Analytical Performance Requirements for Systems for Self-Monitoring of Blood Glucose With Focus on System Accuracy: Relevant Differences Among ISO 15197:2003, ISO 15197:2013, and Current FDA Recommendations
Open accessThe DOI matches its Crossref record, with a small difference in title or year; 35 of the 35 references it cites were found in a registry or an index.
Abstract
In the European Union (EU), the ISO (International Organization for Standardization) 15197 standard is applicable for the evaluation of systems for self-monitoring of blood glucose (SMBG) before the market approval. In 2013, a revised version of this standard was published. Relevant revisions in the analytical performance requirements are the inclusion of the evaluation of influence quantities, for example, hematocrit, and some changes in the testing procedures for measurement precision and system accuracy evaluation, for example, number of test strip lots. Regarding system accuracy evaluation, the most important change is the inclusion of more stringent accuracy criteria. In 2014, the Food and Drug Administration (FDA) in the United States published their own guidance document for the premarket evaluation of SMBG systems with even more stringent system accuracy criteria than stipulated by ISO 15197:2013. The establishment of strict accuracy criteria applicable for the premarket evaluation is a possible approach to further improve the measurement quality of SMBG systems. However, the system accuracy testing procedure is quite complex, and some critical aspects, for example, systematic measurement difference between the reference measurement procedure and a higher-order procedure, may potentially limit the apparent accuracy of a given system. Therefore, the implementation of a harmonized reference measurement procedure for which traceability to standards of higher order is verified through an unbroken, documented chain of calibrations is desirable. In addition, the establishment of regular and standardized post-marketing evaluations of distributed test strip lots should be considered as an approach toward an improved measurement quality of available SMBG systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Hispanic Americans have values of A1C that are intermediate between Blacks and Whites.136”
From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [136]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
Standards of Medical Care in Diabetes
Power, D. (2006) · Diabetes Care · doi:10.2337/diacare.29.02.06.dc05-1593 · Cited by 8,673
Standards of medical care in diabetes—2014
American Diabetes Association (2014) · Diabetes Care · doi:10.2337/dc14-s014 · Cited by 4,572
Basevi, V., Di Mario, S., Morciano, C., Nonino, F., & Magrini, N. (2011) · Diabetes Care · doi:10.2337/dc11-0174 · Cited by 3,055
Standards of medical care in diabetes—2011
American Diabetes Association (2011) · Diabetes Care · doi:10.2337/dc11-s011 · Cited by 2,880
Polonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Petersen, B., Schweitzer, M., & Wagner, R. S. (2011) · Diabetes Care · doi:10.2337/dc10-1732 · Cited by 429
American Association of Diabetes Educators 2017
Carracher, A. M., Marathe, P. H., & Close, K. L. (2017) · Journal of Diabetes · doi:10.1111/1753-0407.12603 · Cited by 300
Self-Monitoring of Blood Glucose
American Diabetes Association (1994) · Diabetes Care · doi:10.2337/diacare.17.1.81 · Cited by 255
Consensus Statement on Self-Monitor ing of Blood Glucose
(1987) · Diabetes Care · doi:10.2337/diacare.10.1.95 · Cited by 248
Boyd, J. C., & Bruns, D. E. (2001) · Clinical Chemistry · doi:10.1093/clinchem/47.2.209 · Cited by 227
System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195 · cited 1×
“Over the last years a number of evaluation studies were published showing that not all available SMBG systems comply with currently still applicable ISO 15197:2003 accuracy standards.(14,16,18,23,43) Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version o…”
BackgroundSelf-citationBergenstal, R. M., & Gavin, J. R. (2005) · The American Journal of Medicine · doi:10.1016/j.amjmed.2005.07.055 · Cited by 166
Tack, C., Pohlmeier, H., Behnke, T., Schmid, V., Grenningloh, M., Forst, T., & Pfützner, A. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0170 · Cited by 92 · cited 1×
“5 mg/ dl) is comparably strict, and published data indicate that many available systems may not reliably achieve this accuracy.(15,16,19,25) The FDA also proposes that 3 test strip lots shall be included in the accuracy evaluation.”
ResultBaumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., & Freckmann, G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600511 · Cited by 90 · cited 3×
“Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version of this standard.(14,16,23,44) However, accuracy data of such postmarketing evaluations do not allow for a general conclusion regarding the measurement performance of a given SMBG system as this evalua…”
BackgroundSelf-citationFreckmann, G., Baumstark, A., Schmid, C., Pleus, S., Link, M., & Haug, C. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0208 · Cited by 77 · cited 3×
“Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version of this standard.(14,16,23,44) However, accuracy data of such postmarketing evaluations do not allow for a general conclusion regarding the measurement performance of a given SMBG system as this evalua…”
BackgroundResultSelf-citationPlasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 74 · cited 1×
“On the contrary, in 2 studies, systematic differences of up to 8% between a GOD method and a HK method were observed.(29,30) In 2 recent SMBG system accuracy evaluations following the requirements of the ISO 15197 standard, accuracy results differed depending on the applied reference measurement procedure.”
MethodBrazg, R. L., Klaff, L. J., & Parkin, C. G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700117 · Cited by 68 · cited 2×
“Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version of this standard.(14,16,23,44) However, accuracy data of such postmarketing evaluations do not allow for a general conclusion regarding the measurement performance of a given SMBG system as this evalua…”
BackgroundStandardized Evaluation of Nine Instruments for Self-Monitoring of Blood Glucose
Kristensen, G. B. B., Monsen, G., Skeie, S., & Sandberg, S. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0034 · Cited by 56
Accuracy and Precision Evaluation of Seven Self-Monitoring Blood Glucose Systems
Kuo, C.-Y., Hsu, C.-T., Ho, C.-S., Su, T.-E., Wu, M.-H., & Wang, C.-J. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0223 · Cited by 55 · cited 1×
“Over the last years a number of evaluation studies were published showing that not all available SMBG systems comply with currently still applicable ISO 15197:2003 accuracy standards.(14,16,18,23,43) Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version o…”
BackgroundKlaff, L. J., Brazg, R., Hughes, K., Tideman, A. M., Schachner, H. C., Stenger, P., Pardo, S., Dunne, N., & Parkes, J. L. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0069 · Cited by 46 · cited 1×
“5 mg/ dl) is comparably strict, and published data indicate that many available systems may not reliably achieve this accuracy.(15,16,19,25) The FDA also proposes that 3 test strip lots shall be included in the accuracy evaluation.”
Result(1994) · doi:10.2337/diacare.17.6.595 · Cited by 32
(2015) · doi:10.1177/1932296815580161 · Cited by 31 · cited 1×
“In 2 recent SMBG system accuracy evaluations following the requirements of the ISO 15197 standard, accuracy results differed depending on the applied reference measurement procedure.(31,32) Thus, the reference measurement procedure used in such evaluations can be the deciding factor whether accuracy data obtained for…”
MethodSelf-citationAccuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial
Halldorsdottir, S., Warchal-Windham, M. E., Wallace, J. F., Pardo, S., Parkes, J. L., & Simmons, D. A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700520 · Cited by 31
Virdi, N., Daskiran, M., Nigam, S., Kozma, C., & Raja, P. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0047 · Cited by 27
Assessing the Quality of Publications Evaluating the Accuracy of Blood Glucose Monitoring Systems
Thorpe, G. H. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0265 · Cited by 27 · cited 1×
“However, most of these studies show remarkable deviations from the complex testing procedure recommended by the ISO standard, for example, number of samples included, sample type, and investigated glucose concentration categories.(27) According to ISO 15197, evaluation of SMBG system accuracy should be performed with…”
ResultAssessment of three frequently used blood glucose monitoring devices in clinical routine
Züger, T., Schuler, V., Stettler, C., Diem, P., & Christ, E. (2012) · Swiss Medical Weekly · doi:10.4414/smw.2012.13631 · Cited by 26
Hasslacher, C., Kulozik, F., & Platten, I. (2013) · Clinical Laboratory · doi:10.7754/clin.lab.2012.120710 · Cited by 26 · cited 1×
“Unsurprisingly, a remarkable number of systems failed to meet the more stringent accuracy criteria of the new version of this standard.(14,16,23,44) However, accuracy data of such postmarketing evaluations do not allow for a general conclusion regarding the measurement performance of a given SMBG system as this evalua…”
BackgroundA comparison of five glucometers in South Africa
Essack, Y., Hoffman, M., Rensburg, M., van Wyk, J., Meyer, C., & Erasmus, R. T. (2009) · Journal of Endocrinology Metabolism and Diabetes of South Africa · doi:10.1080/22201009.2009.10872202 · Cited by 26
Glucose Measurement: Time for a Gold Standard
Hagvik, J. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100205 · Cited by 25
Andreis, E., Küllmer, K., & Appel, M. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814523886 · Cited by 22
Hasslacher, C., Kulozik, F., & Platten, I. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814522804 · Cited by 22
Regulation of medical devices used in diabetology in Europe: Time for reform?
Boulton, A. J. M., & Prato, S. D. (2012) · Diabetologia · doi:10.1007/s00125-012-2580-7 · Cited by 16
Heinemann, L., Freckmann, G., & Koschinsky, T. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700230 · Cited by 14 · cited 1×
“Currently, the premarket evaluation of SMBG systems is a 1-time procedure which is usually conducted by the manufacturer itself and often criticized to be a nontransparent process.(41,42) Since independent premarket evaluations are required neither in the United States nor in the European Union, a selection of test st…”
Self-citationAccuracy Evaluation of Blood Glucose Monitoring Systems in Children on Overnight Closed-Loop Control
DeSalvo, D. J., Shanmugham, S., Ly, T. T., Wilson, D. M., & Buckingham, B. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532238 · Cited by 10
Research works printed but not linked: 4
- 10. Richtlinie 98/79/EG Des Europäischen Parlaments Und Des Rates Vom 27. October 1998. Über In-Vitro-Diagnostika. 1998 [Google Scholar]
- 15. Freckmann G, Baumstark A, Zschornack E, et al. Evaluation of the system accuracy of 34 blood glucose monitoring systems according to EN ISO 15197. Poster presented at: ADA; June 8-12, 2012; Philadelphia, PA. [Google Scholar]
- 32. Freckmann G, Link M, Baumstark A, Schmid C, Pleus S, Haug C. System accuracy evaluation of 10 SMBG systems with 3 lots each following ISO 15197:2013 against 2 different comparison methods. Paper presented at: EASD; September 15-19, 2014; Vienna, Austria. [Google Scholar]
- 33. American Diabetes Association. Self-monitoring of blood glucose. Diabetes Care. 1996;19(suppl 1):S62-S66. [Google Scholar]
Cited by
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
“…waves, called surface plasmon polaritons (SPPs), are excited by an electromagnetic field radiated onto a thin layer of a highly conductive and chemically inert metal such as gold. decaying (evanescent) electric field that is highly sensitive to changes surrounding medium (SPR resonance peak).”
BackgroundEnzyme‐Based Glucose Sensor: From Invasive to Wearable Device
Lee, H., Hong, Y. J., Baik, S., Hyeon, T., & Kim, D. (2018) · Advanced Healthcare Materials · doi:10.1002/adhm.201701150 · Cited by 778
Blonde, L., Umpierrez, G. E., Reddy, S. S., McGill, J. B., Berga, S. L., Bush, M., Chandrasekaran, S., DeFronzo, R. A., Einhorn, D., Galindo, R. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.08.002 · Cited by 507
Bailey, T. S., Grunberger, G., Bode, B. W., Handelsman, Y., Hirsch, I. B., Jovanovič, L., Roberts, V. L., Rodbard, D., Tamborlane, W. V., & Walsh, J. S. A. (2016) · Endocrine Practice · doi:10.4158/ep151124.cs · Cited by 151
Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816641433 · Cited by 115
Self-citation
Metadata
- Authors
- Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L.
- Affiliations
- Institut für Diabetes-Technologie (Germany) (DE)
- Edingen-Neckarhausen, Germany
- Profil Institute for Metabolic Research (DE)
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2015-04-14
- Volume
- 9
- Issue
- 4
- Pages
- 885-894
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4525642
- OpenAlex
- W2001788756
- Cited by
- 70
- References
- 35
- Citations by year
- 2026: 6 · 2025: 7 · 2024: 6 · 2023: 8 · 2022: 7 · 2021: 7 · 2020: 5 · 2019: 5 · 2018: 10 · 2017: 4 · 2016: 4 · 2015: 1
- FWCI
- 0.96
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Universität Ulm
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- self-monitoring of blood glucose, system accuracy, ISO 15197, FDA guidance, measurement precision, influence quantities, hematocrit, reference measurement procedure, traceability, calibration chain, post-marketing evaluation, measurement quality
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Europe, Hematocrit, Humans, Hypoglycemia, United States, United States Food and Drug Administration, Blood Glucose Self-Monitoring, Reproducibility of Results, Guidelines as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.