Touch-Based Fingertip Blood-Free Reliable Glucose Monitoring: Personalized Data Processing for Predicting Blood Glucose Concentrations
Abstract
Diabetes prevalence has been rising exponentially, increasing the need for reliable noninvasive approaches for glucose monitoring. Different biofluids have been explored recently for replacing current blood finger-stick glucose strips with noninvasive painless sensing devices. While sweat has received considerable attention, there are mixed reports on correlating the sweat results with blood glucose levels. Here, we demonstrate a new rapid and reliable approach that combines a simple touch-based fingertip sweat electrochemical sensor with a new algorithm that addresses for personal variations toward the accurate estimate of blood glucose concentrations. The new painless and simple glucose self-testing protocol leverages the fast sweat rate on the fingertip for rapid assays of natural perspiration, without any sweat stimulation, along with the personalized sweat-response-to-blood concentration translation. A reliable estimate of the blood glucose sensing concentrations can thus be realized through a simple one-time personal precalibration. Such system training leads to a substantially improved accuracy with a Pearson correlation coefficient higher than 0.95, along with an overall mean absolute relative difference of 7.79%, with 100% paired points residing in the A + B region of the Clarke error grid. The speed and simplicity of the touch-based blood-free fingertip sweat assay, and the elimination of periodic blood calibrations, should lead to frequent self-testing of glucose and enhanced patient compliance toward the improved management of diabetes.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Saeedi, P., Petersohn, I., Salpea, P., Malanda, B., Karuranga, S., Unwin, N., Colagiuri, S., Guariguata, L., Motala, A. A., Ogurtsova, K., et al. (2019) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2019.107843 · Cited by 10,792
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373
Self-citationAn epidermal patch for the simultaneous monitoring of haemodynamic and metabolic biomarkers
Sempionatto, J. R., Lin, M., Yin, L., De la paz, E., Pei, K., Sonsa-ard, T., de Loyola Silva, A. N., Khorshed, A. A., Zhang, F., Tostado, N., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00685-1 · Cited by 658
Self-citationCorrelation Between Sweat Glucose and Blood Glucose in Subjects with Diabetes
Moyer, J., Wilson, D., Finkelshtein, I., Wong, B., & Potts, R. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0262 · Cited by 570
Wireless smart contact lens for diabetic diagnosis and therapy
Keum, D. H., Kim, S.-K., Koo, J., Lee, G.-H., Jeon, C., Mok, J. W., Mun, B. H., Lee, K. J., Kamrani, E., Joo, C.-K., et al. (2020) · Science Advances · doi:10.1126/sciadv.aba3252 · Cited by 483
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475
Self-citationNyein, H. Y. Y., Bariya, M., Kivimäki, L., Uusitalo, S., Liaw, T. S., Jansson, E., Ahn, C. H., Hangasky, J. A., Zhao, J., Lin, Y., et al. (2019) · Science Advances · doi:10.1126/sciadv.aaw9906 · Cited by 388
Wearable Sensors for Biochemical Sweat Analysis
Bandodkar, A. J., Jeang, W. J., Ghaffari, R., & Rogers, J. A. (2019) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061318-114910 · Cited by 387
Physiological mechanisms determining eccrine sweat composition
Baker, L. B., & Wolfe, A. S. (2020) · European Journal of Applied Physiology · doi:10.1007/s00421-020-04323-7 · Cited by 367
Non‐invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016
Heikenfeld, J. (2016) · Electroanalysis · doi:10.1002/elan.201600018 · Cited by 294
Cited by
Skin-Interfaced Wearable Sweat Sensors for Precision Medicine
Min, J., Tu, J., Xu, C., Lukas, H., Shin, S., Yang, Y., Solomon, S. A., Mukasa, D., & Gao, W. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.2c00823 · Cited by 712
Wearable Electrochemical Glucose Sensors in Diabetes Management: A Comprehensive Review
Saha, T., Del Caño, R., Mahato, K., De la Paz, E., Chen, C., Ding, S., Yin, L., & Wang, J. (2023) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00078 · Cited by 442
Self-citationLi, Q.-F., Chen, X., Wang, H., Liu, M., & Peng, H.-L. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c20543 · Cited by 255
Moon, J., Teymourian, H., De la Paz, E., Sempionatto, J. R., Mahato, K., Sonsa‐ard, T., Huang, N., Longardner, K., Litvan, I., & Wang, J. (2021) · Angewandte Chemie International Edition · doi:10.1002/anie.202106674 · Cited by 103
Self-citationA review of sampling, energy supply and intelligent monitoring for long-term sweat sensors
Yang, P., Wei, G., Liu, A., Huo, F., & Zhang, Z. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00165-9 · Cited by 77
Metadata
- Authors
- Sempionatto, J. R., Moon, J.-M., & Wang, J.
- Venue
- ACS Sensors
- Publisher
- American Chemical Society (ACS)
- Published
- 2021-04-19
- Volume
- 6
- Issue
- 5
- Pages
- 1875-1883
- ISSN
- 2379-3694 (ISSN-L)
- OpenAlex
- W3155150693
- Cited by
- 208
- References
- 14
- FWCI
- 10.44
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Touch-Based Fingertip Blood-Free Reliable Glucose Monitoring: Personalized Data Processing for Predicting Blood Glucose Concentrations. https://app.cochat.ai/papers/10.1021/acssensors.1c00139
- The paper
- Sempionatto, J. R., Moon, J.-M., & Wang, J. (2021). Touch-Based Fingertip Blood-Free Reliable Glucose Monitoring: Personalized Data Processing for Predicting Blood Glucose Concentrations. ACS Sensors. https://doi.org/10.1021/acssensors.1c00139
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.