Improved in Vivo Performance of Amperometric Oxygen ( P O2) Sensing Catheters via Electrochemical Nitric Oxide Generation/Release
Open accessAbstract
A novel electrochemically controlled release method for nitric oxide (NO) (based on electrochemical reduction of nitrite ions) is combined with an amperometric oxygen sensor within a dual lumen catheter configuration for the continuous in vivo sensing of the partial pressure of oxygen (PO2) in blood. The on-demand electrochemical NO generation/release method is shown to be fully compatible with amperometric PO2 sensing. The performance of the sensors is evaluated in rabbit veins and pig arteries for 7 and 21 h, respectively. Overall, the NO releasing sensors measure both venous and arterial PO2 values more accurately with an average deviation of −2 ± 11% and good correlation (R2 = 0.97) with in vitro blood measurements, whereas the corresponding control sensors without NO release show an average deviation of −31 ± 28% and poor correlation (R2 = 0.43) at time points >4 h after implantation in veins and >6 h in arteries. The NO releasing sensors induce less thrombus formation on the catheter surface in both veins and arteries (p < 0.05). This electrochemical NO generation/release method could offer a new and attractive means to improve the biocompatibility and performance of implantable chemical sensors.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Applying different potentials on electrodes can modulate the NO generation rate, offering a steady, controllable, physiologically relevant NO flux compared with chemically based NO generation.
“Applying different potentials on the electrodes can modulate the rate of NO generation, which offers a steady, controllable, and physiologically relevant flux of NO, compared to the chemically-based NO generation.”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [258]
Sensors with electrochemical NO generation provided very accurate oxygen responses in rabbit veins and pig arteries.
“In both cases, the sensors with electrochemical NO generation provided very accurate oxygen responses, while control sensors deviated from the real values by 30–40 % after 5 h of implantation because the local oxygen was consumed by the activated platelets and other cells trapped in the thrombus formed on the surface of the control sensors.”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [258]
This electrochemical approach has been used to develop multi-lumen catheters with one lumen dedicated to generating NO electrochemically to reduce thrombus and microbial biofilm formation on catheter surfaces.
“This approach has been utilized to develop a new generation of multi-lumen catheters in which one lumen is dedicated to generating NO electrochemically to reduce thrombus and microbial biofilm formation on the surface of the catheters.147,258”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [258]
The control sensor deviation was because local oxygen was consumed by activated platelets and other cells trapped in thrombus on the control sensor surfaces.
“In both cases, the sensors with electrochemical NO generation provided very accurate oxygen responses, while control sensors deviated from the real values by 30–40 % after 5 h of implantation because the local oxygen was consumed by the activated platelets and other cells trapped in the thrombus formed on the surface of the control sensors.”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [258]
Control sensors deviated from real oxygen values by 30-40% after 5 h of implantation.
“In both cases, the sensors with electrochemical NO generation provided very accurate oxygen responses, while control sensors deviated from the real values by 30–40 % after 5 h of implantation because the local oxygen was consumed by the activated platelets and other cells trapped in the thrombus formed on the surface of the control sensors.”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [258]
Ren et al. demonstrated a dual lumen catheter combining an amperometric oxygen sensor with electrochemical NO generation using Cu(II)TPMA as mediator for nitrite reduction to NO.
“In a separate study, Ren et al. recently demonstrated the use of a dual lumen catheter configuration to combine an amperometric oxygen sensor (in one lumen) with the electrochemical NO generation in the other lumen using Cu(II)TPMA as the mediator for reduction of nitrite ions to NO258 (see Figure 8).”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [258]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
NIH Image to ImageJ: 25 years of image analysis
Schneider, C. A., Rasband, W. S., & Eliceiri, K. W. (2012) · Nature Methods · doi:10.1038/nmeth.2089 · Cited by 64,638 · see citing works
Nitric oxide synthases: regulation and function
Forstermann, U., & Sessa, W. C. (2012) · European Heart Journal · doi:10.1093/eurheartj/ehr304 · Cited by 4,409 · see citing works
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559 · see citing works
Nitric oxide release: part III. Measurement and reporting.
Coneski, P. N., & Schoenfisch, M. (2012) · Chemical Society Reviews · doi:10.1039/c2cs15271a · Cited by 1,023 · see citing works
Serum Lactate as a Predictor of Mortality in Emergency Department Patients with Infection
Shapiro, N. I., Howell, M. D., Talmor, D. S., Nathanson, L. A., Lisbon, A., Wolfe, R. E., & Weiss, J. W. (2005) · Annals of Emergency Medicine · doi:10.1016/j.annemergmed.2004.12.006 · Cited by 750 · see citing works
Biosensors for real-time in vivo measurements
Wilson, G. S., & Gifford, R. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.003 · Cited by 660 · see citing works
The potential of nitric oxide releasing therapies as antimicrobial agents
Schairer, D. O., Chouake, J. S., Nosanchuk, J. D., & Friedman, A. (2012) · Virulence · doi:10.4161/viru.20328 · Cited by 570 · see citing works
The Chemistry of S-Nitrosothiols
Williams, D. L. H. (1999) · Accounts of Chemical Research · doi:10.1021/ar9800439 · Cited by 499 · see citing works
Molecular basis of biomaterial-mediated foreign body reactions
Hu, W.-J., Eaton, J. W., Ugarova, T. P., & Tang, L. (2001) · Blood · doi:10.1182/blood.v98.4.1231 · Cited by 431 · see citing works
Estimation of nitric oxide production and reactionrates in tissue by use of a mathematical model
Vaughn, M. W., Kuo, L., & Liao, J. C. (1998) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1998.274.6.h2163 · Cited by 338 · see citing works
Kinetics of the Reaction of Nitric Oxide with Oxygen in Aqueous Solutions
Lewis, R. S., & Deen, W. M. (1994) · Chemical Research in Toxicology · doi:10.1021/tx00040a013 · Cited by 293 · see citing works
Wang, G., Zhu, Y., Halushka, P. V., Lincoln, T., & Mendelsohn, M. E. (1998) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.95.9.4888 · Cited by 271 · see citing works
Diffusivity and Solubility of Nitric Oxide in Water and Saline
Zacharia, I. G., & Deen, W. M. (2005) · Annals of Biomedical Engineering · doi:10.1007/s10439-005-8980-9 · Cited by 252 · see citing works
Oxygen Therapy in Critical Illness
Martín, D., & Grocott, M. P. W. (2012) · Critical Care Medicine · doi:10.1097/ccm.0b013e31826a44f6 · Cited by 227 · see citing works
Progress Toward Clinical Application of the Nitric Oxide–Releasing Diazeniumdiolates
Keefer, L. K. (2003) · The Annual Review of Pharmacology and Toxicology · doi:10.1146/annurev.pharmtox.43.100901.135831 · Cited by 189 · see citing works
The role of venous blood gas in the Emergency Department
Bloom, B., Gründlingh, J., Bestwick, J. P., & Harris, T. (2013) · European Journal of Emergency Medicine · doi:10.1097/mej.0b013e32836437cf · Cited by 157 · see citing works
Schoenfisch, M. H., Mowery, K. A., Rader, M. V., Baliga, N., Wahr, J. A., & Meyerhoff, M. E. (2000) · Analytical Chemistry · doi:10.1021/ac991370c · Cited by 121 · see citing works
Self-citationFrost, M. C., Rudich, S. M., Zhang, H., Maraschio, M., & Meyerhoff, M. E. (2002) · Analytical Chemistry · doi:10.1021/ac025944g · Cited by 93 · see citing works
Self-citationBurns, K. E., & McLaren, A. (2009) · Canadian Respiratory Journal · doi:10.1155/2009/751507 · Cited by 83 · see citing works
Ren, H., Wu, J., Xi, C., Lehnert, N., Major, T., Bartlett, R. H., & Meyerhoff, M. E. (2014) · ACS Applied Materials & Interfaces · doi:10.1021/am406066a · Cited by 69 · see citing works
Self-citationGanter, M. T., & Zollinger, A. (2003) · British Journal of Anaesthesia · doi:10.1093/bja/aeg176 · Cited by 67 · see citing works
Improving the biocompatibility of in vivo sensors via nitric oxide release
Shin, J. H., & Schoenfisch, M. H. (2006) · The Analyst · doi:10.1039/b600129g · Cited by 66 · see citing works
WU, Y., ROJAS, A., GRIFFITH, G., SKRZYPCHAK, A., LAFAYETTE, N., BARTLETT, R., & MEYERHOFF, M. (2007) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2006.09.025 · Cited by 65 · see citing works
Self-citationWu, Y., & Meyerhoff, M. E. (2008) · Talanta · doi:10.1016/j.talanta.2007.06.022 · Cited by 65 · see citing works
Self-citationThromboresistant/anti-biofilm catheters via electrochemically modulated nitric oxide release
Ren, H., Colletta, A., Koley, D., Wu, J., Xi, C., Major, T. C., Bartlett, R. H., & Meyerhoff, M. E. (2015) · Bioelectrochemistry · doi:10.1016/j.bioelechem.2014.12.003 · Cited by 55 · see citing works
Self-citationWu, Y., Rojas-Peña, Á., Griffith, G. W., et al. (2007) · PubMed · doi:10.1016/j.snb.2006.09.02 · Cited by 46 · see citing works
Soller, B. R., Idwasi, P. O., Balaguer, J., Levin, S., Simsir, S. A., Salm, T. J. V., Collette, H., & Heard, S. O. (2003) · Critical Care Medicine · doi:10.1097/01.ccm.0000086999.21673.6a · Cited by 42 · see citing works
Yan, Q., Major, T. C., Bartlett, R. H., & Meyerhoff, M. E. (2011) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2011.04.026 · Cited by 41 · see citing works
Self-citationSol–gel derived nitric oxide-releasing oxygen sensors
Marxer, S. M., Robbins, M. E., & Schoenfisch, M. H. (2005) · The Analyst · doi:10.1039/b412079e · Cited by 23 · see citing works
Frost, M. C., Wolf, A. K., & Meyerhoff, M. E. (2013) · Royal Society of Chemistry eBooks · doi:10.1039/9781849737302-00129 · Cited by 6 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- Frost M. C.; Wolf A. K.; Meyerhoff M. E. In Detection Challenges in Clinical Diagnostics; The Royal Society of Chemistry: Cambridge, UK, 2013; pp 129– 155.
Cited by
Sorted by citations
Biosensor nanoengineering: Design, operation, and implementation for biomolecular analysis
Purohit, B., Vernekar, P. R., Shetti, N. P., & Chandra, P. (2020) · Sensors International · doi:10.1016/j.sintl.2020.100040 · Cited by 425 · see citing works
Wo, Y., Brisbois, E. J., Bartlett, R. H., & Meyerhoff, M. E. (2016) · Biomaterials Science · doi:10.1039/c6bm00271d · Cited by 251 · see citing works
Self-citationRecent advances in electrocatalytic nitrite reduction
Zhang, X., Wang, Y., Wang, Y., Guo, Y., Xie, X., Yu, Y., & Zhang, B. (2022) · Chemical Communications · doi:10.1039/d1cc06690k · Cited by 184 · see citing works
A Review of Recent Advancements in Sensor‐Integrated Medical Tools
Park, J., Seo, B., Jeong, Y., & Park, I. (2024) · Advanced Science · doi:10.1002/advs.202307427 · Cited by 89 · see citing works
Brisbois, E. J., Major, T. C., Goudie, M. J., Bartlett, R. H., Meyerhoff, M. E., & Handa, H. (2016) · Acta Biomaterialia · doi:10.1016/j.actbio.2016.04.025 · Cited by 87 · see citing works
Self-citationIn Vivo Chemical Sensors: Role of Biocompatibility on Performance and Utility
Soto, R. J., Hall, J. R., Brown, M. D., Taylor, J. B., & Schoenfisch, M. H. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b04251 · Cited by 85 · see citing works
Hopkins, S. P., Pant, J., Goudie, M. J., Schmiedt, C. W., & Handa, H. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b08647 · Cited by 85 · see citing works
Electrophoretic deposition: a versatile tool against biomaterial associated infections
Bakhshandeh, S., & Yavari, S. A. (2018) · Journal of Materials Chemistry B · doi:10.1039/c7tb02445b · Cited by 69 · see citing works
Can Calibration-Free Sensors Be Realized?
Bakker, E. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00247 · Cited by 63 · see citing works
BioMEMS for biosensors and closed-loop drug delivery
Coffel, J., & Nuxoll, E. E. (2018) · International Journal of Pharmaceutics · doi:10.1016/j.ijpharm.2018.01.030 · Cited by 44 · see citing works
Konopińska, K. K., Schmidt, N. J., Hunt, A. P., Lehnert, N., Wu, J., Xi, C., & Meyerhoff, M. E. (2018) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.8b05917 · Cited by 33 · see citing works
Self-citationVadgama, P. M. (2020) · Sensors · doi:10.3390/s20113149 · Cited by 33 · see citing works
Ren, H., Bull, J. L., & Meyerhoff, M. E. (2016) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.6b00215 · Cited by 31 · see citing works
Self-citationAssessing and improving the biocompatibility of microfluidic artificial lungs
Thompson, A. J., Lindsay, J. T., Major, T. C., Jeakle, M. M. P., Lautner‐Csorba, O., Goudie, M. J., Handa, H., Rojas-Peña, Á., & Potkay, J. A. (2020) · Acta Biomaterialia · doi:10.1016/j.actbio.2020.05.008 · Cited by 27 · see citing works
Self-citationNitric oxide release for improving performance of implantable chemical sensors – A review
Cha, K. H., Wang, X., & Meyerhoff, M. E. (2017) · Applied Materials Today · doi:10.1016/j.apmt.2017.10.002 · Cited by 27 · see citing works
Self-citationHighly Efficient Electrosynthesis of Nitric Oxide for Biomedical Applications
Jin, J., Mao, J., Wu, W., Jiang, Y. Y., Ma, W., Yu, P., & Mao, L. (2022) · Angewandte Chemie International Edition · doi:10.1002/anie.202210980 · Cited by 26 · see citing works
Tang, L., Wu, P., Zhuang, H., Qin, Z., Yu, P., Fu, K., Qiu, P., Liu, Y., & Zhou, Y. (2023) · International Journal of Biological Macromolecules · doi:10.1016/j.ijbiomac.2023.124564 · Cited by 24 · see citing works
Jin, S., Huang, J., Chen, X., Gu, H. S., Li, D., Zhang, A., Liu, X., & Chen, H. (2019) · ACS Applied Bio Materials · doi:10.1021/acsabm.9b00969 · Cited by 22 · see citing works
Wang, X., Garcia, C. T., Gong, G., Wishnok, J. S., & Tannenbaum, S. R. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b04049 · Cited by 18 · see citing works
Cha, K. H., & Meyerhoff, M. E. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00430 · Cited by 16 · see citing works
Self-citationMcCabe, M., Hála, P., Rojas-Peña, Á., Lautner‐Csorba, O., Major, T. C., Ren, H., Bartlett, R. H., Brisbois, E. J., & Meyerhoff, M. E. (2019) · Talanta · doi:10.1016/j.talanta.2019.06.077 · Cited by 12 · see citing works
Self-citationZhang, Q., & Meyerhoff, M. E. (2021) · Electroanalysis · doi:10.1002/elan.202100174 · Cited by 11 · see citing works
Self-citationZhang, Q., Murray, G. P., Hill, J. E., Harvey, S. L., Rojas-Peña, Á., Choi, J. D., Zhou, Y., Bartlett, R. H., & Meyerhoff, M. E. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c02979 · Cited by 11 · see citing works
Self-citationElectrochemical generation of nitric oxide for medical applications
White, C. J., Lehnert, N., & Meyerhoff, M. E. (2021) · Electrochemical Science Advances · doi:10.1002/elsa.202100156 · Cited by 10 · see citing works
Self-citationLong–term Drifts in Sensitivity Caused by Biofouling of an Amperometric Oxygen Sensor
Patrick, M., Grillot, J. M., Derden, Z. M., & Paul, D. W. (2016) · Electroanalysis · doi:10.1002/elan.201600653 · Cited by 10 · see citing works
Zhang, Q., Stachelek, S. J., Inamdar, V. V., Alferiev, I. S., Nagaswami, C., Weisel, J. W., Hwang, J.-H., & Meyerhoff, M. E. (2020) · Colloids and Surfaces B Biointerfaces · doi:10.1016/j.colsurfb.2020.111060 · Cited by 8 · see citing works
Self-citationHighly Efficient Electrosynthesis of Nitric Oxide for Biomedical Applications
Jin, J., Mao, J., Wu, W., Jiang, Y. Y., Ma, W., Yu, P., & Mao, L. (2022) · Angewandte Chemie · doi:10.1002/ange.202210980 · Cited by 7 · see citing works
Dongmo, S., Leyk, J., Dosche, C., Richter‐Landsberg, C., Wollenberger, U., & Wittstock, G. (2016) · Electroanalysis · doi:10.1002/elan.201600267 · Cited by 4 · see citing works
Improving the hemocompatibility of catheters via NO release/generation
Wo, Y., Brisbois, E. J., Bartlett, R. H., & Meyerhoff, M. E. (2017) · Elsevier eBooks · doi:10.1016/b978-0-08-100497-5.00013-6 · Cited by 2 · see citing works
Self-citationZhang, W., Alsuwayni, B., Liu, J., Wu, Q., Mei, Z., Hou, S., Zhang, Z., Du, X., Yi, S., Li, S., et al. (2026) · Chemical Science · doi:10.1039/d6sc02701f
- Authors
- Ren, H., Coughlin, M. A., Major, T. C., Aiello, S., Rojas Pena, A., Bartlett, R. H., & Meyerhoff, M. E.
- Affiliations
- University of Michigan (US)
- University of Michigan , , , ,
- Corresponding author
- Meyerhoff, Mark E.
- Venue
- Analytical Chemistry
- Publisher
- American Chemical Society (ACS)
- Published
- 2015-07-23
- Volume
- 87
- Issue
- 16
- Pages
- 8067-8072
- ISSN
- 0003-2700 (ISSN-L)
- 1520-6882
- PMC
- PMC4556045
- OpenAlex
- W2419451964
- Cited by
- 30
- References
- 29
- Citations by year
- 2026: 1 · 2024: 1 · 2023: 1 · 2022: 3 · 2021: 2 · 2020: 5 · 2019: 2 · 2018: 4 · 2017: 4 · 2016: 7
- FWCI
- 2.41
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Analytical Chemistry (publishedVersion, other-oa)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Heart, Lung, and Blood Institute (HL-119403-01)
- National Institute of Biomedical Imaging and Bioengineering (EB-000783)
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Physical Sciences › Chemistry › Electrochemistry › Electrochemical Analysis and Applications
- Keywords
- nitric oxide release, amperometric oxygen sensor, oxygen partial pressure, implantable catheters, double-lumen catheter, thrombus formation, biocompatibility
- MeSH terms
- Animals, Electrochemistry, Monitoring, Physiologic, Nitric Oxide, Oxygen, Rabbits, Swine, Biosensing Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review