Multi-centre Randomised Controlled Trial of a Smartphone-based Event Recorder Alongside Standard Care Versus Standard Care for Patients Presenting to the Emergency Department with Palpitations and Pre-syncope: The IPED (Investigation of Palpitations in the ED) study
Open accessAbstract
BACKGROUND: Patients with palpitations and pre-syncope commonly present to Emergency Departments (EDs) but underlying rhythm diagnosis is often not possible during the initial presentation. This trial compares the symptomatic rhythm detection rate of a smartphone-based event recorder (AliveCor) alongside standard care versus standard care alone, for participants presenting to the ED with palpitations and pre-syncope with no obvious cause evident at initial consultation. METHODS: Multi-centre open label, randomised controlled trial. Participants ≥ 16 years old presenting to 10 UK hospital EDs were included. Participants were randomised to either (a) intervention group; standard care plus the use of a smartphone-based event recorder or (b) control group; standard care alone. Primary endpoint was symptomatic rhythm detection rate at 90 days. Trial registration number NCT02783898 (ClinicalTrials.gov). FINDINGS: Two hundred forty-three participants were recruited over an 18-month period. A symptomatic rhythm was detected at 90 days in 69 (n = 124; 55.6%; 95% CI 46.9-64.4%) participants in the intervention group versus 11 (n = 116; 9.5%; 95% CI 4.2-14.8) in the control group (RR 5.9, 95% CI 3.3-10.5; p < 0.0001). Mean time to symptomatic rhythm detection in the intervention group was 9.5 days (SD 16.1, range 0-83) versus 42.9 days (SD 16.0, range 12-66; p < 0.0001) in the control group. The commonest symptomatic rhythms detected were sinus rhythm, sinus tachycardia and ectopic beats. A symptomatic cardiac arrhythmia was detected at 90 days in 11 (n = 124; 8.9%; 95% CI 3.9-13.9%) participants in the intervention group versus 1 (n = 116; 0.9%; 95% CI 0.0-2.5%) in the control group (RR 10.3, 95% CI 1.3-78.5; p = 0.006). INTERPRETATION: Use of a smartphone-based event recorder increased the number of patients in whom an ECG was captured during symptoms over five-fold to more than 55% at 90 days. This safe, non-invasive and easy to use device should be considered part of on-going care to all patients presenting acutely with unexplained palpitations or pre-syncope. FUNDING: This study was funded by research awards from Chest, Heart and Stroke Scotland (CHSS) and British Heart Foundation (BHF) which included funding for purchasing the devices. MR was supported by an NHS Research Scotland Career Researcher Clinician award.
AI snapshot
Written from the paper’s full text.
TL;DR
In ED patients with unexplained palpitations or pre-syncope, adding a smartphone-based AliveCor event recorder to standard care increased symptomatic rhythm detection at 90 days from 9.5% to 55.6% and symptomatic cardiac arrhythmia detection from 0.9% to 8.9%. The device was safe and easy to use, but the trial was open-label and used a central ECG reading service not available in routine practice.
Key findings
- Symptomatic rhythm was detected at 90 days in 69/124 (55.6%; 95% CI 46.9–64.4%) intervention participants versus 11/116 (9.5%; 95% CI 4.2–14.8) control participants (RR 5.9, 95% CI 3.3–10.5; p < 0.0001).
- Symptomatic cardiac arrhythmia was detected at 90 days in 11/124 (8.9%; 95% CI 3.9–13.9%) intervention participants versus 1/116 (0.9%; 95% CI 0.0–2.5%) control participants (RR 10.3, 95% CI 1.3–78.5; p = 0.006).
- Mean time to symptomatic rhythm detection was 9.5 days (SD 16.1, range 0–83) in the intervention group versus 42.9 days (SD 16.0, range 12–66) in the control group (p < 0.0001); commonest symptomatic rhythms were sinus rhythm (53; 66.3%), sinus tachycardia (19; 23.8%) and ectopic beats (13; 16.3%).
- Eighty of 92 (87.0%) participants found the AliveCor monitor easy to use; median overall healthcare utilisation cost was £108 in the intervention group versus £0 in standard care (p = 0.0001), and cost per symptomatic rhythm diagnosis was £474 versus £1395.
Methods
- Multi-centre open label randomised controlled trial in EDs and Acute Medical Units of 10 UK hospitals; 243 participants were recruited, with 126 allocated to intervention and 117 to control, leaving 124 and 116 available for analysis after one loss to follow-up in each group.
- Participants were aged ≥16 years presenting with palpitations or pre-syncope and an underlying ECG rhythm that remained undiagnosed after ED assessment; intervention participants received standard care plus the AliveCor smartphone-based event recorder, while control participants received standard care alone, with 90-day follow-up via hospital records, GP records, telephone and questionnaire.
- Primary endpoint was symptomatic rhythm detection at 90 days; analysis was intention-to-treat using comparison of proportions, Fisher/χ² tests, Kaplan–Meier curves and log-rank tests, with p < 0.05; sample size was powered for 25% versus 10% detection (110 per group plus 10% dropout).
Limitations
- The paper states that a large proportion of recruitment occurred in office hours largely by research staff in research-active hospitals, and that the central ECG reading service used is not available in routine practice.
- The paper states that only one device was studied, that patients volunteering for a study of new technology may be more motivated to use the device, and that performance may not be as good in a non-study setting; it also notes that lack of smartphone ownership was the commonest reason for non-participation and that age of non-participants was not recorded.
- Evident from the text: the trial was open-label and the primary outcome was assessed by each local study team, which could introduce detection or reporting bias; the intervention group also had more subsequent ED presentations for palpitations/pre-syncope (12/124, 9.7% vs 3/116, 2.6%; p = 0.031), which the authors discuss as possibly reflecting lack of immediate reassurance.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 21 cited sources checked · 0 supported · 0 flagged
- Findingown data
Mean time to symptomatic rhythm detection was 9.5 days (SD 16.1) in the intervention group versus 42.9 days (SD 16.0) in the control group (p < 0.0001).
“Mean time to symptomatic rhythm detection in the intervention group was 9.5 days (SD 16.1, range 0–83) versus 42.9 days (SD 16.0, range 12–66; p < 0.0001) in the control group.”
Quote in the paperSubstantial claim - Findingown data
At 90 days, a symptomatic rhythm was detected in 55.6% (69/124) of intervention participants versus 9.5% (11/116) of control participants (RR 5.9, 95% CI 3.3-10.5; p < 0.0001).
“A symptomatic rhythm was detected at 90 days in 69 (n = 124; 55.6%; 95% CI 46.9–64.4%) participants in the intervention group versus 11 (n = 116; 9.5%; 95% CI 4.2–14.8) in the control group (RR 5.9, 95% CI 3.3–10.5; p < 0.0001).”
- Also stated in: § Interpretation · plain
“Use of a smartphone-based event recorder increased the number of patients in whom an ECG was captured during symptoms over five-fold to more than 55% at 90 days.”
- Also stated in: § Results · plain
“A symptomatic rhythm was detected at 90 days in 69 (n = 124; 55.6%; 95% CI 46.9–64.4%) participants in the intervention group versus 11 (n = 116; 9.5%; 95% CI 4.2–14.8) in the control group (RR 5.9, 95% CI 3.3–10.5; p < 0.0001).”
Quote in the paperSubstantial claim - Also stated in: § Interpretation · plain
- Findingown data
At 90 days, a symptomatic cardiac arrhythmia was detected in 8.9% (11/124) of intervention participants versus 0.9% (1/116) of control participants (RR 10.3, 95% CI 1.3-78.5; p = 0.006).
“A symptomatic cardiac arrhythmia was detected at 90 days in 11 (n = 124; 8.9%; 95% CI 3.9–13.9%) participants in the intervention group versus 1 (n = 116; 0.9%; 95% CI 0.0–2.5%) in the control group (RR 10.3, 95% CI 1.3–78.5; p = 0.006).”
- Also stated in: § Funding · plain
“The smartphone-based event recorder also increased the number of patients diagnosed with cardiac arrhythmia.”
- Also stated in: § Results · plain
“A symptomatic cardiac arrhythmia was detected at 90 days in 11 (n = 124; 8.9%; 95% CI 3.9–13.9%) participants in the intervention group versus 1 (n = 116; 0.9%; 95% CI 0.0–2.5%) in the control group (RR 10.3, 95% CI 1.3–78.5; p = 0.006).”
Quote in the paperSubstantial claim - Also stated in: § Funding · plain
- Findingnone given
The only adverse events recorded were those directly related to the use of the smartphone-based event recorder and application.
“The only adverse events recorded were those directly related to the use of the smartphone-based event recorder and application.”
Quote in the paperSubstantial claim - Findingown data
The mean time to symptomatic rhythm detection was 9.5 days (SD 16.1) in the intervention group versus 42.9 days (SD 16.0) in the standard care group (p < 0.0001).
“The mean time to symptomatic rhythm detection in the intervention group was 9.5 days (SD 16.1, range 0–83) versus 42.9 days (SD 16.0, range 12–66) in the standard care group (p < 0.0001).”
Quote in the paperSubstantial claim
- Findingown data
The commonest symptomatic rhythms detected were sinus rhythm (53 participants; 66.3%), sinus tachycardia (19; 23.8%) and ectopic beats (13; 16.3%).
“Commonest symptomatic rhythms detected were sinus rhythm (in 53 participants; 66.3%), sinus tachycardia (19; 23.8%) and ectopic beats (13; 16.3%).”
Quote in the paperNot a substantial claim - Findingown data
There was no difference between intervention and control groups in the number of GP presentations due to palpitations or pre-syncope (p = 0.312).
“number of GP presentations due to palpitations or pre-syncope (p = 0.312)”
Quote in the paperNot a substantial claim - Findingown data
There was no difference between intervention and control groups in the number of participants with one or more inpatient hospital days due to palpitations or pre-syncope (p > 0.999).
“There was no difference in the number of participants with one or more inpatient hospital days (over all admissions) due to palpitations or pre-syncope in the intervention group (2; n = 122; 2 patients with no data; 1.6%; 95% CI 0.0–3.8%) compared to the control group (1; n = 116; 0.9%; 95% CI 0.0–2.5%; p > 0.999)”
Quote in the paperNot a substantial claim - Findingown data
Cost per symptomatic rhythm diagnosis was £921 less per patient per symptomatic rhythm in the intervention group (£474) compared to the control group (£1395).
“Cost per symptomatic rhythm diagnosis was £921 less per patient per symptomatic rhythm in the intervention group (£474) compared to the control group (£1395).”
Quote in the paperSubstantial claim - Findingown data
There was no difference between intervention and control groups in the number of outpatient presentations due to palpitations or pre-syncope (p = 0.058).
“number of outpatient presentations due to palpitations or pre-syncope (p = 0.058)”
Quote in the paperNot a substantial claim - Findingown data
There was no difference between intervention and control groups in the number of ECGs performed due to palpitations or pre-syncope (p = 0.143).
“number of ECGs performed due to palpitations or pre-syncope (p = 0.143)”
Quote in the paperNot a substantial claim - Findingown data
Serious outcome at 90 days occurred in 11 (8.9%) intervention participants versus 2 (1.7%) control participants (p = 0.02).
“Serious outcome at 90 days in the intervention group was 11 (8.9%) versus 2 (1.7%) in the control group (p = 0.02).”
Quote in the paperSubstantial claim - Findingown data
There were more ED presentations after the index visit due to palpitations/pre-syncope in the intervention group (12/124; 9.7%) compared to the control group (3/116; 2.6%; p = 0.031).
“There were more ED presentations (after index visit) due to palpitations/pre-syncope in the intervention group (12/124; 9.7%; 95% CI 4.5–14.9% with 1 or more non index ED presentations) compared to the control group (3/116; 2.6%; 95% CI 0.0–5.5%; p = 0.031).”
Quote in the paperNot a substantial claim - Findingown data
The mean time to symptomatic cardiac arrhythmia detection was 9.9 days (SD 15.6) in the intervention group versus 48.0 days in the control group (p = 0.0004).
“The mean time to symptomatic cardiac arrhythmia detection in the intervention group was 9.9 days (SD 15.6, range 1–55) versus 48.0 days (1 participant) in the control group (p = 0.0004).”
Quote in the paperSubstantial claim - Findingown data
At 90 days, 12 intervention participants were subsequently undergoing or planning to undergo treatment for symptomatic cardiac arrhythmia versus 6 control participants (p = 0.192).
“At 90 days, 12 participants in the intervention group were subsequently undergoing (or planning to undergo) treatment for symptomatic cardiac arrhythmia versus 6 in the control group (p = 0.192).”
Quote in the paperNot a substantial claim - Findingown data
Median overall healthcare utilisation cost was £108 (IQR 99.0-246.50) in the intervention group versus £0 (IQR 0-120.0) in the standard care group (p = 0.0001).
“Median overall healthcare utilisation cost (primary/community/secondary care and intervention costs) in the intervention group was £108 (IQR 99.0–246.50, range 99–2697) versus £0 in the standard care group (IQR 0–120.0, range 0–4161; p = 0.0001).”
Quote in the paperSubstantial claim - Findingown data
The commonest reason for a patient not to be able to participate was that they did not possess a smartphone.
“We found that the commonest reason for a patient not to be able to participate was that they did not possess a smartphone.”
Quote in the paperNot a substantial claim - Premisecitation
Around 1 in 10 patients have a cardiac arrhythmia as their symptomatic rhythm.
“However around 1 in 10 patients do have a cardiac arrhythmia as their symptomatic rhythm [3].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3] Dimarco A.D. Onwordi E.N. Murphy C.F. Diagnostic utility of real-time smartphone ECG in the initial investigation of palpitations Brit J Cardio 2018
Not checked in the free run.
- Premisecitation
AliveCor has not yet been assessed against standard care for the broader investigation of palpitations and arrhythmia assessment.
“Whilst AliveCor has now undergone assessment against conventional care for AF detection [25], it has yet to be assessed against standard care for the broader investigation of palpitations and arrhythmia assessment.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [25] Halcox J.P.J. Wareham K. Cardew A. Assessment of remote heart rhythm sampling using the AliveCor heart monitor to screen for atrial fibrillation: the REHEARSE-AF study Circulation 136 2017 1784 1794 28851729 source
Not checked in the free run.
- Premisecitation
Once captured, the symptomatic rhythm underlying about 9 in 10 episodes is benign (e.g., normal sinus rhythm, sinus tachycardia or frequent ectopics).
“Once captured, the symptomatic rhythm underlying about 9 in 10 episodes is benign, e.g., normal sinus rhythm, sinus tachycardia or frequent ectopics (extra or skipped heart beats) [3].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [3] Dimarco A.D. Onwordi E.N. Murphy C.F. Diagnostic utility of real-time smartphone ECG in the initial investigation of palpitations Brit J Cardio 2018
Not checked in the free run.
- Premisecitation
There have been no studies in an acute or ED population investigating smartphone-based event recorders for palpitations and arrhythmia assessment.
“There have been no studies in an acute or ED population, where large numbers of patients present [2].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [2] Probst M.A. Mower W.R. Kanzaria H.K. Hoffman J.R. Buch E.F. Sun B.C. Analysis of emergency department visits for palpitations (from the National Hospital Ambulatory Medical Care Survey) Am J Cardiol 113 10 2014 1685 1690 24698469 PMC4011931 source
Not checked in the free run.
- Premisecitation
Recommended first-line investigation with 12-lead ECG and conventional ambulatory monitoring (Holter or event monitoring) are of limited efficacy due to the infrequency of symptoms in many patients.
“Recommended first line investigation of 12-lead ECG [4] and conventional ambulatory monitoring (Holter or event monitoring) [5], [6] are of limited efficacy due to the infrequency of symptoms in many patients [5], [7], [8], [9], [10].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [4] Blomstrom-Lundqvist C. Sheinman M.M. Aliot E.M. ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias – executive summary J Am Coll Cardiol 42 2003 1493 1531 14563598
Not checked in the free run.
- cites [5] Raviele A. Giada F. Bergfeldt L. Management of patients with palpitations: a position paper from the European Heart Rhythm Association Europace 13 2011 920 934 21697315 source
Not checked in the free run.
- cites [6] National Institute for Health and Care Excellence. NICE clinical knowledge summaries. Palpitations. London: NICE, May 2015. https://cks.nice.org.uk/palpitations#!topicsummary (accessed 16th August 2018).
Not checked in the free run.
- Interpretationreasoning
The detected symptomatic rhythms are clinically significant because they diagnose the underlying cause of the patient's symptoms.
“These are clinically significant rhythms as they diagnose the underlying cause of the patient's symptoms.”
Quote in the paperSubstantial claim - Interpretationnone given
The symptomatic rhythms detected are clinically significant because they diagnose the underlying cause of the patient's symptoms.
“These are clinically significant rhythms as they diagnose the underlying cause of the patient's symptoms.”
Quote in the paperSubstantial claim - Interpretationreasoning
In patients presenting with palpitations or near syncope, incorporating a patient-activated detection device into routine practice may overcome some of the current difficulties in diagnosis caused by the normalisation of cardiac rhythm by the time the patient undergoes a clinical assessment.
“In patients presenting with palpitations or near syncope the incorporation of a patient-activated detection device into routine practice may overcome some of the current difficulties in diagnosis caused by the normalisation of cardiac rhythm by the time the patient undergoes a clinical assessment.”
Quote in the paperSubstantial claim- Rests on an assumption: The normalisation of cardiac rhythm by the time the patient undergoes clinical assessment is a major cause of current difficulties in diagnosis.
- Premisecitation
In the Arrhythmia Alliance study, of 1500 people of all ages who were distributed AliveCor Heart Monitors, only one returned their monitor because it caused them to worry and check their heart rate too often.
“when the Arrhythmia Alliance [14] distributed AliveCor Heart Monitors to 1500 people of all ages, only one returned their monitor because it caused them to worry and check their heart rate too often.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] National Institute for Health and Care Excellence. AliveCor heart monitor and AliveECG app (Kardia Mobile) for detecting atrial fibrillation. NICE advice MIB35. London: NICE, 2015. https://www.nice.org.uk/advice/mib35/chapter/technology-overview (accessed 16th August 2018).
Not checked in the free run.
- Study design and conductcitation
The study had 80% power to detect an absolute 15 percentage point improvement in symptomatic rhythm detection at 90 days, assuming a 25% rate with the device versus 10% with standard care, with 110 participants per group.
“Using a symptomatic rhythm detection rate at 90 days of 25% [4] versus standard care (10%), we estimated that 110 participants in each group would have 80% power to determine an absolute 15 percentage point improvement in symptomatic rhythm detection.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [4] Blomstrom-Lundqvist C. Sheinman M.M. Aliot E.M. ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias – executive summary J Am Coll Cardiol 42 2003 1493 1531 14563598
Not checked in the free run.
- Interpretationcitation
Most symptomatic rhythms in patients with palpitations or pre-syncope are benign.
“This study confirms previous evidence that most symptomatic rhythms in patients with palpitations or pre-syncope are benign [28].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [28] Weber B.E. Kapoor W.H. Evaluations and outcomes of patients with palpitations Am J Med 100 1996 138 148 8629647 source
Not checked in the free run.
- Classified as probably not a substantial claim. 22
- May combine two claims. 20
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataThe commonest symptomatic rhythms detected were sinus rhythm, sinus tachycardia and ectopic beats.
“The commonest symptomatic rhythms detected were sinus rhythm, sinus tachycardia and ectopic beats.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSymptomatic cardiac arrhythmias detected were atrial fibrillation (8 intervention, 0 control), SVT (3 intervention, 0 control), sinus bradycardia (0 intervention, 1 control) and atrial flutter (1 intervention, 0 control).
“Symptomatic cardiac arrhythmias were AF (8 intervention, 0 standard care), SVT (3 intervention, 0 standard care), sinus bradycardia (0 intervention, 1 standard care) and atrial flutter (1 intervention, 0 standard care).”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataEighty of 92 (87.0%) participants found the AliveCor monitor easy to use.
“Eighty of 92 (87.0%) participants found the AliveCor monitor easy to use.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataAmong 80 participants with a symptomatic rhythm detected, 12 had a symptomatic cardiac arrhythmia and 68 had sinus rhythm, sinus tachycardia or ectopics.
“Eighty participants had a symptomatic rhythm detected with 12 of these having a symptomatic cardiac arrhythmia (atrial fibrillation or flutter, SVT and sinus bradycardia) and 68 having sinus rhythm, sinus tachycardia or ectopics [Table 4].”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThe only death in the study was in the intervention group in a participant known to have treated congenital structural heart disease, and the death was thought unrelated to his initial presentation to the ED.
“The only death in the study was in the intervention group in a participant known to have treated congenital structural heart disease whose death was thought unrelated to his initial presentation to the ED.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationIf patients are referred to cardiology services, investigation usually starts with a Holter monitor but non-compliance and lack of extended monitoring reduces diagnostic yield to less than 20%.
“If patients are referred to cardiology services for assessment, investigation usually starts with a Holter monitor but non-compliance and lack of extended monitoring reduces diagnostic yield to less than 20% [11].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [11] Cheung CC1, Kerr CR, Krahn AD. Comparing 14-day adhesive patch with 24-h Holter monitoring. Future Cardiol. 2014; 10(3): 319–22. 24976467 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe AliveCor device is supported for clinical use by a NICE technology appraisal.
“The device is supported for clinical use by a National Institute for Health and Care Excellence (NICE) technology appraisal [14]”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] National Institute for Health and Care Excellence. AliveCor heart monitor and AliveECG app (Kardia Mobile) for detecting atrial fibrillation. NICE advice MIB35. London: NICE, 2015. https://www.nice.org.uk/advice/mib35/chapter/technology-overview (accessed 16th August 2018).
Not checked in the free run.
May combine two claims.
PremisecitationThe AliveCor automatic diagnostic algorithm has a sensitivity of 96.6% and specificity of 94% for correctly interpreting AF versus normal sinus rhythm.
“the automatic diagnostic algorithm has excellent sensitivity (96.6%) and specificity (94%) for correctly interpreting AF versus normal sinus rhythm [15].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] William A.D. Kanbour M. Callahan T. Assessing the accuracy of an automated atrial fibrillation detection algorithm using smartphone technology: the iREAD study Heart Rhythm 2018 30143448 source
Not checked in the free run.
May combine two claims.
Interpretationown dataUse of a smartphone-based event recorder increases the symptom-rhythm correlation rate over five-fold at 90 days with a reduced cost per diagnosis.
“Use of a smartphone-based event recorder increases the symptom–rhythm correlation rate over five-fold at 90 days with a reduced cost per diagnosis.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductcitationThe trial was registered at ClinicalTrials.gov (NCT02783898) and the protocol was published.
“This trial was registered at ClinicalTrials.gov (trial registration number NCT02783898), and the protocol published in Trials [26].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [26] Reed MJ, Grubb NR, Lang CC, et al. Multi-centre randomised controlled trial of a smart phone based event recorder alongside standard care versus standard care for patients presenting to the Emergency Department with palpitations and pre-syncope - the IPED (Investigation of Palpitations in the ED) study: Study protocol for a randomised controlled trial. Trials 2018; 19: 711. nndoi: 10.1186/s13063-018-3098-1 PMC6311046 30594256 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataOne participant in each group was lost to follow-up, leaving 124 participants in the intervention group and 116 in the control group available for analysis.
“One participant in each group was lost to follow-up leaving 124 participants available for analysis in the study group and 116 in the control group.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataECGs were reviewed by the duty Consultant Emergency Physician and a trial team Emergency Physician, with central cardiology opinion in cases of disagreement.
“The duty Consultant Emergency Physician at the coordinating Edinburgh centre along with a trial team Emergency Physician reviewed the ECG.”
- Also stated in: § Procedures · plain
“In cases of disagreement, the central cardiology team were contacted for further opinion.”
Quote in the paperNot a substantial claim- Also stated in: § Procedures · plain
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationIn previous work in ED patients, 74% found it acceptable to use a smartphone to monitor their health, 79% to use a medical device connecting to a smartphone to monitor their health, and 77% reported that they would feel confident to use such technology.
“This reinforces data from previous work in ED patients which showed that 74% found it acceptable to use a smartphone to monitor their health, 79% to use a medical device connecting to a smartphone to monitor their health, and 77% reported that they would feel confident to use such technology [27].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [27] Summers A. Reed M.J. An evaluation of patient ownership and use and acceptability of smartphone technology within the emergency department Eur J Emerg Med 25 3 2018 224 225 29683883 source
Not checked in the free run.
May combine two claims.
Study design and conductown dataSecondary outcomes were not considered in the power calculation, increasing the risk of type I error.
“Statistical significance was determined as p < 0.05 for all outcomes with an acknowledgment of increased type I error risk in the secondary outcomes not considered in the power calculation.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataAllocation was concealed using sealed opaque envelopes prepared by a central administrator not involved in the study.
“Sealed opaque envelopes containing either ‘Standard Care plus Device’ or ‘Standard Care’ cards were prepared by a central administrator not involved in the study.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductown dataOne participant was removed from the study after randomisation because they did not meet the inclusion criteria.
“One participant was removed from the study by the local study team after being randomised, as they did not meet the inclusion criteria.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductreasoningRandomisation was by permuted block randomisation by site, with blocks randomly labelled using random number generation.
“randomisation was by permuted block randomisation by site.”
- Also stated in: § Randomisation · plain
“Blocks were randomly labelled using random number generation with site-specific study participation numbers and sent to each local study team.”
Quote in the paperNot a substantial claim- Also stated in: § Randomisation · plain
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Study design and conductnone givenThe sponsor deemed that a data monitoring committee was not required.
“The sponsor deemed that a data monitoring committee was not required.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductnone givenAll participants were analysed on an intention-to-treat basis.
“All participants were analysed on an intention to treat basis.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationIn the Arrhythmia Alliance study, older people were noted to be regular users of mobile technology and gave positive feedback about the system.
“in the Arrhythmia Alliance study, older people were noted to be regular users of mobile technology and gave positive feedback about the system [14].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [14] National Institute for Health and Care Excellence. AliveCor heart monitor and AliveECG app (Kardia Mobile) for detecting atrial fibrillation. NICE advice MIB35. London: NICE, 2015. https://www.nice.org.uk/advice/mib35/chapter/technology-overview (accessed 16th August 2018).
Not checked in the free run.
May combine two claims.
PremisecitationPreviously unexplained symptoms can cause anxiety and have a significant impact on quality of life even in the absence of underlying arrhythmia.
“Even in the absence of underlying arrhythmia, previously unexplained symptoms can cause anxiety and can have a significant impact on quality of life [29].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [29] Barsky A.J. Cleary P.D. Coeytaux R.R. The clinical course of palpitations in medical outpatients Arch Intern Med 155 1995 1782 1788 7654112 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenThe study used a central ECG reading service not available in routine practice.
“and the use of a central ECG reading service not available in routine practice.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone givenPatients choosing to take part in a study of new technology may be more motivated to use the device, and the device may not perform as well in a non-study setting.
“It is also possible that patients choosing to take part in a study of new technology may be more motivated to use the device, and it may not perform as well in a non-study setting.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenA large proportion of recruitment occurred in office hours largely by research staff in research active hospitals.
“Potential limitations of our study include a large proportion of recruitment occurring in office hours largely by research staff in research active hospitals”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Interpretationnone givenThe results of this study will be generalisable across many forms of patient-activated, symptom-based home ECG recording devices.
“We think that the results of this study will be generalisable across many forms of patient-activated, symptom-based home ECG recording devices.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
Study design and conductnone givenThe study's randomised, prospective design with systematic data collection is a key strength.
“The randomised, prospective design with systematic data collection is a key strength of the study.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationPrevious research shows 64% of those aged 50-75 and 30% over 75 years of age own a smartphone.
“our previous research shows 64% of those aged 50–75 and 30% over 75 years of age own a smartphone [27]”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [27] Summers A. Reed M.J. An evaluation of patient ownership and use and acceptability of smartphone technology within the emergency department Eur J Emerg Med 25 3 2018 224 225 29683883 source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing), S4 (based).
Interpretationnone givenSubtle differences in design or incorporation into clinical workflow may influence the effectiveness of patient-activated home ECG recording devices.
“Subtle differences in design or incorporation into clinical workflow may have an influence on effectiveness”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataThe AliveCor technology performs effectively and safely.
“This study suggests that the AliveCor technology performs effectively and safely.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationnone givenThe pragmatic design with potential variation in standard care between sites ensures the findings are generalisable across all types of standard care in the UK National Health Service without compromising validity.
“Whilst there was a potential variation in standard care between sites, this element of pragmatic design ensures our findings are generalisable across all types of standard care in the UK National Health Service without compromising validity.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataA smartphone-based event recorder is safe, non-invasive and easy to use.
“A safe, non-invasive and easy to use smartphone-based event recorder should be considered part of on-going care of all patients presenting acutely with unexplained palpitations or pre-syncope.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThis study demonstrates the ability of a smartphone-based event recorder to improve clinical care and patient experience for those suffering undiagnosed palpitations and pre-syncope.
“In summary, this study demonstrates the ability of a smartphone-based event recorder to improve clinical care and patient experience for those suffering undiagnosed palpitations and pre-syncope.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThese conditions account for up to one-fifth of all referrals to cardiology clinics.
“where currently these conditions account for up to one-fifth of all referrals) [30], [31], [32], [33]”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [30] Abbott A.V. Diagnostic approach to palpitations Am Fam Physician 71 2005 743 750 15742913
Not checked in the free run.
- cites [31] Messineo F.C. Ventricular ectopic activity: prevalence and risk Am J Cardiol 64 1989 53J 56J 2480710 source
Not checked in the free run.
- cites [32] Kroenke K. Arrington M.E. Mangelsdorff A.D. The prevalence of symptoms in medical outpatients and the adequacy of therapy Arch Intern Med 150 1990 1685 1689 2383163 source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenOnly one type of device was studied.
“Only one type of device was studied”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S4 (based).
Interpretationnone givenThese findings are likely to be generalisable from Emergency Medicine to General/Internal/Acute Medicine and General (Family) Practice in a broad range of developed healthcare systems.
“These findings are likely to be generalisable from Emergency Medicine to General/Internal/Acute Medicine and General (Family) Practice in a broad range of developed healthcare systems.”
Quote in the paperNot a substantial claim
The paper relies on this without support.
Diagnosis of the underlying rhythm is difficult because many patients are fully recovered by the time they are seen in the ED, and examination and presenting ECG are commonly normal.
“Diagnosis of the underlying rhythm is difficult as many patients are fully recovered by the time that they are seen in the ED. Examination and presenting electrocardiogram (ECG) are commonly normal.”
The paper relies on this without support.
The only way to establish the underlying heart rhythm is to capture an ECG whilst the patient has symptoms.
“The only way to establish the underlying heart rhythm is to capture an ECG whilst the patient has symptoms.”
The paper relies on this without support.
Many patients go for years without diagnosis due to the difficulty in capturing the underlying heart rhythm.
“Many patients go for years without diagnosis due to the difficulty in capturing the underlying heart rhythm.”
The paper relies on this without support.
Traditional event recorders, external continuous loop recorders and implantable loop recorders are expensive and not recommended for a patient group who rarely have malignant arrhythmias and may have prolonged periods between episodes.
“Traditional event recorders, external continuous loop recorders and implantable loop recorders are expensive and not recommended for a patient group who rarely have malignant arrhythmias and may have prolonged periods between episodes.”
The paper relies on this without support.
The healthcare system studied does not serve this patient group well.
“which in the healthcare system studied does not serve this patient group well.”
The paper relies on this without support.
The smartphone-based event recorder is easy to use.
“This safe, non-invasive and easy to use device should be considered part of on-going care to all patients presenting acutely with unexplained palpitations or pre-syncope.”
The paper relies on this without support.
The AliveCor app rhythm analysis algorithm automatically reports any ECG recorded as normal, atrial fibrillation or unclassified with excellent sensitivity and specificity.
“The AliveCor app rhythm analysis algorithm automatically reports any ECG recorded as normal, atrial fibrillation or unclassified with excellent sensitivity and specificity.”
The paper relies on this without support.
The normalisation of cardiac rhythm by the time the patient undergoes clinical assessment is a major cause of current difficulties in diagnosis.
“may overcome some of the current difficulties in diagnosis caused by the normalisation of cardiac rhythm by the time the patient undergoes a clinical assessment.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The IPED trial showed that a single-lead ECG monitor can improve outpatient arrhythmia detection in patients who present to the emergency room with palpitations or presyncope and no clear aetiology.
“The IPED trial58 showed that a single-lead ECG monitor can improve outpatient arrhythmia detection in patients who present to the emergency room with palpitations or presyncope and no clear aetiology”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [58]
In the IPED study, the mean time to symptomatic rhythm detection was 9.5 days in the intervention group compared with 42.9 days in the control group.
“The mean time to symptomatic rhythm detection in the intervention group was 9.5 days compared with 42.9 days in the control group58.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [58]
In the IPED study, at 90 days a symptomatic rhythm was detected in 55.6% of participants in the intervention group compared with 9.5% in the control group.
“At 90 days, a symptomatic rhythm was detected in 55.6% of participants in the intervention group compared with only 9.5% in the control group.”
From Smart wearable devices in cardiovascular care: Where we are and how to move forward (2021) · cites this paper as [58]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
The study death is assigned to opposite groups
MajorTable 4 places the only death in the control group, while the results text says it occurred in the intervention group. This reverses the group assignment of a serious safety outcome.
“| All cause death | 0 | 1 | 1 |”
“The only death in the study was in the intervention group in a participant known to have treated congenital structural heart disease whose death was thought unrelated to his initial presentation to the ED.”
Life-threatening arrhythmia total does not equal the group counts
MinorTable 4 reports zero life-threatening arrhythmias in each group but one in total.
Reported 1 · recomputed 0
“| Life-threatening arrhythmia[ventricular tachyarrhythmia, complete or 3rd degree heart block, second degree heart block type II, pause > 6 s, symptomatic bradycardia < 40 beats per minute] | 0 | 0 | 1 |”
Two clinician-rating percentages are overstated
MinorFor both ratings 4 and 5, the total is 32 of 237, or approximately 13.5%, rather than the reported 13.9%.
Reported 13.9 · recomputed 13.50211
“| ED clinician rating of likelihood of any underlying cardiac arrhythmia | 4 | 15 (n = 121) | 12.4 | 17 (n = 116) | 14.7 | 32 (n = 237) | 13.9 |”
Cites
Blomstrom-Lundqvist (2003) · doi:10.1161/01.cir.0000091380.04100.84 · Cited by 955 · see citing works
Blomström‐Lundqvist, C., Scheinman, M. M., Aliot, E., Alpert, J. S., Calkins, H. G., Camm, A. J., Campbell, W. B., Haines, D. E., Kück, K., Lerman, B. B., et al. (2003) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2003.08.013 · Cited by 681 · see citing works
Halcox, J. P. J., Wareham, K., Cardew, A., Gilmore, M., Barry, J. P., Phillips, C., & Gravenor, M. B. (2017) · Circulation · doi:10.1161/circulationaha.117.030583 · Cited by 580 · see citing works
Barrett, P. M., Komatireddy, R., Haaser, S., Topol, S., Sheard, J., Encinas, J., Fought, A. J., & Topol, E. J. (2014) · The American Journal of Medicine · doi:10.1016/j.amjmed.2013.10.003 · Cited by 498 · see citing works
Lowres, N., Neubeck, L., Salkeld, G., Krass, I., McLachlan, A. J., Redfern, J., Bennett, A. A., Briffa, T., Bauman, A., Martinez, C., et al. (2014) · Thrombosis and Haemostasis · doi:10.1160/th14-03-0231 · Cited by 458 · see citing works
The Prevalence of Symptoms in Medical Outpatients and the Adequacy of Therapy
Arrington, M. E. (1990) · Archives of Internal Medicine · doi:10.1001/archinte.1990.00040031685016 · Cited by 367 · see citing works
Lau, J. K., Lowres, N., Neubeck, L., Brieger, D. B., Sy, R. W., Galloway, C. D., Albert, D. E., & Freedman, S. B. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2013.01.220 · Cited by 326 · see citing works
Diagnostic utility of a novel leadless arrhythmia monitoring device.
Turakhia, M. P., Hoang, D. D., Zimetbaum, P., Miller, J. D., Froelicher, V. F., Kumar, U. N., Xu, X., Yang, F., & Heidenreich, P. A. (2013) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2013.04.017 · Cited by 258 · cited 1× · see citing works
“Recent technological advances have led to several novel ECG monitoring devices appearing on the market [12,13].”
BackgroundWireless smartphone ECG enables large-scale screening in diverse populations
HABERMAN, Z. C., JAHN, R. T., BOSE, R., TUN, H., SHINBANE, J. S., DOSHI, R. N., CHANG, P. M., & SAXON, L. A. (2015) · Journal of Cardiovascular Electrophysiology · doi:10.1111/jce.12634 · Cited by 212 · see citing works
Evaluation and outcomes of patients with palpitations
Weber, B. E., & Kapoor, W. N. (1996) · The American Journal of Medicine · doi:10.1016/s0002-9343(97)89451-x · Cited by 196 · see citing works
Desteghe, L., Raymaekers, Z., Lutin, M., Vijgen, J. M., Dilling-Boer, D. M., Koopman, P., Schurmans, J., Vanduynhoven, P., Dendale, P., & Heidbüchel, H. (2016) · EP Europace · doi:10.1093/europace/euw025 · Cited by 171 · see citing works
William, A. D., Kanbour, M., Callahan, T., Bhargava, M., Varma, N., Rickard, J., Saliba, W., Wolski, K., Hussein, A., Lindsay, B. D., et al. (2018) · Heart Rhythm · doi:10.1016/j.hrthm.2018.06.037 · Cited by 154 · cited 1× · see citing works
“6%) and specificity (94%) for correctly interpreting AF versus normal sinus rhythm [15].”
BackgroundRaviele, A., Giada, F., Bergfeldt, L., Blanc, J. J., Blomstrom-Lundqvist, C., Mont, L., Morgan, J. M., Raatikainen, M. J. P., Steinbeck, G., Viskin, S., et al. (2011) · Europace · doi:10.1093/europace/eur130 · Cited by 149 · see citing works
The Evolving Role of Ambulatory Arrhythmia Monitoring in General Clinical Practice
Zimetbaum, P. J., & Josephson, M. E. (1999) · Annals of Internal Medicine · doi:10.7326/0003-4819-130-10-199905180-00020 · Cited by 139 · see citing works
Tarakji, K. G., Wazni, O. M., Callahan, T. D., Kanj, M. H., Hakim, A. H., Wolski, K., Wilkoff, B. L., Saliba, W. I., & Lindsay, B. D. (2014) · Heart Rhythm · doi:10.1016/j.hrthm.2014.11.015 · Cited by 138 · see citing works
Zimetbaum, P. J., Kim, K. Y., Josephson, M. E., Goldberger, A. L., & Cohen, D. J. (1998) · Annals of Internal Medicine · doi:10.7326/0003-4819-128-11-199806010-00002 · Cited by 133 · see citing works
Diagnostic approach to palpitations.
Abbott, A. V. (2005) · PubMed · Cited by 83 · see citing works
Schreiber, D., Sattar, A., Drigalla, D., & Higgins, S. (2014) · Western Journal of Emergency Medicine · doi:10.5811/westjem.2013.11.18973 · Cited by 65 · see citing works
Ventricular ectopic activity: prevalence and risk
Messineo, F. C. (1989) · The American Journal of Cardiology · doi:10.1016/0002-9149(89)91200-9 · Cited by 64 · see citing works
Albert, D. J. (2016) · EP Europace · doi:10.1093/europace/euw218 · Cited by 64 · see citing works
The clinical course of palpitations in medical outpatients
Barsky, A. J. (1995) · Archives of Internal Medicine · doi:10.1001/archinte.1995.00430160124012 · Cited by 61 · see citing works
Comparing 14-day Adhesive Patch With 24-h Holter Monitoring
Cheung, C. C., Kerr, C. R., & Krahn, A. D. (2014) · Future Cardiology · doi:10.2217/fca.14.24 · Cited by 46 · see citing works
Outcomes in Presyncope Patients: A Prospective Cohort Study
Thiruganasambandamoorthy, V., Stiell, I. G., Wells, G. A., Vaidyanathan, A., Mukarram, M. A., & Taljaard, M. (2014) · Annals of Emergency Medicine · doi:10.1016/j.annemergmed.2014.07.452 · Cited by 45 · see citing works
Scherr, D., Dalal, D., Henrikson, C. A., Spragg, D. D., Berger, R. D., Calkins, H., & Cheng, A. (2008) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-008-9251-0 · Cited by 45 · see citing works
Probst, M. A., Mower, W. R., Kanzaria, H. K., Hoffman, J. R., Buch, E. F., & Sun, B. C. (2014) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2014.02.020 · Cited by 43 · see citing works
Macinnes, M., Martin, N., Fulton, H., & McLEOD, K. A. (2018) · Archives of Disease in Childhood · doi:10.1136/archdischild-2018-314901 · Cited by 34 · see citing works
Validation of a smartphone‐based event recorder for arrhythmia detection
Narasimha, D., Hanna, N., Beck, H., Chaskes, M. J., Glover, R., Gatewood, R. P., Bourji, M., Gudleski, G. D., Danzer, S., & Curtis, A. B. (2018) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13317 · Cited by 27 · see citing works
The natural history of palpitations in a family practice.
Knudson, M. P. (1987) · PubMed · Cited by 25 · see citing works
Reed, M. J., Grubb, N. R., Lang, C. C., O’Brien, R., Simpson, K., Padarenga, M., Grant, A., & Tuck, S. (2018) · Trials · doi:10.1186/s13063-018-3098-1 · Cited by 17 · see citing works
Self-citationDiagnostic utility of real-time smartphone ECG in the initial investigation of palpitations
(2018) · British Journal of Cardiology · doi:10.5837/bjc.2018.006 · Cited by 14 · see citing works
Mitchell, A. R. J., & Page, P. L. (2015) · BMJ Innovations · doi:10.1136/bmjinnov-2014-000029 · Cited by 13 · see citing works
Newham, W. G., & Tayebjee, M. H. (2017) · Journal of Atrial Fibrillation · doi:10.4022/jafib.1514 · Cited by 12 · see citing works
Summers, A. M., & Reed, M. J. (2018) · European Journal of Emergency Medicine · doi:10.1097/mej.0000000000000500 · Cited by 7 · cited 2× · see citing works
“This reinforces data from previous work in ED patients which showed that 74% found it acceptable to use a smartphone to monitor their health, 79% to use a medical device connecting to a smartphone to monitor their health, and 77% reported that theywould feel confident to use such technology [27].”
BackgroundMethodSelf-citation
Flagged first, then by citations
1 of the 34 references couldn’t be matched to a known paper.
Research works printed but not linked: 1
- Blomstrom-Lundqvist, 2003, ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias – executive summary, J Am Coll Cardiol, 42, 1493
Cited by
Sorted by citations
Smart wearable devices in cardiovascular care: Where we are and how to move forward
Bayoumy, K., Gaber, M., Elshafeey, A., Mhaimeed, O., Dineen, E. H., Marvel, F. A., Martin, S. S., Muse, E. D., Turakhia, M. P., Tarakji, K. G., et al. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-021-00522-7 · Cited by 829 · see citing works
How to use digital devices to detect and manage arrhythmias: an EHRA practical guide
Svennberg, E., Tjong, F., Goette, A., Akoum, N., Di Biase, L., Bordachar, P., Boriani, G., Burri, H., Conte, G., Deharo, J. C., et al. (2022) · Europace · doi:10.1093/europace/euac038 · Cited by 261 · see citing works
Artificial intelligence in the diagnosis and management of arrhythmias
Nagarajan, V. D., Su-Lin, L., Robertus, J. L., Nienaber, C. A., Trayanova, N. A., & Ernst, S. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab544 · Cited by 139 · see citing works
The Use of Wearable ECG Devices in the Clinical Setting: A Review
Kamga, P., Mostafa, R., & Zafar, S. (2022) · Current Emergency and Hospital Medicine Reports · doi:10.1007/s40138-022-00248-x · Cited by 94 · see citing works
Varma, N., Cygankiewicz, I., Turakhia, M. P., Heidbuchel, H., Hu, Y.-F., Chen, L. Y., Couderc, J.-P., Cronin, E. M., Estep, J. D., Grieten, L., et al. (2021) · Circulation: Arrhythmia and Electrophysiology · doi:10.1161/circep.120.009204 · Cited by 74 · see citing works
“…group, suggesting that at least in presyncope, patient-activated rhythm detection using a commercially available mHealth device is productive (Reed 2019).93 Rhythms reported by devices that rely on HRs will likely require validation with a medical-grade system to provide an ECG tracing…”
MethodVarma, N., Cygankiewicz, I., Turakhia, M. P., Heidbüchel, H., Hu, Y., Chen, L. Y., Couderc, J., Cronin, E. M., Estep, J. D., Grieten, L., et al. (2021) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12795 · Cited by 50 · see citing works
A review of smartphone point‐of‐care adapter design
Alawsi, T., & Mahdi, Z. F. (2019) · Engineering Reports · doi:10.1002/eng2.12039 · Cited by 46 · see citing works
Carrington, M., Providencia, R. A., Chahal, C. A. A., Ricci, F., Epstein, A. E., Gallina, S., Fedorowski, A., Sutton, R., & Khanji, M. Y. (2022) · European Heart Journal Open · doi:10.1093/ehjopen/oeac072 · Cited by 43 · see citing works
An electrocardiogram-based AI algorithm for early detection of pulmonary hypertension
DuBrock, H. M., Wagner, T. E., Carlson, K., Carpenter, C. L., Awasthi, S., Attia, Z. I., Frantz, R. P., Friedman, P. A., Kapa, S., Annis, J., et al. (2024) · European Respiratory Journal · doi:10.1183/13993003.00192-2024 · Cited by 38 · see citing works
Varma, N., Cygankiewicz, I., Turakhia, M. P., Heidbüchel, H., Hu, Y., Chen, L. Y., Couderc, J., Cronin, E. M., Estep, J. D., Grieten, L., et al. (2021) · Journal of Arrhythmia · doi:10.1002/joa3.12461 · Cited by 37 · see citing works
Impact of Mobile Health Devices for the Detection of Atrial Fibrillation: Systematic Review
Biersteker, T. E., Schalij, M. J., & Treskes, R. W. (2021) · JMIR mhealth and uhealth · doi:10.2196/26161 · Cited by 32 · see citing works
Diederichsen, S. Z., Xing, L. Y., Frodi, D. M., Kongebro, E. K., Haugan, K., Graff, C., Højberg, S., Krieger, D. W., Brandes, A., Køber, L., et al. (2023) · JAMA Cardiology · doi:10.1001/jamacardio.2022.5526 · Cited by 28 · see citing works
Leveraging electronic health records to streamline the conduct of cardiovascular clinical trials
Khan, M. S., Usman, M., Talha, K. M., Spall, H. G. C. V., Greene, S. J., Vaduganathan, M., Khan, S. S., Mills, N. L., Ali, Z. A., Mentz, R. J., et al. (2023) · European Heart Journal · doi:10.1093/eurheartj/ehad171 · Cited by 22 · see citing works
Unleashing the Potential for Patient-Generated Health Data (PGHD)
Nazi, K. M., Newton, T., & Armstrong, C. M. (2024) · Journal of General Internal Medicine · doi:10.1007/s11606-023-08461-4 · Cited by 20 · see citing works
Garikapati, K., Turnbull, S., Bennett, R. G., Campbell, T. G., Kanawati, J., Wong, M. S., Thomas, S. P., Chow, C. K., & Kumar, S. (2022) · Heart Lung and Circulation · doi:10.1016/j.hlc.2022.08.001 · Cited by 17 · see citing works
Varma, N., Cygankiewicz, I., Turakhia, M. P., Heidbuchel, H., Hu, Y., Chen, L. Y., Couderc, J.-P., Cronin, E. M., Estep, J. D., Grieten, L., et al. (2021) · Cardiovascular Digital Health Journal · doi:10.1016/j.cvdhj.2020.11.004 · Cited by 16 · see citing works
Krzowski, B., Skoczylas, K., Osak, G., Żurawska, N., Peller, M., Kołtowski, Ł., Zych, A., Główczyńska, R., Lodziński, P., Grabowski, M., et al. (2021) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztab040 · Cited by 15 · see citing works
QT Interval Monitoring with Handheld Heart Rhythm ECG Device in COVID-19 Patients
Minguito‐Carazo, C., Echarte‐Morales, J., Benito‐González, T., Castillo-García, S. D., Rodríguez-Santamarta, M., Sánchez-Muñoz, E., Maniega, C. G., García, R. B., Menéndez-Suárez, P., Gonzalez, S., et al. (2021) · Global Heart · doi:10.5334/gh.916 · Cited by 14 · see citing works
Enhancing Healthcare Access–Smartphone Apps in Arrhythmia Screening: Viewpoint
Książczyk, M., Dębska-Kozłowska, A., Warchoł, I., & Lubiński, A. (2021) · JMIR mhealth and uhealth · doi:10.2196/23425 · Cited by 14 · see citing works
Varma, N., Cygankiewicz, I., Turakhia, M. P., Heidbüchel, H., Hu, Y., Chen, L. Y., Couderc, J., Cronin, E. M., Estep, J. D., Grieten, L., et al. (2021) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztab001 · Cited by 16 · see citing works
Remote Monitoring in Cardiovascular Diseases
Pelter, M. N., Quer, G., & Pandit, J. A. (2023) · Current Cardiovascular Risk Reports · doi:10.1007/s12170-023-00726-1 · Cited by 10 · see citing works
Wearables in Sports Cardiology
Beavers, D. L., & Chung, E. H. (2022) · Clinics in Sports Medicine · doi:10.1016/j.csm.2022.02.004 · Cited by 7 · see citing works
Utility of a Handheld, Single-Lead ECG Device for Diagnosis of Cardiac Arrhythmias
Turnbull, S., Garikapati, K., Bennett, R. G., Campbell, T. G., Kotake, Y., Mahajan, R., Marschner, S. L., Byth, K., Chow, C. K., & Kumar, S. (2023) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2023.03.428 · Cited by 7 · see citing works
Introduction of Wearable Device in Cardiovascular Field for Monitoring Arrhythmia
Kim, N. H., & Ko, J. S. (2021) · Chonnam Medical Journal · doi:10.4068/cmj.2021.57.1.1 · Cited by 7 · see citing works
The Democratization of Diagnosis: Bringing the Power of Medical Diagnosis to the Masses
Bose, R., & Saxon, L. A. (2019) · EClinicalMedicine · doi:10.1016/j.eclinm.2019.03.004 · Cited by 6 · see citing works
Iqhrammullah, M., Abdullah, A., Hermansyah, Ichwansyah, F., Rani, H. A., Alina, M., Simanjuntak, A. M. T., Rampengan, D. D. C. H., Al‐Gunaid, S. T., Gusti, N., et al. (2025) · Journal of Arrhythmia · doi:10.1002/joa3.70087 · Cited by 5 · see citing works
The use of digital health in heart rhythm care
Tchapmi, D. P., Agyingi, C., Egbe, A., Marcus, G. M., & Noubiap, J. J. (2023) · Expert Review of Cardiovascular Therapy · doi:10.1080/14779072.2023.2226868 · Cited by 5 · see citing works
Turnbull, S., Garikapati, K., Bennett, R. G., Campbell, T. G., Kotake, Y., Silva, K. D., Mahajan, R., Wong, M. S., Kazi, S., Marschner, S. L., et al. (2024) · Heart Lung and Circulation · doi:10.1016/j.hlc.2024.05.008 · Cited by 5 · see citing works
Sex and gender reporting in UK emergency medicine trials from 2010 to 2023: a systematic review
Astin-Chamberlain, R., Pott, J., Cole, E. M., & Bloom, B. (2024) · Emergency Medicine Journal · doi:10.1136/emermed-2024-214054 · Cited by 5 · see citing works
Arrhythmia in an athlete diagnosed by smartphone electrocardiogram: a case report
PHILLIPS, D. I. W., O’Callaghan, P., & Zaidi, A. (2021) · European Heart Journal - Case Reports · doi:10.1093/ehjcr/ytab186 · Cited by 5 · see citing works
Varma, N., Cygankiewicz, I., & Turakhia, M. P. (2021) · Russian Journal of Cardiology · doi:10.15829/1560-4071-2021-4420 · Cited by 6 · see citing works
Cullen, J., Reed, M. J., Muir, A., Murphy, R., Pollard, V., Zangana, G. A. S., Krupej, S., Askham, S., Holdsworth, P., & Davies, L. (2021) · European Journal of Emergency Medicine · doi:10.1097/mej.0000000000000859 · Cited by 4 · see citing works
Self-citationWever, W. D., Gruwez, H., Dhont, S., Pison, L., Vandervoort, P. M., & Haemers, P. (2025) · European Heart Journal Acute Cardiovascular Care · doi:10.1093/ehjacc/zuaf060 · Cited by 4 · see citing works
Reed, M. J., Muir, A., Cullen, J., Murphy, R., Pollard, V., Zangana, G., Krupej, S., Askham, S., Holdsworth, P., & Davies, L. (2021) · Medicina · doi:10.3390/medicina57020147 · Cited by 4 · see citing works
Self-citationHashim, U. N., Salahuddin, L. B., Ikram, R. R. R., Rabaah, U., Choon, N. H., & Naim, M. H. (2021) · International Journal of Advanced Computer Science and Applications · doi:10.14569/ijacsa.2021.0120120 · Cited by 4 · see citing works
Wearing Your Heart on Your Wrist
See, V. Y., & Kwong, H. J. (2020) · JACC Case Reports · doi:10.1016/j.jaccas.2019.12.037 · Cited by 4 · see citing works
Makowska, A., Ananthakrishnan, G., Christ, M., & Dehmer, M. (2025) · Healthcare · doi:10.3390/healthcare13040408 · Cited by 3 · see citing works
Leńska‐Mieciek, M., Kuls-Oszmaniec, A., Dociak, N., Kowalewski, M., Sarwiński, K., Osiecki, A., & Fiszer, U. (2022) · Journal of Clinical Medicine · doi:10.3390/jcm11030665 · Cited by 3 · see citing works
Clinical Validation of a Smartphone-based Handheld ECG Device: A Validation Study
Ahmadi‐Renani, S., Gharebaghi, M., kamalian, E., Hajghassem, H., Ghanbari, A., Karimi, A., Mansoury, B., Dayari, M. S., Nemati, M. K., Karimi, A., et al. (2022) · Critical Pathways in Cardiology A Journal of Evidence-Based Medicine · doi:10.1097/hpc.0000000000000303 · Cited by 3 · see citing works
Abnormal heart rate detection through real-time heart monitoring application
Hashim, U. N., Salahuddin, L. B., Hashim, U. R., Naim, M. H., Ikram, R. R. R., & Rahim, F. A. (2023) · Bulletin of Electrical Engineering and Informatics · doi:10.11591/beei.v12i4.4933 · Cited by 3 · see citing works
Use of Smartphones and Wearables for Arrhythmia Monitoring
Sanders, D. J., Wasserlauf, J., & PASSMAN, R. S. (2021) · Cardiac Electrophysiology Clinics · doi:10.1016/j.ccep.2021.04.004 · Cited by 3 · see citing works
Ojha, U., Ayathamattam, J., Okonkwo, K., & Ogunmwonyi, I. (2022) · Current Cardiology Reviews · doi:10.2174/1573403x18666220421110935 · Cited by 2 · see citing works
Utilidad de un registrador de eventos vinculado al teléfono móvil de pacientes con palpitaciones
Alcaraz, M. E. S., Santiago, F. B., Blanco, P. J. F., Alberola, A. G., Pascual-Figal, D. A., & Manzano‐Fernández, S. (2021) · Revista Española de Cardiología · doi:10.1016/j.recesp.2021.05.006 · Cited by 1 · see citing works
Usefulness of a smartphone-based event recorder for ambulatory patients with palpitations
Alcaraz, M. E. S., Santiago, F. B., Flores‐Blanco, P. J., Alberola, A. G., Pascual-Figal, D. A., & Manzano‐Fernández, S. (2021) · Revista Española de Cardiología (English Edition) · doi:10.1016/j.rec.2021.05.014 · Cited by 1 · see citing works
Williams, J., Mills, M. T., & Warriner, D. (2021) · British Journal of Hospital Medicine · doi:10.12968/hmed.2021.0419 · Cited by 1 · see citing works
Rafiq, Q., Christie, L., & Morgan, H. M. (2023) · medRxiv · doi:10.1101/2023.02.16.23286022 · Cited by 1 · see citing works
Chest Pain, Dyspnea, Palpitations
Hanna, E. B. (2022) · Wiley eBooks · doi:10.1002/9781119832737.ch27
Probst, M. A., & Carpenter, C. R. (2023) · Wiley eBooks · doi:10.1002/9781119616870.ch22
Contemporary wearable and handheld technology for the diagnosis of cardiac arrhythmias in Singapore
Chia, P., Tan, K. W. K., Ng, S., & Foo, D. C. G. (2023) · Singapore Medical Journal · doi:10.4103/singaporemedj.smj-2023-048
Kumar, S., Turnbull, S., Garikapati, K., Bennett, R. G., Campbell, T. G., Kotake, Y., Silva, K. D., Mahajan, R., Marschner, S. L., & Chow, C. K. (2023) · doi:10.22541/au.169581253.32737151/v1
Holter ECG Monitoring in Assessing Arrhythmias in Patients with Palpitation
Pandey, B., Karki, P., Pandey, N. K., Ghimire, J. P., Gyawali, A., Pandey, S., Gupta, P., Shrestha, B., Pandey, M., Shrestha, S. B., et al. (2023) · Journal of Advances in Internal Medicine · doi:10.3126/jaim.v12i2.62076
Wang, H., Mi, L., Zhang, Y., Ge, L., Lai, J. C. Y., Chen, T., Li, J., Shi, X., Xiu, J., Tang, M., et al. (2024) · PubMed Central · doi:10.12122/j.issn.1673-4254.2024.05.06
González, N. T., Álvarez-Acosta, L., Valdivia-Miranda, D., Gómez, D. D., & Afonso, J. S. H. (2024) · REC CardioClinics · doi:10.1016/j.rccl.2024.12.001
Matta, S., Mikhail, G., Ugona, D., Lazarescu, R., & Taffaro, A. (2025) · Cureus · doi:10.7759/cureus.98658
Rogers, R., Griffiths, C., Walker, K., Baukaite, E., & Kelbrick, M. (2026) · Open Journal of Psychiatry · doi:10.4236/ojpsych.2026.161003
Research in Emergency Medicine Is Not Optional - Why We Must Generate Our Own Evidence
Reed, M. J. (2026) · British Journal of Hospital Medicine · doi:10.31083/bjhm54690
Self-citationPhotoplethysmography for Heart Rate and Rhythm Monitoring
Wever, M. D., Nandan, A. R., Svennberg, E., Duncker, D., Linz, D., Mittal, S., Haemers, P., Pison, L., & PASSMAN, R. S. (2026) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2026.06.034 · Cited by 1 · see citing works
Wearable detection of atrial fibrillation: Absence of evidence or evidence for adoption?
Kwan, T. N., & Kumar, S. (2026) · Trends in Cardiovascular Medicine · doi:10.1016/j.tcm.2026.08.007
- Authors
- Reed, M. J., Grubb, N. R., Lang, C. C., O'Brien, R., Simpson, K., Padarenga, M., Grant, A., Tuck, S., Keating, L., Coffey, F., Jones, L., Harris, T., Lloyd, G., Gagg, J., Smith, J. E., & Coats, T.
- Affiliations
- Edinburgh Royal Infirmary (GB)
- Western General Hospital (GB)
- Royal Berkshire Hospital (GB)
- Nottingham University Hospitals NHS Trust (GB)
- Chesterfield Royal Hospital NHS Foundation Trust (GB)
- Barts Health NHS Trust (GB)
- Royal Devon and Exeter Hospital (GB)
- Musgrove Park Hospital (GB)
- University Hospitals Plymouth NHS Trust (GB)
- University of Leicester (GB)
- Corresponding author
- Reed, Matthew J.
- Grubb, Neil R.
- Lang, Christopher C.
- Simpson, Kirsty
- Padarenga, Mia
- Grant, Alison
- Tuck, Sharon
- Keating, Liza
- Coffey, Frank
- Jones, Lucy
- Harris, Tim
- Lloyd, Gavin
- Gagg, James
- Smith, Jason E.
- Coats, Tim
- ORCID
- Venue
- EClinicalMedicine
- Publisher
- Elsevier BV
- Published
- 2019-02
- Volume
- 8
- Pages
- 37-46
- ISSN
- 2589-5370 (ISSN-L)
- PMC
- PMC6537555
- OpenAlex
- W2920512760
- Semantic Scholar
- 0d529dcbe3f3cabefeac4fc7646452b71ab3cb18
- Cited by
- 57
- References
- 34
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 6 · 2023: 11 · 2022: 8 · 2021: 21 · 2020: 1 · 2019: 2
- FWCI
- 7.64
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- EClinicalMedicine (publishedVersion, cc-by-nc-nd)
- SSRN Electronic Journal (acceptedVersion)
- PubMed Central (publishedVersion, cc-by-nc-nd)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Figshare (submittedVersion, cc-by-nc-nd)
- Leicester Research Archive (University of Leicester) (submittedVersion, cc-by-nc-nd)
- Edinburgh Research Explorer (University of Edinburgh) (submittedVersion, cc-by-nc-nd)
- Royal Devon and Exeter Research Repository (Royal Devon and Exeter) (submittedVersion, cc-by-nc-nd)
- Funders
- Chest, Heart and Stroke Association Scotland (23,056, R15/A164)
- British Heart Foundation
- Topics
- Health Sciences › Medicine › Surgery › Cardiovascular Syncope and Autonomic Disorders
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac Arrhythmias and Treatments
- Keywords
- palpitations, presyncope, AliveCor, emergency department, arrhythmia detection, rhythm detection, cardiac implantable electronic devices, implantable cardioverter-defibrillator, cardiac resynchronization therapy, arrhythmic risk stratification
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review