Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units
Open accessThe DOI matches its Crossref record; 54 of the 55 references it cites were found in a registry or an index; 19 of the 45 sources its claims cite were checked for support.
Abstract
Standard clinical care in neonatal and pediatric intensive-care units (NICUs and PICUs, respectively) involves continuous monitoring of vital signs with hard-wired devices that adhere to the skin and, in certain instances, can involve catheter-based pressure sensors inserted into the arteries. These systems entail risks of causing iatrogenic skin injuries, complicating clinical care and impeding skin-to-skin contact between parent and child. Here we present a wireless, non-invasive technology that not only offers measurement equivalency to existing clinical standards for heart rate, respiration rate, temperature and blood oxygenation, but also provides a range of important additional features, as supported by data from pilot clinical studies in both the NICU and PICU. These new modalities include tracking movements and body orientation, quantifying the physiological benefits of skin-to-skin care, capturing acoustic signatures of cardiac activity, recording vocal biomarkers associated with tonality and temporal characteristics of crying and monitoring a reliable surrogate for systolic blood pressure. These platforms have the potential to substantially enhance the quality of neonatal and pediatric critical care. Soft electronic patches worn on the skin of infants or children in intensive-care units have a wide range of capabilities in aiding critical care, including monitoring of hemodynamic parameters, cardiac activity, movement and crying.
Overview
CoChat Summary
TL;DR
A pair of soft, skin-interfaced wireless biosensors was developed for neonatal and pediatric intensive care, with pilot studies showing clinical-quality vital signs plus added modalities such as movement/orientation, kangaroo-care tracking, seismocardiography, cry analysis, and a systolic blood pressure surrogate. The system was tested in 50 NICU/PICU patients across a wide age range, with comparisons to standard clinical instruments.
Key findings
- In NICU/PICU pilot studies, the wireless system measured HR, SpO2, temperature, and RR with close agreement to standard monitoring: mean differences were −0.02 bpm for HR, 0.11% for SpO2, and 0.21 °C for temperature (n = 3); RR mean difference was 0.11 breaths per minute with SD 1.95 (n = 6, 41 data points).
- SpO2 accuracy root mean square was 2.99%, and a larger comparison included over 0.4 million aggregated HR/SpO2 data points from n = 20 patients, within FDA regulatory guidelines for HR and reflectance-mode SpO2.
- Time-synchronized chest and limb units enabled PAT- and PTT-derived systolic blood pressure estimates versus arterial-line measurements, with mean differences of 1.60 and −0.04 mmHg and SDs of 7.99 and 7.86 mmHg, respectively (n = 5), within ANSI/AAMI SP10 limits.
- Additional modalities included kangaroo-care orientation/activity tracking, SCG, and cry analysis: KC events showed phi/theta of 2.85 ± 0.10 rad and −0.29 ± 0.28 rad over 2.4 h; activity was 0.07 ± 0.02 g/s during rest/KC versus 0.24 ± 0.05 g/s during hands-on care; 11 cry events in n = 3 infants showed duration difference −3.9 ± 13.9 s and fundamental frequency 410.7 ± 47.9 Hz.
Methods
- Design: two mechanically soft, wireless devices—a chest unit for ECG, accelerometry, temperature, and acoustic/vibration signals, and a limb unit for reflectance PPG and temperature—using Bluetooth Low Energy, with modular/embedded/battery-free power options.
- Data/sample: pilot clinical studies in NICU and PICU patients, including 50 patients for safety/skin tolerability across 23–40 weeks gestational age and 1 week–4 years chronological age; vital-sign comparisons used smaller subsets (e.g., n = 3 for HR/SpO2/temperature, n = 6 for RR, n = 20 for aggregated HR/SpO2, n = 5 for BP).
- Analysis: Bland-Altman comparisons against standard clinical monitors; time-synchronized PAT/PTT calibrated against arterial-line SBP; accelerometer filtering for KC orientation/activity and cry detection; MANOVA for position/activity comparisons.
Limitations
- The paper states that in instances of low SpO2 (<90%), mean disagreement was higher than the overall comparison, likely reflecting different SpO2 de-noising algorithms rather than a true measurement difference.
- The paper notes that comparisons reflect real-world vital-sign calculations during clinical care, feeding, imaging, and other procedures, so motion artifact and noise are inherent and processed differently between the experimental system and clinical comparator.
- Evident from the text: key validation cohorts were small (e.g., n = 3 for HR/SpO2/temperature, n = 5 for BP, n = 3 for KC/cry), thermal safety testing used a subset of 3 patients over 24 h, and BP estimation required initial calibration against 5 min of arterial-line SBP data.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures






Verification
Claims the paper makes
9 Valid · 9 Not valid · 13 Not checked · 63 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Ultimately, both PAT and PTT depend on vascular system geometry, elasticity, SBP, and other factors30,33,34.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Time-synchronized Bi-nodal Communication for Non-invasive Blood Pressure Monitoring
What the sources say
- Partly supportedcites [30] Ma Y, Relation between blood pressure and pulse wave velocity for human arteries. PNAS 115, 11144– 11149 (2018). 30322935 PMC6217416 sourceExperimental — accuracy not yet measured
The In Vitro Experiments section relates pulse transit time to pulse wave velocity and reports pulse arrival times measured with ECG and PPG. It also states that the pressure–velocity relation depends on arterial material properties and geometry. The reported human measurements concern diastolic blood pressure, however, so the paper does not establish the stated dependence on SBP for both PAT and PTT. (LLM, conf=0.84)
- Not matchedcites [33] Kangaroo mother care: a practical guide, (World Health Organization, Dept. of Reproductive Health and Research, 2003).
Its entry has no title and year that we can search by.
- Not checked — no full text of the sourcecites [34] Liu G-Z, Guo Y-W, Zhu Q-S, Huang B-Y & Wang L Estimation of respiration rate from three-dimensional acceleration data based on body sensor network. Telemed. J. E-Health 17, 705– 711 (2011). 22035321 PMC3203837 source
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Cry analysis can serve as a non-invasive method to analyze the neurophysiological state, often influenced by birth trauma, brain injury or pain stress13.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Advanced Use Cases: Kangaroo Care and Cry Analysis
What the sources say
- Partly supportedcites [13] Shinya Y, Kawai M, Niwa F, Imafuku M & Myowa M Fundamental frequency variation of neonatal spontaneous crying predicts language acquisition in preterm and term infants. Front. Psychol. 8, 2195 (2017). 29312060 PMC5744644 sourceExperimental — accuracy not yet measured
The Introduction (page not specified) describes cry acoustics as a non-invasive index of neonatal neurophysiological states, including pain stress, and links abnormal cry frequency to neurological disturbances. It does not specifically identify birth trauma or brain injury as influences, so those examples are not supported by our copy. (LLM, conf=0.88)
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The fundamental frequency of 410.7 ± 47.9 (Supplementary Fig.24e) is consistent with published results46.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Advanced Use Cases: Kangaroo Care and Cry Analysis
What the sources say
- Could not verifycites [46] Inan Omer T, Novel wearable seismocardiography and machine learning algorithms can assess clinical status of heart failure patients. Circ. Heart Fail. 11, e004313 (2018). 29330154 PMC5769154 sourceExperimental — accuracy not yet measured
The Methods (Pre-Processing and Feature Extraction; page unavailable) discuss frequency components of cardiac seismocardiography signals, not crying. Searches for crying, vocal pitch, and the reported frequency found no passage in our copy addressing whether 410.7 ± 47.9 is consistent with published results. (LLM, conf=0.77)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Continuous monitoring of vital signs is essential, yet existing technologies require the use of multiple electrodes and interfaces attached to the skin with adhesive tapes and connected by wires to electronic processing systems that are often tethered to the wall.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Abstract
What the sources say
- Partly supportedcites [3] Bonner O, Beardsall K, Crilly N & Lasenby J ‘ There were more wires than him’: the potential for wireless patient monitoring in neonatal intensive care. BMJ Innov. 3, 12– 18 (2017). PMC5293857 28250963 sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) describes multiple vital-sign sensors in a neonatal intensive care unit connected by wires to a monitoring unit outside the incubator. The Results (page unavailable) describe adhesive ECG and temperature probes, but say the pulse oximeter is secured with Velcro. The paper does not establish that the monitoring units are tethered to a wall or that all the interfaces use adhesive tape. (LLM, conf=0.87)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“The data show that skin-to-skin contact during KC produced a pronounced, gradual increase in the peripheral skin temperature, consistent with expectation and as demonstrated in previous studies39.”
§ Skin-Interfaced Biosensors and Pilot Studies for Advanced Wireless Physiological Monitoring in Neonatal and Pediatric Intensive Care Units > Results > Advanced Use Cases: Kangaroo Care and Cry Analysis
What the sources say
- Could not verifycites [39] Huntsman RJ, Lowry NJ & Sankaran K Nonepileptic motor phenomena in the neonate. Paediatr. Child Health 13, 680– 684 (2008). 19436521 PMC2606074 sourceExperimental — accuracy not yet measured
The paper’s passages on neonatal tremor and jitteriness (section labeled Abstract; page unavailable) do not address Kangaroo Care or peripheral skin temperature. Searches for skin-to-skin contact, thermoregulation, and a gradual temperature increase found no passage supporting the attributed finding. (LLM, conf=0.78)
boostedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
45 more claims
Create a free account to see all 94 claims and to verify the 2 references still open.
- It may state two separate claims as one. 39
- Probably not a substantial claim. 10
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous reports validated the performance of an FDA-cleared physiological monitoring system against gold standard monitoring platforms [18–23].”
From Pregnancy outcomes in infertility patients diagnosed with sleep disordered breathing with wireless wearable sensors (2022) · cites this paper as [19]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In previous work, we demonstrate the ability for the chest- and finger-mounted sensors to monitor vital signs in both children and adults; the abdominal sensor allows for maternal–fetal monitoring of pregnant women in both high- and low-resource settings as well as the clinical and home settings (22, 23).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [22]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Additionally, as demonstrated by our previous work, neonatal vital signs can also be tracked with the same sensors in both high-resource and low-resource settings as well as support kangaroo care (22, 48).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [22]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Previous wearables have been designed to condense the complicated setups in NICU monitoring into miniaturized wearables for facilitated monitoring4–6; however, they encounter problems with device longevity with either single-use or short-term use before replacement.”
From Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia (2025) · cites this paper as [5]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“and have been used in neonatal and pediatric intensive care units. 4,5”
From Advanced wireless monitoring for children in the cardiac perioperative setting (2023) · cites this paper as [5]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
Scheer, B., Perel, A., & Pfeiffer, U. J. (2002) · Critical Care · doi:10.1186/cc1489 · Cited by 854
Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Self-citationKangaroo Mother Care and Neonatal Outcomes: A Meta-analysis
Boundy, E. O., Dastjerdi, R., Spiegelman, D., Fawzi, W. W., Missmer, S. A., Lieberman, E., Kajeepeta, S., Wall, S., & Chan, G. J. (2016) · Pediatrics · doi:10.1542/peds.2015-2238 · Cited by 661
Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressure
Payne, R. A., Symeonides, C. N., Webb, D. J., & Maxwell, S. R. J. (2006) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00657.2005 · Cited by 418
Liu, Y., Norton, J. J. S., Qazi, R., Zou, Z., Ammann, K. R., Liu, H., Yan, L., Tran, P. L., Jang, K.-I., Lee, J. W., et al. (2016) · Science Advances · doi:10.1126/sciadv.1601185 · Cited by 397
Self-citationXu, J., Murphy, S. L., Kochanek, K. D., Bastian, B. A., & Arias, E. (2018) · PubMed · Cited by 391
Radial Artery Cannulation: A Comprehensive Review of Recent Anatomic and Physiologic Investigations
Brzezinski, M., Luisetti, T., & London, M. J. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181bbd416 · Cited by 354
Research works printed but not linked: 1
- Wheeler DS, Wong HR & Shanley TP Pediatric critical care medicine: basic science and clinical evidence, (Springer, London, 2007).
Cited by
Items 201–300 of 476
Bitter Flavored, Soft Composites for Wearables Designed to Reduce Risks of Choking in Infants
Cho, D., Li, R., Jeong, H., Li, S., Wu, C., Tzavelis, A., Yoo, S., Kwak, S. S., Huang, Y. Y., & Rogers, J. A. (2021) · Advanced Materials · doi:10.1002/adma.202103857 · Cited by 26
Self-citationPhan, D. T., Mondal, S., Tran, L. H., Thien, V. T. M., Nguyen, H. V., Nguyen, C. H., Park, S., Choi, J., & Oh, J. (2021) · Flexible and Printed Electronics · doi:10.1088/2058-8585/ac2c50 · Cited by 26
Breathable, Wireless, Thin-Film Wearable Biopatch Using Noise-Reduction Mechanisms
Rodeheaver, N., Kim, H., Herbert, R., Seo, H., & Yeo, W. (2022) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.1c01107 · Cited by 25
Self-citationFabrication of Silver Electrical Circuits on Textile Substrates by Reactive Inkjet Printing
Xiao, Y., Li, Q., Zhang, C., Zhang, W., Yun, W., & Yang, L. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3166936 · Cited by 23
Overview of Biofluids and Flow Sensing Techniques Applied in Clinical Practice
Yáñez, C., DeMas-Giménez, G., & Royo, S. (2022) · Sensors · doi:10.3390/s22186836 · Cited by 29
Paladhi, A. G., Manohar, M. V., Pal, K., et al. (2022) · Process Biochemistry · doi:10.1016/j.procbio.2022.09.023 · Cited by 24
“Various countries are yet to introduce biosensors due to the privacy concerns limiting the control of existing pandemic [16].”
BackgroundInfant movement classification through pressure distribution analysis
Kulvičius, T., Zhang, D., Nielsen‐Saines, K., Bölte, S., Kraft, M., Einspieler, C., Poustka, L., Wörgötter, F., & Marschik, P. B. (2023) · Communications Medicine · doi:10.1038/s43856-023-00342-5 · Cited by 26
“Some studies applied wearable sensors such as 3 DoF accelerometer sensors [44, 45], 6 DoF inertial motion unit (IMU) sensors [46, 47, 48], or an electromagnetic position tracking system [49, 50].”
A Tapered Soft Robotic Oropharyngeal Swab for Throat Testing: A New Way to Collect Sputa Samples
Xie, Z., Chen, B., Liu, J., Yuan, F., Shao, Z., Yang, H. Q., Domel, A. G., Zhang, J., & Wen, L. (2021) · IEEE Robotics & Automation Magazine · doi:10.1109/mra.2020.3044914 · Cited by 23
Self-citationA Wearable Dual-Modal Patch for Rapid Pre-Hospital Diagnosis of Acute Myocardial Infarction
Yang, K., Dong, Q., Yao, Z., Liang, J., Zhao, J., Wu, L., Zong, S., Cui, Y., & Wang, Z. (2025) · ACS Nano · doi:10.1021/acsnano.5c05461 · Cited by 26
Recent Advances in Materials and Flexible Sensors for Arrhythmia Detection
Guess, M., Zavanelli, N., & Yeo, W. (2022) · Materials · doi:10.3390/ma15030724 · Cited by 27
Davies, C., Lee, J. Y., Walter, J., Kim, D., Yu, L., Park, J., Blake, S., Kalluri, L., Cziraky, M., Stanek, E., et al. (2022) · Journal of Clinical Sleep Medicine · doi:10.5664/jcsm.10194 · Cited by 23
Self-citationLiquid Metal‐Based Sensor Skin Enabling Haptic Perception of Softness
Chen, H., Furfaro, I., Lavado, E. F., & Lacour, S. P. (2023) · Advanced Functional Materials · doi:10.1002/adfm.202308698 · Cited by 28
Toward Stretchable Flexible Integrated Sensor Systems
Cui, S., Han, D., Chen, G., Liu, S., Xu, Y., Yu, Y., & Peng, L. (2024) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c12429 · Cited by 26
Self-citationHan, Y.-M., Seo, J., Lee, D. H., & Yoo, H. (2025) · Micromachines · doi:10.3390/mi16020118 · Cited by 29
“Recently, biosensors that collect amounts of human signals such as oxygen saturation [88,89], blood sugar [90,91], and heart rate [92,93] through integration with smart devices are under development for health diagnostics.”
Automated prediction of extubation success in extremely preterm infants: the APEX multicenter study
Kanbar, L. J., Shalish, W., Onu, C. C., Latremouille, S., Kovács, L., Keszler, M., Chawla, S., Brown, K. A., Precup, D., Kearney, R. E., et al. (2022) · Pediatric Research · doi:10.1038/s41390-022-02210-9 · Cited by 28
“The signal acquisition set-up involved instrumentation and the possibility of recordings of low-quality signals, but the use of small multimodal wireless sensors may make data acquisition more practical and reliable.(44,45) The classifier was developed using a pre-specified observation window of 72 h which may not cap…”
MethodPhotonic skin for photonic-integration-based wearable sensors
Li, H., Li, X., Yang, Y., Xie, F., Han, M., Lin, Z., Wang, Y., Zhang, J., Zhang, S., Zhang, C., et al. (2025) · Optica · doi:10.1364/optica.546458 · Cited by 21
A digital nervous system aiming toward personalized IoT healthcare
Armgarth, A., Pantzare, S., Arvén, P., Lassnig, R., Jinno, H., Gabrielsson, E. O., Kifle, Y., Cherian, D., Sjöström, T. A., Berthou, G., et al. (2021) · Scientific Reports · doi:10.1038/s41598-021-87177-z · Cited by 23
He, K., Jiang, Y., Wang, T., Liu, Z., Wang, M., Pan, L., & Chen, X. (2021) · Aggregate · doi:10.1002/agt2.57 · Cited by 22
Bio-inspired ultra-thin microfluidics for soft sweat-activated batteries and skin electronics
Wu, M., Shi, R., Zhou, J., Wong, T. H., Yao, K., Li, J., Huang, X., Li, D., Gao, Y., Liu, Y., et al. (2022) · Journal of Materials Chemistry A · doi:10.1039/d2ta01154a · Cited by 21
Development of wearable multi-lead ECG measurement device using cubic flocked electrode
Takeshita, T., Yoshida, M., Takei, Y., Ouchi, A., Hinoki, A., Uchida, H., & Kobayashi, T. (2022) · Scientific Reports · doi:10.1038/s41598-022-24043-6 · Cited by 25
Hybrid 3D Printing of a Nature‐Inspired Flexible Self‐Adhesive Biopatch for Multi‐Biosignal Sensing
Alsharif, A. A., Syed, A. M., Li, X., Alsharif, N. A., Lubineau, G., & El‐Atab, N. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202406341 · Cited by 29
Wearable Physiological Multi-Vital Sign Monitoring System With Medical Standard
Yin, S., Xue, N., You, C., Guo, Y., Yao, P., Shi, Y., Liu, T., Yao, L., Zhou, J., Sun, J., et al. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3123084 · Cited by 21
Self-citationWearable network for multilevel physical fatigue prediction in manufacturing workers
Mohapatra, P., Aravind, V., Bisram, M., Lee, Y.-J., Jeong, H., Jinkins, K. R., Gardner, R., Streamer, J., Bowers, B., Cavuoto, L. A., et al. (2024) · PNAS Nexus · doi:10.1093/pnasnexus/pgae421 · Cited by 26
Self-citationGao, Z., Zhang, L., Lei, H., Liu, Y., Gu, H., Xie, L., Lu, B., Ji, H., Wen, Z., & Sun, X. (2025) · ACS Nano · doi:10.1021/acsnano.5c00313 · Cited by 24
Lightweight Video-Based Respiration Rate Detection Algorithm: An Application Case on Intensive Care
Tan, X., Hu, M., Zhai, G., Zhu, Y., Li, W., & Zhang, X.-P. S. (2023) · IEEE Transactions on Multimedia · doi:10.1109/tmm.2023.3286994 · Cited by 25
Shu, S., An, J., Chen, P., Liu, D., Wang, Z., Li, C., Zhang, S., Liu, Y., Luo, J., Zu, L., et al. (2021) · Research · doi:10.34133/2021/9783432 · Cited by 23
Fazio, R. D., Giannoccaro, N. I., Carrasco, M., Velázquez, R., & Visconti, P. (2021) · Frontiers of Information Technology & Electronic Engineering · doi:10.1631/fitee.2100085 · Cited by 24
“Different solutions have been proposed in the scientific literature for remote tracking and monitoring of patient vital signs to detect worsening of their conditions early in the disease process (Chung et al., 2020; Greenhalgh et al., 2020; Menni et al., 2020).”
BackgroundGrooby, E., Sitaula, C., Kwok, T. C., Sharkey, D., Marzbanrad, F., & Malhotra, A. K. (2023) · Pediatric Research · doi:10.1038/s41390-022-02416-x · Cited by 29
“One patch is placed on the hand or foot for PPG and oxygen saturation from optical-based sensors(21,93) and peripheral temperature measurement.(21) The other patch is placed on the back or chest and contains 2 electrodes for ECG,(21,93) tri-axial accelerometer, and temperature measurement unit.”
BackgroundMethodInfluentialDesign and manufacturing of soft electronics for in situ biochemical sensing
Xing, Y., Wang, J., & Li, J. (2024) · International Journal of Extreme Manufacturing · doi:10.1088/2631-7990/ad65a0 · Cited by 22
Flexible, wearable mechano-acoustic sensors for body sound monitoring applications
Dang, T. B., Truong, T., Nguyen, C. C., Listyawan, M. A., Sapers, J. S., Zhao, S., Truong, D. P., Zhang, J., Do, T. N., & Phan, H. (2025) · Nanoscale · doi:10.1039/d4nr05145a · Cited by 24
Park, C., Park, D. Y., Zhang, H., Yang, D. S., Redden, C. R., Yoo, S., Park, T. W., Zhou, M., Yoo, J., Sim, Y., et al. (2025) · Nature Communications · doi:10.1038/s41467-025-60911-1 · Cited by 29
Self-citationChallenges and opportunities in next-generation LED therapeutic devices
Wang, C., Yu, Q., Li, M., Chen, H., Fan, H., Ma, Y., Zhang, Z., Wu, M. X., & Lu, M. (2025) · Light Science & Applications · doi:10.1038/s41377-025-01990-z · Cited by 28
“Moreover, integrating transcutaneous bilirubin sensors 153 – 155 with phototherapy garments 13 enhances the treatment of hyperbilirubinemia (Table 3), while implantable sensors combined with phototherapy devices hold promise for simultaneously monitoring gastrointestinal microbiota and modulating…”
Wu, Z., Li, X., Feng, Z., Wan, C., Ya, L., Li, T., Yang, Q., Liu, X., Ren, M., Li, J., et al. (2023) · Advanced Healthcare Materials · doi:10.1002/adhm.202202629 · Cited by 20
Simple, miniaturized biosensors for wireless mapping of thermoregulatory responses
Oh, S., Yoo, J., Maeng, W., Yoo, S., Yang, T., Slattery, S. M., Pessano, S., Chang, E., Jeong, H., Kim, J., et al. (2023) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2023.115545 · Cited by 18
Self-citationPhotoplethysmography in noninvasive blood pressure monitoring
Mukkamala, R., Hahn, J., & Chandrasekhar, A. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00010-4 · Cited by 26
Deep learning empowered sensor fusion boosts infant movement classification
Kulvičius, T., Zhang, D., Poustka, L., Bölte, S., Jahn, L., Flügge, S., Kraft, M., Zweckstetter, M., Nielsen‐Saines, K., Wörgötter, F., et al. (2025) · Communications Medicine · doi:10.1038/s43856-024-00701-w · Cited by 25
“Although assessors ’ reliability has proven excellent across various sitesand studies 17 – 20 , humanand environmental factorswill remainanissue that could affect individual performance 2 .”
Lee, K.-R., Kim, T., Im, S., Lee, Y. J., Jeong, S., Shin, H., Cho, H., Park, S., Kim, M., Lee, J. G., et al. (2025) · Engineering · doi:10.1016/j.eng.2024.12.031 · Cited by 23
Self-citationWireless, Multifunctional System-Integrated Programmable Soft Robot
Han, S., Shin, J., Lee, J. H., Li, B., Ko, G., Jang, T., Dutta, A., Han, W. B., Yang, S. M., Kim, D., et al. (2025) · Nano-Micro Letters · doi:10.1007/s40820-024-01601-3 · Cited by 21
Self-citationHuang, X., Yuan, Y., Liu, J., He, J., Shi, Y., Gao, S., Wu, J., Xu, X., Zhang, H., Li, P., et al. (2025) · National Science Review · doi:10.1093/nsr/nwaf425 · Cited by 20
S- to X-Band Stretchable Inductors and Filters for Gigahertz Soft and Epidermal Electronics
Lan, Y., Zhang, H., Min, S., Kim, D., Gong, S., Katehi, L. P. B., Xu, Y., & Ma, Z. (2021) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.0c22003 · Cited by 17
Self-citationZhang, C., Cao, Y., Song, Y., Yu, G., Lan, L., Zhou, C. Y., Lin, Z., Wang, L., Li, N., Huang, F., et al. (2023) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.2c15657 · Cited by 15
Bioelectronics for targeted pain management
Flavin, M. T., Foppiani, J., Paul, M. A., Alvarez, A. H., Foster, L., Gavlasova, D., Ma, H., Rogers, J. A., & Lin, S. J. (2025) · Nature Reviews Electrical Engineering · doi:10.1038/s44287-025-00177-3 · Cited by 21
“Wearable devices can leverage protocols such as Bluetooth Low Energy 11,39,46 and near-field communication 40,50 .”
Self-citationWireless Battery‐Free Flexible Sensing System for Continuous Wearable Health Monitoring
Wu, C., Han, L., Dong, Y., Guo, M., Wang, R., & Si, J. (2023) · Advanced Materials Technologies · doi:10.1002/admt.202201662 · Cited by 16
A review: Recent advancements in sensor technology for non-invasive neonatal health monitoring
Mishra, S. K., Khouqeer, G. A., Aamna, B., Alodhayb, A. N., Ibrahim, S. J. A., Hooda, M. K., & Jayaswal, G. (2023) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2023.100332 · Cited by 27
Walter, J. R., Lee, J. Y., Snoll, B., Park, J. B., Kim, D. H., Xu, S., & Barnhart, K. (2022) · Sleep Medicine · doi:10.1016/j.sleep.2022.09.026 · Cited by 15
Self-citationChen, Z., Zhang, F., Qian, X., Zhao, G., & Yan, Z. (2025) · Materials Horizons · doi:10.1039/d5mh00700c · Cited by 16
Tissue-like Bioelectronic Material Strategies for Personalized Closed-Loop Neuroprostheses
Kim, Y., Lee, H., Lee, J., Shin, M., & Son, D. (2025) · ACS Nano · doi:10.1021/acsnano.5c07150 · Cited by 21
Self-citationWireless monitoring devices in hospitalized children: a scoping review
Sénéchal, E., Jeanne, E., Tao, L., Kearney, R. E., Shalish, W., & SantʼAnna, G. (2023) · European Journal of Pediatrics · doi:10.1007/s00431-023-04881-w · Cited by 18
The use of wireless sensors in the neonatal intensive care unit: a study protocol
Sénéchal, E., Radeschi, D. J., Tao, L., Lv, S., Jeanne, E., Kearney, R. E., Shalish, W., & SantʼAnna, G. (2023) · PeerJ · doi:10.7717/peerj.15578 · Cited by 17
TRCCBP: Transformer Network for Radar-Based Contactless Continuous Blood Pressure Monitoring
Jiang, X., Zhang, J., Mu, W., Wang, K., Li, L., & Zhang, L. (2023) · Sensors · doi:10.3390/s23249680 · Cited by 16
“U. Chuan [11] monitored the skin vibrations and extracted pulse waves to estimate BP using electrode patches that attached to the skin.”
MethodMicrofluidics-derived microfibers in flexible bioelectronics
Yang, C., Hou, X., & Zhang, L. (2024) · Materials Futures · doi:10.1088/2752-5724/ad667b · Cited by 18
Wearable Blood Pressure Monitoring Devices: Understanding Heterogeneity in Design and Evaluation
Cheung, M. Y., Sabharwal, A., Coté, G. L., & Veeraraghavan, A. (2024) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2024.3434344 · Cited by 24
Malignant Hyperthermia in PICU—From Diagnosis to Treatment in the Light of Up-to-Date Knowledge
Klincová, M., Štěpánková, D., Schrőderová, I., Klabusayová, E., & Štourač, P. (2022) · Children · doi:10.3390/children9111692 · Cited by 21
“The continuous monitoring of at least electrocardiography, non-invasive blood pressure, SpO2, EtCO2, diuresis and temperature should be the routine in PICU, a modern wireless monitoring system can be used [98].”
MethodRecent advances in polymer-based radiative cooling material: Principle, designs and applications
Yang, X., Yang, Y., Deng, Y., Li, Z., Zhang, L., Wang, Q., Lan, R., & Yang, H. (2025) · Review of Materials Research · doi:10.1016/j.revmat.2025.100043 · Cited by 20
Micro/Nanofluidic‐Enabled Biomedical Devices: Integration of Structural Design and Manufacturing
Yin, M., Kim, Z. A., & Xu, B. (2021) · Advanced NanoBiomed Research · doi:10.1002/anbr.202100117 · Cited by 14
Retracted: AI and IoT Enabled Technologies for Monitoring the Right to Health of Disabled People
Pandey, S., Dixit, A. K., Bahuguna, R., et al. (2022) · International Conference on Contemporary Computing and Informatics (IC3I) · doi:10.1109/ic3i56241.2022.10073046 · Cited by 13
“The Unified Country likewise anticipated toward the beginning of 2009 assessment of disabled during pandemic was 10% so it tends to be effortlessly assumed that the proportion in development from that point till currently is probably going to expand there is an opportunity for raise in constant illnesses about which W…”
BackgroundRetractedGuo, X. M., Avila, R. O., Huang, Y. Y., & Xie, Z. (2021) · International journal of mechanical system dynamics · doi:10.1002/msd2.12017 · Cited by 13
“Chung et al.90 designed a pair of soft, skin‐interfaced wearable device units that can be placed in the chest, back, and limbs of neonates to monitor body motion, vocal biomarkers, and crying frequency.”
BackgroundInfluentialSelf-citationFiber‐type soft bioelectronics for wearable and implantable sensing and therapy
Kim, H., Kim, H. Y., Koo, J. H., Cha, G. D., Kim, D., & Kim, H. J. (2025) · BMEMat · doi:10.1002/bmm2.70043 · Cited by 21
Self-citationIslam, B., McElwain, N. L., Li, J., Davila, M. I., Hu, Y., Hu, K., Bodway, J. M., Dhekne, A., Choudhury, R. R., & Hasegawa‐Johnson, M. (2024) · Sensors · doi:10.3390/s24030901 · Cited by 15
“…in particular, have predominantly focused on wireless skin-interfaced biosensors made of soft, flexible electronics that permit the continuous monitoring of vital signs, including but not limited to heart rate, blood pressure, temperature, respiration, and blood oxygen saturation (see [11–13]).”
BackgroundCheng, L. P., Guo, A., Li, J., Li, M., Lei, Q., Xu, W., Guo, X., & Zhang, J. (2024) · Science China Materials · doi:10.1007/s40843-024-2994-0 · Cited by 15
Soft wearable electronics for evaluation of biological tissue mechanics
Lu, Y., Ma, L., Zhang, H., Mei, Y., Xiong, Z., & Song, E. (2024) · Soft Science · doi:10.20517/ss.2024.29 · Cited by 17
Wearable biosensors for pediatric hospitals: a scoping review
Hyun, A., Takashima, M., Hall, S., Lee, L., Dufficy, M., Ruppel, H., & Ullman, A. J. (2024) · Pediatric Research · doi:10.1038/s41390-024-03693-4 · Cited by 21
Bioelectronic devices for light-based diagnostics and therapies
Kim, H., Sritandi, W., Xiong, Z., & Ho, J. S. (2023) · Biophysics Reviews · doi:10.1063/5.0102811 · Cited by 12
Self-citationContinuous sensing and quantification of body motion in infants: A systematic review
Peng, Z. Q., Kommers, D. R., Liang, R.-H., Long, X., Cottaar, W., Niemarkt, H. J., Andriessen, P., & van Pul, C. (2023) · Heliyon · doi:10.1016/j.heliyon.2023.e18234 · Cited by 16
“[10] designed a soft adhesive patch, also described as an epidermal electronic system for physiological monitoring in infants.”
BackgroundMethodInfluentialSoft, Conductive, and Anti‐Freezing Conducting Polymer Organohydrogels
Zhou, X., Kateb, P., Miquet‐Westphal, F., Lodygensky, G. A., & Cicoira, F. (2023) · Advanced Sensor Research · doi:10.1002/adsr.202300072 · Cited by 12
Ultra-Thin, Soft, Radiative Cooling Interfaces for Advanced Thermal Management in Skin Electronics
Li, J., & Yu, X. (2023) · IEEE International Conference on Nanotechnology (IEEE-NANO) · doi:10.1109/nano58406.2023.10231185 · Cited by 12
“Wireless technologies, especially radio frequency (RF) based techniques, enable battery-free for skin electronics and get rid of many external connections[1, 38, 39].”
BackgroundFlexible Meso Electronics and Photonics Based on Cocoon Silk and Applications
Lu, C., Wang, X., & Liu, X. (2024) · ACS Biomaterials Science & Engineering · doi:10.1021/acsbiomaterials.4c00254 · Cited by 13
Wu, C., Rwei, A. Y., Lee, J. Y., Ouyang, W., Jacobson, L., Shen, H., Luan, H., Xu, Y., Park, J. B., Kwak, S. S., et al. (2021) · Journal of Reconstructive Microsurgery · doi:10.1055/s-0041-1732426 · Cited by 14
Self-citationWang, Y., & Wang, C. (2021) · International Journal of Solids and Structures · doi:10.1016/j.ijsolstr.2021.111382 · Cited by 12
Sensors for Neonatal Monitoring
Güler, Ü., Sen, D., Costanzo, I., Tufan, T. B., & Rhein, L. M. (2022) · Elsevier eBooks · doi:10.1016/b978-0-12-822548-6.00104-7 · Cited by 16
Epidermal Bioelectronics for Management of Chronic Diseases: Materials, Devices and Systems
Qiao, Z., Chen, S., Fan, S., Xiong, Z., & Lim, C. T. (2023) · Advanced Sensor Research · doi:10.1002/adsr.202200068 · Cited by 13
Unlocking the potential of transdermal drug delivery
Li, S., Wu, J., Peng, X., & Feng, X. (2024) · International Journal of Smart and Nano Materials · doi:10.1080/19475411.2024.2366210 · Cited by 16
“Bottom: a wearable cardiac ultrasound imager [241]. h ) Skin-interfaced biosensors for wireless physiological monitoring in neonatal and pediatric intensive-care units [244].”
Zhang, W., Zhang, H., Chen, G., Ma, B., Xia, Y., & Chen, Y. (2025) · Materials Horizons · doi:10.1039/d5mh00460h · Cited by 10
Koshi, T., Takei, A., Nobeshima, T., Kanazawa, S., Nomura, K.-I., & Uemura, S. (2024) · Flexible and Printed Electronics · doi:10.1088/2058-8585/ad2778 · Cited by 10
Wearable approaches for non-invasive monitoring of tendons: A scoping review
Kárason, H., Ritrovato, P., Maffulli, N., Boccaccini, A. R., & Tortorella, F. (2024) · Internet of Things · doi:10.1016/j.iot.2024.101199 · Cited by 18
Orlowska, M. K., Guan, B., Sedev, R. V., Morikawa, Y., Suu, K., & Priest, C. (2020) · Analytical Chemistry · doi:10.1021/acs.analchem.0c03667 · Cited by 10
Assessment of neonatal respiratory rate variability
Coleman, J., Ginsburg, A. S., Macharia, W., Ochieng, R., Chomba, D., Zhou, G., Dunsmuir, D. T., Karlen, W., & Ansermino, J. M. (2022) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-022-00840-2 · Cited by 17
Zhou, L., Aljiffry, A., Lee, Y. J., Matthews, J., Seitter, B., Soltis, I., Huang, Y., Maher, K., & Yeo, W.-H. (2025) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2025.117372 · Cited by 9
Self-citationHeo, H., Park, J., Ko, D., Lee, K., Lee, J., Joo, H., Yang, J., Cha, G. D., Kim, D., & Kim, D. C. (2025) · Nanoscale Horizons · doi:10.1039/d5nh00418g · Cited by 13
Self-citationGe, Q., Li, C., Fan, Z., Xia, B., Zang, C., Chen, L., Zhao, C., Sang, H., & Wang, A. (2024) · New Journal of Chemistry · doi:10.1039/d4nj01082e · Cited by 9
Huang, D., Zeng, Y., Zhu, Y., Song, X., Pan, L., Yang, J., Wang, Y., Lu, H., & Wang, W. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3482569 · Cited by 14
Khan, S. A., Kim, J., Kang, T.-U., Park, G., Lee, S., Park, J., & Kim, W. (2024) · ACS Sensors · doi:10.1021/acssensors.4c01456 · Cited by 11
Self-citationPatient-centric care: Unveiling the potential of wearable electronics in clinical practice
Zhao, P., Xu, J., & Han, M. (2024) · Wearable electronics. · doi:10.1016/j.wees.2024.07.001 · Cited by 9
AI-Based Estimation of End-Systolic Elastance From Arm-Pressure and Systolic Time Intervals
Bikia, V., Adamopoulos, D. N., Pagoulatou, S., Rovas, G., & Stergiopulos, N. (2021) · Frontiers in Artificial Intelligence · doi:10.3389/frai.2021.579541 · Cited by 10
Yousefi, N., Caba, C., Hu, A., et al. (2022) · ACS Applied Electronic Materials · doi:10.1021/acsaelm.2c00983 · Cited by 9
A photoacoustic patch for three-dimensional imaging of hemoglobin and core temperaturePreprint
Gao, X., Chen, X., Hu, H., et al. (2021) · Research Square · doi:10.21203/rs.3.rs-558432/v1 · Cited by 9
Self-citationNear infrared spectroscopy in monitoring of head and neck microvascular free flaps
Czakó, L., Šimko, K., Soviš, M., et al. (2023) · Bratislavské lekárske listy/Bratislava medical journal · doi:10.4149/bll_2023_079 · Cited by 8
Yang, Z., Tian, Y., Zhao, D., Zhao, Y., Zhou, Z., Wu, D., Shi, Z., Lin, P., Zeng, L., & Wu, D. (2024) · Nano Research · doi:10.26599/nr.2025.94907477 · Cited by 10
Flexible and wearable bioelectronics for electrocardiography monitoring: a review
Zheng, C., Wang, Z., Li, L., Ma, Y., & Feng, X. (2025) · Science China Materials · doi:10.1007/s40843-025-3688-9 · Cited by 14
“6a) integrates a Bluetooth system-on-chip with multimodal sensing modules embedded in chest and limb units, enabling the continuous acquisition of ECG, body temperature, and peripheral blood oxygen saturation [137].”
li, S., Kong, C., Wu, E., Yang, C., Gao, Z., Zhou, X., Guo, H., Jiang, S., & Liang, J. (2026) · Soft Science · doi:10.20517/ss.2025.120 · Cited by 13
Noninvasive, wireless and real-time bladder pressure monitoring with biomimetic structured devices
Li, C., Wu, M., Liu, J., Wang, R., Wei, C., Wang, R., Zhou, Y., Jiang, H., Pan, J., & Guo, C. F. (2022) · Applied Materials Today · doi:10.1016/j.apmt.2022.101635 · Cited by 9
Substantial curvature effects on compliant serpentine mechanics
Wang, Y., Zhao, W. D., Du, Y., Dai, Z., Liu, Y., & Xu, F. (2023) · Mechanics of Materials · doi:10.1016/j.mechmat.2023.104732 · Cited by 8
Xu, B., Jiang, F., Zhu, Z., Meng, H., & Xu, L. (2023) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2023.107763 · Cited by 10
Baker, S., Yogavijayan, T., & Kandasamy, Y. (2023) · Healthcare · doi:10.3390/healthcare11243107 · Cited by 13
“…focused on neural network (NN) architectures [13–17 Several early works considered various regression methods, including the work by Chung et al. [24] which used linear regression to link pulse transit time parameters derived from PPG, ECG, and seismograph (SCG) sensors with systolic BP in a…”
BackgroundMethodInfluentialSoft electronic skin for self-deployable tape-spring hinges
Yao, Y., & Ning, X. (2024) · Communications Engineering · doi:10.1038/s44172-024-00163-x · Cited by 9
“Recent advances in soft electronics have demonstrated the successful integration of advanced electronics in thin, lightweight, and fl exible forms that can withstand substantial mechanical strains, akin to the deformations experienced by human skin 13,14 .”
BackgroundWeaning from mechanical ventilation and assessment of extubation readiness
SantʼAnna, G., & Shalish, W. (2024) · Seminars in Perinatology · doi:10.1016/j.semperi.2024.151890 · Cited by 8
Walter, J. R., Lee, J. Y., Yu, L., Kim, B., Martell, K., Opdycke, A., Scheffel, J., Felsl, I., Patel, S., Rangel, S. M., et al. (2024) · Scientific Reports · doi:10.1038/s41598-024-57830-4 · Cited by 14
“…validation 18 The ANNE system has multiple FDA clearances (K211305, K220095, and K223711) for monitoring the following parameters: heart rate (HR), RR, pulse oxygenation, skin temperature, body temperature, apnea–hypopnea index, total sleep time, core body position, and snoring in adults 19–22 .”
BackgroundSelf-citationKhan, Y., Mauriello, M. L., Nowruzi, P., Motani, A., Hon, G., Vitale, N. H., Li, J., Kim, J., Foudeh, A. M., Duvio, D., et al. (2024) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3613904.3643473 · Cited by 13
Self-citationZhang, Y., Lyu, W., Shi, M. X., Ma, Y., & Feng, X. (2022) · Acta Mechanica Sinica · doi:10.1007/s10409-023-23093-x · Cited by 8
A comprehensive wireless neurological and cardiopulmonary monitoring platform for pediatrics
Wong, J. N., Walter, J. R., Conrad, E. C., Seshadri, D. R., Lee, J. Y., González, H. L. L., Reuther, W., Hong, S. J., Pini, N., Marsillio, L. E., et al. (2023) · PLOS Digital Health · doi:10.1371/journal.pdig.0000291 · Cited by 8
Self-citation
Metadata
- Authors
- Chung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., Ryu, D., Kulikova, E., Cao, J., Odland, I. C., Fields, K. B., Hopkins, B., Banks, A., Ogle, C., Grande, D., Park, J. B., Kim, J., Irie, M., Jang, H., Lee, J., Park, Y., Kim, J., Jo, H. H., Hahm, H., Avila, R., Xu, Y., Namkoong, M., Kwak, J. W., Suen, E., Paulus, M. A., Kim, R. J., Parsons, B. V., Human, K. A., Kim, S. S., Patel, M., Reuther, W., Kim, H. S., Lee, S. H., Leedle, J. D., Yun, Y., Rigali, S., Son, T., Jung, I., Arafa, H., Soundararajan, V. R., Ollech, A., Shukla, A., Bradley, A., Schau, M., Rand, C. M., Marsillio, L. E., Harris, Z. L., Huang, Y., Hamvas, A., Paller, A. S., Weese-Mayer, D. E., Lee, J. Y., & Rogers, J. A.
- Affiliations
- Northwestern University (US)
- Lurie Children's Hospital (US)
- Dalian University of Technology (CN)
- Sibel (United States) (US)
- University of Illinois Urbana-Champaign (US)
- North Carolina State University (US)
- Kyung Hee University (KR)
- Corresponding author
- Lee, Jong Yoon
- Rogers, John A.
- ORCID
- Chung, Ha Uk
- Rwei, Alina Y.
- Hourlier-Fargette, Aurélie
- Xu, Shuai
- Lee, KunHyuck
- Dunne, Emma C.
- Xie, Zhaoqian
- Liu, Claire
- Carlini, Andrea
- Kim, Dong Hyun
- Cao, Jingyue
- Odland, Ian C.
- Banks, Anthony
- Grande, Dominic
- Park, Jun Bin
- Kim, Jongwon
- Jang, Hokyung
- Park, Yerim
- Jo, Han Heul
- Avila, Raudel
- Xu, Yeshou
- Namkoong, Myeong
- Kwak, Jean Won
- Human, Kelia A.
- Patel, Manish
- Reuther, William
- Lee, Sung Hoon
- Yun, Yeojeong
- Jung, Inhwa
- Arafa, Hany
- Ollech, Ayelet
- Rand, Casey M.
- Marsillio, Lauren E.
- Harris, Zena L.
- Huang, Yonggang
- Hamvas, Aaron
- Paller, Amy S.
- Weese-Mayer, Debra E.
- Lee, Jong Yoon
- Rogers, John A.
- Venue
- Nature Medicine
- Published
- 2020-03
- Volume
- 26
- Issue
- 3
- Pages
- 418-429
- ISSN
- 1078-8956 (ISSN-L)
- 1546-170X
- PMC
- PMC7315772
- OpenAlex
- W3012163988
- Semantic Scholar
- 9be51710140598d05f71429da3bc79c90db9cace
- Cited by
- 482
- References
- 55
- Citations by year
- 2026: 49 · 2025: 77 · 2024: 76 · 2023: 85 · 2022: 77 · 2021: 89 · 2020: 29
- FWCI
- 21.94
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- cc-by
- www.springer.com/tdm (tdm)
- www.springer.com/tdm (vor)
- Locations
- Nature Medicine (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion, cc-by)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Funders
- Bill and Melinda Gates Foundation (OPP1182909, OPP1193311)
- Gerber Foundation
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- wireless physiological monitoring, neonatal intensive care, pediatric intensive care, non-invasive vital sign monitoring, skin-to-skin care, vocal biomarkers, crying
- MeSH terms
- Child, Child, Preschool, Electrocardiography, Equipment Design, Humans, Infant, Newborn, Intensive Care Units, Neonatal, Monitoring, Physiologic, Skin, Time Factors, Intensive Care Units, Pediatric, Biosensing Techniques, Photoplethysmography, Blood Pressure Monitoring, Ambulatory, Wireless Technology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.