Comparative Accuracy of 17 Point-of-Care Glucose Meters
Open accessThe DOI matches its Crossref record; 40 of the 40 references it cites were found in a registry or an index.
Abstract
BACKGROUND: The accuracy of point-of-care blood glucose (BG) meters is important for the detection of dysglycemia, calculation of insulin doses, and the calibration of continuous glucose monitors. The objective of this study was to compare the accuracy of commercially available glucose meters in a challenging laboratory study using samples with a wide range of reference BG and hemoglobin values. METHODS: Fresh, discarded blood samples from a hospital STAT laboratory were either used without modification, spiked with a glucose solution, or incubated at 37°C to produce 347 samples with an even distribution across reference BG levels from 20 to 440 mg/dl and hemoglobin values from 9 to 16 g/dl. We measured the BG of each sample with 17 different commercially available glucose meters and the reference method (YSI 2300) at the same time. We determined the mean absolute relative difference (MARD) for each glucose meter, overall and stratified by reference BG and by hemoglobin level. RESULTS: The accuracy of different meters widely, exhibiting a range of MARDs from 5.6% to 20.8%. Accuracy was lower in the hypoglycemic range, but was not consistently lower in samples with anemic blood hemoglobin levels. CONCLUSIONS: The accuracy of commercially available glucose meters varies widely. Although the sample mix in this study was much more challenging than those that would be collected under most use conditions, some meters were robust to these challenges and exhibited high accuracy in this setting. These data on relative accuracy and robustness to challenging samples may be useful in informing the choice of a glucose meter.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In a 2017 study, however, only 2 of 17 commercial meters intended for home BGM use met the ISO standard (125).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [125]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, a study involving 17 commercially available home glucometers found that only two met the ISO standard and that accuracy decreased when BG was in the hypoglycemic range [5].”
From Non-Diabetic Hypoglycemia: Evaluation and Management in Adults (2025) · cites this paper as [5]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
1 retracted
Sutheran, H. L., & Reynolds, T. (2015) · Journal of Clinical Pathology · doi:10.1136/jclinpath-2014-202797 · Cited by 2
Retracted(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes
(2005) · New England Journal of Medicine · doi:10.1056/nejmoa052187 · Cited by 5,138
Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Unrecognized Anemia in Patients With Diabetes
Thomas, M. C., MacIsaac, R. J., Tsalamandris, C., Power, D., & Jerums, G. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.1164 · Cited by 347
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195
Tang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00038 · Cited by 188
Research works printed but not linked: 3
- 6. Centers for Disease Control and Prevention. National Diabetes Statistics Report: Estimates of Diabetes and Its Burden in the United States, 2014. Atlanta, GA: US Department of Health and Human Services; 2014. [Google Scholar]
- 10. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes Mellitus. EN ISO 15197:2003. (E). [Google Scholar]
- 11. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013. (E). [Google Scholar]
Cited by
Items 101–127 of 127
Le dépistage et la prise en charge des nouveau-nés à risque d’hypoglycémie
Narvey, M. R., & Marks, S. D. (2019) · Paediatrics & Child Health · doi:10.1093/pch/pxz135 · Cited by 1
Predictive low glucose suspend systems
Forlenza, G. P., & Ekhlaspour, L. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00014-4 · Cited by 1
Self-citationMondal, H., & Mondal, S. (2020) · Indian Journal of Medical Sciences · doi:10.25259/ijms_150_2020 · Cited by 1
“There is a difference in the accuracy of different glucose monitors.[7,8] Hence, the doctors and patients should decide the most appropriate meter for them after reviewing the current literature.”
BackgroundDapare, P. P. M. (2021) · Annals of Medical Laboratory Science · doi:10.51374/annalsmls.2021.1.2.0037 · Cited by 1
MacPherson, B. A. (2018) · The Atrium (University of Guelph) · Cited by 2
Development of glucose biosensors using nanostructured composites
Al-Sagur, H. (2018) · Sheffield Hallam University · doi:10.7190/shu-thesis-00120
Point-of-care testing in diabetes management
Cozma, A., Vonica, C. L., Sitar-Tăut, A., & Fodor, A. (2019) · Revista română de medicină de laborator · doi:10.2478/rrlm-2019-0014
“Moreover, the price of the glucose strips did not correlate with the accuracy of the result (18).”
BackgroundBasal Insulin Adjustments and Continuous Glucose Monitoring During Exercise in Type 1 Diabetes
Zaharieva, D. P. (2018) · YorkSpace (York University)
The development of nanosensors for in vivo detection of physiological molecules
Luo, Y. (2018) · Digital Repository Service (Northeastern University) · doi:10.17760/d20316234
Needle Optimization for Wrist-Based Electronic Mosquito Pilot Human Testing
Teixeira, J., Wilkes, R., Yadid-Pecht, O., & Mintchev, M. P. (2019) · IEEE International Conference on Bioinformatics and Bioengineering (BIBE) · doi:10.1109/bibe.2019.00094
“These include the following four: (a) lack of autonomicity; (b) invasive blood sampling; (c) indiscrete equipment and patient alienation; and (d) inaccurate glucose readings [4, 5, 6].”
Needle and Potentiostat Study for Blood Glucose Monitoring: Third-Generation E-Mosquito
Rojas, J. T. D. C. (2020) · doi:10.11575/prism/37628
The Technology of Glucose Sensors
Chetty, T. (2020) · Springer eBooks · doi:10.1007/978-3-030-42806-8_2
Analytical performance of SMBG systems
Ahn, D. T., & Klonoff, D. C. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00002-8
Cost-Effectiveness of Monitoring Metabolic Control
Smokovski, I. (2020) · Springer eBooks · doi:10.1007/978-3-030-51469-3_5 · Cited by 1
Medical Monitoring of Preexisting DM and GDM
Kintiraki, E., & Goulis, D. G. (2022) · Springer eBooks · doi:10.1007/978-3-030-89243-2_7
Kosteneffektivität der Überwachung der metabolischen Kontrolle
Smokovski, I. (2023) · Springer eBooks · doi:10.1007/978-3-031-27793-1_5
Bonetti, G., Garzaniti, A., VUVAN, G., SALVADORI, S., ZALTIERI, G., & Carta, M. (2025) · La Rivista Italiana della Medicina di Laboratorio - Italian Journal of Laboratory Medicine · doi:10.23736/s1825-859x.25.00277-4
Bond, M. M., Asma, E., Peterson, J., Oparah, L., Alooh, M., Kumara, D., Oden, Z. M., Ezeaka, C., MOLYNEUX, E. M., & Richards-Kortum, R. R. (2025) · BMC Pediatrics · doi:10.1186/s12887-025-05934-9 · Cited by 2
“This performance agrees with that shown by Ehklaspour et al., who reported an over-all MARD (Mean Absolute Relative Difference) of 6.3% for StatStripXpress [22]; in our experiments, the MARD ranged from 3.9 to 7.9% across the tested conditions (Fig.”
InfluentialRozario, N. M., & Shanthi, M. (2025) · IEEE International Conference on Innovations in Engineering and Next-Generation Technologies for Sustainability (ICINVENTS) · doi:10.1109/icinvents64613.2025.11401409
Calibración, trazabilidad e incertidumbre de los equipos de medición en la práctica diagnóstica
Guarneros-García, O., Rojas-Ramirez, S.-R., Zacarías-Santiago, A., & Gutiérrez-Urueta, G. L. (2026) · Enfermería Universitaria · doi:10.22201/feno.23958421e.2023.3.1443
Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation
Ahamed, F., Pal, D., Palepu, S., Somashekara, J., Sahoo, S., Roy, A., Saha, T., Das, S., & Goswami, K. (2026) · Indian Journal of Community Medicine · doi:10.4103/ijcm.ijcm_932_25
“[21,22] Subgroup Bland–Altman analysis showed that the HemoCue® Glucose 201RT system performed better at lower FBS levels, while its performance remained relatively consistent across different PPBS values.”
Comparison of Four Point-of-Care Glucometers for Neonatal Glucose Testing—A Diagnostic Study
Sundar, A., Arnepalli, N., Usha Devi, R., & Sethuraman, G. (2026) · Clinical Therapeutics · doi:10.1016/j.clinthera.2026.06.021
Advanced Technology in Diabetes: Sensors and Pumps
Forlenza, G. P. (2026) · Elsevier eBooks · doi:10.1016/b978-0-443-11985-9.00023-8
Neurodevelopmental Consequences of Neonatal Hypoglycemia
Rosenfeld, E., & León, D. D. D. (2026) · Cambridge University Press eBooks · doi:10.1017/9781009418034.028
Letelier, P. J., Fierro, G., Cuminao, K. A. H., Huenchuleo, A., Boguen, R. E., & Guzmán, N. H. (2026) · Journal of Medical Biochemistry · doi:10.5937/jomb0-64436
“Technical aspects include the chemical basis of the test strips and the algorithms used to calculate plasma glucose (28, 29).”
Ruiz, J., Buros‐Stein, A., Wheelock, A., Stewart, B., Bernique, A. S., Jost, E., Garvey, K. C., Astley, C. M., & Cobry, E. C. (2026) · Diabetes Technology & Therapeutics · doi:10.1177/15209156261486139
Metadata
- Authors
- Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J.
- Affiliations
- Massachusetts General Hospital (US)
- Massachusetts General Hospital Diabetes Research Center, Boston, MA, USA
- Corresponding author
- Russell, Steven J.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2016-10-03
- Volume
- 11
- Issue
- 3
- Pages
- 558-566
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC5505415
- OpenAlex
- W2529767183
- Semantic Scholar
- 5e3e282622eb98bf27bbfad30cef36065db8a325
- Cited by
- 127
- References
- 40
- Citations by year
- 2026: 9 · 2025: 14 · 2024: 12 · 2023: 13 · 2022: 5 · 2021: 13 · 2020: 21 · 2019: 24 · 2018: 16
- FWCI
- 13.40
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (R01DK097657)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- point-of-care glucose meters, blood glucose measurement accuracy, mean absolute relative difference, hemoglobin interference, anemia, hypoglycemia, insulin dose calculation
- MeSH terms
- Data Accuracy, Blood Glucose, Humans, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.