Fascicle specific targeting for selective peripheral nerve stimulation

Open access
Overstreet, Cynthia K, Cheng, Jonathan, Keefer, Edward W·2019·Journal of Neural Engineering

Abstract

OBJECTIVE: Electrical stimulation is a blunt tool for evoking neural activity. Neurons are naturally activated asynchronously and non-uniformly, whereas stimulation drives simultaneous activity within a population of cells. These differences in activation pattern can result in unintended side effects, including muddled sensory percepts and undesirable muscle contractions. These effects can be mitigated by the placement of electrodes in close approximation to nerve fibers and careful selection of the neural interface's location. This work describes the benefits of placing electrodes within specific fascicles of peripheral nerve to form selective neural interfaces for bidirectional neuroprosthetic devices. APPROACH: Chronic electrodes were targeted to individual fascicles of the ulnar and median nerves in the forearm of four human subjects. During the surgical implant procedure, fascicles were dissected from each nerve, and functional testing was used to identify the relative composition of sensory and motor fibers within each. FAST-LIFE arrays, composed of longitudinal intrafascicular arrays and fascicular cuff electrodes, were implanted in each fascicle. The location, quality, and stimulation parameters associated with sensations evoked by electrical stimulation on these electrodes were characterized throughout the 90-180 d implant period. MAIN RESULTS: FAST-LIFE arrays enable selective and chronic electrical stimulation of individual peripheral nerve fascicles. The quality of sensations evoked by stimulation in each fascicle is predictable and distinct; subjects reported tactile and cutaneous sensations during stimulation of sensory fascicles and deeper proprioceptive sensations during stimulation of motor fascicles. Stimulation thresholds and strength-duration time constants were typically higher within sensory fascicles. SIGNIFICANCE: Highly selective, stable neural interfaces can be created by placing electrodes within and around single fascicles of peripheral nerves. This method enables targeting electrodes to nerve fibers that innervate a specific body region or have specific functions. Fascicle-specific interfacing techniques have broad potential to maximize the therapeutic effects of electrical stimulation in many neuromodulation applications. (Clinical Trial ID NCT02994160.).

Integrity signals

Registry

Crossref · 10.1088/1741-2552/ab4370

Cites

45 of 45 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Claims the paper makes

10 of 33 cited sources checked · 6 supported · 1 flagged · accuracy not yet measured

  • Premisecitation

    The location and quality of sensations evoked by stimulation of implanted electrode arrays can change significantly.

    “Additionally, the effects of stimulation can be difficult to predict; both the location and quality of sensations evoked by stimulation of implanted electrode arrays can change significantly [14,15].”
    Quote in the paperSubstantial claim1 of 2 sources checked
    • cites [14] Charkhkar, 2018, High-density peripheral nerve cuffs restore natural sensation to individuals with lower-limb amputations, J. Neural Eng. source
    • cites [15] Dhillon, 2005, Effects of short-term training on sensory and motor function in severed nerves of long-term human amputees, J. Neurophysiol. source
  • Premisecitation

    Some degree of functional separation (afferent versus efferent) is often present in peripheral nerve fascicles.

    “Additionally, some degree of functional separation (afferent versus efferent) is often present [25,26].”
    Quote in the paperSubstantial claim1 of 2 sources checked
    • cites [25] Delgado-Martínez, 2016, Fascicular topography of the human median nerve for neuroprosthetic surgery, Frontiers Neurosci. source

      Results §3.3 reports that motor axons in the human median nerve were concentrated in patches rather than randomly mingled with other axons, supporting some degree of functional separation. It does not establish that afferent–efferent separation is often present across peripheral nerve fascicles; purely motor or sensory fascicles were described as exceptional. (LLM, conf=0.84)

      Experimental — accuracy not yet measured
    • cites [26] Chow, 1985, Surgical significance of the motor fascicular group of the ulnar nerve in the forearm, J. Hand Surg. source
  • Premisecitation

    Median nerve fascicles are less frequently unimodal.

    “however, median nerve fascicles are less frequently unimodal [25].”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [25] Delgado-Martínez, 2016, Fascicular topography of the human median nerve for neuroprosthetic surgery, Frontiers Neurosci. source

      Results §3.3 reports that purely motor or purely sensory fascicles were exceptional at the sampled supraepicondylar levels, supporting the idea that unimodal fascicles were uncommon there. It does not establish a comparison with other nerves or show that this holds throughout the median nerve. (LLM, conf=0.82)

      Experimental — accuracy not yet measured
  • Premisecitation

    Electrical stimulation of peripheral nerves can produce undesirable side effects, such as cocontraction of muscles or unnatural sensory percepts.

    “Stimulation can produce undesirable side effects, such as cocontraction of muscles [11,12] or unnatural sensory percepts [8,13].”
    Quote in the paperSubstantial claim1 of 4 sources checked
    • cites [11] McDonnall, 2004, Selective motor unit recruitment via intrafascicular multielectrode stimulation, Can. J. Physiol. Pharmacol. source
    • cites [12] Polasek, 2007, Human nerve stimulation thresholds and selectivity using a multi-contact nerve cuff electrode, IEEE Trans. Neural Syst. Rehabil. Eng. source
    • cites [8] Tan, 2014, A neural interface provides long-term stable natural touch perception, Sci. Transl. Med. source
    • cites [13] Kiernan, 1997, Paraesthesiae induced by prolonged high frequency stimulation of human cutaneous afferents, J. Physiol. source
  • Premisecitation

    Fascicles contain groupings of nerve fibers organized somatotopically.

    “Fascicles contain groupings of nerve fibers that are organized somatotopically [24].”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [24] Hallin, 1990, Microneurography in relation to intraneural topography: somatotopic organisation of median nerve fascicles in humans, J. Neurol. Neurosurg. Psychiatry source
  • May combine two claims. 14
  • Classified as probably not a substantial claim. 8
  • May combine two claims.

    Premisecitation

    Other electrode designs used within peripheral nerve are less specific but allow access to a greater proportion of the nerve via electrical stimulation.

    “other electrode designs used within peripheral nerve are less specific but allow access to a greater proportion of the nerve via electrical stimulation [33].”
    Quote in the paperSubstantial claim1 of 1 sources checked
    • cites [33] Badia, 2011, Comparative analysis of transverse intrafascicular multichannel, longitudinal intrafascicular and multipolar cuff electrodes for the selective stimulation of nerve fascicles, J. Neural Eng. source

      The Introduction (p. 3) says intrafascicular positioning enhances selectivity, while TIME’s distributed sites let it access several fascicles. The Selectivity of stimulation section (p. 12) reports that TIME had higher selectivity than both LIFE and cuff electrodes. Thus, the intraneural design with broader access was not less selective in this comparison; the less-selective cuffs are placed around, not within, the nerve. (LLM, conf=0.82)

      Checked by a reasoning model — may be wrong.

      Experimental — accuracy not yet measured
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Intrafascicular stimulation has been used for many years to study the function and organization of fibers within peripheral nerve.

    “Intrafascicular stimulation has been used for many years to study the function and organization of fibers within peripheral nerve [24,34].”
    Quote in the paperNot a substantial claim1 of 2 sources checked
    • cites [24] Hallin, 1990, Microneurography in relation to intraneural topography: somatotopic organisation of median nerve fascicles in humans, J. Neurol. Neurosurg. Psychiatry source
    • cites [34] Johansson, 1978, Tactile sensibility in the human hand: receptive field characteristics of mechanoreceptive units in the glabrous skin area, J. Physiol. source
  • May combine two claims.

    Findingown data

    Some sensations were occasionally reported outside the expected region; these oddball sensations were often highly repeatable within a single session.

    “Some sensations were occasionally reported outside this region; these oddball sensations were often highly repeatable within a single session and should not be discarded as false positive reports.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    In partial hand amputees, stimulation of fascicles with lower SSEP activity produced stronger contractions of residual thenar and intrinsic muscles, confirming these fascicles contained a larger proportion of motor fibers.

    “In partial hand amputees, where the distal hand muscles are still present, stimulation of the fascicles with lower SSEP activity produced stronger contractions of the residual thenar and intrinsic muscles of the hand, confirming the hypothesis that these fascicles contained a larger proportion of motor fibers.”
    Quote in the paperSubstantial claim
    • Rests on an assumption: Stronger contraction of residual thenar and intrinsic muscles during fascicle stimulation indicates a larger proportion of motor fibers in that fascicle.
  • May combine two claims.

    Findingnone given

    Although muscle contraction sometimes accompanied the sensation, the sensory descriptions persisted even when no surface EMG was detectable.

    “Although muscle contraction sometimes accompanied the sensation, these sensory descriptions persisted even when no surface EMG was detectable.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Stimulation of both motor and sensory fascicles in the median nerve of the transradial amputee predominantly produced cutaneous sensations; weaker deep proprioceptive sensations were occasionally evoked.

    “Stimulation of both the motor and sensory fascicles in the median nerve of the transradial amputee predominantly produced cutaneous sensations; weaker, deep proprioceptive sensations were occasionally evoked.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    Stimulation of median nerve FAST-LIFE arrays produced smaller sensory maps than the ulnar fascicles, but there was not a significant difference in the area of sensations evoked by the two median nerve arrays.

    “Stimulation of these fascicles produced smaller sensory maps than the ulnar fascicles, but there was not a significant difference in the area of the sensations evoked by stimulation of the two median nerve FAST-LIFE arrays.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Both the sensory and motor fascicles of median nerve produced sensations in the thenar eminence, thumb, and index finger as expected.

    “Both the sensory and motor fascicles of median nerve produced sensations in the thenar eminence, thumb, and index finger as expected.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).

    Findingnone given

    A variety of sensations could also be evoked by stimulation of these electrodes over the same duration.

    “Analysis of longitudinal stimulation data was not included in this work, however a variety of sensations could also be evoked by stimulation of these electrodes over the same duration.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Findingown data

    Stimulation of motor fascicles often evoked proprioceptive sensations, but some nonspecific cutaneous sensations were also evoked by stimulation of these fascicles.

    “Stimulation of motor fascicles often evoked proprioceptive sensations, but some nonspecific cutaneous sensations were also evoked by stimulation of these fascicles.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Findingown data

    These sensations were reported to be vivid, easy to localize, and did not fade appreciably during prolonged stimulation.

    “These sensations were reported to be vivid, easy to localize, and did not fade appreciably during prolonged stimulation.”
    Quote in the paperSubstantial claim
  • May combine two claims.

    Premisecitation

    Peripheral nerve contains both afferent and efferent fibers, and multiple types of sensory fibers whose activation produces sensations with distinct qualities.

    “Peripheral nerve contains both afferent and efferent fibers; additionally, multiple types of sensory fibers are present, and activation of these fibers produce sensations with distinct qualities [21,22].”
    Quote in the paperSubstantial claim0 of 2 sources checked
    • cites [21] Johansson, 1983, Tactile sensory coding in the glabrous skin of the human hand, Trends Neurosci. source
    • cites [22] Torebjörk, 1980, Specific sensations evoked by activity in single identified sensory units in man, Acta Physiol. Scand. source
  • May combine two claims.

    Premisecitation

    Intrafascicular electrodes penetrate the perineurium of individual fascicles and enable highly selective recruitment.

    “Intrafascicular electrodes penetrate the perineurium of individual fascicles and are thus positioned immediately adjacent to the target nerve fibers, enabling highly selective recruitment [20].”
    Quote in the paperSubstantial claim0 of 1 sources checked
    • cites [20] Yoshida, 1993, Selective stimulation of peripheral nerve fibers using dual intrafascicular electrodes, IEEE Trans. Biomed. Eng. source
  • May combine two claims.

    Interpretationown data

    The FAST-LIFE design and fascicle-targeting strategy used in this study enables electrical stimulation of peripheral nerve that is both selective and specific.

    “The FAST-LIFE design and fascicle-targeting strategy utilized in this study enables electrical stimulation of peripheral nerve which is both selective and specific.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Interpretationreasoning

    The variation in locations of ulnar motor sensations between subjects is likely due, in part, to effects of partial hand amputation and revision procedures on palm musculature.

    “This is likely due, in part, to effects of partial hand amputation and revision procedures which affect the musculature in the palm.”
    Quote in the paperNot a substantial claim
  • May combine two claims.

    Study design and conductown data

    The curve fits for the sensory electrodes are approximations (R2 ≅ 0.85), as the subject reported that no pulse durations at or below 0.1 ms produced sensations of comparable quality or intensity on those electrodes.

    “The curve fits for the sensory electrodes are approximations (R 2 ≅ 0.85), as the subject reported that no pulse durations at or below 0.1 ms produced sensations of comparable quality or intensity on those electrodes.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Interpretationreasoning

    The ability to address individual nerve fascicles makes FAST-LIFE an attractive method for interfaces with even larger nerves, such as the sciatic and tibial nerves of the lower limb.

    “The ability to address individual nerve fascicles makes it an attractive method for interfaces with even larger nerves, such as the sciatic and tibial nerves of the lower limb.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing). May combine two claims.

    Interpretationreasoning

    FAST-LIFE arrays represent a platform technology which holds great promise for enabling neuromodulation therapies, with high efficacy and low morbidity, based on the ability to sense and stimulate neural activity throughout the body.

    “Due to their favorable selectivity and specificity, FAST-LIFE arrays represent a platform technology which holds great promise for enabling neuromodulation therapies, with high efficacy and low morbidity, based on the ability to sense and stimulate neural activity throughout the body.”
    Quote in the paperNot a substantial claim
  • Classified as probably not a substantial claim.

    Premisenone given

    Prior reports have not clearly addressed the specificity of stimulation within single fascicles of peripheral nerve.

    “However, these reports have not clearly addressed the specificity of stimulation within single fascicles of peripheral nerve.”
    Quote in the paperSubstantial claim
  • Probably not a substantial claim: may not pass S3 (load-bearing).

    Premisecitation

    Intrafascicular electrode arrays have been implanted in the arms of amputees to evaluate feasibility of sensory and motor neuroprosthetic applications.

    “Additionally, intrafascicular electrode arrays have been implanted in the arms of amputees to evaluate feasibility of sensory and motor neuroprosthetic applications [15,35].”
    Quote in the paperNot a substantial claim0 of 2 sources checked
    • cites [15] Dhillon, 2005, Effects of short-term training on sensory and motor function in severed nerves of long-term human amputees, J. Neurophysiol. source
    • cites [35] Horch, 2011, Object discrimination with an artificial hand using electrical stimulation of peripheral tactile and proprioceptive pathways with intrafascicular electrodes, IEEE Trans. Neural Syst. Rehabil. Eng. source
  • Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).

    Interpretationreasoning

    The selective and specific stimulation of neural tissue enabled by the FAST-LIFE approach is well-suited to a wide variety of neural interfacing applications beyond upper extremity neuroprosthetics.

    “The selective and specific stimulation of neural tissue enabled by the FAST-LIFE approach is also well-suited to a wide variety of neural interfacing applications beyond upper extremity neuroprosthetics.”
    Quote in the paperNot a substantial claim
21 of 21
  • The paper relies on this without support.

    The motor and sensory fascicular groups of the ulnar and median nerves can be identified and dissected based on predictable intraneural topography.

    “the motor and sensory fascicular groups of each nerve were identified and dissected microsurgically based on predictable intraneural topography.”
  • The paper relies on this without support.

    A larger magnitude SSEP response to fascicle stimulation indicates a larger proportion of sensory fibers in that fascicle.

    “The sensory fascicles were positively identified as those that elicited a larger magnitude SSEP response.”
  • The paper relies on this without support.

    Stronger contraction of residual thenar and intrinsic muscles during fascicle stimulation indicates a larger proportion of motor fibers in that fascicle.

    “stimulation of the fascicles with lower SSEP activity produced stronger contractions of the residual thenar and intrinsic muscles of the hand, confirming the hypothesis that these fascicles contained a larger proportion of motor fibers.”
  • The paper relies on this without support.

    The expected cutaneous receptive field of the ulnar nerve is carried within the sensory fascicle.

    “This region closely matches the region expected to represent the cutaneous receptive field of the ulnar nerve, carried within the sensory fascicle.”
  • The paper relies on this without support.

    The sample size of four patients in this study is sufficient to characterize the effects of stimulation of single fascicles within peripheral nerve.

    “The sample size of four patients in this study, though small, is sufficient to characterize the effects of stimulation of single fascicles within peripheral nerve.”
  • The paper relies on this without support.

    Fascicles that produced cutaneous sensations contain a high proportion of afferent sensory fibers.

    “these fascicles are expected to contain a high proportion of afferent sensory fibers.”
  • The paper relies on this without support.

    Fascicles containing efferent motor fibers often contain sensory fibers, particularly muscle spindles and Golgi tendon organs.

    “Fascicles containing efferent motor fibers still produce SSEPs because they often contain sensory fibers, particularly muscle spindles and Golgi tendon organs.”
  • The paper relies on this without support.

    Transverse penetrating electrodes are more likely to move within the fascicle or to displace from the fascicle by nerve motion than longitudinal intrafascicular arrays.

    “Transverse penetrating electrodes are also more likely to move within the fascicle or to displace from the fascicle by nerve motion than longitudinal intrafascicular arrays.”
8 of 8

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Overstreet, C. K., Cheng, J., & Keefer, E. W.
Affiliations
The University of Texas Southwestern Medical Center (US)
Corresponding author
Overstreet, Cynthia K
Venue
Journal of Neural Engineering
Published
2019-11-11
Volume
16
Issue
6
Pages
066040
ISSN
  • 1741-2552 (ISSN-L)
  • 1741-2560
OpenAlex
W2972675070
Cited by
46
References
44
Citations by year
2026: 2 · 2025: 1 · 2024: 9 · 2023: 8 · 2022: 7 · 2021: 10 · 2020: 5 · 2019: 1
FWCI
2.57
Type
journal article
Language
en
Open access
Locations
  • Journal of Neural Engineering (publishedVersion)
  • PubMed (publishedVersion)
Funders
Defense Advanced Research Projects Agency (N66001-15-C-4016)
Topics
  • Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
  • Physical Sciences › Engineering › Biomedical Engineering › Muscle activation and electromyography studies
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Molecular Biology › Planarian Biology and Electrostimulation
Keywords
peripheral nerve stimulation, neural interfaces, intrafascicular electrodes, ulnar nerve, median nerve, stimulation threshold, strength-duration time constant, proprioception
MeSH terms
Adult, Transcutaneous Electric Nerve Stimulation, Electrodes, Implanted, Evoked Potentials, Somatosensory, Female, Humans

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-11.

Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-11.

Cited sources are checked by gpt-6-sol.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review