Non‐invasive Analyte Access and Sensing through Eccrine Sweat: Challenges and Outlook circa 2016
Abstract
Despite the many ergonomic advantages of eccrine perspiration (sweat) compared to other possible biofluids (particularly in “wearable” devices), sweat remains an underrepresented source of biomarker analytes compared to the established biofluids blood, urine, and saliva. Upon closer comparison to other non‐invasive biofluids, the advantages may even extend beyond ergonomics: sweat might provide superior analyte information. A number of challenges, however, have historically kept sweat from its place in the pantheon of clinical samples. These challenges include very low sample volumes (nL to µL), unknown concentration due to evaporation, filtration and dilution of large analytes, mixing of old and new sweat, and the potential for contamination from the skin surface. More recently, rapid progress in “wearable” sweat sampling and sensing devices has resolved several of the historical challenges. However, this recent progress has also been limited to high concentration analytes (µM to mM) sampled at high sweat rates (>1 nL/min/gland, e.g. athletics). Progress will be much more challenging as sweat biosensing moves towards use with sedentary users (low sweat rates or not sweating at all) and/or towards low concentration analytes (pM to nM). Addressing these unresolved challenges will require significant advances in sweat stimulation, sample collection efficiency, compact sensors, and likely more. Fortunately, none of the remaining challenges appear to be fundamentally blocking, and scientific and engineering innovations have the opportunity to enable broader application of sweat biosensing technology.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
A critical turning point occurred in 2016 when Gao published a fully integrated array sweat sensor in Nature that simultaneously and selectively measured sweat metabolites and electrolytes without external analysis.
“A critical turning points occurred in 2016, when Gao 56 published research on a fully integrated array sweat sensor in Nature that simultaneously and selectively measured sweat metabolites and electrolytes without requiring external analysis 57 .”
From A review of sampling, energy supply and intelligent monitoring for long-term sweat sensors (2022) · cites this paper as [57]
Sweat provides an ideal source toward the continuous and noninvasive measurement of biomarkers, such as sodium, chlorine, potassium, lactate, calcium, glucose, ammonia, ethanol, urea, cortisol, and various neuropeptides and cytokines.
“Sweat provides an ideal source toward the continuous and noninvasive measurement of biomarkers, such as sodium, chlorine, potassium, lactate, calcium, glucose, ammonia, ethanol, urea, cortisol, and various neuropeptides and cytokines.18”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [18]
Disadvantages of using eccrine sweat include skin contamination, existence of dried sweat on the glands thereby skewing analyte measurements, and low-sampling rates.
“Disadvantages of using eccrine sweat, include skin contamination, existence of dried sweat on the glands thereby skewing analyte measurements, and low-sampling rates.18”
From Wearable sensors for monitoring the physiological and biochemical profile of the athlete (2019) · cites this paper as [18]
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
Non-invasive wearable electrochemical sensors: a review
Bandodkar, A. J., & Wang, J. (2014) · Trends in Biotechnology · doi:10.1016/j.tibtech.2014.04.005 · Cited by 1,247 · see citing works
Sonner, Z., Wilder, E., Heikenfeld, J., Kasting, G., Beyette, F., Swaile, D., Sherman, F., Joyce, J., Hagen, J., Kelley-Loughnane, N., et al. (2015) · Biomicrofluidics · doi:10.1063/1.4921039 · Cited by 724 · see citing works
Self-citationFolding-Based Electrochemical Biosensors: The Case for Responsive Nucleic Acid Architectures
Lubin, A. A., & Plaxco, K. W. (2010) · Accounts of Chemical Research · doi:10.1021/ar900165x · Cited by 506 · see citing works
Derler, S., & Gerhardt, L. ‐. (2011) · Tribology Letters · doi:10.1007/s11249-011-9854-y · Cited by 428 · see citing works
Adhesive RFID Sensor Patch for Monitoring of Sweat Electrolytes
Rose, D. P., Ratterman, M. E., Griffin, D. K., Hou, L., Kelley-Loughnane, N., Naik, R. R., Hagen, J. A., Papautsky, I., & Heikenfeld, J. C. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2369991 · Cited by 416 · see citing works
Self-citationThe measurement of hormones in saliva: Possibilities and pitfalls
VINING, R. F. W., & McGinley, R. A. (1987) · Journal of Steroid Biochemistry · doi:10.1016/0022-4731(87)90297-4 · Cited by 338 · see citing works
A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration
Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D. (2010) · Analytical Methods · doi:10.1039/b9ay00184k · Cited by 286 · see citing works
Stretchable, Wireless Sensors and Functional Substrates for Epidermal Characterization of Sweat
Huang, X. L., Liu, Y., Chen, K., Shin, W., Lu, C., Kong, G., Patnaik, D., Lee, S., Cortes, J. F., & Rogers, J. A. (2014) · Small · doi:10.1002/smll.201400483 · Cited by 282 · see citing works
Sweat: A sample with limited present applications and promising future in metabolomics
Mena-Bravo, A., & de Castro, M. D. L. (2013) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2013.10.048 · Cited by 237 · see citing works
Raiszadeh, M. M., Ross, M. M., Russo, P. S., Schaepper, M. A., Zhou, W., Deng, J., Ng, D., Dickson, A., Dickson, C., Strom, M., et al. (2012) · Journal of Proteome Research · doi:10.1021/pr2007957 · Cited by 142 · see citing works
On-line sample pre-concentration in microfluidic devices: A review
Giordano, B. C., Burgi, D. S., Hart, S. J., et al. (2012) · Analytica Chimica Acta · doi:10.1016/j.aca.2011.12.050 · Cited by 127 · see citing works
Csősz, É., Emri, G., Kalló, G., Tsaprailis, G., & Tőzsér, J. (2015) · Journal of the European Academy of Dermatology and Venereology · doi:10.1111/jdv.13221 · Cited by 72 · see citing works
MIP-based biomimetic sensor for the electronic detection of serotonin in human blood plasma
Peeters, M., Troost, F. J., van Grinsven, B., Horemans, F., Alenus, J., Murib, M. S., Keszthelyi, D., Ethirajan, A., Thoelen, R., Cleij, T. J., et al. (2012) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2012.05.040 · Cited by 72 · see citing works
Nanoparticle Technology: Addressing the Fundamental Roadblocks to Protein Biomarker Discovery
Luchini, A., Fredolini, C., Espina, B., et al. (2010) · Current Molecular Medicine · doi:10.2174/156652410790963268 · Cited by 61 · see citing works
Nanoduct® sweat testing for rapid diagnosis in newborns, infants and children with cystic fibrosis
Desax, M.-C., Ammann, R. A., Hammer, J., Schoeni, M. H., & Barben, J. (2007) · European Journal of Pediatrics · doi:10.1007/s00431-007-0485-0 · Cited by 56 · see citing works
Artificial microfluidic skin for in vitro perspiration simulation and testing
Hou, L., Hagen, J. A., Wang, X., Papautsky, I., Naik, R. K., Kelley‐Loughnane, N., & Heikenfeld, J. C. (2013) · Lab on a Chip · doi:10.1039/c3lc41231h · Cited by 55 · see citing works
Self-citationMonitoring Pregnant Women's Illicit Opiate and Cocaine Use With Sweat Testing
Brunet, B., Barnes, A. J., Choo, R. E., Mura, P., Jones, H. E., & Huestis, M. A. (2010) · Therapeutic Drug Monitoring · doi:10.1097/ftd.0b013e3181c13aaf · Cited by 38 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Bariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486 · see citing works
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434 · see citing works
Self-citationA soft, wearable microfluidic device for the capture, storage, and colorimetric sensing of sweat
Koh, A., Kang, D., Xue, Y., Lee, S., Pielak, R. M., Kim, J., Hwang, T., Min, S., Banks, A., Bastien, P., et al. (2016) · Science Translational Medicine · doi:10.1126/scitranslmed.aaf2593 · Cited by 1,308 · see citing works
Accessing analytes in biofluids for peripheral biochemical monitoring
Heikenfeld, J., Jajack, A., Feldman, B., Granger, S. W., Gaitonde, S., Begtrup, G., & Katchman, B. A. (2019) · Nature Biotechnology · doi:10.1038/s41587-019-0040-3 · Cited by 758 · see citing works
Self-citationBandodkar, A. J., Gutruf, P., Choi, J., Lee, K., Sekine, Y., Reeder, J. T., Jeang, W. J., Aranyosi, A. J., Lee, S. P., Model, J. B., et al. (2019) · Science Advances · doi:10.1126/sciadv.aav3294 · Cited by 761 · see citing works
Parlak, O., Keene, S. T., Marais, A., Curto, V. F., & Salleo, A. (2018) · Science Advances · doi:10.1126/sciadv.aar2904 · Cited by 603 · see citing works
“Among many bodily fluids, sweat provides a significant amount of information about a patient’s health status and is readily accessible, making it suitable for wearable, noninvasive biosensing (15).”
BackgroundWearable sensors for monitoring the physiological and biochemical profile of the athlete
Seshadri, D. R., Li, R. T., Voos, J. E., Rowbottom, J. R., Alfes, C. M., Zorman, C. A., & Drummond, C. K. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0150-9 · Cited by 537 · see citing works
From pre-diabetes to diabetes: diagnosis, treatments and translational research
Khan, R., Chua, Z., Tan, J., Yang, Y., Liao, Z., & Zhao, Y. (2019) · Medicina · doi:10.3390/medicina55090546 · Cited by 484 · see citing works
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475 · see citing works
Roll-to-Roll Gravure Printed Electrochemical Sensors for Wearable and Medical Devices
Bariya, M., Shahpar, Z., Park, H., Sun, J., Jung, Y., Gao, W., Nyein, H. Y. Y., Liaw, T. S., Tai, L.-C., Ngo, Q. P., et al. (2018) · ACS Nano · doi:10.1021/acsnano.8b02505 · Cited by 417 · see citing works
Wearable Sensors for Biochemical Sweat Analysis
Bandodkar, A. J., Jeang, W. J., Ghaffari, R., & Rogers, J. A. (2019) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061318-114910 · Cited by 387 · see citing works
Lipani, L., Dupont, B. G. R., Doungmene, F., Marken, F., Tyrrell, R. M., Guy, R. H., & Ilie, A. (2018) · Nature Nanotechnology · doi:10.1038/s41565-018-0112-4 · Cited by 340 · see citing works
Hong, Y. J., Lee, H., Kim, J., Lee, M., Choi, H. J., Hyeon, T., & Kim, D. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201805754 · Cited by 229 · see citing works
Wearable Biosensors for Non-Invasive Sweat Diagnostics
Xu, J., Fang, Y., & Chen, J. (2021) · Biosensors · doi:10.3390/bios11080245 · Cited by 208 · see citing works
A wearable patch for continuous monitoring of sweat electrolytes during exertion
Alizadeh, A., Burns, A., Lenigk, R., Gettings, R., Ashe, J., Porter, A., McCaul, M., Barrett, R., Diamond, D., White, P., et al. (2018) · Lab on a Chip · doi:10.1039/c8lc00510a · Cited by 180 · see citing works
Hauke, A., Simmers, P., Ojha, Y. R., Cameron, B. D., Ballweg, R., Zhang, T., Twine, N., Brothers, M., Gomez, E., & Heikenfeld, J. (2018) · Lab on a Chip · doi:10.1039/c8lc01082j · Cited by 156 · see citing works
Self-citationWearable Devices for Precision Medicine and Health State Monitoring
Jeong, I. C., Bychkov, D., & Searson, P. C. (2019) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2871638 · Cited by 155 · see citing works
ZnO Nanorod Integrated Flexible Carbon Fibers for Sweat Cortisol Detection
Madhu, S., Anthuuvan, A. J., Ramasamy, S., Manickam, P., Bhansali, S., Nagamony, P., & Chinnuswamy, V. (2020) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.9b00730 · Cited by 102 · see citing works
Wearable Potentiometric Chloride Sweat Sensor: The Critical Role of the Salt Bridge
Choi, D.-H., Kim, J. S., Cutting, G. R., & Searson, P. C. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03391 · Cited by 100 · see citing works
Wearable physiological systems and technologies for metabolic monitoring
Gao, W., Brooks, G. A., & Klonoff, D. C. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00407.2017 · Cited by 91 · see citing works
Wearable sweat sensing for prolonged, semicontinuous, and nonobtrusive health monitoring
Moonen, E. J. M., Haakma, J. R., Peri, E., Pelssers, E., Mischi, M., & den Toonder, J. M. J. (2020) · VIEW · doi:10.1002/viw.20200077 · Cited by 86 · see citing works
A review of sampling, energy supply and intelligent monitoring for long-term sweat sensors
Yang, P., Wei, G., Liu, A., Huo, F., & Zhang, Z. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00165-9 · Cited by 77 · see citing works
Ultra-simple wearable local sweat volume monitoring patch based on swellable hydrogels
Zhao, F. J., Bonmarin, M., Chen, Z. C., Larson, M., Fay, D., Runnoe, D., & Heikenfeld, J. (2020) · Lab on a Chip · doi:10.1039/c9lc00911f · Cited by 71 · see citing works
Self-citationCheng, S., Gu, Z., Zhou, L., Hao, M., An, H., Song, K., Wu, X., Zhang, K., Zhao, Z., Dong, Y., et al. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.765987 · Cited by 64 · see citing works
Open nanofluidic films with rapid transport and no analyte exchange for ultra-low sample volumes
Twine, N. B., Norton, R. M., Brothers, M. C., Hauke, A., Gomez, E. F., & Heikenfeld, J. (2018) · Lab on a Chip · doi:10.1039/c8lc00186c · Cited by 43 · see citing works
Self-citationEnhancing glucose flux into sweat by increasing paracellular permeability of the sweat gland
Jajack, A., Brothers, M., Kasting, G., & Heikenfeld, J. (2018) · PLOS ONE · doi:10.1371/journal.pone.0200009 · Cited by 35 · see citing works
Self-citation
- Authors
- Heikenfeld, J.
- Affiliations
- University of Cincinnati (US)
- Corresponding author
- Heikenfeld, Jason
- ORCID
- Heikenfeld, Jason
- Venue
- Electroanalysis
- Publisher
- Wiley
- Published
- 2016-03-21
- Volume
- 28
- Issue
- 6
- Pages
- 1242-1249
- ISSN
- 1040-0397 (ISSN-L)
- 1521-4109
- OpenAlex
- W2302659947
- Cited by
- 294
- References
- 18
- Citations by year
- 2026: 20 · 2025: 19 · 2024: 31 · 2023: 31 · 2022: 20 · 2021: 41 · 2020: 33 · 2019: 35 · 2018: 41 · 2017: 17 · 2016: 6
- FWCI
- 15.22
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Electroanalysis (publishedVersion)
- Funders
- Univ. of Cincinnati provided by the National Science Foundation (1347725)
- Air Force Research Labs 711th Human Performance Wing
- USAF (FA8650-15-C-6625)
- Center for Advanced Design and Manufacturing of Integrated Microfluidics (IIP-1362165)
- Univ. of Cincinnati College of Pharmacy
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
- Keywords
- eccrine sweat, wearable sensors, sweat sampling, low-concentration analytes, sweat biomarkers
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review