Wireless Power Transfer Approaches for Medical Implants: A Review
Open accessAbstract
Wireless power transmission (WPT) is a critical technology that provides an alternative for wireless power and communication with implantable medical devices (IMDs). This article provides a study concentrating on popular WPT techniques for IMDs including inductive coupling, microwave, ultrasound, and hybrid wireless power transmission (HWPT) systems. Moreover, an overview of the major works is analyzed with a comparison of the symmetric and asymmetric design elements, operating frequency, distance, efficiency, and harvested power. In general, with respect to the operating frequency, it is concluded that the ultrasound-based and inductive-based WPTs have a low operating frequency of less than 50 MHz, whereas the microwave-based WPT works at a higher frequency. Moreover, it can be seen that most of the implanted receiver’s dimension is less than 30 mm for all the WPT-based methods. Furthermore, the HWPT system has a larger receiver size compared to the other methods used. In terms of efficiency, the maximum power transfer efficiency is conducted via inductive-based WPT at 95%, compared to the achievable frequencies of 78%, 50%, and 17% for microwave-based, ultrasound-based, and hybrid WPT, respectively. In general, the inductive coupling tactic is mostly employed for transmission of energy to neuro-stimulators, and the ultrasonic method is used for deep-seated implants.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“This range minimizes eddy current losses, thereby reducing the risk of localized heating and its subsequent influence on neural stability (Maaß et al., 2017; Haerinia and Shadid, 2020; Mahmood et al., 2022).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Haerinia and Shadid, 2020
In the citing paper
“A value closer to 1 not only maximizes energy transfer but also minimizes potential thermal damage to neural tissues—a concern for BMIs requiring high computational power for real-time neural decoding (Haerinia and Shadid, 2020; Khan et al., 2020).”
From Comparative analysis of energy transfer mechanisms for neural implants (2024) · cites this paper as Haerinia and Shadid, 2020
Citations
Cites
Nagaoka, T., Watanabe, S., Sakurai, K., Kunieda, E., Watanabe, S., Taki, M., & Yamanaka, Y. (2003) · Physics in Medicine and Biology · doi:10.1088/0031-9155/49/1/001 · Cited by 775 · cited 1×
“[65] 2004 Body 400 MHz, 900 MHz, 2400 MHz Visible human, FDTD [66] 2004 Body/51 tissues 30 MHz–3 GHz FDTD [67] 2002 Head/10 tissues 900 MHz, 1800 MHz Visible human, FDTD”
Wireless power transfer to deep-tissue microimplants
Ho, J. S., Yeh, A. J., Neofytou, E., Kim, S., Tanabe, Y., Patlolla, B., Beygui, R. E., & Poon, A. S. Y. (2014) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1403002111 · Cited by 541 · cited 3×
“[82], which overcomes the difficulty of miniaturizing the power source via adaptive electromagnetic energy transport.”
BackgroundInfluentialLow, Z. N., Chinga, R. A., Tseng, R., & Lin, J. (2008) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2008.2010110 · Cited by 528 · cited 1×
“Although, in practice, the implanted coil is significantly smaller than the transmitting coil [70].”
BackgroundPower Approaches for Implantable Medical Devices
Amar, A. B., Kouki, A. B., & Cao, H. (2015) · Sensors · doi:10.3390/s151128889 · Cited by 445 · cited 2×
“The second category is split into two subsections: batteries, such as lithium; and natural harvesting, including biofuel cell, thermoelectricity, piezoelectricity, electrostatic, and electromagnetic [1].”
BackgroundDevelopment of Implantable Medical Devices: From an Engineering Perspective
Joung, Y.-H. (2013) · International Neurourology Journal · doi:10.5213/inj.2013.17.3.98 · Cited by 367 · cited 4×
“The designers of medical implants are currently dealing with challenges in materials, output power, size miniaturization, the efficiency of the wireless link, and cybersecurity [118].”
BackgroundInfluentialWireless Power Transmission With Self-Regulated Output Voltage for Biomedical Implant
Ahn, D., & Hong, S. (2013) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2013.2273472 · Cited by 315 · cited 1×
“13 10 doutT = 21 doutR = 1 [19] 2014 8.”
Performances of an Implanted Cavity Slot Antenna Embedded in the Human Arm
Xia, W., Saitô, K., Takahashi, M., & Ito, K. (2009) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2009.2014579 · Cited by 250 · cited 2×
“Assuming far-field WPT, the budget power link as discussed in [71] can be described as follows: (cid:3047)(cid:3028) (cid:3047)(cid:3034) (cid:3033) (cid:3045)(cid:3028) (cid:2868) (cid:1840) (cid:3042) (cid:3045) (cid:3030) (cid:3031) (cid:1842) (cid:1833) (cid:1838) (cid:1833) (cid:1840) where P…”
BackgroundCharthad, J., Weber, M. J., Chang, T. C., & Arbabian, A. (2015) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2015.2427336 · Cited by 245
Channel model for wireless communication around human body
Ryckaert, J., Doncker, P. D., Meys, R., Hoye, A. D. L., & Donnay, S. (2004) · Electronics Letters · doi:10.1049/el:20040386 · Cited by 243 · cited 1×
“[65] 2004 Body 400 MHz, 900 MHz, 2400 MHz Visible human, FDTD [66] 2004 Body/51 tissues 30 MHz–3 GHz FDTD [67] 2002 Head/10 tissues 900 MHz, 1800 MHz Visible human, FDTD”
Ultrasonic transcutaneous energy transfer for powering implanted devices.
Ozeri, S., & Shmilovitz, D. (2010) · Ultrasonics · doi:10.1016/j.ultras.2009.11.004 · Cited by 241 · cited 2×
“[105] 2010 673 kHz 1000 27 40 [106] 2003 100 kHz 5400 36 40 [25] 2002 1 MHz 2100 20 40 [26] 2001 1 MHz 20 30 * Width and total length of 48 symmetric elements of the spherical transducer array.”
High-Speed OQPSK and Efficient Power Transfer Through Inductive Link for Biomedical Implants
Simard, G., Sawan, M., & Massicotte, D. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2009.2039212 · Cited by 193 · cited 2×
“Typically, FSK [24], PSK [25], ASK [26], OOK [27], LSK [28], PPSK [29],”
BackgroundHigh-Efficiency Transcutaneous Energy Transfer for Implantable Mechanical Heart Support Systems
Knecht, O., Bosshard, R., & Kolar, J. W. (2015) · IEEE Transactions on Power Electronics · doi:10.1109/tpel.2015.2396194 · Cited by 187 · cited 1×
“56 70 doutT = 100 doutR = 18 [30] 2015 2 MHz 1450 27 80 doutT = 140 doutR = 65 [80] 2015 800 kHz 30 w 95 20 doutT = 70 doutR = 34 [81] 2012 742 kHz 85 0–50 doutT = 38 doutR = 16.”
Efficient Wireless Power Transfer System Integrating With Metasurface for Biological Applications
Li, L., Liu, H., Zhang, H., & Xue, W. (2017) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2017.2756580 · Cited by 178
Power-Efficient Impedance-Modulation Wireless Data Links for Biomedical Implants
Mandal, S., & Sarpeshkar, R. (2008) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2008.2005295 · Cited by 163
Peyman, A., Gabriel, C., Grant, E. H., Vermeeren, G., & MARTENS, L. C. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/54/2/004 · Cited by 154 · cited 1×
“[63] 2009 Head/16 tissues 50 MHz–20 GHz Measured properties, FDTD”
BackgroundA Miniaturized Triple-Band Implantable Antenna System for Bio-Telemetry Applications
Zada, M., & Yoo, H. (2018) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2018.2874681 · Cited by 149 · cited 1×
“Zada et al. [84] provided a miniaturized implantable antenna with three frequency bands (902–928, 2400–2483.5, and 1824–1980 MHz) operating at the industrial, scientific, and medical (ISM) band and at the midfield band.”
BackgroundInfluentialBasir, A., & Yoo, H. (2020) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2020.2965938 · Cited by 149 · cited 1×
“Reference Year Frequency Output Power (mW) Efficiency (%) Active Range (mm) Transmitter Dimensions (mm) Receiver Dimensions (mm) [86] 2020 1.”
BackgroundAn Ultrasonically Powered Implantable Micro-Oxygen Generator (IMOG)
Maleki, T., Cao, N., Song, S., Kao, C. H., Ko, S., & Ziaie, B. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2011.2163634 · Cited by 144 · cited 1×
“The empirical results can be obtained in vivo [8,42,43], using a living organism, or in vitro [44–46], outside of a living organism.”
BackgroundWireless Power Transfer for Medical Microsystems
Sun, T., Xie, X., & Wang, Z. (2013) · Springer eBooks · doi:10.1007/978-1-4614-7702-0 · Cited by 138 · cited 1×
“More details on how to derive the efficiency of inductive links can be found in [68].”
BackgroundRobust Wireless Power Transmission to mm-Sized Free-Floating Distributed Implants
Mirbozorgi, S. A., Yeon, P., & Ghovanloo, M. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2663358 · Cited by 131 · cited 1×
“A free-floating neural implant, which is insensitive to the location, is provided as an inductive link in [10] for wireless energy transmission.”
BackgroundAdaptive Transcutaneous Power Transfer to Implantable Devices: A State of the Art Review
Bocan, K. N., & Sejdić, E. (2016) · Sensors · doi:10.3390/s16030393 · Cited by 123 · cited 2×
“The development of implantable electronic devices in biological systems has made it In Figure 1, the power ranges of common IMDs are illustrated [1,6]. power ranges of common The WPT systems for the neurostimulator and the pacemaker are discussed in detail in [7–13].”
BackgroundMeng, M., & Kiani, M. (2016) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2016.2583783 · Cited by 120
Zargham, M., & Gulak, P. G. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2014.2328318 · Cited by 115
Ibrahim, A., & Kiani, M. (2016) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2016.2515541 · Cited by 110
Efficient Four-Coil Wireless Power Transfer for Deep Brain Stimulation
Yang, C., Chang, C., Lee, S.-Y., Chang, S.-J., & Chiou, L. (2017) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2017.2658560 · Cited by 97
Jiang, D., Cirmirakis, D., Schormans, M., Perkins, T. A., Donaldson, N., & Demosthenous, A. (2016) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2016.2580513 · Cited by 97 · cited 1×
“Typically, FSK [24], PSK [25], ASK [26], OOK [27], LSK [28], PPSK [29], Signals 2020 , 1 211 QMPM [30], QPSK [31], and impedance modulation [32] have been used for data communication units in medical applications.”
MethodOmnidirectional Ultrasonic Powering for Millimeter-Scale Implantable Devices
Song, S.-H., Kim, A., & Ziaie, B. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2444854 · Cited by 95 · cited 2×
“Song et al. [95] investigated omnidirectional ultrasonic powering for deep implantable microdevices.”
BackgroundMethodImaging Electric Properties of Biological Tissues by RF Field Mapping in MRI
Zhang, X., Zhu, S., & He, B. (2010) · IEEE Transactions on Medical Imaging · doi:10.1109/tmi.2009.2036843 · Cited by 94
A Wireless Magnetic Resonance Energy Transfer System for Micro Implantable Medical Sensors
Li, X., Zhang, H., Peng, F., Li, Y., Yang, T., Wang, B., & Fang, D.-M. (2012) · Italian National Conference on Sensors · doi:10.3390/s120810292 · Cited by 90 · cited 1×
“56 70 doutT = 100 doutR = 18 [30] 2015 2 MHz 1450 27 80 doutT = 140 doutR = 65 [80] 2015 800 kHz 30 w 95 20 doutT = 70 doutR = 34 [81] 2012 742 kHz 85 0–50 doutT = 38 doutR = 16.”
A Review on Miniaturized Ultrasonic Wireless Power Transfer to Implantable Medical Devices
Taalla, R. V., Arefin, M. S., Kaynak, A., & Kouzani, A. Z. (2018) · IEEE Access · doi:10.1109/access.2018.2886780 · Cited by 89 · cited 2×
“…as lithium; and In this paper, common WPT approaches for IMDs, including inductive coupling, microwave, and ultrasound, are studied. natural harvesting, including HWPT systems, a mixture of two di ff erent methods, are also reviewed. methods were evaluated by Taalla et al. [2] and Shadid et al. [3]”
BackgroundMethodFaerber, J., Cummins, G., Pavuluri, S., Record, P., Rodriguez, A. R. A., Lay, H. S., McPhillips, R., Cox, B., Connor, C., Gregson, R., et al. (2017) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2759254 · Cited by 87 · cited 1×
“The empirical results can be obtained in vivo [8,42,43], using a living organism, or in vitro [44–46], outside of a living organism.”
BackgroundA Q-Modulation Technique for Efficient Inductive Power Transmission
Kiani, M., Lee, B., Yeon, P., & Ghovanloo, M. (2015) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2015.2453201 · Cited by 82 · cited 3×
“56 70 doutT = 100 doutR = 18 [30] 2015 2 MHz 1450 27 80 doutT = 140 doutR = 65 [80] 2015 800 kHz 30 w 95 20 doutT = 70 doutR = 34 [81] 2012 742 kHz 85 0–50 doutT = 38 doutR = 16.”
MethodMiniature ultrasonically powered wireless nerve cuff stimulator
Larson, P. J., & Towe, B. C. (2011) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2011.5910538 · Cited by 79
Mobashsher, A. T., Mahmoud, A. A. O., & Abbosh, A. M. (2016) · Scientific Reports · doi:10.1038/srep20459 · Cited by 78
Mutashar, S., Hannan, M. A., Samad, S. A., & Hussain, A. (2014) · Italian National Conference on Sensors · doi:10.3390/s140711522 · Cited by 78 · cited 6×
“Mainly, determining SAR can be achieved by using numerical techniques and empirical models using fabricated tissue phantoms [4,34–38].”
BackgroundMethodInfluentialA Literature Survey on Wireless Power Transfer for Biomedical Devices
Shadid, R., & Noghanian, S. (2018) · International Journal of Antennas and Propagation · doi:10.1155/2018/4382841 · Cited by 77 · cited 3×
“…as lithium; and In this paper, common WPT approaches for IMDs, including inductive coupling, microwave, and ultrasound, are studied. natural harvesting, including HWPT systems, a mixture of two di ff erent methods, are also reviewed. methods were evaluated by Taalla et al. [2] and Shadid et al. [3]”
BackgroundMethodSelf-citationNear-Field Wireless Power Transfer to Deep-Tissue Implants for Biomedical Applications
Zhang, K., Liu, C., Jiang, Z. H., Zhang, Y., Liu, X., Guo, H., & Yang, X. (2019) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2019.2943424 · Cited by 77 · cited 1×
“Reference Year Frequency Output Power (mW) Efficiency (%) Active Range (mm) Transmitter Dimension (mm) Receiver Dimension (mm) [72] 2020 915 MHz 1.”
BackgroundThe Microbead: A Highly Miniaturized Wirelessly Powered Implantable Neural Stimulating System
Khalifa, A., Karimi, Y., Wang, Q., Montlouis, W., Garikapati, S., Stanaćević, M., Thakor, N. V., & Etienne‐Cummings, R. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2802443 · Cited by 77 · cited 1×
“[75] 2018 2, 4 MHz 126 25 6 doutT = 35 doutR = 20 [9] 2018 1.”
Inductive and Ultrasonic Multi-Tier Interface for Low-Power, Deeply Implantable Medical Devices
Sanni, A., Vilches, A., & Toumazou, C. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2175390 · Cited by 75 · cited 1×
“[116] 2012 200 kHz 8 1 70 doutT = 39 doutR = 33 Inductive and Ultrasonic”
Luo, B., Tao, L., Mai, R., Dai, R., He, Z. Y., & Li, W. (2018) · IET Power Electronics · doi:10.1049/iet-pel.2018.5279 · Cited by 70 · cited 2×
“…transfer from the capacitive link of the hybrid system is given by Equation (13): As shown above, the inductive link will generate the alternating magnetic field through coils, high voltages requires high produced voltage system to construct a hybrid electric power transfer is established in [107].”
MethodAsif, S. M., Iftikhar, A., Braaten, B. D., Ewert, D. L., & Maile, K. (2019) · IEEE Access · doi:10.1109/access.2019.2902981 · Cited by 69 · cited 1×
“[85] took advantage of a microwave-based WPT technique to charge deep medical implants like cardiac pacemakers.”
MethodSantagati, G. E., Dave, N., & Melodia, T. (2020) · IEEE/ACM Transactions on Networking · doi:10.1109/tnet.2019.2949805 · Cited by 68 · cited 2×
“[96] 2020 700 kHz 200 doutR = 10 [97] 2017 1 MHz 0.”
BackgroundPower and Information Transmission to Implanted Medical Device Using Ultrasonic
川鍋, 英., Katane, T., Saotome, H., Saito, O., & Kobayashi, K. (2001) · Japanese Journal of Applied Physics · doi:10.1143/jjap.40.3865 · Cited by 67 · cited 1×
“Typically, FSK [24], PSK [25], ASK [26], OOK [27], LSK [28], PPSK [29], Signals 2020 , 1 211 QMPM [30], QPSK [31], and impedance modulation [32] have been used for data communication units in medical applications.”
MethodKeshvari, J., Keshvari, R., & Lang, S. (2006) · Physics in Medicine and Biology · doi:10.1088/0031-9155/51/6/007 · Cited by 66 · cited 1×
“[64] 2006 Eye, head/15 tissues 900 MHz, 1800 MHz, 2450 MHz FDTD”
A 10.5 cm Ultrasound Link for Deep Implanted Medical Devices
Mazzilli, F., Lafon, C., & Dehollain, C. (2014) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2013.2295403 · Cited by 63
Chang, T. C., Wang, M. L., Charthad, J., Weber, M. J., & Arbabian, A. (2017) · Digest of technical papers/Digest of technical papers - IEEE International Solid-State Circuits Conference · doi:10.1109/isscc.2017.7870460 · Cited by 62
Rahmani, H., & Babakhani, A. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2876239 · Cited by 62 · cited 1×
“Mainly, determining SAR can be achieved by using numerical techniques and empirical models using fabricated tissue phantoms [4,34–38].”
MethodAn Inductive Link-Based Wireless Power Transfer System for Biomedical Applications
Adeeb, M. A., Islam, A. B., Haider, M. R., Tulip, F. S., Ericson, M. N., & ISLAM, S. K. (2012) · Active and Passive Electronic Components · doi:10.1155/2012/879294 · Cited by 61 · cited 1×
“Typically, FSK [24], PSK [25], ASK [26], OOK [27], LSK [28], PPSK [29], Signals 2020 , 1 211 QMPM [30], QPSK [31], and impedance modulation [32] have been used for data communication units in medical applications.”
MethodBraun, J., Tzschätzsch, H., Körting, C., de Schellenberger, A. A., Jenderka, M., Drießle, T., Ledwig, M., & Sack, I. (2017) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26659 · Cited by 61 · cited 1×
“5) GHz Measured properties, Cole–Cole [52] 2018 Brain, liver 200–1600 Hz Measured properties [53] 2018 Muscle, fat, skin 915 MHz and 2 GHz Measured properties”
BackgroundPower and Interactive Information Transmission to Implanted Medical Device Using Ultrasonic
Suzuki, S., Kimura, S., Katane, T., Saotome, H., Saito, O., & Kobayashi, K. (2002) · Japanese Journal of Applied Physics · doi:10.1143/jjap.41.3600 · Cited by 60 · cited 1×
“Typically, FSK [24], PSK [25], ASK [26], OOK [27], LSK [28], PPSK [29], Signals 2020 , 1 211 QMPM [30], QPSK [31], and impedance modulation [32] have been used for data communication units in medical applications.”
MethodA Dual-Band HF and UHF Antenna System for Implanted Neural Recording and Stimulation Devices
Sharma, A., Kampianakis, E., & Reynolds, M. S. (2016) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2016.2585650 · Cited by 59
Sun, G.-L., Muneer, B., Li, Y., & Zhu, Q. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2017.2787649 · Cited by 58
Wireless Power Link Based on Inductive Coupling for Brain Implantable Medical Devices
Manoufali, M., Bialkowski, K. S., Mohammed, B. J., & Abbosh, A. M. (2017) · IEEE Antennas and Wireless Propagation Letters · doi:10.1109/lawp.2017.2778698 · Cited by 58 · cited 2×
“[38] 2018 400 MHz 19, 82 1, 3, 6, 12, 15 d = 18 1 × 1”
In vivo electric conductivity of cervical cancer patients based on maps at 3T MRI
Balidemaj, E., de Boer, P., van Lier, A. L. H. M. W., Remis, R., Stalpers, L. J. A., Westerveld, H. G., Nederveen, A. J., van den Berg, C. A. T., & Crezee, J. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/4/1596 · Cited by 54
Effective Permittivity of Biological Tissue: Comparison of Theoretical Model and Experiment
Li, G., du Ning, & Zhang, L. (2017) · Mathematical Problems in Engineering · doi:10.1155/2017/7249672 · Cited by 54 · cited 1×
“[54] 2017 Blood, liver, fat, brain 10 kHz–10 MHz Bottcher–Bordewijk model, measured properties”
An Implantable Pressure Sensing System With Electromechanical Interrogation Scheme
Kim, A., Powell, C., & Ziaie, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2318023 · Cited by 53 · cited 3×
“et o s are restricte t a s rt tra sfer ista ce, isali ent issues ay cc r [94], a t e a etic fiel i te sit ust be under specified limitations for the safety of the body exposure.”
BackgroundInfluentialChip-Scale Coils for Millimeter-Sized Bio-Implants
Feng, P., Yeon, P., Cheng, Y., Ghovanloo, M., & Constandinou, T. G. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2853670 · Cited by 53 · cited 1×
“The maximum SAR value was studied in [40,41].”
BackgroundAhmadi, M. M., & Ghandi, S. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2851605 · Cited by 51 · cited 2×
“[75] 2018 2, 4 MHz 126 25 6 doutT = 35 doutR = 20 [9] 2018 1.”
A Printed Wearable Dual-Band Antenna for Wireless Power Transfer
Haerinia, M., & Noghanian, S. (2019) · Sensors · doi:10.3390/s19071732 · Cited by 51
Self-citationA Tissue-Channel Transcutaneous Power Transfer Technique for Implantable Devices
Chen, P., Yang, H., Luo, R., & Zhao, B. (2018) · IEEE Transactions on Power Electronics · doi:10.1109/tpel.2018.2791966 · Cited by 50 · cited 1×
“In [39], the body tissue was used as the power transfer channel.”
MethodNear-Field Wireless Power Transfer to Stent-Based Biomedical Implants
Wiseman, G., Redoute, J.-M., Ganesan, K., Rind, G. S., John, S. E., Ronayne, S. M., Opie, N. L., Garrett, D. J., & Prawer, S. (2018) · IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology · doi:10.1109/jerm.2018.2833386 · Cited by 49
Asif, S. M., Iftikhar, A., Hansen, J. W., Khan, M. S., Ewert, D. L., & Braaten, B. D. (2018) · IEEE Access · doi:10.1109/access.2018.2885620 · Cited by 49 · cited 4×
“In Figure 1, the discussed in detail in [7–13].”
BackgroundInfluentialLyu, H., Wang, J., La, J.-H., Chung, J. M., & Babakhani, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2852680 · Cited by 47 · cited 3×
“[8] have developed a stimulator that occupies an area of 5 mm × 7.”
BackgroundA Microwave Ring-Resonator Sensor for Non-Invasive Assessment of Meat Aging
Jilani, M. T., Wen, W. P., Cheong, L. Y., & Rehman, M. Z. U. (2016) · Sensors · doi:10.3390/s16010052 · Cited by 46 · cited 1×
“55) GHz Phantom/ FEM [58] 2016 Muscle 500 MHz–20 GHz Fricke [59] 2015 Eye/6 tissues (0.”
Meng, M., & Kiani, M. (2016) · IEEE Transactions on Circuits & Systems II Express Briefs · doi:10.1109/tcsii.2016.2626151 · Cited by 42 · cited 1×
“Meng et al. [110] developed a hybrid inductive-ultrasonic WPT link to power biomedical implants over bone, air, and tissue.”
BackgroundInductive link design for medical implants
Ali, H., Ahmad, T. J., & Khan, S. A. (2009) · 2009 IEEE Symposium on Industrial Electronics & Applications · doi:10.1109/isiea.2009.5356376 · Cited by 41
Multi-disciplinary challenges in tissue modeling for wireless electromagnetic powering: a review
Bocan, K. N., Mickle, M. H., & Sejdic, E. (2017) · IEEE Sensors Journal · doi:10.1109/jsen.2017.2748338 · Cited by 41 · cited 1×
“An assessment of variation in tissue electromagnetic properties was provided by Bocan et al. [48].”
MethodA Closed-Loop Remote Powering Link for Wireless Cortical Implants
Silay, K. M., Dehollain, C., & Declercq, M. (2013) · IEEE Sensors Journal · doi:10.1109/jsen.2013.2259621 · Cited by 40
Novel coated differentially fed dual‐band fractal antenna for implantable medical devices
Fan, Y., Liu, H., Liu, X., Cao, Y., Li, Z., & Tentzeris, M. M. (2019) · IET Microwaves Antennas & Propagation · doi:10.1049/iet-map.2018.6171 · Cited by 40
Design of 2.45 GHz microwave wireless power transfer system for battery charging applications
Li, K., See, K. Y., Koh, W.-J., & Zhang, J.-W. (2017) · Photonics & Electromagnetics Research Symposium - Fall (PIERS-Fall) · doi:10.1109/piers-fall.2017.8293542 · Cited by 39 · cited 1×
“[38] 2018 400 MHz 19, 82 - 1, 3, 6, 12, 15 doutT = 18 1 × 1 [89] 2018 280 MHz 44 3 30 × 80 [90] 2017 2.”
An 8 Mbps data rate transmission by inductive link dedicated to implantable devices
Lü, Z., & Sawan, M. (2008) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2008.4542103 · Cited by 34 · cited 2×
“QMPM [30], QPSK [31], and impedance modulation [32] have been used for data communication units in medical applications.”
MethodLow-Power Implanted Sensor for Orthodontic Bond Failure Diagnosis and Detection
Quadir, N. A., Albasha, L., Taghadosi, M., Qaddoumi, N. N., & Hatahet, B. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2791426 · Cited by 32
Internet of Things Use Cases for the Healthcare Industry
Raj, P., Chatterjee, J. M., Kumar, A., & Balamurugan, B. (2020) · Springer eBooks · doi:10.1007/978-3-030-37526-3 · Cited by 30
Wireless Power Transfer - Fundamentals and Technologies
Coca, E. (2016) · InTech eBooks · doi:10.5772/61488 · Cited by 26
Design, fabrication, and application of bio-implantable acoustic power transmission
Shih, P., & Shih, W. (2010) · Journal of Microelectromechanical Systems · doi:10.1109/jmems.2010.2042568 · Cited by 26 · cited 2×
“3 30–400 doutT = 8 [12] 2011 1 MHz 23 120 doutR = 8 [104] 2010 35 kHz 1.”
Compensating for Tissue Changes in an Ultrasonic Power Link for Implanted Medical Devices
Vihvelin, H., Leadbetter, J., Bance, M., Brown, J. A., & Adamson, R. B. A. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2015.2421823 · Cited by 26 · cited 2×
“184 [99] 2016 1 MHz 25 3–7 d = 8 [100] 2015 280 kHz 2.”
Lyu, H., Gad, P., Zhong, H., Edgerton, V. R., & Babakhani, A. (2018) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2018.2879357 · Cited by 26
Fornés-Leal, A., Cardona, N., Frasson, M., Castelló-Palacios, S., Nevárez, A., Pons, V., & Garcia‐Pardo, C. (2019) · IEEE Access · doi:10.1109/access.2019.2903481 · Cited by 24
Biocompatible wireless power transferring based on ultrasonic resonance devices
Lee, S. Q., Youm, W., & Hwang, G. (2013) · Proceedings of meetings on acoustics · doi:10.1121/1.4800386 · Cited by 23 · cited 4×
“…WPT system is an e ff ective method for medical applications such as cardiac defibrillators The ultrasonic-based WPT system is an effective method for medical applications such as cardiac defibrillators and deep brain stimulators (DBSs) [93]. defibrillators and deep brain stimulators (DBSs) [93].”
BackgroundMethodInfluentialMazzilli, F., & Dehollain, C. (2016) · Electronics Letters · doi:10.1049/el.2015.4267 · Cited by 23
HYBRID INDUCTIVE POWER TRANSFER AND WIRELESS ANTENNA SYSTEM FOR BIOMEDICAL IMPLANTED DEVICES
Shadid, R., Haerinia, M., Roy, S. K., & Noghanian, S. (2018) · Progress In Electromagnetics Research C · doi:10.2528/pierc18061604 · Cited by 23
Self-citationSuzuki, S., Katane, T., & Saito, O. (2003) · Journal of Artificial Organs · doi:10.1007/s10047-003-0209-9 · Cited by 19 · cited 2×
“[105] 2010 673 kHz 1000 27 40 [106] 2003 100 kHz 5400 36 40 [25] 2002 1 MHz 2100 20 40 [26] 2001 1 MHz 20 30 * Width and total length of 48 symmetric elements of the spherical transducer array.”
Zhao, B., Kuo, N., Liu, B., Li, Y.-A., Iotti, L., & Niknejad, A. M. (2019) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2019.2954772 · Cited by 19
Simulating, Modeling, and Sensing Variable Tissues for Wireless Implantable Medical Devices
Bocan, K. N., Mickle, M. H., & Sejdić, E. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2811497 · Cited by 18 · cited 1×
“5) GHz Measured properties, Cole–Cole [52] 2018 Brain, liver 200–1600 Hz Measured properties [53] 2018 Muscle, fat, skin 915 MHz and 2 GHz Measured properties”
BackgroundA 0.13µm CMOS integrated wireless power receiver for biomedical applications
Zargham, M., & Gulak, P. G. (2013) · European Solid-State Circuits Conference (ESSCIRC) · doi:10.1109/esscirc.2013.6649091 · Cited by 17
Development of highly efficient transducer for wireless power transmission system by ultrasonic
Shigeta, Y., Hori, Y., Fujimori, K., Tsuruta, K., & Nogi, S. (2011) · IEEE MTT-S International Microwave Workshop Series on Innovative Wireless Power Transmission: Technologies, Systems, and Applications (IMWS-IWPT) · doi:10.1109/imws.2011.5877115 · Cited by 17
Li, C., Chen, Q., Xie, Y., & Wu, T. (2014) · Bioelectromagnetics · doi:10.1002/bem.21835 · Cited by 17 · cited 1×
“2) THz FEM [61] 2014 Eye, head/14 tissues (0.”
Yu, Y., Hao, H., Wang, W., & Li, L. (2009) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2009.5332831 · Cited by 17 · cited 2×
“Medical implants like implanted spinal cord stimulators can use a rechargeable battery to improve their capability and reduce overall costs [5].”
BackgroundAnalytical Modeling of Small, Solenoidal, and Implantable Coils With Ferrite Tube Core
Cheng, Y., Chen, G., Xuan, D., Qian, G., Ghovanloo, M., & Wang, G. (2019) · IEEE Microwave and Wireless Components Letters · doi:10.1109/lmwc.2019.2891620 · Cited by 17
Rosa, B. M. G., & Yang, G. Z. (2016) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2016.7516281 · Cited by 16 · cited 2×
“Additionally, the acoustic emission can be used for supplying energy irelessly in the active biological environment [45].”
BackgroundGougheri, H. S., & Kiani, M. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591802 · Cited by 16 · cited 2×
“The empirical results can be obtained in vivo [8,42,43], using a living organism, or in vitro [44–46], outside of a living organism.”
BackgroundMonebhurrun, V., Dale, C., Bolomey, J., & Wiart, J. (2002) · IEEE Transactions on Magnetics · doi:10.1109/20.996193 · Cited by 15 · cited 1×
“[65] 2004 Body 400 MHz, 900 MHz, 2400 MHz Visible human, FDTD [66] 2004 Body/51 tissues 30 MHz–3 GHz FDTD [67] 2002 Head/10 tissues 900 MHz, 1800 MHz Visible human, FDTD”
Fang, B., Feng, T., Zhang, M., & Chakrabartty, S. (2015) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2015.7168750 · Cited by 13 · cited 1×
“6 18 18 doutT = 20 doutR = 20 [101] 2015 3.”
Leadbetter, J., Brown, J., & Adamson, R. B. (2013) · The Journal of the Acoustical Society of America · doi:10.1121/1.4805305 · Cited by 13
Lyu, H., Jian, Z., Liu, X., Sun, Y., & Babakhani, A. (2019) · IEEE Sensors Letters · doi:10.1109/lsens.2019.2896185 · Cited by 13
A Stent-Based Power and Data Link for Sensing Intravascular Biological Indicators
Aldaoud, A., Redouté, J., Ganesan, K., Rind, G. S., John, S. E., Ronayne, S. M., Opie, N. L., Garrett, D. J., & Prawer, S. (2018) · IEEE Sensors Letters · doi:10.1109/lsens.2018.2876350 · Cited by 13
Yang, C.-H., Li, W., & Chen, R. K. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.2972465 · Cited by 13
Analysis of Misalignment Effects on Link Budget of an Implantable Antenna
Haerinia, M., & Noghanian, S. (2019) · URSI International Symposium on Electromagnetic Theory · doi:10.23919/ursi-emts.2019.8931519 · Cited by 12
Self-citationShadid, R., Haerinia, M., & Noghanian, S. (2020) · Sensors · doi:10.3390/s20051368 · Cited by 12 · cited 1×
“The material type and amount needed for muscle phantom fabrication is summarized and measured in [47].”
BackgroundSelf-citationShort range remote powering of implanted electronics for freely moving animals
Kilinc, E. G., Ghanad, M. A., Maloberti, F., & Dehollain, C. (2013) · IEEE International New Circuits and Systems Conference (NEWCAS) · doi:10.1109/newcas.2013.6573608 · Cited by 11
Hao, J., Lv, L., Ju, L., Xie, X., Liu, Y., & Yang, H. (2018) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2018.12.025 · Cited by 11 · cited 1×
“[50] 2019 Muscle, fat, skin 50 MHz, 300 MHz, 700 MHz, and 900 MHz FDTD **”
Remotely powered implantable heart monitoring system for freely moving animals
Kilinc, E. G., Moya, A. C., van Lintel, H., Renaud, P., Maloberti, F., Wang, Q., & Dehollain, C. (2013) · International Workshop on Advances in Sensors and Interfaces (IWASI) · doi:10.1109/iwasi.2013.6576092 · Cited by 10
A low-power PPM demodulator for remotely powered batteryless implantable devices
Kilinc, E. G., Dehollain, C., & Maloberti, F. (2014) · IEEE International Midwest Symposium on Circuits and Systems (MWSCAS) · doi:10.1109/mwscas.2014.6908416 · Cited by 10
A Wireless Rechargeable Implantable System for Monitoring and Pacing the Gut in Small Animals
Javan‐Khoshkholgh, A., Sassoon, J. C., & Farajidavar, A. (2019) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2019.8919125 · Cited by 10
Study of Bending Effects on a Dual-Band Implantable Antenna
Haerinia, M., & Noghanian, S. (2019) · Digest - IEEE Antennas and Propagation Society. International Symposium · doi:10.1109/apusncursinrsm.2019.8888320 · Cited by 10 · cited 1×
“Therefore, the bending of a flexible substrate and the misalignment e ff ect must be investigated precisely [112].”
BackgroundSelf-citationExperimental validation of a hybrid Wireless Power Transfer - Power Line Communication system
Barmada, S., Tucci, M., Raugi, M., Dionigi, M., & Mezzanotte, P. (2016) · International Symposium on Power Line Communications and Its Applications · doi:10.1109/isplc.2016.7476259 · Cited by 9 · cited 1×
“In [108], the authors proposed a coupled WPT–Power Line Communication (PLC) system that consists of a two-coil resonator system.”
BackgroundAn ultra-low-power monitoring system for inductively coupled biomedical implants
Keikhosravy, K., Kamalinejad, P., Mirabbasi, S., Takahata, K., & Leung, V. C. M. (2013) · IEEE International Symposium on Circuits and Systems proceedings · doi:10.1109/iscas.2013.6572333 · Cited by 8 · cited 2×
“This reduces the cost, improves productivity, and provides compatibility and reliability of prototypes [14,15].”
BackgroundMethodA Multi-Transmitter Configuration for High-Safety Wireless Power Transfer Applications
Brizi, D., Fontana, N., Barmada, S., & Monorchio, A. (2019) · CINECA IRIS Institutial research information system (University of Pisa) · Cited by 8
Inductive power transmission to millimeter-sized biomedical implants using printed spiral coils
Ibrahim, A., & Kiani, M. (2016) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591801 · Cited by 7 · cited 2×
“The maximum SAR value was studied in [40,41].”
BackgroundDahdouh, S., Varsier, N., Ochoa, M. N., Wiart, J., Peyman, A., & Isabelle Bloch, A. (2016) · Physics in Medicine and Biology · doi:10.1088/0031-9155/61/4/1500 · Cited by 7 · cited 1×
“[55] 2016 Muscle, bladder, cervix 128 MHz Measured properties, Cole–Cole [56] 2016 Body/14 tissues 2.”
Design of Hybrid Wireless Power Transfer and Dual Ultrahigh-Frequency Antenna System
Haerinia, M., & Noghanian, S. (2019) · URSI International Symposium on Electromagnetic Theory · doi:10.23919/ursi-emts.2019.8931514 · Cited by 7 · cited 2×
“[109] decreased the size of the compact system, at the same time implementing multi-functionality.”
BackgroundMethodSelf-citationTerahertz signal propagation analysis inside the human skin
Abdelaziz, A. F., Abbasi, Q. H., Yang, K., Qaraqe, K. A., Hao, Y., & Alomainy, A. (2015) · IEEE International Conference on Wireless and Mobile Computing, Networking and Communications (WiMob) · doi:10.1109/wimob.2015.7347935 · Cited by 6
Wireless power beam device using microwave power transfer
Ishizaki, T., & Nishikawa, K. (2014) · IEEE Wireless Power Transfer Conference · doi:10.1109/wpt.2014.6839622 · Cited by 6
Modeling and simulation of inductive-based wireless power transmission systems
Haerinia, M. (2018) · De Gruyter eBooks · doi:10.1515/9783110445053-012 · Cited by 6
Self-citationDiao, Y., Leung, S.-W., Chan, K. H., Sun, W., Siu, Y. M., & Kong, R. Y. C. (2015) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncv514 · Cited by 5 · cited 1×
“55) GHz Phantom/ FEM [58] 2016 Muscle 500 MHz–20 GHz Fricke [59] 2015 Eye/6 tissues (0.”
ANALYSIS OF IMPLANTABLE ULTRASONIC COUPLING WIRELESS POWER TRANSMISSION SYSTEM
Yan, X., Zhu, Z., Liu, G., & Zhao, X. (2019) · Progress In Electromagnetics Research M · doi:10.2528/pierm19011001 · Cited by 2 · cited 2×
“The field model of the ultrasonic coupling wireless transmission system [92].”
Background
Research works printed but not linked: 3
- Lyu, 2018, A 430-MHz wirelessly powered implantable pulse generator with intensity/rate control and sub-1 µA quiescent current consumption, IEEE Trans. Biomed. Circuits Syst., 13, 180
- Institute of Electrical and Electronics Engineers (1999). IEEE Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields, 3 kHz to 300 GHz, IEEE.
- Lee, S.Q., Youm, W., Hwang, G., and Moon, K.S. (2015, January 12–16). Wireless power transferring and charging for implantable medical devices based on ultrasonic resonance. Proceedings of the 22nd International Congress on Sound and Vibration, Florence, Italy.
Cited by
Sorted by citations
Sagar, A., Kashyap, A., Nasab, M. A., Padmanaban, S., Bertoluzzo, M., Kumar, A., & Blaabjerg, F. (2023) · IEEE Access · doi:10.1109/access.2023.3300475 · Cited by 238
Advanced Energy Harvesters and Energy Storage for Powering Wearable and Implantable Medical Devices
Gao, Z., Zhou, Y., Zhang, J., Foroughi, J., Peng, S., Baughman, R. H., Wang, Z. L., & Wang, C. (2024) · Advanced Materials · doi:10.1002/adma.202404492 · Cited by 250
Wireless Power Transfer Technologies, Applications, and Future Trends: A Review
Alabsi, A., Hawbani, A., Wang, X., Al‐Dubai, A., Hu, J., Aziz, S. A., Kumar, S., Zhao, L., Shvetsov, A. V., & Alsamhi, S. H. (2024) · IEEE Transactions on Sustainable Computing · doi:10.1109/tsusc.2024.3380607 · Cited by 140
“Such surveys would offer insights into the latest developments and help guide the development and deployment of WPT technologies and applications.”
Energy Harvesting in Implantable and Wearable Medical Devices for Enduring Precision Healthcare
Shuvo, M. M. H., Titirsha, T., Amin, N., & ISLAM, S. K. (2022) · Energies · doi:10.3390/en15207495 · Cited by 137
Battery‐Free and Wireless Technologies for Cardiovascular Implantable Medical Devices
Zhang, J., Das, R., Zhao, J., Mirzai, N., Mercer, J. R., & Heidari, H. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202101086 · Cited by 112
Metadata
- Authors
- Haerinia, M., & Shadid, R.
- Affiliations
- University of North Dakota (US)
- Applied Science Private University (JO)
- Corresponding author
- Haerinia, Mohammad
- Venue
- Signals
- Publisher
- MDPI AG
- Published
- 2020-12-16
- Volume
- 1
- Issue
- 2
- Pages
- 209-229
- ISSN
- 2624-6120 (ISSN-L)
- OpenAlex
- W3112882954
- Cited by
- 174
- References
- 118
- Citations by year
- 2026: 15 · 2025: 41 · 2024: 41 · 2023: 36 · 2022: 33 · 2021: 8
- FWCI
- 6.61
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Signals (publishedVersion, cc-by)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Wireless Power Transfer Systems
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Induction Heating and Inverter Technology
- Keywords
- wireless power transfer, biomedical implants, inductive coupling, microwave power transmission, power transfer efficiency, nerve stimulator
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review