Photon-Counting Computed Tomography for Vascular Imaging of the Head and Neck: First In Vivo Human Results
Open accessThe DOI matches its Crossref record; 34 of the 34 references it cites were found in a registry or an index.
Abstract
Purpose: The purpose of this study was to evaluate image quality of a spectral photon-counting detector (PCD) computed tomography (CT) system for evaluation of major arteries of the head and neck compared with conventional single-energy CT scans using energy-integrating detectors (EIDs). Methods: In this institutional review board–approved study, 16 asymptomatic subjects (7 men) provided informed consent and received both PCD and EID contrast-enhanced CT scans of the head and neck (mean age, 58 years; range, 46–75 years). Tube settings were (EID: 120 kVp/160 mA vs PCD: 140 kVp/108 mA) for all volunteers. Quantitative analysis included measurements of mean attenuation, image noise, and contrast-to-noise ratio (CNR). Spectral PCD data were used to reconstruct virtual monoenergetic images and iodine maps. A head phantom was used to validate iodine concentration measurements in PCD images only. Two radiologists blinded to detector type independently scored the image quality of different segments of the arteries, as well as diagnostic acceptability, image noise, and severity of artifacts of the PCD and EID images. Reproducibility was assessed with intraclass correlation coefficient. Linear mixed models that account for within-subject correlation of analyzed arterial segments were used. Linear regression and Bland-Altman analysis with 95% limits of agreement were used to calculate the accuracy of material decomposition. Results: Photon-counting detector image quality scores were significantly higher compared with EID image quality scores with lower image noise (P < 0.01) and less image artifacts (P < 0.001). Photon-counting detector image noise was 9.1% lower than EID image noise (8.0 ± 1.3 HU vs 8.8 ± 1.5 HU, respectively, P < 0.001). Arterial segments showed artifacts on EID images due to beam hardening that were not present on PCD images. On PCD images of the head phantom, there was excellent correlation (R 2 = 0.998) between actual and calculated iodine concentrations without significant bias (bias: −0.4 mg/mL [95% limits of agreements: −1.1 to 0.4 mg/mL]). Iodine maps had 20.7% higher CNR compared with nonspectral PCD (65.2 ± 9.0 vs 54.0 ± 4.5, P = 0.01), and virtual monoenergetic image at 70 keV showed similar CNR to nonspectral images (52.6 ± 4.2 vs 54.0 ± 4.5, P = 0.39). Conclusions: Photon-counting CT has the potential to improve the image quality of carotid and intracranial CT angiography compared with single-energy EID CT.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Simultaneous Inference in General Parametric Models
Hothorn, T., Bretz, F., & Westfall, P. (2008) · Biometrical Journal · doi:10.1002/bimj.200810425 · Cited by 13,847 · cited 1×
“The Tukey test for multiple comparisons of means was used to compare the EID and PCD attenuation values for the different arterial segments using the “lme4” and “multcomp” functions in R Statistical Software.(27,28) The”
MethodMixed-effects modeling with crossed random effects for subjects and items
Baayen, R. H., Davidson, D. J., & Bates, D. M. (2008) · Journal of Memory and Language · doi:10.1016/j.jml.2007.12.005 · Cited by 8,540 · cited 1×
“Linear mixed models that accounted for within-subject correlations were used to estimate the effects of detector subsystem, laterality, and acquisition order on HU values in the analyzed arterial segments.(26) The fixed-effect predictors included indicator variables for detector subsystem (EID vs PCD CT), left versus…”
MethodForming inferences about some intraclass correlation coefficients.
McGraw, K. O., & Wong, S. P. (1996) · Psychological Methods · doi:10.1037/1082-989x.1.1.30 · Cited by 6,903 · cited 1×
“Unauth This paper can be cited using the date of access and the un measures of consistency was used to calculate ICC values.(29) Linear regression and Bland-Altman analysis with 95% limits of agreement (LOAs) was used to calculate the agreement between the true and estimated iodine concentrations of the iodine test vi…”
MethodStatistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220 · cited 1×
“Linear regression and Bland-Altman analysis with 95% limits of agreement (LOAs) was used to calculate the agreement between the true and estimated iodine concentrations of the iodine test vials.(30)We plotted differences against the calibrated true iodine concentrations as a criterion standard method as proposed by Kr…”
MethodFitting Linear Mixed-Effects Models Using lme4
Bates, D. M., Maechler, M., Bolker, B. M., & Walker, S. (2014) · arXiv (Cornell University) · doi:10.3929/ethz-b-000105397 · Cited by 2,541
Energy-selective reconstructions in X-ray computerised tomography
Alvarez, R. E., & Macovski, A. (1976) · Physics in Medicine & Biology · doi:10.1088/0031-9155/21/5/002 · Cited by 2,013 · cited 1×
“Based on the unique energy-dependent behavior of materials, different materials can be highlighted (eg, iodine vs calcium).(24,25)”
BackgroundDual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications
McCollough, C. H., Leng, S., Yu, L., & Fletcher, J. G. (2015) · Radiology · doi:10.1148/radiol.2015142631 · Cited by 1,598
Material differentiation by dual energy CT: initial experience
Johnson, T. R. C., Krauß, B., Sedlmair, M., Grasruck, M., Bruder, H., Morhard, D., Fink, C., Weckbach, S., Lenhard, M., Schmidt, B., et al. (2007) · European Radiology · doi:10.1007/s00330-006-0517-6 · Cited by 1,510 · cited 1×
“Regions of interest of approximately 20 mm(2) were carefully placed inside the iodine calibration vials, and the iodine concentration (milligram per milliliter) was recorded.(23) The software generated virtual monoenergetic images (VMIs) with photon energies of 40 to 140 keVat 10-keV intervals for the human subjects.”
MethodVision 20/20: Single photon counting x‐ray detectors in medical imaging
Taguchi, K., & Iwanczyk, J. S. (2013) · Medical Physics · doi:10.1118/1.4820371 · Cited by 948 · cited 1×
“Based on the unique energy-dependent behavior of materials, different materials can be highlighted (eg, iodine vs calcium).(24,25)”
BackgroundSchlomka, J. P., Roessl, E., Dorscheid, R., Dill, S., Martens, G., Istel, T., Bäumer, C., Herrmann, C., Steadman, R., Zeitler, G., et al. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/15/002 · Cited by 801
Beam hardening in X-ray reconstructive tomography
Brooks, R. A., & Chiro, G. D. (1976) · Physics in Medicine and Biology · doi:10.1088/0031-9155/21/3/004 · Cited by 721 · cited 1×
“This increases the mean energy of the x-ray beam.(4,5) Higher peak kilovoltage (kVp) settings (and therefore a smaller fraction of low-energy photons) result in reduced beam-hardening artifacts.”
BackgroundMicrosurgical anatomy of the supraclinoid portion of the internal carotid artery
Gibo, H., Lenkey, C., & Rhoton, A. L. (1981) · Journal of Neurosurgery · doi:10.3171/jns.1981.55.4.0560 · Cited by 440 · cited 1×
“1).(22) Readers used commercially available PACS software (VuePACS v.”
MethodWhy Bland–Altman plots should use X, not (Y+X)/2 when X is a reference method
Krouwer, J. S. (2008) · Statistics in Medicine · doi:10.1002/sim.3086 · Cited by 415
Grant, K. L., Flohr, T. G., Krauss, B., Sedlmair, M., Thomas, C., & Schmidt, B. (2014) · Investigative Radiology · doi:10.1097/rli.0000000000000060 · Cited by 307 · cited 1×
“Recent enhancements in noiseoptimized virtual monoenergetic reconstructions may further improve CNR between iodine and surrounding tissue.(32,37)”
BackgroundSelf-citationGutjahr, R., Halaweish, A. F., Yu, Z., Leng, S., Yu, L., Li, Z., Jorgensen, S. M., Ritman, E. L., Kappler, S., & McCollough, C. H. (2016) · Investigative Radiology · doi:10.1097/rli.0000000000000251 · Cited by 282
Self-citationSymons, R., Krauss, B., Sahbaee, P., Cork, T. E., Lakshmanan, M. N., Bluemke, D. A., & Pourmorteza, A. (2017) · Medical Physics · doi:10.1002/mp.12301 · Cited by 217 · cited 1×
“Additional thresholds would allow, for example, water, gadolinium, iodine, and calcium to be separated using k-edge imaging and material decomposition techniques.(13,14,40) Fifth, ultra–high-resolution imaging (eg, 0.”
BackgroundSelf-citationAbdominal Imaging with Contrast-enhanced Photon-counting CT: First Human Experience
Pourmorteza, A., Symons, R., Sandfort, V., Mallek, M., Fuld, M. K., Henderson, G., Jones, E. C., Malayeri, A. A., Folio, L. R., & Bluemke, D. A. (2016) · Radiology · doi:10.1148/radiol.2016152601 · Cited by 207 · cited 1×
“Recently, a prototype whole-body PCD CT scanner has been developed based on a dual-source CT scanner with 2 independent EID and PCD subsystems with identical x-ray sources.(8,12,15) The aim of this study was to evaluate the image quality of a spectral PCD CT system compared with conventional EID CT for contrast-enhanc…”
MethodSelf-citationA new look at CT dose measurement: Beyond CTDI
Dixon, R. L. (2003) · Medical Physics · doi:10.1118/1.1576952 · Cited by 202 · cited 1×
“Unauth This paper can be cited using the date of access and the un of the PCD subsystem in the prototype, which increases the off-collimation radiation to the object; such off-collimation radiation does not affect the image quality but increments the total radiation dose.(20) Identical collimations would result in sim…”
BackgroundFeasibility of Dose-reduced Chest CT with Photon-counting Detectors: Initial Results in Humans
Symons, R., Pourmorteza, A., Sandfort, V., Ahlman, M. A., Cropper, T., Mallek, M., Kappler, S., Ulzheimer, S., Mahesh, M., Jones, E. C., et al. (2017) · Radiology · doi:10.1148/radiol.2017162587 · Cited by 188 · cited 1×
“4 mGy).(18,19) This is due to the smaller z-axis collimation”
BackgroundSelf-citationDual-contrast agent photon-counting computed tomography of the heart: initial experience
Symons, R., Cork, T. E., Lakshmanan, M. N., Evers, R. W., Davies-Venn, C., Rice, K. A., Thomas, M. L., Liu, C., Kappler, S. G., Ulzheimer, S., et al. (2017) · International journal of cardiac imaging · doi:10.1007/s10554-017-1104-4 · Cited by 143 · cited 1×
“Additional thresholds would allow, for example, water, gadolinium, iodine, and calcium to be separated using k-edge imaging and material decomposition techniques.(13,14,40) Fifth, ultra–high-resolution imaging (eg, 0.”
BackgroundSelf-citationKitagawa, K., George, R. T., Arbab‐Zadeh, A., et al. (2010) · Radiology · doi:10.1148/radiol.10091399 · Cited by 126
High-Definition CT Gemstone Spectral Imaging of the Brain
Lin, X. Z., Miao, F., Li, J. Y., et al. (2011) · Journal of Computer Assisted Tomography · doi:10.1097/rct.0b013e3182058d5c · Cited by 124
Low-dose lung cancer screening with photon-counting CT: a feasibility study
Symons, R., Cork, T. E., Sahbaee, P., Fuld, M. K., Kappler, S. G., Folio, L. R., Bluemke, D. A., & Pourmorteza, A. (2016) · Physics in Medicine and Biology · doi:10.1088/1361-6560/62/1/202 · Cited by 120 · cited 1×
“Recently, a prototype whole-body PCD CT scanner has been developed based on a dual-source CT scanner with 2 independent EID and PCD subsystems with identical x-ray sources.(8,12,15) The aim of this study was to evaluate the image quality of a spectral PCD CT system compared with conventional EID CT for contrast-enhanc…”
MethodSelf-citationEnergy-sensitive photon counting detector-based X-ray computed tomography
Taguchi, K. (2017) · Radiological Physics and Technology · doi:10.1007/s12194-017-0390-9 · Cited by 95
Yu, Z., Leng, S., Li, Z., & McCollough, C. H. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/18/6707 · Cited by 94
Neuhaus, V., Abdullayev, N., Hokamp, N. G., Pahn, G., Kabbasch, C., Mpotsaris, A., Maintz, D. C., & Borggrefe, J. (2017) · Investigative Radiology · doi:10.1097/rli.0000000000000367 · Cited by 88
An Inaccuracy in Computed Tomography: The Energy Dependence of CT Values
Zatz, L. M., & Alvarez, R. E. (1977) · Radiology · doi:10.1148/124.1.91 · Cited by 74 · cited 1×
“This increases the mean energy of the x-ray beam.(4,5) Higher peak kilovoltage (kVp) settings (and therefore a smaller fraction of low-energy photons) result in reduced beam-hardening artifacts.”
BackgroundMultidetector-row CT angiography in the study of atherosclerotic carotid arteries
Saba, L., Sanfilippo, R., Pirisi, R., Pascalis, L., Montisci, R., & Mallarini, G. (2007) · Neuroradiology · doi:10.1007/s00234-007-0244-y · Cited by 63
A Spectral CT Method to Directly Estimate Basis Material Maps From Experimental Photon-Counting Data
Schmidt, T. G., Barber, R. F., & Sidky, E. Y. (2017) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2017.2696338 · Cited by 59
Quantitative Spectral K-Edge Imaging in Preclinical Photon-Counting X-Ray Computed Tomography
de Vries, A., Roessl, E., Kneepkens, E., Thran, A., Brendel, B., Martens, G., Proska, R., Nicolay, K., & Grüll, H. (2014) · Investigative Radiology · doi:10.1097/rli.0000000000000126 · Cited by 36 · cited 1×
“Additional thresholds would allow, for example, water, gadolinium, iodine, and calcium to be separated using k-edge imaging and material decomposition techniques.(13,14,40) Fifth, ultra–high-resolution imaging (eg, 0.”
BackgroundMachida, H., TAKEUCHI, H., Tanaka, I., et al. (2013) · American Journal of Neuroradiology · doi:10.3174/ajnr.a3320 · Cited by 35
Estimation of signal and noise for a whole-body research photon-counting CT system
Li, Z., Leng, S., Yu, Z., Kappler, S. G., & McCollough, C. H. (2017) · Journal of medical imaging · doi:10.1117/1.jmi.4.2.023505 · Cited by 13
Self-citation
Cited by
Photon-counting CT: Technical Principles and Clinical Prospects
Willemink, M. J., Persson, M., Pourmorteza, A., Pelc, N. J., & Fleischmann, D. (2018) · Radiology · doi:10.1148/radiol.2018172656 · Cited by 1,326
Self-citationWillemink, M. J., & Noël, P. B. (2019) · European Radiology · doi:10.1007/s00330-018-5810-7 · Cited by 603
Flohr, T., Petersilka, M., Henning, A., Ulzheimer, S., Ferda, J., & Schmidt, B. (2020) · Physica Medica · doi:10.1016/j.ejmp.2020.10.030 · Cited by 581
Self-citationLell, M. M., & Kachelrieß, M. (2020) · Investigative Radiology · doi:10.1097/rli.0000000000000601 · Cited by 254
Leng, S., Rajendran, K., Gong, H., Zhou, W., Halaweish, A. F., Henning, A., Kappler, S., Baer, M., Fletcher, J. G., & McCollough, C. H. (2018) · Investigative Radiology · doi:10.1097/rli.0000000000000488 · Cited by 208
Self-citation
Metadata
- Authors
- Symons, R., Reich, D. S., Bagheri, M., Cork, T. E., Krauss, B., Ulzheimer, S., Kappler, S., Bluemke, D. A., & Pourmorteza, A.
- Venue
- Investigative Radiology
- Published
- 2018-03
- Volume
- 53
- Issue
- 3
- Pages
- 135-142
- ISSN
- 0020-9996 (ISSN-L)
- 1536-0210
- OpenAlex
- W2758647230
- Cited by
- 191
- References
- 34
- FWCI
- 5.70
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.