A Flexible, Low Noise Reflective PPG Sensor Platform for Ear-Worn Heart Rate Monitoring
Abstract
This paper presents a novel ear-worn reflective photoplethysmography (PPG) sensor that addresses the mechanical complexities of coupling the sensor to the surface of the skin and a detection circuit that minimises ambient noise artefacts. The flexible optoelectronic transducer structure can adapt to a variety of skin surface contours. Light emitting diode (LED) modulation and a unique integrating photocurrent demodulator reduce susceptibility to wideband noise and allow subtraction of ambient light from the desired PPG signal. Experimental results demonstrate that the sensitivity is robust to sensor location and application pressure variations. Simulations also show that the photodetection method is resilient against high levels of wideband noise.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Cardiac-related components have negligible frequency components above 15Hz.
“As the cardiac-related components have negligible frequency components above 15Hz44, we separated noise above 15Hz and signal below 15Hz from PPG data using 3rd order Butterworth filter.”
From A Multimodal Dataset for Mixed Emotion Recognition (2024) · cites this paper as [44]
The arrangement of the source and detector in Patterson et al.'s flexible PPG sensor contributes to eliminating interference of ambient noise to the signal.
“Apart from its electronic system design, the arrangement of the source and detector in this sensor also contributes to eliminate the interference of ambient noise to the signal.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [201]
Patterson et al.'s opto-electronic modulation scheme helps eliminate ambient light noise through multiplexing and sampling of light level by keeping LEDs off for a fixed time period and later subtracting it out from the desired signals.
“The opto-electronic modulation scheme in this flexible PPG sensor helps to eliminate ambient light noise through the multiplexing technique and sampling of light level by keeping the LEDs off for a fixed time period and later subtracting it out from the desired signals.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [201]
An ear-worn, flexible, low-power PPG sensor for heart rate monitoring was introduced by Patterson et al.
“An ear-worn, flexible, low-power PPG sensor for heart rate monitoring was introduced by Patterson et al. [26].”
From A review of wearable sensors and systems with application in rehabilitation (2012) · cites this paper as [26]
The ear-worn PPG sensor is suited for long-term monitoring due to its location and unobtrusive design.
“The sensor is suited for long-term monitoring due to its location and unobtrusive design.”
From A review of wearable sensors and systems with application in rehabilitation (2012) · cites this paper as [26]
Cites
Yang, G. (2006) · Springer eBooks · doi:10.1007/1-84628-484-8 · Cited by 563 · see citing works
Self-citationStojmenović, I. (2005) · Cited by 278 · see citing works
Spectrophotometric monitoring of arterial oxygen saturation in the fingertip
Yoshiya, I., Shimada, Y., & Tanaka, K. (1980) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442476 · Cited by 262 · see citing works
Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography
Mendelson, Y., & Ochs, B. D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.7286 · Cited by 251 · see citing works
Artifact-resistant power-efficient design of finger-ring plethysmographic sensors
Rhee, S., Yang, B.-H., & Asada, H. H. (2001) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.930904 · Cited by 235 · see citing works
McCombie, D. B., Reisner, A. T., & Asada, H. H. (2006) · 2006 International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2006.260590 · Cited by 118 · see citing works
Multichannel reflective PPG earpiece sensor with passive motion cancellation
Wang, L., Lo, B. P., & Yang, G.-Z. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.910900 · Cited by 114 · see citing works
Artifact reduction in photoplethysmography
Hayes, M., & Smith, P. R. (1998) · Applied Optics · doi:10.1364/ao.37.007437 · Cited by 71 · see citing works
Wong, A., Pun, K.-P., Zhang, Y.-T., & Hung, K. (2005) · IEEE Transactions on Circuits and Systems I: Regular Papers · doi:10.1109/tcsi.2005.857767 · Cited by 63 · see citing works
An Efficient Motion-Resistant Method for Wearable Pulse Oximeter
Yan, Y.-S., & Zhang, Y.-T. (2008) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2007.902173 · Cited by 62 · see citing works
A light emitting diode skin reflectance oximeter
Cohen, A., & Wadsworth, N. (1972) · Medical & Biological Engineering & Computing · doi:10.1007/bf02474218 · Cited by 37 · see citing works
Exercise and Sports Cardiology
Thompson, P. D., & Fernández, A. B. (2016) · WORLD SCIENTIFIC (EUROPE) eBooks · doi:10.1142/q0045-vol2 · Cited by 14 · see citing works
Branche, P. C., Johnston, W. S., Pujary, C. J., et al. (2004) · Proceedings of the IEEE Annual Northeast Bioengineering Conference · doi:10.1109/nebc.2004.1300072 · Cited by 12 · see citing works
Respiratory monitoring using photoplethysmography-evaluation in the postoperative care unit
Johansson, A., Nilsson, L. M., Kálmán, S., et al. (2002) · Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1998.746181 · Cited by 7 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
A review of wearable sensors and systems with application in rehabilitation
Patel, S., Park, H., Bonato, P., Chan, L., & Rodgers, M. (2012) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/1743-0003-9-21 · Cited by 2,396 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
“Year published 2009 [20] 2008 [26] 2007 [19] 2009 [27] 2010 [22] 2012 [24] 2004 [23] 2009 [21]”
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
A Multimodal Dataset for Mixed Emotion Recognition
Yang, P., Liu, N., Liu, X., Shu, Y., Ji, W., Ren, Z., Sheng, J., Yu, M., Yi, R., Zhang, D., et al. (2024) · Scientific Data · doi:10.1038/s41597-024-03676-4 · Cited by 37 · see citing works
- Authors
- Patterson, J. A. C., McIlwraith, D. C., & Yang, G.-Z.
- Affiliations
- Royal Society (GB)
- ORCID
- Yang, Guang-Zhong
- Venue
- 2009 Sixth International Workshop on Wearable and Implantable Body Sensor Networks
- Publisher
- IEEE
- Published
- 2009-06
- Pages
- 286-291
- ISSN
- 2376-8886 (ISSN-L)
- OpenAlex
- W2158314230
- Cited by
- 91
- References
- 14
- Citations by year
- 2026: 2 · 2025: 2 · 2024: 10 · 2023: 11 · 2022: 3 · 2021: 9 · 2020: 4 · 2019: 6 · 2018: 8 · 2017: 10 · 2016: 3 · 2015: 8 · 2014: 6 · 2013: 2 · 2012: 3
- FWCI
- 4.86
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- Wearable and Implantable Body Sensor Networks (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- heart rate monitoring, LED modulation, broadband noise, wearable sensing
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review