Optimization of preanalytical conditions and analysis of plasma glucose. 1. Impact of the new WHO and ADA recommendations on diagnosis of diabetes mellitus
The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.
Abstract
The new diagnostic criteria for type 2 diabetes from the American Diabetes Association (ADA) and World Health Organization (WHO) recommend measurements on plasma and a lowering of the glucose threshold for diabetes by 0.8 mmol/L. This narrows the distance between the upper end of the reference limit and the discriminatory level to a degree where analytical quality becomes critical. The quality demands for the preanalytical and analytical phase and their consequences on diagnostic performance have to be established in the new technical system, measuring in plasma rather than in capillary whole blood. Because of the instability of glucose in blood samples it is necessary to clarify the influence of different preanalytical and analytical factors on the number of false-positive and false-negative classifications. Thus the aim of the present study was to find optimal conditions for sampling, additives, storage, transport and analysis of plasma glucose combining feasibility with an analytical bias close to zero and a within-imprecision around 1%. We have documented the analytical performance of the method itself and its traceability to an international standard. The preanalytical conditions, such as influence of antiglycolytic agent NaF, conditions for plasma separation, storage temperature and storage time before and after plasma separation were investigated. In conclusion, we recommend that blood should be drawn in tubes containing heparin and NaF and kept on ice water for not mo…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
Report of the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus
The Expert Committee on the Diagnosis and Classification of Diabetes Mellitus (1997) · Diabetes Care · doi:10.2337/diacare.20.7.1183 · Cited by 6,220
Die enzymatische Bestimmung von Glucose und Fructose nebeneinander
Schmidt, F. H. (1961) · Klinische Wochenschrift · doi:10.1007/bf01506150 · Cited by 503
Proposals for Setting Generally Applicable Quality Goals Solely Based on Biology
Fraser, C. G., Petersen, P. H., Libeer, J.-C., & Ricos, C. (1997) · Annals of Clinical Biochemistry: International Journal of Laboratory Medicine · doi:10.1177/000456329703400103 · Cited by 370
Analytical performance characteristics should be judged against objective quality specifications
Fraser, C. G., & Petersen, P. H. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.3.321 · Cited by 194
What is Blood Glucose: Can it be Measured?
Burrin, J. M., & Alberti, K. G. M. M. (1990) · Diabetic Medicine · doi:10.1111/j.1464-5491.1990.tb01370.x · Cited by 105
Proposal for standardizing direct-reading biosensors for blood glucose
Fogh-Andersen, N., & D’Orazio, P. (1998) · Clinical Chemistry · doi:10.1093/clinchem/44.3.655 · Cited by 84
Contant, G., Gouault-Heilmann, M., & Martinoli, J. L. (1983) · Thrombosis Research · doi:10.1016/0049-3848(83)90337-7 · Cited by 80
Storage of whole blood: effect of temperature on the measured concentration of analytes in serum.
Rehak, N. N., & Chiang, B. (1988) · Clinical Chemistry · doi:10.1093/clinchem/34.10.2111 · Cited by 76
Felding, P., Tryding, N., Petersen, P. H., & Hørder, M. (1980) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365518009091972 · Cited by 69
The stability of blood, plasma and serum constituents during simulated transport
Felding, P., Petersen, P. H., & Hørder, M. (1981) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365518109092012 · Cited by 49
Research works printed but not linked: 7
- Chan AYW, 1989, Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem, 35, 315
- Alberti KG, 1998, Part 1. Diagnosis and classification of diabetes mellitus. Diabet Med, 15, 539
- JOrgensen LGM, Stahl M,, 2001, The impact of the new WHO and ADA recommen-dations on the diagnosis of Diabetes Mellitus. 2. The plasma glucose reference interval in a non-risk population. Scand J Clin Lab Invest, 61, 181
- Hyltoft Petersen P, 2001, Evaluation of systematic and random factors in measurements of fasting plasma glucose as the basis for analytical quality specifications in the diagnosis of diabetes. 3. Impact of the new WHO and ADA recommendation on diagnosis of dia-betes mellitus. Scand J Clin Lab Invest, 61, 191
- Hyltoft Petersen P, 1997, Graphical interpretation of analytical data from comparison of a field method with a Reference Method by use of difference plots. J Clin Chem, 43, 2039
- Gowans EMS, 1988, Analytical goals for the acceptance of common reference intervals for laboratories throughout a geographic area. Scand J Clin Lab Invest, 48, 757
- Sebastian-Gambaro MA, 1997, Eur J Clin Chem Clin Biochem, 35, 845
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Diabetes Care · doi:10.2337/dc11-9998 · Cited by 718
“6 mg/dL) (64), or approximately 2% (65), or 0.”
A1c versus glucose testing: a comparison
Sacks, D. B. (2011) · Diabetes Care · doi:10.2337/dc10-1546 · Cited by 461
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
“6 mg/dL) (64), or approximately 2% (65), or 0.”
Prevalence and Management of Diabetes in Korean Adults
Choi, Y. J., Kim, H. C., Kim, H. M., Park, S. W., Kim, J., & Kim, D. J. (2009) · Diabetes Care · doi:10.2337/dc08-2228 · Cited by 166
Acidification of blood is superior to sodium fluoride alone as an inhibitor of glycolysis
Gambino, R., Piscitelli, J., Ackattupathil, T. A., Theriault, J. L., Andrin, R. D., Sanfilippo, M. L., & Etienne, M. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.121707 · Cited by 160
Diabetes Melito: Diagnóstico, Classificação e Avaliação do Controle Glicêmico
Gross, J. L., Silveiro, S. P., Camargo, J. L., Reichelt, A. J., & de Azevedo, M. J. (2002) · Arquivos Brasileiros de Endocrinologia & Metabologia · doi:10.1590/s0004-27302002000100004 · Cited by 465
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 208
“6 mg/dL) (64), or approximately 2% (65), or 0.”
Objective Criteria for Partitioning Gaussian-distributed Reference Values into Subgroups
Lahti, A., Petersen, P. H., Boyd, J. C., Fraser, C. G., & Jørgensen, N. B. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.2.338 · Cited by 137
“Every laboratory that takes part in the production of the common reference intervals and wishes to use them must fulfill the required quality specifications (11, 12).”
BackgroundTanner, M., Kent, N., Smith, B., Fletcher, S., & Lewer, M. (2008) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1258/acb.2007.007183 · Cited by 125
Comparability of venous and capillary glucose measurements in blood
Colagiuri, S., Sandbæk, A., Carstensen, B., Christensen, J., Glumer, C., Lauritzen, T., & Borch‐Johnsen, K. (2003) · Diabetic Medicine · doi:10.1046/j.1464-5491.2003.01048.x · Cited by 101
Performance Evaluation of Blood Glucose Monitoring Devices
Chen, E. T., Nichols, J. H., Duh, S.-H., & Hortin, G. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526969 · Cited by 96
“As described by Stahl et al., 37 special measures are required to achieve accurate analysis of glucose concentrations.”
Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640
Twomey, P. J. (2004) · Journal of Clinical Pathology · doi:10.1136/jcp.2003.013417 · Cited by 74
Glucose: A Simple Molecule That Is Not Simple to Quantify
Gambino, R. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2007.094466 · Cited by 56
Petersen, P. H., Jørgensen, L. G. M., Brandslund, I., De Fine Olivarius, N., & Stahl, M. (2005) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510500236135 · Cited by 51
Daly, N., Flynn, I., Carroll, C., Farren, M., McKeating, A., & Turner, M. J. (2016) · Clinical Chemistry · doi:10.1373/clinchem.2015.247478 · Cited by 50
Stahl, M., Brandslund, I., Jørgensen, L. G. M., Petersen, P. H., Borch-Johnsen, K., & Olivarius, N. D. F. (2002) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/003655102753611799 · Cited by 50
Self-citationP. Hyltoft Petersen, I. B. (2001) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/003655101300133630 · Cited by 49
What information on quality specifications should be communicated to clinicians, and how?
Plebani, M. (2004) · Clinica Chimica Acta · doi:10.1016/j.cccn.2004.03.019 · Cited by 48
Haeckel, R., Brinck, U., Colic, D., Janka, H.-U., Püntmann, I., Schneider, J., & Viebrock, C. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.6.936 · Cited by 44
Felding, P., Rustad, P., Mårtensson, A., Kairisto, V., Franzson, L., Petersen, P. H., & Uldall, A. (2004) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510410006630 · Cited by 43
Waring, W. S., Evans, L. E., & Kirkpatrick, C. T. (2007) · Journal of Clinical Pathology · doi:10.1136/jcp.2006.039925 · Cited by 42
Which sample tube should be used for routine glucose determination?
Bonetti, G., Cancelli, V., Coccoli, G., Piccinelli, G., Brugnoni, D., Caimi, L., & Carta, M. (2015) · Primary care diabetes · doi:10.1016/j.pcd.2015.11.003 · Cited by 40
Nielsen, A. A., Petersen, P. H., Green, A., Christensen, C., Christensen, H., & Brandslund, I. (2014) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2013-0337 · Cited by 41
Prerequisites for establishing common reference intervals
Petersen, P. H., & Rustad, P. (2004) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510410006298 · Cited by 38
Sandbæk, A., Lauritzen, T., Borch‐Johnsen, K., Mai, K., & Christiansen, J. S. (2005) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01491.x · Cited by 37
Citric/citrate buffer: an effective antiglycolytic agent
del Pino, I. G., Constanso, I., Mourín, L. V., Safont, C. B., & Vázquez, P. R. (2013) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0735 · Cited by 33
“Glycolysis ex-vivo [1] is the main interference factor in glycemia determination when analysis is delayed, the specimen is not kept cold enough [2] and centrifugation cannot be carried out in a short period of time after venipuncture, especially in presence of leukocytosis, thrombocytosis or elevated hematocrit.”
BackgroundBình, T. Q., Phương, P. T., Nhung, B. T., Thoang, D. D., Thang, P., Long, T. K., & Thanh, D. V. (2012) · BMC Public Health · doi:10.1186/1471-2458-12-939 · Cited by 31
“However, although this process we used can minimize the preanalytical impact on detecting diabetes, it may cause about 2% underestimate of prediabetes prevalence [34].”
MethodWatkinson, P., Barber, V. S., Amira, E., James, T., Taylor, R. P., & Young, J. D. (2012) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1258/acb.2011.011162 · Cited by 32
Controlled storage conditions prolong stability of biochemical components in whole blood
Stahl, M., & Brandslund, I. (2005) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2005.036 · Cited by 30
“The same result was previously found for glucose measurements in plasma (5).”
BackgroundBennett, K. A., Turner, L. M., Millward, S., Moss, S. E. W., & Hall, A. J. (2017) · Conservation Physiology · doi:10.1093/conphys/cox013 · Cited by 29
“The American Diabetes Association recommends readings from glucometers should be within 15% of laboratory-derived values to be usable (Stahl et al., 2001).”
IMPORTANCE OF THE PRE-ANALYTICAL PHASE IN BLOOD GLUCOSE ANALYSIS
Janssen, K., & Delanghe, J. (2010) · Acta Clinica Belgica · doi:10.1179/acb.2010.068 · Cited by 22
“sodium fl uoride) have a delayed activity, a tube containing a glycolytic inhibitor should be placed in icewater immediately after blood collection and the separation should take place within 30 minutes (8, 15).”
BackgroundUpper reference limit, analytical quality specifications and clinical use of haemoglobin A1C
Jørgensen, L. G. M., Brandslund, I., Stahl, M., Petersen, P. H., Iversen, S., Klitgaard, N. A., & Olivarius, N. D. F. (2002) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/003655102764654349 · Cited by 23
Lyon, A. W. (2004) · Canadian Medical Association Journal · doi:10.1503/cmaj.1040138 · Cited by 21
The clinical impact of screening for gestational diabetes
Schytte, T., Jørgensen, L. G. M., Brandslund, I., Petersen, P. H., & Andersen, B. N. (2004) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2004.209 · Cited by 16
Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus
Stahl, M., & Brandslund, I. (2003) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365510310002590 · Cited by 17
Long-term stability of glucose: glycolysis inhibitor vs. gel barrier tubes
Winter, T., Hannemann, A., Suchsland, J., Nauck, M. A., & Petersmann, A. (2018) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2017-0860 · Cited by 14
RESULTS OF SELFMONITORING ON GLUCOMETER SYSTEMS ADVANCE AND OPTIUM IN DAILY ROUTINE
Chlup, R., Payne, M., Zapletalová, J., Komenda, S., Doubravová, B., Řezníčková, M., Chlupová, L., & Sečkař, P. (2005) · Biomedical Papers · doi:10.5507/bp.2005.015 · Cited by 16
A Potential Issue with Screening Prediabetes or Diabetes Using Serum Glucose: A Delay in Diagnosis
Kang, J. G., Park, C., Ihm, S., & Park, S. W. (2016) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2016.40.5.414 · Cited by 15
The Effect of the New ADA and WHO Guidelines on the Number of Diagnosed Cases of Diabetes Mellitus
Jørgensen, L. G. M., Brandslund, I., Petersen, P. H., Olivarius, N. D. F., & Stahl, M. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.191 · Cited by 12
“Cobas Mira instrument (Roche, Basel, Switzerland), described in detail previously (7, 9).”
BackgroundMethodInfluentialImportance and Status of HBA1C in T2DM and its Indian Perspective
Sarmah, D., & Sharma, B. (2012) · Asian journal of biomedical and pharmaceutical sciences · doi:10.15272/ajbps.v2i12.60 · Cited by 12
Jørgensen, L. G. M., Schytte, T., Brandslund, I., Stahl, M., Petersen, P. H., & Andersen, B. N. (2003) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2003.031 · Cited by 10
“03 (3%) for capillary whole blood measurement also at the level of 5 mmol/l (12).”
BackgroundImpact of Glucose Measuring Systems and Sample Type on Diagnosis Rates of Diabetes Mellitus
Blaurock, M., Kallner, A., Menzel, S., Masuch, A., Nauck, M. A., & Petersmann, A. (2018) · Diabetes Therapy · doi:10.1007/s13300-018-0495-0 · Cited by 12
“[11] and those of the WHO for performing P-glucose testing were also followed 2032 Diabetes Ther (2018) 9:2029–2041”
BackgroundSteele, A. M., Wensley, K. J., Brewer, E., Shields, B. M., Hattersley, A. T., & McDonald, T. J. (2013) · Practical Diabetes · doi:10.1002/pdi.1758 · Cited by 12
Hagvik, J. (2008) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2008.142 · Cited by 7
“should be stored on ice for not more than 1 h in a heparin/sodium fluoride tube until centrifugation (4).”
BackgroundYuan, Q., Wang, L., Zhu, X., Wu, Z., He, S., Zhang, H., & Lin, D. (2025) · Journal of Affective Disorders · doi:10.1016/j.jad.2025.121026 · Cited by 9
Nakanga, W. P., Balungi, P., Niwaha, A. J., Shields, B. M., Hughes, P. J., Andrews, R. C., Donald, T. J. M., Nyirenda, M., & Hattersley, A. T. (2022) · PLoS ONE · doi:10.1371/journal.pone.0264432 · Cited by 7
Should haemoglobin A1cbe used for the diagnosis of diabetes mellitus in South Africa?
George, J. A. (2011) · Journal of Endocrinology Metabolism and Diabetes of South Africa · doi:10.1080/22201009.2011.10872263 · Cited by 9
Preanalytical issues related to routine and diagnostic glucose tests: Results from a survey in Spain
del Pino, I. G., Ibarz, M., Rioja, R. G., Salas, P., Segovia, M., Puente, J. J., Marzana, I., Ventura, M., Martínez, D., Llopis, M. A., et al. (2020) · Biochemia Medica · doi:10.11613/bm.2020.010704 · Cited by 8
Effect of delay on measurement of blood glucose levels in young subjects with type 1 diabetes
Elleri, D., Acerini, C. L., Allen, J. M., Larsen, A. M. F., Wilinska, M. E., Dunger, D. B., & Hovorka, R. (2009) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2009.06.025 · Cited by 4
Bakkebø, H., Haaland, K. L., Hoff, K. S., & Schwettmann, L. (2024) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2024.2318617 · Cited by 5
“This is in agreement with earlier studies [23].”
ResultCreation of a low-risk reference group and referenceinterval of fasting venous plasma glucose
Jørgensen, L. G. M., Brandslund, I., Petersen, P. H., Stahl, M., & Olivarius, N. D. F. (2004) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2004.135 · Cited by 3
Lee, Y.-W., Cha, Y. J., Chae, S. L., Song, J., Yun, Y., Park, H., Seong, M., Whang, D. H., Kim, H. S., Kim, J., et al. (2009) · Annals of Laboratory Medicine · doi:10.3343/kjlm.2009.29.6.524 · Cited by 4
Zhang, L., Ran, Y., Zhu, Y., & Zhen, Q. (2020) · Laboratory Medicine · doi:10.1093/labmed/lmaa051 · Cited by 3
Olivarius, N. D. F., & Jørgensen, L. G. M. (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2006.1859_1.x · Cited by 3
Trends in obesity and type 2 diabetes : ethnic aspects and links to adipokines
Lilja, M. (2011) · DiVA at Umeå University (Umeå University) · Cited by 4
Özcürümez, M. K., Weninger, J., Coşkun, A., Arzideh, F., Streichert, T., Torge, A., Sowa, J., Quast, C., Canbay, A. E., Plebani, M., et al. (2024) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2024-0970 · Cited by 2
“Our model therefore appears to agree in principle with the current recommendations for glucose testing [26, 27].”
Jørgensen, L. G. M., Petersen, P. H., & Brandslund, I. (2006) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2006.057 · Cited by 2
Screening for gestational diabetes mellitus
Skitek, M. (2005) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2005.114 · Cited by 2
Welke omstandigheden kunnen een gemeten nuchtere bloedsuiker beïnvloeden?
Tavenier, D., & Reenders, K. (2010) · Bohn Stafleu van Loghum eBooks · doi:10.1007/978-90-313-7342-0_5 · Cited by 2
Kughapriya, P., & Elanchezhian, J. A. (2019) · Indian Journal of Medical Biochemistry · doi:10.5005/jp-journals-10054-0080 · Cited by 4
Reshma, D. C. (2019) · International Journal of Clinical Biochemistry and Research · doi:10.18231/j.ijcbr.2019.066 · Cited by 2
Samuel, J. K. R., G, K. V., & Chandrasekaran, S. (2022) · International Journal of Clinical Biochemistry and Research · doi:10.18231/j.ijcbr.2022.041 · Cited by 1
Sacks, D. B., Arnold, M. A., Metzger, B. E., & Nathan, D. M. (2012) · Cited by 1
Birkeland, K. I., Skeie, S., & Hanssen, K. F. (2009) · Norsk Epidemiologi · doi:10.5324/nje.v16i1.200
“Analyse i plasma (glass med Na fluorid/heparin) (eller serum) anbefales, sentrifugering og utskilling av plasma bør skje innen 30-60 minutter og plasma bør sendes til et laboratorium for våtkjemisk analyse (5,6).”
BackgroundMethodActions du fluorure de sodium et du mono-iodoacétate de sodium comme antiglycolytiques
Mavioga, E. M.-K., Verret, C., Berre, J. P. L., Ceppa, F., & Burnat, P. (2011) · Revue Francophone des Laboratoires · doi:10.1016/s1773-035x(11)70768-7
Könnun á stöðugleika glúkósa í þremur tegundum blóðsýnaglasa
Jónsdóttir, Þ. G., Hjaltadóttir, G., & Þorsteinsdóttir, I. (2012) · Hirsla (Landspítali University Hospital)
Sztefko, K. (2009) · Homo Politicus (Academy of Humanities and Economics in Lodz) · Cited by 2
Overestimation of Hypoglycemia in Infants with a High Hematocrit
Kemperman, H., van Solinge, W. W., & de Vooght, K. M. K. (2016) · The Journal of Applied Laboratory Medicine · doi:10.1373/jalm.2016.020164
A New Material Based on Clay and its Successful Application to Fix the Glucose
Torkia, B., Abedelkader, B., & Haouzi, A. (2017) · Oriental Journal Of Chemistry · doi:10.13005/ojc/330145
Effect of Preanalytical factors on Variation in Glucose Levels
Shabana, S., & Pradeepa, N. (2018) · IJAR - Indian Journal of Applied Research
Methoden der Stoffwechselkontrolle
(2006) · Springer eBooks · doi:10.1007/3-540-26602-x_10
Welke omstandigheden kunnen een gemeten nuchtere bloedsuiker beïnvloeden ?
Tavenier, D., & Reenders, K. (2006) · Bohn Stafleu van Loghum eBooks · doi:10.1007/978-90-313-8808-0_1045
Abara, U. S., Ibanga, I. E., Nwapi, L. L., Ekpo, M. I., Akpan, H. A., Nkechi, O., Mkpani, J. O., Joseph, N. I., Konglam, N. I., Nkwocha, B. C., et al. (2023) · Journal of Advances in Medical and Pharmaceutical Sciences · doi:10.9734/jamps/2023/v25i1595
Diabetes mellitus - laboratory diagnostics and monitoring of patients state
Friedecky, B., Kratochvíla, J., Springer, D., Prázný, M., & Zima, T. (2016) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2016.008 · Cited by 4
Diabetes mellitus - laboratory diagnostics and follow-up of patient condition
Friedecky, B., Zima, T., Kratochvíla, J., & Springer, D. (2012) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2012.021 · Cited by 4
Diagnostics of diabetes using plasma glucose. Unresolved problem of laboratory medicine
Friedecky, B., Palička, V., & Kratochvíla, J. (2017) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2017.031 · Cited by 1
Diabetes mellitus - laboratory diagnostics and patient status monitoring
Friedecky, B., Kratochvíla, J., Springer, D., Prázný, M., Pelikánová, T., Zima, T., & Racek, J. (2019) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2019.007 · Cited by 7
Metadata
- Authors
- M. Stahl, L. G. M. J.
- Venue
- Scandinavian Journal of Clinical and Laboratory Investigation
- Publisher
- Informa UK Limited
- Published
- 2001-01
- Volume
- 61
- Issue
- 3
- Pages
- 169-179
- ISSN
- 0036-5513 (ISSN-L)
- 1502-7686
- OpenAlex
- W2147966535
- Semantic Scholar
- 4f96cb128be838de276cc4317c4e6f0ad5097dfc
- Cited by
- 79
- References
- 16
- Citations by year
- 2025: 1 · 2024: 2 · 2023: 3 · 2022: 2 · 2020: 1 · 2019: 4 · 2018: 3 · 2017: 3 · 2016: 3 · 2015: 2 · 2014: 1 · 2013: 2 · 2012: 6
- FWCI
- 3.78
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Scandinavian Journal of Clinical and Laboratory Investigation (publishedVersion)
- PubMed (publishedVersion)
- Research at the University of Copenhagen (University of Copenhagen) (submittedVersion)
- University of Southern Denmark Research Portal (University of Southern Denmark) (submittedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- plasma glucose, type 2 diabetes diagnosis, diagnostic criteria, glucose threshold, sample storage, sample transport, plasma separation, sodium fluoride, glucose stability, analytical bias
- MeSH terms
- Anticoagulants, Blood Glucose, Centrifugation, Chemistry, Clinical, Diabetes Mellitus, Type 2, Humans, Societies, Medical, Specimen Handling, World Health Organization, Reproducibility of Results, Cryopreservation, Guidelines as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.