Student or educator? Sign up with your school email and get 6 months of Pro free.
CoChat

The DOI matches its Crossref record; 56 of the 56 references it cites were found in a registry or an index.

High-density peripheral nerve cuffs restore natural sensation to individuals with lower-limb amputations

Open access

Abstract

Objective . Sensory input in lower-limb amputees is critically important to maintaining balance, preventing falls, negotiating uneven terrain, responding to unexpected perturbations, and developing the confidence required for societal participation and public interactions in unfamiliar environments. Despite noteworthy advances in robotic prostheses for lower-limb amputees, such as microprocessor knees and powered ankles, natural somatosensory feedback from the lost limb has not yet been incorporated in current prosthetic technologies. Approach . In this work, we report eliciting somatic sensation with neural stimulation delivered by chronically-implanted, non-penetrating nerve cuff electrodes in two transtibial amputees. High-density, flexible, 16-contact nerve cuff electrodes were surgically implanted for the selective activation of sensory fascicles in the nerves of the posterior thigh above the knee. Electrical pulses at safe levels were delivered to the nerves by an external stimulator via percutaneous leads attached to the cuff electrodes. Main results . The neural stimulation was perceived by participants as sensation originating from the missing limb. We quantitatively and qualitatively ascertained the intensity, modality as well as the location and stability of the perceived sensations. Stimulation through individual contacts within the nerve cuffs evoked repeatable sensations of various modalities and at discrete locations projected to the missing toes, foot and ankle, as well as in the residual limb. In addition, we observed a high overlap in reported locations between distal versus proximal cuffs suggesting that the same sensory responses could be elicited from more proximal points on the nerve. Significance . Based on these findings, the high-density cuff technology is suitable for restoring natural sensation to lower-limb amputees and could be utilized in developing a neuroprosthesis with natural sensory feedback. The overlap in reported locations between proximal and distal cuffs indicates that our approach might be applicable to transfemoral amputees where distal muscles and branches of sciatic nerve are not available.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1088/1741-2552/aac964

Cites

56 of 56 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 (a) Illustration of implanted system and its components. The C-FINEs each had 16 contacts and wrapped around the nerve ((a) (i) and (ii)). Access to the C-FINEs was through percutaneous leads routed to the upper thigh where they exited the body. The connection between the C-FINEs and the percutaneous leads was through inline connectors (a)(iv). An external stimulator was utilized to deliver electrical stimulation to the nerves (a)(iii). (b) The location of implanted C-FINEs for each subject. The C-FINEs were implanted to allow access to proximal and distal points of the nerves in the popliteal fossa.
Figure 1 (a) Illustration of implanted system and its components. The C-FINEs each had 16 contacts and wrapped around the nerve ((a) (i) and (ii)). Access to the C-FINEs was through percutaneous leads routed to the upper thigh where they exited the body. The connection between the C-FINEs and the percutaneous leads was through inline connectors (a)(iv). An external stimulator was utilized to deliver electrical stimulation to the nerves (a)(iii). (b) The location of implanted C-FINEs for each subject. The C-FINEs were implanted to allow access to proximal and distal points of the nerves in the popliteal fossa.CC-BY · From the PDF
Figure 2 (a) Locations reported by LE01 when electrodes in the sciatic C-FINE were stimulated at threshold pulse width and pulse amplitude. Color indicates different contacts, numbered 1 through 16, where contacts 1 and 16 were on the medial side of the nerve while contacts 8 and 9 were on the lateral side of the nerve. (b) Overlap of locations reported by LE01 during stimulation of the electrodes on the sciatic nerve with locations reported during stimulation of electrodes on the common peroneal (blue) and tibial (yellow) nerves.
Figure 2 (a) Locations reported by LE01 when electrodes in the sciatic C-FINE were stimulated at threshold pulse width and pulse amplitude. Color indicates different contacts, numbered 1 through 16, where contacts 1 and 16 were on the medial side of the nerve while contacts 8 and 9 were on the lateral side of the nerve. (b) Overlap of locations reported by LE01 during stimulation of the electrodes on the sciatic nerve with locations reported during stimulation of electrodes on the common peroneal (blue) and tibial (yellow) nerves.CC-BY · From the PDF
Figure 3 The locations reported by LE02 at threshold pulse width and pulse amplitude levels for the tibial C-FINE. Color indicates different contacts, numbered 1 through 16, where contacts 1 and 16 were on the lateral side of the nerve while contacts 8 and 9 were on the medial side of the nerve. The dotted arrows indicate a pre-movement sensation where the subject reported that his ankle was about to move down but that he did not feel any movement.
Figure 3 The locations reported by LE02 at threshold pulse width and pulse amplitude levels for the tibial C-FINE. Color indicates different contacts, numbered 1 through 16, where contacts 1 and 16 were on the lateral side of the nerve while contacts 8 and 9 were on the medial side of the nerve. The dotted arrows indicate a pre-movement sensation where the subject reported that his ankle was about to move down but that he did not feel any movement.CC-BY · From the PDF
Figure 4 Changes in the perceived modality with increases in stimulation charge density. (a) Data shown from representative C-FINE contacts in both subjects. (b) The distribution of tactile versus proprioceptive sensations as a function of charge density normalized to the value at threshold for the all of the contacts that responded in LE01 ' s sciatic C-FINE and LE02 ' s tibial C-FINE.
Figure 4 Changes in the perceived modality with increases in stimulation charge density. (a) Data shown from representative C-FINE contacts in both subjects. (b) The distribution of tactile versus proprioceptive sensations as a function of charge density normalized to the value at threshold for the all of the contacts that responded in LE01 ' s sciatic C-FINE and LE02 ' s tibial C-FINE.CC-BY · From the PDF
Image 5
CC-BY · From the PDF
Figure 5 Intensity discrimination as a function of PW. (a) The intensity discrimination data (black dots) and the fitted curve (dashed line) for contact S12 in LE01. The data points indicate the probability of presented stimuli to be judged correctly, i.e., stronger or weaker than the reference PW. (b) Combined discrimination curves for both subjects ( n = 2 contacts per subject, solid line is the mean and the shaded area denotes the standard error of the mean).
Figure 5 Intensity discrimination as a function of PW. (a) The intensity discrimination data (black dots) and the fitted curve (dashed line) for contact S12 in LE01. The data points indicate the probability of presented stimuli to be judged correctly, i.e., stronger or weaker than the reference PW. (b) Combined discrimination curves for both subjects ( n = 2 contacts per subject, solid line is the mean and the shaded area denotes the standard error of the mean).CC-BY · From the PDF
Figure 6 Strength-duration curves measured 2 and 3 months post-surgery for LE01 (a) and LE02 (b) with default 50 ms inter-pulse intervals (IPIs). We also compared SD curves at 10 ms versus 50 ms IPIs for LE01 (c) and LE02 (d) to examine changes in SD time constant ( τ SD) and base current ( I 0).
Figure 6 Strength-duration curves measured 2 and 3 months post-surgery for LE01 (a) and LE02 (b) with default 50 ms inter-pulse intervals (IPIs). We also compared SD curves at 10 ms versus 50 ms IPIs for LE01 (c) and LE02 (d) to examine changes in SD time constant ( τ SD) and base current ( I 0).CC-BY · From the PDF
Figure 7 The distance to the phantom big toe from the end of the residuum over time reported by LE01. Distances reported before (blue) and after (red) stimulation was applied in a given day are shown as points for measurements from individual days while the average across a month are shown as bars. The phantom toe moved farther away from the end of the residuum during month 1 and the beginning of month 2, then remained closer to the floor in subsequent months compared to month 1. ¤ indicates a significant difference compared to month 1 (post-hoc Tukey test ( p < 0.05) across measurements before and after stimulation). ‡ indicates a significant difference between months 4 -6 and months 10 -12 (post-hoc Tukey test ( p < 0.05) across measurements before and after stimulation). There is also a significant difference between measurements before and after stimulation (main effect, p = 0.011).
Figure 7 The distance to the phantom big toe from the end of the residuum over time reported by LE01. Distances reported before (blue) and after (red) stimulation was applied in a given day are shown as points for measurements from individual days while the average across a month are shown as bars. The phantom toe moved farther away from the end of the residuum during month 1 and the beginning of month 2, then remained closer to the floor in subsequent months compared to month 1. ¤ indicates a significant difference compared to month 1 (post-hoc Tukey test ( p < 0.05) across measurements before and after stimulation). ‡ indicates a significant difference between months 4 -6 and months 10 -12 (post-hoc Tukey test ( p < 0.05) across measurements before and after stimulation). There is also a significant difference between measurements before and after stimulation (main effect, p = 0.011).CC-BY · From the PDF
Figure 8 The yield for each of the C-FINEs (indicated by line style and marker type) over time after surgery across two subjects, LE01 (blue) and LE02 (black).
Figure 8 The yield for each of the C-FINEs (indicated by line style and marker type) over time after surgery across two subjects, LE01 (blue) and LE02 (black).CC-BY · From the PDF
Figure 9 The charge density thresholds to elicit sensation in LE01 (a) and LE02 (b) over time for each of the C-FINEs (indicated by line style and marker type). Markers indicate the mean for a given month post-surgery and error bars represent standard error of the mean. For the sake of clarity, only the top or bottom half of the error bars are shown.
Figure 9 The charge density thresholds to elicit sensation in LE01 (a) and LE02 (b) over time for each of the C-FINEs (indicated by line style and marker type). Markers indicate the mean for a given month post-surgery and error bars represent standard error of the mean. For the sake of clarity, only the top or bottom half of the error bars are shown.CC-BY · From the PDF

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 56 references

Cited by

See all 131

Metadata

Authors
Charkhkar, H., Shell, C. E., Marasco, P. D., Pinault, G. J., Tyler, D. J., & Triolo, R. J.
Affiliations
  • Louis Stokes Cleveland VA Medical Center (US)
  • Cleveland Clinic (US)
Corresponding author
Charkhkar, Hamid
Venue
Journal of Neural Engineering
Published
2018-07-02
Volume
15
Issue
5
Pages
056002
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W2806169161
Cited by
131
References
56
Citations by year
2026: 10 · 2025: 22 · 2024: 12 · 2023: 22 · 2022: 15 · 2021: 21 · 2020: 8 · 2019: 20 · 2018: 1
FWCI
5.71
Type
journal article
Language
en
Open access
Licence
Locations
  • Journal of Neural Engineering (publishedVersion, cc-by)
  • PubMed (publishedVersion)
Funders
Defense Advanced Research Projects Agency (N66001-15-C-4038)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Health Sciences › Medicine › Anesthesiology and Pain Medicine › Pain Management and Treatment
  • Physical Sciences › Engineering › Biomedical Engineering › Prosthetics and Rehabilitation Robotics
Keywords
lower limb amputation, transtibial amputees, sensory feedback, neurostimulation, somatosensory feedback, neural prostheses, sciatic nerve, percutaneous leads, sensory perception
MeSH terms
Aged, Amputation, Traumatic, Amputees, Artificial Limbs, Electric Stimulation, Electrodes, Humans, Male, Middle Aged, Peripheral Nerves, Phantom Limb, Prosthesis Design, Sensation Disorders, Sensory Thresholds, Lower Extremity, Neural Prostheses

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: High-density peripheral nerve cuffs restore natural sensation to individuals with lower-limb amputations. https://app.cochat.ai/papers/10.1088/1741-2552/aac964
The paper
Charkhkar, H., Shell, C. E., Marasco, P. D., Pinault, G. J., Tyler, D. J., & Triolo, R. J. (2018). High-density peripheral nerve cuffs restore natural sensation to individuals with lower-limb amputations. Journal of Neural Engineering. https://doi.org/10.1088/1741-2552/aac964

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.