Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units
Open accessThe DOI matches its Crossref record; 54 of the 55 references it cites were found in a registry or an index; 19 of the 45 sources its claims cite were checked for support.
Abstract
Standard clinical care in neonatal and pediatric intensive-care units (NICUs and PICUs, respectively) involves continuous monitoring of vital signs with hard-wired devices that adhere to the skin and, in certain instances, can involve catheter-based pressure sensors inserted into the arteries. These systems entail risks of causing iatrogenic skin injuries, complicating clinical care and impeding skin-to-skin contact between parent and child. Here we present a wireless, non-invasive technology that not only offers measurement equivalency to existing clinical standards for heart rate, respiration rate, temperature and blood oxygenation, but also provides a range of important additional features, as supported by data from pilot clinical studies in both the NICU and PICU. These new modalities include tracking movements and body orientation, quantifying the physiological benefits of skin-to-skin care, capturing acoustic signatures of cardiac activity, recording vocal biomarkers associated with tonality and temporal characteristics of crying and monitoring a reliable surrogate for systolic blood pressure. These platforms have the potential to substantially enhance the quality of neonatal and pediatric critical care. Soft electronic patches worn on the skin of infants or children in intensive-care units have a wide range of capabilities in aiding critical care, including monitoring of hemodynamic parameters, cardiac activity, movement and crying.
Overview
CoChat Summary
TL;DR
A pair of soft, skin-interfaced wireless biosensors was developed for neonatal and pediatric intensive care, with pilot studies showing clinical-quality vital signs plus added modalities such as movement/orientation, kangaroo-care tracking, seismocardiography, cry analysis, and a systolic blood pressure surrogate. The system was tested in 50 NICU/PICU patients across a wide age range, with comparisons to standard clinical instruments.
Key findings
- In NICU/PICU pilot studies, the wireless system measured HR, SpO2, temperature, and RR with close agreement to standard monitoring: mean differences were −0.02 bpm for HR, 0.11% for SpO2, and 0.21 °C for temperature (n = 3); RR mean difference was 0.11 breaths per minute with SD 1.95 (n = 6, 41 data points).
- SpO2 accuracy root mean square was 2.99%, and a larger comparison included over 0.4 million aggregated HR/SpO2 data points from n = 20 patients, within FDA regulatory guidelines for HR and reflectance-mode SpO2.
- Time-synchronized chest and limb units enabled PAT- and PTT-derived systolic blood pressure estimates versus arterial-line measurements, with mean differences of 1.60 and −0.04 mmHg and SDs of 7.99 and 7.86 mmHg, respectively (n = 5), within ANSI/AAMI SP10 limits.
- Additional modalities included kangaroo-care orientation/activity tracking, SCG, and cry analysis: KC events showed phi/theta of 2.85 ± 0.10 rad and −0.29 ± 0.28 rad over 2.4 h; activity was 0.07 ± 0.02 g/s during rest/KC versus 0.24 ± 0.05 g/s during hands-on care; 11 cry events in n = 3 infants showed duration difference −3.9 ± 13.9 s and fundamental frequency 410.7 ± 47.9 Hz.
Methods
- Design: two mechanically soft, wireless devices—a chest unit for ECG, accelerometry, temperature, and acoustic/vibration signals, and a limb unit for reflectance PPG and temperature—using Bluetooth Low Energy, with modular/embedded/battery-free power options.
- Data/sample: pilot clinical studies in NICU and PICU patients, including 50 patients for safety/skin tolerability across 23–40 weeks gestational age and 1 week–4 years chronological age; vital-sign comparisons used smaller subsets (e.g., n = 3 for HR/SpO2/temperature, n = 6 for RR, n = 20 for aggregated HR/SpO2, n = 5 for BP).
- Analysis: Bland-Altman comparisons against standard clinical monitors; time-synchronized PAT/PTT calibrated against arterial-line SBP; accelerometer filtering for KC orientation/activity and cry detection; MANOVA for position/activity comparisons.
Limitations
- The paper states that in instances of low SpO2 (<90%), mean disagreement was higher than the overall comparison, likely reflecting different SpO2 de-noising algorithms rather than a true measurement difference.
- The paper notes that comparisons reflect real-world vital-sign calculations during clinical care, feeding, imaging, and other procedures, so motion artifact and noise are inherent and processed differently between the experimental system and clinical comparator.
- Evident from the text: key validation cohorts were small (e.g., n = 3 for HR/SpO2/temperature, n = 5 for BP, n = 3 for KC/cry), thermal safety testing used a subset of 3 patients over 24 h, and BP estimation required initial calibration against 5 min of arterial-line SBP data.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures






Verification
Claims the paper makes
9 Valid · 9 Not valid · 13 Not checked · 63 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Ultimately, both PAT and PTT depend on vascular system geometry, elasticity, SBP, and other factors30,33,34.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Time-synchronized Bi-nodal Communication for Non-invasive Blood Pressure Monitoring
What the sources say
- Partly supportedcites [30] Ma Y, Relation between blood pressure and pulse wave velocity for human arteries. PNAS 115, 11144– 11149 (2018). 30322935 PMC6217416 sourceExperimental — accuracy not yet measured
The In Vitro Experiments section relates pulse transit time to pulse wave velocity and reports pulse arrival times measured with ECG and PPG. It also states that the pressure–velocity relation depends on arterial material properties and geometry. The reported human measurements concern diastolic blood pressure, however, so the paper does not establish the stated dependence on SBP for both PAT and PTT. (LLM, conf=0.84)
- Not matchedcites [33] Kangaroo mother care: a practical guide, (World Health Organization, Dept. of Reproductive Health and Research, 2003).
Its entry has no title and year that we can search by.
- Not checked — no full text of the sourcecites [34] Liu G-Z, Guo Y-W, Zhu Q-S, Huang B-Y & Wang L Estimation of respiration rate from three-dimensional acceleration data based on body sensor network. Telemed. J. E-Health 17, 705– 711 (2011). 22035321 PMC3203837 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Cry analysis can serve as a non-invasive method to analyze the neurophysiological state, often influenced by birth trauma, brain injury or pain stress13.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Advanced Use Cases: Kangaroo Care and Cry Analysis
What the sources say
- Partly supportedcites [13] Shinya Y, Kawai M, Niwa F, Imafuku M & Myowa M Fundamental frequency variation of neonatal spontaneous crying predicts language acquisition in preterm and term infants. Front. Psychol. 8, 2195 (2017). 29312060 PMC5744644 sourceExperimental — accuracy not yet measured
The Introduction (page not specified) describes cry acoustics as a non-invasive index of neonatal neurophysiological states, including pain stress, and links abnormal cry frequency to neurological disturbances. It does not specifically identify birth trauma or brain injury as influences, so those examples are not supported by our copy. (LLM, conf=0.88)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The fundamental frequency of 410.7 ± 47.9 (Supplementary Fig.24e) is consistent with published results46.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Advanced Use Cases: Kangaroo Care and Cry Analysis
What the sources say
- Could not verifycites [46] Inan Omer T, Novel wearable seismocardiography and machine learning algorithms can assess clinical status of heart failure patients. Circ. Heart Fail. 11, e004313 (2018). 29330154 PMC5769154 sourceExperimental — accuracy not yet measured
The Methods (Pre-Processing and Feature Extraction; page unavailable) discuss frequency components of cardiac seismocardiography signals, not crying. Searches for crying, vocal pitch, and the reported frequency found no passage in our copy addressing whether 410.7 ± 47.9 is consistent with published results. (LLM, conf=0.77)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Continuous monitoring of vital signs is essential, yet existing technologies require the use of multiple electrodes and interfaces attached to the skin with adhesive tapes and connected by wires to electronic processing systems that are often tethered to the wall.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Abstract
What the sources say
- Partly supportedcites [3] Bonner O, Beardsall K, Crilly N & Lasenby J ‘ There were more wires than him’: the potential for wireless patient monitoring in neonatal intensive care. BMJ Innov. 3, 12– 18 (2017). PMC5293857 28250963 sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) describes multiple vital-sign sensors in a neonatal intensive care unit connected by wires to a monitoring unit outside the incubator. The Results (page unavailable) describe adhesive ECG and temperature probes, but say the pulse oximeter is secured with Velcro. The paper does not establish that the monitoring units are tethered to a wall or that all the interfaces use adhesive tape. (LLM, conf=0.87)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The data show that skin-to-skin contact during KC produced a pronounced, gradual increase in the peripheral skin temperature, consistent with expectation and as demonstrated in previous studies39.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Advanced Use Cases: Kangaroo Care and Cry Analysis
What the sources say
- Could not verifycites [39] Huntsman RJ, Lowry NJ & Sankaran K Nonepileptic motor phenomena in the neonate. Paediatr. Child Health 13, 680– 684 (2008). 19436521 PMC2606074 sourceExperimental — accuracy not yet measured
The paper’s passages on neonatal tremor and jitteriness (section labeled Abstract; page unavailable) do not address Kangaroo Care or peripheral skin temperature. Searches for skin-to-skin contact, thermoregulation, and a gradual temperature increase found no passage supporting the attributed finding. (LLM, conf=0.78)
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
45 more claims
Create a free account to see all 94 claims and to verify the 2 references still open.
- It may state two separate claims as one. 39
- Probably not a substantial claim. 10
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [19]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [22]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Additionally, as demonstrated by our previous work, neonatal vital signs can also be tracked with the same sensors in both high-resource and low-resource settings as well as support kangaroo care (22, 48).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [22]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous wearables have been designed to condense the complicated setups in NICU monitoring into miniaturized wearables for facilitated monitoring4–6; however, they encounter problems with device longevity with either single-use or short-term use before replacement.”
From Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia (2025) · cites this paper as [5]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [5]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
Scheer, B., Perel, A., & Pfeiffer, U. J. (2002) · Critical Care · doi:10.1186/cc1489 · Cited by 854
Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Self-citationKangaroo Mother Care and Neonatal Outcomes: A Meta-analysis
Boundy, E. O., Dastjerdi, R., Spiegelman, D., Fawzi, W. W., Missmer, S. A., Lieberman, E., Kajeepeta, S., Wall, S., & Chan, G. J. (2016) · Pediatrics · doi:10.1542/peds.2015-2238 · Cited by 661
Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure
Payne, R. A., Symeonides, C. N., Webb, D. J., & Maxwell, S. R. J. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00657.2005 · Cited by 418
Liu, Y., Norton, J. J. S., Qazi, R., Zou, Z., Ammann, K. R., Liu, H., Yan, L., Tran, P. L., Jang, K.-I., Lee, J. W., et al. (2016) · Science Advances · doi:10.1126/sciadv.1601185 · Cited by 397
Self-citationXu, J., Murphy, S. L., Kochanek, K. D., Bastian, B. A., & Arias, E. (2018) · PubMed · Cited by 391
Radial Artery Cannulation: A Comprehensive Review of Recent Anatomic and Physiologic Investigations
Brzezinski, M., Luisetti, T., & London, M. J. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181bbd416 · Cited by 354
Research works printed but not linked: 1
- Wheeler DS, Wong HR & Shanley TP Pediatric critical care medicine: basic science and clinical evidence, (Springer, London, 2007).
Cited by
Items 101–200 of 476
Multimodal Wearable Sensor Sheet for Health-Related Chemical and Physical Monitoring
Hozumi, S., Honda, S., Arie, T., Akita, S., & Takei, K. (2021) · ACS Sensors · doi:10.1021/acssensors.1c00281 · Cited by 80
Materials design and integration strategies for soft bioelectronics in digital healthcare
Kim, H. J., Koo, J. H., Lee, S., Hyeon, T., & Kim, D. (2025) · Nature Reviews Materials · doi:10.1038/s41578-025-00819-w · Cited by 136
Self-citationCarbon nanotube-based flexible high-speed circuits with sub-nanosecond stage delays
Long, G., Jin, W., Xia, F., Wang, Y., Bai, T., Chen, X., Liang, X., Peng, L., & Hu, Y. (2022) · Nature Communications · doi:10.1038/s41467-022-34621-x · Cited by 74
Lin, S., Hu, S., Song, W., Gu, M., Liu, J., Song, J., Liu, Z., Li, Z., Huang, K., Wu, Y., et al. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00158-8 · Cited by 71
Kwak, S. S., Yoo, S., Avila, R. O., Chung, H. U., Jeong, H., Liu, C. Y., Vogl, J. L., Kim, J., Yoon, H., Park, Y., et al. (2021) · Advanced Materials · doi:10.1002/adma.202103974 · Cited by 74
Self-citationDong, J., Hou, J., Peng, Y., Zhang, Y., Liu, H., Long, J., Park, S., Liu, T. X., & Huang, Y. (2024) · Advanced Materials · doi:10.1002/adma.202409071 · Cited by 87
Self-citationTough and Highly Efficient Underwater Self‐Repairing Hydrogels for Soft Electronics
Fu, R., Guan, Y., Xiao, C., Fan, L., Wang, Z., Li, Y., Yu, P., Tu, L., Tan, G., Zhai, J., et al. (2022) · Small Methods · doi:10.1002/smtd.202101513 · Cited by 67
Challenges and prospects of visual contactless physiological monitoring in clinical study
Huang, B., Hu, S., Liu, Z., Lin, C., Su, J., Zhao, C., Wang, L., & Wang, W. (2023) · npj Digital Medicine · doi:10.1038/s41746-023-00973-x · Cited by 92
“5, the current trend in the development of neonatal physiological signal monitoring instruments is shifting from wired contact to wireless contact monitoring 118 – 120 , and ultimately towards contactless measurement.”
BackgroundFlexible, wearable biosensors for digital health
Zhu, P., Peng, H., & Rwei, A. Y. (2022) · Medicine in Novel Technology and Devices · doi:10.1016/j.medntd.2022.100118 · Cited by 82
Self-citationHighly Sensitive, Stretchable, and Robust Strain Sensor Based on Crack Propagation and Opening
Wu, S., Moody, K., Kollipara, A., & Zhu, Y. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c16741 · Cited by 74
Application and potential of artificial intelligence in neonatal medicine
Kwok, T. C., Henry, C., Saffaran, S., Meeus, M., Bates, D. G., Laere, D. V., Boylan, G. B., Boardman, J. P., & Sharkey, D. (2022) · Seminars in Fetal and Neonatal Medicine · doi:10.1016/j.siny.2022.101346 · Cited by 91
Yang, A., Song, J., Liu, H., Zhao, Z., Li, L., & Yan, F. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202215037 · Cited by 74
Skin-interfaced wireless biosensors for perinatal and paediatric health
Kim, J., Yoo, S., Liu, C. Y., Kwak, S. S., Walter, J. R., Xu, S., & Rogers, J. A. (2023) · Nature Reviews Bioengineering · doi:10.1038/s44222-023-00090-0 · Cited by 74
Self-citationYokus, M. A., & Daniele, M. A.-A. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113249 · Cited by 66
Fabric computing: Concepts, opportunities, and challenges
Chen, M., Liu, J., Li, P., Gharavi, H., Hao, Y., Ouyang, J., Hu, J., Hu, L., Hou, C., Humar, I., et al. (2022) · The Innovation · doi:10.1016/j.xinn.2022.100340 · Cited by 77
Liu, C., Kim, J., Kwak, S. S., Hourlier‐Fargette, A., Avila, R., Vogl, J., Tzavelis, A., Chung, H. U., Lee, J. Y., Kim, D. H., et al. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100383 · Cited by 57
Self-citationLiu, Z., Li, S., Lin, S., Shi, Y., Yang, P., Chen, X., & Wang, Z. L. (2022) · Nano Letters · doi:10.1021/acs.nanolett.2c00767 · Cited by 58
Decoding tissue biomechanics using conformable electronic devices
Yoon, H., Kim, J., Sadat, D., Barrett, A., Ko, S. H., & Dağdeviren, C. (2024) · Nature Reviews Materials · doi:10.1038/s41578-024-00729-3 · Cited by 80
Self-citationNear-Field Communication in Biomedical Applications
Kang, S.-G., Song, M.-S., Kim, J., Lee, J. W., & Kim, J. (2021) · Sensors · doi:10.3390/s21030703 · Cited by 66
Self-citationIntelligent fibers and textiles for wearable biosensors
Fu, W., Yu, B., Ji, D., Zhou, Z., Li, X., Wang, R., Lu, W., Sun, Y., & Dai, Y. (2024) · Responsive materials · doi:10.1002/rpm.20240018 · Cited by 78
Xu, S., Rwei, A. Y., Vwalika, B., Chisembele, M. P., Stringer, J. S. A., Ginsburg, A. S., & Rogers, J. A. (2021) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00001-7 · Cited by 57
Self-citationFlexible near-infrared organic photodetectors for emergent wearable applications
Wang, Z., Cheng, S., Fukuda, K., Hu, W., Xu, X., & Someya, T. (2024) · Wearable electronics. · doi:10.1016/j.wees.2024.04.001 · Cited by 63
Wearable Sensors, Data Processing, and Artificial Intelligence in Pregnancy Monitoring: A Review
Liu, L., Pu, Y., Fan, J., Yan, Y., Liu, W., Luo, K., Wang, Y., Zhao, G., Chen, T. P., Poenar, D. P., et al. (2024) · Sensors · doi:10.3390/s24196426 · Cited by 86
“Another important study by Chung et al. [53] delineated a pioneering approach in the realm of neonatal and pediatric care through the deployment of skin-interfaced biosensors for wireless physiological monitoring as shown in Figure 10.”
BackgroundGhanim, R. W., Kaushik, A., Park, J., & Abramson, A. (2023) · Device · doi:10.1016/j.device.2023.100092 · Cited by 61
Self-citationResponsive materials and mechanisms as thermal safety systems for skin-interfaced electronic devices
Yoo, S., Yang, T., Park, M., Jeong, H., Lee, Y. J., Cho, D., Kim, J., Kwak, S. S., Shin, J., Park, Y., et al. (2023) · Nature Communications · doi:10.1038/s41467-023-36690-y · Cited by 63
Self-citationZhuang, Q., Yao, K., Song, X., Zhang, Q., Zhang, C., Wang, H., Yang, R., Zhao, G., Li, S., Shu, H., et al. (2025) · Science Advances · doi:10.1126/sciadv.adu3146 · Cited by 66
“Nevertheless, existing wearable technologies suffer from unad-dressed gaps in clinical translations and adoptions ( 31 – 33 ), and the achievement of ICU-grade accuracy is rare ( 20 , 21 , 34 ).”
He, X., Cheng, J., Li, Z., Ye, H., Wei, X., Li, H., Wang, R., Zhang, Y., Yang, H. Y., Guo, C. F., et al. (2022) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c18954 · Cited by 54
Real-time personal healthcare data analysis using edge computing for multimodal wearable sensors
Matsumura, G., Honda, S., Kikuchi, T., Mizuno, Y., Hara, H., Kondo, Y., Nakamura, H., Watanabe, S., Hayakawa, K., Nakajima, K., et al. (2024) · Device · doi:10.1016/j.device.2024.100597 · Cited by 75
Self-powered wearable electronics
Tan, P., Zou, Y., Fan, Y., & Li, Z. (2020) · Wearable Technologies · doi:10.1017/wtc.2020.3 · Cited by 61
Li, J., Chu, H., Chen, Z., Yiu, C. K., Qu, Q., Li, Z.-Y., & Yu, X. (2024) · ACS Nano · doi:10.1021/acsnano.4c04291 · Cited by 51
Convertino, V. A., Schauer, S. G., Weitzel, E. K., Cardin, S., Stackle, M. E., Talley, M. J., Sawka, M. N., & Inan, O. T. (2020) · Sensors · doi:10.3390/s20226413 · Cited by 54
“Recent developments in device fabrication have allowed PPG sensing systems to be flexible and skin-interfaced for comfortable use in long-term care scenarios [81,82].”
BackgroundSkin-interfacing wearable biosensors for smart health monitoring of infants and neonates
Zhou, L., Guess, M., Kim, K. R., & Yeo, W. (2024) · Communications Materials · doi:10.1038/s43246-024-00511-6 · Cited by 64
“Ina clinical test on 50 neonates observing changes in skin condition (erythema, dryness, or breakdown) after 15 minutes of EES application, the average change was negligible to slightly improved 17 .”
InfluentialSelf‐Healing Materials for Bioelectronic Devices
Liu, C. Y., Kelley, S. O., & Wang, Z. (2024) · Advanced Materials · doi:10.1002/adma.202401219 · Cited by 77
Self-citationLu, D., Li, S., Yang, Q., Arafa, H., Xu, Y., Yan, Y., Ostojich, D., Bai, W., Guo, H., Wu, C., et al. (2022) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2022.114145 · Cited by 51
Self-citationZhang, G., Li, P., Wang, X., et al. (2022) · Advanced Functional Materials · doi:10.1002/adfm.202208900 · Cited by 50
Recent Advances in Bioinspired Hydrogels: Materials, Devices, and Biosignal Computing
Zhu, Y., Haghniaz, R., Hartel, M. C., Mou, L., Tian, X., Garrido, P. R., Wu, Z., Hao, T., Guan, S., Ahadian, S., et al. (2021) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.1c00741 · Cited by 57
Self-citationWireless Battery-free and Fully Implantable Organ Interfaces
Bhatia, A., Hanna, J., Stuart, T., Kasper, K. A., Clausen, D. M., & Gutruf, P. (2024) · Chemical Reviews · doi:10.1021/acs.chemrev.3c00425 · Cited by 54
Hauer, L., Naga, A., Badr, R. G. M., Pham, J. T., Wong, W. S. Y., & Vollmer, D. (2024) · Soft Matter · doi:10.1039/d4sm00346b · Cited by 58
Recent progress in flexible tactile sensor systems: from design to application
Zhu, J., Zhou, C., & Zhang, M. (2021) · Soft Science · doi:10.20517/ss.2021.02 · Cited by 60
Flexible and Stretchable Ultrasonic Transducer Array Conformed to Complex Surfaces
Liu, W., Zhu, C., & Wu, D. (2020) · IEEE Electron Device Letters · doi:10.1109/led.2020.3045037 · Cited by 47
Chen, S., Ouyang, Q., Meng, X., Yang, Y., Li, C., Miao, X.-B., Chen, Z., Zhao, G., Lei, Y., Ghanem, B., et al. (2025) · Science Advances · doi:10.1126/sciadv.adv2406 · Cited by 58
Anti‐Motion Artifacts Iontronic Sensor for Long‐Term Accurate Fingertip Pulse Monitoring
You, J., Lu, M., Dazhen, L., Gao, M., Zhang, R., Li, W., Lei, F., Ren, W., Li, G., & Yang, J. (2025) · Advanced Science · doi:10.1002/advs.202414425 · Cited by 50
Evolvable Skin Electronics by In Situ and In Operando Adaptation
Kim, K. K., Choi, J., Kim, J., Nam, S., & Ko, S. H. (2021) · Advanced Functional Materials · doi:10.1002/adfm.202106329 · Cited by 47
Self-citationAdvances in olfactory augmented virtual reality towards future metaverse applications
Zhang, Z., Guo, X., & Lee, C. (2024) · Nature Communications · doi:10.1038/s41467-024-50261-9 · Cited by 56
Recent progress of skin-integrated electronics for intelligent sensing
Li, D., Yao, K., Gao, Z., Liu, Y., & Yu, X. (2021) · Light Advanced Manufacturing · doi:10.37188/lam.2021.004 · Cited by 52
Marks, H. L., Bucknor, A. E. M., Roussakis, E., Nowell, N. H., Kamali, P., Cascales, J. P., Kazei, D. D., Lin, S. J., & Evans, C. L. (2020) · Science Advances · doi:10.1126/sciadv.abd1061 · Cited by 43
“Recently, skin interfaced wireless sensors for wearable (%stO 2 ) monitoring in neonates have been pioneered by Chung et al. ( 19 , 20 ) and show great promise for overcoming the typical setbacks of continuous (%stO 2 ) monitoring, once made cost effective and widely available.”
BackgroundRecent advances in passive cooling materials for thermal management in flexible electronics
Li, J., Zhou, Y., Jiang, C., Lei, D., & Yu, X. (2024) · Journal of Materials Chemistry C · doi:10.1039/d4tc02523g · Cited by 51
Materials and device strategies to enhance spatiotemporal resolution in bioelectronics
Zhang, J., Cheng, Z., Li, P., & Tian, B. (2025) · Nature Reviews Materials · doi:10.1038/s41578-025-00798-y · Cited by 71
“…m Days to months Tracking embryonic development and physiological changes with wearable bioelectronic sensors a,293,294 Flexible polymer sensors 294 , wireless fetal monitoring systems 293 , electroactive polymers 295 Plant growth 1 mm–30 m Days to years Using imperceptible bioelectronic…”
Fully Photonic Integrated Wearable Optical Interrogator
Li, H., An, Z., Zhang, S., Zuo, S., Zhu, W., Zhang, S., Huang, B., Cao, L., Zhang, C., Zhang, Z., et al. (2021) · ACS Photonics · doi:10.1021/acsphotonics.1c01236 · Cited by 43
Advanced thermal sensing techniques for characterizing the physical properties of skin
Madhvapathy, S. R., Arafa, H., Patel, M. J., Winograd, J., Kong, J., Zhu, J., Xu, S., & Rogers, J. A. (2022) · Applied Physics Reviews · doi:10.1063/5.0095157 · Cited by 48
Self-citationFlexible Electronics and Healthcare Applications
Wang, Z., Sun, L., Ni, Y., Liu, L., & Xu, W. (2021) · Frontiers in Nanotechnology · doi:10.3389/fnano.2021.625989 · Cited by 44
“Flexible transistors, flexible capacitors and flexible sensors are typical flexible electronic components; all have been integrated into complex circuits for applications in health care (Ray et al., 2019; Chung et al., 2020; Hanna et al., 2020; Jin et al., 2020; Ma et al., 2020; Madhvapathy et al., 2020; Sun et al., 2…”
BackgroundSpatiotemporal hemodynamic monitoring via configurable skin-like microfiber Bragg grating group
Zhu, H., Luo, J., Dai, Q., Zhu, S., Yang, H., Zhou, K., Zhan, L., Xu, B., Chen, Y., Lu, Y., et al. (2023) · Opto-Electronic Advances · doi:10.29026/oea.2023.230018 · Cited by 40
Liu, C. Y., Kim, J., Yang, D. S., Cho, D., Yoo, S., Madhvapathy, S. R., Jeong, H., Yang, T., Luan, H., Avila, R. O., et al. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202302256 · Cited by 47
Self-citationZeng, T., Wu, Y., & Lei, M. (2024) · Advanced Composites and Hybrid Materials · doi:10.1007/s42114-024-00949-9 · Cited by 50
E-Skin and Its Advanced Applications in Ubiquitous Health Monitoring
Sun, X., Guo, X., Gao, J., Wu, J., Huang, F., Zhang, J., Huang, F., Lu, X. J., Shi, Y., & Pan, L. (2024) · Biomedicines · doi:10.3390/biomedicines12102307 · Cited by 49
“…the neonate’s Guo et al. achieved accurate and rapid measurement of applied mechanical pressure and recognition of multiple movement patterns of the infant by combining sensors for infant care with the assistance of artificial intelligence and deep learning algorithms [204]. condition (Figure 16).”
BackgroundMethodInfluentialWearable devices for patient monitoring in the intensive care unit
Angelucci, A., Greco, M., Cecconi, M., & Aliverti, A. (2025) · Intensive Care Medicine Experimental · doi:10.1186/s40635-025-00738-8 · Cited by 60
“A study conducted in a newborn and pediatric population [18] assessed a soft skin-interfaced biosensor employing advanced wireless technology for the monitoring of several vital parameters in neonatal and pediatric ICUs.”
Self-Powered Wireless Flexible Ionogel Wearable Devices
Li, W., Lin, K., Chen, L., Yang, D., Ge, Q., & Wang, Z. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c19744 · Cited by 54
Li, Q., Liu, S., Wang, J., Mbola, N. M., Meng, Z., Wang, X., & Xue, M. (2022) · Journal of Materials Chemistry C · doi:10.1039/d2tc01310j · Cited by 40
Mamun, M. M. R. K., & Sherif, A. (2022) · Bioengineering · doi:10.3390/bioengineering10010027 · Cited by 44
“Like smartphones, smartwatches and fitness trackers are equipped with advanced sensor technologies and analysis capabilities to provide alerts for different health conditions [115–118].”
BackgroundMiniaturized retractable thin-film sensor for wearable multifunctional respiratory monitoring
Li, C., Xu, Z., Xu, S., Wang, T., Zhou, S., Sun, Z., Wang, Z. L., & Tang, W. (2023) · Nano Research · doi:10.1007/s12274-023-5420-1 · Cited by 39
Self-citationA perspective on flexible sensors in developing diagnostic devices
Wang, L., Jiang, K., & Shen, G. (2021) · Applied Physics Letters · doi:10.1063/5.0057020 · Cited by 36
Huttunen, O., Behfar, M. H., Hiitola‐Keinänen, J., & Hiltunen, J. A. (2022) · Advanced Materials Technologies · doi:10.1002/admt.202101496 · Cited by 44
Biocompatible Electronic Skins for Cardiovascular Health Monitoring
Du, Y., Kim, J. H., Kong, H., Li, A. A., Jin, M., Kim, D. H., & Wang, Y. (2024) · Advanced Healthcare Materials · doi:10.1002/adhm.202303461 · Cited by 43
Self-citationLighting the Path to Precision Healthcare: Advances and Applications of Wearable Photonic Sensors
Song, R., Cho, S., Khan, S. A., Park, I., & Gao, W. F. (2025) · Advanced Materials · doi:10.1002/adma.202419161 · Cited by 54
Vogl, J. L., Dunne, E. C., Liu, C. Y., Bradley, A., Rwei, A. Y., Lonergan, E. K., Hopkins, B. S., Kwak, S. S., Simon, C. D., Rand, C. M., et al. (2021) · Developmental Psychobiology · doi:10.1002/dev.22100 · Cited by 42
Self-citationBiointerfaced sensors for biodiagnostics
Zohar, O., Khatib, M., Omar, R., Vishinkin, R., Broza, Y. Y., & Haick, H. (2021) · View · doi:10.1002/viw.20200172 · Cited by 37
Wang, H., Wang, J., Chen, D., Song, G., Liu, Y., Wang, Z., Zhang, X., Guo, Q., & Yang, J. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3141457 · Cited by 38
Early brain activity: Translations between bedside and laboratory
Luhmann, H. J., Kanold, P. O., Molnár, Z., & Vanhatalo, S. (2022) · Progress in Neurobiology · doi:10.1016/j.pneurobio.2022.102268 · Cited by 35
Huang, Q., Wang, J., Li, C., Zhu, J., Wang, W., Huang, Y., Zhang, Y., Jiao, H., Zhang, S., Meng, H., et al. (2022) · npj Flexible Electronics · doi:10.1038/s41528-022-00190-8 · Cited by 40
Self-citationFlexible and Ultra‐Sensitive Planar Supercapacitive Pressure Sensor Based on Porous Ionic Foam
Tang, Y.-D., Wang, P., Li, G., Wang, G., Yu, W., Meng, C., & Guo, S. (2022) · Advanced Engineering Materials · doi:10.1002/adem.202200814 · Cited by 36
Transfer printing technologies for soft electronics
Huang, Z., & Lin, Y. (2022) · Nanoscale · doi:10.1039/d2nr04283e · Cited by 35
Soft electronics for advanced infant monitoring
Yuan, M., Long, Y., Liu, T., Liu, J., Qiu, S., Lin, T. Y., Xu, F., & Fang, Y. (2024) · Materials Today · doi:10.1016/j.mattod.2024.03.005 · Cited by 35
Encapsulated stretchable amphibious strain sensors
Wu, S., Kim, D., Tang, X., King, M. W., & Zhu, Y. (2024) · Materials Horizons · doi:10.1039/d4mh00757c · Cited by 39
Self-citationTaccola, S., Poliziani, A., Santonocito, D., Mondini, A., Denk, C., Ide, A. N., Oberparleiter, M., Greco, F., & Mattoli, V. (2021) · Sensors · doi:10.3390/s21041197 · Cited by 40
Flexible organic photodetectors and their use in wearable systems
Guo, H., Saifi, S., Fukuda, K., Cheng, H., Lou, Z., & Xu, X. (2021) · Digital Signal Processing · doi:10.1016/j.dsp.2021.103145 · Cited by 37
Jia, H., Gao, Y., Zhou, J., Li, J., Yiu, C. K., Park, W., Yang, Z., & Yu, X. (2024) · Nano Energy · doi:10.1016/j.nanoen.2024.109796 · Cited by 36
Anisotropic hydrogel microelectrodes for intraspinal neural recordings in vivo
Huang, S., Xiao, R., Lin, S., Wu, Z., Lin, C., Jang, G., Hong, E., Gupta, S. A., Lu, F., Chen, B., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-56450-4 · Cited by 51
Biomimetic electroactive materials and devices for regenerative engineering
Liu, Y., Bai, Y., Heng, B. C., Wang, Y., Zhang, X., Nguyen, T. D., & Deng, X. (2025) · Nature Reviews Electrical Engineering · doi:10.1038/s44287-025-00145-x · Cited by 45
Vialess heterogeneous skin patch for multimodal monitoring and stimulation
Lee, H., Song, S., Yea, J., Ha, J., Oh, S., Jekal, J., Hong, M. S., Won, C., Jung, H. H., Keum, H., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-55951-6 · Cited by 48
Advances in Biosignal Sensing and Signal Processing Methods with Wearable Devices
Matthews, J., Kim, J., & Yeo, W. (2022) · Analysis & Sensing · doi:10.1002/anse.202200062 · Cited by 42
Self-citationAI‐Based Metamaterial Design for Wearables
Yigci, D., Ahmadpour, A., & Taşoğlu, S. (2023) · Advanced Sensor Research · doi:10.1002/adsr.202300109 · Cited by 35
Nazir, A., Nazir, A., Jamal, M. S. W., Sadiq, S. U. R., Aman, S., Mustapha, M. J., Lawal, S. O., Abdulkareem, M. O., & Mustapha, F. B. (2025) · Health Science Reports · doi:10.1002/hsr2.71486 · Cited by 48
Wang, Y., Li, Z., & Hu, Q. (2021) · Nano Today · doi:10.1016/j.nantod.2021.101127 · Cited by 38
Basodan, R. A. M., Park, B., & Chung, H. (2021) · Flexible and Printed Electronics · doi:10.1088/2058-8585/ac32a9 · Cited by 36
Self-citationUltraflexible tattoo electrodes for epidermal and in vivo electrophysiological recording
Wei, B., Wang, Z., Guo, H., Xie, F., Cheng, S., Lou, Z., Zhou, C., Ji, H., Zhang, M., Wang, X., et al. (2023) · Cell Reports Physical Science · doi:10.1016/j.xcrp.2023.101335 · Cited by 36
Qin, M., Bian, Y., Wang, C.-Y., Sun, J., Shi, W., Liu, K., Zheng, Y., Zhang, F., Liu, G., Shao, M., et al. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202403770 · Cited by 31
Liu, Z., Chen, J., Jo, M., Vogel, V., Chen, J., Rogers, J. A., & Li, S. (2026) · Nature Reviews Bioengineering · doi:10.1038/s44222-025-00391-6 · Cited by 54
Self-citationNear‐Field Communication Powered Hydrogel‐Based Smart Contact Lens
Quintero, A. V., Sekiguchi-Arai, R., Yamazaki, Y., Satô, T., & Smet, H. D. (2020) · Advanced Materials Technologies · doi:10.1002/admt.202000702 · Cited by 28
Thermoplastic Elastomers for Wireless, Skin‐Interfaced Electronic, and Microfluidic Devices
Wu, Y., Liu, C. Y., Lapiere, M., Ciatti, J. L., Yang, D. S., Berkovich, J., Model, J. B., Banks, A. R., Ghaffari, R., Chang, J., et al. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202300732 · Cited by 34
Self-citationReduced Graphene Oxide Tattoo as Wearable Proximity Sensor
Kedambaimoole, V., Kumar, N., Shirhatti, V., Nuthalapati, S., Kumar, S., Nayak, M. M., Sen, P., Akinwande, D., & Rajanna, K. (2021) · Advanced Electronic Materials · doi:10.1002/aelm.202001214 · Cited by 28
Energy Harvesting Untethered Soft Electronic Devices
Kim, K. K., Choi, J., & Ko, S. H. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202002286 · Cited by 30
Recent Advances in High-Throughput Nanomaterial Manufacturing for Hybrid Flexible Bioelectronics
Zavanelli, N., Kim, J., & Yeo, W. (2021) · Materials · doi:10.3390/ma14112973 · Cited by 32
Self-citationPiezoelectric approaches for wearable continuous blood pressure monitoring: a review
Yi, Z., Zhang, W., & Yang, B. T. (2022) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/ac87ba · Cited by 31
Pathway of transient electronics towards connected biomedical applications
Dutta, A., & Cheng, H. (2023) · Nanoscale · doi:10.1039/d2nr06068j · Cited by 27
Liquid Metal‐Based High‐Density Interconnect Technology for Stretchable Printed Circuits
Wang, B., Prasad, S., Hellman, O., Li, H., Fridberger, A., & Hjort, K. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202309707 · Cited by 32
Flexible hybrid electronics: Enabling integration techniques and applications
Wu, H., Huang, Y., & Yin, Z. (2022) · Science China Technological Sciences · doi:10.1007/s11431-022-2074-8 · Cited by 32
Self-citationMachine learning-assisted flexible wearable device for tyrosine detection
Bao, Q., Li, G., Cheng, W., Yang, Z., Qu, Z., Wei, J., & Lin, L. (2023) · RSC Advances · doi:10.1039/d3ra02900j · Cited by 25
Du, Y., Zhou, C., Feng, Y., & Qiu, L. (2024) · npj Flexible Electronics · doi:10.1038/s41528-024-00373-5 · Cited by 33
Smart Bioelectronics for Real-Time Diagnosis and Therapy of Body Organ Functions
Guo, L., Lee, H. K., Oh, S., Koirala, G. R., & Kim, T. (2025) · ACS Sensors · doi:10.1021/acssensors.5c00024 · Cited by 24
Wireless Qi-Powered, Multinodal and Multisensory Body Area Network for Mobile Health
Dautta, M., Jimenez, A., Dia, K. K. H., Rashid, N., Faruque, M. A. A., & Tseng, P. (2020) · IEEE Internet of Things Journal · doi:10.1109/jiot.2020.3040713 · Cited by 27
Metadata
- Authors
- Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., Ryu, D., Kulikova, E., Cao, J., Odland, I. C., Fields, K. B., Hopkins, B., Banks, A., Ogle, C., Grande, D., Park, J. B., Kim, J., Irie, M., Jang, H., Lee, J., Park, Y., Kim, J., Jo, H. H., Hahm, H., Avila, R., Xu, Y., Namkoong, M., Kwak, J. W., Suen, E., Paulus, M. A., Kim, R. J., Parsons, B. V., Human, K. A., Kim, S. S., Patel, M., Reuther, W., Kim, H. S., Lee, S. H., Leedle, J. D., Yun, Y., Rigali, S., Son, T., Jung, I., Arafa, H., Soundararajan, V. R., Ollech, A., Shukla, A., Bradley, A., Schau, M., Rand, C. M., Marsillio, L. E., Harris, Z. L., Huang, Y., Hamvas, A., Paller, A. S., Weese-Mayer, D. E., Lee, J. Y., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- Lurie Children's Hospital (US)
- Dalian University of Technology (CN)
- Sibel (United States) (US)
- University of Illinois Urbana-Champaign (US)
- North Carolina State University (US)
- Kyung Hee University (KR)
- Corresponding author
- Lee, Jong Yoon
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Rwei, Alina Y.
- Hourlier-Fargette, Aurélie
- Xu, Shuai
- Lee, KunHyuck
- Dunne, Emma C.
- Xie, Zhaoqian
- Liu, Claire
- Carlini, Andrea
- Kim, Dong Hyun
- Cao, Jingyue
- Odland, Ian C.
- Banks, Anthony
- Grande, Dominic
- Park, Jun Bin
- Kim, Jongwon
- Jang, Hokyung
- Park, Yerim
- Jo, Han Heul
- Avila, Raudel
- Xu, Yeshou
- Namkoong, Myeong
- Kwak, Jean Won
- Human, Kelia A.
- Patel, Manish
- Reuther, William
- Lee, Sung Hoon
- Yun, Yeojeong
- Jung, Inhwa
- Arafa, Hany
- Ollech, Ayelet
- Rand, Casey M.
- Marsillio, Lauren E.
- Harris, Zena L.
- Huang, Yonggang
- Hamvas, Aaron
- Paller, Amy S.
- Weese-Mayer, Debra E.
- Lee, Jong Yoon
- Rogers, John A.
- Venue
- Nature Medicine
- Published
- 2020-03
- Volume
- 26
- Issue
- 3
- Pages
- 418-429
- ISSN
- 1078-8956 (ISSN-L)
- 1546-170X
- PMC
- PMC7315772
- OpenAlex
- W3012163988
- Semantic Scholar
- 9be51710140598d05f71429da3bc79c90db9cace
- Cited by
- 482
- References
- 55
- Citations by year
- 2026: 49 · 2025: 77 · 2024: 76 · 2023: 85 · 2022: 77 · 2021: 89 · 2020: 29
- FWCI
- 21.94
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- cc-by
- www.springer.com/tdm (tdm)
- www.springer.com/tdm (vor)
- Locations
- Nature Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion, cc-by)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909, OPP1193311)
- Gerber Foundation
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- wireless physiological monitoring, neonatal intensive care, pediatric intensive care, non-invasive vital sign monitoring, skin-to-skin care, vocal biomarkers, crying
- MeSH terms
- Child, Child, Preschool, Electrocardiography, Equipment Design, Humans, Infant, Newborn, Intensive Care Units, Neonatal, Monitoring, Physiologic, Skin, Time Factors, Intensive Care Units, Pediatric, Biosensing Techniques, Photoplethysmography, Blood Pressure Monitoring, Ambulatory, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.