A review of sampling, energy supply and intelligent monitoring for long-term sweat sensors
Open accessAbstract
Sweat is a biofluid with rich information that can reflect an individual’s state of health or activity. But the real-time in situ sweat sensors lack the ability of long-term monitoring. Against this background, this article provides a holistic review on the necessary process and methods for sweat sensing, including sweat collection, composition analysis, energy supply, and data processing. The impacts of the environment in stimulating sweat production, providing energy supply, and intelligent health monitoring are discussed. Based on the review of previous endeavors, the future development in material, structure and artificial intelligence application of long-term sweat monitoring is envisioned.
AI snapshot
Written from the paper’s full text.
TL;DR
This review surveys the components needed for long-term wearable sweat sensing: sweat sampling/control, sweat composition and analytes, sustainable energy supply, and AI/ML-based intelligent monitoring. It argues that long-term sweat monitoring remains limited by collection efficiency, contamination, energy autonomy, unclear sweat–health relationships, and sensor accuracy.
Key findings
- Sweat secretion rates can be lower than 10 nL min⁻¹ cm⁻² at low-secretion sites such as arm or leg, but can reach 100 nL min⁻¹ cm⁻² at high-secretion areas like palm and foot.
- Sweat pH is generally reported as 4.2–7.8; sweat lactate concentration ranges from ~6 mM to ~100 mM; sweat cortisol is reported as 2.21 × 10⁻⁵ to 3.86 × 10⁻⁴ mM versus blood cortisol generally 1.24 × 10⁻⁴ to 4.0 × 10⁻⁴ mM.
- For energy, woven lithium-ion fiber batteries reportedly retain 90.5% capacity after 500 charge–discharge cycles and over 80% after 100,000 bending cycles; TENG devices in the literature can supply energy for up to 72 h.
- The review identifies a shift from uncontrolled sweat collection toward precise, controlled microfluidic/wettability-based sampling, and highlights biofuel cells, TENG, PENG, and WTEG as promising self-powered options, while noting that hybrid energy supply and energy management remain challenging.
Methods
- Narrative review of wearable sweat sensors, covering sweat production/collection, composition analysis, energy supply/storage, and intelligent monitoring/data processing.
- Bibliometric analysis using Web of Science for “wearable sweat sensor” and VOSviewer to create a network visualization map of the top 50 keywords.
- No primary experimental data are reported; the paper synthesizes prior studies and discusses future directions in materials, structures, and artificial intelligence.
Limitations
- The paper states that long-term sweat monitoring is still uncertain and affected by environmental contamination, equipment reliability, biocompatibility, energy supply, and system design; it also states that sweat composition differs by stimulation method and that no clear relationship exists between sweat signals and human health parameters.
- The paper states that wearable sensor screening accuracy is unlikely to match special medical devices, so such sensors may provide only rough trend information; it also raises privacy and data-security risks for intelligent monitoring.
- Evident from the text: this is a narrative review, not a systematic review; the bibliometric search is limited to Web of Science and the topic term “wearable sweat sensor,” with no formal inclusion/exclusion criteria or quality appraisal described.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 19 cited sources checked · 0 supported · 0 flagged
- Findingreasoning
Control of paper bases and fabrics for sweat collection remains a problem.
“They have high hydrophilicity, a natural advantage for sweat collection, but control remains a problem.”
Quote in the paperSubstantial claim - Findingnone given
Sweat contains an abundance of small molecules.
“sweat is a biological fluid with an abundance of small molecules that make it a potential choice for biofuel cells.”
Quote in the paperSubstantial claim - Premisecitation
Sweat contains abundant biomarkers such as ions and metabolites that can be used to track metabolism, fatigue, and stress.
“Among the biological fluids, sweat contains an abundance of relevant biomarkers, such as ions and metabolites, that can be leveraged for tracking the user's metabolism, fatigue, and stress [17][18][19][20]”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [17][18][19][20]
Not checked in the free run.
- Interpretationnone given
In vitro sweat collection methods including sweat-absorbent patches, armbands and microcatheters have the problem that sweat is lost during in vitro analysis after collection.
“However, all methods have the problem that sweat is lost during the in vitro analysis after collection.”
Quote in the paperSubstantial claim - Premisecitation
The secretion mechanism of lactic acid into sweat is complex and still unclear.
“However, the secretion mechanism of lactic acid into sweat is complex and still unclear 79 .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [79] Jia, W. et al. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Anal. Chem. 85, 6553–6560 (2013). source
Not checked in the free run.
- Premisecitation
A critical turning point occurred in 2016 when Gao published a fully integrated array sweat sensor in Nature that simultaneously and selectively measured sweat metabolites and electrolytes without external analysis.
“A critical turning points occurred in 2016, when Gao 56 published research on a fully integrated array sweat sensor in Nature that simultaneously and selectively measured sweat metabolites and electrolytes without requiring external analysis 57 .”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites Gao 56
Not checked in the free run.
- cites [57] Heikenfeld, J. Non-invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016. Electroanalysis 28, 1242–1249 (2016). source
Not checked in the free run.
- Premisereasoning
Unlike ions, metabolites need to be precisely monitored using enzymes.
“Unlike ions, these metabolites need to be precisely monitored using enzymes.”
Quote in the paperNot a substantial claim - Interpretationnone given
Wearable sweat sensors are still far from being realized for commercial health monitoring.
“Despite the research to date, a wearable sweat sensor is still far from being realized for commercial health monitoring.”
Quote in the paperNot a substantial claim - Premisecitation
The whole body-washdown method is the gold standard for sweat analysis.
“This rigorous and standard method is the gold standard for sweat analysis.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Bandodkar et al. 62
Not checked in the free run.
- Premisenone given
Different components of sweat can reflect different health conditions.
“Different components can reflect different health conditions.”
Quote in the paperSubstantial claim - Interpretationreasoning
Due to limitations of enzyme activity, the life of an enzyme sensor will be lower than that of an ion-selective sensor.
“Still, due to the limitations of enzyme activity, the life of the enzyme sensor will be lower than that of the ion-selective sensor.”
Quote in the paperNot a substantial claim - Premisecitation
Iontophoresis is a useful method to obtain sufficient sweat when sweat amount may not meet monitoring needs.
“In such circumstances, iontophoresis, whereby sweat glands are stimulated through delivery of an agonist using a low electrical current, is an useful method to obtain sufficient sweat 60,61 .”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [60] Emaminejad, S. et al. Autonomous sweat extraction and analysis applied to cystic fibrosis and glucose monitoring using a fully integrated wearable platform. Proc. Natl Acad. Sci. USA. 114, 4625–4630 (2017). source
Not checked in the free run.
- cites [61] Kim, J. et al. Noninvasive Alcohol Monitoring Using a Wearable Tattoo-Based Iontophoretic-Biosensing System. ACS Sens. 1, 1011–1019 (2016). source
Not checked in the free run.
- Interpretationreasoning
Sweat collection is more complicated than standard body fluid analysis such as blood analysis or urinalysis.
“Sweat collection is more complicated than standard body fluid analysis, such as blood analysis or urinalysis.”
Quote in the paperSubstantial claim - Interpretationreasoning
It is difficult for a battery-free system to meet the needs of continuous monitoring for an extended period.
“Therefore, it is difficult for a battery-free system to meet the needs of continuous monitoring for an extended period.”
Quote in the paperSubstantial claim - Interpretationnone given
NFC has a short interaction distance compared with other communication methods such as Bluetooth and WiFi.
“In addition, compared with other communication methods, such as Bluetooth and WiFi, its short interaction distance is limiting.”
Quote in the paperSubstantial claim - Interpretationnone given
Traditional batteries are not light and thin enough to meet the needs of flexible wearables.
“However, traditional batteries are not light and thin enough to meet the needs of flexible wearables.”
Quote in the paperSubstantial claim - Interpretationreasoning
NFC passive data transmission mode cannot achieve continuous data monitoring.
“However, this acquisition mode cannot achieve continuous data monitoring.”
Quote in the paperSubstantial claim - Interpretationreasoning
In the case of insufficient monitoring accuracy, wearable sensors can still only provide rough trend information of health targets.
“Therefore, in the case of insufficient monitoring accuracy, wearable sensors can still only provide rough trend information of health targets.”
Quote in the paperSubstantial claim - Interpretationreasoning
Coordination of multiple energy supply systems brings challenges in terms of energy management and complexity of structure.
“However, this also brings about challenges in terms of energy management and the complexity of structure.”
Quote in the paperNot a substantial claim - Premisecitation
TENG devices reported in the literature can supply energy for up to 72 h.
“TENG devices reported in the literature can supply energy for up to 72 h.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Xia et al. 106
Not checked in the free run.
- Interpretationreasoning
Biofuel cells are one of the most suitable components of self-powered sweat sensors.
“Therefore, biofuel cells are one of the most suitable components of self-powered sweat sensors.”
Quote in the paperSubstantial claim - Premisereasoning
Lactic acid in sweat can meet the needs of low-power sensors.
“The commonly used battery type is a lactic acid fuel cell, and lactic acid in sweat can meet the needs of low-power sensors.”
Quote in the paperSubstantial claim - Interpretationnone given
Current paper-based self-powered sensors still face many problems, including limited mechanical stability, moisture resistance and lifespan.
“Unfortunately, current paper-based self-powered sensors still face many problems, including limited mechanical stability, moisture resistance and lifespan.”
Quote in the paperSubstantial claim
- May combine two claims. 29
- Classified as probably not a substantial claim. 55
May combine two claims.
Findingown dataBibliometric search on Web of Science identified relevant research on wearable sweat sensors since 2005, but only a few researchers focused on sweat sensing.
“Searching on the subject of 'wearable sweat sensor' on the Web of Science website identified the presence of relevant research since 2005, but only a few researchers focused on sweat sensing.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataVOSviewer keyword map core keywords are 'wearable,' 'sensor,' 'biosensor' and 'sweat.'
“The core keywords observed are 'wearable,' 'sensor,' 'biosensor' and 'sweat.'”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
FindingreasoningResearch on wearable sweat sensors has grown rapidly in the past five years.
“This significant work attracted the attention of many researchers and in the past five years, research on wearable sweat sensors has grown rapidly.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Findingnone givenProblems of low collection efficiency and poor results still exist.
“However, problems of low collection efficiency and poor results still exist.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Findingnone givenPaper bases and fabrics have high hydrophilicity.
“They have high hydrophilicity, a natural advantage for sweat collection, but control remains a problem.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Findingnone givenTraditional microfluidic technology remains a major contender for sweat collection but faces the problem of processing efficiency.
“Traditional microfluidic technology remains a major contender for sweat collection but faces the problem of processing efficiency.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenRegarding sweat collection, a shift has occurred from extensive uncontrolled collection to more precise and controlled collection.
“Regarding sweat collection, a shift has occurred from extensive uncontrolled collection to more precise and controlled collection.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenSpecific differences occur in the composition of sweat obtained by different methods used to stimulate perspiration.
“In addition, specific differences occur in the composition of sweat obtained by different methods used to stimulate perspiration; exploring and identifying these differences is a key area for future research.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenContamination on the skin surface is a key challenge affecting the accuracy of the device.
“Contamination on the skin surface is a key challenge affecting the accuracy of the device.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenCurrent wearable self-powered devices have good compatibility with wearable sweat sensors.
“Current wearable self-powered devices have good compatibility with wearable sweat sensors.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenNoninvasive sweat analysis increases the chance of personal information theft.
“Careful consideration must be given to the risk with intelligent monitoring of personal privacy being maintained, especially with noninvasive sweat analysis, which increases the chance of personal information theft.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Findingnone givenA single energy supply element cannot meet the demands faced with scenarios that require data transmission.
“However, the problem is not only the energy supply of the device itself but also that a single energy supply element cannot meet the demands faced with scenarios that require data transmission.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Findingnone givenSweat production is strongly coupled to movement and temperature.
“In addition, because sweat production is strongly coupled to movement and temperature, PENG (piezoelectric nanogenerators) and WTNG (wearable thermoelectric generators) are energy supplies that should be further explored.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenLimitations remain in the accuracy of existing sensors.
“While limitations remain in the accuracy of existing sensors, obtaining data is only one component of the original intention of sweat sensing and health monitoring.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Findingnone givenTENG has shown its potential in sweat sensing and monitoring.
“At the same time, TENG, which has received much attention recently, has also shown its potential in sweat sensing and monitoring.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Findingnone givenCurrent analysis of sweat data is limited to visual presentation or simple statistical analysis.
“The current analysis of sweat data is limited to visual presentation or simple statistical analysis.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingnone givenBoth lactic acid and glucose have been demonstrated as potential fuels for the self-powered sensor.
“Both lactic acid and glucose have been demonstrated as potential fuels for the self-powered sensor.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenCurrent real-time in situ sweat sensors lack the ability for long-term monitoring.
“But the real-time in situ sweat sensors lack the ability of long-term monitoring.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationnone givenCurrent research has not found that pH value of sweat alone can reflect health or disease, but pH is an essential reference.
“Although current research has not found that pH value of sweat alone can reflect the characteristics of health or disease, pH value is an essential reference.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationSweat secretion rates can be lower than 10 nL min−1 cm−2 at low secretion sites such as an arm or leg, but can reach 100 nL min−1 cm−2 at high secretion areas like the palm and foot.
“Sweat secretion rates can be lower than 10 nL min -1 cm -2 at low secretion sites such as an arm or leg, but can reach 100 nL min -1 cm -2 at high secretion areas like the palm and foot 5 .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [5] Nyein, H. Y. Y. et al. A wearable patch for continuous analysis of thermoregulatory sweat at rest. Nat. Commun. 12, 1823 (2021). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenThe ions in sweat are thought to reflect the level of water and salt metabolism in the human body.
“The ions in sweat are thought to reflect the level of water and salt metabolism in the human body.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningSuch in vitro collection methods cannot be accurately time-stamped, making it difficult to match sweat analysis results to the subject's state.
“Further, such collection methods cannot be accurately time-stamped, making it difficult to match the results of the sweat analysis to the state of the subject.”
Quote in the paperSubstantial claimMay combine two claims.
Premisenone givenSodium ions can indirectly reflect the sweat rate of the human body and provide a reference for calibration of sweat monitoring.
“Sodium ions can indirectly reflect the sweat rate of the human body and thus also provide a reference for calibration of sweat monitoring.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenMetabolites in sweat are represented by lactic acid and glucose.
“The metabolites in sweat are represented by lactic acid and glucose.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe current mechanism of potassium ion secretion in sweat is still uncertain, so the final potassium concentration in sweat may be maintained within the plasma potassium concentration range.
“However, the current mechanism of potassium ion secretion in sweat is still uncertain, so the final potassium concentration in sweat may be maintained within the plasma potassium concentration range 76 .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [76] Baker, L. B. Sweating Rate and Sweat Sodium Concentration in Athletes: A Review of Methodology and Intra/Interindividual Variability. Sport. Med. 47, 111–128 (2017). source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
Interpretationnone givenCore issues for long-term sweat monitoring include sweat collection and analysis, energy supply and management, and establishing the relationship between monitoring data and human health.
“The core issues requiring attention relate to long-term monitoring, including the mechanism for collection and analysis of sweat, the supply and management of energy to the device, and the establishment of the relationship between monitoring data and human health.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenSweat and blood concentrations of calcium are correlated.
“Sweat and blood concentrations of calcium are correlated.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Premisenone givenSodium is the most common ion in human sweat and the primary indicator for levels of human water and salt metabolism.
“The most common ion in human sweat, sodium is the primary indicator for levels of human water and salt metabolism.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenExcessive loss of sodium and potassium can lead to hyponatremia, muscle cramps, and dehydration.
“Excessive loss of sodium and potassium can lead to hyponatremia, muscle cramps, and dehydration.”
Quote in the paperNot a substantial claimMay combine two claims.
Premisenone givenThe concentration of ions in sweat is much lower than in blood, but they can still reflect the state of the human.
“Although the concentration of ions in sweat is much lower than in blood, they can still reflect the state of the human.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationAbnormally high sweat chloride concentration is observed in cystic fibrosis patients, and sweat chloride analysis has been adopted as the gold standard for cystic fibrosis diagnosis.
“Abnormally high sweat chloride concentration is observed in cystic fibrosis (CF) patients, and sweat chloride analysis has been adopted as the gold standard for cystic fibrosis diagnosis 75”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [75] O’Sullivan, B. P. & Freedman, S. D. Cystic fibrosis. Lancet 373, 1891–1904 (2009). source
Not checked in the free run.
Probably not a substantial claim: may not pass S4 (based).
Interpretationnone givenA closed sweat collection system can effectively avoid external contamination.
“The advantage to this method is that a closed sweat collection system can effectively avoid external contamination.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Premisenone givenIt is generally believed that the pH value of human sweat is in the range of 4.2-7.8.
“It is generally believed that the pH value of human sweat is in the range of 4.2-7.8.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenLocal sweat differences are likely to occur based on factors such as different parts of the human body, different diets, and climate.
“At the same time, local sweat differences are likely occur based on factors such as different parts of the human body, different diets, and climate.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone givenThe concentration of cortisol in sweat ranges from 2.21 × 10−5 to 3.86 × 10−4 mM, with the largest concentration typically occurring in the morning.
“The concentration of cortisol in sweat ranges from 2.21 × 10 -5 to 3.86 × 10 -4 mM, with the largest concentration typically occurring in the morning.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationLi-ion batteries have high energy density, high power density and a long life cycle.
“Li-ion batteries have high energy density, high power density and a long life cycle.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Liu 97
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Premisenone givenUA is a risk factor for cardiovascular disease, type 2 diabetes, and renal disease; it is also widely used in clinical settings to manage gout.
“UA is a risk factor for cardiovascular disease, type 2 diabetes, and renal disease; it is also widely used in clinical settings to manage gout, the most common inflammatory arthritis, which affects tens of millions of people worldwide.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based). May combine two claims.
Premisenone givenHeavy metal ions are not ordinary human sweat metabolites, but the primary way to metabolize heavy metal ions is through sweat.
“Heavy metal ions are not ordinary human sweat metabolites, but the primary way to metabolize heavy metal ions is through sweat.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationStudies have shown a high correlation between the concentrations of alcohol in sweat and blood, which enables continuous noninvasive blood alcohol monitoring by monitoring sweat alcohol concentration.
“Studies have shown a high correlation between the concentrations of alcohol in sweat and blood, which enables continuous noninvasive blood alcohol monitoring to be achieved by monitoring sweat alcohol concentration 91,92 .”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [91] Gamella, M. et al. A novel non-invasive electrochemical biosensing device for in situ determination of the alcohol content in blood by monitoring ethanol in sweat. Anal. Chim. Acta 806, 1–7 (2014). source
Not checked in the free run.
- cites [92] Davis-Martin, R. E., Alessi, S. M. & Boudreaux, E. D. Alcohol Use Disorder in the Age of Technology: A Review of Wearable Biosensors in Alcohol Use Disorder Treatment. Front. Psychiatry 12, 642813 (2021). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenMultichannel sweat composition analysis has become a relatively mature technology.
“Multichannel sweat composition analysis has become a relatively mature technology, but the perceptual parameters required differ according to the specific health monitoring task.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone givenIntrinsically flexible batteries still have certain technical difficulties, but structurally flexible batteries realized by island structures are relatively easy to implement.
“Intrinsically flexible batteries still have certain technical difficulties, but structurally flexible batteries realized by island structures are relatively easy to implement.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenCompared with the activity of the whole body, sweat lactic acid may have a more direct relationship with the activity of the sweat glands themselves.
“Compared with the activity of the whole body, sweat lactic acid may have a more direct relationship with the activity of the sweat glands themselves.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenScreening accuracy with wearable sensors is unlikely to match that required of special medical devices.
“The challenges with this are that the targets that can be monitored are limited and screening accuracy with wearable sensors is unlikely to match that required of special medical devices.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenFlexible batteries are currently the best technical option for traditional sensors.
“Flexible batteries are currently the best technical option for traditional sensors.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone givenA battery-free system is a simpler and more stable structure with a lower cost.
“A battery-free system is a simpler and more stable structure with a lower cost.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenDeciding on the appropriate parameter to monitor depends on medical data providing sufficient theoretical analysis to identify the parameter and provide quantitative data regarding normal levels.
“Deciding on the appropriate parameter to monitor, however, depends on the medical data to provide sufficient theoretical analysis to not only identify the parameter, but also to provide quantitative data regarding normal levels.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenThe concentration range of lactic acid in sweat is ~6 mM to ~100 mM.
“The concentration range of lactic acid in sweat ranges from ~6 mM to ~100 mM.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationA correlation between sweat glucose and blood glucose exists, and obtaining blood glucose values by detecting glucose in sweat is expected to replace conventional diabetes blood glucose monitoring practices.
“A correlation between sweat glucose and blood glucose exists and obtaining blood glucose values by detecting glucose in sweat is thus expected to replace conventional diabetes blood glucose monitoring practices 83,84”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [83] Zhou, W. et al. Recent advances in flexible sweat glucose biosensors. J. Phys. D. Appl. Phys. 54, 423001 (2021). source
Not checked in the free run.
- cites [84] Yunos, M. F. A. M. & Nordin, A. N. Non-invasive glucose monitoring devices: A review. Bull. Electr. Eng. Inform. 9, 2609–2618 (2020). source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S4 (based).
Interpretationnone givenMedical knowledge may not be able to provide such information.
“Unfortunately, medical knowledge may not be able to provide such information.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe cortisol concentration in the blood is generally 1.24 × 10−4 to 4.0 × 10−4 mM.
“In comparison, the cortisol concentration in the blood is generally 1.24 × 10 -4 to 4.0 × 10 -4 mM 88 .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [88] Raul, J. S., Cirimele, V., Ludes, B. & Kintz, P. Detection of physiological concentrations of cortisol and cortisone in human hair. Clin. Biochem. 37, 1105–1111 (2004). source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Interpretationnone givenThe difference between sweat and blood cortisol concentrations may be related to type II 11β-hydroxysteroid-dehydrogenase (HSD) in the sweat ducts that converts cortisol into corticosterone.
“This difference may be related to the type II 11β-hydroxysteroid-dehydrogenase (HSD) in the sweat ducts that converts cortisol into corticosterone.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenSweat lactic acid can indirectly reflect the physical state of exercise.
“Nonetheless, sweat lactic acid can indirectly reflect the physical state of exercise.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationnone givenExisting self-powered systems can meet the energy requirements of the body surface sensor, but research on design is required to achieve a stable supply of energy.
“While existing selfpowered systems can meet the energy requirements of the body surface sensor, research on design is required to achieve a stable supply of energy.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Interpretationnone givenMost of the energy sources for TENG fabrics are low-frequency human motion and transformation of the motion to energy for a sensor remains to be studied.
“However, most of the energy sources for TENG fabrics are low-frequency human motion and transformation of the motion to energy for a sensor remains to be studied.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
InterpretationreasoningBecause TENG energy supply requires mechanical movement, the relationship between exercise and sweat in human health monitoring makes it a good alternative.
“Because its energy supply requires mechanical movement, the relationship between exercise and sweat in human health monitoring makes it a good alternative.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
InterpretationreasoningTypically, sweat production is accompanied by higher exercise intensity or ambient temperature, providing an energy source for various wearable self-powered devices.
“Typically, sweat production is accompanied by higher exercise intensity or ambient temperature, providing an energy source for various wearable self-powered devices that can, in turn be used for sweat sensor monitoring.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
Interpretationnone givenThe use of biofuel cells as a self-powered technology for sweat sensing has been studied and needs to now be tested in the context of applied use.
“In summary, the use of biofuel cells as a self-powered technology for sweat sensing has been studied and needs to now be tested in the context of applied use.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Premisenone givenTENG is one of the most popular devices in use for self-powered wearable systems.
“TENG is also one of the most popular devices in use for selfpowered wearable systems.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationA hybrid energy supply could further enhance the stability of energy supply.
“It is worth noting that a hybrid energy supply could further enhance the stability of energy supply 112 .”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [112] García Núñez, C., Manjakkal, L. & Dahiya, R. Energy autonomous electronic skin. npj Flex. Electron 3, 1 (2019). source
Not checked in the free run.
May combine two claims.
InterpretationreasoningA major advantage of the biofuel cell is that it can use the very substances the sensor is detecting, reducing the system's complexity and ensuring its stability.
“A major advantage of the biofuel cell is that it can use the very substances the sensor is detecting, reducing the system's complexity and ensuring its stability.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationWoven lithium-ion fiber batteries have a capacity retention that reaches 90.5% after 500 charge-discharge cycles, comparable to commercial batteries such as pouch cells, and over 80% capacity was maintained after bending the fiber for 100,000 cycles.
“These batteries have a capacity retention that reaches 90.5% after 500 charge-discharge cycles, comparable to commercial batteries such as pouch cells.Over 80% capacity was maintained after bending the fiber for 100,000 cycles.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites Peng et al. 99
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenSelf-powered devices such as TENG, PENG, and WTEG all show potential in the field of sweat sensors.
“Self-powered devices such as TENG, PENG, and WTEG all show potential in the field of sweat sensors.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
InterpretationreasoningWith proper design, self-powered flexible sensors can meet the long-term monitoring goals of most applications.
“With proper design, self-powered flexible sensors can meet the long-term monitoring goals of most applications.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenFor textile wearable TENG, as long as the structure was not damaged, its long-term stability was satisfactory.
“As long as the structure was not damaged, its long-term stability was satisfactory.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S4 (based).
InterpretationreasoningThe correlation between the production of sweat and environmental factors makes it especially feasible to match the self-supply of the environment.
“The correlation between the production of sweat and environmental factors makes it especially feasible to match the self-supply of the environment.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing). May combine two claims.
Interpretationnone givenAlthough the potential of textile TENG for sweat monitoring has been verified, more specific selectivity and stability research is required for further exploration.
“Although the potential has been verified, more specific selectivity and stability research is required for further exploration.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim. May combine two claims.
InterpretationreasoningThe coordination of multiple energy supply systems can cover a broader range of sensor device usage requirements and avoid the limitations of a single self-supply system.
“The coordination of multiple energy supply systems can cover a broader range of sensor device usage requirements and avoid the limitations of a single selfsupply system.”
Quote in the paperSubstantial claim
The paper relies on this without support.
Sweat composition obtained by iontophoresis differs from the composition of sweat spontaneously produced by the human body.
“However, sweat composition obtained by iontophoresis differs from the composition of sweat spontaneously produced by the human body.”
The paper relies on this without support.
Long-term continuous monitoring is required for wearable sensors to obtain sufficient data to accurately measure human health.
“For wearable sensors, long-term continuous monitoring is required to obtain sufficient data to accurately measure human health.”
The paper relies on this without support.
Due to current technology and data limitations, it is still difficult to directly obtain critical parameters that affect the state of the human body.
“However, due to current technology and data limitations, it is still difficult to directly obtain critical parameters that affect the state of the human body.”
The paper relies on this without support.
Most researchers currently study simple heart rate, electrocardiograms, blood oxygen levels, and other simple parameters.
“Most researchers currently study simple heart rate, electrocardiograms, blood oxygen levels, and other simple parameters.”
The paper relies on this without support.
The human body is a complex system and a variety of monitorable data is required to establish a comprehensive model of its health.
“The human body is a complex system and a variety of monitorable data is required to establish a comprehensive model of its health.”
The paper relies on this without support.
AI can be integrated into wearable sensors to improve the applied performance of wearable sensors for health monitoring.
“Artificial intelligence (AI) can be integrated into wearable sensors to improve the applied performance of wearable sensors for health monitoring.”
The paper relies on this without support.
The AI biosensor is a rapidly growing area of research.
“The AI biosensor is a rapidly growing area of research.”
The paper relies on this without support.
At present, there is no clear relationship between sweat signals and human health parameters.
“At present, there is no clear relationship between sweat signals and human health parameters; yet many scholars have studied the relationship between sweat sugar and blood sugar.”
The paper relies on this without support.
Many scholars have studied the relationship between sweat sugar and blood sugar.
“At present, there is no clear relationship between sweat signals and human health parameters; yet many scholars have studied the relationship between sweat sugar and blood sugar.”
The paper relies on this without support.
The analysis of blood glucose is a relatively mature applied model, for which physiological model-based algorithms and datadriven algorithms are used.
“The analysis of blood glucose is a relatively mature applied model, for which physiological model-based algorithms and datadriven algorithms are used.”
The paper relies on this without support.
The physiological model of the human body depends on many factors (e.g., gender, age, weight, meals, external environment).
“The physiological model of the human body depends on many factors (e.g., gender, age, weight, meals, external environment).”
The paper relies on this without support.
In the classic biological model, glucose, insulin, and glucagon must be considered comprehensively.
“In the classic biological model, glucose, insulin, and glucagon must be considered comprehensively.”
The paper relies on this without support.
Change in blood sugar levels has a certain degree of hysteresis, and the change of glucose in sweat depends on, and lags behind, the blood sugar, making it difficult to establish the model.
“At the same time, change in blood sugar levels has a certain degree of hysteresis, and the change of glucose in sweat depends on, and lags behind, the blood sugar, making it difficult to establish the model.”
The paper relies on this without support.
The complexity of the human body further makes the modeling accuracy insufficient.
“The complexity of the human body further makes the modeling accuracy insufficient.”
The paper relies on this without support.
In recent years, data-driven algorithms have been increasingly applied to various model predictions and this can build models even in the case of complex and unknown relationships of information that are more inline with reality.
“In recent years, data-driven algorithms have been increasingly applied to various model predictions and this can build models even in the case of complex and unknown relationships of information that are more inline with reality.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Despite progress in wearable biosensors, there is a gap in the literature in finding a relationship between biofluid signals and human health parameters.
“In addition, despite all the progress made in wearable biosensors, there is a gap in the literature in finding a relationship between biofluid signals and human health parameters [251].”
From The emergence of AI-based wearable sensors for digital health technology: a review (2023) · cites this paper as [251]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Reference 98 describes a battery, not a biofuel cell
MinorThe text calls the cited device a biofuel cell, while reference 98 identifies it as a fruit-based battery.
“Wang et al. 98 proposed an environmentally friendly and degradable biofuel cell that cleverly uses the kirigami process to design the device structure.”
“98. Wang, Z. et al. Fully transient stretchable fruit-based battery as safe and environmentally friendly power source for wearable electronics. EcoMat 3, e12073 (2021).”
Bandodkar study is dated 2013, but reference 62 is from 2014
MinorThe cited reference is dated 2014, not 2013. Its title also describes a tattoo sodium sensor rather than the whole-body washdown method attributed to it.
“Bandodkar et al. 62 achieved in situ sweat sensing in 2013, using a whole body-washdown method since local sweat collection cannot fully reflect human sweat.”
“62. Bandodkar, A. J. et al. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 54, 603-609 (2014).”
Figure 6c cites a thermoelectric study for a TENG
MinorThe caption describes a textile-based TENG and dates reference 105 to 2018 in Nano Energy. Reference 105 instead describes a thermoelectric generator published in Applied Energy in 2020.
“c A flexible and textile-based self-powered sweat sensor based on TENG.Ref. 105 , Copyright 2018, Nano Energy.”
“105. Yuan, J. & Zhu, R. A fully self-powered wearable monitoring system with sys- tematically optimized flexible thermoelectric generator. Appl. Energy 271, 115250 (2020).”
Figure 6a gives a different publication for reference 100
MinorThe caption assigns reference 100 to a 2021 Advanced Functional Materials publication, whereas the reference list places it in Joule.
“a Self-powered devices, a wearable BFC harvester powered by finger touch.Ref. 100 , Copyright 2021, Advanced Functional Materials.”
“100. Yin, L. et al. A passive perspiration biofuel cell: High energy return on invest- ment. Joule 5, 1888-1904 (2021).”
Figure 6b gives a different year and journal for reference 103
MinorThe caption identifies reference 103 as a 2015 Biosensors and Bioelectronics work; the listed reference is a 2020 Science Advances paper.
“b The FTENG for harvesting energy from human motion.Ref. 103 Copyright 2015, Biosensors and Bioelectronics.”
“100. Song, Y. et al. Wireless battery-free wearable sweat sensor powered by human motion. Sci. Adv. 6, eaay9842 (2020).”
Figure 6d gives a different year and journal for reference 111
MinorThe caption identifies reference 111 as a 2018 Nano Energy publication, but the reference list gives a 2021 Advanced Functional Materials publication.
“d A scalable, flexible magnetoelectrical clothing generator.Ref. 111 , Copyright 2018, Nano Energy.”
“111. Wang, R. et al. Magnetoelectrical Clothing Generator for High-Performance Transduction from Biomechanical Energy to Electricity. Adv. Funct. Mater. 2107682, 1-10 (2021).”
Woven-battery study is attributed to a different author
MinorThe text attributes reference 99 to Peng et al., but the reference lists He et al.
“Peng et al. 99 proposed woven lithium-ion fiber batteries, which can be prepared in large quantities and have good stability.”
“99. He, J. et al. Scalable production of high-performing woven lithium-ion fibre batteries. Nature 597, 57-63 (2021).”
Cites
LeCun, Y., Bengio, Y., & Hinton, G. (2015) · Nature · doi:10.1038/nature14539 · Cited by 84,126 · see citing works
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
Wearable biosensors for healthcare monitoring
Kim, J., Campbell, A. S., de Ávila, B. E.-F., & Wang, J. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0045-y · Cited by 3,373 · see citing works
Lab-on-Skin: A Review of Flexible and Stretchable Electronics for Wearable Health Monitoring
Liu, Y., Pharr, M., & Salvatore, G. A. (2017) · ACS Nano · doi:10.1021/acsnano.7b04898 · Cited by 1,725 · see citing works
Bio-Integrated Wearable Systems: A Comprehensive Review
Ray, T. R., Choi, J., Bandodkar, A. J., Krishnan, S., Gutruf, P., Tian, L., Ghaffari, R., & Rogers, J. A. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00573 · Cited by 1,350 · see citing works
O'Sullivan, B. P., & Freedman, S. D. (2009) · The Lancet · doi:10.1016/s0140-6736(09)60327-5 · Cited by 1,320 · see citing works
Wearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315 · see citing works
A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat
Yang, Y., Song, Y., Bo, X., Min, J., Pak, O. S., Zhu, L., Wang, M., Tu, J., Kogan, A., Zhang, H., et al. (2020) · Nature Biotechnology · doi:10.1038/s41587-019-0321-x · Cited by 1,270 · see citing works
Flexible and Stretchable Energy Storage: Recent Advances and Future Perspectives
Liu, W., Song, M., Kong, B., & Cui, Y. (2016) · Advanced Materials · doi:10.1002/adma.201603436 · Cited by 1,204 · see citing works
Flexible Sensing Electronics for Wearable/Attachable Health Monitoring
Wang, X., Liu, Z., & Zhang, T. (2017) · Small · doi:10.1002/smll.201602790 · Cited by 1,083 · see citing works
Electrochemical Tattoo Biosensors for Real-Time Noninvasive Lactate Monitoring in Human Perspiration
Jia, W., Bandodkar, A. J., Valdés-Ramírez, G., Windmiller, J. R., Yang, Z., Ramírez, J., Chan, G., & Wang, J. (2013) · Analytical Chemistry · doi:10.1021/ac401573r · Cited by 862 · see citing works
Emaminejad, S., Gao, W., Wu, E., Davies, Z. A., Yin Yin Nyein, H., Challa, S., Ryan, S. P., Fahad, H. M., Chen, K., Shahpar, Z., et al. (2017) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1701740114 · Cited by 839 · see citing works
Bandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav3294 · Cited by 761 · see citing works
Accessing analytes in biofluids for peripheral biochemical monitoring
Heikenfeld, J., Jajack, A., Feldman, B., Granger, S. W., Gaitonde, S., Begtrup, G., & Katchman, B. A. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0040-3 · Cited by 758 · see citing works
Wearable Electrochemical Sensors and Biosensors: A Review
Windmiller, J. R., & Wang, J. (2012) · Electroanalysis · doi:10.1002/elan.201200349 · Cited by 699 · see citing works
Nyein, H. Y. Y., Gao, W., Shahpar, Z., Emaminejad, S., Challa, S., Chen, K., Fahad, H. M., Tai, L.-C., Ota, H., Davis, R. W., et al. (2016) · ACS Nano · doi:10.1021/acsnano.6b04005 · Cited by 669 · see citing works
Wireless battery-free wearable sweat sensor powered by human motion
Song, Y., Min, J., Yu, Y., Wang, H., Yang, Y., Zhang, H., & Gao, W. (2020) · Science Advances · doi:10.1126/sciadv.aay9842 · Cited by 652 · see citing works
Noninvasive Alcohol Monitoring Using a Wearable Tattoo-Based Iontophoretic-Biosensing System
Kim, J., Jeerapan, I., Imani, S., Cho, T. N., Bandodkar, A., Cinti, S., Mercier, P. P., & Wang, J. (2016) · ACS Sensors · doi:10.1021/acssensors.6b00356 · Cited by 599 · see citing works
Scalable production of high-performing woven lithium-ion fibre batteries
He, J., Lu, C., Jiang, H., Han, F., Shi, X., Wu, J., Wang, L., Chen, T., Wang, J., Zhang, Y., et al. (2021) · Nature · doi:10.1038/s41586-021-03772-0 · Cited by 561 · see citing works
Multifunctional conductive hydrogel-based flexible wearable sensors
Wang, L., Xu, T., & Zhang, X. (2021) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2020.116130 · Cited by 507 · see citing works
Bandodkar, A. J., Molinnus, D., Mirza, O., Guinovart, T., Windmiller, J. R., Valdés-Ramírez, G., Andrade, F. J., Schöning, M. J., & Wang, J. (2014) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2013.11.039 · Cited by 490 · see citing works
Investigation of Cortisol Dynamics in Human Sweat Using a Graphene-Based Wireless mHealth System
Torrente-Rodríguez, R. M., Tu, J., Yang, Y., Min, J., Wang, M., Song, Y., Yu, Y., Xu, C., Ye, C., IsHak, W. W., et al. (2020) · Matter · doi:10.1016/j.matt.2020.01.021 · Cited by 479 · see citing works
Bioinspired Janus Textile with Conical Micropores for Human Body Moisture and Thermal Management
Dai, B., Li, K., Shi, L., Wan, X., Liu, X., Zhang, F., Jiang, L., & Wang, S. (2019) · Advanced Materials · doi:10.1002/adma.201904113 · Cited by 456 · see citing works
A wearable patch for continuous analysis of thermoregulatory sweat at rest
Nyein, H. Y. Y., Bariya, M., Tran, B., Ahn, C. H., Brown, B. J., Ji, W., Davis, N., & Javey, A. (2021) · Nature Communications · doi:10.1038/s41467-021-22109-z · Cited by 426 · see citing works
Bandodkar, A. J., You, J., Kim, N.-H., Gu, Y. Q., Kumar, R., Mohan, A. M. V., Kurniawan, J. F., Imani, S., Nakagawa, T., Parish, B., et al. (2017) · Energy & Environmental Science · doi:10.1039/c7ee00865a · Cited by 413 · see citing works
Baker, L. B. (2017) · Sports Medicine · doi:10.1007/s40279-017-0691-5 · Cited by 403 · see citing works
Artificial intelligence biosensors: Challenges and prospects
Jin, X., Liu, C., Xu, T., Su, L., & Zhang, X. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112412 · Cited by 399 · see citing works
Reviews of wearable healthcare systems: Materials, devices and system integration
Lou, Z., Wang, L., Jiang, K., Wei, Z., & Shen, G. (2019) · Materials Science and Engineering R Reports · doi:10.1016/j.mser.2019.100523 · Cited by 389 · see citing works
Wearable Sensors for Biochemical Sweat Analysis
Bandodkar, A. J., Jeang, W. J., Ghaffari, R., & Rogers, J. A. (2019) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061318-114910 · Cited by 387 · see citing works
Wearable flexible sweat sensors for healthcare monitoring: a review
Chung, M., Fortunato, G., & Radacsi, N. (2019) · Journal of The Royal Society Interface · doi:10.1098/rsif.2019.0217 · Cited by 378 · see citing works
Battery-free, wireless sensors for full-body pressure and temperature mapping
Han, S., Kim, J., Won, S. M., Ma, Y., Kang, D., Xie, Z., Lee, K., Chung, H. U., Banks, A. R., Min, S., et al. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aan4950 · Cited by 356 · see citing works
Detection of physiological concentrations of cortisol and cortisone in human hair
Raul, J.-S., Cirimele, V., Ludes, B., & Kintz, P. (2004) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2004.02.010 · Cited by 348 · see citing works
Wearable potentiometric ion sensors
Parrilla, M., Cuartero, M., & Crespo, G. A. (2018) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2018.11.024 · Cited by 334 · see citing works
Energy autonomous electronic skin
García Núñez, C., Manjakkal, L., & Dahiya, R. (2019) · npj Flexible Electronics · doi:10.1038/s41528-018-0045-x · Cited by 332 · see citing works
Zhao, Y., Zhai, Q., Dong, D., An, T., Gong, S. P., Shi, Q., & Cheng, W. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b00152 · Cited by 330 · see citing works
Choi, J., Kang, D., Han, S., Kim, S. B., & Rogers, J. A. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201601355 · Cited by 317 · see citing works
Advanced Materials for Health Monitoring with Skin‐Based Wearable Devices
Jin, H., Abu‐Raya, Y. S., & Haick, H. (2017) · Advanced Healthcare Materials · doi:10.1002/adhm.201700024 · Cited by 317 · see citing works
Methylxanthine Drug Monitoring with Wearable Sweat Sensors
Tai, L., Gao, W., Chao, M., Bariya, M., Ngo, Q. P., Shahpar, Z., Nyein, H. Y. Y., Park, H., Sun, J., Jung, Y., et al. (2018) · Advanced Materials · doi:10.1002/adma.201707442 · Cited by 310 · see citing works
Kumar, P. M., & Gandhi, U. D. (2017) · Computers & Electrical Engineering · doi:10.1016/j.compeleceng.2017.09.001 · Cited by 300 · see citing works
Jao, Y.-T., Yang, P., Chiu, C.-M., Lin, Y.-J., Chen, S.-W., Choi, D., & Lin, Z. (2018) · Nano Energy · doi:10.1016/j.nanoen.2018.05.071 · Cited by 296 · see citing works
Self-citationAn IoT-based framework for early identification and monitoring of COVID-19 cases
Otoom, M., Otoum, N. A., Alzubaidi, M. A., Etoom, Y., & Banihani, R. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102149 · Cited by 296 · see citing works
Non‐invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016
Heikenfeld, J. (2016) · Electroanalysis · doi:10.1002/elan.201600018 · Cited by 294 · see citing works
Sensors and Functionalities of Non-Invasive Wrist-Wearable Devices: A Review
Kamišalić, A., Fister, I., Turkanović, M., & Karakatič, S. (2018) · Sensors · doi:10.3390/s18061714 · Cited by 272 · see citing works
Recent advances in 3D printing technologies for wearable (bio)sensors
Kalkal, A., Kumar, S., Kumar, P., Pradhan, R., Willander, M., Packirisamy, G., Kumar, S., & Malhotra, B. D. (2021) · Additive manufacturing · doi:10.1016/j.addma.2021.102088 · Cited by 238 · see citing works
Bio-sensing textile based patch with integrated optical detection system for sweat monitoring
Morris, D., Coyle, S. M., Wu, Y., Lau, K. T., Wallace, G. G., & Diamond, D. (2009) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2009.02.032 · Cited by 234 · see citing works
Wearable Chemical Sensors: Emerging Systems for On-Body Analytical Chemistry
Sempionatto, J. R., Jeerapan, I., Krishnan, S., & Wang, J. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b04668 · Cited by 230 · see citing works
Advanced Wearable Microfluidic Sensors for Healthcare Monitoring
Li, S., Ma, Z., Cao, Z., Pan, L., & Shi, Y. (2019) · Small · doi:10.1002/smll.201903822 · Cited by 214 · see citing works
An automatic and non-invasive physical fatigue assessment method for construction workers
Yu, Y., Li, H., Yang, X., Kong, L., Luo, X., & Wong, A. Y. L. (2019) · Automation in Construction · doi:10.1016/j.autcon.2019.02.020 · Cited by 212 · see citing works
Sempionatto, J. R., Moon, J., & Wang, J. (2021) · ACS Sensors · doi:10.1021/acssensors.1c00139 · Cited by 209 · see citing works
Wearable Biosensors for Non-Invasive Sweat Diagnostics
Xu, J., Fang, Y., & Chen, J. (2021) · Biosensors · doi:10.3390/bios11080245 · Cited by 208 · see citing works
Wang, R., Zhai, Q., An, T., Gong, S. P., & Cheng, W. (2020) · Talanta · doi:10.1016/j.talanta.2020.121484 · Cited by 199 · see citing works
The Detection of Cortisol in Human Sweat
Russell, E., Koren, G., Rieder, M. J., & Uum, S. H. M. V. (2013) · Therapeutic Drug Monitoring · doi:10.1097/ftd.0b013e31829daa0a · Cited by 195 · see citing works
Smart healthcare framework for ambient assisted living using IoMT and big data analytics techniques
Syed, L., Jabeen, S., Manimala, S., & Alsaeedi, A. (2019) · Future Generation Computer Systems · doi:10.1016/j.future.2019.06.004 · Cited by 195 · see citing works
Flexible and Superwettable Bands as a Platform toward Sweat Sampling and Sensing
He, X., Xu, T., Gu, Z., Gao, W. F., Xu, L., Pan, T., & Zhang, X. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.8b05875 · Cited by 193 · see citing works
Integrated Smart Janus Textile Bands for Self-Pumping Sweat Sampling and Analysis
He, X., Yang, S., Pei, Q., Song, Y., Liu, C., Xu, T., & Zhang, X. (2020) · ACS Sensors · doi:10.1021/acssensors.0c00563 · Cited by 186 · see citing works
Keene, S. T., Fogarty, D., Cooke, R., Casadevall, C., Salleo, A., & Parlak, O. (2019) · Advanced Healthcare Materials · doi:10.1002/adhm.201901321 · Cited by 185 · see citing works
Review—Textile Based Chemical and Physical Sensors for Healthcare Monitoring
Hatamie, A., Angizi, S., Kumar, S., Pandey, C. M., Simchi, A., Willander, M., & Malhotra, B. D. (2020) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ab6827 · Cited by 184 · see citing works
Energy Autonomous Sweat‐Based Wearable Systems
Manjakkal, L., Yin, L., Nathan, A., Wang, J., & Dahiya, R. (2021) · Advanced Materials · doi:10.1002/adma.202100899 · Cited by 170 · see citing works
Cactus‐Spine‐Inspired Sweat‐Collecting Patch for Fast and Continuous Monitoring of Sweat
Son, J., Bae, G. Y., Lee, S., Lee, G., Kim, S. W., Kim, D., Chung, S., & Cho, K. (2021) · Advanced Materials · doi:10.1002/adma.202102740 · Cited by 165 · see citing works
Biospired Janus Silk E-Textiles with Wet–Thermal Comfort for Highly Efficient Biofluid Monitoring
He, X., Fan, C. G., Xu, T., & Zhang, X. (2021) · Nano Letters · doi:10.1021/acs.nanolett.1c03426 · Cited by 165 · see citing works
Soft, thin skin-mounted power management systems and their use in wireless thermography
Lee, J. W., Xu, R., Lee, S. M., Jang, K., Yang, Y., Banks, A. R., Yu, K. J., Kim, J., Xu, S. H., Ma, S., et al. (2016) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1605720113 · Cited by 163 · see citing works
Zhai, Q., Yap, L. W., Wang, R., Gong, S. P., Guo, Z., Liu, Y., Lyu, Q., Wang, J., Simon, G. P., & Cheng, W. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c00274 · Cited by 163 · see citing works
Windows into human health through wearables data analytics
Witt, D. R., Kellogg, R., Snyder, M. P., & Dunn, J. (2019) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2019.01.001 · Cited by 160 · see citing works
Yuan, J., & Zhu, R. (2020) · Applied Energy · doi:10.1016/j.apenergy.2020.115250 · Cited by 160 · see citing works
Wen, F., He, T., Liu, H., Chen, H., Zhang, T., & Lee, C. (2020) · Nano Energy · doi:10.1016/j.nanoen.2020.105155 · Cited by 160 · see citing works
The role of sampling in wearable sweat sensors
Liu, C., Xu, T., Wang, D., & Zhang, X. (2020) · Talanta · doi:10.1016/j.talanta.2020.120801 · Cited by 153 · see citing works
Kim, S., Lee, B.-R., Reeder, J. T., Seo, S. H., Lee, S.-U., Hourlier‐Fargette, A., Shin, J., Sekine, Y., Jeong, H., Oh, Y. S., et al. (2020) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2012700117 · Cited by 153 · see citing works
Wearable Potentiometric Sensors for Medical Applications
Cuartero, M., Parrilla, M., & Crespo, G. A. (2019) · Sensors · doi:10.3390/s19020363 · Cited by 148 · see citing works
Choi, J., Xue, Y., Xia, W. J., Ray, T. R., Reeder, J. T., Bandodkar, A. J., Kang, D., Xu, S., Huang, Y. Y., & Rogers, J. A. (2017) · Lab on a Chip · doi:10.1039/c7lc00525c · Cited by 147 · see citing works
Recent Progress in Wearable Biosensors: From Healthcare Monitoring to Sports Analytics
Ye, S., Feng, S., Huang, L. J., & Bian, S. (2020) · Biosensors · doi:10.3390/bios10120205 · Cited by 146 · see citing works
Kinetic investigation into pH-dependent color of anthocyanin and its sensing performance
Tang, B., He, Y., Liu, J., Zhang, J., Li, J., Zhou, J., Ye, Y., Wang, J., & Wang, X. (2019) · Dyes and Pigments · doi:10.1016/j.dyepig.2019.107643 · Cited by 143 · see citing works
Xu, G., Cheng, C., Yuan, W., Liu, Z., Zhu, L., Li, X., Lu, Y., Chen, Z., Liu, J., Cui, Z., et al. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.126743 · Cited by 141 · see citing works
Gamella, M., Campuzano, S., Manso, J., Rivera, G. G. D., López-Colino, F., Reviejo, A. J., & Pingarrón, J. M. (2014) · Analytica Chimica Acta · doi:10.1016/j.aca.2013.09.020 · Cited by 140 · see citing works
Heo, S. Y., Kim, J., Gutruf, P., Banks, A. R., Wei, P., Pielak, R. M., Balooch, G., Shi, Y., Araki, H., Rollo, D., et al. (2018) · Science Translational Medicine · doi:10.1126/scitranslmed.aau1643 · Cited by 138 · see citing works
Flexible, self-healable, adhesive and wearable hydrogel patch for colorimetric sweat detection
Wang, L., Xu, T., He, X., & Zhang, X. (2021) · Journal of Materials Chemistry C · doi:10.1039/d1tc03905a · Cited by 135 · see citing works
A passive perspiration biofuel cell: High energy return on investment
Yin, L., Moon, J.-M., Sempionatto, J. R., Lin, M., Cao, M., Trifonov, A., Zhang, F., Lou, Z., Jeong, J.-M., Lee, S.-J., et al. (2021) · Joule · doi:10.1016/j.joule.2021.06.004 · Cited by 134 · see citing works
Skin-Patchable Electrodes for Biosensor Applications: A Review
Shetti, N. P., Mishra, A. K., Basu, S., Mascarenhas, R. J., Kakarla, R. R., & Aminabhavi, T. M. (2020) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.9b01659 · Cited by 129 · see citing works
Stress Monitoring and Recent Advancements in Wearable Biosensors
Samson, C., & Koh, A. (2020) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.01037 · Cited by 129 · see citing works
Recent Advances in Electrochemical Sensors for Wearable Sweat Monitoring: A Review
Yeung, K. K., Huang, T., Hua, Y., Zhang, K., Yuen, M. M. F., & Gao, Z. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3074311 · Cited by 128 · see citing works
Soft, skin-interfaced wearable systems for sports science and analytics
Ray, T. R., Choi, J., Reeder, J. T., Lee, S. P., Aranyosi, A. J., Ghaffari, R., & Rogers, J. A. (2019) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2019.01.003 · Cited by 123 · see citing works
Li, G., & Wen, D. (2020) · Journal of Materials Chemistry B · doi:10.1039/c9tb02474c · Cited by 119 · see citing works
Review—Towards Wearable Sensor Platforms for the Electrochemical Detection of Cortisol
Sekar, M. K., Sriramprabha, R., Sekhar, P. K., Bhansali, S., Ponpandian, N., Manickam, P., & Viswanathan, C. (2020) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ab7e24 · Cited by 118 · see citing works
IoT-based Healthcare Monitoring System for War Soldiers using Machine Learning
Gondalia, A., Dixit, D., Parashar, S., Raghava, V., Sengupta, A., & Sarobin, V. R. (2018) · Procedia Computer Science · doi:10.1016/j.procs.2018.07.075 · Cited by 117 · see citing works
A 3D paper-based enzymatic fuel cell for self-powered, low-cost glucose monitoring
Fischer, C., Fraiwan, A., & Choi, S. (2015) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2015.12.020 · Cited by 114 · see citing works
Yang, B., Xiong, Y., Ma, K., Liu, S., & Tao, X. (2020) · EcoMat · doi:10.1002/eom2.12054 · Cited by 107 · see citing works
Near-Field Communication Sensors
Cao, Z., Chen, P., Ma, Z., Li, S., Gao, X., Wu, R., Pan, L., & Shi, Y. (2019) · Sensors · doi:10.3390/s19183947 · Cited by 102 · see citing works
Wearable Sweat Sensors: Background and Current Trends
Kaya, T., Liu, G., Ho, J. C.-T., Yelamarthi, K., Miller, K. C., Edwards, J. E., & Stannard, A. B. (2018) · Electroanalysis · doi:10.1002/elan.201800677 · Cited by 98 · see citing works
Aziz, O., Klenk, J., Schwickert, L., Chiari, L., Becker, C., Park, E. J., Mori, G., & Robinovitch, S. N. (2017) · PLoS ONE · doi:10.1371/journal.pone.0180318 · Cited by 92 · see citing works
Wearable physiological systems and technologies for metabolic monitoring
Gao, W., Brooks, G. A., & Klonoff, D. C. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00407.2017 · Cited by 91 · see citing works
Electrochemical Sensing of Cannabinoids in Biofluids: A Noninvasive Tool for Drug Detection
Klimuntowski, M., Alam, M. M., Singh, G., & Howlader, M. M. R. (2020) · ACS Sensors · doi:10.1021/acssensors.9b02390 · Cited by 88 · see citing works
A review of wearable biosensors for sweat analysis
Jo, S., Sung, D., Kim, S., & Koo, J. (2021) · Biomedical Engineering Letters · doi:10.1007/s13534-021-00191-y · Cited by 84 · see citing works
Annular flexible thermoelectric devices with integrated-module architecture
Qu, D., Huang, X., Li, X., Wang, H., & Chen, G. (2020) · npj Flexible Electronics · doi:10.1038/s41528-020-0064-2 · Cited by 82 · see citing works
Sweat detection theory and fluid driven methods: A review
Yu, H., & Sun, J. (2020) · Nanotechnology and Precision Engineering · doi:10.1016/j.npe.2020.08.003 · Cited by 80 · see citing works
Wearable self-powered biosensors
Reid, R. C., & Mahbub, I. (2019) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2019.10.002 · Cited by 75 · see citing works
Smartphone with optical, physical, and electrochemical nanobiosensors
Seo, S. E., Tabei, F., Park, S. J., Askarian, B., Kim, K. H., Moallem, G., Chong, J. W., & Kwon, O. S. (2019) · Journal of Industrial and Engineering Chemistry · doi:10.1016/j.jiec.2019.04.037 · Cited by 72 · see citing works
Sweat Lactate, Ammonia, and Urea in Rugby Players
Alvear-Ordenes, I., García-López, D., De Paz, J. A., & González-Gallego, J. (2005) · International Journal of Sports Medicine · doi:10.1055/s-2004-830380 · Cited by 67 · see citing works
Challenges and opportunities in wearable technology for biochemical analysis in sweat
McCaul, M. V., Glennon, T., & Diamond, D. (2017) · Current Opinion in Electrochemistry · doi:10.1016/j.coelec.2017.06.001 · Cited by 65 · see citing works
Bhide, A., Muthukumar, S., & Prasad, S. (2018) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2018.06.065 · Cited by 65 · see citing works
Advances in diagnostic microfluidics
Burklund, A., Tadimety, A., Nie, Y., Hao, N., & Zhang, J. X. J. (2020) · Advances in clinical chemistry · doi:10.1016/bs.acc.2019.08.001 · Cited by 63 · see citing works
Machine-learning enabled wireless wearable sensors to study individuality of respiratory behaviors
Chen, A., Zhang, J., Zhao, L., Rhoades, R. D., Kim, D., Wu, N., Liang, J., & Chae, J. (2020) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2020.112799 · Cited by 62 · see citing works
Recent advances of triboelectric nanogenerator based applications in biomedical systems
Xia, X., Liu, Q., Zhu, Y., & Zi, Y. (2020) · EcoMat · doi:10.1002/eom2.12049 · Cited by 61 · see citing works
Tang, Q., Guo, H., Peng, Y., & Hu, C. (2020) · EcoMat · doi:10.1002/eom2.12060 · Cited by 60 · see citing works
Yokus, M. A., & Daniele, M. A.-A. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113249 · Cited by 60 · see citing works
Wang, Z., Li, X., Yang, Z., Guo, H., Tan, Y. J., Susanto, G. J., Cheng, W. M., Yang, W., & Tee, B. C. K. (2021) · EcoMat · doi:10.1002/eom2.12073 · Cited by 59 · see citing works
Fan, R., & Andrew, T. L. (2020) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ab67b0 · Cited by 56 · see citing works
Coronaviruses and people with intellectual disability: an exploratory data analysis
Tummers, J., Catal, C., Tobi, H., Teki̇nerdoğan, B., & Leusink, G. L. (2020) · Journal of Intellectual Disability Research · doi:10.1111/jir.12730 · Cited by 56 · see citing works
Noninvasive glucose detection in exhaled breath condensate
Tankasala, D., & Linnes, J. C. (2019) · Translational research · doi:10.1016/j.trsl.2019.05.006 · Cited by 54 · see citing works
Davis-Martin, R. E., Alessi, S. M., & Boudreaux, E. D. (2021) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2021.642813 · Cited by 54 · see citing works
Wang, R., Du, Z., Xia, Z., Liu, J., Li, P., Wu, Z., Yue, Y., Xiang, Y., Meng, J., Liu, D., et al. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202107682 · Cited by 54 · see citing works
Sweat lactate secretion during exercise in relation to women's aerobic capacity
Lamont, L. S. (1987) · Journal of Applied Physiology · doi:10.1152/jappl.1987.62.1.194 · Cited by 46 · see citing works
Recent Progress in Self-Powered Skin Sensors
Rao, J., Chen, Z., Zhao, D., Yin, Y., Wang, X., & Fang, Y. (2019) · Sensors · doi:10.3390/s19122763 · Cited by 43 · see citing works
Recent advances in flexible sweat glucose biosensors
Zhou, W., He, Q., Ye, H., Ye, C., Wu, X., & Chu, J. (2021) · Journal of Physics D: Applied Physics · doi:10.1088/1361-6463/ac14ef · Cited by 40 · see citing works
Bordbar, M. M., Barzegar, H., Tashkhourian, J., Bordbar, M., & Hemmateenejad, B. (2020) · Analytica Chimica Acta · doi:10.1016/j.aca.2020.10.029 · Cited by 34 · see citing works
Veeralingam, S., Khandelwal, S., & Badhulika, S. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2984807 · Cited by 32 · see citing works
Kanoga, S., Hoshino, T., & Asoh, H. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102817 · Cited by 30 · see citing works
Fully-Conformable Porous Polyethylene Nanofilm Sweat Sensor for Sports Fatigue
Oktavius, A. K., Gu, Q., Wihardjo, N., Winata, O., Sunanto, S. W., Li, J., & Gao, P. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3054989 · Cited by 27 · see citing works
Sweating Out the Circadian Rhythm: A Technical Review
Upasham, S., Churcher, N. K. M., Rice, P., & Prasad, S. (2021) · ACS Sensors · doi:10.1021/acssensors.0c02622 · Cited by 25 · see citing works
Pali, M. L., Jagannath, B., Lin, K., Sankhala, D., Upasham, S., Muthukumar, S., & Prasad, S. (2021) · Bioengineering & Translational Medicine · doi:10.1002/btm2.10241 · Cited by 25 · see citing works
Shitanda, I., & Tsujimura, S. (2021) · Journal of Physics Energy · doi:10.1088/2515-7655/abebcb · Cited by 24 · see citing works
Kwon, J., Park, H. W., Park, Y.-B., & Kim, N. (2017) · International Journal of Precision Engineering and Manufacturing-Green Technology · doi:10.1007/s40684-017-0039-5 · Cited by 21 · see citing works
Quan, T. M., Doike, T., Bui, D. C., Hayashi, K., Arata, S., Kobayashi, A., Islam, M. Z., & Niitsu, K. (2019) · IEEE International Conference on Artificial Intelligence Circuits and Systems (AICAS) · doi:10.1109/aicas.2019.8771604 · Cited by 21 · see citing works
Non-invasive glucose monitoring devices: A review
Mohamad Yunos, M. F. A., & Nordin, A. N. (2020) · Bulletin of Electrical Engineering and Informatics · doi:10.11591/eei.v9i6.2628 · Cited by 20 · see citing works
Sweat-Based in Vitro Diagnostics (IVD): From Sample Collection to Point-of-Care Testing (POCT)
Sarwar, M., Rodriguez, P., & Li, C.-Z. (2019) · Journal of Analysis and Testing · doi:10.1007/s41664-019-00097-w · Cited by 19 · see citing works
WILSON, E. K. (2019) · Engineering · doi:10.1016/j.eng.2019.04.008 · Cited by 19 · see citing works
Wearable pH sensing beyond the Nernst limit
Yang, Y., & Gao, W. F. (2018) · Nature Electronics · doi:10.1038/s41928-018-0166-1 · Cited by 17 · see citing works
Shin, S.-H., & Kim, S.-J. (2021) · Metals · doi:10.3390/met11010150 · Cited by 3 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- Matzeu, G., Florea, L. & Diamond, D. Advances in wearable chemical sensor design for monitoring biological fluids. Sensors Actuators. B Chem. 211, 403–418 (2015).
- Nguyen, B., Coelho, Y., Bastos, T. & Krishnan, S. Trends in human activity recognition with focus on machine learning and power requirements. Mach. Learn. Appl 5, 100072 (2021).
- Wu, H., Zhang, Y., Kjøniksen, A. L., Zhou, X. X. & Zhou, X. X. Wearable Biofuel Cells: Advances from Fabrication to Application. Adv. Funct. Mater. 2103976, 1–38 (2021).
Cited by
Sorted by citations
The emergence of AI-based wearable sensors for digital health technology: a review
Shajari, S., Kuruvinashetti, K., Komeili, A., & Sundararaj, U. (2023) · Sensors · doi:10.3390/s23239498 · Cited by 535 · see citing works
Arya, S. S., Dias, S. B., Jelinek, H. F., Hadjileontiadis, L. J., & Pappa, A. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115387 · Cited by 161 · see citing works
Ma, C., Shang, X., Sun, M., Bo, X., Bai, J., Du, Y., & Zhou, M. (2025) · ACS Sensors · doi:10.1021/acssensors.4c03396 · Cited by 147 · see citing works
3D-printed microneedle-based potentiometric sensor for pH monitoring in skin interstitial fluid
Parrilla, M., Vanhooydonck, A., Johns, M., Watts, R. T., & Wael, K. D. (2022) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2022.133159 · Cited by 113 · see citing works
A Bionic Skin for Health Management: Excellent Breathability, In Situ Sensing, and Big Data Analysis
Shi, S., Ming, Y., Wu, H., Zhi, C., Yang, L., Meng, S., Si, Y., Wang, D., Fei, B., & Hu, J. (2023) · Advanced Materials · doi:10.1002/adma.202306435 · Cited by 111 · see citing works
Smart Skin‐Adhesive Patches: From Design to Biomedical Applications
Wong, S. H. D., Deen, G. R., Bates, J. S., Maiti, C., Lam, C. Y. K., Pachauri, A., AlAnsari, R., Bělský, P., Yoon, J., & Dodda, J. M. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202213560 · Cited by 95 · see citing works
Chen, Y., Li, G., Mu, W., Wan, X., Lu, D., Gao, J., & Wen, D. (2023) · Analytical Chemistry · doi:10.1021/acs.analchem.2c05613 · Cited by 92 · see citing works
Niu, J., Lin, S., Chen, D., Wang, Z., Cao, C., Gao, A., Cui, S., Liu, Y., Hong, Y., Zhi, X., et al. (2023) · Small · doi:10.1002/smll.202306769 · Cited by 81 · see citing works
Farzin, M. A., Naghib, S. M., & Rabiee, N. (2024) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.3c01633 · Cited by 78 · see citing works
Liu, Y., Li, J., Xiao, S., Liu, Y., Bai, M., Gong, L., Zhao, J., & Chen, D. (2023) · Biosensors · doi:10.3390/bios13070726 · Cited by 77 · see citing works
Mo, L., Ma, X., Fan, L., Xin, J. H., & Yu, H. (2022) · Chemical Engineering Journal · doi:10.1016/j.cej.2022.140473 · Cited by 72 · see citing works
Triboelectric Nanogenerator Enabled Sweat Extraction and Power Activation for Sweat Monitoring
Xu, G., Huang, X., Shi, R., Yang, Y., Wu, P., Zhou, J., He, X., Li, J., Zen, Y., Jiao, Y., et al. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202310777 · Cited by 64 · see citing works
Microfluidic wearable electrochemical sweat sensors for health monitoring
Balaji, R., & Liao, Y. (2022) · Biomicrofluidics · doi:10.1063/5.0116648 · Cited by 60 · see citing works
Perspiration permeable, textile embeddable microfluidic sweat sensor
Liu, M., Wang, S., Xiong, Z., Zheng, Z., Ma, N., Li, L., Gao, Q., Ge, C., Wang, Y., & Zhang, T. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115504 · Cited by 52 · see citing works
A Comprehensive Review of the Recent Developments in Wearable Sweat-Sensing Devices
Ibrahim, N. F. A., Sabani, N. B., Johari, S., Manaf, A. A., Wahab, A. A., Zakaria, Z., & Noor, A. M. (2022) · Sensors · doi:10.3390/s22197670 · Cited by 49 · see citing works
Yang, M., Sun, N., Lai, X., Li, Y., Zhao, X., Wu, J., & Zhou, W. (2023) · Micromachines · doi:10.3390/mi14081497 · Cited by 43 · see citing works
Machine learning-powered wearable interface for distinguishable and predictable sweat sensing
Zhou, Z., He, X., Xiao, J., Pan, J., Li, M., Xu, T., & Zhang, X. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116712 · Cited by 40 · see citing works
Jia, S., Gao, H., Xue, Z., & Meng, X. (2022) · Biosensors · doi:10.3390/bios12111057 · Cited by 39 · see citing works
Recent Advances in Two-Dimensional MXene-Based Electrochemical Biosensors for Sweat Analysis
Ganesan, S., Ramajayam, K., Kokulnathan, T., & Palaniappan, A. (2023) · Molecules · doi:10.3390/molecules28124617 · Cited by 37 · see citing works
Sweat‐powered, skin‐adhesive multimodal sensor for long‐term and real‐time sweat monitoring
He, X., Li, Z., Huang, X., Zhang, Q., Zeng, Y., Li, J., Yiu, C. K., Yang, Y., Zhou, J., Xu, G., et al. (2024) · BMEMat · doi:10.1002/bmm2.12124 · Cited by 33 · see citing works
A self-powered wireless sweat-analysis patch for real-time monitoring physiological status
Cao, H., Lin, R., Long, Z., Xing, L., & Xue, X. (2024) · Nano Energy · doi:10.1016/j.nanoen.2024.109411 · Cited by 31 · see citing works
A Step Forward for Smart Clothes─Fabric-Based Microfluidic Sensors for Wearable Health Monitoring
Zhang, T., Ratajczak, A. M., Chen, H., Terrell, J. A., & Chen, C. (2022) · ACS Sensors · doi:10.1021/acssensors.2c01827 · Cited by 31 · see citing works
Ning, Q., Feng, S., Sun, Q., Yu, R., Li, T., Xu, H., Cui, D., & Wang, K. (2023) · Nano Research · doi:10.1007/s12274-023-6197-6 · Cited by 29 · see citing works
Cheng, Z., Wen, Y., Xie, Z., Zhang, M., Feng, Q., Wang, Y., Liu, D., Cao, Y., & Mao, Y. (2024) · Chemical Engineering Journal · doi:10.1016/j.cej.2024.158760 · Cited by 29 · see citing works
Hydrogel-based sensors for multimodal health monitoring: from material design to intelligent sensing
Yu, Y., Liang, X., Ruan, H., Wang, T., Li, Y., & Wen, Z. (2025) · Nanoscale · doi:10.1039/d5nr03553h · Cited by 28 · see citing works
Sensing patches for biomarker identification in skin-derived biofluids
Chen, W., Fu, D., Ma, T., Chen, M., Wang, F., Chen, G., & Wang, Z. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116326 · Cited by 26 · see citing works
Self-powered microfluidic-based sensor for noninvasive sweat analysis
Peringeth, K., Ganguly, A., Pal, A., Chowdhury, J. R., Kaswan, K., Ho, H.-Y., Yu, J., Kao, F., & Lin, Z. (2024) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2024.136859 · Cited by 28 · see citing works
Rao, L. T., Mandal, C. K., & Patolsky, F. (2025) · Advanced Healthcare Materials · doi:10.1002/adhm.202503096 · Cited by 30 · see citing works
Sweat analysis for urea sensing: trends and challenges
Futane, A., Senthil, M., Jayashree, S., Srinivasan, A., Kalpana, R., & Narayanamurthy, V. (2023) · Analytical Methods · doi:10.1039/d3ay01089a · Cited by 25 · see citing works
Advanced Microfluidic‐Based Wearable Electrochemical Sensors for Continuous Biochemical Monitoring
Park, S., Kim, S., Lee, S.-J., Tsukruk, V. V., Park, S., & Lim, H. (2025) · Advanced Electronic Materials · doi:10.1002/aelm.202500010 · Cited by 27 · see citing works
Method for the measurement of triboelectric charge transfer at solid–liquid interface
Chen, Q., Cheng, B., Wang, T., Shang, H., & Shao, T. (2023) · Friction · doi:10.1007/s40544-022-0740-z · Cited by 24 · see citing works
Artificial Intelligence for Risk Assessment on Primary Prevention of Coronary Artery Disease
Chen, S., Loguercio, S., Chen, K., Lee, S. E., Park, J., Liu, S., Sadaei, H. J., & Torkamani, A. (2023) · Current Cardiovascular Risk Reports · doi:10.1007/s12170-023-00731-4 · Cited by 24 · see citing works
Jin, M., Su, P., Huang, X., Zhang, R., He, X., Wang, Z. B., Su, C., Katona, J. M., & Ye, Y. (2024) · Small · doi:10.1002/smll.202311380 · Cited by 23 · see citing works
Advances in Fiber-Based Wearable Sensors for Personal Digital Health Monitoring
Liu, J., Zhang, J., Liu, J., Sun, W., Li, W., Shen, H., Wang, L., & Li, G. (2023) · Materials · doi:10.3390/ma16237428 · Cited by 22 · see citing works
Halbach magnet arrays in electromagnetic kinetic energy harvesters: A review
Asadi, M., Ahmadi, R., & Abazari, A. M. (2024) · Energy Conversion and Management X · doi:10.1016/j.ecmx.2024.100544 · Cited by 21 · see citing works
Future directions: what LIES ahead for smart biochemical wearables in health monitoring?
Kharb, S. (2025) · Frontiers in Analytical Science · doi:10.3389/frans.2024.1509815 · Cited by 19 · see citing works
Hao, Z., Fang, X., Wang, Z., Huang, C., Li, F., Luo, Y., Ying, Y., Ma, S., Jia, Y., Wang, X., et al. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202400947 · Cited by 19 · see citing works
Yang, M., Sun, N., Lai, X., Zhao, X., & Zhou, W. (2023) · Biosensors · doi:10.3390/bios14010017 · Cited by 18 · see citing works
Novel Sweat-Based Wearable Device for Advanced Monitoring of Athletic Physiological Biometrics
Aguilar-Torán, J., Rabost-Garcia, G., Toinga-Villafuerte, S., Álvarez-Carulla, A., Colmena-Rubil, V., Fajardo-Garcia, A., Cardona-Bonet, A., Casals‐Terré, J., Muñoz-Pascual, X., Miribel‐Català, P. L., et al. (2023) · Sensors · doi:10.3390/s23239473 · Cited by 15 · see citing works
Bio-inspired hierarchical wearable patch for fast sweat collection
Wu, T., Yang, P., Xie, X., Cao, X., Deng, Y., Ding, X., & Zhang, Z. (2024) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2024.116430 · Cited by 14 · see citing works
Self-citationHydrogel tissue adhesive: Adhesion strategy and application
An, B., Cui, H., Wang, M., Li, Z., & Li, J. (2025) · Colloids and Surfaces B Biointerfaces · doi:10.1016/j.colsurfb.2025.114755 · Cited by 14 · see citing works
Recent Advances in Wearable Sweat Sensor Development
Zhang, T., Kabandana, G. K. M., Terrell, J. A., Chen, H., & Chen, C. (2025) · Wiley Interdisciplinary Reviews Nanomedicine and Nanobiotechnology · doi:10.1002/wnan.70006 · Cited by 13 · see citing works
Jin, J., Won, J., Kim, D., Arumugasamy, S. K., Park, S., & Kim, T.-W. (2026) · Nano-Micro Letters · doi:10.1007/s40820-025-02056-w · Cited by 10 · see citing works
A Battery-Less and Low-Cost RFID Sensor for Unassisted Multilevel Detection of Sweat Loss
López-Soriano, S., Vilajosana, X., & Melià‐Seguí, J. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3485592 · Cited by 8 · see citing works
Sweat as a diagnostic biofluid: analytical advances and future directions
Bordin, D. M., McCauley, J. I., de Campos, E. G., Bishop, D., & Martinis, B. S. D. (2025) · Journal of Pharmaceutical Analysis · doi:10.1016/j.jpha.2025.101473 · Cited by 9 · see citing works
Wang, R., Yang, X., Tian, F., Zhu, Z., Teng, P., Zhang, Y., Wen, X., Li, X., Guan, A., Wang, Y., et al. (2025) · Optics & Laser Technology · doi:10.1016/j.optlastec.2025.112975 · Cited by 7 · see citing works
Sardesai, A. U., Greyling, C. F., Lin, K., Kumar, R. M., Muthukumar, S., & Prasad, S. (2023) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2023.100377 · Cited by 7 · see citing works
Yang, M., Cheng, J., Zhao, X., Cai, J., & Zhou, W. (2024) · Analytical Sciences · doi:10.1007/s44211-024-00705-5 · Cited by 6 · see citing works
Zhu, X., Tang, X., Ng, C. A., Li, L., Lai, Y., Miller, G. W., Jiang, C., Barchowsky, A., Buchanich, J. M., & Gao, P. (2025) · Reviews of Environmental Contamination and Toxicology · doi:10.1007/s44169-025-00088-2 · Cited by 6 · see citing works
Yan, Y., Xiao, T., Lin, M., Yue, W., Qu, J., Chen, Y., Zhang, Z., Meng, J., Pan, D., Li, F., et al. (2025) · Biosensors · doi:10.3390/bios15120790 · Cited by 5 · see citing works
Self-citationNovel and rapid analytical platform development enabled by advances in 3D printing
Malinick, A. S., Ebel, C. P., Stuart, D. D., Valiulis, S. N., Hanson, V. A., & Cheng, Q. J. (2024) · Frontiers in Analytical Science · doi:10.3389/frans.2024.1505510 · Cited by 5 · see citing works
Su, M., Zhu, L., Zhou, Y., Ding, Z., Yao, Y., Li, M., Qiu, Q., Peng, W., Gao, H., & Yu, C. (2025) · Analytical Chemistry · doi:10.1021/acs.analchem.5c01954 · Cited by 5 · see citing works
Machine Learning-Enabled Biosensors in Clinical Decision Making
Verma, S., Shukla, R. P., & Dutta, G. (2022) · Springer eBooks · doi:10.1007/978-981-19-7130-3_7 · Cited by 6 · see citing works
Tamilselvi, T., Lakshmi, D., Lavanya, R., & Revathi, K. (2023) · International Journal on Recent and Innovation Trends in Computing and Communication · doi:10.17762/ijritcc.v11i3.6331 · Cited by 4 · see citing works
Wearable sweat induction strategies for noninvasive biosensing
He, Y., Qiao, Y., Shang, J., Gan, H., Zhu, X., He, X., Zhang, H., & Qigong, L. (2025) · Talanta · doi:10.1016/j.talanta.2025.129294 · Cited by 3 · see citing works
From lab to field: progress and prospect of sweat sensing technology in sports monitoring
Gao, N., Deng, X., Lei, Y., Jiang, P., & Wu, Y. (2026) · Microchimica Acta · doi:10.1007/s00604-026-07835-9 · Cited by 3 · see citing works
Su, M., Zhou, A., Chen, Z., Li, M., Yao, Y., Ding, Z., Zhou, Y., Qiu, Q., & Yu, C. (2026) · Microchemical Journal · doi:10.1016/j.microc.2026.117726 · Cited by 2 · see citing works
Shen, H., Pang, Y., Yuan, D., Zhang, J., Guo, H., Wang, C., He, H., Ling, M., Chen, L., & Huang, H. (2024) · Advanced Materials Technologies · doi:10.1002/admt.202301803 · Cited by 2 · see citing works
Stretchable and bioadhesive sensing platform for stable decoding of epidermal sweat information
Wang, H., Zhou, B., Li, Y., Feng, D., Shi, M., Zhang, W., Li, J., & Shang, X. (2025) · Microchemical Journal · doi:10.1016/j.microc.2025.115758 · Cited by 1 · see citing works
Zhang, Y., Zhong, B., Yuan, Y., Xu, K., Yu, Y., Lou, Z., & Wang, L. (2026) · Advanced Materials Technologies · doi:10.1002/admt.202502662 · Cited by 2 · see citing works
Ion sensing in body fluids: Recent progress and perspectives
Ling, H., Yan, J., Han, J., (294193), J. C., Guo, J., & Li, D. (2026) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2026.118955 · Cited by 1 · see citing works
Wearable Health Monitoring System for Sweat Analysis
Pagar, V., Bhadane, P. K., & Shaligram, A. D. (2024) · Communications in computer and information science · doi:10.1007/978-3-031-54547-4_31 · Cited by 1 · see citing works
Wang, J., Wang, S., & Zhang, J. (2025) · Flexible and Printed Electronics · doi:10.1088/2058-8585/adb8fa · Cited by 1 · see citing works
State of the Art of Wearable Sensors for Healthcare
Conti, M., Orcioni, S., Giuliani, R., Gabrielli, L., Gaiduk, M., Madrid, N. M., & Seepold, R. (2022) · Lecture notes in bioengineering · doi:10.1007/978-3-031-16855-0_6
Wearable Health Monitoring System Based on Human Sweat Analysis using Electrochemical Sensors
Pagar, V., Bhadane, P. K., & Shaligram, A. D. (2023) · Romanian Journal of Biophysics · doi:10.59277/rjb.2023.4.02
Sharmila, V. C., Varun, B., Josh, P. A. A., Atmuri, A., & Prasanth, K. (2024) · International Conference on Mobile Networks and Wireless Communications (ICMNWC) · doi:10.1109/icmnwc63764.2024.10872139
Alsultan, M. A., López-Soriano, S., & Melià‐Seguí, J. (2025) · European Conference on Antennas and Propagation · doi:10.23919/eucap63536.2025.11000019
Structure design and application analysis of sweat detection device based on bionic microchannel
Liu, H., Lin, G., Li, W., & Shi, Y. (2025) · Microchemical Journal · doi:10.1016/j.microc.2025.114738
Affective Computing—Student Emotion Recognition
Zhang, C., & Chen, C. (2026) · Emerald Group Publishing Limited eBooks · doi:10.1108/978-1-80686-209-220261007
Laser-Induced Breakdown Spectroscopy as a Tool for the Detection of Trace Elements in Sweat
Sreeja, K., V, P, S. T., & Anoop, K. K. (2026) · Journal of Condensed Matter · doi:10.61343/jcm.v4i01.218
A High-Sensitivity Flexible Sweat pH Sensing Patch Based on a Screen-Printed PANI Sensing Layer
Guo, C., Dang, Z., Yang, H., Yang, S., Wen, D., Li, Y., Liu, H., & He, Y. (2026) · Journal of Macromolecular Science Part B · doi:10.1080/00222348.2026.2710125
Wei, X., Tao, J., Wang, X., 解巧凤, Li, Y., Zheng, B., Dong, J., & Zheng, H. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3721337
Li, Y., Tian, Y., Luo, J., Wang, J., Cheng, J., Suo, C., Wang, C., Li, Z., Li, H., & Chen, X. (2026) · Advanced Composites and Hybrid Materials · doi:10.1007/s42114-026-02067-0
Qi, Q., Duan, J., Lei, Y., Zhou, S., Liu, Q. S., Wang, Q., Gao, N., Sun, W., Zhan, Z., & Wu, Y. (2026) · Biosensors · doi:10.3390/bios16090523
Triboelectric sweat sensors for wearable health monitoring
Li, X., He, H., Liang, R., Ye, Z., Yu, K., Luo, Q., Zhang, S., Cai, C., Liu, T., Wang, L., et al. (2026) · Matter · doi:10.1016/j.matt.2026.103047
Cao, H., Zhuang, X., Zhu, G., Mu, T., & Ma, Y. (2026) · Chemosensors · doi:10.3390/chemosensors14100231
Zhang, C., Liang, A., Dong, W., Chen, X., Liu, G., & Zhou, M. (2026) · Chemical Engineering Journal · doi:10.1016/j.cej.2026.183248
- Authors
- Yang, P., Wei, G., Liu, A., Huo, F., & Zhang, Z.
- Affiliations
- Shanghai Jiao Tong University (CN)
- Second Military Medical University (CN)
- UNSW Sydney (AU)
- Nanjing Tech University (CN)
- Corresponding author
- Wei, Gaofeng
- Venue
- npj Flexible Electronics
- Published
- 2022-06-03
- Volume
- 6
- Issue
- 1
- ISSN
- 2397-4621 (ISSN-L)
- OpenAlex
- W4282053000
- Cited by
- 77
- References
- 126
- Citations by year
- 2026: 14 · 2025: 17 · 2024: 17 · 2023: 21 · 2022: 8
- FWCI
- 4.58
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- npj Flexible Electronics (publishedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- State Key Laboratory of Mechanical System and Vibration (MSVZD202108)
- he open project of National Research Center for Translational Medicine,·Shanghai
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Keywords
- sweat sensors, long-term monitoring, sweat collection, energy supply, wearable health monitoring, artificial intelligence, flexible electronics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review