Correlation of fingerstick blood glucose measurements with GlucoWatch biographer glucose results in young subjects with type 1 diabetes.
The DOI matches its Crossref record.
Abstract
OBJECTIVE: The purpose of this study was to compare measurements of glucose obtained via iontophoretic extraction with the GlucoWatch automatic glucose biographer (Cygnus, Inc., Redwood City, CA) with capillary blood glucose values that were determined 1) in a controlled outpatient clinic setting and 2) in a home setting. RESEARCH DESIGN AND METHODS: There were 76 GlucoWatch biographers used on 28 different young adults (21 women and 7 men) with type 1 diabetes (age 30.9 +/- 6.9 years and duration of diabetes 18.4 +/- 8.1 years [mean +/- SD]) in a controlled outpatient clinic setting. Some subjects participated on multiple days. Subjects wore two GlucoWatch biographers, each on the forearm (ventral aspect). Comparisons were made to HemoCue blood glucose analyzer (Aktiebolgat Leo, Helsingborg, Sweden) capillary blood glucose measurements. In addition, GlucoWatch biographers (one each day for 3 consecutive days) were used by 12 subjects (8 women, 4 men) in a home setting. Comparisons were made to capillary blood glucose values determined using the One Touch Profile meter (Johnson & Johnson, New Brunswick, NJ). RESULTS: GlucoWatch biographer glucose values correlated well with capillary blood glucose values determined using the HemoCue analyzer in the clinic setting (r = 0.90, 1,554 paired data points) and using the One Touch Profile meter in the home setting (r = 0.85, 204 paired data points). When 36 subjects wore two biographers simultaneously, the correlation between the two biographers was r = 0.94. The error grid analysis demonstrated that > 96% of biographer glucose values determined in the clinic or home setting were in the clinically acceptable A and B regions. CONCLUSIONS: This study confirms the accuracy and precision of glucose values as determined using the GlucoWatch biographer in clinic and home settings.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Type 1 diabetes: new perspectives on disease pathogenesis and treatment
Atkinson, M. A., & Eisenbarth, G. S. (2001) · The Lancet · doi:10.1016/s0140-6736(01)05415-0 · Cited by 1,491
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Sacks, D. B., Bruns, D. E., Goldstein, D. E., Maclaren, N. K., McDonald, J. M., & Parrott, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.436 · Cited by 1,130
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Fluorescence-based glucose sensors
Pickup, J. C., Hussain, F., Evans, N. D., Rolinski, O. J., & Birch, D. J. S. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.10.002 · Cited by 616
“Another available technology for ose sensing in diabetes, reverse iontophoresis (Tamada e l., 1999; Garg et al., 1999), also has special problems su s skin irritation and inaccuracies, which have restri ts widespread uptake into diabetes management (Pickup, 004).”
Non-Invasive Blood Glucose Monitoring Technology: A Review
Tang, L., Chang, S. J., Chen, C.-J., & Liu, J.-T. (2020) · Sensors · doi:10.3390/s20236925 · Cited by 481
A history of blood glucose meters and their role in self-monitoring of diabetes mellitus
Clarke, S. F., & Foster, J. R. (2012) · British Journal of Biomedical Science · doi:10.1080/09674845.2012.12002443 · Cited by 456
Tierney, M. J., Tamada, J. A., Potts, R. O., Jovanovic, L., & Garg, S. (2001) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00189-0 · Cited by 390
Self-citationMechanisms of sampling interstitial fluid from skin using a microneedle patch
Samant, P. P., & Prausnitz, M. R. (2018) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1716772115 · Cited by 363
Lipani, L., Dupont, B. G. R., Doungmene, F., Marken, F., Tyrrell, R. M., Guy, R. H., & Ilie, A. (2018) · Nature Nanotechnology · doi:10.1038/s41565-018-0112-4 · Cited by 340
Hadgraft, J. (2001) · International Journal of Pharmaceutics · doi:10.1016/s0378-5173(01)00731-1 · Cited by 297
Sensors for glucose monitoring: technical and clinical aspects
Koschinsky, T., & Heinemann, L. (2001) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.188 · Cited by 288
Tura, A., Maran, A., & Pacini, G. (2007) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2006.10.027 · Cited by 283
“Comparisons were made to capillary blood glucose (One Touch Profile meter) [52].”
MethodPhysiological Differences Between Interstitial Glucose and Blood Glucose Measured in Human Subjects
Kulcu, E., Tamada, J. A., Reach, G., Potts, R. O., & Lesho, M. J. (2003) · Diabetes Care · doi:10.2337/diacare.26.8.2405 · Cited by 251
Self-citationWearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 237
Coming of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 223
“Minimally invasive, realtime continuous glucose measurement [5, 6] is progressing to accurate, insulin-dosing approved, factory-calibrated systems [7].”
BackgroundCan Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 213
“To-day, almost all minimal and noninvasive glucose detection methods, such as iontophoresis 7,8 , hypodermic needles 9,10 , electrochemical/ enzymatic sensors, 11–15 ultrasound, 16 and optical approaches, 17,18 have focused on estimating glucose in the interstitial fluid (ISF) compartment of the…”
In vivo glucose monitoring: the clinical reality and the promise
Pickup, J. C., Hussain, F., Evans, N. D., & Sachedina, N. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.08.016 · Cited by 203
Continuous Subcutaneous Glucose Monitoring in Children With Type 1 Diabetes
Chase, H. P., Kim, L. M., Owen, S. L., MacKenzie, T. A., Klingensmith, G. J., Murtfeldt, R., & Garg, S. K. (2001) · Pediatrics · doi:10.1542/peds.107.2.222 · Cited by 194
Self-citationKoschwanez, H. E., & Reichert, W. M. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.034 · Cited by 181
In-Vivo Electrochemistry: What Can We Learn about Living Systems?
Wilson, G. S., & Johnson, M. A. (2008) · Chemical Reviews · doi:10.1021/cr068082i · Cited by 178
Monitoring glycemic control: the importance of self-monitoring of blood glucose
Renard, E. (2005) · The American Journal of Medicine · doi:10.1016/j.amjmed.2005.07.052 · Cited by 99
Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00304-9 · Cited by 76
“, 1988), in a liquid extracted from the subcutaneous tissue by iontophoresis (Garg et al., 1999), or microdialysis (Sternberg et al.”
BackgroundForst, T., Caduff, A., Talary, M., Weder, M., Brändle, M., Kann, P., Flacke, F., Friedrich, C., & Pfützner, A. (2006) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2006.8.94 · Cited by 51
Metadata
- Authors
- Garg, S. K., Potts, R. O., Ackerman, N. R., Fermi, S. J., Tamada, J. A., & Chase, H. P.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 1999-10-01
- Volume
- 22
- Issue
- 10
- Pages
- 1708-1714
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2033633407
- Cited by
- 201
- References
- 0
- FWCI
- 9.57
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.