Absolute frequency measurement of the 88 Sr + clock transition using a GPS link to the SI second
Open accessThe DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index.
Abstract
We report the results of a recent measurement of the absolute frequency of the 5 s 2 S 1 / 2 – 4 d 2 D 5 / 2 transition of the 88 S r + ion. The optical frequency was measured against the international atomic time realization of the SI second on the geoid as obtained by frequency transfer using a global positioning system link and the precise point positioning technique. The measurement campaign yielded more than 100 h of frequency data. It was performed with improvements to the stability and accuracy of the single-ion clock compared to the last measurement made in 2012. The single ion clock uncertainty is evaluated at 1.5 × 10 − 17 when contributions from acousto-optic modulator frequency chirps and servo errors are taken into account. The stability of the ion clock is 3 × 10 − 15 at 1 s averaging, a factor of three better than in the previous measurement. The results from the two measurement campaigns are in good agreement. The uncertainty of the measurement, primarily from the link to the SI second, is 0.75 Hz ( 1.7 × 10 − 15 ). The frequency measured for the S – D clock transition of 88 S r + is ν 0 = 444 779 044 095 485.27 ( 75 ) Hz.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
Frequency Ratio of Al+and Hg+Single-Ion Optical Clocks; Metrology at the 17th Decimal Place
Rosenband, T., Hume, D. B., Schmidt, P. O., Chou, C. W., Brusch, A., Lorini, L., Oskay, W. H., Drullinger, R. E., Fortier, T. M., Stalnaker, J. E., et al. (2008) · Science · doi:10.1126/science.1154622 · Cited by 1,450
Godun, R. M., Nisbet-Jones, P. B. R., Jones, J. M., King, S. A., Johnson, L. A. M., Margolis, H. S., Szymaniec, K., Lea, S. N., Bongs, K., & Gill, P. (2014) · Physical Review Letters · doi:10.1103/physrevlett.113.210801 · Cited by 468
Improved Limit on a Temporal Variation ofmp/mefrom Comparisons ofYb+and Cs Atomic Clocks
Huntemann, N., Lipphardt, B., Tamm, C., Gerginov, V., Weyers, S., & Peik, E. (2014) · Physical Review Letters · doi:10.1103/physrevlett.113.210802 · Cited by 366
Hertz-Level Measurement of the Optical Clock Frequency in a Single 88 Sr + Ion
Margolis, H. S., Barwood, G. P., Huang, G., Klein, H. A., Lea, S. N., Szymaniec, K., & Gill, P. (2004) · Science · doi:10.1126/science.1105497 · Cited by 301
Porsev, S. G., & Derevianko, A. (2006) · Physical Review A · doi:10.1103/physreva.74.020502 · Cited by 279
Fortier, T. M., Ashby, N., Bergquist, J. C., Delaney, M. J., Diddams, S. A., Heavner, T. P., Hollberg, L., Itano, W. M., Jefferts, S. R., Kim, K., et al. (2007) · Physical Review Letters · doi:10.1103/physrevlett.98.070801 · Cited by 226
Shift ofS122hyperfine splittings due to blackbody radiation
Itano, W. M., Lewis, L. L., & Wineland, D. J. (1982) · Physical Review A · doi:10.1103/physreva.25.1233 · Cited by 219
Geodetic techniques for time and frequency comparisons using GPS phase and code measurements
Ray, J., & Senior, K. (2005) · Metrologia · doi:10.1088/0026-1394/42/4/005 · Cited by 173
Schrama, C. A., Peik, E., Smith, W. W., & WALTHER, H. (1993) · Optics Communications · doi:10.1016/0030-4018(93)90318-y · Cited by 172
1 × 10−16frequency transfer by GPS PPP with integer ambiguity resolution
Petit, G., Kanj, A., Loyer, S., Delporte, J., Mercier, F., & Perosanz, F. (2015) · Metrologia · doi:10.1088/0026-1394/52/2/301 · Cited by 168
Madej, A. A., Dubé, P., Zhou, Z., Bernard, J. E., & Gertsvolf, M. (2012) · Physical Review Letters · doi:10.1103/physrevlett.109.203002 · Cited by 155
Self-citationElectric Quadrupole Shift Cancellation in Single-Ion Optical Frequency Standards
Dubé, P., Madej, A. A., Bernard, J. E., Marmet, L., Boulanger, J.-S., & Cundy, S. (2005) · Physical Review Letters · doi:10.1103/physrevlett.95.033001 · Cited by 152
Self-citationEvaluation of systematic shifts of the88Sr+single-ion optical frequency standard at the10−17level
Dubé, P., Madej, A. A., Zhou, Z., & Bernard, J. E. (2013) · Physical Review A · doi:10.1103/physreva.87.023806 · Cited by 127
Self-citationLaser frequency stabilization to a single ion
Peik, E., Schneider, T., & Tamm, C. (2006) · Journal of Physics B: Atomic, Molecular and Optical Physics · doi:10.1088/0953-4075/39/1/012 · Cited by 125 · cited 1×
“For an alternative implementation of cavity drift rate compensation see, for example, [25].”
BackgroundDubé, P., Madej, A. A., Tibbo, M., & Bernard, J. E. (2014) · Physical Review Letters · doi:10.1103/physrevlett.112.173002 · Cited by 106
Self-citationWhen should we change the definition of the second?
Gill, P. (2011) · Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences · doi:10.1098/rsta.2011.0237 · Cited by 101
Madej, A. A., Bernard, J. E., Dubé, P., Marmet, L., & Windeler, R. S. (2004) · Physical Review A · doi:10.1103/physreva.70.012507 · Cited by 92
Self-citationAgreement between two88Sr+optical clocks to 4 parts in 1017
Barwood, G. P., Huang, G., Klein, H. A., Johnson, L. A. M., King, S. A., Margolis, H. S., Szymaniec, K., & Gill, P. (2014) · Physical Review A · doi:10.1103/physreva.89.050501 · Cited by 90
A laser frequency lock referenced to a single trapped ion
Bernard, J. E., Marmet, L., & Madej, A. A. (1998) · Optics Communications · doi:10.1016/s0030-4018(98)00121-7 · Cited by 88
Self-citationBlackbody-radiation shift in a88Sr+ion optical frequency standard
Jiang, D., Arora, B., Safronova, M. S., & Clark, C. W. (2009) · Journal of Physics B: Atomic, Molecular and Optical Physics · doi:10.1088/0953-4075/42/15/154020 · Cited by 79
Dubé, P., Madej, A. A., Bernard, J. E., Marmet, L., & Shiner, A. D. (2009) · Applied Physics B · doi:10.1007/s00340-009-3390-6 · Cited by 72
Self-citationPetit, G. (2009) · Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS) · doi:10.1109/freq.2009.5168153 · Cited by 61
Sr88+single-ion optical clock with a stability approaching the quantum projection noise limit
Dubé, P., Madej, A. A., Shiner, A., & Jian, B. (2015) · Physical Review A · doi:10.1103/physreva.92.042119 · Cited by 50
Self-citationThe Canadian Geodetic Vertical Datum of 2013 (CGVD2013)
Veronneau, M., & Huang, J. (2016) · Geomatica · doi:10.5623/cig2016-101 · Cited by 47
Margolis, H. S., & Gill, P. (2015) · Metrologia · doi:10.1088/0026-1394/52/5/628 · Cited by 40
Is the time right for a redefinition of the second by optical atomic clocks?
Gill, P. (2016) · Journal of Physics: Conference Series · doi:10.1088/1742-6596/723/1/012053 · Cited by 37
Influence of Chirped Excitation Pulses in an Optical Clock With Ultracold Calcium Atoms
Degenhardt, C., Nazarova, T., Lisdat, C., Stoehr, H., Sterr, U., & Riehle, F. (2005) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2004.843388 · Cited by 31 · cited 2×
“Modulation of the rf power causes temperature variations in the TeO2 crystal of the AOM that can result in frequency shifts [6, 29].”
BackgroundResultHachisu, H., Petit, G., & Ido, T. (2017) · Applied Physics B · doi:10.1007/s00340-016-6603-9 · Cited by 31
Broadband, unpolarized repumping and clearout light sources for Sr^+ single-ion clocks
Fordell, T., Lindvall, T., Dubé, P., Madej, A. A., Wallin, A. E., & Merimaa, M. (2015) · Optics Letters · doi:10.1364/ol.40.001822 · Cited by 21
Self-citationPorsev, S. G., & Derevianko, A. (2012) · Physical Review A · doi:10.1103/physreva.86.029904 · Cited by 19
Lindvall, T., Merimaa, M., Tittonen, I. J., & Madej, A. A. (2012) · Physical Review A · doi:10.1103/physreva.86.033403 · Cited by 18
Atomic Quadrupole Moment Measurement Using Dynamic Decoupling
Shaniv, R., Akerman, N., & Ozeri, R. (2016) · Physical Review Letters · doi:10.1103/physrevlett.116.140801 · Cited by 17
Optical Atomic Clocks Could Redefine Unit of Time
Riehle, F. (2012) · Physics · doi:10.1103/physics.5.126 · Cited by 13
Bernard, J. E., Dubé, P., Madej, A. A., Cundy, S., Boulanger, J.-S., Jiang, J., & Jones, D. J. (2010) · The European Physical Journal D · doi:10.1140/epjd/e2010-00052-5 · Cited by 12
Self-citationStatus of the atomic fountain clock at the National Research Council of Canada
Beattie, S., Alcock, J. A., Jian, B., Gertsvolf, M., & Bernard, J. E. (2016) · Journal of Physics Conference Series · doi:10.1088/1742-6596/723/1/012008 · Cited by 6 · cited 1×
“Furthermore, a Cs fountain clock is expected to become available at NRC in the near future for direct absolute frequency measurements at the low 10−16 level [38].”
BackgroundSelf-citationMeasurement of the static scalar polarizability of the 88Sr+ clock transition
Dubé, P., Madej, A. A., Tibbo, M., & Bernard, J. E. (2014) · European Frequency and Time Forum · doi:10.1109/eftf.2014.7331531 · Cited by 4 · cited 1×
“The suppression level is limited by the dif ference between the restoring forces in the axial and radial directions (Mathieu parameters qr and qz) rather than by the uncertainty of α ∆ 0; the optimum Ω0 is slightly different for motion along the axial direction than it is for motion along the radial direction [23, 24…”
BackgroundSelf-citation
Research works printed but not linked: 1
- Mireault, 2008, Online precise point positioning: a new, timely service from Natural Resources Canada, GPS World, 19, 59
Cited by
Mehlstäubler, T. E., Grosche, G., Lisdat, C., Schmidt, P. O., & Denker, H. (2018) · Reports on Progress in Physics · doi:10.1088/1361-6633/aab409 · Cited by 263
Pizzocaro, M., Bregolin, F., Barbieri, P., Rauf, B., Levi, F., & Calonico, D. (2020) · Metrologia · doi:10.1088/1681-7575/ab50e8 · Cited by 63
Hobson, R., Bowden, W., Vianello, A., Silva, A., Baynham, C. F. A., Margolis, H. S., Baird, P. E. G., Gill, P., & Hill, I. R. (2020) · Metrologia · doi:10.1088/1681-7575/abb530 · Cited by 58
First accuracy evaluation of the NRC-FCs2 primary frequency standard
Beattie, S., Jian, B., Alcock, J. A., Gertsvolf, M., Hendricks, R., Szymaniec, K., & Gibble, K. (2020) · Metrologia · doi:10.1088/1681-7575/ab7c54 · Cited by 52
Self-citationThe CIPM list ‘Recommended values of standard frequencies’: 2021 update
Margolis, H. S., Panfilo, G., Petit, G., Oates, C. W., Ido, T., & Bize, S. (2024) · Metrologia · doi:10.1088/1681-7575/ad3afc · Cited by 48
Metadata
- Authors
- Dubé, P., Bernard, J. E., & Gertsvolf, M.
- Affiliations
- National Research Council Canada (CA)
- Corresponding author
- Dubé, Pierre
- Venue
- Metrologia
- Publisher
- IOP Publishing
- Published
- 2017-03-05
- Volume
- 54
- Issue
- 3
- Pages
- 290-298
- ISSN
- 0026-1394 (ISSN-L)
- 1681-7575
- OpenAlex
- W2605209510
- Cited by
- 28
- References
- 36
- Citations by year
- 2025: 1 · 2024: 1 · 2023: 4 · 2022: 4 · 2021: 1 · 2020: 4 · 2019: 5 · 2018: 5 · 2017: 3
- FWCI
- 2.28
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- publishingsupport.iopscience.iop.org/iop-standard/v1 (vor)
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- Locations
- Metrologia (publishedVersion)
- NPARC (submittedVersion)
- Funders
- National Research Council Canada
- Natural Resources Canada
- Topics
- Physical Sciences › Physics and Astronomy › Atomic and Molecular Physics, and Optics › Advanced Frequency and Time Standards
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced Electrical Measurement Techniques
- Keywords
- optical frequency measurement, SI second, frequency transfer, precise point positioning, frequency stability, frequency metrology
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.