Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality
Open accessAbstract
Purpose: To evaluate the image quality of virtual monochromatic images synthesized from dual‐source dual‐energy computed tomography (CT) in comparison with conventional polychromatic single‐energy CT for the same radiation dose. Methods: In dual‐energy CT, besides the material‐specific information, one may also synthesize monochromatic images at different energies, which can be used for routine diagnosis similar to conventional polychromatic single‐energy images. In this work, the authors assessed whether virtual monochromatic images generated from dual‐source CT scanners had an image quality similar to that of polychromatic single‐energy images for the same radiation dose. First, the authors provided a theoretical analysis of the optimal monochromatic energy for either the minimum noise level or the highest iodine contrast to noise ratio (CNR) for a given patient size and dose partitioning between the low‐ and high‐energy scans. Second, the authors performed an experimental study on a dual‐source CT scanner to evaluate the noise and iodine CNR in monochromatic images. A thoracic phantom with three sizes of attenuating rings was used to represent four adult sizes. For each phantom size, three dose partitionings between the low‐energy (80 kV) and the high‐energy (140 kV) scans were used in the dual‐energy scan. Monochromatic images at eight energies (40 to 110 keV) were generated for each scan. Phantoms were also scanned at each of the four polychromatic single energy (80, 100, 120, and 140 kV) with the same radiation dose. Results: The optimal virtual monochromatic energy depends on several factors: phantom size, partitioning of the radiation dose between low‐ and high‐energy scans, and the image quality metrics to be optimized. With the increase of phantom size, the optimal monochromatic energy increased. With the increased percentage of radiation dose on the low energy scan, the optimal monochromatic energy decreased. When maximizing the iodine CNR in monochromatic images, the optimal energy was lower than that when minimizing noise level. When the total radiation dose was equally distributed between low and high energy in dual‐energy scans, for minimum noise, the optimal energies were 68, 71, 74, and 77 keV for small, medium, large, and extra‐large (xlarge) phantoms, respectively; for maximum iodine CNR, the optimal energies were 66, 68, 70, 72 keV. With the optimal monochromatic energy, the noise level was similar to and the CNR was better than that in a single‐energy scan at 120 kV for the same radiation dose. Compared to an 80 kV scan, however, the iodine CNR in monochromatic images was lower for the small, medium, and large phantoms. Conclusions: In dual‐source dual‐energy CT, optimal virtual monochromatic energy depends on patient size, dose partitioning, and the image quality metric optimized. With the optimal monochromatic energy, the noise level was similar to and the iodine CNR was better than that in 120 kV images for the same radiation dose. Compared to single‐energy 80 kV images, the iodine CNR in virtual monochromatic images was lower for small to large phantom sizes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Energy-selective reconstructions in X-ray computerised tomography
Alvarez, R. E., & Macovski, A. (1976) · Physics in Medicine & Biology · doi:10.1088/0031-9155/21/5/002 · Cited by 2,013
Material differentiation by dual energy CT: initial experience
Johnson, T. R. C., Krauß, B., Sedlmair, M., Grasruck, M., Bruder, H., Morhard, D., Fink, C., Weckbach, S., Lenhard, M., Schmidt, B., et al. (2007) · European Radiology · doi:10.1007/s00330-006-0517-6 · Cited by 1,510
Evaluation of a prototype dual‐energy computed tomographic apparatus. I. Phantom studies
Kalender, W. A., Perman, W. H., Vetter, J. R., & Klotz, E. (1986) · Medical Physics · doi:10.1118/1.595958 · Cited by 451
Dual Energy CT Characterization of Urinary Calculi: Initial In Vitro and Clinical Experience
Graser, A., Johnson, T. R. C., Bader, M., Staehler, M., Haseke, N., Nikolaou, K., Reiser, M. F., Stief, C. G., & Becker, C. R. (2008) · Investigative Radiology · doi:10.1097/rli.0b013e318157a144 · Cited by 341
Siegel, M. J., Schmidt, B., Bradley, D., Suess, C., & Hildebolt, C. (2004) · Radiology · doi:10.1148/radiol.2332032107 · Cited by 294
Dose Reduction in Pediatric CT: A Rational Approach
Boone, J. M., Geraghty, E. M., Seibert, J. A., & Wootton-Gorges, S. L. (2003) · Radiology · doi:10.1148/radiol.2282020471 · Cited by 253
Glazebrook, K. N., Guimarães, L. S., Murthy, N. S., Black, D. F., Bongartz, T., J. Manek, N., Leng, S., Fletcher, J. G., & McCollough, C. H. (2011) · Radiology · doi:10.1148/radiol.11102485 · Cited by 242
Self-citationAutomatic selection of tube potential for radiation dose reduction in CT: A general strategy
Yu, L., Li, H., Fletcher, J. G., & McCollough, C. H. (2010) · Medical Physics · doi:10.1118/1.3264614 · Cited by 230
Self-citationFunama, Y., Awai, K., Nakayama, Y., Kakei, K., Nagasue, N., Shimamura, M., Sato, N., Sultana, S., Morishita, S., & Yamashita, Y. (2005) · Radiology · doi:10.1148/radiol.2373041643 · Cited by 226
Noise considerations in dual energy CT scanning
Kelcz, F., Joseph, P. M., & Hilal, S. K. (1979) · Medical Physics · doi:10.1118/1.594520 · Cited by 221
Cited by
Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications
McCollough, C. H., Leng, S., Yu, L., & Fletcher, J. G. (2015) · Radiology · doi:10.1148/radiol.2015142631 · Cited by 1,598
Self-citationPhoton-counting CT: Technical Principles and Clinical Prospects
Willemink, M. J., Persson, M., Pourmorteza, A., Pelc, N. J., & Fleischmann, D. (2018) · Radiology · doi:10.1148/radiol.2018172656 · Cited by 1,326
Dual-Energy CT–Based Monochromatic Imaging
Yu, L., Leng, S., & McCollough, C. H. (2012) · American Journal of Roentgenology · doi:10.2214/ajr.12.9121 · Cited by 609
Self-citationState of the Art: Dual-Energy CT of the Abdomen
Marin, D., Boll, D. T., Mileto, A., & Nelson, R. C. (2014) · Radiology · doi:10.1148/radiol.14131480 · Cited by 368
Grant, K. L., Flohr, T. G., Krauss, B., Sedlmair, M., Thomas, C., & Schmidt, B. (2014) · Investigative Radiology · doi:10.1097/rli.0000000000000060 · Cited by 307
Metadata
- Authors
- Yu, L., Christner, J. A., Leng, S., Wang, J., Fletcher, J. G., & McCollough, C. H.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2011-11-09
- Volume
- 38
- Issue
- 12
- Pages
- 6371-6379
- ISSN
- 0094-2405 (ISSN-L)
- 2473-4209
- OpenAlex
- W2065526112
- Cited by
- 343
- References
- 16
- FWCI
- 7.11
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality. https://app.cochat.ai/papers/10.1118/1.3658568
- The paper
- Yu, L., Christner, J. A., Leng, S., Wang, J., Fletcher, J. G., & McCollough, C. H. (2011). Virtual monochromatic imaging in dual‐source dual‐energy CT: Radiation dose and image quality. Medical Physics. https://doi.org/10.1118/1.3658568
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.