interfaces Seventh International Conference on Microelectronics for Neural, Fuzzy and Bio-Inspired Systems 1999
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
PremisecitationLCP is a biocompatible and flexible substrate.
“Liquid crystal polymer (LCP) is also biocompatible and flexible substrate [22] [23] [24].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [22] Edell DJ Farrell B Implantable devices having a liquid crystal polymer substrate U. S. P. a. T. Office United States of America 2003
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [23] Lee CJ Oh SJ Song JK Kim SJ Neural signal recording using microelectrode arrays fabricated on liquid crystal polymer material Materials Science and Engineering: C 24 265 268 January 2004
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [24] McCamley MK Artenstein AW Crawford GP Woltman SJ Crawford GP Jay GD Liquid Crystal Biosensors: A New Approach to Medical Diagnostic Devices Liquid Crystals: Frontiers in Biomedical Applications Hackensack, NJ World Scientific 2007
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
Probably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenThe oxidation peak at around 0.1V SCE arose from the rearrangement of the iridium oxide lattice structure.
“We postulate that this peak arose from the rearrangement of the iridium oxide lattice structure.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThe sputtered iridium oxide film on LCP was robust and could provide a good solution for stimulation electrodes that require a flexible substrate and large charge injection capability.
“Finally, the sputtered iridium oxide film on LCP was robust and could provide a good solution for stimulation electrodes that require a flexible substrate and large charge injection capability [12] [13].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [12] Slavcheva E Vitushinsky R Mokwa W Schnakenberg U Sputtered Iridium Oxide Films as Charge Injection Material for Functional Electrostimulation Journal of The Electrochemical Society 151 E226 E237 2004
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
- cites [13] Cogan SF Guzelian AA Agnew WF Yuen TGH McCreery DB Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation Journal of Neuroscience Methods 137 141 150 2004 15262054 source
Not checked in the free run.
May combine two claims.
Interpretationown dataThe LCP substrate itself was observed to be rough and could have contributed to both the amorphous and spongy film structure and the lack of delamination.
“The LCP substrate itself was observed to be rough and could have contributed to both the amorphous and spongy film structure and the lack of delamination.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationDiscrepancies in CVs have been observed with exposure of different faces of the platinum crystal lattice in solution.
“Such discrepancies have been observed in CVs involving the exposure of different faces of the platinum crystal lattice in solution [9].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [9] Agnew WF McCreery DB Neural Prostheses: Fundamental Studies Englewood Cliffs, NJ Prentice Hall 1990
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
May combine two claims.
InterpretationreasoningOperating the electrode at voltages that avoid charge imbalance allows the charge injected into the electrode to be recovered and minimizes the possibility of generating harmful species in biological tissue.
“By operating the electrode at voltages that avoid charge imbalance, the charge injected into the electrode can be recovered and the possibility of generating harmful species in biological tissue can be minimized.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningThe electrochemical properties of the electrode should not change with changes to the fabrication process, since the electrochemistry occurs at the iridium oxide.
“However, the electrochemical properties of the electrode should not change, since the electrochemistry occurs at the iridium oxide.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
LCP is chemically inert, facilitating long-term biological implantation.
“LCP is also chemically inert facilitating long-term biological implantation.”
The paper relies on this without support.
10mA is ten times higher than the normal amplitude used for peripheral nerve electrodes.
“The 10mA current amplitude was chosen since preliminary working electrode voltage measurements indicated that the working electrode was operating within the water window, and 10mA is ten times higher than the normal amplitude used for peripheral nerve electrodes.”
The paper relies on this without support.
The accelerated soak and stimulation protocol at 67°C for approximately 700 hours is equivalent to eight months at human body temperature.
“equivalent stimulation and soak time of eight months at human body temperature”
The paper relies on this without support.
Charge imbalance is associated with net charge consumption for or output from electrochemical reactions that are not reversed.
“Charge imbalance is associated with net charge consumption for or output from electrochemical reactions that are not reversed.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Iridium oxide coating is corrosion resistant and exhibits, in general, good biocompatibility and higher charge delivery capacity, but its exact properties depend greatly on the fabrication method, which affects the stability of the coating.
“Iridium oxide (IrO2) coating is corrosion resistant and exhibits, in general, good biocompatibility and higher charge delivery capacity (Wang et al. 2009). Nevertheless, its exact properties depend greatly on the fabrication method – activated, sputtered and electrodeposited variants have been reported – which, in turn, affects the stability of the coating.”
From Realizing flexible bioelectronic medicines for accessing the peripheral nerves – technology considerations (2018) · cites this paper as Wang et al. 2009
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Cogan is cited as reference 1, which lists different authors
MinorThe cited reference [1] is by Lyons and colleagues, not Cogan. This makes the attribution of the reported electrochemical peaks inconsistent with the paper's reference list.
“Similar multiple peaks had been reported by Cogan et al. [1] and Slavcheva et al. [2].”
“1. Lyons GM Sinkjaer T Burridge JH Wilcox DJ A Review on Portable FES-based Neural Orthoses for the Correction of Drop Foot”
Cites
(1996) · Choice Reviews Online · doi:10.5860/choice.33-4512 · Cited by 640 · see citing works
Brain stimulation for epilepsy
Theodore, W. H., & Fisher, R. S. (2004) · The Lancet Neurology · doi:10.1016/s1474-4422(03)00664-1 · Cited by 416 · see citing works
Sawyer, D. T., Sobkowiak, A., & Roberts, J. L. (1995) · Cited by 409 · see citing works
Functionally selective peripheral nerve stimulation with a flat interface nerve electrode
Tyler, D. J., & Durand, D. M. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806840 · Cited by 398 · cited 1× · see citing works
“Finally, the sputtered iridium oxide film on the LCP was robust and could provide a good solution for stimulation electrodes that require a flexible substrate and large charge injection capability [33], [ 34 ].”
MethodSelf-citationIn vitro electrical properties for iridium oxide versus titanium nitride stimulating electrodes
Weiland, J. D., Anderson, D. J., & Humayun, M. S. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.805487 · Cited by 382 · see citing works
Weiland, J. D., Liu, W., & Humayun, M. S. (2005) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.7.060804.100435 · Cited by 371 · see citing works
Micromachined, Polyimide-Based Devices for Flexible Neural Interfaces
Stieglitz, T., Beutel, H., Schuettler, M., & Meyer, J.-U. (2000) · Biomedical Microdevices · doi:10.1023/a:1009955222114 · Cited by 343 · see citing works
Deep Brain Stimulation for Neurologic and Neuropsychiatric Disorders
Wichmann, T., & DeLong, M. R. (2006) · Neuron · doi:10.1016/j.neuron.2006.09.022 · Cited by 331 · see citing works
Chemical kinetics and catalysis
(2002) · Focus on Catalysts · doi:10.1016/s1351-4180(02)00754-7 · Cited by 326 · see citing works
Neural prostheses : fundamental studies
Agnew, W. F., & McCreery, D. B. (1990) · Medical Entomology and Zoology · Cited by 316 · see citing works
A review of portable FES-based neural orthoses for the correction of drop foot
Lyons, G. M., Sinkjaer, T., Burridge, J. H., & Wilcox, D. J. (2002) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2002.806832 · Cited by 277 · cited 1× · see citing works
“In functional electrical stimulation, electrical stimulation has been used effectively in the treatment of foot drop [1], the restoration”
Electrodeposited iridium oxide for neural stimulation and recording electrodes
Meyer, R. D., Cogan, S. F., Nguyen, T. H., & Rauh, R. D. (2001) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/7333.918271 · Cited by 255 · see citing works
Selective activation of small motor axons by quasitrapezoidal current pulses
Fang, Z.-P., & Mortimer, J. T. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.76383 · Cited by 222 · see citing works
Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation
Cogan, S. F., Guzelian, A. A., Agnew, W. F., Yuen, T. G. H., & McCreery, D. B. (2004) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.02.019 · Cited by 210 · see citing works
Hydrogen adsorption and oxide formation on platinum single crystal electrodes
Yamamoto, K., Kolb, D. M., Kötz, R., & Lehmpfuhl, G. (1979) · Journal of Electroanalytical Chemistry and Interfacial Electrochemistry · doi:10.1016/s0022-0728(79)80380-0 · Cited by 191 · see citing works
Weiland, J. D., & Humayun, M. S. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2314733 · Cited by 186 · see citing works
Liquid crystal polymer (LCP) for MEMS: processes and applications
Wang, X., Engel, J. M., & Liu, C. (2003) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/13/5/314 · Cited by 170 · see citing works
Self-citationWang, G.-J., Yang, J., Volkow, N. D., Telang, F., Ma, Y., Zhu, W., Wong, C. T., Tomasi, D., Thanos, P. K., & Fowler, J. S. (2006) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0601977103 · Cited by 168 · see citing works
Metz, S., Bertsch, A., Bertrand, D., & Renaud, P. (2004) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2003.11.021 · Cited by 167 · see citing works
Cogan, S. F., Troyk, P. R., Ehrlich, J., & Plante, T. D. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2005.851503 · Cited by 162 · see citing works
Sputtered Iridium Oxide Films as Charge Injection Material for Functional Electrostimulation
Slavcheva, E., Vitushinsky, R., Mokwa, W., & Schnakenberg, U. (2004) · Journal of The Electrochemical Society · doi:10.1149/1.1747881 · Cited by 154 · see citing works
Sputtered iridium oxide films (SIROFs) for low-impedance neural stimulation and recording electrodes
Cogan, S. F., Plante, T. D., & Ehrlich, J. (2005) · The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2004.1404158 · Cited by 138 · see citing works
Improved Electrodeposited Iridium Oxide pH Sensor Fabricated on Etched Titanium Substrates
Marzouk, S. A. M. (2003) · Analytical Chemistry · doi:10.1021/ac0261404 · Cited by 136 · see citing works
THE USE OF ELECTRICAL DEVICES FOR THE TREATMENT OF BLADDER DYSFUNCTION: A REVIEW OF METHODS
van BALKEN, M. R., VERGUNST, H., & BEMELMANS, B. L. H. (2004) · Journal of Urology · doi:10.1097/01.ju.0000134418.21959.98 · Cited by 121 · see citing works
Mailley, S. C., Hyland, M., Mailley, P., McLaughlin, J. M., & McAdams, E. T. (2002) · Materials Science and Engineering: C · doi:10.1016/s0928-4931(02)00098-x · Cited by 84 · see citing works
The Role of Oxygen Reduction in Electrical Stimulation of Neural Tissue
Morton, S. L., Daroux, M. L., & Mortimer, J. T. (1994) · Journal of The Electrochemical Society · doi:10.1149/1.2054671 · Cited by 55 · see citing works
Neural signal recording using microelectrode arrays fabricated on liquid crystal polymer material
Lee, C. J., Oh, S., Song, J. K., & Kim, S. J. (2003) · Materials Science and Engineering C · doi:10.1016/j.msec.2003.09.143 · Cited by 48 · see citing works
A planar IrO multichannel stimulating electrode for use in neural prostheses
Tanghe, S. J., Najafi, K., & Wise, K. D. (1990) · Sensors and Actuators B: Chemical · doi:10.1016/0925-4005(90)80250-4 · Cited by 47 · see citing works
History of Cochlear Implants and Auditory Brainstem Implants
Møller, A. R. (2006) · Advances in Oto-Rhino-Laryngology · doi:10.1159/000094455 · Cited by 37 · see citing works
Brain stimulation for epilepsy
Theodore, W. H. (2005) · Nature Clinical Practice Neurology · doi:10.1038/ncpneuro0051 · Cited by 23 · see citing works
Flexible, polyimide-based neural interfaces
Stieglitz, T., Beutel, H., Keller, R., Schüettler, M., & Meyer, J.-U. (1999) · International Conference on Microelectronics for Neural, Fuzzy and Bio-Inspired Systems (MicroNeuro) · doi:10.1109/mn.1999.758853 · Cited by 20 · see citing works
Functional Electrical Stimulation in Tetraplegic Patients to Restore Hand Function
Edlich, R., Degnan, G. G., & Wind, T. C. (2002) · Journal of Long-Term Effects of Medical Implants · doi:10.1615/jlongtermeffmedimplants.v12.i3.40 · Cited by 12 · see citing works
Comprehensive Cyclic Voltammetry Characterization of AIROF Microelectrodes
Hu, Z., Troyk, P. R., & Cogan, S. F. (2005) · 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference · doi:10.1109/iembs.2005.1615662 · Cited by 12 · see citing works
(1993) · Choice Reviews Online · doi:10.5860/choice.31-0958 · Cited by 11 · see citing works
Stresses associated with diffusion in polyimide and polyacrylics films
Xu, T., & Farris, R. J. (2005) · Journal of Applied Polymer Science · doi:10.1002/app.22807 · Cited by 9 · see citing works
Liquid Crystal Biosensors: A New Approach to Medical Diagnostic Devices
McCamley, M. K., Artenstein, A. W., & Crawford, G. P. (2007) · WORLD SCIENTIFIC eBooks · doi:10.1142/9789812778871_0004 · Cited by 2 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- edell, 2003, implantable devices having a liquid crystal polymer substrate, U S Patent 6 643552
- mccamley, 2007, Proc Liquid Crystals Frontiers in Biomedical Applications
- hayden, 2002, new liquid crystal polymer flex circuits to meet demanding reliability and end-use applications, Int Conf Adv Packag Syst
Cited by
Sorted by citations
Hassler, C., Boretius, T., & Stieglitz, T. (2011) · Journal of Polymer Science Part B: Polymer Physics · doi:10.1002/polb.22169 · Cited by 501 · see citing works
Emerging Modalities and Implantable Technologies for Neuromodulation
Won, S. M., Song, E., Reeder, J. T., & Rogers, J. A. (2020) · Cell · doi:10.1016/j.cell.2020.02.054 · Cited by 293 · see citing works
Recent Advances in Materials, Devices, and Systems for Neural Interfaces
Won, S. M., Song, E., Zhao, J., Li, J., Rivnay, J., & Rogers, J. A. (2018) · Advanced Materials · doi:10.1002/adma.201800534 · Cited by 206 · see citing works
Polymer integration for packaging of implantable sensors
Qin, Y., Howlader, M. M. R., Deen, M. J., Haddara, Y. M., & Selvaganapathy, P. R. (2014) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2014.05.063 · Cited by 163 · see citing works
The future of neuroscience: flexible and wireless implantable neural electronics
McGlynn, E., Nabaei, V., Ren, E., Galeote‐Checa, G., Das, R., Curia, G., & Heidari, H. (2021) · Advanced Science · doi:10.1002/advs.202002693 · Cited by 115 · see citing works
Monolithic Encapsulation of Implantable Neuroprosthetic Devices Using Liquid Crystal Polymers
Lee, S. W., Min, K. S., Jeong, J., Kim, J., & Kim, S. J. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2136341 · Cited by 104 · see citing works
Cruz, A. M., Abad, L., Carretero, N. M., Moral‐Vico, J., Fraxedas, J., Lozano, P., Subías, G., Padial, V., Carballo, M., Collazos-Castro, J. E., et al. (2012) · The Journal of Physical Chemistry C · doi:10.1021/jp212275q · Cited by 85 · see citing works
Bedri, A., Li, D., Khurana, R., Bhuwalka, K., & Goel, M. (2020) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3313831.3376869 · Cited by 83 · see citing works
Advances in Materials for Recent Low-Profile Implantable Bioelectronics
Chen, Y., Kim, Y., Tillman, B. W., Yeo, W., & Chun, Y. (2018) · Materials · doi:10.3390/ma11040522 · Cited by 60 · see citing works
Giagka, V., & Serdijn, W. A. (2018) · Bioelectronic Medicine · doi:10.1186/s42234-018-0010-y · Cited by 58 · see citing works
Flexible Liquid Crystal Polymer Technologies from Microwave to Terahertz Frequencies
Zhou, Z., Li, W., Qian, J., Liu, W., Wang, Y., Zhang, X., Guo, Q., Yashchyshyn, Y., Wang, Q., Shi, Y., et al. (2022) · Molecules · doi:10.3390/molecules27041336 · Cited by 58 · see citing works
Lee, Y. J., Kim, H.-J., Do, S. H., Kang, J. Y., & Lee, S. H. (2016) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2016.06.169 · Cited by 54 · see citing works
Maeng, J., Chakraborty, B., Geramifard, N., Kang, T. H., Rihani, R. T., Joshi‐Imre, A., & Cogan, S. F. (2019) · Journal of Biomedical Materials Research Part B Applied Biomaterials · doi:10.1002/jbm.b.34442 · Cited by 51 · see citing works
Liquid crystal polymer (LCP)-based neural prosthetic devices
Gwon, T. M., Kim, C., Shin, S., Park, J. H., Kim, J. H., & Kim, S. J. (2016) · Biomedical Engineering Letters · doi:10.1007/s13534-016-0229-z · Cited by 47 · see citing works
Yin, P., Liu, Y., Xiao, L., & Zhang, C. (2021) · Polymers · doi:10.3390/polym13162834 · Cited by 42 · see citing works
The neural tissue around SU-8 implants: A quantitative in vivo biocompatibility study
Márton, G., Tóth, E. Z., Wittner, L., Fiáth, R., Pinke, D., Orbán, G., Meszéna, D., Pál, I., Győri, E. L., Bereczki, Z., et al. (2020) · Materials Science and Engineering C · doi:10.1016/j.msec.2020.110870 · Cited by 41 · see citing works
Cogan, S. F., Ehrlich, J., Plante, T. D., Gingerich, M. D., & Shire, D. B. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2050690 · Cited by 35 · see citing works
Structural and functional imaging of brains
Liu, Z., Zhu, Y. J., Zhang, L., Jiang, W., Liu, Y., Tang, Q., Cai, X., Li, J., Wang, L., Tao, C.-L., et al. (2022) · Science China Chemistry · doi:10.1007/s11426-022-1408-5 · Cited by 33 · see citing works
Liquid crystal elastomers as substrates for 3D, robust, implantable electronics
Maeng, J., Rihani, R. T., Javed, M., Rajput, J. S., Kim, H., Bouton, I. G., Criss, T. A., Pancrazio, J. J., Black, B. J., & Ware, T. H. (2020) · Journal of Materials Chemistry B · doi:10.1039/d0tb00471e · Cited by 31 · see citing works
Iridium Oxide–Electrodeposited Nanoporous Gold Multielectrode Array with Enhanced Stimulus Efficacy
Kim, Y. H., Kim, G. H., Kim, M. S., & Jung, S.-D. (2016) · Nano Letters · doi:10.1021/acs.nanolett.6b03473 · Cited by 30 · see citing works
Jeong, J., Min, K. S., & Kim, S. J. (2019) · Microelectronic Engineering · doi:10.1016/j.mee.2019.111096 · Cited by 29 · see citing works
A Polymer-Based Multichannel Cochlear Electrode Array
Min, K. S., Oh, S. H., Park, M., Jeong, J., & Kim, S. J. (2014) · Otology & Neurotology · doi:10.1097/mao.0000000000000292 · Cited by 28 · see citing works
Towards smart self-clearing glaucoma drainage device
Park, H., Raffiee, A. H., John, S. W. M., Ardekani, A. M., & Lee, H. (2018) · Microsystems & Nanoengineering · doi:10.1038/s41378-018-0032-3 · Cited by 28 · see citing works
Kang, X., Liu, J., Tian, H., Yang, B. T., NuLi, Y., & Yang, C. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.11.050 · Cited by 27 · see citing works
Materials analyses and electrochemical impedance of implantable metal electrodes
Howlader, M. M. R., Alam, A. U., Sharma, R., & Deen, M. J. (2015) · Physical Chemistry Chemical Physics · doi:10.1039/c4cp05899b · Cited by 25 · see citing works
Novel four-sided neural probe fabricated by a thermal lamination process of polymer films
Shin, S., Kim, J., Jeong, J., Gwon, T. M., Lee, S., & Kim, S. J. (2016) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2016.12.017 · Cited by 24 · see citing works
Development of an iridium-based pH sensor for bioanalytical applications
Bause, S., Decker, M., Gerlach, F., Näther, J., Köster, F., Neubauer, P. B., & Vonau, W. (2017) · Journal of Solid State Electrochemistry · doi:10.1007/s10008-017-3721-1 · Cited by 23 · see citing works
Kang, X., Liu, J., Tian, H., Yang, B. T., NuLi, Y., & Yang, C. (2014) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/24/2/025015 · Cited by 19 · see citing works
LCP MEMS implantable pressure sensor for Intracranial Pressure measurement
Sattayasoonthorn, P., Suthakorn, J., Chamnanvej, S., Miao, J., & Kottapalli, A. G. P. (2013) · IEEE International Conference on Nano/Molecular Medicine and Engineering (NANOMED) · doi:10.1109/nanomed.2013.6766317 · Cited by 17 · see citing works
Nanofibrous PEDOT-Carbon Composite on Flexible Probes for Soft Neural Interfacing
Vajrala, V. S., Saunier, V., Nowak, L. G., Emmanuel Flahaut, B., Bergaud, C., & Maziz, A. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.780197 · Cited by 13 · see citing works
In vitro assessment of long-term reliability of low-cost μΕCoG arrays
Palopoli-Trojani, K., Woods, V., Chiang, C., Trumpis, M., & Viventi, J. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591728 · Cited by 13 · see citing works
On the feasibility of a liquid crystal polymer pressure sensor for intracranial pressure measurement
Sattayasoonthorn, P., Suthakorn, J., & Chamnanvej, S. (2019) · Biomedizinische Technik/Biomedical Engineering · doi:10.1515/bmt-2018-0029 · Cited by 13 · see citing works
Penetrating microelectrode arrays with low-impedance sputtered iridium oxide electrode coatings
Cogan, S. F., Ehrlich, J., Plante, T. D., & Wagenen, R. V. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5335359 · Cited by 11 · see citing works
Material–tissue interfaces in implantable systems
Stieglitz, T., & Schüettler, M. (2013) · Elsevier eBooks · doi:10.1533/9780857096289.1.39 · Cited by 11 · see citing works
Vajrala, V. S., Elkhoury, K., Pautot, S., Bergaud, C., & Maziz, A. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115182 · Cited by 9 · see citing works
Sputtered Iridium Oxide as Electrode Material for Subretinal Stimulation
Haas, J., Rudorf, R., Becker, M. T., Daschner, R., Drzyzga, A., Burkhardt, C. J., & Stett, A. (2020) · Sensors and Materials · doi:10.18494/sam.2020.2903 · Cited by 8 · see citing works
Electrodes for electrical conduction block of the peripheral nerve
Bhadra, N., & Vrabec, T. L. (2015) · Elsevier eBooks · doi:10.1016/b978-0-323-26208-8.00010-8 · Cited by 7 · see citing works
Microfabricated nerve–electrode interfaces in neural prosthetics and neural engineering
Song, Y., Ibrahim, A. M. S., Rabie, A. N., Han, J., & Lin, S. J. (2013) · Biotechnology and Genetic Engineering Reviews · doi:10.1080/02648725.2013.801231 · Cited by 6 · see citing works
Development of IrO2 bio-ink for ink-jet printing application
Hsieh, Y.-C., Wang, H.-Y., Tso, K.-C., Chang, C., Chen, C.-S., Cheng, Y., & Wu, P. (2019) · Ceramics International · doi:10.1016/j.ceramint.2019.05.206 · Cited by 6 · see citing works
Analysis of pressure variations in a low-pressure nickel–hydrogen battery: Part 1
Purushothaman, B., & Wainright, J. S. (2012) · Journal of Power Sources · doi:10.1016/j.jpowsour.2012.01.150 · Cited by 5 · see citing works
Pan, Y., Noda, T., Sasagawa, K., Tokuda, T., Kanda, H., Fujikado, T., & Ohta, J. (2012) · Journal of Physics Conference Series · doi:10.1088/1742-6596/352/1/012005 · Cited by 5 · see citing works
Ramesh, V., Johny, J., Jakobi, J., Stuckert, R., Rehbock, C., & Barcikowski, S. (2024) · ChemPhysChem · doi:10.1002/cphc.202300623 · Cited by 4 · see citing works
Chen, Y., Rasmussen, M., & Scherson, D. A. (2010) · Electrochemical and Solid-State Letters · doi:10.1149/1.3355611 · Cited by 4 · see citing works
Electrical interfaces for recording, stimulation, and sensing
Hess‐Dunning, A., & Zorman, C. A. (2015) · Elsevier eBooks · doi:10.1016/b978-0-323-26208-8.00002-9 · Cited by 4 · see citing works
Supracortical Microstimulation: Advances in Microelectrode Design and In Vivo Validation
Schmitz, C. M., Smith, J. E., Rachinskiy, I., Pesaran, B., Vitale, F., Sommer, M. A., & Viventi, J. (2025) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-103023-072855 · Cited by 3 · see citing works
Developing clinical grade flexible implantable electronics
Liu, Y., Balsamo, D., & Degenaar, P. A. (2022) · Flexible and Printed Electronics · doi:10.1088/2058-8585/aca779 · Cited by 3 · see citing works
Direct Current Block of Peripheral Nerve: Electrode and Waveform Development
Vrabec, T. L. (2016) · OhioLink ETD Center (Ohio Library and Information Network) · Cited by 3 · see citing works
Chen, Q., Huang, N., Tong, C., Liu, P., Que, S., Xue, H., Xiao, X., & Song, E. (2026) · npj Biosensing · doi:10.1038/s44328-026-00109-7 · Cited by 2 · see citing works
Porous Polypropylene Membrane Based pH Sensing for Skin Monitoring
Chawang, K., Chou, S.-C., Bing, S., Wu, P., & Chiao, J. ‐. (2022) · IEEE Access · doi:10.1109/access.2022.3215588 · Cited by 2 · see citing works
Optomechatronics for Biomedical Optical Imaging: An Overview
Cho, H. (2015) · MATEC Web of Conferences · doi:10.1051/matecconf/20153204001 · Cited by 2 · see citing works
Characterization of Spinal Cord Stimulation Electrode for Chronic Implant in Animal Models
Cavalcanti, L. R. C., Filho, G., de Medeiros, R. E., Diniz, H., Damasceno, I. Z., Morya, E., & Pereira, H. S. G. (2023) · International Electronic Conference on Biosensors (IECB) · doi:10.3390/iecb2023-14579 · Cited by 1 · see citing works
Sattayasoonthorn, P., Suthakorn, J., & Chamnanvej, S. (2019) · International Journal of Power Electronics and Drive Systems/International Journal of Electrical and Computer Engineering · doi:10.11591/ijece.v9i5.pp4044-4052 · Cited by 1 · see citing works
Thin Film Coatings as Electrodes in Neuroscience
Khan, S. P., Mian, A., & Newaz, G. M. (2013) · Biological and medical physics series · doi:10.1007/978-94-007-2592-8_9
Boron-Doped Diamond Electrodes for Neural Stimulation
Kim, S. (2014) · OhioLink ETD Center (Ohio Library and Information Network)
Neuroflex: Intraneural and Extraneural Flexible Sensor Architectures for Neural Probing
Patil, A. C., & Thakor, N. V. (2022) · Handbook of Neuroengineering · doi:10.1007/978-981-15-2848-4_16-2
Neuroflex: Intraneural and Extraneural Flexible Sensor Architectures for Neural Probing
Patil, A. C., & Thakor, N. V. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_16
- Authors
- Wang, K., Liu, C.-C., & Durand, D. M.
- Affiliations
- Case Western Reserve University (US)
- ORCID
- Durand, D.M.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2009-01
- Volume
- 56
- Issue
- 1
- Pages
- 6-14
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- PMC
- PMC2738844
- OpenAlex
- W2154069304
- Semantic Scholar
- 50f65e3db048ec40fda75664ddf9220de6d2fc7a
- Cited by
- 57
- References
- 36
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 1 · 2023: 2 · 2022: 6 · 2021: 3 · 2020: 5 · 2019: 5 · 2018: 4 · 2017: 1 · 2016: 6 · 2015: 5 · 2014: 4 · 2013: 5 · 2012: 3
- FWCI
- 1.11
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Neurological Disorders and Stroke (R01NS032845)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
- Life Sciences › Neuroscience › Neurology › Vagus Nerve Stimulation Research
- Keywords
- iridium oxide films, liquid crystal polymer substrate, sputtered thin films, water window, oxygen reduction reaction
- MeSH terms
- Electric Stimulation, Electrochemistry, Electrodes, Electric Stimulation Therapy, Iridium, Microscopy, Electron, Scanning, Oxygen, Polymers, Prosthesis Design
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review