The UVA/PADOVA Type 1 Diabetes Simulator: New Features
Open accessThe DOI matches its Crossref record; 25 of the 25 references it cites were found in a registry or an index; 10 of the 22 sources its claims cite were checked for support.
Abstract
Recent studies have provided new insights into nonlinearities of insulin action in the hypoglycemic range and into glucagon kinetics as it relates to response to hypoglycemia. Based on these data, we developed a new version of the UVA/PADOVA Type 1 Diabetes Simulator, which was submitted to FDA in 2013 (S2013). The model of glucose kinetics in hypoglycemia has been improved, implementing the notion that insulin-dependent utilization increases nonlinearly when glucose decreases below a certain threshold. In addition, glucagon kinetics and secretion and action models have been incorporated into the simulator: glucagon kinetics is a single compartment; glucagon secretion is controlled by plasma insulin, plasma glucose below a certain threshold, and glucose rate of change; and plasma glucagon stimulates with some delay endogenous glucose production. A refined statistical strategy for virtual patient generation has been adopted as well. Finally, new rules for determining insulin to carbs ratio (CR) and correction factor (CF) of the virtual patients have been implemented to better comply with clinical definitions. S2013 shows a better performance in describing hypoglycemic events. In addition, the new virtual subjects span well the real type 1 diabetes mellitus population as demonstrated by good agreement between real and simulated distribution of patient-specific parameters, such as CR and CF. S2013 provides a more reliable framework for in silico trials, for testing glucose sensors and insulin augmented pump prediction methods, and for closed-loop single/dual hormone controller design, testing, and validation.
Overview
CoChat Summary
TL;DR
The paper presents S2013, a new version of the UVA/PADOVA Type 1 Diabetes Simulator, adding hypoglycemia-aware glucose utilization, glucagon secretion/action, updated joint parameter distributions, and stricter virtual-subject generation. The authors argue that S2013 better describes hypoglycemic events and better matches real CR and CF distributions, while remaining FDA-submitted for preclinical in silico use.
Key findings
- The original S2008 simulator included a population of 300 in silico subjects: 100 adults, 100 adolescents, and 100 children; the paper says S2008 had been used by 32 research groups and reported in 63 peer-reviewed publications.
- S2013 adds a hypoglycemia model with a threshold set at 60 mg/dl and incorporates glucagon kinetics/secretion/action, including a 1-compartment glucagon kinetics model and a 2-compartment subcutaneous glucagon transit model.
- The new simulator was assessed on 32 T1DM subjects in an unpublished clamp dataset and on 13 data sets from 11 adults with T1DM in a bihormonal closed-loop clinical trial; the paper states the model fits glucose data well before the threshold and underestimates data afterward as expected.
- In simulation results for 100 adults, 100 adolescents, and 100 children, the paper reports that the distributions of CR and CF “well reproduce” real patients, and that children had higher CFs than adolescents and adults; the table reports mean weight, insulin needs, CHO ratio, and fasting plasma glucose for each group.
Methods
- The paper is a simulator-development paper comparing S2013 to S2008, describing changes to the glucose-insulin model, counterregulation, parameter distributions, and virtual patient generation.
- It uses simulation and model identification/assessment on two empirical datasets: 32 T1DM subjects undergoing hyperinsulinemic euglycemic and hypoglycemic clamps, and 11 adults with T1DM (13 data sets) from a 27-hour bihormonal closed-loop study with 3 carbohydrate-rich meals.
- Main analyses include fitting the hypoglycemia model, simulating 50 g CHO meal tests for CR estimation, using the 1700 rule for CF, and filtering virtual subjects by CR thresholds, steady-state glucose > 300 mg/dl, and Mahalanobis distance below the 95% percentile.
Limitations
- The authors explicitly state that the subcutaneous glucagon model “suffers some important limitations,” because it was hard to isolate insulin’s effect on glucagon response since insulin is constantly changing, and glucagon is given repeatedly in the study.
- They note that both S2008 and S2013 were validated and accepted by FDA for a single meal scenario only; multiple-meal scenarios can be simulated, but results would not be realistic because the simulator does not include time-varying parameters.
- Evident from the text: the paper repeatedly presents validation and comparison results, but it does not report formal performance metrics such as error estimates, p-values, or confidence intervals for the simulator comparisons.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
6 Valid · 2 Not valid · 5 Not checked · 24 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The development of artificial pancreas control algorithms in the past few years has been accelerated by computer simulation; for example, in silico testing has provided directions for clinical studies and ruling out of ineffective control scenarios in a cost-effective manner. 1,2”
What the sources say
- Supportedcites [1] matched to “Diabetes: Models, Signals, and Control” (2009) source
Section IV.D says in silico testing supported authorization of a clinical trial, helped rule out ineffective control scenarios, and produced preclinical results in a fraction of the time and cost of animal trials. This supports the claim that simulation accelerated artificial-pancreas algorithm development, while the paper cautions that simulation does not replace clinical trials. (LLM, conf=0.93)
- Partly supportedcites [2] matched to “Artificial Pancreas: Past, Present, Future” (2011) sourceExperimental — accuracy not yet measured
In “A step forward: in silico experiments replacing animal trials” (page unavailable), the paper says simulation is already used to test control algorithms efficiently and at a fraction of the time and cost of animal trials. But it says simulation *will* greatly accelerate future artificial-pancreas development; it does not establish that algorithm development had already accelerated in the past few years. (LLM, conf=0.85)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“This model was assessed on a clinical data set kindly provided by Dr Damiano's group (Boston University), which was collected in a bihormonal closed-loop clinical trial. 15”
What the sources say
- Could not verifycites [15] matched to “Duration of type I diabetes affects glucagon and glucose responses to insulin-induced hypoglycemia.” (1984)Experimental — accuracy not yet measured
The Patients and Methods passages (page not provided) describe insulin-tolerance tests and measurements of glucose and pancreatic glucagon, not an assessment of a subcutaneous glucagon transport model. Searches for the model, Dr. Damiano’s group, and a bihormonal closed-loop trial found no passage addressing the attributed claim. (LLM, conf=0.76)
plainn = 11- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“This phenomenon has been described during hyperinsulemic clamps in T1DM. 9,10”
What the sources say
- Partly supportedcites [9] matched to “Modeling Insulin‐Glucose Dynamics during Insulin Induced Hypoglycemia. Evaluation of Glucose Counterregulation” (1999) sourceExperimental — accuracy not yet measured
The Discussion (page unavailable) reports that, during hyperinsulinemic clamps in patients with type 1 diabetes, a lower glucose nadir was associated with faster model-estimated insulin utilization. It does not describe insulin action increasing within a patient as glucose crosses a particular threshold; the reported association is across subjects. (LLM, conf=0.85)
- Not checked — no full text of the sourcecites [10] matched to “Dynamic network model of glucose counterregulation in subjects with insulin-requiring diabetes” (2000) source
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“The choice to include glucagon secretion was based on our recent data in T1DM 23 showing a nonnegligible glucagon secretion.”
What the sources say
- Partly supportedcites [23] matched to “Diurnal Pattern of Insulin Action in Type 1 Diabetes” (2013) sourceExperimental — accuracy not yet measured
The Results section (page unavailable) reports measurable plasma glucagon concentrations in people with type 1 diabetes and a postprandial increase that differed by meal. That supports including glucagon in a model, but the paper reports concentrations rather than directly measuring secretion or establishing its magnitude as “nonnegligible.” (LLM, conf=0.82)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“S2013 shows a better performance in describing hypoglycemic events.”
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
13 more claims
Create a free account to see all 37 claims and every source verdict.
- It may state two separate claims as one. 17
- Probably not a substantial claim. 5
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“On the contrary, the Hovorka model and the UVAPadova model were originally formulated to represent the physiological behaviour of T1DM individuals, and no modification is necessary for the purpose of the present work.”
From A comparison among three maximal mathematical models of the glucose-insulin system (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This model is indeed included in the EGP module of the UVa/Padova T1D simulator [32–34].”
From Modeling the effect of glucagon on endogenous glucose production in healthy individuals under meal-like conditions (2025) · cites this paper as [33]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“While the Hovorka model and the UVAPadova Simulator only describe the glucose metabolism of a subject with type 1 diabetes, the Sorensen model was formulated to simulate the behaviour of both normal and diabetic individuals.”
From A comparison among three maximal mathematical models of the glucose-insulin system (2021) · cites this paper as [23]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“The training set, which spanned a wide range of pre-meal conditions and patient characteristics, was created in silico by the UVa/Padova T1D Simulator [38], a simulation tool that includes a model of glucose-insulin-glucagon dynamics in subjects with T1D, accepted by the FDA to substitute preclinical trials of certain insulin treatments.”
From Advanced Diabetes Management Using Artificial Intelligence and Continuous Glucose Monitoring Sensors (2020) · cites this paper as [38]
The cited source says what the claim attributes to it.
In the citing paper
“Another example is compartmental glucose-insulin models, which have been used to generate virtual cohorts to evaluate automated insulin pumps [38,39]”
From Credibility assessment of in silico clinical trials for medical devices (2024) · cites this paper as [38]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ T1DM Simulator 2013 (S2013)
“which was collected in a bihormonal closed-loop clinical trial. 15”
§ References
“15. Lorenzi M, Bohannon N, Tsalikian E, et al. Duration of type 1 diabetes affects glucagon and glucose responses to insulininduced hypoglycemia. West J Med . 1984;141:467-471.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ T1DM Simulator 2013 (S2013)
“only subjects who met the following criteria have been included in the new in silico population: (1) CR ≤ 30 g/U for adult and adolescents and CR ≤ 40 g/U for children”
§ Simulation Results
“| CHO ratio (g/U) | 15.9 (5.3) | 7.2 | 29.5 | 16.0 (5.9) | 5.3 | 32.2 | 23.6 (6.6) | 8.0 | 33.7 |”
Citations
Cites
Quantitative estimation of insulin sensitivity.
Bergman, R. N., Ider, Y. Z., Bowden, C. R., & Cobelli, C. (1979) · American Journal of Physiology-Endocrinology and Metabolism · doi:10.1152/ajpendo.1979.236.6.e667 · Cited by 1,874 · cited 1×
“where G(t) is plasma glucose concentration, I(t) is plasma insulin concentration, GIR(t) is exogenous glucose infusion rate, SG, SI, and p 2 are model parameters, and risk is defined as in (11-12).”
BackgroundSelf-citationIn Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes
Kovatchev, B. P., Breton, M., Dalla Man, C., & Cobelli, C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300106 · Cited by 760
Self-citationArtificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Self-citationDiabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
Self-citationA Bihormonal Closed-Loop Artificial Pancreas for Type 1 Diabetes
El-Khatib, F. H., Russell, S. J., Nathan, D. M., Sutherlin, R. G., & Damiano, E. R. (2010) · Science Translational Medicine · doi:10.1126/scitranslmed.3000619 · Cited by 403
A Comparison of Childhood and Adult Type I Diabetes Mellitus
Karjalainen, J., Salmela, P., Ilonen, J., Surcel, H.-M., & Knip, M. (1989) · New England Journal of Medicine · doi:10.1056/nejm198904063201401 · Cited by 279
Age of Onset and Type of Diabetes
Laakso, M., & Pyörälä, K. (1985) · Diabetes Care · doi:10.2337/diacare.8.2.114 · Cited by 245
Evaluating the Efficacy of Closed-Loop Glucose Regulation via Control-Variability Grid Analysis
Magni, L., Raimondo, D. M., Man, C. D., Breton, M., Patek, S., De Nicolao, G., Cobelli, C., & Kovatchev, B. P. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200414 · Cited by 207
Self-citationIncidence of Type 1 and Type 2 Diabetes in Adults Aged 30–49 Years
Bruno, G., Runzo, C., Cavallo-Perin, P., Merletti, F., Rivetti, M., Pinach, S., Novelli, G., Trovati, M., Cerutti, F., Pagano, G., et al. (2005) · Diabetes Care · doi:10.2337/diacare.28.11.2613 · Cited by 187
Closing the Loop: The Adicol Experience
Hovorka, R., Chassin, L. J., Wilinska, M. E., Canonico, V., Akwi, J. A., Federici, M. O., Massi-Benedetti, M., Hutzli, I., Zaugg, C., Kaufmann, H., et al. (2004) · Diabetes Technology & Therapeutics · doi:10.1089/152091504774197990 · Cited by 154
Research works printed but not linked: 4
- Bergman RN, 1979, Am J Physiol, 236, E667
- Lorenzi M, 1984, West J Med, 141, 467
- Walsh J, Pumping Insulin. Everything You Need for Success with an Insulin Pump, 2000, 105
- Davidson PC, 2003, Diabetes Technol Ther, 5, 237
Cited by
Items 101–200 of 786
Liu, C., Vehí, J., Avari, P., Reddy, M., Oliver, N. S., Georgiou, P., & Herrero, P. (2019) · Sensors · doi:10.3390/s19194338 · Cited by 48
“To build the employed mixed-meal model library within the UVa-Padova T1DM simulator [36], the scientific literature was reviewed for clinical trials studying the effect of meal composition on non-diabetic subjects, which included mean population plasma glucose and plasma insulin concentration data, meal composition, a…”
MethodAnnuzzi, G., Apicella, A., Arpaïa, P., Bozzetto, L., Criscuolo, S., Benedetto, E. D., Pesola, M., Prevete, R., & Vallefuoco, E. (2023) · IEEE Access · doi:10.1109/access.2023.3244712 · Cited by 55
“…[30], [31], [32], [33], [34], [35], [36] have employed ANNs in the prediction of postprandial blood glucose values, using both data from real patients with T1D and data computer-simulated patients (the latter named virtual patients ) [37], such as those obtained with UVA/Padova simulator [38].”
MethodToffanin, C., Aiello, E. M., Favero, S. D., Cobelli, C., & Magni, L. (2019) · Journal of Process Control · doi:10.1016/j.jprocont.2019.03.007 · Cited by 43
Self-citationGlucose-insulin model identified in free-living conditions for hypoglycaemia prevention
Toffanin, C., Favero, S. D., Aiello, E. M., Messori, M., Cobelli, C., & Magni, L. (2018) · Journal of Process Control · doi:10.1016/j.jprocont.2018.02.003 · Cited by 39
Self-citationVISENTIN, R., Cobelli, C., & Man, C. D. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2020.0110 · Cited by 45
Self-citationSchiavon, M., Man, C. D., Kudva, Y. C., Basu, A., & Cobelli, C. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700606 · Cited by 52
Self-citationReinforcement Learning Models and Algorithms for Diabetes Management
Yau, K. A., Chong, Y.-W., Fan, X., Wu, C., Saleem, Y., & Lim, P.-C. (2023) · IEEE Access · doi:10.1109/access.2023.3259425 · Cited by 51
“Although these factors affect glucose variability and glucose-insulin dynamics, they have not been incorporated in FDA-approved simulators, such as UVA/ Padova [66].”
BackgroundMethodA meta-learning approach to personalized blood glucose prediction in type 1 diabetes
Langarica, S., Rodriguez-Fernández, M., Núñez, F., & Doyle, F. J. (2023) · Control Engineering Practice · doi:10.1016/j.conengprac.2023.105498 · Cited by 46
Mughal, I. S., Patané, L., & Caponetto, R. (2024) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2024.100937 · Cited by 46
Real-Time Model-Based Fault Detection of Continuous Glucose Sensor Measurements
Turksoy, K., Roy, A., & Çınar, A. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2535412 · Cited by 38
Visentin, R., Giegerich, C., Jäger, R., Dahmen, R., Boss, A., Grant, M., Dalla Man, C., Cobelli, C., & Klabunde, T. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0128 · Cited by 36
“Mahalanobis distance lower than that corresponding to the 95% percentile was considered, paralleling what was done in Dalla Man et al.7 Each individual parameter set describing the TI/Gen2 PK was randomly assigned to each insilico subject.”
BackgroundMethodInfluentialSelf-citationAutomatic adaptation of basal therapy for Type 1 diabetic patients: A Run-to-Run approach
Toffanin, C., Messori, M., Cobelli, C., & Magni, L. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.09.002 · Cited by 37
“Due to the switching structure of the updating law (1), the classic stability analysis of R2R algorithms (see Francois et al. [2003], Owens et al.”
MethodInfluentialSelf-citationSanz, R., García, P., Díez, J.-L., & Bondía, J. (2020) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2020.2975147 · Cited by 42
“It should be remarked that Gb is a control parameter defining the operating conditions in the absence of disturbances, which was set to Gb = 100 mg/dL, and the corresponding basal insulin infusions for each patient ub j are obtained from the equilibrium equations of the simulator model [33].”
BackgroundMethodInfluentialParcerisas, A., Contreras, I., Delecourt, A., Bertachi, A. H., Beneyto, A., Conget, I., Viñals, C., Giménez, M., & Vehí, J. (2022) · Sensors · doi:10.3390/s22041665 · Cited by 43
Meneghetti, L., Facchinetti, A., & Favero, S. D. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3004270 · Cited by 38
“The simulated data were obtained using the Padova/UVA Type 1 Diabetes simulator [49], an accurate physiology simulator that was approved by Food and Drug Administration as a substitute of animal testing prior to AP clinical trials on humans.”
MethodHughes, J., Gautier, T., Colmegna, P., Fabris, C., & Breton, M. D. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820973193 · Cited by 34
“1330 Journal of Diabetes Science and Technology 15(6)”
BackgroundMethodSelf-citationCao, Z., Gondhalekar, R., Dassau, E., & Doyle, F. J. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2783238 · Cited by 33
Artificial Pancreas: Clinical Study in Latin America Without Premeal Insulin Boluses
Sánchez‐Peña, R., Colmegna, P. H., Garelli, F., Battista, H. D., García-Violini, D., Moscoso-Vásquez, M., Rosales, N., Fushimi, E., Campos-Náñez, E., Breton, M. D., et al. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818786488 · Cited by 42
Self-citationFrom bed to bench: How in silico medicine can help ageing research
Viceconti, M., & Dall’Ara, E. (2018) · Mechanisms of Ageing and Development · doi:10.1016/j.mad.2018.07.001 · Cited by 37
Lim, M. H., Lee, W. H., Jeon, B., & Kim, S. (2021) · IEEE Access · doi:10.1109/access.2021.3100007 · Cited by 47
“The UVA/Padova type 1 diabetes simulator [29], [30] is such an”
BackgroundPostprandial Glucose Regulation via KNN Meal Classification in Type 1 Diabetes
Aiello, E. M., Toffanin, C., Messori, M., Cobelli, C., & Magni, L. (2018) · IEEE Control Systems Letters · doi:10.1109/lcsys.2018.2844179 · Cited by 35
“The new approach is tested in silico by using the 2013 version of the UVA/PADOVA Simulator [7].”
MethodSelf-citationPhenomenological-Based model of human stomach and its role in glucose metabolism
Lema-Pérez, L., Garcia‐Tirado, J., Builes-Montaño, C. E., & Álvarez, H. (2018) · Journal of Theoretical Biology · doi:10.1016/j.jtbi.2018.10.024 · Cited by 33
Huang, W. H., Pang, I., Bai, J., Cui, B., Qi, X., & Zhang, S. (2025) · Advanced Intelligent Systems · doi:10.1002/aisy.202400822 · Cited by 43
Model Identification using Continuous Glucose Monitoring Data for Type 1 Diabetes
Boiroux, D., Hagdrup, M., Mahmoudi, Z., Poulsen, N. K., Madsen, H., & Jørgensen, J. B. (2016) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2016.07.279 · Cited by 31
“They usually consist either of a system of ordinary differential equations (ODEs) (e.g. the models developed by Hovorka et al. (2004); Kanderian et al. (2009); Dalla Man et al. (2014)), or a system of stochastic differential equations (SDEs) (Duun-Henriksen et al. (2013)).”
BackgroundOviedo, S., Contreras, I., Bertachi, A. H., Quirós, C., Giménez, M., Conget, I., & Vehí, J. (2019) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2019.06.025 · Cited by 32
An augmented subcutaneous type 1 diabetic patient modelling and design of adaptive glucose control
Nath, A., Deb, D., & Dey, R. (2020) · Journal of Process Control · doi:10.1016/j.jprocont.2019.08.010 · Cited by 32
Noguer, J., Contreras, I., Mujahid, O., Beneyto, A., & Vehí, J. (2022) · Sensors · doi:10.3390/s22134944 · Cited by 42
DeJournett, L., & DeJournett, J. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816653967 · Cited by 32
Offset-free MPC strategy for nonzero regulation of linear impulsive systems
Villa-Tamayo, M. F., Caicedo, M. A., & Rivadeneira, P. S. (2020) · ISA Transactions · doi:10.1016/j.isatra.2020.01.005 · Cited by 30
Modeling of Diabetes and Its Clinical Impact
Fritzen, K., Heinemann, L., & Schnell, O. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818785642 · Cited by 34
Meneghetti, L., Susto, G. A., & Favero, S. D. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819881452 · Cited by 29
Auto-Regressive Time Delayed jump neural network for blood glucose levels forecasting
D’Antoni, F., Merone, M., Piemonte, V., Iannello, G., & Soda, P. (2020) · Knowledge-Based Systems · doi:10.1016/j.knosys.2020.106134 · Cited by 32
Wang, W., Wang, S., Wang, X., Liu, D., Geng, Y., & Wu, T. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3015199 · Cited by 33
“In [30] and [31], a two-compartment model of glucose kinetics was developed by Cobelli for in silico simulations.”
MethodSliding-mode disturbance observers for an artificial pancreas without meal announcement
Sala-Mira, I., Díez, J.-L., Ricarte, B., & Bondía, J. (2019) · Journal of Process Control · doi:10.1016/j.jprocont.2019.03.008 · Cited by 34
Adaptive Impulsive Offset-Free MPC to Handle Parameter Variations for Type 1 Diabetes Treatment
Villa-Tamayo, M. F., & Rivadeneira, P. S. (2020) · Industrial & Engineering Chemistry Research · doi:10.1021/acs.iecr.9b05979 · Cited by 29
Towards a Verified Artificial Pancreas: Challenges and Solutions for Runtime Verification
Cameron, F., Fainekos, G., Maahs, D. M., & Sankaranarayanan, S. (2015) · Lecture notes in computer science · doi:10.1007/978-3-319-23820-3_1 · Cited by 39
“of the FDA approved T1DM simulator that can now be used as an alternative to animal testing [46].”
MethodColmegna, P. H., Sánchez‐Peña, R., Gondhalekar, R., Dassau, E., & Doyle, F. J. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816638857 · Cited by 34
“Recently, models of PA have been developed in Schiavon,1 Breton,13 Dalla Man et al,14 Jacobs et al,15 Ben Brahim et al,16 Schiavon et al,17 and references therein.”
BackgroundMethodInfluentialSanz, A. J. L., Doyle, F. J., & Dassau, E. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816680632 · Cited by 32
Embedded Model Predictive Control for a Wearable Artificial Pancreas
Chakrabarty, A., Healey, E., Shi, D., Zavitsanou, S., Doyle, F. J., & Dassau, E. (2019) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2019.2939122 · Cited by 31
“tor [44].”
A predictive computational platform for optimizing the design of bioartificial pancreas devices
Ernst, A. U., Wang, L., Worland, S. C., Marfil‐Garza, B. A., Wang, X., Liu, W. H., Chiu, A., Kin, T., O’Gorman, D., Steinschneider, S., et al. (2022) · Nature Communications · doi:10.1038/s41467-022-33760-5 · Cited by 34
“Like-wise, recently, an FDA-approved simulator has demonstrated the ability tomodelglycemiclevelsinsimulatedpatientsfollowingselectedroutes of insulin administration (NCT03093636) 63 .”
BackgroundCappon, G., Prendin, F., Facchinetti, A., Sparacino, G., & Favero, S. D. (2023) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2023.3276193 · Cited by 41
“…concentration; α (min) is the delay between the plasmatic and interstitial glucose concentration compartments; and ρ ( G ) is a function, introduced by [52], that allows to better describe Authorized licensed use limited to the terms of the applicable license agreement with IEEE.”
BackgroundDigital Twins in Type 1 Diabetes: A Systematic Review
Cappon, G., & Facchinetti, A. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241262112 · Cited by 44
Astillo, P. V., Choudhary, G., Duguma, D. G., Kim, J., & You, I. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3063173 · Cited by 30
Improved postprandial glucose control with a customized Model Predictive Controller
Messori, M., Ellis, M. J., Cobelli, C., Christofides, P. D., & Magni, L. (2015) · Proceedings of the ... American Control Conference/Proceedings of the American Control Conference · doi:10.1109/acc.2015.7172136 · Cited by 29
“The results presented in this section are based on an in– silico study of 100 adult virtual patients of the UVA/Padova simulator [12].”
BackgroundMethodInfluentialSelf-citationHidalgo, J. I., Botella, M., Velasco, J. M., Garnica, Ó., Cervigón, C., Martínez, R. R., Aramendi, A., Maqueda, E., & Lanchares, J. (2019) · Applied Soft Computing · doi:10.1016/j.asoc.2019.105923 · Cited by 26
Digital Twins and Automation of Care in the Intensive Care Unit
Chase, J. G., Zhou, C., Knopp, J. L., Möeller, K., Benyó, B. I., Desaive, T., Wong, J. H. K., Malinen, S. K., Näswall, K., Shaw, G. M., et al. (2023) · Wiley eBooks · doi:10.1002/9781119857433.ch17 · Cited by 35
A comparison among three maximal mathematical models of the glucose-insulin system
Pompa, M., Panunzi, S., Borri, A., & De Gaetano, A. (2021) · PLOS ONE · doi:10.1371/journal.pone.0257789 · Cited by 24
“The equation from UVAPadova S2017 [24] is replaced by that in UVAPadova (2014) [23], because reported with an error.”
BackgroundMethodInfluentialZhu, T., Li, K., & Georgiou, P. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3303367 · Cited by 50
“Particularly, we choose FQE as the OPE method with the implementation by Paine et al. [47], considering it provided accurate and robust estimation in healthcare settings [49].”
InfluentialRosales, N., Battista, H. D., & Garelli, F. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.103106 · Cited by 30
Statistical Verification of Cyber-Physical Systems using Surrogate Models and Conformal Inference
Qin, X., Xia, Y., Zutshi, A., Fan, C., & Deshmukh, J. V. (2022) · ACM/IEEE International Conference on Cyber-Physical Systems (ICCPS) · doi:10.1109/iccps54341.2022.00017 · Cited by 45
Krishnamoorthy, D., & Doyle, F. J. (2022) · IEEE Control Systems Letters · doi:10.1109/lcsys.2022.3179330 · Cited by 27
HyMn: Mining linear hybrid automata from input output traces of cyber-physical systems
Lamrani, I., Banerjee, A., & Gupta, S. K. S. (2018) · IEEE International Conference on Industrial Cyber-Physical Systems (ICPS) · doi:10.1109/icphys.2018.8387670 · Cited by 35
“Benefits of using HyMn inferred patient specific parameters: We executed the AP system with two parameter configurations: a) taking statistical average parameters obtained from a large pool of T1D subjects as discussed in [13], and b) obtaining the patient specific parameters for a given subject using the HyMN approac…”
MethodFalco, I. D., Cioppa, A. D., Giugliano, A., Marcelli, A., Koutný, T., Krčma, M., Scafuri, U., & Tarantino, E. (2019) · Applied Soft Computing · doi:10.1016/j.asoc.2019.01.020 · Cited by 24
Ramkissoon, C. M., Bertachi, A. H., Beneyto, A., Bondía, J., & Vehí, J. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2898558 · Cited by 29
“The UVA/Padova Simulator [14] coupled with an exercise model [15] was used to build an 15-day scenario with 10 adult subjects.”
MethodA case study on using a large language model to analyze continuous glucose monitoring data
Healey, E., Tan, A. L. M., Flint, K. L., Ruiz, J., & Kohane, I. S. (2025) · Scientific Reports · doi:10.1038/s41598-024-84003-0 · Cited by 36
“This patient simulator has been FDA-accepted for use in pre-clinical testing of insulin dosing algorithms and it includes a type 1 diabetes (T1D) adult patient model and an interstitial glucose sensing model to simulate CGM data 18,19 .”
Effect of BGM Accuracy on the Clinical Performance of CGM: An In-Silico Study
Campos-Náñez, E., & Breton, M. D. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817710476 · Cited by 26
Self-citationModeling the Error of the Medtronic Paradigm Veo Enlite Glucose Sensor
Biagi, L., Ramkissoon, C. M., Facchinetti, A., Leal, Y., & Vehí, J. (2017) · Sensors · doi:10.3390/s17061361 · Cited by 27
“, the University of Virgina (UVA)/Padova T1D simulator [12].”
BackgroundMapping data to virtual patients in type 1 diabetes
Colmegna, P. H., Wang, K., Garcia‐Tirado, J., & Breton, M. D. (2020) · Control Engineering Practice · doi:10.1016/j.conengprac.2020.104605 · Cited by 26
Self-citationHybrid Knowledge and Data Driven Synthesis of Runtime Monitors for Cyber-Physical Systems
Zhou, X., Ahmed, B., Aylor, J. H., Asare, P., & Alemzadeh, H. (2023) · IEEE Transactions on Dependable and Secure Computing · doi:10.1109/tdsc.2023.3242653 · Cited by 35
“For APS, we integrated two widely-used APS controllers (OpenAPS [60] and Basal-Bolus [61]) with two different patient glucose simulators, including Glucosym [62] and UVA-Padova Type 1 Diabetes Simulator [63].”
BackgroundMethodReal-time detection of Glucose Sensor and Insulin Pump Faults in an Artificial Pancreas.
Favero, S. D., Monaro, M., Facchinetti, A., Tagliavini, A., Sparacino, G., & Cobelli, C. (2014) · IFAC Proceedings Volumes · doi:10.3182/20140824-6-za-1003.02733 · Cited by 28
Self-citation(2018) · doi:10.1177/1932296817735341 · Cited by 23
Schiavon, M., Man, C. D., & Cobelli, C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0248 · Cited by 23
Self-citationClinical impact of blood glucose monitoring accuracy: an in-silico study
Campos-Náñez, E., Fortwaengler, K., & Breton, M. D. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817710474 · Cited by 22
Self-citationCamerlingo, N., Vettoretti, M., Del Favero, S., Cappon, G., Sparacino, G., & Facchinetti, A. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0139 · Cited by 22
Artificial Pancreas: Evaluating the ARG Algorithm Without Meal Announcement
Fushimi, E., Colmegna, P. H., Battista, H. D., Garelli, F., & Sánchez‐Peña, R. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819864585 · Cited by 27
Goodwin, G. C., Seron, M. M., Medioli, A. M., Smith, T. A., King, B. R., & Smart, C. E. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.101813 · Cited by 24
Franco, R., de Loza, A. F., Ríos, H., Cassany, L., Gucik‐Derigny, D., Cieslak, J., Henry, D., & Olçomendy, L. (2021) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2020.3046420 · Cited by 24
“ference approximations (FDAs) for preclinical validation of insulin therapy [19].”
BackgroundTowards Safe and Robust Closed-Loop Artificial Pancreas Using Improved PID-Based Control Strategies
Alshalalfah, A.-L., Hamad, G. B., & Mohamed, O. A. (2021) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2021.3058355 · Cited by 32
Noaro, G., Zhu, T., Cappon, G., Facchinetti, A., & Georgiou, P. (2023) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2023.3249571 · Cited by 27
“The proposed algorithm is trained and tested by means of a well-established computer simulation software: the UVA/Padova T1D Simulator (T1DS).”
BackgroundMethodInfluentialPersonalized LSTM-based alarm systems for hypoglycemia and hyperglycemia prevention
Iacono, F. D., Magni, L., & Toffanin, C. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105167 · Cited by 25
Hettiarachchi, C., Malagutti, N., Nolan, C. J., Suominen, H., & Daskalaki, E. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.105839 · Cited by 36
Optimal regulation of blood glucose level in Type I diabetes using insulin and glucagon
Shirin, A., Rossa, F. D., Klickstein, I. S., Russell, J. J., & Sorrentino, F. (2019) · PLoS ONE · doi:10.1371/journal.pone.0213665 · Cited by 27
“We consider the model in [3, 4] which is a system of nonlinear ordinary differential equations (ODEs).”
BackgroundMethodCappon, G., Facchinetti, A., Sparacino, G., Georgiou, P., & Herrero, P. (2019) · Sensors · doi:10.3390/s19143168 · Cited by 29
“It has been accepted by the FDA as a substitute to pre-clinical trials for closed-loop insulin delivery strategies and open-loop therapies [17].”
BackgroundMethodInfluentialClinically Accurate Prediction of Glucose Levels in Patients with Type 1 Diabetes
Amar, Y., Shilo, S., Oron, T., Amar, E., Phillip, M., & Segal, E. (2020) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0435 · Cited by 27
“…models published to date mostly use linear predictors of glucose levels, 12 including model predictive control, proportional–integral– derivative control, and fuzzy logic control, which rely mostly on medical logic, 13 in recent years, an increasing number of data-driven models have been published.”
Yet Another Glucose Variability Index: Time for a Paradigm Change?
Cobelli, C., & Facchinetti, A. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0397 · Cited by 22
Self-citationPanunzi, S., Pompa, M., Borri, A., Piemonte, V., & De Gaetano, A. (2020) · PLoS ONE · doi:10.1371/journal.pone.0237215 · Cited by 21
Zone-MPC Automated Insulin Delivery Algorithm Tuned for Pregnancy Complicated by Type 1 Diabetes
Özaslan, B., Deshpande, S., Doyle, F. J., & Dassau, E. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2021.768639 · Cited by 24
Yousefi, M., van Heusden, K., West, N., Mitchell, I. M., Ansermino, J. M., & Dumont, G. A. (2018) · Control Engineering Practice · doi:10.1016/j.conengprac.2018.11.009 · Cited by 26
Kriventsov, S., Lindsey, A., & Hayeri, A. (2020) · JMIR Diabetes · doi:10.2196/18660 · Cited by 23
Nyman, E., Wesseling-Rozendaal, Y. J. W., Helmlinger, G., Hamrén, B., Kjellsson, M. C., Strålfors, P., van Riel, N. A. W., Gennemark, P., & Cedersund, G. (2016) · Interface Focus · doi:10.1098/rsfs.2015.0075 · Cited by 24
“The same model, but with insulin secretion replaced with external insulin inputs, has also been applied to data from type 1 diabetic patients, and approved by the Food and Drug Administration to be used instead of test animals when testing closed-loops algorithms for insulin administration [4,5].”
BackgroundMethodHerrgårdh, T., Li, H., Nyman, E., & Cedersund, G. (2021) · Frontiers in Physiology · doi:10.3389/fphys.2021.619254 · Cited by 23
Modeling Glucagon Action in Patients With Type 1 Diabetes
Emami, A., Youssef, J. E., Rabasa-Lhoret, R., Pineau, J., Castle, J. R., & Haidar, A. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2593630 · Cited by 20
“Clinical trials are an integral part of the development process of closed-loop systems but are time consuming, resource demanding, and costly [18], [19].”
BackgroundParameter-Invariant Monitor Design for Cyber–Physical Systems
Weimer, J. E., Ivanov, R., Chen, S., Roederer, A., Sokolsky, O. V., & Lee, I. (2017) · Proceedings of the IEEE · doi:10.1109/jproc.2017.2723847 · Cited by 22
“There are multiple ways to model insulin-glucose physiology ranging from low-fidelity minimal models [125] to a high-fidelity maximal models such as the FDA-accepted UVa/Padova Type 1 Diabetes Mellitus Metabolic Simulator (T1DMS) [143].”
MethodControl-Oriented Model With Intra-Patient Variations for an Artificial Pancreas
Moscoso-Vásquez, M., Colmegna, P. H., Rosales, N., Garelli, F., & Sánchez‐Peña, R. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.2969389 · Cited by 24
“The baseline control-oriented LPV model used in this work is the one developed in [14], [46] that is based on the UVA/Padova metabolic simulator [33].”
MethodLi, P., Yu, L., Fang, Q., & Lee, S.-Y. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1436-y · Cited by 22
“Those control-relevant models are important for evaluating a control strategy for patients [9] or acting as a component in decision support systems to providing prediction directly to the user [31].”
BackgroundWendt, S. L., Ranjan, A., Møller, J. K., Schmidt, S., Knudsen, C. B., Holst, J. J., Madsbad, S., Madsen, H., Nørgaard, K., & Jørgensen, J. B. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817693254 · Cited by 20
“The ability of insulin to suppress the glycogenolytic response to glucagon at high insulin concentration is not reflected in previously published models of glucose-glucagon dynamics [18-20].”
BackgroundBertachi, A. H., Beneyto, A., Ramkissoon, C. M., & Vehí, J. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0392 · Cited by 23
Deep Multi-Output Forecasting: Learning to Accurately Predict Blood Glucose Trajectories
Fox, I., Ang, L. P., Jaiswal, M., Pop‐Busui, R., & Wiens, J. (2018) · PubMed · doi:10.48550/arxiv.1806.05357 · Cited by 24
Robust H∞ control of T–S fuzzy blood glucose regulation system via adaptive event-triggered scheme
Yan, S., Chu, L.-L., & Cai, Y. (2023) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2023.104643 · Cited by 21
Layered Meta-Learning Algorithm for Predicting Adverse Events in Type 1 Diabetes
D’Antoni, F., Petrosino, L., Marchetti, A., Bacco, L., Pieralice, S., Vollero, L., Pozzilli, P., Piemonte, V., & Merone, M. (2023) · IEEE Access · doi:10.1109/access.2023.3237992 · Cited by 24
Deichmann, J., Bachmann, S., Burckhardt, M., Pfister, M., Szinnai, G., & Kaltenbach, H. (2023) · PLoS Computational Biology · doi:10.1371/journal.pcbi.1010289 · Cited by 25
A simple modeling framework for prediction in the human glucose–insulin system
Sirlanci, M., Levine, M. E., Wang, C. C. L., Albers, D. J., & Stuart, A. M. (2023) · Chaos An Interdisciplinary Journal of Nonlinear Science · doi:10.1063/5.0146808 · Cited by 25
Robust Data-Driven Control of Artificial Pancreas Systems Using Neural Networks
Dutta, S., Kushner, T., & Sankaranarayanan, S. (2018) · Lecture notes in computer science · doi:10.1007/978-3-319-99429-1_11 · Cited by 26
Cobelli, C., & Man, C. D. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211015268 · Cited by 24
Self-citationA Model of Self-Monitoring Blood Glucose Measurement Error
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817698498 · Cited by 21
Self-citationApplication of Robust Fixed Point Control in Case of T1DM
Eigner, G., Horvath, P., Tar, J. K., Rudas, I. J., & Kovács, L. (2015) · Conference proceedings/Conference proceedings - IEEE International Conference on Systems, Man, and Cybernetics · doi:10.1109/smc.2015.430 · Cited by 19
“Also, this models are the base of the UVA/Padova simulator [18].”
MethodMessori, M., Cobelli, C., & Magni, L. (2015) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2015.09.148 · Cited by 20
“…the latest MPC for AP used in in-vivo clinical trials is presented in (Toffanin et al., 2013), whose process model is the linearization around a basal fictitious equilibrium of the average nonlinear model describing the adult virtual population of the UVA/Padova simulator (Dalla Man et al., 2014).”
BackgroundMethodSelf-citationMPC Model Individualization in Free-Living Conditions: A Proof-of-Concept Case Study
Toffanin, C., Favero, S. D., Aiello, E. M., Messori, M., Cobelli, C., & Magni, L. (2017) · IFAC-PapersOnLine · doi:10.1016/j.ifacol.2017.08.271 · Cited by 18
Self-citationCappon, G., Marturano, F., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818777524 · Cited by 18
Roversi, C., Vettoretti, M., Favero, S. D., Facchinetti, A., Choudhary, P., & Sparacino, G. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211012392 · Cited by 19
“A suitable tool for this purpose is the T1D patient decision simulator (T1D-PDS), based on the well-known FDA-approved UVA/Padova T1D physiological model, 24,25 a simulation tool that allows performing in silico trials in T1D and was shown to reproduce realistic treatment scenarios.”
Method
Metadata
- Authors
- Man, C. D., Micheletto, F., Lv, D., Breton, M., Kovatchev, B., & Cobelli, C.
- Affiliations
- University of Padua (IT)
- University of Virginia (US)
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2014-01
- Volume
- 8
- Issue
- 1
- Pages
- 26-34
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4454102
- OpenAlex
- W2129995532
- Cited by
- 791
- References
- 25
- Citations by year
- 2026: 49 · 2025: 85 · 2024: 76 · 2023: 89 · 2022: 69 · 2021: 76 · 2020: 91 · 2019: 77 · 2018: 66 · 2017: 47 · 2016: 30 · 2015: 23 · 2014: 10
- FWCI
- 29.50
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Padua Research Archive (University of Padova) (submittedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Pancreatic function and diabetes
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- type 1 diabetes, hypoglycemia, glucose kinetics, virtual patients, correction factors, in silico clinical trials, artificial pancreas, glucose sensor
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-09, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-09 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.