Optimal Frequency for Wireless Power Transmission Into Dispersive Tissue
The DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index.
Abstract
RF wireless interface enables remotely-powered implantable devices. Current studies in wireless power transmission into biological tissue tend to operate below 10 MHz due to tissue absorption loss, which results in large receive antennas. This paper examines the range of frequencies that will optimize the tradeoff between received power and tissue absorption. It first models biological tissue as a dispersive dielectric in a homogeneous medium and performs full-wave analysis to show that the optimal frequency is above 1 GHz for small receive coil and typical transmit-receive separations. Then, it includes the air-tissue interface and models human body as a planarly layered medium. The optimal frequency is shown to remain in the GHz-range. Finally, electromagnetic simulations are performed to include the effect of load impedance and look at the matched power gain. The optimal frequency is in the GHz-range for mm-sized transmit antenna and shifts to the sub-GHz range for cm-sized transmit antenna. The multiple orders of magnitude increase in the operating frequency enables dramatic miniaturization of implantable devices.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Theoretical studies have shown that energy transfer implemented with these characteristics can be 2-3 orders of magnitude more efficient than near-field systems (23,26).”
From Wireless power transfer to deep-tissue microimplants (2014) · cites this paper as [26]
Citations
Cites
Greiner, W. (1998) · Theoretical Physics · doi:10.1007/978-1-4612-0587-6 · Cited by 6,657
Gabriel, S., Lau, R. W., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/003 · Cited by 4,175 · cited 4×
“1 The real part begins to vary 1 The parameters in [23] are for the 4-term Cole-Cole model which is a variant of the Debye relaxation model.”
BackgroundMethodInfluentialWaves and Fields in Inhomogenous Media
Chew, W. C. (1999) · doi:10.1109/9780470547052 · Cited by 3,986
Classical Electrodynamics, 3rd ed .
Jackson, J. D., & Fox, R. F. (1999) · American Journal of Physics · doi:10.1119/1.19136 · Cited by 3,073
Implanted Antennas Inside a Human Body: Simulations, Designs, and Characterizations
Kim, J., & Rahmat-Samii, Y. (2004) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2004.832018 · Cited by 757
Design of Implantable Microstrip Antenna for Communication With Medical Implants
Soontornpipit, P., Furse, C. M., & Chung, Y. C. (2004) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2004.831976 · Cited by 472
Geometric approach for coupling enhancement of magnetically coupled coils
Zierhofer, C. M., & Hochmair, E. S. (1996) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.503178 · Cited by 420
Design of radio-frequency powered coils for implant instruments
Ko, W. H., Liang, S. P., & Fung, C. D. F. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457921 · Cited by 374
A neuro-stimulus chip with telemetry unit for retinal prosthetic device
Liu, W., Vichienchom, K., Clements, M., DeMarco, S. C., Hughes, C., McGucken, E., Humayun, M. S., De Juan, E., Weiland, J. D., & Greenberg, R. (2000) · IEEE Journal of Solid-State Circuits · doi:10.1109/4.871327 · Cited by 371
Feedback Analysis and Design of RF Power Links for Low-Power Bionic Systems
Baker, M. W., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.893180 · Cited by 317
Research works printed but not linked: 4
- donaldson, 1986, frequency-hopping in r.f. energy-transfer links, Electron Wireless World, 24
- galbraith, 1984, An implantable multichannel neural stimulator
- schuder, 1961, high-level electromagnetic energy transfer through a closed chest wall, IRE Int Conv Record, 9, 119
- 0, IE3D Version 12 2
Cited by
Items 101–200 of 469
Stuart, T., Kasper, K. A., Iwerunmor, I. C., McGuire, D. T., Peralta, R., Hanna, J., Johnson, M., Farley, M., LaMantia, T., Udorvich, P., et al. (2021) · Science Advances · doi:10.1126/sciadv.abj3269 · Cited by 46
A Low-Power 13.56 MHz RF Front-End Circuit for Implantable Biomedical Devices
Lee, S.-Y., Hong, J.-H., Hsieh, C., Liang, M., & Kung, J.-Y. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2212276 · Cited by 43
An RF-Powered FDD Radio for Neural Microimplants
Rajavi, Y., Taghivand, M., Aggarwal, K., Ma, A., & Poon, A. S. Y. (2017) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2016.2645601 · Cited by 43
“Unlike the common practice of RF-powering at the HF band through inductive coupling, wireless powering at low gigahertz frequencies has proven to deliver more power and penetrate deeper in tissue for small receiving coils [8]–[10].”
BackgroundSelf-citationNear Field Wireless Powering of Deep Medical Implants
Campi, T., Cruciani, S., Santis, V. D., Maradei, F., & Feliziani, M. (2019) · Energies · doi:10.3390/en12142720 · Cited by 43
“However, in MWP the frequency is much higher than 10 MHz and antennas are used for millimeter size receivers.”
BackgroundMethodInfluentialData Packet Transmission Through Fat Tissue for Wireless IntraBody Networks
Asan, N. B., Pérez-Penichet, C., Shah, S. R. M., Noreland, D., Hassan, E., Rydberg, A., Blokhuis, T. J., Voigt, T., & Augustine, R. (2017) · IEEE Journal of Electromagnetics RF and Microwaves in Medicine and Biology · doi:10.1109/jerm.2017.2766561 · Cited by 42
“tion due to the lossy nature of biological tissues [30], [31].”
BackgroundUWB Path Loss Models for Ingestible Devices
Perez-Simbor, S., Andreu, C., Garcia‐Pardo, C., Frasson, M., & Cardona, N. (2019) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2019.2891717 · Cited by 42
“Even though those are the defined frequencies for the different scenarios, many works have studied the optimum frequency bands for IB2OB communications [5]–[7].”
BackgroundSelf-Sustainable Biomedical Devices Powered by RF Energy: A Review
Alkhalaf, H. Y., Ahmad, M. Y., & Ramiah, H. (2022) · Sensors · doi:10.3390/s22176371 · Cited by 49
“In order to obtain high PTE in a deep implantation position, an appropriate operating frequency is selected depending on the depth of IMDs and tissue layer properties [182].”
BackgroundElectric near-field coupling for wireless power transfer in biomedical applications
Jegadeesan, R., Guo, Y., & Je, M. (2013) · IEEE MTT-S International Microwave Workshop Series on RF and Wireless Technologies for Biomedical and Healthcare Applications · doi:10.1109/imws-bio.2013.6756140 · Cited by 49
“This model has also been put to use in [4] for inductive links in biomedical implants.”
MethodMulti-disciplinary challenges in tissue modeling for wireless electromagnetic powering: a review
Bocan, K. N., Mickle, M. H., & Sejdic, E. (2017) · IEEE Sensors Journal · doi:10.1109/jsen.2017.2748338 · Cited by 41
“ious tissues at frequencies of interest, often in conjunction with simulated tissue models [13], [56].”
BackgroundWireless Powering Efficiency of Deep-Body Implantable Devices
Soares, Í. V., Gao, M., Cil, E., Šipuš, Z., Skrivervik, A. K., Ho, J. S., & Nikolayev, D. (2022) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2022.3231492 · Cited by 50
“The scattering parameters were measured using a two-port vector network analyzer (VNA) calibrated in the frequency range from 100 MHz to 5 GHz, in which the transmitter was connected to port 1 and the receiver to port 2.”
BackgroundResultInfluentialShaw, T., Mandal, B., Mitra, D., Rangaiah, P. K. B., Perez, M. D., & Augustine, R. (2023) · AEU - International Journal of Electronics and Communications · doi:10.1016/j.aeue.2023.155010 · Cited by 48
A Wideband Dual-Antenna Receiver for Wireless Recording From Animals Behaving in Large Arenas
Lee, S. B., Yin, M., Manns, J. R., & Ghovanloo, M. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2247603 · Cited by 41
“However, such frequencies are not useful for implantable devices due to significant absorption of high-frequency electromagnetic fields in the tissue [17].”
BackgroundBoutry, C. M., Chandrahalim, H., Streit, P., Schinhammer, M., Hänzi, A. C., & Hierold, C. (2012) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2012.08.039 · Cited by 39
Wireless Power Transmission Through Concrete Using Circuits Resonating at Utility Frequency of 60 Hz
Ishida, H., & Furukawa, H. (2014) · IEEE Transactions on Power Electronics · doi:10.1109/tpel.2014.2322876 · Cited by 39
“Since resonant energy transfer was first developed by the WiTricity project [2], there has been a great deal of research carried out using this and other approaches [3]– [5].”
BackgroundTissue Variability and Antennas for Power Transfer to Wireless Implantable Medical Devices
Bocan, K. N., Mickle, M. H., & Sejdić, E. (2017) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2017.2723391 · Cited by 38
“When the two antennas are not isolated, they must be analyzed simultaneously to account for loading effects of the implant in addition to effects of the tissue medium [16], [17], [20].”
BackgroundMethodInfluentialNoncontact Electromagnetic Wireless Recognition for Prosthesis Based on Intelligent Metasurface
Wang, H., Zhou, Y., Li, H., Liu, G., Yin, S. M., Li, P. J., Dong, S. Y., Gong, C. Y., Wang, S. Y., Li, Y. B., et al. (2022) · Advanced Science · doi:10.1002/advs.202105056 · Cited by 38
Investigation of Near-Field Wireless Power Transfer in the Presence of Lossy Dielectric Materials
Yoon, I., & Ling, H. (2012) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2012.2215296 · Cited by 36
Shah, I. A., Zada, M., Basir, A., Shah, S. A. A., Iman, U. R., Lim, Y., & Yoo, H. (2024) · IEEE Internet of Things Journal · doi:10.1109/jiot.2024.3491735 · Cited by 46
“In the case of deeply implanted medical devices, the selection of optimal frequency for maximum PTE is still a subject of ongoing research [16], primarily due to the complex anatomical structure of the human body [28], [29].”
The injectable neurostimulator: an emerging therapeutic device
Li, X., Serdijn, W. A., Zheng, W., Tian, Y., & Zhang, B. (2015) · Trends in biotechnology · doi:10.1016/j.tibtech.2015.04.001 · Cited by 38
Distributed sensor and actuator networks for closed-loop bioelectronic medicine
Bhave, G., Chen, J., Singer, A., Sharma, A., & Robinson, J. T. (2021) · Materials Today · doi:10.1016/j.mattod.2020.12.020 · Cited by 37
Wireless power transfer system applied to an active implantable medical device
Campi, T., Cruciani, S., Feliziani, M., & Hirata, A. (2014) · IEEE Wireless Power Transfer Conference · doi:10.1109/wpt.2014.6839612 · Cited by 36
Optimal Frequency for Wireless Power Transmission Into the Body: Efficiency Versus Received Power
Freeman, D. K., & Byrnes, S. J. (2019) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2019.2905672 · Cited by 33
“tissue heating at these frequencies [7]–[9].”
BackgroundTruong, B. D., & Roundy, S. (2020) · Smart Materials and Structures · doi:10.1088/1361-665x/ab90a2 · Cited by 35
“[53], 2010 RF 1 GHz 2× 2 mm(2) 40 mm 0.”
Moon, E., Barrow, M., Lim, J., Lee, J., Nason, S. R., Costello, J., Kim, H. S., Chestek, C., Jang, T., Blaauw, D., et al. (2021) · ACS Photonics · doi:10.1021/acsphotonics.1c00160 · Cited by 31
“11,12 Ultra-sound o ff ers a path to high e ffi ciency wireless power transfer, 1 − 3,8, 13 comparable to”
BackgroundDual-Band Transmitter for Wireless Power Transfer to Biomedical Implant Applications
Le-Huu, H., & Seo, C. (2023) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2023.3246163 · Cited by 29
“To cope with the problems of tissue losses and enhance PTE to a miniature Rx, an optimal operating frequency (fopt) equation was introduced and theoretically proved in [7], [11], and[12].”
BackgroundMethodWireless channel characterization for mm-size neural implants
Mark, M., Björninen, T., Chen, Y., Venkatraman, S., Ukkonen, L., Sydanheimo, L. T., Carmena, J. M., & Rabaey, J. M. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5626695 · Cited by 35
“Recent research has shown that, depending on the exact size of the antennas, the ideal frequency for wireless power transfer to a mm-sized antenna through the human cranium lies somewhere between 100 MHz and a few GHz [ 3 ].”
BackgroundWireless data links for next-generation networked micro-implantables
Wang, M. L., Baltsavias, S., Chang, T. C., Weber, M. J., Charthad, J., & Arbabian, A. (2018) · Custom Integrated Circuits Conference · doi:10.1109/cicc.2018.8357096 · Cited by 30
“2 – to first order, this implies lower signal attenuation for near-field links [18].”
BackgroundAnalysis of the Human Body as an Antenna for Wireless Implant Communication
Kibret, B., Teshome, A. K., & Lai, D. T. H. (2016) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2016.2526070 · Cited by 30
“alently efficient magnetic link can be obtained by operating in the low-GHz range and applying mm-sized coils [4], [5].”
BackgroundHa‐Van, N., & Seo, C. (2019) · IEEE Access · doi:10.1109/access.2019.2933260 · Cited by 30
Gasperini, D., Costa, F., Daniel, L., Manara, G., & Genovesi, S. (2022) · IEEE Antennas and Propagation Magazine · doi:10.1109/map.2022.3195466 · Cited by 28
“Even though lower frequencies are often preferred due to their greater penetration inside tissues, it has been shown that, when the transmitted antenna is shorter than a wavelength, the maximum transmitted power occurs in the gigahertz range [29], [30].”
BackgroundAnalytic Approximation of In-Body Path Loss for Implanted Antennas
Gao, M., Šipuš, Z., & Skrivervik, A. K. (2023) · IEEE Open Journal of Antennas and Propagation · doi:10.1109/ojap.2023.3276686 · Cited by 28
“perspective [23], [24], [25], [26], [27], [28], [29], [30], [31], [32], [33], [34].”
Goodarzy, F., Skafidas, E. S., & Gambini, S. (2014) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2014.2346487 · Cited by 31
“ternal electromagnetic field with a pressure field generated by an ultrasound transducer [65], [68], [73]–[77].”
BackgroundPowering mm-Size Wireless Implants for Brain-Machine Interfaces
Michael, M. (2011) · UC Berkeley · Cited by 28
Wang, W., Hemour, S., & Wu, K. (2014) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2014.2305401 · Cited by 27
“sorption [11], [12].”
Conformal antennas for miniature in-body devices: The quest to improve radiation performance
Nikolayev, D., Zhadobov, M., Karban, P., & Sauleau, R. (2017) · URSI Radio Science Bulletin · doi:10.23919/ursirsb.2017.8409427 · Cited by 30
Energy Harvesting and Power Delivery for Implantable Medical Devices
Tsui, C.-Y., Li, X., & Ki, W. (2013) · Foundations and Trends® in Electronic Design Automation · doi:10.1561/1000000029 · Cited by 25
A Systematic Method for Efficient Wireless Powering to Implantable Biomedical Devices
Cao, X., Sato, H., Xu, K., Jiang, W. T., Gong, S., & Chen, Q. (2023) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2023.3240005 · Cited by 26
“ciency and potential for dramatic miniaturization [17],”
BackgroundWireless Microstimulators for Neural Prosthetics
Şahin, M., & Pikov, V. (2011) · Critical Reviews in Biomedical Engineering · doi:10.1615/critrevbiomedeng.v39.i1.50 · Cited by 31
A Study of Multi-Layer Spiral Inductors for Remote Powering of Implantable Sensors
Olivo, J., Carrara, S., & Micheli, G. D. (2013) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2225620 · Cited by 28
Is submucosal bladder pressure monitoring feasible?
Basu, A. S., Majerus, S. J. A., Ferry, E. K., Makovey, I., Zhu, H., & Damaser, M. S. (2018) · Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · doi:10.1177/0954411918754925 · Cited by 26
Nie, Z., & Yang, Y. W. (2018) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2018.2794385 · Cited by 25
“Recently, the WPT techniques with operation frequency from sub- to low GHz have been drawing attention due to their much higher transfer efficiency than that of conventional near-field coupling [16]–[18].”
BackgroundSelf-reconfigurable micro-implants for cross-tissue wireless and batteryless connectivity
Abdelhamid, M. R., Chen, R., Cho, J., Chandrakasan, A. P., & Adib, F. (2020) · ACM Annual International Conference on Mobile Computing and Networking (MobiCom) · doi:10.1145/3372224.3419216 · Cited by 34
“contributions are orthogonal and can be combined with these past proposals if one wishes to operate the micro-implants at their other frequency bands (albeit operating at higher frequencies would introduce more attenuation [40]).”
BackgroundA Broadband High-Efficiency Rectifier for Mid-Field Wireless Power Transfer
Le, G., Nguyen, N. T. B., Au, N. D., & Seo, C. (2021) · IEEE Microwave and Wireless Components Letters · doi:10.1109/lmwc.2021.3077566 · Cited by 24
Efficient Compact Radiative Near-Field Wireless Power Transfer to Miniature Biomedical Implants
Le-Huu, H., Bui, G. T., & Seo, C. (2023) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2023.3298350 · Cited by 28
“significant loss due to the high values of tissue’s relative permittivity ( ε r ) and conductivity ( σ ) [8], [9].”
BackgroundAn Injection-Locked Wireless Power Transfer Transmitter With Automatic Maximum Efficiency Tracking
Feng, G., & Sit, J. (2020) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2020.3000102 · Cited by 23
“However, the energy absorbed by biological tissues and silicon substrates also increases with frequency [19].”
BackgroundAn Anatomical Model for the Simulation and Development of Subcutaneous Implantable Wireless Devices
Green, R. B., Hays, M. R., Mangino, M. J., & Topsakal, E. (2020) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2020.3008619 · Cited by 23
High-efficiency CMOS rectifier for fully integrated mW wireless power transfer
Zargham, M., & Gulak, P. G. (2012) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2012.6271912 · Cited by 22
“[1], [8].”
A NOVEL RECONFIGURABLE SPIRAL-SHAPED MONOPOLE ANTENNA FOR BIOMEDICAL APPLICATIONS
Salim, M., & Pourziad, A. (2015) · Progress In Electromagnetics Research Letters · doi:10.2528/pierl15083103 · Cited by 22
“Since each part of the body has an optimum frequency, for on-body wireless communication it is necessary to have different operating frequencies which enable the antenna to operate on various parts of the body [13].”
BackgroundA Single-Connector Stent Antenna for Intravascular Monitoring Applications
Liu, C., Chen, S.-C., & Hsiao, H. M. (2019) · Sensors · doi:10.3390/s19214616 · Cited by 22
“In fact, in the human body, electromagnetic waves (EM) have different pen tration attenuations and variable power loss boundaries at different f quencies [23,32].”
BackgroundInfluentialTheoretical Modeling and Design Guidelines for a New Class of Wearable Bio-Matched Antennas
Blauert, J., & Kiourti, A. (2019) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2019.2948727 · Cited by 22
“Indeed, numerous papers have demonstrated that it is possible to have reliable and optimally efficient data links within microwave frequencies to such devices [19]-[22].”
BackgroundEfficient Wireless Power Transfer With Capacitively Segmented RF Coils
Stoecklin, S., Yousaf, A., Gidion, G., & Reindl, L. M. (2020) · IEEE Access · doi:10.1109/access.2020.2971176 · Cited by 22
“[27], who proved that systems with mm-sized receiver coils, typically implemented as single-turn loop coils, reach their optimum efficiency in the ultra-high frequency (UHF) range at around 1GHz.”
BackgroundRF energy harvesting for intraoral orthodontic force monitoring
Li, P., Zhu, L., Ding, Y., Long, Z., Yang, Y., Pan, J., Gu, M., Wang, W., & Yang, Z. (2023) · Nano Energy · doi:10.1016/j.nanoen.2023.109244 · Cited by 23
Zhao, Y., Tang, L. J., Rennaker, R. L., Hutchens, C., & Ibrahim, T. S. (2013) · PLoS ONE · doi:10.1371/journal.pone.0077759 · Cited by 28
“sized implanted antennas is in the gigahertz range [46,47].”
BackgroundSoltani, N., ElAnsary, M., Xu, J., Filho, J. S., & Genov, R. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3125618 · Cited by 22
Soft, wireless and subdermally implantable recording and neuromodulation tools
Cai, L., & Gutruf, P. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abe805 · Cited by 24
A compact, low-power, fully analog implantable microstimulator
Khalifa, A., Zhang, J. J., Leistner, M., & Etienne‐Cummings, R. (2016) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2016.7539084 · Cited by 21
“This common belief has recently been challenged and it has been shown that the optimal operating radio-frequency (RF) for wirelesslypowered neural implants (often defined as the frequency at which the ratio of received power to total tissue absorption is maximal) is in the low GHz-range [16], [8].”
BackgroundA Multipolar Analysis of Near-Field Absorption and Scattering Processes
Liberal, I., Ederra, Í., Gonzalo, R., & Ziolkowski, R. W. (2013) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2013.2272453 · Cited by 20
“, the selection of the frequency of operation [10] and transmitter current distribution [11] for power transfer”
BackgroundOn the use of smart stents for monitoring in-stent restenosis
Keikhosravy, K., Zargaran-Yazd, A., & Mirabbasi, S. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346653 · Cited by 21
“As expected (and as analyzed in [16]), there is an optimum frequency for power gain transmission for smart stent at which the power transmission gain reaches –28 dB.”
BackgroundMiniaturized Wirelessly Powered and Controlled Implants for Multisite Stimulation
Habibagahi, I., Jang, J., & Babakhani, A. (2023) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2022.3233368 · Cited by 21
“The carrier frequency choice should be based on the application, safety, bandwidth (BW), and minimum required distance [25], [26], [27].”
BackgroundAC Computing Methodology for RF-Powered IoT Devices
Wan, T., Karimi, Y., Stanaćević, M., & Salman, E. (2019) · IEEE Transactions on Very Large Scale Integration (VLSI) Systems · doi:10.1109/tvlsi.2019.2894531 · Cited by 19
“the optimal frequency shifts to the order of 1 GHz [43].”
BackgroundHa‐Van, N., & Seo, C. (2018) · IEEE Wireless Power Transfer Conference · doi:10.1109/wpt.2018.8639272 · Cited by 18
“Beside the WPT techniques such as an inductive coupling system for very short distances of a few millimeters [6], an electromagnetic wave system with the huge absorption rate in the human body [7], a magnetic resonance WPT system is the most suitable type for implantable biomedical system.”
MethodOmi, A. I., Jiang, A., & Chatterjee, B. (2025) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2025.3531995 · Cited by 20
Maini, L., Genovés, V., Furrer, R., Cesarovic, N., Hierold, C., & Roman, C. I. (2024) · Microsystems & Nanoengineering · doi:10.1038/s41378-023-00632-x · Cited by 21
“Second, EM waves are strongly absorbed by the surrounding tissue 10 .”
BackgroundMethodCampi, T., Cruciani, S., Santilli, G. P., & Feliziani, M. (2015) · IEEE Wireless Power Transfer Conference · doi:10.1109/wpt.2015.7140172 · Cited by 18
Implanted Miniaturized Antenna for Brain Computer Interface Applications: Analysis and Design
Zhao, Y., Rennaker, R. L., Hutchens, C., & Ibrahim, T. S. (2014) · PLoS ONE · doi:10.1371/journal.pone.0103945 · Cited by 18
“It is noted however, that the ratio between received RF power and tissue absorption depends on the input impedance of the receive antenna [15].”
BackgroundAhmad, A., & Choi, D. (2025) · IEEE Internet of Things Journal · doi:10.1109/jiot.2025.3560366 · Cited by 17
“However, GHz-band operation faces substantial challenges when deployed within human tissue as a result of the elevated relative permittivity ((cid:2) r ) and conductivity (σ) of the tissue, causing significant signal loss [25], [43].”
Ultra-Fast and Efficient Design Method Using Deep Learning for Capacitive Coupling WPT System
Keshavarz, R., Majidi, E., Raza, A., & Shariati, N. (2023) · IEEE Transactions on Power Electronics · doi:10.1109/tpel.2023.3319505 · Cited by 20
“Specifically, when the Tx/Rx are limited to millimeter-size, high-frequency WPT systems are inevitable [20], [21], [22], [23], [24].”
BackgroundOptimal design of energy efficient inductive links for powering implanted devices
Cabrera, F. L., & de Sousa, F. R. (2014) · IEEE Topical Conference on Biomedical Wireless Technologies, Networks, and Sensing Systems (BioWireleSS) · doi:10.1109/biowireless.2014.6827731 · Cited by 17
“Most previous works use frequencies in the MHz range, however, the possibility of powering implants in frequencies of up to a few GHz has been studied [6].”
BackgroundChang, C., Hou, K.-C., Shieh, L.-J., Hung, S.-H., & Chiou, J. (2012) · Solid-State Electronics · doi:10.1016/j.sse.2012.05.020 · Cited by 18
Rothfuss, M. A., Unadkat, J. V., Gimbel, M. L., Mickle, M. H., & Sejdić, E. (2016) · Ultrasound in Medicine & Biology · doi:10.1016/j.ultrasmedbio.2016.11.005 · Cited by 17
Study of midfield wireless power transfer for implantable medical devices
Keerthi., K. S., Ilango., K., & Manjula., G. N. (2018) · 2018 2nd International Conference on Biomedical Engineering (IBIOMED) · doi:10.1109/ibiomed.2018.8534820 · Cited by 17
Das, D., Xu, Z., Nasrollahpour, M., Martos-Repath, I., Zaeimbashi, M., Khalifa, A., Mittal, A., Cash, S. S., Sun, N. X., Shrivastava, A., et al. (2023) · IEEE Open Journal of Circuits and Systems · doi:10.1109/ojcas.2023.3259233 · Cited by 18
“References [59] and [60] have shown that low gigahertz frequencies are well-suited for wireless power transfer to millimeter-sized implantable devices.”
BackgroundA 0.13µm CMOS integrated wireless power receiver for biomedical applications
Zargham, M., & Gulak, P. G. (2013) · European Solid-State Circuits Conference (ESSCIRC) · doi:10.1109/esscirc.2013.6649091 · Cited by 17
“The consequence of these two competing effects leads to an optimum frequency that maximizes the WPT efficiency through biological media [3], [4].”
BackgroundWireless Power Transmission Enabled by a Triboelectric Nanogenerator via a Magnetic Interaction
Cao, S., Zhang, H., Guo, R., Zhang, W., & Sang, S. (2019) · Energy Technology · doi:10.1002/ente.201900503 · Cited by 17
New compact design for short range wireless power transmission at 1GHz using H-slot resonators
Hekal, S. S., & Abdel-Rahman, A. B. (2015) · European Conference on Antennas and Propagation · Cited by 16
Shaw, T., Mandal, B., Samanta, G., Voigt, T., Mitra, D., & Augustine, R. (2024) · Scientific Reports · doi:10.1038/s41598-024-70591-4 · Cited by 19
Design of miniaturized wireless power receivers for mm-sized implants
Kim, C., Ha, S., Akinin, A., Park, J., Kubendran, R. C., Wang, H., Mercier, P. P., & Cauwenberghs, G. (2017) · Custom Integrated Circuits Conference · doi:10.1109/cicc.2017.7993703 · Cited by 18
“It is important to note that link efficiency from TX coil to RX coil is inversely proportional to the cube of the distance between TX and RX coils [18] and the cube of the square root of the area of RX coil [21] for the case of mm-sized RX coil.”
BackgroundAn energy harvested ultra-low power transceiver for Internet of Medical Things
Rajavi, Y., Taghivand, M., Aggarwal, K., Ma, A., & Poon, A. S. Y. (2016) · European Solid-State Circuits Conference (ESSCIRC) · doi:10.1109/esscirc.2016.7598260 · Cited by 16
“Contrary to the common practice of powering implants at LMHz through inductive coupling, a gigahertz wireless link has proven to deliver more power at greater depths inside a tissue, hence eliminating the LMHz coil antenna in favour of a smaller gigahertz antenna [3]-[5].”
Self-citationWang, Z., & Mirabbasi, S. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2904487 · Cited by 18
Biswas, D. K., Tasneem, N. T., & Mahbub, I. (2019) · IEEE Texas Symposium on Wireless and Microwave Circuits and Systems (WMCS) · doi:10.1109/wmcas.2019.8732531 · Cited by 16
“The optimization process described previously is executed to obtain the maximum PTE for different frequencies ranging from 10 MHz to 50 MHz as most of the power transfer through body considers the optimal frequency to be below hundreds of MHz [16], [17].”
MethodAhsan, F., Chi, T., Cho, R., Sheth, S. A., Goodman, W. K., & Aazhang, B. (2022) · Journal of Neural Engineering · doi:10.1088/1741-2552/ac7894 · Cited by 20
Barbruni, G. L., Cordara, C., Carminati, M., Carrara, S., & Ghezzi, D. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3359481 · Cited by 16
Wireless subcutaneous power supplies for recharging implanted electronic devices
Lyu, S., Chen, Z., Ding, Y., & Wang, Y. (2024) · Device · doi:10.1016/j.device.2024.100423 · Cited by 19
S-MRUT: sectored-multiring ultrasonic transducer for selective powering of brain implants
Hosseini, S., Laursen, K., Rashidi, A., Mondal, T., Corbett, B., & Moradi, F. (2021) · IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control · doi:10.1109/tuffc.2020.3001084 · Cited by 15
“(RF) [9], inductive [10], and, recently, ultrasonic [11], [12] links are among the popular approaches for powering brain implants.”
BackgroundMass fabrication and delivery of 3D multilayer μTags into living cells
Chen, L. Y., Parizi, K. B., Kosuge, H., Milaninia, K. M., McConnell, M. V., Wong, H. ‐. P., & Poon, A. S. Y. (2013) · Scientific Reports · doi:10.1038/srep02295 · Cited by 14
Self-citationWireless power transfer in human tissue via Conformal Strongly Coupled Magnetic Resonance
Hu, H., & Georgakopoulos, S. V. (2015) · IEEE Wireless Power Transfer Conference · doi:10.1109/wpt.2015.7140150 · Cited by 15
“However, inductive coupling has limited range and poor efficiency [6], [7].”
BackgroundAn RF-Powered Crystal-Less Double-Mixing Receiver for Miniaturized Biomedical Implants
Cai, M., Wang, Z., Luo, Y., & Mirabbasi, S. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2868351 · Cited by 15
“On the other hand, the loop or coil antenna structures are usually preferred for implementation inside the human body due to their compact geometry, which is normally in the range of millimeter scale [3], [19]”
BackgroundLamkaddem, A., Yousfi, A. E., Huang, Y., González‐Posadas, V., & Segovia-Vargas, D. (2025) · IEEE Sensors Journal · doi:10.1109/jsen.2025.3587334 · Cited by 17
“The best frequency to use for WPT is in the gigahertz (GHz) range when the size of the transmitting (Tx) antenna is much smaller than the wavelength, as investigated in [36].”
Wireless Power Transfer System Design for Biomedical Implants at 2.45 GHz
Shaw, T., & Mitra, D. (2019) · Asia-Pacific Microwave Conference (APMC) · doi:10.1109/apmc46564.2019.9038770 · Cited by 14
“In contradiction to the near-field and resonant based WPT system generally operated in the MHz frequency region, it was elaborated in [7] that the optimal frequency to obtain better performance from the WPT system, particularly in the case of bio-implantable scenario exists in the GHz and Sub-GHz range.”
BackgroundCho, Y., Basir, A., Lim, Y., & Yoo, H. (2023) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2023.3308628 · Cited by 14
“In previous studies, a mathematical model based on the current density rather than a physical antenna was used for the Tx because it allowed for the exclusion of the effect of the antenna shape [12], [13], [14], [15], [16], [17], [18].”
BackgroundMethodInfluentialBarbruni, G. L., Rodino, F., Ros, P. M., Demarchi, D., Ghezzi, D., & Carrara, S. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3357626 · Cited by 17
Fu, G., & Sonkusale, S. (2015) · Applied Physics Letters · doi:10.1063/1.4921523 · Cited by 14
Self-folding microcube antennas for wireless power transfer in dispersive media
Anacleto, P., Gultepe, E., Gomes, S., Mendes, P. M., & Gracias, D. H. (2016) · TECHNOLOGY · doi:10.1142/s2339547816500047 · Cited by 14
Rahmani, H., & Babakhani, A. (2016) · Proceedings of IEEE Sensors/Proceedings of IEEE Sensors ... · doi:10.1109/icsens.2016.7808979 · Cited by 14
“It has been shown that the optimum frequency for power transmission into biological tissues is at low GHz frequency ranges [2].”
BackgroundDIY Electromagnetic Phantoms for Biomedical Wireless Power Transfer Experiments
van Nunen, T., Huismans, E., Mestrom, R., Bentum, M., & Visser, H. J. (2019) · IEEE Wireless Power Transfer Conference · doi:10.1109/wptc45513.2019.9055704 · Cited by 14
“However, incorporating the displacement current in the analysis leads to an optimal operating frequency that can be in the GHz-range [1, 2].”
BackgroundMethodCho, J., Seong, G., Chang, Y., & Kim, C. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.667447 · Cited by 19
“Furthermore, a higher carrier frequency allows for a shorter wavelength that provides high spatial-resolution and thus improved WPT energy efficiency (Poon et al., 2010).”
BackgroundA microwave powered injectable neural stimulator
Towe, B. C., Larson, P. J., & Gulick, D. W. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6347117 · Cited by 21
“The use of microwave energy has been considered by Poon [6-8] for power transfer in biomedical applications.”
BackgroundPark, S. I. (2013) · Electromagnetic waves · doi:10.2528/pier12110902 · Cited by 14
“However, it has been shown that biological tissues such as skin, muscle, and bone are closer to dielectrics than a conductor at the low giga-hertz frequency ranges as in [10].”
InfluentialWireless power link design using silicon-embedded inductors for brain-machine interface
Wu, R., Raju, S., Chan, M., Sin, J. K. O., & Yue, C. P. (2012) · International Symposium on VLSI Design, Automation and Test · doi:10.1109/vlsi-dat.2012.6212648 · Cited by 14
“While wireless data transmission is well studied and is the foundation of today’s telecommunication industry, wireless power transmission, especially to a brain implant, remains to be a challenging research topic and is being actively pursued.”
BackgroundWireless Power Transfer System Design Using Zero-Index Metamaterial for Implantable Medical Devices
Shaw, T., Mandal, B., Mitra, D., Rangaiah, P. K. B., Perez, M. D., & Augustine, R. (2023) · European Conference on Antennas and Propagation · doi:10.23919/eucap57121.2023.10133695 · Cited by 13
“in [12] that in the case of an implantable WPT system, better performance has been obtained at the frequency of gigahertz and sub-gigahertz range with an electrical dimension of the transmit antenna much smaller than a wavelength and comparable to a wavelength, respectively".”
Background
Metadata
- Authors
- Poon, A. S. Y., O'Driscoll, S., & Meng, T. H.
- Affiliations
- Stanford University (US)
- ORCID
- Poon, Ada S Y
- Venue
- IEEE Transactions on Antennas and Propagation
- Published
- 2010-05
- Volume
- 58
- Issue
- 5
- Pages
- 1739-1750
- ISSN
- 0018-926X (ISSN-L)
- 1558-2221
- OpenAlex
- W2110545851
- Semantic Scholar
- e6a37dc9b793041a01befbefb870dde9882038d3
- Cited by
- 469
- References
- 25
- Citations by year
- 2026: 7 · 2025: 25 · 2024: 32 · 2023: 31 · 2022: 30 · 2021: 42 · 2020: 23 · 2019: 38 · 2018: 27 · 2017: 36 · 2016: 34 · 2015: 32 · 2014: 33 · 2013: 31 · 2012: 29
- FWCI
- 19.92
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Antennas and Propagation (publishedVersion)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Energy Harvesting in Wireless Networks
- Keywords
- wireless power transfer, biomedical implants, optimal operating frequency, air-tissue interface, load impedance matching, RF power transmission
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.