Bio-inspired active soft orthotic device for ankle foot pathologies
The DOI matches its Crossref record; 24 of the 24 references it cites were found in a registry or an index.
Abstract
We describe the design of an active soft ankle-foot orthotic device powered by pneumatic artificial muscles for treating gait pathologies associated with neuromuscular disorders. The design is inspired by the biological musculoskeletal system of a human foot and a lower leg, and mimics the muscle-tendon-ligament structure. A key feature of the device is that it is fabricated with flexible and soft materials that provide assistance without restricting degrees of freedom at the ankle joint. Three pneumatic artificial muscles assist dorsiflexion as well as inversion and eversion. The prototype is also equipped with various embedded sensors for gait training and gait pattern analysis. The prototype is capable of 12° dorsiflexion from a resting position of an ankle joint and a 20° dorsiflexion from plantarflexion. Results of early feedback control experiments show controllability of ankle joint angles. Ultimately, we envision a system that not only can provide physical support to improve mobility but also can increase safety and stability during walking, while enhancing muscle usage and encouraging rehabilitation.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Figure 6E presents an EGaIn based-strain sensor that can be applied on top of the ankle to measure the joint’s angle to provide feedback on the foot’s motion during rehabilitation.[201]”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [201]
Citations
Cites
Miller, F. (2005) · Cited by 2,000
Adaptive control of a variable-impedance ankle-foot orthosis to assist drop-foot gait
Blaya, J. A., & Herr, H. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2003.823266 · Cited by 741 · cited 1×
“Previous work has shown the efficacy of adaptive control with active devices [4].”
BackgroundSmooth Vertical Surface Climbing With Directional Adhesion
Kim, S., Spenko, M., Trujillo, S., Heyneman, B., Santos, D., & Cutkosky, M. R. (2008) · IEEE Transactions on Robotics · doi:10.1109/tro.2007.909786 · Cited by 661 · cited 1×
“The differential tendon mechanism of Muscle 2, similar to the foot design of a wall-climbing robot [18], distributes the pulling force from one tendon cable to four anchoring points.”
BackgroundHyperelastic pressure sensing with a liquid-embedded elastomer
Park, Y.-L., Majidi, C., Kramer, R., Bérard, P., & Wood, R. J. (2010) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/20/12/125029 · Cited by 513 · cited 2×
“While the current pressure sensors are used as on-off switches in the current prototype, they can be replaced with more accurate sensors, such as [25], that can read actual foot pressures.”
BackgroundSelf-citationAn Ankle-Foot Orthosis Powered by Artificial Pneumatic Muscles
Ferris, D. P., Czerniecki, J. M., & Hannaford, B. (2005) · Journal of Applied Biomechanics · doi:10.1123/jab.21.2.189 · Cited by 376 · cited 1×
“McKibben pneumatic artificial muscle actuators with carbon fiber reinforced foot and leg frames have been implemented for supplementing weak muscles for both dorsiflexion and plantarflexion [8], [11].”
MethodMuscular Atrophy Following Immobilisation: A Review
Appell, H.-J. (1990) · Sports Medicine · doi:10.2165/00007256-199010010-00005 · Cited by 362 · cited 1×
“However, long-term use of this type of passive orthoses makes the user more physically dependent on the device, as the immobilization or misuse of specific muscles may induce weakness and atrophy [1], [10], [19].”
BackgroundUser‐adaptive control of a magnetorheological prosthetic knee
Herr, H., & Wilkenfeld, A. (2003) · Industrial Robot: An International Journal · doi:10.1108/01439910310457706 · Cited by 293 · cited 1×
“While active devices require relatively complicated systems compared to passive braces, with appropriate control, they can provide higher independence, as seen with active prosthetic legs [6], [14], [16], [21].”
BackgroundMechanical performance of artificial pneumatic muscles to power an ankle–foot orthosis
Gordon, K. E., Sawicki, G. S., & Ferris, D. P. (2006) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2005.05.018 · Cited by 223 · cited 1×
“McKibben pneumatic artificial muscle actuators with carbon fiber reinforced foot and leg frames have been implemented for supplementing weak muscles for both dorsiflexion and plantarflexion [8], [11].”
MethodHow does the human brain deal with a spinal cord injury?
Bruehlmeier, M., Dietz, V., Leenders, K. L., Roelcke, U., Missimer, J., & Curt, A. (1998) · European Journal of Neuroscience · doi:10.1046/j.1460-9568.1998.00454.x · Cited by 214 · cited 1×
“Neural adaptation of motor representation with spinal cord injury or immobilization has been discussed in [5], [12].”
BackgroundManufacture of Passive Dynamic Ankle–Foot Orthoses Using Selective Laser Sintering
Faustini, M. C., Neptune, R. R., Crawford, R. H., & Stanhope, S. J. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.912638 · Cited by 209 · cited 1×
“The most widely used solution is to wear a passive orthotic device, such as an ankle-foot orthosis (AFO) [3], [7], [22], which forces the joint angle to be close to 90◦.”
Background
Research works printed but not linked: 3
- kastner, 1999, What are the benefits of the Otto Bock C-Ieg? A comparative gait analysis of C-Ieg, 3R45 and 3R80 (Was kann das C-Ieg?-Ganganalytischer vergleigh von C-Ieg, 3R45 and 3R80), Med Orthop Tech, 119, 131
- dietl, 1997, The application of electronics in prosthetics for lower extremities (Der einsatz von elektronik bei prothesen ZUI versorgung der unteren extremitat), Med Orthop Tech, 117, 31
- krebs, 2005, Rehabilitation robotics, orthotics, and prosthetics; chapter 48, Textbook of Neural Repair and Rehabilitation
Cited by
Design and Fabrication of Soft Artificial Skin Using Embedded Microchannels and Liquid Conductors
Park, Y.-L., Chen, B.-R., & Wood, R. J. (2012) · IEEE Sensors Journal · doi:10.1109/jsen.2012.2200790 · Cited by 776
“[30], [44], human-friendly robots for human-robot interactions (HRI) [45], and human-computer interface (HCI).”
BackgroundSelf-citationFlexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics
Heng, W., Solomon, S., & Gao, W. (2022) · Advanced Materials · doi:10.1002/adma.202107902 · Cited by 609
Design and control of a bio-inspired soft wearable robotic device for ankle–foot rehabilitation
Park, Y.-L., Chen, B.-R., Pérez-Arancibia, N. O., Young, D., Stirling, L., Wood, R. J., Goldfield, E. C., & Nagpal, R. (2014) · Bioinspiration & Biomimetics · doi:10.1088/1748-3182/9/1/016007 · Cited by 497
“More details on the control hardware can be found in [2].”
BackgroundSelf-citationA biologically inspired soft exosuit for walking assistance
Asbeck, A. T., De Rossi, S. M. M., Holt, K. G., & Walsh, C. J. (2015) · The International Journal of Robotics Research · doi:10.1177/0278364914562476 · Cited by 450
“Two of these robots are active soft orthotics that can provide slowly varying forces to the ankle (Park et al., 2011) or ankle and knee (Stirling et al.”
BackgroundSoft robotic sleeve supports heart function
Roche, E. T., Horvath, M. A., Wamala, I., Alazmani, A., Song, S.-E., Whyte, W., Machaidze, Z., Payne, C. J., Weaver, J. C., Fishbein, G., et al. (2017) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf3925 · Cited by 425
“Ecoflex 00-30 silicone was selected as the matrix for fabrication of the soft actuated material because it can generate large strains and because its elastic modulus (125 kPa) (14) is within the range of reported values for passivemyocardial tissue (42).”
Background
Metadata
- Authors
- Park, Y.-L., Chen, B.-R., Young, D., Stirling, L., Wood, R. J., Goldfield, E., & Nagpal, R.
- Affiliations
- Harvard University (US)
- Boston Children's Hospital (US)
- Venue
- 2011 IEEE/RSJ International Conference on Intelligent Robots and Systems
- Publisher
- IEEE
- Published
- 2011-09
- Pages
- 4488-4495
- ISSN
- 2153-0858 (ISSN-L)
- OpenAlex
- W2112427939
- Semantic Scholar
- 56152583e9b0c6375deed54abd785b0f21a74827
- Cited by
- 130
- References
- 24
- Citations by year
- 2026: 3 · 2025: 6 · 2024: 5 · 2023: 7 · 2022: 5 · 2021: 8 · 2020: 14 · 2019: 9 · 2018: 9 · 2017: 12 · 2016: 12 · 2015: 9 · 2014: 8 · 2013: 11 · 2012: 9
- FWCI
- 20.29
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS) (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Prosthetics and Rehabilitation Robotics
- Physical Sciences › Engineering › Biomedical Engineering › Soft Robotics and Applications
- Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
- Keywords
- ankle-foot orthosis, pneumatic artificial muscles, soft robotics, biomimetic design, neuromuscular disorders, gait rehabilitation, ankle joint, gait analysis, ankle inversion and eversion, feedback control
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.