Comparative Accuracy of 17 Point-of-Care Glucose Meters
Open accessThe DOI matches its Crossref record; 40 of the 40 references it cites were found in a registry or an index.
Abstract
BACKGROUND: The accuracy of point-of-care blood glucose (BG) meters is important for the detection of dysglycemia, calculation of insulin doses, and the calibration of continuous glucose monitors. The objective of this study was to compare the accuracy of commercially available glucose meters in a challenging laboratory study using samples with a wide range of reference BG and hemoglobin values. METHODS: Fresh, discarded blood samples from a hospital STAT laboratory were either used without modification, spiked with a glucose solution, or incubated at 37°C to produce 347 samples with an even distribution across reference BG levels from 20 to 440 mg/dl and hemoglobin values from 9 to 16 g/dl. We measured the BG of each sample with 17 different commercially available glucose meters and the reference method (YSI 2300) at the same time. We determined the mean absolute relative difference (MARD) for each glucose meter, overall and stratified by reference BG and by hemoglobin level. RESULTS: The accuracy of different meters widely, exhibiting a range of MARDs from 5.6% to 20.8%. Accuracy was lower in the hypoglycemic range, but was not consistently lower in samples with anemic blood hemoglobin levels. CONCLUSIONS: The accuracy of commercially available glucose meters varies widely. Although the sample mix in this study was much more challenging than those that would be collected under most use conditions, some meters were robust to these challenges and exhibited high accuracy in this setting. These data on relative accuracy and robustness to challenging samples may be useful in informing the choice of a glucose meter.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In a 2017 study, however, only 2 of 17 commercial meters intended for home BGM use met the ISO standard (125).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [125]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“However, a study involving 17 commercially available home glucometers found that only two met the ISO standard and that accuracy decreased when BG was in the hypoglycemic range [5].”
From Non-Diabetic Hypoglycemia: Evaluation and Management in Adults (2025) · cites this paper as [5]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
1 retracted
Sutheran, H. L., & Reynolds, T. (2015) · Journal of Clinical Pathology · doi:10.1136/jclinpath-2014-202797 · Cited by 2
Retracted(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Intensive diabetes treatment and cardiovascular disease in patients with type 1 diabetes
(2005) · New England Journal of Medicine · doi:10.1056/nejmoa052187 · Cited by 5,138
Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Tonyushkina, K., & Nichols, J. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300446 · Cited by 406
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Unrecognized Anemia in Patients With Diabetes
Thomas, M. C., MacIsaac, R. J., Tsalamandris, C., Power, D., & Jerums, G. (2003) · Diabetes Care · doi:10.2337/diacare.26.4.1164 · Cited by 347
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195
Tang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00038 · Cited by 188
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
Gandhi, G. Y., Kovalaske, M., Kudva, Y., Walsh, K., Elamin, M. B., Beers, M., Coyle, C., Goalen, M., Murad, M. S., Erwin, P. J., et al. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500419 · Cited by 134
S. Karon, B., Griesmann, L., Scott, R., Bryant, S. C., Dubois, J. A., Shirey, T. L., Presti, S., & Santrach, P. J. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0257 · Cited by 132
Tack, C., Pohlmeier, H., Behnke, T., Schmid, V., Grenningloh, M., Forst, T., & Pfützner, A. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0170 · Cited by 92
Baumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., & Freckmann, G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600511 · Cited by 90
Real-time continuous glucose monitoring system for treatment of diabetes: a systematic review
Hoeks, L. B. E. A., Greven, W. L., & de Valk, H. W. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2010.03177.x · Cited by 88
Freckmann, G., Baumstark, A., Schmid, C., Pleus, S., Link, M., & Haug, C. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0208 · Cited by 77
Self-Monitoring of Blood Glucose: The Use of the First or the Second Drop of Blood
Hortensius, J., Slingerland, R. J., Kleefstra, N., Logtenberg, S. J. J., Groenier, K. H., Houweling, S. T., & Bilo, H. J. G. (2011) · Diabetes Care · doi:10.2337/dc10-1694 · Cited by 76
Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 74
Hematocrit Interference of Blood Glucose Meters for Patient Self-Measurement
Ramljak, S., Lock, J. P., Schipper, C., Musholt, P. B., Forst, T., Lyon, M., & Pfützner, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700123 · Cited by 72
Brazg, R. L., Klaff, L. J., & Parkin, C. G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700117 · Cited by 68
Klaff, L. J., Brazg, R., Hughes, K., Tideman, A. M., Schachner, H. C., Stenger, P., Pardo, S., Dunne, N., & Parkes, J. L. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0069 · Cited by 46
Musholt, P. B., Schipper, C., Thomé, N., Ramljak, S., Schmidt, M., Forst, T., & Pfützner, A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500520 · Cited by 46
The Business of Self-Monitoring of Blood Glucose: A Market Profile
Hughes, M. D. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300530 · Cited by 41
Blood Glucose: Measurement in the Point-of-Care Setting
Kiechle, F. L., & Main, R. I. (2000) · Laboratory Medicine · doi:10.1309/4bf1-et6t-wfe3-m7xa · Cited by 39
Freckmann, G., Pleus, S., Link, M., Baumstark, A., Schmid, C., Högel, J., & Haug, C. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0043 · Cited by 37
(2015) · doi:10.1177/1932296815580161 · Cited by 31
Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial
Halldorsdottir, S., Warchal-Windham, M. E., Wallace, J. F., Pardo, S., Parkes, J. L., & Simmons, D. A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700520 · Cited by 31
Francescato, M. P., Geat, M., Stel, G., & Cauci, S. (2011) · Clinica Chimica Acta · doi:10.1016/j.cca.2011.10.012 · Cited by 29
The Accuracy of Home Glucose Meters in Hypoglycemia
Sonmez, A., Yilmaz, Z., Uckaya, G., Kilic, S., Tapan, S., Taslipinar, A., Aydogdu, A., Yazici, M., Yilmaz, M. I., Serdar, M., et al. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0183 · Cited by 23
Precision and accuracy of two blood glucose meters: FreeStyle Flash versus One Touch Ultra
Rivers, S. M., Kane, M. P., Bakst, G., Busch, R. S., & Hamilton, R. A. (2006) · American Journal of Health-System Pharmacy · doi:10.2146/ajhp050473 · Cited by 22
Evaluation of hematocrit interference with MyStar extra and seven competitive devices
Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814565790 · Cited by 22
[Reference values for hematologic laboratory tests and hematologic disorders in the aged].
Takubo, T., & TATSUMI, N. (2000) · PubMed · Cited by 18
Pfützner, A., Hengesbach, C., Demircik, F., Schipper, C., Forst, T. M., & Musholt, P. B. (2013) · Current Medical Research and Opinion · doi:10.1185/03007995.2013.858617 · Cited by 17
Pfützner, A., Schipper, C., Ramljak, S., Flacke, F., Sieber, J., Forst, T., & Musholt, P. B. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700122 · Cited by 16
Fogh‐Andersen, N., D’Orazio, P., Kuwa, K., Külpmann, W. R., Mager, G., & Larsson, L. (2000) · PubMed Central · Cited by 15
Huang, T.-Y., Chang, H.-W., Tsao, M.-F., Chuang, S.-M., Ni, C.-C., Sue, J.-W., Lin, H.-C., & Hsu, C.-T. (2014) · Clinica Chimica Acta · doi:10.1016/j.cca.2014.02.023 · Cited by 15
Pfützner, A., Musholt, P. B., Schipper, C., Demircik, F., Hengesbach, C., Flacke, F., Sieber, J., & Forst, T. M. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700613 · Cited by 13
Accuracy Evaluation of Blood Glucose Monitoring Systems in Children on Overnight Closed-Loop Control
DeSalvo, D. J., Shanmugham, S., Ly, T. T., Wilson, D. M., & Buckingham, B. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532238 · Cited by 10
Sutheran, H. L., & Reynolds, T. (2016) · Journal of Clinical Pathology · doi:10.1136/jclinpath-2015-203339 · Cited by 10
Research works printed but not linked: 3
- 6. Centers for Disease Control and Prevention. National Diabetes Statistics Report: Estimates of Diabetes and Its Burden in the United States, 2014. Atlanta, GA: US Department of Health and Human Services; 2014. [Google Scholar]
- 10. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes Mellitus. EN ISO 15197:2003. (E). [Google Scholar]
- 11. International Organization for Standardization. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013. (E). [Google Scholar]
Cited by
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Advances in technology for management of type 1 diabetes
Beck, R. W., Bergenstal, R. M., Laffel, L. M., & Pickup, J. C. (2019) · The Lancet · doi:10.1016/s0140-6736(19)31142-0 · Cited by 242
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
“8%.(146) Indeed, only one-third of the meters assessed”
BackgroundMethod
Metadata
- Authors
- Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J.
- Affiliations
- Massachusetts General Hospital (US)
- Massachusetts General Hospital Diabetes Research Center, Boston, MA, USA
- Corresponding author
- Russell, Steven J.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2016-10-03
- Volume
- 11
- Issue
- 3
- Pages
- 558-566
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC5505415
- OpenAlex
- W2529767183
- Semantic Scholar
- 5e3e282622eb98bf27bbfad30cef36065db8a325
- Cited by
- 127
- References
- 40
- Citations by year
- 2026: 9 · 2025: 14 · 2024: 12 · 2023: 13 · 2022: 5 · 2021: 13 · 2020: 21 · 2019: 24 · 2018: 16
- FWCI
- 13.40
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (R01DK097657)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- point-of-care glucose meters, blood glucose measurement accuracy, mean absolute relative difference, hemoglobin interference, anemia, hypoglycemia, insulin dose calculation
- MeSH terms
- Data Accuracy, Blood Glucose, Humans, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review