Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality

Open access
Yu, Lifeng, Christner, Jodie A., Leng, Shuai, Wang, Jia, Fletcher, Joel G., McCollough, Cynthia H.·2011·Medical Physics

Abstract

Purpose: To evaluate the image quality of virtual monochromatic images synthesized from dual‐source dual‐energy computed tomography (CT) in comparison with conventional polychromatic single‐energy CT for the same radiation dose. Methods: In dual‐energy CT, besides the material‐specific information, one may also synthesize monochromatic images at different energies, which can be used for routine diagnosis similar to conventional polychromatic single‐energy images. In this work, the authors assessed whether virtual monochromatic images generated from dual‐source CT scanners had an image quality similar to that of polychromatic single‐energy images for the same radiation dose. First, the authors provided a theoretical analysis of the optimal monochromatic energy for either the minimum noise level or the highest iodine contrast to noise ratio (CNR) for a given patient size and dose partitioning between the low‐ and high‐energy scans. Second, the authors performed an experimental study on a dual‐source CT scanner to evaluate the noise and iodine CNR in monochromatic images. A thoracic phantom with three sizes of attenuating rings was used to represent four adult sizes. For each phantom size, three dose partitionings between the low‐energy (80 kV) and the high‐energy (140 kV) scans were used in the dual‐energy scan. Monochromatic images at eight energies (40 to 110 keV) were generated for each scan. Phantoms were also scanned at each of the four polychromatic single energy (80, 100, 120, and 140 kV) with the same radiation dose. Results: The optimal virtual monochromatic energy depends on several factors: phantom size, partitioning of the radiation dose between low‐ and high‐energy scans, and the image quality metrics to be optimized. With the increase of phantom size, the optimal monochromatic energy increased. With the increased percentage of radiation dose on the low energy scan, the optimal monochromatic energy decreased. When maximizing the iodine CNR in monochromatic images, the optimal energy was lower than that when minimizing noise level. When the total radiation dose was equally distributed between low and high energy in dual‐energy scans, for minimum noise, the optimal energies were 68, 71, 74, and 77 keV for small, medium, large, and extra‐large (xlarge) phantoms, respectively; for maximum iodine CNR, the optimal energies were 66, 68, 70, 72 keV. With the optimal monochromatic energy, the noise level was similar to and the CNR was better than that in a single‐energy scan at 120 kV for the same radiation dose. Compared to an 80 kV scan, however, the iodine CNR in monochromatic images was lower for the small, medium, and large phantoms. Conclusions: In dual‐source dual‐energy CT, optimal virtual monochromatic energy depends on patient size, dose partitioning, and the image quality metric optimized. With the optimal monochromatic energy, the noise level was similar to and the iodine CNR was better than that in 120 kV images for the same radiation dose. Compared to single‐energy 80 kV images, the iodine CNR in virtual monochromatic images was lower for small to large phantom sizes.

Overview

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1118/1.3658568

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 16 references

Cited by

See all 343

Metadata

Authors
Yu, L., Christner, J. A., Leng, S., Wang, J., Fletcher, J. G., & McCollough, C. H.
Venue
Medical Physics
Publisher
Wiley
Published
2011-11-09
Volume
38
Issue
12
Pages
6371-6379
ISSN
  • 0094-2405 (ISSN-L)
  • 2473-4209
OpenAlex
W2065526112
Cited by
343
References
16
FWCI
7.11
Type
journal article
Language
en
Open access
bronze

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality. https://app.cochat.ai/papers/10.1118/1.3658568
The paper
Yu, L., Christner, J. A., Leng, S., Wang, J., Fletcher, J. G., & McCollough, C. H. (2011). Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality. Medical Physics. https://doi.org/10.1118/1.3658568

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.