Potential Stability of All-Solid-State Ion-Selective Electrodes Using Conducting Polymers as Ion-to-Electron Transducers
The DOI matches its Crossref record; 34 of the 34 references it cites were found in a registry or an index.
Abstract
Demanding analytical applications such as on-line process analysis and clinical analysis require robust, reliable, and maintenance-free ion sensors of high potential stability. In this work the stability of the electrode potential of all-solid-state ion-selective electrodes using conducting polymers as ion-to-electron transducers is critically evaluated by using chronopotentiometry and electrochemical impedance spectroscopy. This study is focused on the relationship between the potential stability of the electrode and the capacitance of the solid contact where ion-to-electron transduction takes place. The influence of this capacitance on the potential stability of all-solid-state ion-selective electrodes is studied experimentally by using conducting polymer layers of different thickness as solid contacts in potassium ion-selective electrodes based on a solvent polymeric membrane. Because of its excellent environmental stability, the conducting polymer poly(3,4-ethylenedioxythiophene) (PEDOT) is used as a model compound for the solid contact material. Chronopotentiometry is found to be a convenient and fast experimental method to critically evaluate the potential stability of different types of ion-selective electrodes.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Also, the high impedance nature of the ISE-based sensors37 required the use of high-impedance voltage buffers to ensure accurate open voltage measurement as intended.”
From Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis (2016) · cites this paper as [37]
Citations
Cites
Handbook of conducting polymers
Skotheim, T. A., Elsenbaumer, R. L., & Reynolds, J. R. (1986) · M. Dekker eBooks · Cited by 8,079
Poly(alkylenedioxythiophene)s—new, very stable conducting polymers
Heywang, G., & Jonas, F. (1992) · Advanced Materials · doi:10.1002/adma.19920040213 · Cited by 1,027
Cadogan, A., Gao, Z., Lewenstam, A., Ivaska, A., & Diamond, D. (1992) · Analytical Chemistry · doi:10.1021/ac00045a007 · Cited by 428
Coated wire ion-selective electrodes
Cattrall, R. W., & Freiser, H. (1971) · Analytical Chemistry · doi:10.1021/ac60307a032 · Cited by 387
Horvai, G., Graf, E., Toth, K., Pungor, E., & Buck, R. P. (1986) · Analytical Chemistry · doi:10.1021/ac00126a034 · Cited by 235
GUSTAFSSON, J., LIEDBERG, B., & INGANAS, O. (1994) · Solid State Ionics · doi:10.1016/0167-2738(94)90403-0 · Cited by 213
Technical applications for conductive polymers
Jonas, F., & Heywang, G. (1994) · Electrochimica Acta · doi:10.1016/0013-4686(94)e0057-7 · Cited by 202
Single-piece all-solid-state ion-selective electrode
Bobacka, J., Lindfors, T., McCarrick, M., Ivaska, A., & Lewenstam, A. (1995) · Analytical Chemistry · doi:10.1021/ac00116a034 · Cited by 188
Self-citationBobacka, J., McCarrick, M., Lewenstam, A., & Ivaska, A. (1994) · The Analyst · doi:10.1039/an9941901985 · Cited by 175
Self-citationCharacterization of electroactive polymer layers by electrochemical impedance spectroscopy (EIS)
Musiani, M. M. (1990) · Electrochimica Acta · doi:10.1016/0013-4686(90)80023-h · Cited by 148
Research works printed but not linked: 7
- Nikolskii B. P., 1985, Ion-Sel. Electrode Rev., 7, 39
- Buck R. P., 1976, Anal. Chem., 48, 39R
- Momma T., 1995, Sens. Acuators, B, 24, 728
- Pandey P. C., 1998, Sens. Actuators, B, 46, 65
- Glarum S. H., 1987, J. Electrochem. Soc., 134, 147
- Bobacka J., 1991, Synth. Met., 44, 34
- Koryta, J.; Dvorak, J.; Kavan, L. Principles of Electrochemistry , 2nd ed.; Wiley: Chichester, U.K., 1993; p 38.
Cited by
Items 401–500 of 686
3D-Drawn Supports for Ion-Selective Electrodes
Kalisz, J., Węgrzyn, K., Maksymiuk, K., & Michalska, A. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.1c05431 · Cited by 17
Realization of a PEDOT:PSS/Graphene Oxide On-Chip Pseudo-Reference Electrode for Integrated ISFETs
Tintelott, M., Kremers, T., Ingebrandt, S., Pachauri, V., & Vu, X. T. (2022) · Sensors · doi:10.3390/s22082999 · Cited by 17
“It is a polymer–polyelectrolyte complex that offers both ion and electron conductivity with semiconducting and redox-active charge conduction properties [18,19], making it a popular material, often used in organic electrochemical transistors (OECTs) [20], light-emitting diodes (LEDs) [21], biohybrid synapse [22], neur…”
BackgroundSolid-contact ionophore-based electrode for determination of pH in acidic media
Samsonova, E. N., Lutov, V. M., & Mikhelson, K. N. (2008) · Journal of Solid State Electrochemistry · doi:10.1007/s10008-008-0557-8 · Cited by 17
Li, J., & Qin, W. (2019) · Analytica Chimica Acta · doi:10.1016/j.aca.2019.04.003 · Cited by 18
Criscuolo, F., Taurino, I., Dam, V. A. T., Catthoor, F., Zevenbergen, M. A. G., Carrara, S., & Micheli, G. D. (2019) · Sensors · doi:10.3390/s19102260 · Cited by 16
Liu, Y., Wang, X., Zeng, X., Waterhouse, G. I. N., Jiang, X., Zhang, Z., & Yu, L. (2023) · Talanta · doi:10.1016/j.talanta.2023.124436 · Cited by 16
A Review of the Carbon-Based Solid Transducing Layer for Ion-Selective Electrodes
Wang, P., Liu, H., Zhou, S., Chen, L., Yu, S., & Wei, J. (2023) · Molecules · doi:10.3390/molecules28145503 · Cited by 19
“Extensive theoretical and practical evidence demonstrates that a sufficiently large oxidation–reduction or electric double-layer (EDL) capacitance serves as a guarantee for electrode potential stability [90].”
BackgroundLiu, Z.-H., Cai, X. M., Dai, H.-H., Zhao, Y., Gao, Z., Yang, Y., Liu, Y.-Z., Yang, M., Li, M., Li, P., et al. (2024) · Analytical Chemistry · doi:10.1021/acs.analchem.4c00590 · Cited by 15
Perspective of Printed Solid‐State Ion Sensors toward High Sensitivity and Selectivity
Bao, C., & Kim, W. S. (2020) · Advanced Engineering Materials · doi:10.1002/adem.202000116 · Cited by 17
Yewale, R., Damlin, P., Salomäki, M., & Kvarnström, C. (2020) · Materials Today Communications · doi:10.1016/j.mtcomm.2020.101398 · Cited by 18
Bondar, A. V., Keresten, V. M., & Mikhelson, K. N. (2021) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2021.131231 · Cited by 16
Fan, J., Wang, H., Zeng, X., Su, L., & Zhang, X. (2022) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2022.116323 · Cited by 15
Perspective on the coulometric transduction principle for ion-selective electrodes
Bobacka, J. (2024) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2024.135674 · Cited by 15
Self-citationHussein, L. A. E.-A., Magdy, N., & Yamani, H. Z. (2017) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2017.03.021 · Cited by 15
Subnanomolar detection of ions using thin voltammetric membranes with reduced Exchange capacity
Xu, K., Crespo, G. A., & Cuartero, M. (2020) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2020.128453 · Cited by 15
Design, synthesis and application of carbazole macrocycles in anion sensors
Rüütel, A., Yrjänä, V., Kadam, S. A., Saar, I., Ilisson, M., Darnell, A., Haav, K., Haljasorg, T., Toom, L., Bobacka, J., et al. (2020) · Beilstein Journal of Organic Chemistry · doi:10.3762/bjoc.16.157 · Cited by 15
Self-citationCheong, Y. H., & Lisak, G. (2021) · ACS Sensors · doi:10.1021/acssensors.1c01413 · Cited by 15
Zhang, H., Sun, L., Song, C., Liu, Y., Xuan, X., Wang, F., Zhong, J., & Sun, L. (2021) · Sensor Review · doi:10.1108/sr-03-2021-0081 · Cited by 15
Huang, S., Shih, W.-L., Chen, Y., Wu, Y.-M., & Chen, L. (2021) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2021.100088 · Cited by 15
Zhang, Y., Tang, Y., Liang, R., Zhong, L., Xu, J., Lu, H., Xu, X., Han, T., Bao, Y., Ma, Y., et al. (2022) · Membranes · doi:10.3390/membranes12090903 · Cited by 15
McElhoney, K. M., O’Neil, G. D., Chaniotakis, N. A., & Kounaves, S. P. (2012) · Electroanalysis · doi:10.1002/elan.201200264 · Cited by 15
Lenar, N., Piech, R., & Paczosa‐Bator, B. (2021) · Materials · doi:10.3390/ma14051308 · Cited by 15
“One of the leading trends in designing new ion-selective electrodes was finding electrodes with high stability, repeatability, and reproducibility of the potential [4,6].”
BackgroundMethodInfluentialLenar, N., Piech, R., & Paczosa‐Bator, B. (2021) · Materials · doi:10.3390/ma14081891 · Cited by 15
“However, since the PEDOT:PSS polymer itself does not exhibit high electrical capacity, ion-selective electrodes with such mediation layer are also characterized by rather low (~200 μF) capacitance [9].”
BackgroundMethodInfluentialГорелов, И. П., Ryasenskii, S. S., Картамышев, С. В., & Fedorova, M. V. (2005) · Journal of Analytical Chemistry · doi:10.1007/s10809-005-0057-2 · Cited by 15
Ionophores in polymeric membranes for selective ion recognition; impedance studies
Lisowska‐Oleksiak, A., Lesinska, U., Nowak, A. P., & Bocheńska, M. (2006) · Electrochimica Acta · doi:10.1016/j.electacta.2005.04.078 · Cited by 15
Pejcic, B., & MARCO, R. D. (2004) · Electrochimica Acta · doi:10.1016/j.electacta.2004.03.029 · Cited by 16
Disposable blood potassium sensors based on screen-printed thick film electrodes
Xu, H., Yang, X., Wang, Y., Zheng, J., Luo, Z., & Li, G. (2010) · Measurement Science and Technology · doi:10.1088/0957-0233/21/5/055802 · Cited by 14
“capacitance of the solid-contact material [13].”
Hydrous Cerium Dioxide-Based Materials as Solid-Contact Layers in Potassium-Selective Electrodes
Lenar, N., Piech, R., & Paczosa‐Bator, B. (2022) · Membranes · doi:10.3390/membranes12040349 · Cited by 14
“Electrical parameters are calculated based on the results obtained from each step using the Rtotal = 4Edc/2I equation, where 4Edc stands for the value of the potential jumps between the odd and even steps and I is the set current value (for calculating the resistance Rtotal value) and C = I(dt/dEdc) equation, where Ed…”
MethodChai, D., Sun, Y., Li, Z., Yang, H., Mao, S., Tang, J., Gong, W., & Zeng, X. (2022) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2022.132055 · Cited by 14
Kozma, J., Papp, S., & Gyurcsányi, R. E. (2022) · Bioelectrochemistry · doi:10.1016/j.bioelechem.2022.108352 · Cited by 14
Liu, Y., Zeng, X., Waterhouse, G. I. N., Jiang, X., Zhang, Z., & Yu, L. (2023) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2023.117472 · Cited by 14
New Materials Used for the Development of Anion-Selective Electrodes—A Review
Wardak, C., Morawska, K., & Pietrzak, K. (2023) · Materials · doi:10.3390/ma16175779 · Cited by 17
“(3) The equilibrium of the ion-electron conductivity should be reversible and stable [19].”
BackgroundHassan, S. S. M., El-Shalakany, H. H., Fathy, M. A., & Kamel, A. H. (2024) · Microchimica Acta · doi:10.1007/s00604-024-06394-1 · Cited by 14
“(3) and (4) [42]. where E represents potential change, I is the applied current (1.0 nA), R is bulk membrane resistance, t is time change, and C is redox capacitance.”
InfluentialFlexible ammonium ion-selective electrode based on inkjet-printed graphene solid contact
Krivačić, S., Boček, Ž., Zubak, M., Kojić, V., & Kassal, P. (2024) · Talanta · doi:10.1016/j.talanta.2024.126614 · Cited by 17
Soares, R. R. A., Miliao, G. L., Pola, C. C., Jing, D., Opare‐Addo, J., Smith, E. A., Claussen, J. C., & Gomes, C. L. (2024) · Microchimica Acta · doi:10.1007/s00604-024-06672-y · Cited by 14
“The methodology employed to determine the low-frequency capacitance in ISEs involves the application of a small constant current sufficient to polarize the electrode within the range of nanoamperes (nA) to picoamperes (pA) [25].”
Rustamlı, E., Göker, S., Tarkuç, S., Udum, Y. A., & Toppare, L. K. (2015) · Journal of The Electrochemical Society · doi:10.1149/2.0751509jes · Cited by 15
Synthesis and characterization of high-integrity solid-contact polymeric ion sensors
MARCO, R. D., Jee, E., Prince, K. E., Pretsch, E., & Bakker, E. (2008) · Journal of Solid State Electrochemistry · doi:10.1007/s10008-008-0600-9 · Cited by 14
Best, F. R. G., Koch, J., Lilienkamp, G., Körkemeyer, F., Maier, H. J., Caro, J., & Lange, K. (2018) · International Journal of Electrochemistry · doi:10.1155/2018/9875438 · Cited by 14
“Chronopotentiometry is found to be an appropriate and fast method to evaluate the electrochemical stability of such electrodes [69].”
MethodKałuża, D., Michalska, A., & Maksymiuk, K. (2019) · Electroanalysis · doi:10.1002/elan.201900380 · Cited by 14
Neo, Z. H., Seah, G. E. K. K., Ng, S. H., Safanama, D., Seng, D. H. L., & Goh, S. S. (2022) · Analytical Chemistry · doi:10.1021/acs.analchem.2c02119 · Cited by 16
Yao, Y., Wen, Y., Xu, J., Zhang, L., Duan, X., Lu, L., & Xia, H. (2013) · Monatshefte für Chemie - Chemical Monthly · doi:10.1007/s00706-013-1003-x · Cited by 14
Shvedene, N. V., Rzhevskaia, A. V., & Плетнев, И. В. (2012) · Talanta · doi:10.1016/j.talanta.2012.07.057 · Cited by 15
Bondar, A. V., Keresten, V. M., & Mikhelson, K. N. (2022) · Journal of Analytical Chemistry · doi:10.1134/s1061934822020046 · Cited by 13
Bajaber, M. A., & Kamel, A. H. (2022) · Polymers · doi:10.3390/polym14224814 · Cited by 13
“In line with Bobacka’s method [52], the short-term potential stability of the presented electrodes was investigated.”
MethodBao, H., Ye, J., Zhao, X., & Zhang, Y. (2023) · Molecules · doi:10.3390/molecules28073242 · Cited by 15
“The semicircle is related to the bulk impedance of the ion−selective sensitive membrane, and also to the contact resistance of the solid contact layer and the interface of the ion−selective sensitive membrane [46].”
BackgroundBao, Y., Yan, J., Hu, J., & Li, J. (2023) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2023.133997 · Cited by 13
Rousseau, C. R., Chipangura, Y. E., Stein, A., & Bühlmann, P. (2024) · Langmuir · doi:10.1021/acs.langmuir.3c03027 · Cited by 14
Solid‐contact Acetate‐selective Electrode Based on a 1,3‐bis(carbazolyl)urea‐ionophore
Martin, K., Kadam, S. A., Mattinen, U., Bobacka, J., & Leito, I. (2019) · Electroanalysis · doi:10.1002/elan.201800790 · Cited by 13
Self-citationAbd‐Rabboh, H. S. M., Kamel, A. H., & Amr, A. E. E. (2020) · Chemosensors · doi:10.3390/chemosensors8040093 · Cited by 13
“The low-frequency capacitance values for C/Ca2+-ISEs, C/PPY/Ca2+-ISEs, and C/GO/Ca2+-ISEs were calculated from the equation (ΔE/Δt = I/C) [29], and found to be 8.9 ± 0.8, 54.6 ± 1.1 and 88.4 ± 1.3 μF, respectively.”
MethodInfluentialZhang, B., Li, J., & Qin, W. (2023) · Talanta · doi:10.1016/j.talanta.2023.124444 · Cited by 13
Shanaah, H. H., Ameen, Z., Jaafar, K., Hefnawy, A., Abd‐Rabboh, H. S. M., & Kamel, A. H. (2023) · ChemElectroChem · doi:10.1002/celc.202300153 · Cited by 13
Potentiometric detection of apomorphine in human plasma using a 3D printed sensor
Elhassan, M. M., Glasco, D. L., Sheelam, A., Mahmoud, A. M., Hegazy, M. A. M., Mowaka, S., & Bell, J. G. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115971 · Cited by 13
Paczosa‐Bator, B., Piech, R., & Cabaj, L. (2012) · Electrochimica Acta · doi:10.1016/j.electacta.2012.08.051 · Cited by 13
Bhattacharya, G., Mathur, A. K., Pal, S., McLaughlin, J. A., & Roy, S. S. (2016) · International Journal of Electrochemical Science · doi:10.20964/2016.07.86 · Cited by 13
Amr, A. E. E., Al-Omar, M. A., Kamel, A. H., & Elsayed, E. A. (2019) · Molecules · doi:10.3390/molecules24050920 · Cited by 13
“To evaluate the potential stability in the proposed sensors using either PEDOT/PSS or SWCNTs as solid contact materials, constant current chronopotentiometry was used according to Bobacka’s protocol [39].”
MethodEzzat, S., Ahmed, M. A., Amr, A. E. E., Al-Omar, M. A., Kamel, A. H., & Khalifa, N. M. (2019) · Materials · doi:10.3390/ma12182924 · Cited by 13
“3 μF by using the equation ∆E/∆t = I/C [47].”
InfluentialBaranowska‐Korczyc, A., Jaworska, E., Strawski, M., Paterczyk, B., Maksymiuk, K., & Michalska, A. (2020) · The Analyst · doi:10.1039/d0an00830c · Cited by 14
Abd‐Rabboh, H. S. M., Amr, A. E. E., Kamel, A. H., Al-Omar, M. A., & Sayed, A. Y. A. (2021) · RSC Advances · doi:10.1039/d0ra09903a · Cited by 13
Spectroelectrochemical and potentiometric studies of functionalised electroactive polymers
Tarábek, J., Rapta, P., Jähne, E., Ferse, D., Adler, H. P., Maumy, M., & Dunsch, L. (2004) · Electrochimica Acta · doi:10.1016/j.electacta.2004.10.020 · Cited by 13
Design and synthesis of a task-specific ionic liquid as a transducer in potentiometric sensors
Ping, J. F., Wang, Y., Ying, Y. B., & Wu, J. F. (2013) · RSC Advances · doi:10.1039/c3ra23367g · Cited by 13
A Batteryless Chronic Wound Monitoring System With 13.56-MHz Energy Harvesting
Cho, H.-W., Jo, S.-H., Yoon, J. H., Goh, T. S., Choi, B. G., & Yoo, H.-J. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2923847 · Cited by 13
Keresten, V. M., Solovyeva, E. V., & Mikhelson, K. N. (2021) · Membranes · doi:10.3390/membranes11050344 · Cited by 12
Kamel, A. H., Amr, A. E. E., Al-Omar, M. A., & Almehizia, A. A. (2020) · Membranes · doi:10.3390/membranes10100279 · Cited by 12
“The high electric capacity obtained after inserting the MWCNT layer in MIP (IPU)/MWCNTs-ISEs or MIP (DU)/MWCNTs-ISEs plays an essential role in resisting the external current’s polarization [51].”
BackgroundSu, Y., Liu, T., Song, C., Ai-Qiao, F. A. N., Zhu, N., Sun, B., & Yang, C. (2021) · Chemosensors · doi:10.3390/chemosensors9070155 · Cited by 12
Synergy between the Assembly of Individual PEDOT Chains and Their Interaction with Light
Ivanko, I., Mahun, A., Kobera, L., Černochová, Z., Pavlová, E., Toman, P., Pientka, Z., Štěpánek, P., & Tomšík, E. (2021) · Macromolecules · doi:10.1021/acs.macromol.1c01975 · Cited by 12
Solid-Contact Potentiometric Cell with Symmetry
Zdrachek, E. A., Forrest, T., & Bakker, E. (2021) · Analytical Chemistry · doi:10.1021/acs.analchem.1c04722 · Cited by 12
AC-Electrogravimetry Investigation in Electroactive Thin Films
Gabrielli, C., & Perrot, H. (2009) · Modern aspects of electrochemistry · doi:10.1007/978-0-387-49586-6_5 · Cited by 18
Janrungroatsakul, W., Lertvachirapaiboon, C., Ngeontae, W., Aeungmaitrepirom, W., Chailapakul, O., Ekgasit, S., & Tuntulani, T. (2013) · The Analyst · doi:10.1039/c3an01385e · Cited by 12
An ionic liquid-based sensor for diclofenac determination in water
Brennan, E., Futvoie, P., Cassidy, J., & Schazmann, B. (2017) · International Journal of Environmental & Analytical Chemistry · doi:10.1080/03067319.2017.1333607 · Cited by 12
Jaworska, E., Michalska, A., & Maksymiuk, K. (2020) · Electroanalysis · doi:10.1002/elan.201900650 · Cited by 12
Abd‐Rabboh, H. S. M., & Kamel, A. H. (2020) · Analytical Sciences · doi:10.2116/analsci.20p143 · Cited by 12
Lenar, N., Piech, R., Wyrwa, J., & Paczosa‐Bator, B. (2021) · Membranes · doi:10.3390/membranes11040259 · Cited by 12
Han, T., Song, T., Gan, S., Han, D., Bobacka, J., Niu, L., & Ivaska, A. U. (2022) · Chinese Journal of Chemistry · doi:10.1002/cjoc.202200500 · Cited by 12
Self-citationAll-solid-state ion-selective electrodes with ion-to-electron transducers for dopamine determination
Холошенко, Н. М., Ryasenskii, S. S., & Горелов, И. П. (2006) · Pharmaceutical Chemistry Journal · doi:10.1007/s11094-006-0122-7 · Cited by 12
Grygołowicz‐Pawlak, E., Pałys, B., Biesiada, K., Olszyna, A. R., & Malinowska, E. (2008) · Analytica Chimica Acta · doi:10.1016/j.aca.2008.07.019 · Cited by 12
Robinson, E. E. A., Fathy, M. A., & Bühlmann, P. (2025) · ACS Sensors · doi:10.1021/acssensors.4c03702 · Cited by 11
Miniature Ion‐selective Electrodes with Mesoporous Carbon Black as Solid Contact
Zhang, Z., & Papautsky, I. (2021) · Electroanalysis · doi:10.1002/elan.202100088 · Cited by 11
An all-solid-state thin-layer laminated cell for calibration-free coulometric determination of K+
Tatsumi, S., Omatsu, T., Maeda, K., Mousavi, M. P. S., Whitesides, G. M., & Yoshida, Y. (2022) · Electrochimica Acta · doi:10.1016/j.electacta.2022.139946 · Cited by 11
Abd‐Rabboh, H. S. M., Hashem, H. M., Shagri, L. M. S. A., Amr, A. E.-G. E., Almehizia, A. A., Naglah, A. M., & Kamel, A. H. (2022) · Membranes · doi:10.3390/membranes12050446 · Cited by 11
“Constant-current chronopotentiometry measurements were conducted to test the short-term potential stability of the presented electrodes and to evaluate the double-layer capacitance of both MWCNTs as an ion-to electron transducer [42].”
BackgroundMethodNiemiec, B., Piech, R., & Paczosa‐Bator, B. (2023) · Molecules · doi:10.3390/molecules28114313 · Cited by 11
“Chronopotentiograms were obtained according to the method proposed by Bobacka [34].”
MethodChapter 4 Ion sensors with conducting polymers as ion-to-electron transducers
Bobacka, J., & Ivaska, A. U. (2007) · Comprehensive analytical chemistry · doi:10.1016/s0166-526x(06)49004-8 · Cited by 11
Self-citationA solid-contact pH-selective electrode based on tridodecylamine as hydrogen neutral ionophore
Zhang, J., Guo, Y., Li, S., & Xu, H. (2016) · Measurement Science and Technology · doi:10.1088/0957-0233/27/10/105101 · Cited by 11
Ocaña, C., Muñoz-Correas, M., Abramova, N., & Bratov, A. V. (2020) · Sensors · doi:10.3390/s20051348 · Cited by 11
“From these measurements, we can also estimate the bulk resistance of the electrode (solid-contact and membrane) and redox capacitance of the solid-contact [34].”
BackgroundMethodInfluentialEffect of dissolved CO2 on the potential stability of all-solid-state ion-selective electrodes
Han, J. H., Cui, G., Kim, S. J., Han, S. H., Cha, G. S., & Nam, H. (2001) · The Analyst · doi:10.1039/b105585m · Cited by 11
Malinowski, P., Grzegrzółka, I., Michalska, A., & Maksymiuk, K. (2010) · Journal of Solid State Electrochemistry · doi:10.1007/s10008-010-1073-1 · Cited by 11
Gosti, C., Mousavi, Z., Fiore, L., Mazzaracchio, V., Olivieri, F., Gentile, G., Arduini, F., & Bobacka, J. (2025) · ACS Sensors · doi:10.1021/acssensors.5c02226 · Cited by 10
Self-citationRobust Potentiometric Microelectrodes for In Situ Sensing of Ion Fluxes with High Sensitivity
Li, Y., Zhu, X., Ding, J., & Qin, W. (2023) · Analytical Chemistry · doi:10.1021/acs.analchem.3c03267 · Cited by 10
Choi, K. R., Honig, M. L., & Bühlmann, P. (2024) · Analytical Chemistry · doi:10.1021/acs.analchem.4c00204 · Cited by 10
Ion-Selective Electrode for Nitrates Based on a Black PCV Membrane
Lenar, N., Drużyńska, M., Piech, R., & Paczosa‐Bator, B. (2024) · Molecules · doi:10.3390/molecules29153473 · Cited by 10
“The stability of the potentiometric response of all-solid-state sensors was improved thanks to the application of a redox-active, hydrophobic material of high surface area acting as a connector between an ionic and electronic conductor [7].”
BackgroundInfluentialPrinted Potentiometric Ammonium Sensors for Agriculture Applications
Toor, A., Goodrich, P. J., Anthony, T. L., Beckstoffer, C., Jegan, H., Silver, W. L., & Arias, A. C. (2024) · ACS Omega · doi:10.1021/acsomega.4c05746 · Cited by 10
Abd‐Rabboh, H. S. M., Amr, A. E.-G. E., Almehizia, A. A., Naglah, A. M., & Kamel, A. H. (2022) · Polymers · doi:10.3390/polym14101942 · Cited by 12
Choi, K. R., Honig, M. L., & Bühlmann, P. (2023) · The Analyst · doi:10.1039/d3an02047a · Cited by 10
Non-enzymatic paper-based analytical device for direct potentiometric detection of urine creatinine
Kamel, A. H., Abd‐Rabboh, H. S. M., & Bajaber, M. A. (2024) · Microchimica Acta · doi:10.1007/s00604-024-06203-9 · Cited by 11
“…(Δ E /Δ t ), bulk membrane resistance ( R b ), and interfacial double-layer capacitance ( C L ) using the plot of the potential values with time [31]. where Δ E , I , R , Δ t , and C L represent the change in potential, applied current (1 nA), bulk membrane resistance, (1) Δ E = IR (2) Δ E ∕Δ t…”
Shishkanová, T. V., Broncová, G., Kronďák, M., Sýkora, D., & Král, V. (2011) · Journal of Materials Science · doi:10.1007/s10853-011-5735-x · Cited by 10
Wang, T., Cui, C., Huang, Y., Fan, Y., Xu, Z., Sarge, L., Bagtzoglou, C., Brückner, C., Gao, P., & Li, B. (2022) · Environmental Science Nano · doi:10.1039/d1en00966d · Cited by 13
Kamal, A., Raj, R., Kumar, V., & Mahajan, R. K. (2015) · Electrochimica Acta · doi:10.1016/j.electacta.2015.03.014 · Cited by 10
Reduced graphene oxide in the construction of solid-state bromide-selective electrode
Shvedene, N. V., Rzhevskaia, A. V., Anuchin, N. M., Kapitanova, O. O., Баранов, А. Н., & Плетнев, И. В. (2015) · Journal of Analytical Chemistry · doi:10.1134/s1061934815030193 · Cited by 10
Impedance spectroscopic investigations of zeolite–polydimethylsiloxane electrodes
Matysik, S., Matysik, F., Schulze, K. D., & Einicke, W. ‐. (2002) · Electrochimica Acta · doi:10.1016/s0013-4686(02)00666-7 · Cited by 10
Flow Injection Analysis of Iodate Reduction on PEDOT Modified Electrode
Balamurugan, A., & Chen, S. (2008) · Electroanalysis · doi:10.1002/elan.200804261 · Cited by 10
Metadata
- Authors
- Bobacka, J.
- Affiliations
- Åbo Akademi University (FI)
- Corresponding author
- Bobacka, Johan
- ORCID
- Bobacka, Johan
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 1999-10-01
- Volume
- 71
- Issue
- 21
- Pages
- 4932-4937
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- OpenAlex
- W2000694408
- Cited by
- 686
- References
- 34
- Citations by year
- 2026: 29 · 2025: 33 · 2024: 54 · 2023: 39 · 2022: 46 · 2021: 49 · 2020: 47 · 2019: 51 · 2018: 33 · 2017: 28 · 2016: 23 · 2015: 23 · 2014: 23 · 2013: 19 · 2012: 25
- FWCI
- 4.77
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Analytical Chemistry (publishedVersion)
- PubMed (publishedVersion)
- Åbo Akademi University Research Portal (submittedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
- Keywords
- potential stability, conducting polymers, ion-to-electron transduction, PEDOT, potassium ion-selective electrode, chronopotentiometry, electrochemical impedance spectroscopy, solvent polymeric membranes
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.