Evaluation of poly(3,4-ethylenedioxythiophene)/carbon nanotube neural electrode coatings for stimulation in the dorsal root ganglion
Open accessThe DOI matches its Crossref record; 79 of the 79 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: The dorsal root ganglion is an attractive target for implanting neural electrode arrays that restore sensory function or provide therapy via stimulation. However, penetrating microelectrodes designed for these applications are small and deliver low currents. For long-term performance of microstimulation devices, novel coating materials are needed in part to decrease impedance values at the electrode-tissue interface and to increase charge storage capacity. APPROACH: Conductive polymer poly(3,4-ethylenedioxythiophene) (PEDOT) and multi-wall carbon nanotubes (CNTs) were coated on the electrode surface and doped with the anti-inflammatory drug, dexamethasone. Electrode characteristics and the tissue reaction around neural electrodes as a result of stimulation, coating and drug release were characterized. Hematoxylin and eosin staining along with antibodies recognizing Iba1 (microglia/macrophages), NF200 (neuronal axons), NeuN (neurons), vimentin (fibroblasts), caspase-3 (cell death) and L1 (neural cell adhesion molecule) were used. Quantitative image analyses were performed using MATLAB. MAIN RESULTS: Our results indicate that coated microelectrodes have lower in vitro and in vivo impedance values. Significantly less neuronal death/damage was observed with coated electrodes as compared to non-coated controls. The inflammatory response with the PEDOT/CNT-coated electrodes was also reduced. SIGNIFICANCE: This study is the first to report on the utility of these coatings in stimulation applications. Our results indicate PEDOT/CNT coatings may be valuable additions to implantable electrodes used as therapeutic modalities.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“PEDOT/CNT films were assessed for their utility in modifying a PNS interface, and it was found that coated electrodes exhibited improved neuronal responses at the interface and decreased expression of neuronal apoptosis markers following implantation into the DRG of rats.43”
From Regenerative Electrode Interfaces for Neural Prostheses. (2016) · cites this paper as [43]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Poly(3,4-ethylenedioxythiophene) and Its Derivatives: Past, Present, and Future
Groenendaal, L., Jonas, F., Freitag, D., Pielartzik, H., & Reynolds, J. R. (2000) · Advanced Materials · doi:10.1002/(sici)1521-4095(200004)12:7<481::aid-adma481>3.0.co;2-c · Cited by 3,420 · cited 1×
“Furthermore, PEDOT has improved electrochemical, mechanical and thermal stability necessary for use in chronic implants [31].”
BackgroundNeural Stimulation and Recording Electrodes
Cogan, S. F. (2008) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.10.061807.160518 · Cited by 2,243 · cited 2×
“Among the various conducting polymers, poly(3,4-ethylenedioxythiophene) (PEDOT) has been shown to be an excellent material for neural stimulation as a result of its superior impedance and charge injection capacity when compared to thin film platinum (Pt) electrodes [7, 28, 29].”
BackgroundElectrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078 · cited 1×
“stabilize the electrode potential from any ‘ratcheting’ effects [46].”
BackgroundApplications of Carbon Nanotubes in Drug Delivery
Kaur, J., Gill, G. S., & Jeet, K. (2019) · Characterization and Biology of Nanomaterials for Drug Delivery · doi:10.1016/b978-0-12-814031-4.00005-2 · Cited by 2,019
(2006) · Advanced Materials · doi:10.1002/(issn)1521-4095 · Cited by 1,662
Stimuli-Responsive Polymer Nanocomposites Inspired by the Sea Cucumber Dermis
Capadona, J. R., Shanmuganathan, K., Tyler, D. J., Rowan, S. J., & Weder, C. (2008) · Science · doi:10.1126/science.1153307 · Cited by 962 · cited 1×
“To address the mechanical mismatch between current electrode materials and brain tissue, a number of groups are also investigating soft materials with mechanical properties similar to those of central nervous system (CNS) or peripheral nervous system (PNS) tissue [18-23].”
BackgroundBiran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893 · cited 1×
“Biocompatibility issues resulting in immune and inflammatory reactions prevent functional integration with the surrounding neural tissue and cause chronic neuronal degeneration [8, 9].”
BackgroundConducting‐Polymer Nanotubes for Controlled Drug Release
Abidian, M. R., Kim, D. ‐., & Martin, D. C. (2006) · Advanced Materials · doi:10.1002/adma.200501726 · Cited by 869 · cited 2×
“Additionally, anti-inflammatory agents like dexamethasone have been successfully incorporated and electrically released from conducting polymers to inhibit the inflammatory tissue response [27, 32].”
BackgroundFunctionalization density dependence of single-walled carbon nanotubes cytotoxicity in vitro
Sayes, C. M., Liang, F., Hudson, J. L., Mendez, J., Guo, W., Beach, J. M., Moore, V. C., Doyle, C. D., West, J. L., Billups, W. E., et al. (2006) · Toxicology Letters · doi:10.1016/j.toxlet.2005.08.011 · Cited by 856 · cited 1×
“In addition, CNTs have low toxicity, no immunogenicity [36-38] and can act as a nanoreservoir for targeted drug delivery using controlled drug release [39, 40].”
BackgroundBhirde, A. A., Patel, V., Gavard, J., Zhang, G., Sousa, A. A., Masedunskas, A., Leapman, R. D., Weigert, R., Gutkind, J. S., & Rusling, J. F. (2009) · ACS Nano · doi:10.1021/nn800551s · Cited by 853 · cited 1×
“In addition, CNTs have low toxicity, no immunogenicity [36-38] and can act as a nanoreservoir for targeted drug delivery using controlled drug release [39, 40].”
BackgroundKilbride, B. E., Coleman, J. N., Fraysse, J., Fournet, P., Cadek, M., Drury, A., Hutzler, S., Roth, S., & Blau, W. J. (2002) · Journal of Applied Physics · doi:10.1063/1.1506397 · Cited by 789 · cited 1×
“, no delamination or cracking) even after prolonged and aggressive stimulation [34–36].”
BackgroundSecondary injury mechanisms in traumatic spinal cord injury: a nugget of this multiply cascade
Oyinbo, C. (2011) · Acta Neurobiologiae Experimentalis · doi:10.55782/ane-2011-1848 · Cited by 786 · cited 1×
“, cytokines, inflammation, excitotoxicity and free radical damage) [58, 60, 61].”
BackgroundAcute Spinal Cord Injury, Part I: Pathophysiologic Mechanisms
Dumont, R. J., Okonkwo, D. O., Verma, S., Hurlbert, R. J., Boulos, P. T., Ellegala, D. B., & Dumont, A. S. (2001) · Clinical Neuropharmacology · doi:10.1097/00002826-200109000-00002 · Cited by 732 · cited 2×
“Apoptosis, one type of programmed cell death, occurs following nervous system injury [58, 59] and can be triggered by a variety of insults that can lead to apoptotic cell death including those that impact cells away from the interface (i.”
BackgroundKozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721 · cited 1×
“Furthermore, ultrasmall electrodes may be needed to extend early benefits to long-term performance [23, 24].”
BackgroundSelf-citationDumortier, H., Lacotte, S., Pastorin, G., Marega, R., Wu, W., Bonifazi, D., Briand, J.-P., Prato, M., Muller, S., & Bianco, A. (2006) · Nano Letters · doi:10.1021/nl061160x · Cited by 697 · cited 1×
“In addition, CNTs have low toxicity, no immunogenicity [36-38] and can act as a nanoreservoir for targeted drug delivery using controlled drug release [39, 40].”
BackgroundAtom movements: Diffusion and mass transport in solids
Philibert, J. (1991) · International Linear Collider · Cited by 694
Philibert, J. (2020) · EDP Sciences eBooks · doi:10.1051/978-2-7598-0172-5 · Cited by 687
Schmidt, E. M., Bak, M. J., Hambrecht, F. T., Kufta, C. V., O'Rourke, D. K., & Vallabhanath, P. (1996) · Brain · doi:10.1093/brain/119.2.507 · Cited by 672
Bennett, D. L. H., Michael, G. J., Ramachandran, N., Munson, J. B., Averill, S., Yan, Q., McMahon, S. B., & Priestley, J. V. (1998) · The Journal of Neuroscience · doi:10.1523/jneurosci.18-08-03059.1998 · Cited by 627 · cited 1×
“In peripheral tissues like the DRG, it is believed that immune cells invade following injury [53, 54] and some microglia/macrophages express Iba1.”
BackgroundWadhwa, R., Lagenaur, C. F., & Cui, X. T. (2006) · Journal of Controlled Release · doi:10.1016/j.jconrel.2005.10.027 · Cited by 496 · cited 2×
“The intermediate impedance of the PCD film can be attributed to the additional doping of the free dexamethasone which has been shown to yield films with less substantial impedance decreases [33, 42].”
BackgroundMcConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480 · cited 1×
“Biocompatibility issues resulting in immune and inflammatory reactions prevent functional integration with the surrounding neural tissue and cause chronic neuronal degeneration [8, 9].”
BackgroundChronic Neural Recording Using Silicon-Substrate Microelectrode Arrays Implanted in Cerebral Cortex
Vetter, R. J., Williams, J. C., Hetke, J. F., Nunamaker, E. A., & Kipke, D. R. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.826680 · Cited by 475 · cited 1×
“For example, high voltage pulses to disrupt scar tissue [11, 12] and localized delivery of anti-inflammatory drugs and neurotrophic factors [13-17] have been investigated.”
BackgroundOchoa, J., & Torebjörk, E. (1983) · The Journal of Physiology · doi:10.1113/jphysiol.1983.sp014873 · Cited by 467 · cited 1×
“In the nervous system, spatial specificity is particularly important as discrete and graded sensations can be evoked through stimulation [2-6].”
BackgroundZhong, Y., & Bellamkonda, R. V. (2007) · Brain Research · doi:10.1016/j.brainres.2007.02.024 · Cited by 321 · cited 6×
“This is similar to the results obtained with silicon neural probes with and without dexamethasone at four weeks post-implantation [53] and may be due to drug depletion or stabilization of the inflammatory response as previously proposed.”
BackgroundResultInfluentialJain, A., Kim, Y.-T., McKeon, R. J., & Bellamkonda, R. V. (2006) · Biomaterials · doi:10.1016/j.biomaterials.2005.07.008 · Cited by 308 · cited 1×
“For example, high voltage pulses to disrupt scar tissue [11, 12] and localized delivery of anti-inflammatory drugs and neurotrophic factors [13-17] have been investigated.”
BackgroundVisual sensations produced by intracortical microstimulation of the human occipital cortex
Bak, M., Girvin, J. P., Hambrecht, F. T., Kufta, C. V., Loeb, G. E., & Schmidt, E. M. (1990) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442682 · Cited by 299 · cited 1×
“with a rationale for these parameters established in work by Schmidt and colleagues [44, 45].”
Prasad, A., Xue, Q.-S., Sankar, V., Nishida, T., Shaw, G., Streit, W. J., & Sanchez, J. C. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/5/056015 · Cited by 298 · cited 2×
“electrode) impact electrical properties as well [10].”
BackgroundHu, P., Bembrick, A. L., Keay, K. A., & McLachlan, E. M. (2007) · Brain, Behavior, and Immunity · doi:10.1016/j.bbi.2006.10.013 · Cited by 279 · cited 1×
“In peripheral tissues like the DRG, it is believed that immune cells invade following injury [55, 56] and some microglia/macrophages express Iba1.”
BackgroundPoly (3,4-Ethylenedioxythiophene) for Chronic Neural Stimulation
Cui, X. T., & Zhou, D. D. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.909811 · Cited by 277 · cited 1×
“Among the various conducting polymers, poly(3,4-ethylenedioxythiophene) (PEDOT) has been shown to be an excellent material for neural stimulation as a result of its superior impedance and charge injection capacity when compared to thin film platinum (Pt) electrodes [7, 28, 29].”
BackgroundMacefield, G., Gandevia, S. C., & Burke, D. (1990) · The Journal of Physiology · doi:10.1113/jphysiol.1990.sp018247 · Cited by 274 · cited 1×
“In the nervous system, spatial specificity is particularly important as discrete and graded sensations can be evoked through stimulation [2-6].”
BackgroundHighly stable carbon nanotube doped poly(3,4-ethylenedioxythiophene) for chronic neural stimulation
Luo, X., Weaver, C. L., Zhou, D. D., Greenberg, R., & Cui, X. T. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2011.04.051 · Cited by 252 · cited 3×
“, no delamination or cracking) even after prolonged and aggressive stimulation [33-35].”
BackgroundMethodSelf-citationCarbon nanotube nanoreservior for controlled release of anti-inflammatory dexamethasone
Luo, X., Matranga, C., Tan, S., Alba, N., & Cui, X. T. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2011.05.020 · Cited by 249 · cited 4×
“For example, multiwall CNTs loaded with dexamethasone were able to reduce the extent of microglial activation in response to lipopolysaccharide treatment to a similar extent as an exogenously administered drug [41].”
BackgroundMethodSelf-citationThompson, B. C., Richardson, R. T., Moulton, S. E., Evans, A. J., O'Leary, S., Clark, G. M., & Wallace, G. G. (2010) · Journal of Controlled Release · doi:10.1016/j.jconrel.2009.09.016 · Cited by 229 · cited 1×
“Alternative drugs or combinations thereof may prove more effective in controlling the immune response and glial activation; certainly, combinatorial approaches are effective for promoting neurite outgrowth [56].”
BackgroundShain, W., Spataro, L., Dilgen, J., Haverstick, K., Retterer, S., Isaacson, M., Saltzman, M., & Turner, J. N. (2003) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2003.814800 · Cited by 227 · cited 3×
“7 μM [52, 53] indicating that the hour long stimulation of the PCD coated-electrodes yields a biologically significant amount of dexamethasone release over the course of this study.”
BackgroundResultA carbon-fiber electrode array for long-term neural recording
Guitchounts, G., Markowitz, J. E., Liberti, W. A., & Gardner, T. J. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/4/046016 · Cited by 224 · cited 1×
“with ultrasmall electrodes [24, 25], there has yet to be a”
BackgroundSpataro, L., Dilgen, J., Retterer, S., Spence, A. J., Isaacson, M., Turner, J. N., & Shain, W. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2004.08.037 · Cited by 215 · cited 3×
“This is consistent with studies demonstrating dexamethasone's ability to inhibit the immune response [16, 51, 52].”
BackgroundResultKozai, T. D. Y., Vazquez, A. L., Weaver, C. L., Kim, S.-G., & Cui, X. T. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/6/066001 · Cited by 215 · cited 1×
“While this in vitro and early chronic in vivo work demonstrates a proof-of-concept platform for drug-releasing stimulation electrodes, recent work suggests that the first 12 h to 21 d post-implant may be the most dynamic and most useful for interventions aimed at promoting long-term function [10, 54]; sustained dexame…”
BackgroundSelf-citationSubstrate dependent stability of conducting polymer coatings on medical electrodes
Green, R. A., Hassarati, R. T., Bouchinet, L., Lee, C. S., Cheong, G. L. M., Yu, J. F., Dodds, C. W., Suaning, G. J., Poole-Warren, L. A., & Lovell, N. H. (2012) · Biomaterials · doi:10.1016/j.biomaterials.2012.05.017 · Cited by 205
Electrochemical polymerization of conducting polymers in living neural tissue
Richardson-Burns, S. M., Hendricks, J. L., & Martin, D. C. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/2/l02 · Cited by 195 · cited 1×
“, brain, heart, skeletal muscle) has been investigated for a variety of applications including vagal nerve stimulation, retinal and cochlear implants, spinal cord stimulation and deep brain stimulation [1].”
BackgroundKozai, T. D. Y., & Kipke, D. R. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.08.002 · Cited by 190
Self-citationMechanically adaptive intracortical implants improve the proximity of neuronal cell bodies
Harris, J. P., Capadona, J. R., Miller, R. H., Healy, B. C., Shanmuganathan, K., Rowan, S. J., Weder, C., & Tyler, D. J. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/6/066011 · Cited by 190 · cited 1×
“To address the mechanical mismatch between current electrode materials and brain tissue, a number of groups are also investigating soft materials with mechanical properties similar to those of central nervous system (CNS) or peripheral nervous system (PNS) tissue [18-23].”
BackgroundWilks, S. (2009) · Frontiers in Neuroengineering · doi:10.3389/neuro.16.007.2009 · Cited by 188 · cited 1×
“They exhibit both fast and high charge delivery capacities as a result of the high ionic conductivity and large electroactive surface area [26] resulting in low impedance and more effective charge transfer [27].”
BackgroundKozai, T. D. Y., Marzullo, T. C., Hooi, F., Langhals, N. B., Majewska, A. K., Brown, E. B., & Kipke, D. R. (2010) · Journal of Neural Engineering · doi:10.1088/1741-2560/7/4/046011 · Cited by 184 · cited 1×
“Another possible explanation for this variability in tissue reaction is the extent of injury to key vascular structures caused by the implant [62, 63].”
BackgroundSelf-citationLinear-dendritic drug conjugates forming long-circulating nanorods for cancer-drug delivery
Zhou, Z., Ma, X., Jin, E., Tang, J., Sui, M., Shen, Y., Van Kirk, E. A., Murdoch, W. J., & Radosz, M. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.04.012 · Cited by 178
L1.1 Is Involved in Spinal Cord Regeneration in Adult Zebrafish
Becker, C. G., Lieberoth, B. C., Morellini, F., Feldner, J., Becker, T., & Schachner, M. (2004) · The Journal of Neuroscience · doi:10.1523/jneurosci.2420-04.2004 · Cited by 178 · cited 1×
“L1 has been shown to be up-regulated following nervous system injury both in vitro (reviewed in [64]) and in vivo [65, 66].”
BackgroundDrugs released from polymers: diffusion and elimination in brain tissue
Saltzman, W. M., & Radomsky, M. L. (1991) · Chemical Engineering Science · doi:10.1016/0009-2509(91)80036-x · Cited by 177 · cited 1×
“Using an instantaneous source diffusion model [49] and a dexamethasone diffusion coefficient in brain tissue [50], the average concentration of dexamethasone was determined within 500 μm of the implant one day post-stimulation.”
MethodFabrication of Responsive, Softening Neural Interfaces
Ware, T., Simon, D., Arreaga‐Salas, D. E., Reeder, J., Rennaker, R., Keefer, E. W., & Voit, W. (2012) · Advanced Functional Materials · doi:10.1002/adfm.201200200 · Cited by 165 · cited 1×
“To address the mechanical mismatch between current electrode materials and brain tissue, a number of groups are also investigating soft materials with mechanical properties similar to those of central nervous system (CNS) or peripheral nervous system (PNS) tissue [18-23].”
BackgroundPolyaniline nanotube composites: A high-resolution printable conductor
Blanchet, G. B., Fincher, C. R., & Gao, F. (2003) · Applied Physics Letters · doi:10.1063/1.1553991 · Cited by 161 · cited 1×
“, no delamination or cracking) even after prolonged and aggressive stimulation [33-35].”
BackgroundA multi-step method for preparation of porcine small intestinal submucosa (SIS)
Luo, J., Chen, W., Chen, X., Qin, T., Huang, Y., Xie, H., Li, X., Qian, Z., & Yang, Z. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.09.017 · Cited by 161
In vivo deployment of mechanically adaptive nanocomposites for intracortical microelectrodes
Harris, J. P., Hess, A. E., Rowan, S. J., Weder, C., Zorman, C. A., Tyler, D. J., & Capadona, J. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/4/046010 · Cited by 151 · cited 1×
“To address the mechanical mismatch between current electrode materials and brain tissue, a number of groups are also investigating soft materials with mechanical properties similar to those of central nervous system (CNS) or peripheral nervous system (PNS) tissue [18-23].”
BackgroundAnti-apoptotic and neuroprotective effects of edaravone following transient focal ischemia in rats
Amemiya, S., Kamiya, T., Nito, C., Inaba, T., Kato, K., Ueda, M., Shimazaki, K., & Katayama, Y. (2005) · European Journal of Pharmacology · doi:10.1016/j.ejphar.2005.04.036 · Cited by 151 · cited 1×
“, cytokines, inflammation, excitotoxicity and free radical damage) [60, 62, 63].”
BackgroundControlled release of anti-inflammatory agent α-MSH from neural implants
Zhong, Y., & Bellamkonda, R. V. (2005) · Journal of Controlled Release · doi:10.1016/j.jconrel.2005.05.012 · Cited by 139 · cited 1×
“[13–17] have been investigated.”
McCreery, D., Pikov, V., & Troyk, P. R. (2010) · Journal of Neural Engineering · doi:10.1088/1741-2560/7/3/036005 · Cited by 137 · cited 3×
“The stimulation protocol employed was modeled after those utilized for intracortical microstimulation of visual prosthetic devices [45] and for sensorimotor cortex stimulation [43] and differs from that used in prior studies [42].”
MethodResultInfluentialVallbo, Å. B. (1981) · Brain Research · doi:10.1016/0006-8993(81)90517-5 · Cited by 130 · cited 1×
“In the nervous system, spatial specificity is particularly important as discrete and graded sensations can be evoked through stimulation [2-6].”
BackgroundRepeated voltage biasing improves unit recordings by reducing resistive tissue impedances
Johnson, M. D., Otto, K. J., & Kipke, D. R. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847373 · Cited by 130 · cited 1×
“For example, high voltage pulses to disrupt scar tissue [11, 12] and localized delivery of anti-inflammatory drugs and neurotrophic factors [13-17] have been investigated.”
BackgroundIon conducting polymer microelectrodes for interfacing with neural networks.
Nyberg, T., Shimada, A., & Torimitsu, K. (2007) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2006.08.008 · Cited by 117 · cited 1×
“Among the various conducting polymers, poly(3,4-ethylenedioxythiophene) (PEDOT) has been shown to be an excellent material for neural stimulation as a result of its superior impedance and charge injection capacity when compared to thin film platinum (Pt) electrodes [7, 29, 30].”
BackgroundZhou, H., Cheng, X., Rao, L., Li, T., & Duan, Y. Y. (2013) · Acta Biomaterialia · doi:10.1016/j.actbio.2013.01.042 · Cited by 105 · cited 1×
“In the resulting PEDOT/CNT (PC) coating, CNTs were well-dispersed and formed a network-like structure (figures 2(B), (D)) unlike that observed by others [49].”
ResultChen, J., Bernreuther, C., Dihné, M., & Schachner, M. (2005) · Journal of Neurotrauma · doi:10.1089/neu.2005.22.896 · Cited by 89 · cited 1×
“For example, in mouse models of Parkinson’s disease and in spinal cord injury models L1 has been shown to improve cell survival [69, 70].”
BackgroundLimb-state feedback from ensembles of simultaneously recorded dorsal root ganglion neurons
Weber, D. J., Stein, R. B., Everaert, D. G., & Prochazka, A. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/3/s04 · Cited by 89 · cited 1×
“Although the DRG is an attractive site for recording or stimulating primary afferent neurons to provide sensory feedback [2, 47], studies investigating the tissue reaction in response to stimulation are lacking.”
BackgroundSelf-citationJaquins‐Gerstl, A., Shu, Z., Zhang, J., Liu, Y., Weber, S. G., & Michael, A. C. (2011) · Analytical Chemistry · doi:10.1021/ac200782h · Cited by 78 · cited 1×
“Although scarring would not be expected in this relatively short-term study, dexamethasone has also been shown to have inhibitory effects on this related process [16, 52, 53, 57].”
BackgroundGreen, R. A., Williams, C. M., Lovell, N. H., & Poole-Warren, L. A. (2008) · Journal of Materials Science Materials in Medicine · doi:10.1007/s10856-008-3376-7 · Cited by 69 · cited 1×
“In addition, CNTs have low toxicity, no immunogenicity [36-38] and can act as a nanoreservoir for targeted drug delivery using controlled drug release [39, 40].”
BackgroundAn ultra-compliant, scalable neural probe with molded biodissolvable delivery vehicle
Gilgunn, P. J., Khilwani, R., Kozai, T. D. Y., Weber, D. J., Cui, X. T., Erdos, G., Ozdoganlar, O. B., & Fedder, G. K. (2012) · 2012 IEEE 25th International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/memsys.2012.6170092 · Cited by 65 · cited 2×
“Furthermore, ultrasmall electrodes may be needed to extend early benefits to long-term performance [23, 24].”
BackgroundSelf-citationThe role of methane in mammalian physiology—is it a gasotransmitter?
Boros, M., Tuboly, E., Mészáros, A., & Amann, A. (2015) · Journal of Breath Research · doi:10.1088/1752-7155/9/1/014001 · Cited by 62
Ohara, S., Weiss, N., & Lenz, F. A. (2004) · Journal of Neurophysiology · doi:10.1152/jn.00648.2003 · Cited by 61 · cited 1×
“In the nervous system, spatial specificity is particularly important as discrete and graded sensations can be evoked through stimulation [2-6].”
BackgroundCui, Y.-F., Hargus, G., Xu, J.-C., Schmid, J. S., Shen, Y.-Q., Glatzel, M., Schachner, M., & Bernreuther, C. (2009) · Brain · doi:10.1093/brain/awp290 · Cited by 61 · cited 1×
“For example, in mouse models of Parkinson's disease and in spinal cord injury models L1 has been shown to improve cell survival [67, 68].”
BackgroundKolarcik, C. L., Bourbeau, D., Azemi, E., Rost, E., Zhang, L., Lagenaur, C. F., Weber, D. J., & Cui, X. T. (2012) · Acta Biomaterialia · doi:10.1016/j.actbio.2012.06.034 · Cited by 59 · cited 1×
“The kill zone was identified by the area absent of NF200 staining, and quantification of this area was performed as described previously [46].”
MethodSelf-citationBecker, T., Lieberoth, B. C., Becker, C. G., & Schachner, M. (2005) · Molecular and Cellular Neuroscience · doi:10.1016/j.mcn.2005.07.008 · Cited by 58 · cited 1×
“L1 has been shown to be up-regulated following nervous system injury both in vitro (reviewed in [64]) and in vivo [65, 66].”
BackgroundGaunt, R. A., Hokanson, J. A., & Weber, D. J. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/055009 · Cited by 53 · cited 3×
“The stimulation paradigm was designed to activate sensory neurons in the DRG based on previous acute studies [2].”
BackgroundMethodSelf-citationEvolution of Brain Impedance in Dystonic Patients Treated by GPi Electrical Stimulation
Hemm, S., Vayssière, N., Mennessier, G., Cif, L., Zanca, M., Ravel, P., Frèrebeau, P., & Coubes, P. A. (2004) · Neuromodulation Technology at the Neural Interface · doi:10.1111/j.1094-7159.2004.04009.x · Cited by 52 · cited 1×
“observed transiently by others [59].”
Nanostructured Coatings for Improved Charge Delivery to Neurons
Kozai, T. D. Y., Alba, N. A., Zhang, H., Kotov, N. A., Gaunt, R. A., & Cui, X. T. (2014) · Springer eBooks · doi:10.1007/978-1-4899-8038-0_4 · Cited by 48
Self-citationCharacterization of Implantable Microfabricated Fluid Delivery Devices
Rathnasingham, R., Kipke, D. R., Bledsoe, S. C., & McLaren, J. D. (2004) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.820311 · Cited by 47 · cited 1×
“For example, high voltage pulses to disrupt scar tissue [11, 12] and localized delivery of anti-inflammatory drugs and neurotrophic factors [13-17] have been investigated.”
BackgroundPaterniti, I., Genovese, T., Crisafulli, C., Mazzon, E., Di Paola, R., Galuppo, M., Bramanti, P., & Cuzzocrea, S. (2009) · Naunyn-Schmiedeberg's Archives of Pharmacology · doi:10.1007/s00210-009-0414-z · Cited by 39 · cited 1×
“Apoptosis, one type of programmed cell death, occurs following nervous system injury [60, 61] and can be triggered by a variety of insults that can lead to apoptotic cell death including those that impact cells away from the interface (i.”
BackgroundThe Injured and Regenerating Nervous System: Immunoglobulin Superfamily Members as Key Players
Irintchev, A., & Schachner, M. (2012) · The Neuroscientist · doi:10.1177/1073858411419047 · Cited by 37 · cited 1×
“L1 has been shown to be up-regulated following nervous system injury both in vitro (reviewed in [64]) and in vivo [65, 66].”
BackgroundThe roles of cell adhesion molecules on the formation of peripheral myelin.
Takeda, Y., Murakami, Y., Asou, H., & Uyemura, K. (2001) · The Keio Journal of Medicine · doi:10.2302/kjm.50.240 · Cited by 32 · cited 1×
“peripheral myelin formation [71].”
Multi-site incorporation of bioactive matrices into MEMS-based neural probes
Williams, J. C., Holecko, M. M., Massia, S. P., Rousche, P. J., & Kipke, D. R. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/4/l03 · Cited by 30 · cited 1×
“For example, high voltage pulses to disrupt scar tissue [11, 12] and localized delivery of anti-inflammatory drugs and neurotrophic factors [13-17] have been investigated.”
BackgroundBlood supply of spinal nerve roots
Petterson, C., & Olsson, Y. (1989) · Acta Neuropathologica · doi:10.1007/bf00687706 · Cited by 29 · cited 1×
“of injury to key vascular structures caused by the implant [64, 65].”
BackgroundInside Front Cover: Conducting‐Polymer Nanotubes for Controlled Drug Release (Adv. Mater. 4/2006)
Abidian, M. R., Kim, D., & Martin, D. C. (2006) · Advanced Materials · doi:10.1002/adma.200690018 · Cited by 15
Luo, X., Matranga, C., Tan, S., et al. (2011) · Cited by 15
Self-citationApplications of carbon nanotubes in drug delivery
Bianco, A., Kostarelos, K., & Prato, M. (2005) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2005.10.005
Research works printed but not linked: 2
- 10.1002/(issn)1521-4095 · DOI · Paper page
- 10.1111/ner.2004.7.issue-2 · doi:10.1111/ner.2004.7.issue-2
Cited by
Brain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies
Kozai, T. D. Y., Jaquins-Gerstl, A. S., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn500256e · Cited by 622
Self-citationOrganic Bioelectronics: Bridging the Signaling Gap between Biology and Technology
Simon, D. T., Gabrielsson, E. O., Tybrandt, K., & Berggren, M. (2016) · Chemical Reviews · doi:10.1021/acs.chemrev.6b00146 · Cited by 568
Current and Novel Polymeric Biomaterials for Neural Tissue Engineering
Boni, R., Ali, A., Shavandi, A., & Clarkson, A. N. (2018) · Journal of Biomedical Science · doi:10.1186/s12929-018-0491-8 · Cited by 439
A Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389
Self-citationConductive biomaterials for muscle tissue engineering
Dong, R., Ma, P. X., & Guo, B. (2020) · Biomaterials · doi:10.1016/j.biomaterials.2019.119584 · Cited by 381
Metadata
- Authors
- Kolarcik, C. L., Catt, K., Rost, E., Albrecht, I. N., Bourbeau, D., Du, Z., Kozai, T. D. Y., Luo, X., Weber, D. J., & Tracy Cui, X.
- Affiliations
- University of Pittsburgh (US)
- Qingdao University of Science and Technology (CN)
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2014-12-08
- Volume
- 12
- Issue
- 1
- Pages
- 016008
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- PMC
- PMC4433052
- OpenAlex
- W2045925821
- Semantic Scholar
- 57158d0a2f630484004ccda825e83c570fb81a52
- Cited by
- 126
- References
- 78
- Citations by year
- 2026: 1 · 2025: 8 · 2024: 8 · 2023: 14 · 2022: 11 · 2021: 13 · 2020: 9 · 2019: 16 · 2018: 12 · 2017: 14 · 2016: 6 · 2015: 10 · 2014: 2
- FWCI
- 4.88
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- iopscience.iop.org/info/page/text-and-data-mining (tdm)
- iopscience.iop.org/page/copyright (vor)
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- U.S. Department of Agriculture
- National Institutes of Health (TL1TR000145, 4r01ns072342-05, UL1TR000005)
- National Institute of Neurological Disorders and Stroke (R01-NS072342, R01NS062019)
- National Center for Advancing Translational Sciences
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Materials Science › Polymers and Plastics › Conducting polymers and applications
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Keywords
- PEDOT, carbon nanotubes, dorsal root ganglia, microelectrode stimulation, electrode impedance, charge storage capacity, neuroinflammation, microglia/macrophages, neuronal cell death
- MeSH terms
- Animals, Anti-Inflammatory Agents, Dexamethasone, Drug Implants, Electric Conductivity, Equipment Design, Female, Ganglia, Spinal, Materials Testing, Microelectrodes, Polymers, Sensitivity and Specificity, Reproducibility of Results, Rats, Sprague-Dawley, Bridged Bicyclo Compounds, Heterocyclic, Equipment Failure Analysis, Coated Materials, Biocompatible, Nanotubes, Carbon, Rats, Nanocapsules, Spinal Cord Stimulation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.