Optimal Frequency for Wireless Power Transmission Into Dispersive Tissue
The DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
RF wireless interface enables remotely-powered implantable devices. Current studies in wireless power transmission into biological tissue tend to operate below 10 MHz due to tissue absorption loss, which results in large receive antennas. This paper examines the range of frequencies that will optimize the tradeoff between received power and tissue absorption. It first models biological tissue as a dispersive dielectric in a homogeneous medium and performs full-wave analysis to show that the optimal frequency is above 1 GHz for small receive coil and typical transmit-receive separations. Then, it includes the air-tissue interface and models human body as a planarly layered medium. The optimal frequency is shown to remain in the GHz-range. Finally, electromagnetic simulations are performed to include the effect of load impedance and look at the matched power gain. The optimal frequency is in the GHz-range for mm-sized transmit antenna and shifts to the sub-GHz range for cm-sized transmit antenna. The multiple orders of magnitude increase in the operating frequency enables dramatic miniaturization of implantable devices.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Theoretical studies have shown that energy transfer implemented with these characteristics can be 2-3 orders of magnitude more efficient than near-field systems (23,26).”
From Wireless power transfer to deep-tissue microimplants (2014) · cites this paper as [26]
Citations
Cites
Greiner, W. (1998) · Theoretical Physics · doi:10.1007/978-1-4612-0587-6 · Cited by 6,657
Gabriel, S., Lau, R. W., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/003 · Cited by 4,175 · cited 4×
“1 The real part begins to vary 1 The parameters in [23] are for the 4-term Cole-Cole model which is a variant of the Debye relaxation model.”
BackgroundMethodInfluentialWaves and Fields in Inhomogenous Media
Chew, W. C. (1999) · doi:10.1109/9780470547052 · Cited by 3,986
Classical Electrodynamics, 3rd ed .
Jackson, J. D., & Fox, R. F. (1999) · American Journal of Physics · doi:10.1119/1.19136 · Cited by 3,073
Implanted Antennas Inside a Human Body: Simulations, Designs, and Characterizations
Kim, J., & Rahmat-Samii, Y. (2004) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2004.832018 · Cited by 757
Design of Implantable Microstrip Antenna for Communication With Medical Implants
Soontornpipit, P., Furse, C. M., & Chung, Y. C. (2004) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2004.831976 · Cited by 472
Geometric approach for coupling enhancement of magnetically coupled coils
Zierhofer, C. M., & Hochmair, E. S. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.503178 · Cited by 420
Design of radio-frequency powered coils for implant instruments
Ko, W. H., Liang, S. P., & Fung, C. D. F. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457921 · Cited by 374
A neuro-stimulus chip with telemetry unit for retinal prosthetic device
Liu, W., Vichienchom, K., Clements, M., DeMarco, S. C., Hughes, C., McGucken, E., Humayun, M. S., De Juan, E., Weiland, J. D., & Greenberg, R. (2000) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.871327 · Cited by 371
Feedback Analysis and Design of RF Power Links for Low-Power Bionic Systems
Baker, M. W., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.893180 · Cited by 317
Research works printed but not linked: 4
- donaldson, 1986, frequency-hopping in r.f. energy-transfer links, Electron Wireless World, 24
- galbraith, 1984, An implantable multichannel neural stimulator
- schuder, 1961, high-level electromagnetic energy transfer through a closed chest wall, IRE Int Conv Record, 9, 119
- 0, IE3D Version 12 2
Cited by
Items 301–400 of 469
Sileo, L., Martiradonna, L., Arcuti, P., Monti, G., Tasco, V., Maschio, M. D., Pruzzo, G., Bozzini, B., Tarricone, L., & Vittorio, M. D. (2013) · Microelectronic Engineering · doi:10.1016/j.mee.2013.02.089 · Cited by 5
Percutaneous and transcutaneous connections
Qu, Y., Vilathgamuwa, D. M., Bolle, E. C. L., & Jayathurathnage, P. K. S. (2017) · Elsevier eBooks · doi:10.1016/b978-0-12-810491-0.00021-7 · Cited by 6
Zhan, B., Xia, W., Xiong, C., & Liu, S. (2021) · International Conference on Electronic Packaging Technology (ICEPT) · doi:10.1109/icept52650.2021.9568095 · Cited by 5
“With the rapid development of technology, in the past few decades, many methods have been proposed for charging flexible wearable electronics, such as: near-field inductive coupling [1], near-field capacitive coupling [2], ultrasound [3], field electromagnetic coupling [4] and far-field electromagnetic coupling [5], e…”
MethodOptimization of Wireless Power Transfer via Magnetic Resonance in Different Media
Jonah, O. (2013) · doi:10.25148/etd.fi13042506 · Cited by 4
“A full wave analysis was performed in [90] in order to identify the optimal frequency for power transmission in biological media at GHz-range and at a depth of 2 cm.”
MethodAboualalaa, M., & Pokharel, R. K. (2025) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2025.3587000 · Cited by 5
“However, for low displacement current and small-size WPT systems, working at sub-gigahertz is e ffi cient [10], [26], [27].”
Zhang, L., Tang, L., Liu, L., Zhao, S., Guo, M., Aw, K. C., & Hu, A. P. (2024) · IEEE Journal of Emerging and Selected Topics in Industrial Electronics · doi:10.1109/jestie.2024.3408317 · Cited by 5
A 402 MHz and 1.73-VCE Resonance Regulating Rectifier With On-Chip Antennas for Bioimplants
Liu, G., & Hu, Y. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3523913 · Cited by 4
“According to the simulation result, the RX antenna can achieve 2.2 mW under the maximum specific absorption rate (SAR) stipulated by the FCC (1.6 W/kg), consistent with the discussion in [8].”
InfluentialCompact Dual-Band Dual-Polarized Implantable Antenna for Biomedical Applications
Huang, J., Liu, X., Fan, Y., Li, K., & Lin, J. (2024) · IEEE International Symposium on Microwave, Antenna, Propagation and EMC Technologies for Wireless Communications · doi:10.1109/mape62875.2024.10813699 · Cited by 3
“For implantable antennas, a balance between induced current loss and dielectric loss can be struck in these bands, realizing relatively better efficiency and gain [2], [3].”
Closed-Form Approximations for RF Radiation from Implantable Medical Devices
Gao, M., Šipuš, Z., Capstick, M. H., Kuster, N., & Skrivervik, A. K. (2025) · European Conference on Antennas and Propagation · doi:10.23919/eucap63536.2025.10999524 · Cited by 3
“This approach ultimately enables the development of design strategies and benchmarks for implantable antennas and provides rapid assessment for customized designs [9], [10], [11], [12].”
Energy Transfer for Implantable Electronics in the Electromagnetic Midfield
Ho, J. S., & Poon, A. S. Y. (2014) · Cited by 3
Self-citationAdaptive transmitting coil array for optimal power transfer in deeply implanted medical devices
Jian, J., & Stanaćević, M. (2016) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2016.7538976 · Cited by 3
“Recently, more researchers turn to high frequency region from hundreds of MHz to GHz [3].”
BackgroundCrosstalk-free magnetic MIMO communication using heterogeneous antenna array
Kim, H., & Choi, J.-W. (2016) · URSI Asia-Pacific Radio Science Conference · doi:10.1109/ursiap-rasc.2016.7883543 · Cited by 3
“Recently, many research groups have tried to solve these limitations [5-9].”
BackgroundAhmadi-Farsani, J., Sadjedi, H., & Ghaznavi‐Ghoushchi, M. B. (2018) · Integration · doi:10.1016/j.vlsi.2018.03.007 · Cited by 3
Optimized Transmitting Sources for Radiative Wireless Power Transmission With Lossy Media
Kim, J., & Nam, S. (2021) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2021.3118847 · Cited by 3
Guerrero, E., Rocha-Flores, P. E., Gutiérrez‐Heredia, G., Cogan, S. F., Voit, W., & Maeng, J. (2022) · Flexible and Printed Electronics · doi:10.1088/2058-8585/ac8492 · Cited by 2
Soares, Í. V., & Nikolayev, D. (2022) · 2022 Wireless Power Week (WPW) · doi:10.1109/wpw54272.2022.9853957 · Cited by 3
Fiber-Optic Intracranial Pressure Monitoring System Using Wi-Fi—An In Vivo Study
da Silva, E. B., Hamasaki, E. E., Smaili, H. Y., Wozniak, A., Tristão, E. S. Y., Loureiro, M. D. P., Milano, J. B., de Meneses, M. S., de Oliveira, R. M., & Ramina, R. (2022) · Neurosurgery · doi:10.1227/neu.0000000000002250 · Cited by 3
Joshi, A. C., Dash, J. C., & Sarkar, D. (2023) · IEEE Microwaves, Antennas, and Propagation Conference (MAPCON) · doi:10.1109/mapcon58678.2023.10506221 · Cited by 3
“It is shown in [10] that the optimum performance of radiative WPT is obtained in the gigahertz and sub-gigahertz frequency range hence the size of the implant can be reduced significantly.”
BackgroundA Time-Efficient Model for Estimating Far-Field Wireless Power Transfer to Biomedical Implants
O'Callaghan, B., Gawade, D. R., Kumar, S., O’Hare, D., & Buckley, J. L. (2024) · European Conference on Antennas and Propagation · doi:10.23919/eucap60739.2024.10500964 · Cited by 3
Link Budget Estimation for Implantable Antennas: From In-Body Coupling to Free-Space Radiation
Gao, M., Raman, S., Šipuš, Z., & Skrivervik, A. K. (2024) · European Conference on Antennas and Propagation · doi:10.23919/eucap60739.2024.10501759 · Cited by 3
Morag, Y., & Levron, Y. (2019) · Instruments · doi:10.3390/instruments3020026 · Cited by 3
“Among the existing literature, references [2,30,31] did address HF aspects of induction power transfer.”
BackgroundMethodResultInfluentialA Novel Optimal Transmitting Source for Midfield Wireless Powering of Micro-implant
Nguyen, N. T. B., Ha‐Van, N., Le, M. T., & Seo, C. (2020) · IEEE Wireless Power Transfer Conference · doi:10.1109/wptc48563.2020.9295627 · Cited by 3
Kan‐Dapaah, K., Rahbar, N., Theriault, C., & Soboyejo, W. O. (2015) · Journal of Applied Physics · doi:10.1063/1.4918271 · Cited by 3
High-efficiency transcutaneous wireless energy transfer for biomedical applications
Xue, R.-F., Khannur, P. B., Cha, H., Zheng, Y., & Je, M. (2011) · Defense Science Research Conference and Expo (DSR) · doi:10.1109/dsr.2011.6026860 · Cited by 5
“However, as frequency lies in nna and shifts to tenna [10].”
BackgroundElectromagnetic Energy and Data Transfer in Biological Tissues Using Loop Antennas
Rezaei, N., Majumdar, D., Cockburn, B. F., & Schlegel, C. B. (2013) · Procedia Computer Science · doi:10.1016/j.procs.2013.06.124 · Cited by 3
Avestruz, A.-T., Rinehart, M., & Leeb, S. B. (2014) · IEEE ... Workshop on Control and Modeling for Power Electronics · doi:10.1109/compel.2014.6877192 · Cited by 3
In-vitro RF characterization of implantable telemetry system
Fricke, K., Wang, Z., & Sobot, R. (2014) · Analog Integrated Circuits and Signal Processing · doi:10.1007/s10470-014-0422-x · Cited by 3
“Extensive research has been documented outlying various tissue properties and standards [3–5, 11–13, 19, 21].”
BackgroundDing, K., Yu, Y., & Lin, H. R. (2015) · IEEE International Conference on Communication Technology (ICCT) · doi:10.1109/icct.2015.7399891 · Cited by 3
Wireless power transmission in human tissue for nerve stimulation
Salama, R. S., Kharkovsky, S., Liyanapathirana, R., & Gunawardana, U. (2016) · IET Microwaves Antennas & Propagation · doi:10.1049/iet-map.2015.0285 · Cited by 3
Millimeter-Scale True 3-D Antenna-in-Package Structures for Near-Field Power Transfer
Gadfort, P., & Franzon, P. D. (2014) · IEEE Transactions on Components Packaging and Manufacturing Technology · doi:10.1109/tcpmt.2014.2349983 · Cited by 3
“For simplicity, the surrounding material will be assumed to be air, but it is possible to include models of human tissues [2] for biomedical applications.”
BackgroundWireless power transfer for microchip implants
Yang, Y. W., Nie, Z., Qi, X., & Tan, J. (2016) · Progress in Electromagnetic Research Symposium · doi:10.1109/piers.2016.7735870 · Cited by 3
Inductive and ultrasonic wireless power transmission to biomedical implants
Ibrahim, A., Meng, M., & Kiani, M. (2017) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2017.8050939 · Cited by 5
“To improve PTE in mm-sized implants, fp has been increased to the sub-GHz range and below [6]-[9].”
BackgroundEffect of A Superstrate on On-Head Matched Antennas for Biomedical Applications
Rokunuzzaman, M., Ahmed, A., Alomainy, A., & Rowe, W. S. T. (2020) · Electronics · doi:10.3390/electronics9071099 · Cited by 3
“This frequency was targeted as the lower microwave frequency region is reported to be more suitable compared to higher microwave frequencies to achieve increased signal propagation inside human tissue [22].”
BackgroundSun, Y., Wang, X., Li, Y., & Mengjun, W. (2021) · Global Symposium on Millimeter-Waves · doi:10.1109/gsmm53250.2021.9512013 · Cited by 3
Magnetoelectric (ME) Antenna for On-chip Implantable Energy Harvesting
Nasrollahpour, M., Zaeimbashi, M., Khalifa, A., Liang, X., Chen, H., Sun, N., Abrishami, S. M. S., Martos-Repath, I., Emam, S., Cash, S. S., et al. (2021) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc46164.2021.9629823 · Cited by 3
“It is shown in [8] that low GHz frequencies are the optimum frequency for energy harvesting in body tissues.”
BackgroundTrezise, T. (2011) · UVic’s Research and Learning Repository (University of Victoria) · Cited by 5
Joshi, A. C., Dash, J. C., & Sarkar, D. (2022) · doi:10.36227/techrxiv.21641654 · Cited by 5
State-of-the-Art Technology on Highly Miniaturized Free-Floating Neural Implants
Khalifa, A., & Etienne‐Cummings, R. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_114 · Cited by 3
Optimal position of the transmitter coil for wireless power transfer to the implantable device
Jian, J., & Stanaćević, M. (2014) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2014.6945128 · Cited by 2
“Recently, it was shown that the optimal transmission frequency through a biological tissue in case of mm-size coils is in GHz range [6].”
BackgroundNikolayev, D., Zhadobov, M., Karban, P., & Sauleau, R. (2017) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-981-10-5122-7_59 · Cited by 2
“The radiation efficiency η of an in-body antenna is frequency-dependent [11].”
BackgroundSong, L., & Rahmat‐Samii, Y. (2021) · 2021 International Applied Computational Electromagnetics Society (ACES-China) Symposium · doi:10.23919/aces-china52398.2021.9581908 · Cited by 2
Analytical Approach to Matching Layer Design for Electric Field Maximization in Biological Tissues
Gasperini, D., Costa, F., Daniel, L., Manara, G., & Genovesi, S. (2022) · URSI Atlantic Radio Science Conference · doi:10.23919/at-ap-rasc54737.2022.9814183 · Cited by 2
“Indeed, according to [12], a good trade off between received power and tissue absorption can be obtained in the GHz-range, if the radiating antenna is shorter than a wavelength.”
BackgroundRadiation Performance of Antennas Implanted in Small Animals for Neuroengineering
Gao, M., Rosenthal, J., Wu, K., Šipuš, Z., Lacour, S. P., & Skrivervik, A. K. (2023) · European Conference on Antennas and Propagation · doi:10.23919/eucap57121.2023.10133552 · Cited by 2
Bui, G. T., Nguyen, D., & Seo, C. (2023) · European Microwave Conference (EuMC) · doi:10.23919/eumc58039.2023.10290292 · Cited by 2
“In biomedical implants with tissue environment, the frequency will be shifted due to variations in the electrical properties [7].”
BackgroundUltimate performance of microwave tissue ablation
Lee, S., Lee, J., Poon, A. S. Y., & Kim, S. (2024) · Physical Review Applied · doi:10.1103/physrevapplied.22.014018 · Cited by 2
Self-citationAnalytical Equivalent Circuit Extraction of Foreign Metal Objects in WPT Systems
Shirai, R., 中尾, 悟., & Hashimoto, M. (2024) · IEEE Access · doi:10.1109/access.2024.3500008 · Cited by 2
NON-COIL, OPTIMAL SOURCES FOR WIRELESS POWERING OF SUB-MILLIMETER IMPLANTABLE DEVICES
Kim, S., Ho, J. S., & Poon, A. S. Y. (2017) · Electromagnetic waves · doi:10.2528/pier16092301 · Cited by 2
“Recently, it has been shown that by operating at GHz-range frequencies, much higher power transfer efficiency can be achieved for a mm-scaled receiver [6].”
Self-citationDesign, modeling, and fabrication of mm3 three-dimensional integrated antennas
Gadfort, P., & Franzon, P. D. (2012) · IEEE Electronic Components and Technology Conference (ECTC) · doi:10.1109/ectc.2012.6249081 · Cited by 2
Investigation of material effects on near-field wireless power transfer
Yoon, I., & Ling, H. (2012) · doi:10.1109/aps.2012.6349072 · Cited by 2
“There are several works that have been reported thus far regarding material effects on wireless power transfer based on numerical simulations or measurements [5]-[7].”
BackgroundPalacios, S., Ramesh, N., Rida, A., Tentzeris, M. M., & Nikolaou, S. (2013) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/aps.2013.6711755 · Cited by 2
Implantable intraocular pressure sensor using principles of electromagnetic radiation
Arango, M. A. P., Reyes, J. C. B., Rodrigez, F. R., Unigarro, E. A., Florian, A. U. A., & Segura-Quijano, F. E. (2014) · IEEE IberoAmerican Congress on Sensors (IBERSENSOR) · doi:10.1109/ibersensor.2014.6995543 · Cited by 2
Noubir, G., & Javaheri, H. (2012) · Cited by 2
Dual band electrically small non-uniform pitch ellipsoidal helix antenna for cardiac pacemakers
Huang, H., Chen, P., Mauro Ferrari, B., Hu, Y., & Akinwande, D. (2013) · IEEE Topical Conference on RF/Microwave Power Amplifiers for Radio and Wireless Applications (PAWR) · doi:10.1109/pawr.2013.6490220 · Cited by 2
Testbed implementation of near-field magnetic MIMO communication system using SDR
Hwang, S., Kim, H., Kim, K. T., & Choi, J.-W. (2016) · International Symposium on Antennas and Propagation · Cited by 2
Ultrasonic wireless powering link of visual cortical prosthesis implant
Jaafar, B., Neasham, J. A., Chester, G., & Degenaar, P. A. (2017) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2017.8325115 · Cited by 2
“Different wireless power transfer systems have been proposed where the Transcutaneous wireless power transfer has been extensively targeted such as RF link [3], inductive link [4], optical link [5] and ultrasonic link [6].”
BackgroundMagneto-Inductive Powering and Uplink of In-Body Microsensors: Feasibility and High-Density Effects
Dumphart, G., Bitachon, B. I., & Wittneben, A. (2019) · IEEE Wireless Communications and Networking Conference · doi:10.1109/wcnc.2019.8885956 · Cited by 4
“The chosen 750MHz are close to the suggestion of ≈ 1GHz by [17] and sufficiently lowfrequency to penetrate a few cm of tissue.”
BackgroundRadiation Performance of Highly Miniaturized Implantable Devices
Nikolayev, D., Zhadobov, M., Joseph, W., Martens, L., & Sauleau, R. (2019) · European Microwave Conference (EuMC) · doi:10.23919/eumc.2019.8910944 · Cited by 2
“Next, we add 2-mm skin and 5-mm fat layers to the model as proposed in [33].”
MethodImplantable Interface for an Arm Neuroprosthesis
Imenes, K., Blystad, L.-C., Marchetti, L., Hønsvall, B. K., Øhlckers, P. A., Rabbani, S., Moldovan, C. A., Ionescu, O. N., Franti, E. D., Dascălu, M., et al. (2021) · European Microelectronics and Packaging Conference · doi:10.23919/empc53418.2021.9585011 · Cited by 2
Mitcheson, P. D., & Pucci, N. (2024) · Elsevier eBooks · doi:10.1016/b978-0-323-95478-5.00035-2 · Cited by 3
Design of Analog CMOS Circuits for Batteryless Implantable Telemetry Systems
Gannes, D., & Govindasamy, K. (2014) · Cited by 2
Pless, B., & Goodman, A. M. (2018) · Elsevier eBooks · doi:10.1016/b978-0-12-805353-9.00022-x · Cited by 2
Rittiplang, A., & Pijitrojana, W. (2016) · IJIREEICE · doi:10.17148/ijireeice.2016.4113 · Cited by 2
Wireless Implant Communications Using the Human Body
Teshome, A. K., Kibret, B., & Lai, D. T. H. (2017) · IGI Global eBooks · doi:10.4018/978-1-5225-2255-3.ch550 · Cited by 2
Miniature antennas for biomedical applications
Nikolayev, D. (2017) · theses.fr (ABES) · doi:10.70675/6323cff1za131z4e8bzb5dbzfa2f4d2e65ce · Cited by 3
Shaw, T., Mandal, B., & Augustine, R. (2025) · European Conference on Antennas and Propagation · doi:10.23919/eucap63536.2025.10999887 · Cited by 1
“While higher-frequency signals can theoretically support increased data rates, they are typically unsuitable for IBC due to the high attenuation caused by the lossy nature of biological tissues, which severely degrade signal integrity [11], [12].”
Advanced energy storage systems as power sources for biosensing technologies
Pereira, C. R., Pereira, A. M., Teixeira, J. S., Sousa, A. R., Queirós, G. P., Costa, R. S., Nunes, M., & Rocha, M. (2025) · Progress in molecular biology and translational science · doi:10.1016/bs.pmbts.2025.05.012 · Cited by 1
Battery-less implantable sensor for real-time intrasac pressure monitoring in EVAR-treated AAAs
Saroglia, G., Diana, S., Marocco, S., Metaldi, M., Nicoloso, S., Tozzi, M., & Stefanini, I. (2025) · Advanced Sensor and Energy Materials · doi:10.1016/j.asems.2025.100170 · Cited by 1
Mocan, B., Mocan, M., Fulea, M., Murar, M., Mate, Z., Calboréan, A., & Bintintan, V. V. (2026) · Diagnostics · doi:10.3390/diagnostics16091318 · Cited by 1
“Ultra-high-frequency (UHF) RFID at 860–960 MHz transitions from reactive near-field to radiative far-field coupling, enabling extended free-space ranges (several metres) and the smallest transponder form factors [32].”
Smart sensors for the early detection of periprosthetic joint infection: A translational perspective
Koucheki, R., Tang, J., Lex, J. R., Lameire, D. L., Raleigh, M. J., Nousiainen, M. T., Wolfstadt, J. I., Whyne, C., & Ravi, B. (2026) · Acta Biomaterialia · doi:10.1016/j.actbio.2026.06.012 · Cited by 1
Wearable and Implantable Technologies for Cardiovascular Health Informatics
Zheng, Y., Poon, C. C. Y., & Zhang, Y. (2013) · Springer eBooks · doi:10.1007/978-1-4614-8495-0_22 · Cited by 1
Beam focused slot antenna for microchip implants
Tanabe, Y., Wong, H., Kim, S., Ho, J. S., & Poon, A. S. Y. (2012) · International Symposium on Antennas and Propagation · Cited by 1
Self-citationDual band electrically small non-uniform pitch ellipsoidal helix antenna for cardiac pacemakers
Huang, H., Chen, P., Mauro Ferrari, B., Hu, Y., & Akinwande, D. (2013) · IEEE Topical Meeting on Silicon Monolithic Integrated Circuits in RF Systems (SiRF) · doi:10.1109/sirf.2013.6489487 · Cited by 1
“It has been shown that 1 GHz is the optimal frequency range for wireless power transmission through a lossy and dispersive dielectric of human body tissue [7].”
Transcutaneous power delivery and safety
Jegadeesan, R., Guo, Y., Yen, S.-C., & Thakor, N. V. (2014) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/aps.2014.6904928 · Cited by 1
“Most published works use low frequency (up to few tens of MHz) operation in accordance with classic work [3], however recent work [4] recommends high frequency of operation to reduce tissue losses.”
BackgroundOPTIMIZATION ON WIRELESS POWER TRANSFER
Halim, M. A. A., Razak, M. S., Yasin, M. N. M., Khairunizam, W., Fareq, M., & Fitra, M. (2015) · Jurnal Teknologi · doi:10.11113/jt.v77.6776 · Cited by 1
“At this point we take a short detour to find the directivity of the infinitesimal dipole [3].”
BackgroundEfficient dual-band energy harvesting system for implantable biosensors
Mouris, B. A., Soliman, A. M., Ali, T. A., Eshrah, I. A., & Badawi, A. (2016) · International Symposium on Antenna Technology and Applied Electromagnetics · doi:10.1109/antem.2016.7550242 · Cited by 1
Bocan, K. N., & Sejdić, E. (2016) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2016.7539159 · Cited by 1
Dual band electrically small non-uniform pitch ellipsoidal helix antenna for cardiac pacemakers
Huang, H., Chen, P., Mauro Ferrari, B., Hu, Y., & Akinwande, D. (2013) · IEEE Radio and Wireless Symposium (RWS) · doi:10.1109/rws.2013.6486729 · Cited by 1
“It has been shown that 1 GHz is the optimal frequency range for wireless power transmission through a lossy and dispersive dielectric of human body tissue [7].”
BackgroundPerformance analysis of ultra small antenna in body-air channel
Kang, J., Lee, J., Suh, J., Kim, S. J., & Kim, S.-C. (2017) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/apusncursinrsm.2017.8072240 · Cited by 1
“The standard in-body pathloss channel model [2] is established from simulation results, and the optimal frequency for implantable device has been derived using Debye biological media model [3].”
MethodFast Estimation of Wireless Charging System in a Large Tissue
Jeong, N. S., Wilken-Resman, E., & von Novak, B. (2018) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/apusncursinrsm.2018.8608271 · Cited by 1
“On the other hand, meta-surface lenses have been designed and demonstrated at a low GHz in the mid-field range [4].”
BackgroundWireless Neuromodulation: From Bench to Bedside
Perryman, L. T. (2019) · IntechOpen eBooks · doi:10.5772/intechopen.85501 · Cited by 1
The Wireless Power Transfer System Applied to an Implantable Deep Brain Stimulator
Wei, J., Chen, X., Jin, H. S., & Li, J. (2019) · International Congress on Image and Signal Processing (CISP) · doi:10.1109/cisp-bmei48845.2019.8965928 · Cited by 2
“INTRODUCTION Traditional implantable deep brain stimulators require regular surgery to replace batteries because of limited battery capacity [1-7].”
BackgroundWireless Bioelectronic Interfaces Electromagnetic Performance and Safety
Kim, H., Dong, Z., & Ho, J. S. (2023) · Handbook of Neuroengineering · doi:10.1007/978-981-16-5540-1_24 · Cited by 1
Towards Localization of Miniaturized Medical Robots With Microwaves
Lin, Y.-H., Daguerre, H., Lavrenko, A., & Misra, S. (2024) · IEEE Sensors Letters · doi:10.1109/lsens.2024.3412400 · Cited by 1
“Signals in this frequency range are able to reach a deeply situatedtargetinthehumanbody[17].”
BackgroundA 400 MHz Voltage-Mode-Based Fully Integrated Regulating Rectifier for Deep Tissue Bio-implants
Liu, G., & Hu, Y. (2024) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas58744.2024.10558169 · Cited by 1
“7(b), the RX antenna can achieve 2.2 mW under the maximum specific absorption rate (SAR) stipulated by FCC (1.6 W/kg), consistent with the discussion in [6].”
MethodInfluentialFernandes, J., Mendes, P. M., & Abreu, C. (2015) · International Conference on Signal Processing and Integrated Networks (SPIN) · doi:10.1109/spin.2015.7095439 · Cited by 1
Wireless powering of microchip implants by a cross-slot antenna
Tanabe, Y., Ho, J. S., Wong, H., & Poon, A. S. Y. (2012) · Asia-Pacific Microwave Conference (APMC) · doi:10.1109/apmc.2012.6421617 · Cited by 1
“Recently, it was shown that when the receiver is much smaller than the transmitter, high efficiency can be obtained by operating in the low-GHz range despite increased tissue loss [2].”
BackgroundSelf-citationLee, J., Song, K., & Choi, J. (2016) · Microwave and Optical Technology Letters · doi:10.1002/mop.29735 · Cited by 1
Implantable Wireless Medical Devices for Gastroesophageal Applications
Rao, S., & Chiao, J. ‐. (2017) · Cambridge University Press eBooks · doi:10.1017/9781107297302.001 · Cited by 1
Wireless Power Transfer to Implantable Medical Devices With Multi-Layer Planar Spiral Coils
Artan, N. S., & Amineh, R. K. (2018) · Advances in wireless technologies and telecommunication book series · doi:10.4018/978-1-5225-5870-5.ch009 · Cited by 1
Compact coplanar waveguide fed wideband directional antenna for medical diagnosis
Rokunuzzaman, M., Ahmed, A., Baum, T. C., & Rowe, W. S. T. (2020) · Journal of Physics D Applied Physics · doi:10.1088/1361-6463/abb0bb · Cited by 1
Wireless Power Transfer to Implantable Medical Devices With Multi-Layer Planar Spiral Coils
Artan, N. S., & Amineh, R. K. (2021) · IGI Global eBooks · doi:10.4018/978-1-7998-8050-9.ch023 · Cited by 1
Optimal Frequency for Implantable Mid-field Wireless Power Transfer
Tang, D.-F., & Chen, Z.-Y. (2022) · International Conference on Emerging Technologies · doi:10.1109/icet55676.2022.9825368 · Cited by 1
Compact rectenna array on-package for wirelessly powered small device
Zhang, J. W. (2015) · doi:10.32657/10356/62196 · Cited by 1
Ghosal, S., De, A., Shubair, R. M., & Chakrabarty, A. (2019) · arXiv (Cornell University) · doi:10.48550/arxiv.1907.12127 · Cited by 1
Freedman, D. S., Şahin, M., & Towe, B. C. (2022) · Springer eBooks · doi:10.1007/978-1-0716-1006-0_605 · Cited by 1
Zhao, Y. (2015) · D-Scholarship@Pitt (University of Pittsburgh)
New Concepts in Human Telemetry
Carrara, S. (2013) · Springer eBooks · doi:10.1007/978-1-4614-1674-6_4
Swain, B., Kumar, K., Kar, D. P., Nayak, P. P., Bhuyan, S., & Sharma, R. (2015) · International Conference on Electrical, Electronics, Signals, Communication and Optimization · doi:10.1109/eesco.2015.7253787
A novel biotelemetry system to monitor human vital signs
Arcuti, P., Monti, G., Tarricone, L., Sileo, L., Martiradonna, L., & Vittorio, M. D. (2012) · International Symposium on Electromagnetic Compatibility - EMC Europe (EMC Europe) · doi:10.1109/emceurope.2012.6396791
“As reported in [7], the values of the operating frequency that allow optimizing the tradeoff between tissue absorption and received power depend on the antenna dimensions.”
Background
Metadata
- Authors
- Poon, A. S. Y., O'Driscoll, S., & Meng, T. H.
- Affiliations
- Stanford University (US)
- ORCID
- Poon, Ada S Y
- Venue
- IEEE Transactions on Antennas and Propagation
- Published
- 2010-05
- Volume
- 58
- Issue
- 5
- Pages
- 1739-1750
- ISSN
- 0018-926X (ISSN-L)
- 1558-2221
- OpenAlex
- W2110545851
- Semantic Scholar
- e6a37dc9b793041a01befbefb870dde9882038d3
- Cited by
- 469
- References
- 25
- Citations by year
- 2026: 7 · 2025: 25 · 2024: 32 · 2023: 31 · 2022: 30 · 2021: 42 · 2020: 23 · 2019: 38 · 2018: 27 · 2017: 36 · 2016: 34 · 2015: 32 · 2014: 33 · 2013: 31 · 2012: 29
- FWCI
- 19.92
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Antennas and Propagation (publishedVersion)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Energy Harvesting in Wireless Networks
- Keywords
- wireless power transfer, biomedical implants, optimal operating frequency, air-tissue interface, load impedance matching, RF power transmission
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.