Day-and-Night Closed-Loop Glucose Control in Patients With Type 1 Diabetes Under Free-Living Conditions: Results of a Single-Arm 1-Month Experience Compared With a Previously Reported Feasibility Study of Evening and Night at Home

Open access
Renard, Eric, Farret, Anne, Kropff, Jort, Bruttomesso, Daniela, Messori, Mirko, Place, Jerome, Visentin, Roberto, Calore, Roberta +13 more·2016·Diabetes Care

Abstract

OBJECTIVE: After testing of a wearable artificial pancreas (AP) during evening and night (E/N-AP) under free-living conditions in patients with type 1 diabetes (T1D), we investigated AP during day and night (D/N-AP) for 1 month. RESEARCH DESIGN AND METHODS: Twenty adult patients with T1D who completed a previous randomized crossover study comparing 2-month E/N-AP versus 2-month sensor augmented pump (SAP) volunteered for 1-month D/N-AP nonrandomized extension. AP was executed by a model predictive control algorithm run by a modified smartphone wirelessly connected to a continuous glucose monitor (CGM) and insulin pump. CGM data were analyzed by intention-to-treat with percentage time-in-target (3.9–10 mmol/L) over 24 h as the primary end point. RESULTS: Time-in-target (mean ± SD, %) was similar over 24 h with D/N-AP versus E/N-AP: 64.7 ± 7.6 vs. 63.6 ± 9.9 (P = 0.79), and both were higher than with SAP: 59.7 ± 9.6 (P = 0.01 and P = 0.06, respectively). Time below 3.9 mmol/L was similarly and significantly reduced by D/N-AP and E/N-AP versus SAP (both P < 0.001). SD of blood glucose concentration (mmol/L) was lower with D/N-AP versus E/N-AP during whole daytime: 3.2 ± 0.6 vs. 3.4 ± 0.7 (P = 0.003), morning: 2.7 ± 0.5 vs. 3.1 ± 0.5 (P = 0.02), and afternoon: 3.3 ± 0.6 vs. 3.5 ± 0.8 (P = 0.07), and was lower with D/N-AP versus SAP over 24 h: 3.1 ± 0.5 vs. 3.3 ± 0.6 (P = 0.049). Insulin delivery (IU) over 24 h was higher with D/N-AP and SAP than with E/N-AP: 40.6 ± 15.5 and 42.3 ± 15.5 vs. 36.6 ± 11.6 (P = 0.03 and P = 0.0004, respectively). CONCLUSIONS: D/N-AP and E/N-AP both achieved better glucose control than SAP under free-living conditions. Although time in the different glycemic ranges was similar between D/N-AP and E/N-AP, D/N-AP further reduces glucose variability.

Integrity signals

Registry

Crossref · 10.2337/dc16-0008

Cites

26 of 26 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • In July 2016, Diabetes Care published a symposium featuring several outpatient AID studies using DiAs running in two different in silico designed algorithm configurations.

    “In July 2016, Diabetes Care published a symposium, 129 which featured several outpatient AID studies using DiAs running in two different in silico designed algorithm configurations.[123][124][125]”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Renard, E., Farret, A., Kropff, J., Bruttomesso, D., Messori, M., Place, J., Visentin, R., Calore, R., Toffanin, C., Di Palma, F., Lanzola, G., Magni, P., Boscari, F., Galasso, S., Avogaro, A., Keith-Hynes, P., Kovatchev, B., Del Favero, S., Cobelli, C., Magni, L., & DeVries, J. H.
Affiliations
  • Department of Endocrinology, Diabetes, Nutrition, Montpellier University Hospital; INSERM Clinical Investigation Centre 1411; Institute of Functional Genomics, CNRS UMR 5203, INSERM U1191, University of Montpellier, Montpellier, France
  • Centre National de la Recherche Scientifique (FR)
  • Amsterdam UMC Location University of Amsterdam (NL)
  • University of Padua (IT)
  • University of Pavia (IT)
  • Department of Civil Engineering and Architecture, University of Pavia, Pavia, Italy
  • University of Virginia (US)
  • Department of Information Engineering, University of Padova, Padova, Italy
Venue
Diabetes Care
Published
2016-04-17
Volume
39
Issue
7
Pages
1151-1160
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W2346512890
Cited by
104
References
26
Citations by year
2026: 1 · 2025: 2 · 2024: 1 · 2023: 6 · 2022: 8 · 2021: 9 · 2020: 11 · 2019: 16 · 2018: 20 · 2017: 22 · 2016: 8
FWCI
22.14
Type
journal article
Language
en
Open access
Locations
  • Diabetes Care (publishedVersion)
  • PubMed (publishedVersion)
  • Pure Amsterdam UMC (publishedVersion, other-oa)
  • Padua Research Archive (University of Padova) (submittedVersion)
  • Data Archiving and Networked Services (DANS) (submittedVersion)
  • HAL (submittedVersion)
Funders
European Commission, Framework Program 7 (FP7-ICT-2009-4 grant number 247138)
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
Keywords
type 1 diabetes, closed-loop glucose control, model predictive control, continuous glucose monitoring, insulin pump, sensor-augmented pump, time in target, hypoglycemia, blood glucose variability, free-living conditions, smartphone control, insulin delivery, insulin resistance
MeSH terms
Adult, Algorithms, Blood Glucose, Circadian Rhythm, Diabetes Mellitus, Type 1, Feasibility Studies, Female, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged, Social Conditions, Blood Glucose Self-Monitoring, Cross-Over Studies, Pancreas, Artificial, Young Adult

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review