Regenerative Electrode Interfaces for Neural Prostheses.
Open accessAbstract
This review categorizes tissue-engineering strategies for regenerative electrode interfaces and discusses selective interfacing with specific neuronal subpopulations.
AI snapshot
Written from the paper’s full text.
TL;DR
This review surveys tissue-engineering strategies intended to improve electrode–neuron integration in peripheral and central neural prostheses, grouping them into architectural redesign, bioactive/material coatings, and biohybrid cell or tissue grafts. It argues that regenerative interfaces could improve device stability, longevity, and subtype selectivity, while noting that translation and long-term durability remain unresolved.
Key findings
- The review organizes regenerative interfaces into three broad approaches: (1) electrode architecture with finer geometries or topographical guidance, (2) material coatings and incorporated bioactive molecules, and (3) cellular grafts or biohybrid devices; these are discussed separately for the peripheral and central nervous systems.
- In the CNS, implantation is linked to neuronal loss near the electrode surface, with about a 40% decrease in neuronal density reported within the estimated recordable radius; in the PNS, axons have been observed to retract more than 1 cm from the wound after penetrating electrode implantation.
- Examples include flexible polyimide sieve electrodes that supported regenerating fibers in all 30 implanted rats; microchannel electrode arrays with roughly 100 × 100 µm channels that supported miniature fascicles up to 5 mm; PDMS/SU-8 microchannel scaffolds that recorded in vivo for 5 months; and regenerative peripheral nerve interfaces that recorded compound muscle action potentials over 7 months and 14 months, with unpublished rodent viability beyond 20 months and 50–100× signal amplification compared with state-of-the-art peripheral nerve interfaces.
- L1-coated probes were associated with no evident reduction in neuronal density around implants over 8 weeks and improved density/neurofilament staining within the first 100 µm; a laminin nanoscale coating reduced gliosis at 4 weeks but did not affect local neuronal density; parylene-C coated versus uncoated arrays showed no significant differences over 12 weeks.
Methods
- Narrative review, not a primary experimental study. It synthesizes published work on regenerative electrode interfaces for the PNS and CNS, with literature summarized in Supplementary Figs. S1 and S2.
- The main organizing analysis is thematic: approaches are categorized by strategy (architecture, coatings/biologics, biohybrid devices) and by nervous system region, with selected in vitro and in vivo examples.
- No systematic search strategy, inclusion criteria, or quantitative meta-analysis is described.
Limitations
- Stated: the review notes ongoing challenges including translating in vitro coating benefits to in vivo outcomes, long-term electrical and mechanical stability of bioactive conductive polymers, limited survival or detection of grafted cells beyond early time points, and unclear mechanisms behind improved neuronal responses.
- Stated: several cited findings are preliminary or proof-of-concept, and some data are unpublished; longer-term studies are called for.
- Evident: as a narrative review, it lacks a formal systematic search and risk-of-bias assessment, and its conclusions rely on selected examples rather than comprehensive quantitative synthesis.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
48 of 112 cited sources checked · 27 supported · 0 flagged · accuracy not yet measured
- Interpretationreasoning
Signal instability and signal loss may undermine the detection and decoding efficacy of brain-machine interfaces and closed-loop neurostimulation applications in which the stimulus is conditioned on a recorded neural signal.
“This may undermine the detection and decoding efficacy of brain-machine interfaces and closed-loop neurostimulation applications in which the stimulus is conditioned on a recorded neural signal.1,15”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [1] 1. Rosin, B., Slovik, M., Mitelman, R., Rivlin-Etzion, M., Haber, S.N., Vaadia, E., et al. Closed-loop deep brain stimulation is superior in ameliorating parkinsonism. Neuron 72, 370, 2011. source
- cites [15] 15. Perge, J.A., Homer, M.L., Malik, W.Q., Cash, S., Eskandar, E., Friehs, G., et al. Intra-day signal instabilities affect decoding performance in an intracortical neural interface system. J Neural Eng 10, 036004, 2013. source
The passage immediately before Methods reports that within-day signal changes caused decoding errors and degraded neural cursor control in people with tetraplegia. The paper supports the concern for brain–machine interface decoding, but the cited passages do not address detection efficacy or closed-loop neurostimulation conditioned on a recorded signal. Page numbers are not provided in our copy. (LLM, conf=0.87)
Experimental — accuracy not yet measured
- Premisecitation
Neural injury and inflammation that accompany device implantation are key contributing factors to signal loss and instability.
“but the neural injury and inflammation that accompany device implantation are key contributing factors.”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [16] 16. Prasad, A., Xue, Q.-S., Sankar, V., Nishida, T., Shaw, G., Streit, W.J., et al. Comprehensive characterization and failure modes of tungsten microwire arrays in chronic neural implants. J Neural Eng 9, 056015, 2012. source
- cites [17] 17. Prasad, A., Xue, Q.-S., Dieme, R., Sankar, V., Mayrand, R.C., Nishida, T., et al. Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants. Front Neuroeng 7, 2, 2014. source
The Discussion identifies vascular disruption and intraparenchymal hemorrhage during implantation as a major biotic factor likely contributing to electrode failure. But the paper reports that electrode performance was not correlated with microglial activation (neuroinflammation), so it does not support inflammation as a key contributor to signal loss. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
After implantation in severed rat sciatic nerves, Bellamkonda microchannel scaffolds guided the growth of axons and supporting cells along the length of the embedded microchannels.
“scaffolds guided the growth of axons and supporting cells along the length of the embedded microchannels”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [33] 33. Srinivasan, A., Tahilramani, M., Bentley, J.T., Gore, R.K., Millard, D.C., Mukhatyar, V.J., et al. Microchannel-based regenerative scaffold for chronic peripheral nerve inter- facing in amputees. Biomaterials 41, 151, 2015. source
The implantation methods describe scaffolds placed in transected rat sciatic nerves, and Results §3.3 reports axon regeneration inside most of the larger microchannels. Results §3.2 shows Schwann cells migrating along microchannels, but that experiment used cultured tissue rather than implanted nerves; Results §3.5 locates axons and Schwann cells together distal to the implant. The copy therefore does not establish that supporting cells grew along the length of the embedded channels after implantation. (LLM, conf=0.86)
Experimental — accuracy not yet measured
- Premisecitation
Koppes et al. observed positive effects of electrically guided proximal Schwann cells on neurite outgrowth.
“Koppes et al. observed positive effects of electrically guided proximal Schwann cells on neurite outgrowth.57”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [57] 57. Koppes, A.N., Nordberg, A.L., Paolillo, G.M., Goodsell, N.M., Darwish, H.A., Zhang, L., et al. Electrical stimula- tion of schwann cells promotes sustained increases in neurite outgrowth. Tissue Eng Part A 20, 494, 2014. source
The Results report that electrically prestimulated Schwann cells supported significantly greater neurite outgrowth than unstimulated Schwann cells. The study describes cultured Schwann cells, but does not identify them as proximal Schwann cells or show that electrical guidance, rather than prestimulation, caused the increased outgrowth. (LLM, conf=0.84)
Experimental — accuracy not yet measured
- Premisecitation
Bellamkonda microchannel scaffolds stimulated the growth of essential Schwann cells and structural fibroblasts.
“scaffolds stimulated the growth of essential Schwann cells and structural fibroblasts, potentially enhancing the integrity of regenerated neural tissue”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [33] 33. Srinivasan, A., Tahilramani, M., Bentley, J.T., Gore, R.K., Millard, D.C., Mukhatyar, V.J., et al. Microchannel-based regenerative scaffold for chronic peripheral nerve inter- facing in amputees. Biomaterials 41, 151, 2015. source
The Results (§3.5) describe regenerating axons and Schwann cells within fibroblast sheaths, and the Discussion (§4) says the scaffold supports a microenvironment for proliferating Schwann cells and chronic nerve stability. These passages support cell-associated regeneration and structural organization, but do not show that the scaffold stimulated fibroblast growth specifically. Page numbers were not supplied. (LLM, conf=0.84)
Experimental — accuracy not yet measured
- Premisecitation
NGF incorporated as a codopant in the electrochemical polymerization of PEDOT and PPy resulted in increased neurite outgrowth of PC-12 cells on the NGF-containing polymer substrates.
“NGF has been incorporated as a codopant in the electrochemical polymerization of PEDOT and PPy, where PC-12 cells responded with increased neurite outgrowth on the NGFcontaining polymer substrates.53”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [53] 53. Kim, D.H., Richardson-Burns, S.M., Hendricks, J.L., Sequera, C., and Martin, D.C. Effect of immobilized nerve growth factor on conductive polymers: electrical properties and cellular response. Adv Funct Mater 17, 79, 2007. source
- Premisecitation
Abidian and coauthors reported enhanced neurite outgrowth of peripheral dorsal root ganglion neurons on PEDOT and PPy nanotubes compared with conductive polymer film substrates.
“Abidian and coauthors reported enhanced neurite outgrowth of peripheral dorsal root ganglion (DRG) neurons on PEDOT and PPy nanotubes in comparison to conductive polymer film substrates.37”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [37] 37. Abidian, M.R., Corey, J.M., Kipke, D.R., and Martin, D.C. Conducting-polymer nanotubes improve electrical proper- ties, mechanical adhesion, neural attachment and neurite outgrowth of neural electrodes. Small 6, 421, 2010. source
- Premisecitation
Electrical signals were recorded in vivo over a period of 5 months from the Bellamkonda microchannel array.
“electrical signals were recorded in vivo over a period of 5 months.33”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [33] 33. Srinivasan, A., Tahilramani, M., Bentley, J.T., Gore, R.K., Millard, D.C., Mukhatyar, V.J., et al. Microchannel-based regenerative scaffold for chronic peripheral nerve inter- facing in amputees. Biomaterials 41, 151, 2015. source
The Methods (§2.7) say signals were acquired after 20 weeks of implantation, and the Results (§3.8–3.9) report recordings at five months. The paper supports in vivo recording from the microchannel interfaces at that time point, but not recordings throughout a five-month period. Page numbers are unavailable in our copy. (LLM, conf=0.88)
Experimental — accuracy not yet measured
- Premisecitation
Stoyanova et al. 3D bifurcating electrode arrays showed promise in facilitating bifurcation of large regenerated fascicles into smaller, almost singularly separated axonal fibers in vivo.
“When tested in vivo, these electrode arrays showed promise in facilitating bifurcation of large regenerated fascicles into smaller (almost singularly) separated axonal fibers”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [35] 35. Stoyanova, I.I., van Wezel, R.J.A., and Rutten, W.L.C. In vivo testing of a 3D bifurcating microchannel scaffold in- ducing separation of regenerating axon bundles in periph- eral nerves. J Neural Eng 10, 066018, 2013. source
The Results and Discussion report that the implanted 3D bifurcating scaffold separated regenerating axonal bundles in vivo, with 91% separation into secondary channels. But the paper describes an implanted microchannel scaffold, not an electrode array, and presents contact with single axons or small groups as a future possibility rather than an observed outcome (Discussion; page unavailable). (LLM, conf=0.88)
Experimental — accuracy not yet measured
- Premisecitation
There was no evidence of neuroma formation with the Bellamkonda microchannel array.
“there was no evidence of neuroma formation”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [33] 33. Srinivasan, A., Tahilramani, M., Bentley, J.T., Gore, R.K., Millard, D.C., Mukhatyar, V.J., et al. Microchannel-based regenerative scaffold for chronic peripheral nerve inter- facing in amputees. Biomaterials 41, 151, 2015. source
- Premisecitation
The microscale geometries presented in (A-C) are each associated with an improved tissue response to the implant.
“The microscale geometries presented in (A-C) are each associated with an improved tissue response to the implant.”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [66] 66. Seymour, J.P., and Kipke, D.R. Neural probe design for reduced tissue encapsulation in CNS. Biomaterials 28, 3594, 2007. source
- cites [70] 70. Skousen, J.L., Merriam, S.M.E., Srivannavit, O., Perlin, G., Wise, K.D., and Tresco, P.A. Reducing surface area while maintaining implant penetrating profile lowers the brain foreign body response to chronically implanted planar sil- icon microelectrode arrays. Prog Brain Res 194, 167, 2011. source
- cites [71] 71. Kozai, T.D.Y., Langhals, N.B., Patel, P.R., Deng, X., Zhang, H., Smith, K.L., et al. Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nat Mater 11, 1065, 2012. source
The passage on chronic tissue responses (indexed as Abstract, page unavailable) reports that the ultrasmall MTE probes elicited a reduced tissue response compared with conventional silicon electrodes. It does not establish that each of the geometries shown in the citing document’s panels A–C improves tissue response, or isolate geometry from the probes’ other properties as the cause. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Premisecitation
Coated and uncoated microelectrode arrays implanted in rat brains showed no significant differences in neuronal and non-neuronal immunohistochemical outcomes at any time point over 12 weeks.
“microelectrode arrays to uncoated arrays implanted in rat brains using a variety of quantitative immunohistochemical metrics for neuronal and non-neuronal outcomes over a 12week period, and there were no significant differences determined at any time point between the coated and uncoated electrodes.76”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [76] 76. Winslow, B.D., Christensen, M.B., Yang, W.K., Solzba- cher, F., and Tresco, P.A. A comparison of the tissue re- sponse to chronically implanted Parylene-C-coated and uncoated planar silicon microelectrode arrays in rat cortex. Biomaterials 31, 9163, 2010. source
- Premisecitation
Concepts for neural stem and progenitor cell-seeded CNS prostheses have been reported.
“More recently, concepts for neural stem and progenitor cell-seeded CNS prostheses have been reported, where incorporation on the device has been achieved through hydrogel scaffolding 87 or attachment to laminin 88 (Fig. 4).”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [87] 87. Purcell, E.K., Seymour, J.P., Yandamuri, S., Kipke, D.R. In vivo evaluation of a neural stem cell-seeded prosthesis. J Neural Eng 6, 026005, 2009. source
- cites [88] 88. Azemi, E., Gobbel, G.T., and Cui, X.T. Seeding neural progenitor cells on silicon-based neural probes. J Neuro- surg 113, 673, 2010. source
- Premisecitation
Coating electrodes with L1 reduced gliosis surrounding devices implanted in rat motor cortex.
“Likewise, coating electrodes with L1 reduced gliosis and improved neuronal density surrounding devices implanted in rat motor cortex.77”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
- Premisecitation
In vivo, L1-coated electrodes significantly decreased 'kill zones' characterized by reduced expression of neuronal markers surrounding electrodes.
“in vivo evaluation revealed similar results: ''kill zones'' characterized by reduced expression of neuronal markers surrounding electrodes were significantly decreased when L1-coated electrodes were used.81”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [81] 81. Kolarcik, C.L., Bourbeau, D., Azemi, E., Rost, E., Zhang, L., Lagenaur, C.F., et al. In vivo effects of L1 coating on inflammation and neuronal health at the electrode-tissue interface in rat spinal cord and dorsal root ganglion. Acta Biomater 8, 3561, 2012. source
- Premisecitation
Incorporation of neural stem and progenitor cells on CNS prostheses has been achieved through hydrogel scaffolding.
“More recently, concepts for neural stem and progenitor cell-seeded CNS prostheses have been reported, where incorporation on the device has been achieved through hydrogel scaffolding 87 or attachment to laminin 88 (Fig. 4).”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [87] 87. Purcell, E.K., Seymour, J.P., Yandamuri, S., Kipke, D.R. In vivo evaluation of a neural stem cell-seeded prosthesis. J Neural Eng 6, 026005, 2009. source
- Premisecitation
Hydrogel scaffolding with neural stem cells produced an initial benefit to surrounding neuronal density within the first week of transplantation.
“In the case of hydrogel scaffolding, an initial benefit to surrounding neuronal density was noted within the first week of transplantation, which may be related to release of neurotrophic factors by the engrafted cells.89”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [89] 89. Purcell, E.K., Singh, A., and Kipke, D.R. Alginate com- position effects on a neural stem cell-seeded scaffold. Tissue Eng Part C 15, 541, 2009. source
The Discussion reports that alginate-encapsulated neural stem cells secrete neurotrophic factors, and Figure 4 reports a protective effect of their conditioned medium on cultured PC-12 cells. The paper does not report transplantation or a change in surrounding neuronal density within the first week; it describes testing the scaffolded cells’ ability to repair nervous tissue as future work. The proposed mechanism is supported, but the claimed transplant outcome is not. (LLM, conf=0.85)
Experimental — accuracy not yet measured
- Premisecitation
Azemi et al. reported a pronounced improvement in neurofilament staining within the first 100 mm of the probe surface over the entire 8-week study period.
“Azemi et al. reported a pronounced improvement in neuronal density and neurofilament staining within the first 100 mm of the probe surface over the entire 8-week study period.77”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
Results §3.1 reports greater NF-200 staining around L1-modified probes than controls within 100 μm of the interface at all three time points (1, 4, and 8 weeks). It also reports higher neuronal cell counts in that region. The citing sentence says 100 mm, which is not the distance reported by the source. (LLM, conf=0.88)
Experimental — accuracy not yet measured
- Premisecitation
In vitro, L1-modified electrodes decreased neuronal death associated with electrode implantation into the spinal cord and DRG of rats.
“In vitro, L1-modified electrodes decreased the inflammatory glial response and neuronal death associated with electrode implantation into the spinal cord and DRG of rats, 81”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [81] 81. Kolarcik, C.L., Bourbeau, D., Azemi, E., Rost, E., Zhang, L., Lagenaur, C.F., et al. In vivo effects of L1 coating on inflammation and neuronal health at the electrode-tissue interface in rat spinal cord and dorsal root ganglion. Acta Biomater 8, 3561, 2012. source
The Abstract reports decreased neuronal cell death with L1-coated electrodes implanted in rat spinal cord and DRG, and Results §3.1.3 reports the decrease in the DRG. The claim incorrectly calls this an in vitro finding: the paper says the coatings were verified in vitro, but the implantation and neuronal-death findings were in vivo. (LLM, conf=0.90)
Experimental — accuracy not yet measured
- Premisecitation
Azemi et al. reported a pronounced improvement in neuronal density within the first 100 mm of the probe surface over the entire 8-week study period.
“Azemi et al. reported a pronounced improvement in neuronal density and neurofilament staining within the first 100 mm of the probe surface over the entire 8-week study period.77”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
Results §3.1 reports significantly higher neuronal counts and NF-200 staining around L1-modified probes than controls at 1, 4, and 8 weeks. The reported region is within 100 μm of the interface, not 100 mm as printed in the citing sentence; page numbers were not provided. (LLM, conf=0.88)
Experimental — accuracy not yet measured
- Premisecitation
Nanostructured materials can influence synaptic transmission of attached neuronal networks, affecting electrophysiological function of neurons at a synthetic interface.
“Additionally, it has been shown that nanostructured materials can influence the synaptic transmission of attached neuronal networks, thus affecting the electrophysiological function of neurons at a synthetic interface.94”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [94] 94. Cellot, G., Toma, F.M., Kasap Varley, Z., Laishram, J., Villari, A., Quintana, M., et al. Carbon nanotube scaffolds tune synaptic strength in cultured neural circuits: novel frontiers in nanomaterial-tissue interactions. J Neurosci 31, 12945, 2011. source
- Premisecitation
Immunohistochemistry showed evidence of graft cells within the device channel one day following implantation in a rat brain.
“Immunohistochemistry showed evidence of graft cells (white arrowheads) within the device channel one day following implantation in a rat brain”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [87] 87. Purcell, E.K., Seymour, J.P., Yandamuri, S., Kipke, D.R. In vivo evaluation of a neural stem cell-seeded prosthesis. J Neural Eng 6, 026005, 2009. source
- Premisecitation
Coating the electrode site with PEDOT facilitates signal detection.
“Coating the electrode site with poly(3,4-ethylenedioxythiphene) (PED-OT) facilitates signal detection.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [71] 71. Kozai, T.D.Y., Langhals, N.B., Patel, P.R., Deng, X., Zhang, H., Smith, K.L., et al. Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nat Mater 11, 1065, 2012. source
- Premisecitation
Implantation of functionalized devices demonstrated improved recording quality from electrode sites positioned on the lattice edge compared with the faces, verified through computational modeling.
“Further, implantation of functionalized devices demonstrated improved recording quality from electrode sites positioned on the lattice edge as compared to the faces and verified through computational modeling.67”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [67] 67. Seymour, J.P., Langhals, N.B., Anderson, D.J., and Kipke, D.R. Novel multi-sided, microelectrode arrays for im- plantable neural applications. Biomed Microdevices 13, 441, 2011. source
In §3.5 (Neural recordings), acute animal recordings yielded more isolatable units and higher signal amplitudes at edge electrodes than at planar sites. The finite-element model in §3.4 supports a wider viewing angle for edge electrodes, but finds planar electrodes more sensitive when directly facing a neuron; it does not verify an across-the-board improvement in recording quality. (LLM, conf=0.83)
Experimental — accuracy not yet measured
- Premisecitation
Cone electrode-based devices have obtained recordings for several years in human subjects.
“recordings were obtained for several years in human subjects.85,86”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [85] 85. Bartels, J., Andreasen, D., Ehirim, P., Mao, H., Seibert, S., Wright, E.J., et al. Neurotrophic electrode: method of as- sembly and implantation into human motor speech cortex. J Neurosci Methods 174, 168, 2008. source
- cites [86] 86. Kennedy, P. Reliable neural interface: the first quarter century of the neurotrophic electrode. Conf Proc IEEE Eng Med Biol Soc 2012, 3332, 2012. source
- Premisecitation
L1 has reported neuroprotective effects.
“L1 contributes many essential functions to the maintenance of the nervous system, where it is expressed in the majority of CNS neurons and has reported neuroprotective effects.77”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
- Premisecitation
L1 is expressed in the majority of CNS neurons.
“L1 contributes many essential functions to the maintenance of the nervous system, where it is expressed in the majority of CNS neurons and has reported neuroprotective effects.77”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
- Premisecitation
In vitro, L1-modified electrodes decreased the inflammatory glial response associated with electrode implantation into the spinal cord and DRG of rats.
“In vitro, L1-modified electrodes decreased the inflammatory glial response and neuronal death associated with electrode implantation into the spinal cord and DRG of rats, 81”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [81] 81. Kolarcik, C.L., Bourbeau, D., Azemi, E., Rost, E., Zhang, L., Lagenaur, C.F., et al. In vivo effects of L1 coating on inflammation and neuronal health at the electrode-tissue interface in rat spinal cord and dorsal root ganglion. Acta Biomater 8, 3561, 2012. source
The Results report fewer activated microglia on L1-coated probes in vitro, and reduced inflammatory staining at some time points after implantation in rat spinal cord and DRG. But implantation into rats was studied in vivo, not in vitro; the Results also report increased Iba1 and GFAP staining in the DRG at four weeks with L1 coating. The claim therefore conflates the culture and implantation findings and overgeneralizes the glial response. (LLM, conf=0.90)
Experimental — accuracy not yet measured
- Premisecitation
Tissue at the electrode interface demonstrates re-emergence of markers associated with development and re-entry into the cell cycle, indicating plasticity in the local environment of neural prostheses.
“Finally, the tissue at the electrode interface demonstrates the re-emergence of markers associated with development and re-entry into the cell cycle, indicating plasticity in the local environment of neural prostheses.87,92,93”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [87] 87. Purcell, E.K., Seymour, J.P., Yandamuri, S., Kipke, D.R. In vivo evaluation of a neural stem cell-seeded prosthesis. J Neural Eng 6, 026005, 2009. source
The ‘Scaffold condition effects on neuronal and non-neuronal densities’ section (page not provided) reports doublecortin, a marker associated with migrating neuronal precursors, in host tissue one week after implantation. It treats the significance of that finding cautiously and does not report markers of cell-cycle re-entry. The claim is supported for a developmental marker at the interface, but not for cell-cycle re-entry or a demonstrated conclusion about local plasticity. (LLM, conf=0.86)
Experimental — accuracy not yet measured - cites [92] 92. Purcell, E.K., Thompson, D.E., Ludwig, K.A., and Kipke, D.R. Flavopiridol reduces the impedance of neural pros- theses in vivo without affecting recording quality. J Neu- rosci Methods 183, 149, 2009. source
- cites [93] 93. Kozai, T.D.Y., Gugel, Z., Li, X., Gilgunn, P.J., Khilwani, R., Ozdoganlar, O.B., et al. Chronic tissue response to carboxymethyl cellulose based dissolvable insertion needle for ultra-small neural probes. Biomaterials 35, 9255, 2014. source
- Premisecitation
Coating electrodes with L1 improved neuronal density surrounding devices implanted in rat motor cortex.
“Likewise, coating electrodes with L1 reduced gliosis and improved neuronal density surrounding devices implanted in rat motor cortex.77”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
- Findingown data
Unpublished data demonstrate biological viability of the RPNI interface exceeding 20 months in rodents.
“Further, unpublished data demonstrate biological viability of the interface exceeding 20 months in rodents.”
Quote in the paperNot a substantial claim - Premisecitation
A cuff electrode offers reasonable biocompatibility at the expense of selectivity in recording compound signals and stimulation of nerve fascicles.
“As the cuff electrode wraps around the exterior of the nerve, it offers reasonable biocompatibility at the expense of selectivity in recording compound signals and stimulation of nerve fascicles.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [23] 23. Del Valle, J., and Navarro, X. Interfaces with the peripheral nerve for the control of neuroprostheses. Int Rev Neurobiol 109, 63, 2013. source
- Premisecitation
Neuronal loss is most pronounced nearest the electrode surface, with about a 40% decrease in neuronal density reported within the estimated recordable radius of the device following implantation in the brain.
“Importantly, neuronal loss is most pronounced nearest the electrode surface, where a *40% decrease in neuronal density has been reported within the estimated recordable radius of the device following implantation in the brain.18,21”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [18] 18. Biran, R., Martin, D.C., and Tresco, P.A. Neuronal cell loss accompanies the brain tissue response to chronically im- planted silicon microelectrode arrays. Exp Neurol 195, 115, 2005. source
- cites [21] 21. Henze, D.A., Borhegyi, Z., Csicsvari, J., Mamiya, A., Harris, K.D., and Buzsa ´ki, G. Intracellular features pre- dicted by extracellular recordings in the hippocampus in vivo. J Neurophysiol 84, 390, 2000. source
- Premisecitation
Axons have been observed to retract by over 1 cm from the wound site following penetrating electrode implantation.
“axons have been observed to retract by over 1 cm from the wound site following penetrating electrode implantation.22”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [22] 22. Christensen, M.B., Pearce, S.M., Ledbetter, N.M., Warren, D.J., Clark, G.A., and Tresco, P.A. The foreign body re- sponse to the Utah Slant Electrode Array in the cat sciatic nerve. Acta Biomater 10, 4650, 2014. source
- Premisecitation
In the PNS, persistent inflammation, demyelination, Wallerian degeneration, neuroma formation, and encapsulation by connective tissue make maintaining patency in the electrode-tissue interface particularly challenging.
“In the PNS, persistent inflammation, demyelination, Wallerian degeneration, neuroma formation, and encapsulation by connective tissue make maintaining patency in the electrode-tissue interface particularly challenging;”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [22] 22. Christensen, M.B., Pearce, S.M., Ledbetter, N.M., Warren, D.J., Clark, G.A., and Tresco, P.A. The foreign body re- sponse to the Utah Slant Electrode Array in the cat sciatic nerve. Acta Biomater 10, 4650, 2014. source
- Premisecitation
Gomez and Schmidt reported that enhancement of neurite outgrowth from attached PC-12 cells was similar with both NGF-immobilized PPy and soluble NGF.
“Gomez and Schmidt reported covalently linking NGF to an underlying PPy substrate, where enhancement of neurite outgrowth from attached PC-12 cells was similar with both NGF-immobilized PPy and soluble NGF.49”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [49] 49. Gomez, N., and Schmidt, C.E. Nerve growth factor- immobilized polypyrrole: bioactive electrically conducting polymer for enhanced neurite extension. J Biomed Mater Res 81, 135, 2007. source
- Premisecitation
Schwann cells are essential supporters of peripheral nerve tissue growth through secreting trophic factors and assisting regeneration following injury.
“Schwann cells are essential supporters of peripheral nerve tissue growth through secreting trophic factors and assisting regeneration following injury.55”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [55] 55. Namgung, U. The role of Schwann cell-axon interaction in peripheral nerve regeneration. Cells Tissues Organs 200, 6, 2015. source
- Premisecitation
Larger signals were recorded from lower impedance, conducting polymer electrodes compared with standard stainless steel in RPNIs.
“with larger signals recorded from lower impedance, conducting polymer electrodes compared to standard stainless steel.62”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [62] 62. Kung, T.A., Langhals, N.B., Martin, D.C., Johnson, P.J., Cederna, P.S., and Urbanchek, M.G. Regenerative periph- eral nerve interface viability and signal transduction with an implanted electrode. Plast Reconstr Surg 133, 1380, 2014. source
- Premisecitation
Recordings from PNS neurons through an MCEA were validated in rats in an acute setting.
“Recordings from PNS neurons through a MCEA were later validated in rats in an acute setting.32”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [32] 32. Minev, I.R., Chew, D.J., Delivopoulos, E., Fawcett, J.W., and Lacour, S.P. High sensitivity recording of afferent nerve activity using ultra-compliant microchannel elec- trodes: an acute in vivo validation. J Neural Eng 9, 026005, 2012. source
- Premisecitation
The MCEA device allowed regeneration of miniature nerve fascicles up to 5 mm long with their own vascular structures through the implanted device.
“The MCEA device allowed for the regeneration of miniature nerve fascicles up to 5 mm long with their own vascular structures to grow through the implanted device.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [29] 29. Lacour, S.P., Atta, R., FitzGerald, J.J., Blamire, M., Tarte, E., and Fawcett, J. Polyimide micro-channel arrays for peripheral nerve regenerative implants. Sens Actuat A Phys 147, 456, 2008. source
- cites [30] 30. Lacour, S.P., Fitzgerald, J.J., Lago, N., Tarte, E., McMahon, S., and Fawcett, J. Long micro-channel electrode arrays: a novel type of regenerative peripheral nerve interface. IEEE Trans Neural Syst Rehabil Eng 17, 454, 2009. source
- Interpretationreasoning
Regenerative designs that encourage axonal growth in close proximity to electrode sites offer improved biocompatibility while maintaining effective neuron-electrode contact.
“Regenerative designs, which encourage axonal growth in close proximity to electrode sites, are a compromise alternative, offering improved biocompatibility while maintaining effective neuron-electrode contact.”
Quote in the paperSubstantial claim - Premisecitation
A subsequent in vivo assessment demonstrated increased SGN density and reduced thresholds of activation in guinea pigs receiving electrical stimulation from cochlear implants with PPy/p-TS/NT3-coated electrodes compared with subjects receiving PPy/p-TS-coated devices.
“A subsequent assessment in vivo demonstrated increased SGN density and reduced thresholds of activation in guinea pigs receiving electrical stimulation from cochlear implants with PPy/p-TS/NT3-coated electrodes in comparison to subjects receiving PPy/p-TS-coated devices.48”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [48] 48. Richardson, R.T., Wise, A.K., Thompson, B.C., Flynn, B.O., Atkinson, P.J., Fretwell, N.J., et al. Polypyrrole- coated electrodes for the delivery of charge and neuro- trophins to cochlear neurons. Biomaterials 30, 2614, 2009. source
- Premisecitation
Lu et al. reported PC12 attachment and neurite outgrowth on PPy/CNT coatings in vitro.
“Lu et al. reported PC12 attachment and neurite outgrowth on PPy/CNT coatings in vitro”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [42] 42. Lu, Y., Li, T., Zhao, X., Li, M., Cao, Y., Yang, H., et al. Electrodeposited polypyrrole/carbon nanotubes composite films electrodes for neural interfaces. Biomaterials 31, 5169, 2010. source
- Premisecitation
PPy/CNT-coated electrodes were associated with increased neuronal density upon in vivo implantation in rats.
“coated electrodes were associated with increased neuronal density upon in vivo implantation in rats.42”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [42] 42. Lu, Y., Li, T., Zhao, X., Li, M., Cao, Y., Yang, H., et al. Electrodeposited polypyrrole/carbon nanotubes composite films electrodes for neural interfaces. Biomaterials 31, 5169, 2010. source
- Premisecitation
The sieve electrode marked a substantial improvement in selectivity compared with the cuff electrode, which indiscriminately interfaced the entire nerve bundle.
“This marked a substantial improvement in selectivity in comparison to the cuff electrode, which indiscriminately interfaced the entire nerve bundle.23”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] 23. Del Valle, J., and Navarro, X. Interfaces with the peripheral nerve for the control of neuroprostheses. Int Rev Neurobiol 109, 63, 2013. source
- Premisecitation
Behan et al. found that Schwann cell responses were particularly favorable when cells were embedded within the SWCNT hydrogel.
“Behan and colleagues utilized singlewalled carbon nanotubes (SWCNTs) immersed within conductive hydrogels as a substrate for Schwann cell growth and observed responses in both 2D and 3D; they found that responses were particularly favorable when cells were embedded within the SWCNT hydrogel.59”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [59] 59. Behan, B.L., DeWitt, D.G., Bogdanowicz, D.R., Koppes, A.N., Bale, S.S., and Thompson, D.M. Single-walled car- bon nanotubes alter Schwann cell behavior differentially within 2D and 3D environments. J Biomed Mater Res Part A 96, 46, 2011. source
- Premisecitation
Conducting polymer/carbon nanotube composites as electrode coatings have exhibited favorable stability, biocompatibility, and conductivity characteristics.
“studies have reported the development of conducting polymer/carbon nanotube (CNT) composites as electrode coatings, 41 as these materials have exhibited favorable stability, biocompatibility, and conductivity characteristics.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [41] 41. Keefer, E.W., Botterman, B.R., Romero, M.I., Rossi, A.F., and Gross, G.W. Carbon nanotube coating improves neu- ronal recordings. Nat Nanotechnol 3, 434, 2008. source
- Premisenone given
Several approaches to regenerating and preserving neuronal function and density at the CNS device interface have demonstrated success.
“Nonetheless, several approaches to regenerating and preserving neuronal function and density at the device interface have demonstrated success”
Quote in the paperSubstantial claim - Premisecitation
NT3 is a neurotrophin known to support survival of auditory nerve cells, including spiral ganglion neurons.
“neurotrophin-3 (NT3), which is a neurotrophin known to support survival of auditory nerve cells (spiral ganglion neurons [SGNs]).54”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [54] 54. Ernfors, P., Duan, M.L., ElShamy, W.M., and Canlon, B. Protection of auditory neurons from aminoglycoside tox- icity by neurotrophin-3. Nat Med 2, 463, 1996. source
- Premisecitation
Surfaces rendered from conductive polymers PEDOT and PPy can encourage neurite outgrowth from peripheral neurons when used to create scaffolds with neurite-promoting topographical cues.
“surfaces rendered from these materials can encourage neurite outgrowth from peripheral neurons when used to create scaffolds with neurite-promoting topographical cues.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [38] 38. Green, R.A., Lovell, N.H., Wallace, G.G., and Poole- Warren, L.A. Conducting polymers for neural interfaces: challenges in developing an effective long-term implant. Biomaterials 29, 3393, 2008. source
- Premisenone given
Translocation of the free muscle graft results in preliminary degeneration of the muscle graft while awaiting vessel inosculation and revascularization.
“Translocation of the free muscle graft results in preliminary degeneration of the muscle graft while awaiting vessel inosculation and revascularization.”
Quote in the paperNot a substantial claim - Interpretationcitation
The results indicate a synergistic effect of neurotrophin release and electrical stimulation in promoting neuronal survival at the implanted electrode interface.
“These results indicate a synergistic effect of neurotrophin release and electrical stimulation in promoting neuronal survival at the implanted electrode interface.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [47] 47. Richardson, R.T., Thompson, B., Moulton, S., Newbold, C., Lum, M.G., Cameron, A., et al. The effect of poly- pyrrole with incorporated neurotrophin-3 on the promotion of neurite outgrowth from auditory neurons. Biomaterials 28, 513, 2007. source
- cites [48] 48. Richardson, R.T., Wise, A.K., Thompson, B.C., Flynn, B.O., Atkinson, P.J., Fretwell, N.J., et al. Polypyrrole- coated electrodes for the delivery of charge and neuro- trophins to cochlear neurons. Biomaterials 30, 2614, 2009. source
- Premisecitation
Penetrating arrays have excellent integration with individual nerve fibers at the expense of a poorer tissue response due to trauma and inflammation resulting from device implantation.
“For penetrating arrays, the opposite is true: excellent integration with individual nerve fibers comes at the expense of a poorer tissue response due to the trauma and inflammation resulting from device implantation.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] 23. Del Valle, J., and Navarro, X. Interfaces with the peripheral nerve for the control of neuroprostheses. Int Rev Neurobiol 109, 63, 2013. source
- Premisecitation
Compound muscle action potentials were successfully recorded from RPNIs over the entire 7-month evaluation period.
“Compound muscle action potentials were successfully recorded from RPNIs over the entire 7-month evaluation period”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [62] 62. Kung, T.A., Langhals, N.B., Martin, D.C., Johnson, P.J., Cederna, P.S., and Urbanchek, M.G. Regenerative periph- eral nerve interface viability and signal transduction with an implanted electrode. Plast Reconstr Surg 133, 1380, 2014. source
- Premisecitation
Polymer substrates with micron- and nanoscale topographical features can accelerate and guide axonal outgrowth to a desired target.
“Polymer substrates with micron-and nanoscale topographical features can accelerate and guide axonal outgrowth to a desired target, 36”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [36] 36. Purcell, E.K., Naim, Y., Yang, A., Leach, M.K., Velkey, J.M., Duncan, R.K., et al. Combining topographical and genetic cues to promote neuronal fate specification in stem cells. Biomacromolecules 13, 3427, 2012. source
- Interpretationnone given
In amputees, residual muscle pieces within the amputated limb could typically be harvested for use in RPNI construction.
“In the case of amputees, there are typically residual muscle pieces within the amputated limb that could be harvested for use in RPNI construction.”
Quote in the paperSubstantial claim - Premisecitation
Electrical stimulation increased neuritogenesis of rat pheochromocytoma PC12 cells cultured on PPy-coated poly(lactic-co-glycolic acid) nanofiber substrates compared with unstimulated cells.
“stimulation increased neuritogenesis of rat pheochromocytoma (PC12) cells cultured on PPycoated poly(lactic-co-glycolic acid) nanofiber substrates in comparison to unstimulated cells.40”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [40] 40. Lee, J.Y., Bashur, C.A., Goldstein, A.S., and Schmidt, C.E. Polypyrrole-coated electrospun PLGA nanofibers for neural tissue applications. Biomaterials 30, 4325, 2009. source
- Premisecitation
Fascicles within the MCEA maintained a similar structure and function to nerves regenerated through an ordinary silicone tube.
“These fascicles within the MCEA maintained a similar structure and function to nerves regenerated through an ordinary silicone tube.”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [29] 29. Lacour, S.P., Atta, R., FitzGerald, J.J., Blamire, M., Tarte, E., and Fawcett, J. Polyimide micro-channel arrays for peripheral nerve regenerative implants. Sens Actuat A Phys 147, 456, 2008. source
- cites [30] 30. Lacour, S.P., Fitzgerald, J.J., Lago, N., Tarte, E., McMahon, S., and Fawcett, J. Long micro-channel electrode arrays: a novel type of regenerative peripheral nerve interface. IEEE Trans Neural Syst Rehabil Eng 17, 454, 2009. source
- Premisecitation
Castro et al. reported preferential growth of unmyelinated and sensory axons through sieve electrodes implanted in rat sciatic nerves compared with myelinated and motor axons.
“Castro et al.reported differential growth of certain axonal subtypes through sieve electrodes implanted in rat sciatic nerves: preferential growth of unmyelinated and sensory axons occurred in comparison to myelinated and motor axons.28”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [28] 28. Castro, J., Negredo, P., and Avendan ˜o, C. Fiber composi- tion of the rat sciatic nerve and its modification during regeneration through a sieve electrode. Brain Res 1190, 65, 2008. source
- Premisecitation
L1 reduces attachment of glial cells in vitro.
“L1 has also been shown to promote neuronal growth and neurite extension, as well as reducing attachment of glial cells in vitro.79”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [79] 79. Azemi, E., Stauffer, W.R., Gostock, M.S., Lagenaur, C.F., and Cui, X.T. Surface immobilization of neural adhesion molecule L1 for improving the biocompatibility of chronic neural probes: in vitro characterization. Acta Biomater 4, 1208, 2008. source
- Interpretationnone given
While some materials show promise in mitigating the tissue response in vitro, these impacts do not necessarily directly translate to findings observed in vivo.
“However, while some materials show promise in mitigating the tissue response in the in vitro setting, these impacts do not necessarily directly translate to findings observed in vivo.”
Quote in the paperNot a substantial claim - Interpretationnone given
Smaller electrode sites facilitate high-density recording arrays with improved spatial sampling resolution.
“Nevertheless, smaller electrode sites facilitate high-density recording arrays with improved spatial sampling resolution, and there is ample room for further development of this approach.”
Quote in the paperNot a substantial claim - Premisecitation
The laminin-incorporating nanoscale coating demonstrated superior adhesion and neurite outgrowth from cortical neurons cultured in vitro compared with alternative coatings and control wafers.
“The laminin-incorporating coating demonstrated superior adhesion and neurite outgrowth from cortical neurons cultured in vitro compared to alternative coatings and control wafers.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [73] 73. He, W., and Bellamkonda R.V. Nanoscale neuro- integrative coatings for neural implants. Biomaterials 26, 2983, 2005. source
- Premisecitation
Incorporation of neural stem and progenitor cells on CNS prostheses has been achieved through attachment to laminin.
“More recently, concepts for neural stem and progenitor cell-seeded CNS prostheses have been reported, where incorporation on the device has been achieved through hydrogel scaffolding 87 or attachment to laminin 88 (Fig. 4).”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [88] 88. Azemi, E., Gobbel, G.T., and Cui, X.T. Seeding neural progenitor cells on silicon-based neural probes. J Neuro- surg 113, 673, 2010. source
- Premisecitation
Early in vivo results indicated that L1 may aid in decreasing mechanical failure and inflammation of electrodes implanted in the brain.
“Early in vivo results indicated that L1 may aid in decreasing mechanical failure and inflammation of electrodes implanted in the brain.80”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [80] 80. Zhong, Y., Yu, X., Gilbert, R., and Bellamkonda R V. Stabilizing electrode-host interfaces: a tissue engineering approach. J Rehabil Res Dev 38, 627, 2001.
- Premisecitation
Cone electrode-based devices have been validated to obtain long-term recordings in rats and nonhuman primates.
“signals were successfully recorded in rats and nonhuman primates, 83,84”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [83] 83. Kennedy, P.R., Mirra, S.S., and Bakay, R.A.E. The cone electrode: ultrastructural studies following long-term re- cording in rat and monkey cortex. Neurosci Lett 142, 89, 1992. source
- cites [84] 84. Kennedy, P.R., and Bakay, R.A. Activity of single action potentials in monkey motor cortex during long-term task learning. Brain Res 760, 251, 1997. source
- Premisecitation
Wu et al. reported successful culture of hippocampal and cortical neurons on hyaluronic acid-based films fabricated through layer-by-layer assembly.
“Wu et al., who reported successful culture of hippocampal and cortical neurons on hyaluronic acidbased films fabricated through LBL assembly.74”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [74] 74. Wu, Z.-R., Ma, J., Liu, B.-F., Xu, Q.-Y., and Cui, F.-Z. Layer-by-layer assembly of polyelectrolyte films improving cytocompatibility to neural cells. J Biomed Mater Res Part A 81, 355, 2007. source
- Premisecitation
Detection of graft cells following implantation was limited to early time points in transplantation studies.
“However, as is common in transplantation studies, 90 detection of graft cells following implantation was limited to early time points”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [90] 90. Dimmeler, S., Ding, S., Rando, T.A., and Trounson, A. Translational strategies and challenges in regenerative medicine. Nat Med 20, 814, 2014. source
- Premisecitation
Acute-phase production of inflammatory cytokines is increased surrounding smaller devices implanted in the rat brain.
“Somewhat paradoxically, it has been observed that the acute-phase production of inflammatory cytokines is increased surrounding smaller devices implanted in the rat brain;”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [68] 68. Karumbaiah, L., Saxena, T., Carlson, D., Patil, K., Patkar, R., Gaupp, E.A., et al. Relationship between intracortical electrode design and chronic recording function. Bioma- terials 34, 8061, 2013. source
- Interpretationown data
Translating in vitro effects on electrode coatings to in vivo outcomes is challenging because the intact brain provides an extraordinarily complex environment for effective regeneration at the device-tissue interface.
“These results illustrate the challenges posed in translating in vitro effects to in vivo outcomes, where the intact brain poses an extraordinarily complex environment for effective regeneration at the device-tissue interface.”
Quote in the paperSubstantial claim - Interpretationreasoning
Survival of transplanted cells is likely to be a challenge with these approaches.
“and survival of transplanted cells is likely to be a challenge with these approaches.”
Quote in the paperNot a substantial claim - Premisecitation
Skousen et al. found that neuronal loss was mitigated surrounding a lattice silicon neural probe featuring reduced surface area compared with a conventional, single shank silicon electrode implanted in the rat cerebral cortex.
“Skousen et al. found that neuronal loss was mitigated surrounding a lattice silicon neural probe featuring reduced surface area in comparison to a conventional, single shank silicon electrode implanted in the rat cerebral cortex (Fig. 3B).70”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [70] 70. Skousen, J.L., Merriam, S.M.E., Srivannavit, O., Perlin, G., Wise, K.D., and Tresco, P.A. Reducing surface area while maintaining implant penetrating profile lowers the brain foreign body response to chronically implanted planar sil- icon microelectrode arrays. Prog Brain Res 194, 167, 2011. source
- May combine two claims. 32
- Classified as probably not a substantial claim. 8
May combine two claims.
PremisecitationNeuronal loss, signatures of neurodegeneration, and activation of neuronal cell death pathways have been observed in tissue local to devices implanted in the CNS.
“Neuronal loss, signatures of neurodegeneration, and activation of neuronal cell death pathways have been observed in the tissue local to devices implanted in the CNS.[18][19][20]”
Quote in the paperSubstantial claim1 of 3 sources checked- cites [18] 18. Biran, R., Martin, D.C., and Tresco, P.A. Neuronal cell loss accompanies the brain tissue response to chronically im- planted silicon microelectrode arrays. Exp Neurol 195, 115, 2005. source
- cites [19] 19. Kozai, T.D.Y., Li, X., Bodily, L.M., Caparosa, E.M., Ze- nonos, G.A., Carlisle, D.L., et al. Effects of caspase-1 knockout on chronic neural recording quality and longev- ity: insight into cellular and molecular mechanisms of the reactive tissue response. Biomaterials 35, 9620, 2014. source
- cites [20] 20. McConnell, G.C., Rees, H.D., Levey, A.I., Gutekunst, C.A., Gross, R.E., and Bellamkonda, R.V. Implanted neural electrodes cause chronic, local inflammation that is corre- lated with local neurodegeneration. J Neural Eng 2009, 6, 056003. source
May combine two claims.
PremisecitationIn the CNS, small signal amplitudes, signal instability, and signal loss over time are often observed.
“In the CNS, small signal amplitudes, signal instability, and signal loss over time are often observed.14,15”
Quote in the paperSubstantial claim2 of 2 sources checked- cites [14] 14. Chestek, C.A., Gilja, V., Nuyujukian, P., Foster, J.D., Fan, J.M., Kaufman, M.T., et al. Long-term stability of neural prosthetic control signals from silicon cortical arrays in rhesus macaque motor cortex. J Neural Eng 8, 045005, 2011. source
The Results (sections 3.1 and 3.2; page not provided) report declining action-potential amplitudes in three of four cortical arrays and waveform variation over hours and days. The Discussion (section 4; page not provided) says these changes make single units difficult to maintain, but it does not establish that small amplitudes or complete signal loss are often observed across the CNS; recorded neural activity and decoder performance could persist. (LLM, conf=0.85)
Experimental — accuracy not yet measured - cites [15] 15. Perge, J.A., Homer, M.L., Malik, W.Q., Cash, S., Eskandar, E., Friehs, G., et al. Intra-day signal instabilities affect decoding performance in an intracortical neural interface system. J Neural Eng 10, 036004, 2013. source
The Introduction describes commonly observed instability in intracortical recordings, declining spike amplitudes, and fewer recorded action potentials over months. Section 3.3 discusses units with small spike amplitudes, but does not establish that small amplitudes are often observed throughout the CNS. The claim generalizes beyond the intracortical interface recordings the paper discusses. (LLM, conf=0.82)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationCommon sources of recording failures in the CNS include demyelination of adjacent neurons, breakdown of the blood-brain barrier around the implant site, neuronal loss, degradation of the electrode, and formation of a glial sheath around implanted electrodes.
“Common sources of recording failures in the CNS include demyelination of the adjacent neurons, breakdown of the blood-brain barrier around the implant site, neuronal loss, degradation of the electrode, and formation of a glial sheath around implanted electrodes.64,65”
Quote in the paperSubstantial claim1 of 2 sources checked- cites [64] 64. Polikov, V.S., Tresco, P.A., and Reichert, W.M. Response of brain tissue to chronically implanted neural electrodes. J Neurosci Methods 148, 1, 2005. source
- cites [65] 65. Kozai, T.D.Y., Jaquins-Gerstl, A.S. Vazquez, A.L., Mi- chael, A.C., and Cui, X.T. Brain tissue responses to neural implants impact signal sensitivity and intervention strate- gies. ACS Chem Neurosci 6, 48, 2015. source
May combine two claims.
PremisecitationMesenchymal stem cells are a promising alternative cell source for deriving Schwann-like cells and can be readily obtained from bone marrow or adipose tissue.
“mesenchymal stem cells are a promising alternative cell source for deriving Schwann-like cells and can be readily obtained from bone marrow or adipose tissue.56”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [56] 56. Rodrigues, M.C.O., Rodrigues, A.A., Glover, L.E., Vol- tarelli, J., and Borlongan, C.V. Peripheral nerve repair with cultured schwann cells: getting closer to the clinics. Sci- entificWorldJournal 2012, 1, 2012. source
Section 4 says bone-marrow MSCs can differentiate into Schwann-like cells and describes their ease of access; it also identifies adipose tissue as an alternative source and discusses adipose-derived Schwann cells. The Conclusions describe minimally invasive bone-marrow harvesting. Our copy does not establish that adipose-derived MSCs are readily obtained, so that part of the claim is not fully supported. (LLM, conf=0.82)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationLotfi et al. results indicated preferential neurite outgrowth of subtypes of neurons toward the expected target locations, providing proof-of-concept support for selective neuronal regeneration.
“The results presented indicated preferential neurite outgrowth of subtypes of neurons toward the expected target locations, providing a proof-of-concept support for this type of approach.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [34] 34. Lotfi, P., Garde, K., Chouhan, A.K., Bengali, E., and Romero-Ortega, M.I. Modality-specific axonal regenera- tion: toward selective regenerative neural interfaces. Front Neuroeng 4, 1, 2011. source
May combine two claims.
InterpretationcitationThe microchannel array design allowed for a considerable quantity of neural tissue regeneration with minimal scar tissue and inflammation.
“This study concluded that the microchannel array design allowed for a considerable quantity of neural tissue regeneration with minimal scar tissue and inflammation”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [33] 33. Srinivasan, A., Tahilramani, M., Bentley, J.T., Gore, R.K., Millard, D.C., Mukhatyar, V.J., et al. Microchannel-based regenerative scaffold for chronic peripheral nerve inter- facing in amputees. Biomaterials 41, 151, 2015. source
The Discussion reports robust axon regeneration in the larger microchannels and says the implants were generally well tolerated with little to no scar tissue. It does not report minimal inflammation for this scaffold; its references to inflammation concern other electrode designs. Page numbers were not available in our copy. (LLM, conf=0.84)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationRegenerated fibers in the Stoyanova arrays contained both myelinated and unmyelinated fibers of varying diameter, suggesting the possible presence of motor, autonomic, and sensory neurites.
“These regenerated fibers contained both myelinated and unmyelinated fibers of varying diameter, which suggests the possible presence of motor, autonomic, and sensory neurites.”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [35] 35. Stoyanova, I.I., van Wezel, R.J.A., and Rutten, W.L.C. In vivo testing of a 3D bifurcating microchannel scaffold in- ducing separation of regenerating axon bundles in periph- eral nerves. J Neural Eng 10, 066018, 2013. source
May combine two claims.
PremisecitationPEDOT/CNT-coated electrodes exhibited improved neuronal responses at the interface and decreased expression of neuronal apoptosis markers following implantation into the DRG of rats.
“PEDOT/CNT films were assessed for their utility in modifying a PNS interface, and it was found that coated electrodes exhibited improved neuronal responses at the interface and decreased expression of neuronal apoptosis markers following implantation into the DRG of rats.43”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [43] 43. Kolarcik, C.L., Catt, K., Rost, E., Albrecht, I.N., Bourbeau, D., Du, Z., et al. Evaluation of poly(3,4-ethylenediox- ythiophene)/carbon nanotube neural electrode coatings for stimulation in the dorsal root ganglion. J Neural Eng 12, 016008, 2015. source
May combine two claims.
PremisecitationVascular endothelial growth factor is known to promote revascularization, axonal outgrowth, and Schwann cell proliferation.
“Vascular endothelial growth factor (VEGF) is known to promote revascularization, axonal outgrowth, and Schwann cell proliferation.58”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [58] 58. Sondell, M., Lundborg, G., and Kanje, M. Vascular endo- thelial growth factor has neurotrophic activity and stimu- lates axonal outgrowth, enhancing cell survival and Schwann cell proliferation in the peripheral nervous sys- tem. J Neurosci 19, 5731, 1999. source
The Abstract (page not provided) reports VEGF-induced axonal outgrowth and increased Schwann cell proliferation in cultured adult mouse ganglia. It also says VEGF promotes angiogenesis in vivo, but does not specifically establish revascularization—the restoration of a blood supply—so that part of the claim is more specific than the source supports. (LLM, conf=0.82)
Experimental — accuracy not yet measured
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationMicrochannel-based devices guide regenerating axons and parcel outgrowth toward distinct target locations.
“Microchannel-based devices guide regenerating axons and parcel outgrowth toward distinct target locations.”
Quote in the paperNot a substantial claim2 of 3 sources checked- cites [95] 95. Srinivasan, A., Liang, G., and Bellamkonda, R.V. ''Re- generative microchannel electrode array for peripheral nerve interfacing.'' Neural Engineering (NER), 2011 5th International IEEE/EMBS Conference on. IEEE, 2011. source
- cites [34] 34. Lotfi, P., Garde, K., Chouhan, A.K., Bengali, E., and Romero-Ortega, M.I. Modality-specific axonal regenera- tion: toward selective regenerative neural interfaces. Front Neuroeng 4, 1, 2011. source
- cites [35] 35. Stoyanova, I.I., van Wezel, R.J.A., and Rutten, W.L.C. In vivo testing of a 3D bifurcating microchannel scaffold in- ducing separation of regenerating axon bundles in periph- eral nerves. J Neural Eng 10, 066018, 2013. source
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationParylene C is a biocompatible polymer commonly used to insulate recording electrodes, and it has been shown to reduce attachment of microglia and macrophages to an underlying substrate in vitro.
“Likewise, parylene C is a biocompatible polymer commonly used to insulate recording electrodes, and it is has been shown to reduce attachment of microglia and macrophages to an underlying substrate in vitro, 76”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [76] 76. Winslow, B.D., Christensen, M.B., Yang, W.K., Solzba- cher, F., and Tresco, P.A. A comparison of the tissue re- sponse to chronically implanted Parylene-C-coated and uncoated planar silicon microelectrode arrays in rat cortex. Biomaterials 31, 9163, 2010. source
The Introduction describes Parylene-C as a common insulating coating for cortical recording electrodes. The Results (§3.1) and Discussion (§4) report a 95% reduction in activated microglial attachment to Parylene-C-coated coverslips in vitro; the Introduction also describes reduced brain macrophage attachment. Our copy does not explicitly establish the additional characterization of Parylene-C as biocompatible. (LLM, conf=0.83)
Experimental — accuracy not yet measured
May combine two claims.
PremisecitationWhile typical neuronal loss occurred at the device interface surrounding control implants, no such reduction in neuronal density was evident surrounding L1-coated implants.
“Remarkably, while there was evidence of typical neuronal loss at the device interface surrounding control implants, no such reduction in neuronal density was evident surrounding L1-coated implants.”
Quote in the paperSubstantial claim1 of 1 sources checked- cites [77] 77. Azemi, E., Lagenaur, C.F., and Cui, X.T. The surface im- mobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681, 2011. source
May combine two claims.
PremisecitationNext-generation devices fabricated from parylene-insulated carbon fibers with antibiofouling poly(ethylene glycol) methacrylate coatings have demonstrated robust recording quality and a minimal accompanying tissue response.
“Next-generation devices fabricated from parylene-insulated carbon fibers with antibiofouling poly(ethylene glycol) methacrylate coatings have demonstrated robust recording quality and a minimal accompanying tissue response (Fig. 3C).71”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [71] 71. Kozai, T.D.Y., Langhals, N.B., Patel, P.R., Deng, X., Zhang, H., Smith, K.L., et al. Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nat Mater 11, 1065, 2012. source
May combine two claims.
PremisecitationIn a follow-up study, the nanoscale coating in vivo following implantation in rat brains showed a positive impact on reducing gliosis at the 4 week time point, but no apparent effect on the neuronal density local to the implant.
“In a follow-up study, He and colleagues evaluated the nanoscale coating in vivo following implantation in rat brains, where it showed a positive impact on reducing gliosis at the 4 week time point, but no apparent effect on the neuronal density local to the implant.75”
Quote in the paperNot a substantial claim1 of 1 sources checked- cites [75] 75. He, W., McConnell, G.C., and Bellamkonda, R.V. Na- noscale laminin coating modulates cortical scarring re- sponse around implanted silicon microelectrode arrays. J Neural Eng 3, 316, 2006. source
May combine two claims.
PremisecitationSmaller electrode sites necessitated by finer-scale geometries have the drawback of increased impedance, but inclusion of conductive polymers mitigates this effect and allows improved signal detection.
“While the smaller electrode sites necessitated by finer-scale geometries have the drawback of increased impedance, inclusion of conductive polymers mitigates this effect and allows improved signal detection.71,72”
Quote in the paperNot a substantial claim1 of 2 sources checked- cites [71] 71. Kozai, T.D.Y., Langhals, N.B., Patel, P.R., Deng, X., Zhang, H., Smith, K.L., et al. Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nat Mater 11, 1065, 2012. source
- cites [72] 72. Ludwig, K.A., Langhals, N.B., Joseph, M.D., Richardson- Burns, S.M., Hendricks, J.L., and Kipke, D.R. Poly (3,4-ethylenedioxythiophene) (PEDOT) polymer coatings facilitate smaller neural recording electrodes. J Neural Eng 8, 014001, 2011. source
May combine two claims.
Interpretationnone givenBioengineering strategies to stabilize or regenerate the neuronal interface of implanted devices are amenable to the naturally regenerative environment of the PNS, and applications in the more challenging CNS environment have also exhibited success.
“Such strategies are amenable to the naturally regenerative environment of the PNS, but applications in the more challenging environment of the CNS have also exhibited success.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
Premisenone givenInstability and poor patency in the tissue-electrode interface undermine the longevity and performance of neural prostheses.
“but instability and poor patency in the tissue-electrode interface undermines the longevity and performance of these devices”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationA natural spatial separation occurred between regenerating motor and sensory axons, with larger-diameter, slower-growth motor axons tending to occupy the periphery of the sieve and smaller, faster-regenerating sensory axons found primarily in the center.
“A natural spatial separation occurred between regenerating motor and sensory axons, where larger-diameter, slowergrowth motor axons tended to occupy the periphery of the device, and smaller, faster-regenerating sensory axons were found primarily in the center of the sieve.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [27] 27. Lago, N., Ceballos, D., Rodrı ´guez, F.J., Stieglitz, T., and Navarro, X. Long term assessment of axonal regeneration through polyimide regenerative electrodes to interface the peripheral nerve. Biomaterials 26, 2021, 2005. source
May combine two claims.
PremisecitationBioactive NGF delivered from PEDOT doped with p-toluene sulfonate promoted neurite outgrowth from PC12 cells, but results with a laminin peptide-doped material were less encouraging in both physical and biological assessment.
“while bioactive NGF delivered from PEDOT doped with paratoluene sulfonate (p-TS) promoted neurite outgrowth from PC12 cells, results with a laminin peptide-doped material were less encouraging, both in terms of the physical and biological assessment of the material.45”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [45] 45. Green, R.A., Lovell, N.H., and Poole-Warren, L.A. Impact of co-incorporating laminin peptide dopants and neuro- trophic growth factors on conducting polymer properties. Acta Biomater 6, 63, 2010. source
May combine two claims.
PremisecitationIn chronic studies in rats, the muscle is reinnervated by the peripheral nerve, and the formation of new neuromuscular junctions is evident.
“In chronic studies in rats, the muscle is reinnervated by the peripheral nerve, and the formation of new neuromuscular junctions is evident.62”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [62] 62. Kung, T.A., Langhals, N.B., Martin, D.C., Johnson, P.J., Cederna, P.S., and Urbanchek, M.G. Regenerative periph- eral nerve interface viability and signal transduction with an implanted electrode. Plast Reconstr Surg 133, 1380, 2014. source
May combine two claims.
PremisecitationAn initial report demonstrated enhanced neurite outgrowth from SGNs cultured on PPy that incorporated NT3, and outcomes were further improved by applying electrical stimulation in a paradigm mimicking normal cochlear implant function.
“An initial report demonstrated enhanced neurite outgrowth from SGNs cultured on PPy that incorporated NT3, and outcomes were further improved by applying electrical stimulation in a paradigm mimicking normal cochlear implant function.47”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [47] 47. Richardson, R.T., Thompson, B., Moulton, S., Newbold, C., Lum, M.G., Cameron, A., et al. The effect of poly- pyrrole with incorporated neurotrophin-3 on the promotion of neurite outgrowth from auditory neurons. Biomaterials 28, 513, 2007. source
May combine two claims.
PremisecitationIn the PNS, nerve growth factor supports neuronal survival and neurite outgrowth.
“In the PNS, nerve growth factor (NGF) supports neuronal survival and neurite outgrowth, 52”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [52] 52. Chan, K.M., Gordon, T., Zochodne, D.W., and Power, H.A. Improving peripheral nerve regeneration: from molecular mechanisms to potential therapeutic targets. Exp Neurol 261, 826, 2014. source
May combine two claims.
Premisenone givenThe CNS has little natural capacity to regenerate, and insertion trauma, inflammation, and secondary injury response associated with chronically implanted microelectrode arrays present significant challenges to achieving effective long-term device function.
“The CNS has little natural capacity to regenerate, and the insertion trauma, inflammation, and secondary injury response associated with chronically implanted microelectrode arrays present significant challenges to achieving effective long-term device function.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningGrafting muscle to the end of a severed peripheral nerve results in a minimal overall neural regeneration time because peripheral nerves continue to regenerate following injury at about 1 mm per day.
“Since peripheral nerves continue to regenerate following an injury at *1 mm per day, grafting the muscle to the end of the nerve results in a minimal overall neural regeneration time.”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationreasoningBy grafting muscle to the end of the nerve, the standard signal pathway is disrupted, making the RPNI interface appropriate for amputees and distal nerve injuries.
“By grafting muscle to the end of the nerve, the standard signal pathway is disrupted, making this interface appropriate for amputees and distal nerve injuries.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationIn an in vivo assessment of chronically implanted polyimide arrays, regenerating fibers were observed in all 30 implanted rats with little evidence of axonopathy, marking an improvement over silicon arrays.
“In an in vivo assessment of chronically implanted polyimide arrays, regenerating fibers were observed in all 30 implanted rats with little evidence of axonopathy, marking an improvement over silicon arrays.27”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [27] 27. Lago, N., Ceballos, D., Rodrı ´guez, F.J., Stieglitz, T., and Navarro, X. Long term assessment of axonal regeneration through polyimide regenerative electrodes to interface the peripheral nerve. Biomaterials 26, 2021, 2005. source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based). May combine two claims.
Interpretationnone givenGiven a small muscle harvest size of less than three centimeters, there is limited resultant donor site morbidity, which may regenerate from nearby myocytes.
“Given a small muscle harvest size of less than three centimeters, there is limited resultant donor site morbidity, which may regenerate from nearby myocytes.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationA longer-term assessment over 14 months demonstrated that RPNIs had positive histological outcomes and robust signal detection over the duration of the study.
“A longer-term assessment over 14 months again demonstrated that RPNIs had positive histological outcomes and robust signal detection over the duration of the study.63”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [63] 63. Langhals, N.B., Member, I., Woo, S.L., Moon, J.D., Larson J.V, Leach, M.K., et al. Electrically stimulated signals from a long-term regenerative peripheral nerve interface. In Engineering in Medicine and Biology Society (EMBC), 2014 36th Annual International Conference of the IEEE, 1989, 2014. source
May combine two claims.
InterpretationreasoningRPNIs do not record signals as large as healthy muscles, but the muscle tissue serves as a biological amplifier by increasing the signal amplitude by 50-100 times compared with what can currently be recorded from state of the art peripheral nerve interfaces.
“While RPNIs do not record signals as large as healthy muscles, due to the regeneration and reinnervation process, the muscle tissue serves as a biological amplifier by increasing the signal amplitude by 50-100 times compared to what can currently be recorded from state of the art peripheral nerve interfaces.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
InterpretationreasoningSmaller, less invasive device designs may produce a more favorable neuronal response to implanted CNS prostheses.
“It is intuitive that smaller, less invasive device designs may produce a more favorable neuronal response to implanted CNS prostheses.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationcitationImproved recording quality may be related to reduced non-neuronal density near the lateral features, and reactive glia can produce inflammatory cytokines and chemokines with neurotoxic potential.
“This may be related to reduced non-neuronal density near the lateral features; reactive glia can produce inflammatory cytokines and chemokines with neurotoxic potential.18,68,69”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [18] 18. Biran, R., Martin, D.C., and Tresco, P.A. Neuronal cell loss accompanies the brain tissue response to chronically im- planted silicon microelectrode arrays. Exp Neurol 195, 115, 2005. source
- cites [68] 68. Karumbaiah, L., Saxena, T., Carlson, D., Patil, K., Patkar, R., Gaupp, E.A., et al. Relationship between intracortical electrode design and chronic recording function. Bioma- terials 34, 8061, 2013. source
- cites [69] 69. Skousen, J.L., Bridge, M.J., and Tresco, P.A. A strategy to passively reduce neuroinflammation surrounding devices implanted chronically in brain tissue by manipulating de- vice surface permeability. Biomaterials 36, 33, 2015. source
Probably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
PremisecitationSeymour and Kipke reported increased neuronal density adjacent to finer-scale, lateral features of an open-architecture polymer neural probe implanted in the motor cortex of rats compared with tissue adjacent to the main shank.
“In 2007, Seymour and Kipke reported increased neuronal density adjacent to finer-scale, lateral features of FIG. 2. Regenerative peripheral nerve interface (RPNI) to record prosthetic control signals from severed peripheral nerves.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [66] 66. Seymour, J.P., and Kipke, D.R. Neural probe design for reduced tissue encapsulation in CNS. Biomaterials 28, 3594, 2007. source
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationAn alternative strategy for incorporating neural progenitors on neural prostheses immobilizes laminin on the surface of silicon arrays to facilitate cellular attachment to the device.
“An alternative strategy for incorporating neural progenitors on neural prostheses immobilizes laminin on the surface of silicon arrays to facilitate cellular attachment to the device.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [88] 88. Azemi, E., Gobbel, G.T., and Cui, X.T. Seeding neural progenitor cells on silicon-based neural probes. J Neuro- surg 113, 673, 2010. source
Probably not a substantial claim: may not pass S4 (based). May combine two claims.
Interpretationnone givenCurrent designs suffer from poor integration between electrodes and neurons, with an extreme mismatch between the properties of biological and synthetic substrates at the device interface.
“Nonetheless, current designs suffer from poor integration between electrodes and neurons, where an extreme mismatch exists between the properties of the biological and synthetic substrates at the device interface.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningSmaller device designs offer the dual benefit of reduced insertion damage due to a smaller footprint and the capability to tap into neural circuitry with higher spatial resolution in a more distributed manner.
“These designs offer the dual benefit of reduced insertion damage due to a smaller footprint of the implanted device, as well as the capability to tap into neural circuitry with higher spatial resolution in a more distributed manner.”
Quote in the paperSubstantial claimMay combine two claims.
PremisecitationL1 promotes neuronal growth and neurite extension.
“L1 has also been shown to promote neuronal growth and neurite extension, as well as reducing attachment of glial cells in vitro.79”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [79] 79. Azemi, E., Stauffer, W.R., Gostock, M.S., Lagenaur, C.F., and Cui, X.T. Surface immobilization of neural adhesion molecule L1 for improving the biocompatibility of chronic neural probes: in vitro characterization. Acta Biomater 4, 1208, 2008. source
May combine two claims.
PremisecitationThe neurotrophic cone electrode developed by Kennedy was a pioneering example of a biohybrid central neural prosthesis: a section of nerve housed within a glass electrode assembly was used to encourage neurite ingrowth near the gold recording wire.
“The neurotrophic cone electrode developed by Kennedy was a pioneering example of a biohybrid central neural prosthesis: a section of nerve housed within a glass electrode assembly was used to encourage neurite ingrowth near the gold recording wire.82”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [82] 82. Kennedy, P.R. The cone electrode: a long-term electrode that records from neurites grown onto its recording surface. J Neurosci Methods 29, 181, 1989. source
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Most commonly the nerve lesion for regenerative implants is a full nerve transection where the two ends are cleanly separated and a gap forms.
“Most commonly this is in the form of a full nerve transection, where the two ends of the nerve are cleanly separated and a gap forms between the two (Navarro et al., 2005; Thompson et al., 2015).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Thompson et al., 2015
The Biohybrid devices section (page unavailable) describes a regenerative interface attached to a transected nerve, and the Fig. 2 caption describes grafting muscle to the end of a divided or severed nerve. Neither passage says that full transection is the most common lesion for regenerative implants or that it leaves two cleanly separated ends with a gap. (LLM, conf=0.84)
Experimental — accuracy not yet measuredCuff electrodes can be inserted into the gap of a nerve lesion and the nerve allowed to grow through it, rather than wrapping the cuff around a healthy nerve.
“Cuff electrodes can be inserted into the gap of a nerve lesion and the nerve allowed to grow through it, rather than wrapping the cuff around a healthy nerve (Lotfi et al., 2011; Stoyanova et al., 2013; Thompson et al., 2015).”
From Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics (2021) · cites this paper as Thompson et al., 2015
The Peripheral Regenerative Nerve Interfaces section says cuff electrodes wrap around the exterior of a nerve. The Electrode architecture section describes axons regenerating through sieve electrodes, but does not say that a cuff electrode can be placed in a lesion gap for the nerve to grow through. Page numbers are unavailable in our copy. (LLM, conf=0.85)
Experimental — accuracy not yet measuredMicrochannel structure maximises the amplitude of recorded biopotentials and reduces the currents required to activate nerves without reliance on the position of the nodes of Ranvier in myelinated fibers.
“This maximises the amplitude of the recorded biopotentials and reduces the currents required to activate nerves without reliance on the position of the nodes of Ranvier in myelinated fibers [23], [35].”
From Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review (2024) · cites this paper as [23]
The Electrode architecture section (page unavailable) says insulated microchannels may boost recorded signals by reducing shunt loss. It does not establish that they maximise biopotential amplitude, reduce the current needed to activate nerves, or make activation independent of node-of-Ranvier position. (LLM, conf=0.85)
Experimental — accuracy not yet measured
Cites
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Hochberg, L. R., Bacher, D., Jarosiewicz, B., Masse, N. Y., Simeral, J. D., Vogel, J., Haddadin, S., Liu, J., Cash, S. S., van der Smagt, P., et al. (2012) · Nature · doi:10.1038/nature11076 · Cited by 2,719 · see citing works
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · see citing works
Conducting polymers in biomedical engineering
Guimard, N. K., Gomez, N., & Schmidt, C. E. (2007) · Progress in Polymer Science · doi:10.1016/j.progpolymsci.2007.05.012 · Cited by 1,599 · see citing works
Responsive cortical stimulation for the treatment of medically intractable partial epilepsy
Morrell, M. J. (2011) · Neurology · doi:10.1212/wnl.0b013e3182302056 · Cited by 1,282 · see citing works
Restoring natural sensory feedback in real-time bidirectional hand prostheses
Raspopovic, S., Capogrosso, M., Petrini, F. M., Bonizzato, M., Rigosa, J., Di Pino, G., Carpaneto, J., Controzzi, M., Boretius, T., Fernandez, E., et al. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3006820 · Cited by 1,001 · see citing works
Multi-electrode array technologies for neuroscience and cardiology
Spira, M. E., & Hai, A. (2013) · Nature Nanotechnology · doi:10.1038/nnano.2012.265 · Cited by 991 · see citing works
Intracellular Features Predicted by Extracellular Recordings in the Hippocampus In Vivo
Henze, D. A., Borhegyi, Z., Csicsvari, J., Mamiya, A., Harris, K. D., & Buzsáki, G. (2000) · Journal of Neurophysiology · doi:10.1152/jn.2000.84.1.390 · Cited by 942 · see citing works
Biran, R., Martin, D. C., & Tresco, P. A. (2005) · Experimental Neurology · doi:10.1016/j.expneurol.2005.04.020 · Cited by 893 · see citing works
Cochlear Implants: System Design, Integration, and Evaluation
Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., & Feng, H. (2008) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2008.2008250 · Cited by 878 · see citing works
Conducting‐Polymer Nanotubes for Controlled Drug Release
Abidian, M. R., Kim, D., & Martin, D. C. (2006) · Advanced Materials · doi:10.1002/adma.200501726 · Cited by 869 · see citing works
Closed-Loop Deep Brain Stimulation Is Superior in Ameliorating Parkinsonism
Rosin, B., Slovik, M., Mitelman, R., Rivlin-Etzion, M., Haber, S. N., Israel, Z., Vaadia, E., & Bergman, H. (2011) · Neuron · doi:10.1016/j.neuron.2011.08.023 · Cited by 793 · see citing works
A neural interface provides long-term stable natural touch perception
Tan, D. W., Schiefer, M. A., Keith, M. W., Anderson, J. R., Tyler, J., & Tyler, D. J. (2014) · Science Translational Medicine · doi:10.1126/scitranslmed.3008669 · Cited by 751 · see citing works
Sondell, M., Lundborg, G., & Kanje, M. (1999) · The Journal of Neuroscience · doi:10.1523/jneurosci.19-14-05731.1999 · Cited by 750 · see citing works
Polypyrrole-coated electrospun PLGA nanofibers for neural tissue applications
Lee, J. Y., Bashur, C. A., Goldstein, A. S., & Schmidt, C. E. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2009.04.042 · Cited by 735 · see citing works
Kozai, T. D. Y., Langhals, N. B., Patel, P. R., Deng, X., Zhang, H., Smith, K. L., Lahann, J., Kotov, N. A., & Kipke, D. R. (2012) · Nature Materials · doi:10.1038/nmat3468 · Cited by 721 · see citing works
Self-citationCarbon nanotube coating improves neuronal recordings
Keefer, E. W., Botterman, B. R., Romero, M. I., Rossi, A. F., & Gross, G. W. (2008) · Nature Nanotechnology · doi:10.1038/nnano.2008.174 · Cited by 721 · see citing works
Revealing neuronal function through microelectrode array recordings
Obien, M. E. J., Deligkaris, K., Bullmann, T., Bakkum, D. J., & Frey, U. (2015) · Frontiers in Neuroscience · doi:10.3389/fnins.2014.00423 · Cited by 705 · see citing works
Conducting polymers for neural interfaces: Challenges in developing an effective long-term implant
Green, R. A., Lovell, N. H., Wallace, G. G., & Poole-Warren, L. A. (2008) · Biomaterials · doi:10.1016/j.biomaterials.2008.04.047 · Cited by 691 · see citing works
Ludwig, K. A., Uram, J. D., Yang, J., Martin, D. C., & Kipke, D. R. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/1/007 · Cited by 638 · see citing works
Brain Tissue Responses to Neural Implants Impact Signal Sensitivity and Intervention Strategies
Kozai, T. D. Y., Jaquins-Gerstl, A. S., Vazquez, A. L., Michael, A. C., & Cui, X. T. (2015) · ACS Chemical Neuroscience · doi:10.1021/cn500256e · Cited by 622 · see citing works
McConnell, G. C., Rees, H. D., Levey, A. I., Gutekunst, C.-A., Gross, R. E., & Bellamkonda, R. V. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/5/056003 · Cited by 480 · see citing works
Neural probe design for reduced tissue encapsulation in CNS
Seymour, J. P., & Kipke, D. R. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.024 · Cited by 462 · see citing works
Abidian, M. R., Corey, J. M., Kipke, D. R., & Martin, D. C. (2010) · Small · doi:10.1002/smll.200901868 · Cited by 405 · see citing works
Double nerve intraneural interface implant on a human amputee for robotic hand control
Rossini, P. M., Micera, S., Benvenuto, A., Carpaneto, J., Cavallo, G., Citi, L., Cipriani, C., Denaro, L., Denaro, V., Di Pino, G., et al. (2010) · Clinical Neurophysiology · doi:10.1016/j.clinph.2010.01.001 · Cited by 396 · see citing works
Chestek, C. A., Gilja, V., Nuyujukian, P., Foster, J. D., Fan, J. M., Kaufman, M. T., Churchland, M. M., Rivera-Alvidrez, Z., Cunningham, J. P., Ryu, S. I., et al. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/4/045005 · Cited by 350 · see citing works
Gomez, N., & Schmidt, C. E. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.31047 · Cited by 306 · see citing works
Polypyrrole-coated electrodes for the delivery of charge and neurotrophins to cochlear neurons
Richardson, R. T., Wise, A. K., Thompson, B. C., Flynn, B. O., Atkinson, P. J., Fretwell, N. J., Fallon, J. B., Wallace, G. G., Shepherd, R. K., Clark, G. M., et al. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2009.01.015 · Cited by 304 · see citing works
Kim, D. ‐., Richardson‐Burns, S. M., Hendricks, J. L., Sequera, C., & Martin, D. C. (2007) · Advanced Functional Materials · doi:10.1002/adfm.200500594 · Cited by 299 · see citing works
Prasad, A., Xue, Q.-S., Sankar, V., Nishida, T., Shaw, G., Streit, W. J., & Sanchez, J. C. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/5/056015 · Cited by 298 · see citing works
Ludwig, K. A., Langhals, N. B., Joseph, M. D., Richardson-Burns, S. M., Hendricks, J. L., & Kipke, D. R. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/1/014001 · Cited by 283 · see citing works
Self-citationThe L1 Family of Neural Cell Adhesion Molecules: Old Proteins Performing New Tricks
Hortsch, M. (1996) · Neuron · doi:10.1016/s0896-6273(00)80192-0 · Cited by 259 · see citing works
Cell attachment functionality of bioactive conducting polymers for neural interfaces
Green, R. A., Lovell, N. H., & Poole-Warren, L. A. (2009) · Biomaterials · doi:10.1016/j.biomaterials.2009.03.043 · Cited by 258 · see citing works
Richardson, R. T., Thompson, B., Moulton, S., Newbold, C., Lum, M. G., Cameron, A., Wallace, G., Kapsa, R., Clark, G., & O’Leary, S. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2006.09.008 · Cited by 251 · see citing works
Protection of auditory neurons from aminoglycoside toxicity by neurotrophin-3
Ernfors, P., Li Duan, M., Elshamy, W. M., & Canlon, B. (1996) · Nature Medicine · doi:10.1038/nm0496-463 · Cited by 245 · see citing works
Perge, J. A., Homer, M. L., Malik, W. Q., Cash, S., Eskandar, E., Friehs, G., Donoghue, J. P., & Hochberg, L. R. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/3/036004 · Cited by 243 · see citing works
Relationship between intracortical electrode design and chronic recording function
Karumbaiah, L., Saxena, T., Carlson, D., Patil, K., Patkar, R., Gaupp, E. A., Betancur, M., Stanley, G. B., Carin, L., & Bellamkonda, R. V. (2013) · Biomaterials · doi:10.1016/j.biomaterials.2013.07.016 · Cited by 241 · see citing works
Kozai, T. D. Y., Gugel, Z., Li, X., Gilgunn, P. J., Khilwani, R., Ozdoganlar, O. B., Fedder, G. K., Weber, D. J., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.07.039 · Cited by 193 · see citing works
Electrodeposited polypyrrole/carbon nanotubes composite films electrodes for neural interfaces
Lu, Y., Li, T., Zhao, X., Li, M., Cao, Y., Yang, H., & Duan, Y. Y. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.03.022 · Cited by 191 · see citing works
Improving peripheral nerve regeneration: From molecular mechanisms to potential therapeutic targets
Chan, K. M., Gordon, T., Zochodne, D. W., & Power, H. A. (2014) · Experimental Neurology · doi:10.1016/j.expneurol.2014.09.006 · Cited by 189 · see citing works
Weiland, J. D., & Humayun, M. S. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2314733 · Cited by 186 · see citing works
Conductive Single-Walled Carbon Nanotube Substrates Modulate Neuronal Growth
Malarkey, E. B., Fisher, K. A., Bekyarova, E. B., Liu, W., Haddon, R. C., & Parpura, V. (2008) · Nano Letters · doi:10.1021/nl802855c · Cited by 186 · see citing works
Kung, T. A., Langhals, N. B., Martin, D. C., Johnson, P. J., Cederna, P. S., & Urbanchek, M. G. (2014) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0000000000000168 · Cited by 181 · see citing works
Self-citationTranslational strategies and challenges in regenerative medicine
Dimmeler, S., Ding, S., Rando, T. A., & Trounson, A. (2014) · Nature Medicine · doi:10.1038/nm.3627 · Cited by 179 · see citing works
Abiotic-biotic characterization of Pt/Ir microelectrode arrays in chronic implants
Prasad, A., Xue, Q.-S., Dieme, R., Sankar, V., Mayrand, R. C., Nishida, T., Streit, W. J., & Sanchez, J. C. (2014) · Frontiers in Neuroengineering · doi:10.3389/fneng.2014.00002 · Cited by 178 · see citing works
Lago, N., Ceballos, D., J Rodrı́guez, F., Stieglitz, T., & Navarro, X. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.06.025 · Cited by 178 · see citing works
He, W., McConnell, G. C., & Bellamkonda, R. V. (2006) · Journal of Neural Engineering · doi:10.1088/1741-2560/3/4/009 · Cited by 177 · see citing works
Winslow, B. D., Christensen, M. B., Yang, W.-K., Solzbacher, F., & Tresco, P. A. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.05.050 · Cited by 171 · see citing works
Borton, D., Micera, S., Millán, J. D. R., & Courtine, G. (2013) · Science Translational Medicine · doi:10.1126/scitranslmed.3005968 · Cited by 171 · see citing works
Green, R. A., Lovell, N. H., & Poole-Warren, L. A. (2010) · Acta Biomaterialia · doi:10.1016/j.actbio.2009.06.030 · Cited by 170 · see citing works
Azemi, E., Lagenaur, C. F., & Cui, X. T. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2010.09.033 · Cited by 164 · see citing works
Cellot, G., Toma, F. M., Kasap Varley, Z., Laishram, J., Villari, A., Quintana, M., Cipollone, S., Prato, M., & Ballerini, L. (2011) · The Journal of Neuroscience · doi:10.1523/jneurosci.1332-11.2011 · Cited by 158 · see citing works
Nanoscale neuro-integrative coatings for neural implants
He, W., & Bellamkonda, R. V. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2004.08.021 · Cited by 156 · see citing works
Kennedy, P. R. (1989) · Journal of Neuroscience Methods · doi:10.1016/0165-0270(89)90142-8 · Cited by 151 · see citing works
A micromachined silicon sieve electrode for nerve regeneration applications
Akin, T., Najafi, K., Smoke, R. H., & Bradley, R. M. (1994) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.284958 · Cited by 144 · see citing works
Kozai, T. D. Y., Li, X., Bodily, L. M., Caparosa, E. M., Zenonos, G. A., Carlisle, D. L., Friedlander, R. M., & Cui, X. T. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.08.006 · Cited by 136 · see citing works
The foreign body response to the Utah Slant Electrode Array in the cat sciatic nerve
Christensen, M. B., Pearce, S. M., Ledbetter, N. M., Warren, D. J., Clark, G. A., & Tresco, P. A. (2014) · Acta Biomaterialia · doi:10.1016/j.actbio.2014.07.010 · Cited by 135 · see citing works
The Role of Schwann Cell-Axon Interaction in Peripheral Nerve Regeneration
Namgung, U. (2014) · Cells Tissues Organs · doi:10.1159/000370324 · Cited by 135 · see citing works
Kolarcik, C. L., Catt, K., Rost, E., Albrecht, I. N., Bourbeau, D., Du, Z., Kozai, T. D. Y., Luo, X., Weber, D. J., & Tracy Cui, X. (2015) · Journal of Neural Engineering · doi:10.1088/1741-2560/12/1/016008 · Cited by 126 · see citing works
Skousen, J. L., Merriam, S. M. E., Srivannavit, O., Perlin, G., Wise, K. D., & Tresco, P. A. (2011) · Progress in Brain Research · doi:10.1016/b978-0-444-53815-4.00009-1 · Cited by 111 · see citing works
Electrical Stimulation of Schwann Cells Promotes Sustained Increases in Neurite Outgrowth
Koppes, A. N., Nordberg, A. L., Paolillo, G. M., Goodsell, N. M., Darwish, H. A., Zhang, L., & Thompson, D. M. (2014) · Tissue Engineering Part A · doi:10.1089/ten.tea.2013.0012 · Cited by 108 · see citing works
Neurotrophic electrode: Method of assembly and implantation into human motor speech cortex
Bartels, J., Andreasen, D., Ehirim, P., Mao, H., Seibert, S., Wright, E. J., & Kennedy, P. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.06.030 · Cited by 108 · see citing works
Interfaces with the Peripheral Nerve for the Control of Neuroprostheses
del Valle, J., & Navarro, X. (2013) · International Review of Neurobiology · doi:10.1016/b978-0-12-420045-6.00002-x · Cited by 105 · see citing works
Azemi, E., Stauffer, W. R., Gostock, M. S., Lagenaur, C. F., & Cui, X. T. (2008) · Acta Biomaterialia · doi:10.1016/j.actbio.2008.02.028 · Cited by 97 · see citing works
Novel multi-sided, microelectrode arrays for implantable neural applications
Seymour, J. P., Langhals, N. B., Anderson, D. J., & Kipke, D. R. (2011) · Biomedical Microdevices · doi:10.1007/s10544-011-9512-z · Cited by 95 · see citing works
Self-citationAlginate Composition Effects on a Neural Stem Cell–Seeded Scaffold
Purcell, E. K., Singh, A., & Kipke, D. R. (2009) · Tissue Engineering Part C: Methods · doi:10.1089/ten.tec.2008.0302 · Cited by 95 · see citing works
Self-citationMannard, A., Stein, R. B., & Charles, D. (1974) · Science · doi:10.1126/science.183.4124.547 · Cited by 85 · see citing works
The cone electrode: Ultrastructural studies following long-term recording in rat and monkey cortex
Kennedy, P. R., Mirra, S. S., & Bakay, R. A. E. (1992) · Neuroscience Letters · doi:10.1016/0304-3940(92)90627-j · Cited by 81 · see citing works
Clark, G. A., Wendelken, S., Page, D. M., Davis, T., Wark, H. A. C., Normann, R. A., Warren, D. J., & Hutchinson, D. T. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6944001 · Cited by 78 · see citing works
Stieglitz, T., Ruf, H. H., Gross, M., Schüettler, M., & Meyer, J.-U. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(02)00019-2 · Cited by 77 · see citing works
<i>In vivo</i> evaluation of a neural stem cell-seeded prosthesis
Purcell, E. K., Seymour, J. P., Yandamuri, S., & Kipke, D. R. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/2/026005 · Cited by 76 · see citing works
Self-citationMicrochannels as Axonal Amplifiers
FitzGerald, J. J., Lacour, S. P., McMahon, S. B., & Fawcett, J. W. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.909533 · Cited by 76 · see citing works
Peripheral nerve repair with cultured schwann cells: getting closer to the clinics
Rodrigues, M. C. O., Rodrigues, A. A., Glover, L. E., Voltarelli, J., & Borlongan, C. V. (2012) · The Scientific World Journal · doi:10.1100/2012/413091 · Cited by 75 · see citing works
Long Micro-Channel Electrode Arrays: A Novel Type of Regenerative Peripheral Nerve Interface
Lacour, S. P., Fitzgerald, J. J., Lago, N., Tarte, E., McMahon, S., & Fawcett, J. (2009) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2031241 · Cited by 75 · see citing works
Lago, N., Udina, E., Ramachandran, A., & Navarro, X. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.891168 · Cited by 68 · see citing works
Stabilizing electrode-host interfaces: a tissue engineering approach.
(2001) · PubMed · Cited by 68 · see citing works
Microchannel-based regenerative scaffold for chronic peripheral nerve interfacing in amputees
Srinivasan, A., Tahilramani, M., Bentley, J. T., Gore, R. K., Millard, D. C., Mukhatyar, V. J., Joseph, A., Haque, A. S., Stanley, G. B., English, A. W., et al. (2015) · Biomaterials · doi:10.1016/j.biomaterials.2014.11.035 · Cited by 67 · see citing works
Castro, J., Negredo, P., & Avendaño, C. (2008) · Brain Research · doi:10.1016/j.brainres.2007.11.028 · Cited by 63 · see citing works
Flavopiridol reduces the impedance of neural prostheses in vivo without affecting recording quality
Purcell, E. K., Thompson, D. E., Ludwig, K. A., & Kipke, D. R. (2009) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2009.06.026 · Cited by 62 · see citing works
Self-citationSkousen, J. L., Bridge, M. J., & Tresco, P. A. (2015) · Biomaterials · doi:10.1016/j.biomaterials.2014.08.039 · Cited by 61 · see citing works
Kolarcik, C. L., Bourbeau, D., Azemi, E., Rost, E., Zhang, L., Lagenaur, C. F., Weber, D. J., & Cui, X. T. (2012) · Acta Biomaterialia · doi:10.1016/j.actbio.2012.06.034 · Cited by 59 · see citing works
Histopathology of human cochlear implants: Correlation of psychophysical and anatomical measures
Khan, A. M., Whiten, D. M., Nadol, J. B., & Eddington, D. K. (2005) · Hearing Research · doi:10.1016/j.heares.2005.03.003 · Cited by 59 · see citing works
Polyimide micro-channel arrays for peripheral nerve regenerative implants
Lacour, S. P., Atta, R., FitzGerald, J. J., Blamire, M., Tarte, E., & Fawcett, J. (2008) · Sensors and Actuators A: Physical · doi:10.1016/j.sna.2008.05.031 · Cited by 56 · see citing works
Behan, B. L., DeWitt, D. G., Bogdanowicz, D. R., Koppes, A. N., Bale, S. S., & Thompson, D. M. (2011) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32939 · Cited by 56 · see citing works
Tools for Resolving Functional Activity and Connectivity within Intact Neural Circuits
Jennings, J. H., & Stuber, G. D. (2014) · Current Biology · doi:10.1016/j.cub.2013.11.042 · Cited by 54 · see citing works
Seeding neural progenitor cells on silicon-based neural probes: Laboratory investigation
Azemi, E., Gobbel, G. T., & Cui, X. T. (2010) · Journal of Neurosurgery · doi:10.3171/2010.1.jns09313 · Cited by 47 · see citing works
Layer‐by‐layer assembly of polyelectrolyte films improving cytocompatibility to neural cells
Wu, Z., Ma, J., Liu, B., Xu, Q., & Cui, F. (2007) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.30993 · Cited by 43 · see citing works
Modality-Specific Axonal Regeneration: Toward Selective Regenerative Neural Interfaces
Lotfi, P., Garde, K., Chouhan, A. K., Bengali, E., & Romero-Ortega, M. I. (2011) · Frontiers in Neuroengineering · doi:10.3389/fneng.2011.00011 · Cited by 42 · see citing works
Update in facial nerve paralysis: tissue engineering and new technologies
Langhals, N. B., Urbanchek, M. G., Ray, A., & Brenner, M. J. (2014) · Current Opinion in Otolaryngology & Head and Neck Surgery · doi:10.1097/moo.0000000000000062 · Cited by 41 · see citing works
Self-citationMinev, I. R., Chew, D. J., Delivopoulos, E., Fawcett, J. W., & Lacour, S. P. (2012) · Journal of Neural Engineering · doi:10.1088/1741-2560/9/2/026005 · Cited by 36 · see citing works
Activity of single action potentials in monkey motor cortex during long-term task learning
Kennedy, P. R., & Bakay, R. A. E. (1997) · Brain Research · doi:10.1016/s0006-8993(97)00051-6 · Cited by 34 · see citing works
Electrically stimulated signals from a long-term Regenerative Peripheral Nerve Interface
Langhals, N. B., Woo, S. L., Moon, J. D., Larson, J. V., Leach, M. K., Cederna, P. S., & Urbanchek, M. G. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6944004 · Cited by 30 · see citing works
Self-citationCombining Topographical and Genetic Cues to Promote Neuronal Fate Specification in Stem Cells
Purcell, E. K., Naim, Y., Yang, A., Leach, M. K., Velkey, J. M., Duncan, R. K., & Corey, J. M. (2012) · Biomacromolecules · doi:10.1021/bm301220k · Cited by 26 · see citing works
Self-citationA micromachined silicon sieve electrode for nerve regeneration applications
Akın, T., & Najafi, K. (2002) · International Conference on Solid-State Sensors, Actuators and Microsystems · doi:10.1109/sensor.1991.148818 · Cited by 19 · see citing works
Stoyanova, I. I., van Wezel, R. J. A., & Rutten, W. L. C. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/6/066018 · Cited by 18 · see citing works
Regenerative microchannel electrode array for peripheral nerve interfacing
Srinivasan, A., Guo, L., & Bellamkonda, R. V. (2011) · 2011 5th International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2011.5910535 · Cited by 12 · see citing works
In vivoevaluation of a neural stem cell-seeded prosthesis
Purcell, E. K., Seymour, J. P., Yandamuri, S. S., & Kipke, D. R. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2552/6/4/049801 · Cited by 9 · see citing works
Self-citationRegenerative electrodes for interfacing injured peripheral nerves: neurobiological assessment
Lago, N., Udina, E., & Navarro, X. (2006) · IEEE/RAS-EMBS International Conference on Biomedical Robotics and Biomechatronics (BioRob) · doi:10.1109/biorob.2006.1639247 · Cited by 7 · see citing works
Reliable neural interface: The first quarter century of the Neurotrophic Electrode
Kennedy, P. (2012) · 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2012.6346678 · Cited by 4 · see citing works
Weiland, J. D., & Humayun, M. S. (2012) · Springer eBooks · doi:10.1007/978-1-4614-5227-0_15 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Foreign Body Reaction to Implanted Biomaterials and Its Impact in Nerve Neuroprosthetics
Carnicer-Lombarte, A., Chen, S.-T., Malliaras, G. G., & Barone, D. G. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.622524 · Cited by 454 · see citing works
“Most commonly this is in the form of a full nerve transection, where the two ends of the nerve are cleanly separated and a gap forms between the two (Navarro et al., 2005; Thompson et al., 2015).”
BackgroundA Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389 · see citing works
In Vivo Organic Bioelectronics for Neuromodulation
Magnus Berggren, A., Głowacki, E. D., Simon, D. T., Stavrinidou, E., & Tybrandt, K. (2022) · Chemical Reviews · doi:10.1021/acs.chemrev.1c00390 · Cited by 123 · see citing works
Interfaces with the peripheral nervous system for the control of a neuroprosthetic limb: a review
Yildiz, K. A., Shin, A. Y., & Kaufman, K. R. (2020) · Journal of NeuroEngineering and Rehabilitation · doi:10.1186/s12984-020-00667-5 · Cited by 113 · see citing works
Invasive Intraneural Interfaces: Foreign Body Reaction Issues
Lotti, F., Ranieri, F., Vadalà, G., Zollo, L., & Pino, G. D. (2017) · Frontiers in Neuroscience · doi:10.3389/fnins.2017.00497 · Cited by 117 · see citing works
“…factors (Korecka et al., 2007), such as nerve growth factor, glial derived neurotrophic factor, insulin growth factor (Fine et al., 2002; Apel et al., 2010; Madduri et al., 2010; Yan et al., 2012), that mimic the microenvironment, helping cells to restore the damaged nerves (Thompson et al., 2016).”
BackgroundMaria, F. D., Lodola, F., Zucchetti, E., Benfenati, F., & Lanzani, G. (2018) · Chemical Society Reviews · doi:10.1039/c7cs00860k · Cited by 95 · see citing works
Update on Peripheral Nerve Electrodes for Closed-Loop Neuroprosthetics
Rijnbeek, E. H., Eleveld, N., & Olthuis, W. (2018) · Frontiers in Neuroscience · doi:10.3389/fnins.2018.00350 · Cited by 91 · see citing works
“The nerve then regenerates through the pores, after which APs can be evoked or recorded using the conductive material inside the pores (Thompson et al., 2015).”
BackgroundYoung, A. T., Cornwell, N., & Daniele, M. A.-A. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201700239 · Cited by 70 · see citing works
Advances in Carbon-Based Microfiber Electrodes for Neural Interfacing
Hejazi, M., Tong, W., Ibbotson, M. R., Prawer, S., & Garrett, D. J. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.658703 · Cited by 68 · see citing works
“One example is to temporarily improve the electrode stiffness by coating the flexible fibers with materials such as silk and sucrose, which dissolve into the surrounding tissue after electrode insertion (Tien et al., 2013; Apollo et al., 2015).”
BackgroundHejazi, M. A., Tong, W., Stacey, A. D., Soto-Breceda, A., Ibbotson, M. R., Yunzab, M., Maturana, M. I., Almasi, A., Jung, Y. J., Sun, S. H., et al. (2019) · Biomaterials · doi:10.1016/j.biomaterials.2019.119648 · Cited by 65 · see citing works
Cutrone, A., & Micera, S. (2019) · Advanced Healthcare Materials · doi:10.1002/adhm.201801345 · Cited by 57 · see citing works
Spatiotemporal patterns of gene expression around implanted silicon electrode arrays
Thompson, C. H., Saxena, A., Heelan, N., Salatino, J. W., & Purcell, E. K. (2021) · Journal of Neural Engineering · doi:10.1088/1741-2552/abf2e6 · Cited by 47 · see citing works
“These observations, in combination with other early reports [21, 25, 26], provided motivation for the design of next-generation neural interfaces with improved biological integration [18, 27, 28].”
ResultSelf-citationBeyond Tissue replacement: The Emerging role of smart implants in healthcare
Abyzova, E., Dogadina, E., Rodriguez, R. D. D., Petrov, I., Kolesnikova, Y., Zhou, M., Liu, C., & Sheremet, E. S. (2023) · Materials Today Bio · doi:10.1016/j.mtbio.2023.100784 · Cited by 56 · see citing works
“For example, electrodes serve as an active element in neural interfaces by applying a potential or current to tissues [41], while electrochemical electrodes act as tissue-electronics interfaces in monitoring [42], and force and pressure sensors measure mechanical loads [40,43].”
BackgroundThompson, C. H., Riggins, T. E., Patel, P. R., Chestek, C. A., Li, W., & Purcell, E. K. (2020) · Journal of Neural Engineering · doi:10.1088/1741-2552/ab7030 · Cited by 38 · see citing works
“Likewise, modifying electrodes with polymer coatings, biologics, topographical cues [137, 148] and related approaches often remains within the purview of the specific lab originating the technique.”
BackgroundSelf-citationAurand, E. R., Usmani, S., Medelin, M., Scaini, D., Bosi, S., Rosselli, F. B., Donato, S., Tromba, G., Prato, M., & Ballerini, L. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201700550 · Cited by 35 · see citing works
Biohybrid neural interfaces: improving the biological integration of neural implants
Boulingre, M., Lara, R. P., & Green, R. A. (2023) · Chemical Communications · doi:10.1039/d3cc05006h · Cited by 39 · see citing works
Gołda‐Cępa, M., Chorylek, A., Chytrosz-Wróbel, P., Brzychczy‐Włoch, M., Jaworska, J., Kasperczyk, J., Hakkarainen, M., Engvall, K., & Kotarba, A. (2016) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.6b08025 · Cited by 38 · see citing works
Fascicular nerve stimulation and recording using a novel double-aisle regenerative electrode
Delgado-Martínez, I., Righi, M., Santos, D., Cutrone, A., Bossi, S., D’Amico, S., Del Valle, J., Micera, S., & Navarro, X. (2017) · Journal of Neural Engineering · doi:10.1088/1741-2552/aa6bac · Cited by 35 · see citing works
Kim, D., Park, D., Kim, T. H., Chung, J. J., Jung, Y., & Kim, S. H. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100107 · Cited by 30 · see citing works
Purushothaman, A. E., Thakur, K., & Kandasubramanian, B. (2019) · International Journal of Polymeric Materials · doi:10.1080/00914037.2019.1581197 · Cited by 33 · see citing works
Regenerative Peripheral Nerve Interfaces for Advanced Control of Upper Extremity Prosthetic Devices
Kumar, N. G., Kung, T. A., & Cederna, P. S. (2021) · Hand Clinics · doi:10.1016/j.hcl.2021.04.005 · Cited by 25 · see citing works
Strategies for neural control of prosthetic limbs: from electrode interfacing to 3D printing
Ngan, C., Kapsa, R. M. I., & Choong, P. (2019) · Materials · doi:10.3390/ma12121927 · Cited by 24 · see citing works
“These systems can be described as sieves or hollow electrode array systems, which are placed between two ends of a severed nerve [35].”
BackgroundAlterations in Ion Channel Expression Surrounding Implanted Microelectrode Arrays in the BrainPreprint
Salatino, J. W., Kale, A. P., & Purcell, E. K. (2019) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/518811 · Cited by 16 · see citing works
Self-citationKim, H., Dingle, A. M., Ness, J. P., Baek, D.-H., Bong, J., Lee, I. K., Shulzhenko, N. O., Zeng, W., Israel, J. S., Pisaniello, J. A., et al. (2020) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2020.108602 · Cited by 19 · see citing works
The Pursuit of Chronically Reliable Neural Interfaces: A Materials Perspective
Guo, L. (2016) · Frontiers in Neuroscience · doi:10.3389/fnins.2016.00599 · Cited by 19 · see citing works
“…neural interfaces is being revived, combining state-of-the-art stretchable electronics with tissue engineering approaches for a better-integrated electrode-tissue interface (Figure 1D; Lacour et al., 2010; Clements et al., 2013; Musick et al., 2015; Srinivasan et al., 2015; Thompson et al., 2016).”
BackgroundInfluentialA model of electrical impedance tomography implemented in nerve-cuff for neural-prosthetics control
Hope, J., Vanholsbeeck, F., & McDaid, A. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aab73a · Cited by 16 · see citing works
An ASIC for Recording and Stimulation in Stacked Microchannel Neural Interfaces
Lancashire, H. T., Jiang, D., Demosthenous, A., & Donaldson, N. (2019) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2019.2891284 · Cited by 15 · see citing works
“MNIs achieve this with microchannels (small diameter tubes) through which axons are placed [21] or regenerate [22].”
BackgroundStankova, N. E., Nikolov, A. S., Iordanova, E., et al. (2021) · Polymers · doi:10.3390/polym13173004 · Cited by 12 · see citing works
Active Neural Interface Circuits and Systems for Selective Control of Peripheral Nerves: A Review
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2024) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2024.3430038 · Cited by 11 · see citing works
“…is made above the outer layer of the nerve (epineurium), (2) intra-fascicular or intra-neural, which tap into the nerve fascicles with penetrative structures [20], [21], and (3) regenerative, where the interface architecture requires axonal growth through fine-scale device geometries [22], [23].”
Spatiotemporal expression of RNA-seq identified proteins at the electrode interface
Thompson, C. H., Evans, B. M., Zhao, D. X., & Purcell, E. K. (2023) · Acta Biomaterialia · doi:10.1016/j.actbio.2023.04.028 · Cited by 10 · see citing works
Self-citationKrutko, M., Poling, H. M., Sheth, M., Kongsomros, S., Bryan, A. E., Sharma, M., Singh, A., Reza, H. A., Wikenheiser‐Brokamp, K. A., Takebe, T., et al. (2025) · Advanced Materials Technologies · doi:10.1002/admt.202401513 · Cited by 12 · see citing works
Veith, A. P., Li, X., Modi, H., Abbaspour, A. R., Luan, L., Xie, C., & Baker, A. B. (2021) · Biomaterials · doi:10.1016/j.biomaterials.2021.120924 · Cited by 9 · see citing works
Mechanism of peripheral nerve modulation and recent applications
Shin, H., Kang, M., & Lee, S. (2021) · International Journal of Optomechatronics · doi:10.1080/15599612.2021.1978601 · Cited by 9 · see citing works
“Connecting the cut planes of the nerve with the electrode, the axons regenerate through the pores that have a metal contact layer, which makes higher selectivity compared with other types of electrodes.([41]) Since the electrode at the invasive method is adjacent to the axon, it is possible to exceed the threshold val…”
BackgroundNeuroethical considerations of high-density electrode arrays
Cabrera, L. Y., Sadle, C. J., & Purcell, E. K. (2019) · Nature Biomedical Engineering · doi:10.1038/s41551-019-0407-2 · Cited by 10 · see citing works
Self-citationAn Active Microchannel Neural Interface for Implantable Electrical Stimulation and Recording
Habibollahi, M., Jiang, D., Lancashire, H. T., & Demosthenous, A. (2025) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2025.3533612 · Cited by 8 · see citing works
“Regenerative interfaces such as sieve electrodes, double-aisle electrodes, and microchannel nerve interfaces (MNIs) can achieve very high selectivity following regeneration of transected nerve sections through the device [26], [27].”
Wu, Y., Zhang, Q., Wang, H., & Wang, M. (2020) · Materials Chemistry Frontiers · doi:10.1039/d0qm00279h · Cited by 9 · see citing works
Designification of Neurotechnological Devices through 3D Printed Functional Materials
Castro, N. J., & Hutmacher, D. W. (2017) · Advanced Functional Materials · doi:10.1002/adfm.201703905 · Cited by 8 · see citing works
Mandeville, R. M., Sanchez, B., Johnston, B. R., Bazarek, S. F., Thum, J. A., Birmingham, A., See, R. B., Leochico, C. F. D., Kumar, V., Dowlatshahi, A. S., et al. (2023) · Journal of Neural Engineering · doi:10.1088/1741-2552/acdbeb · Cited by 7 · see citing works
“[227] Thompson C H, Zoratti M J, Langhals N B and Purcell E K”
Background7.32 Engineering the Neural Interface
Kelly, A., Ballerini, L., Lowery, M. M., & Biggs, M. J. P. (2017) · Elsevier eBooks · doi:10.1016/b978-0-12-803581-8.09322-x · Cited by 7 · see citing works
Bonato, M., Galletta, V., Chiaramello, E., et al. (2025) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2025.108862 · Cited by 6 · see citing works
Krutko, M., Poling, H. M., Bryan, A. E., Sharma, M., Singh, A., Reza, H. A., Wikenheiser‐Brokamp, K. A., Takebe, T., Helmrath, M. A., Harris, G. M., et al. (2024) · bioRxiv (Cold Spring Harbor Laboratory) · doi:10.1101/2024.08.16.608345 · Cited by 7 · see citing works
Improved wetting of gold active braze alloy on diamond for use in medical implants
Edalati, K., Stamp, M. E. M., Ganesan, K., Stacey, A. D., Martin-Hardy, G., Fontaine, R., Prawer, S., & Garrett, D. J. (2020) · Diamond and Related Materials · doi:10.1016/j.diamond.2020.108089 · Cited by 6 · see citing works
Gunapu, D. V. S. K., Mudigunda, S. V., Das, A., Rengan, A. K., & Vanjari, S. R. K. (2020) · Journal of Applied Electrochemistry · doi:10.1007/s10800-020-01447-8 · Cited by 6 · see citing works
“Functional Electrode stimulation (FES) garnered a wide range of attention amongst the neuroprosthesis community over the past decade owing to its ability to mimic neural behaviour, thereby acting as a potential replacement for neurons that lost their functionality due to underlying disease conditions [1].”
BackgroundInterfacing peripheral nerve with macro-sieve electrodes following spinal cord injury
Ray, W. Z., Birenbaum, N. K., & MacEwan, M. R. (2017) · Neural Regeneration Research · doi:10.4103/1673-5374.208565 · Cited by 5 · see citing works
“Regenerative electrodes, such as the MSE, interface peripheral nerve that has regenerated across the electrode interface from two surgically opposed nerve stumps (Thompson et al., 2016).”
STEER: 3D Printed Guide for Nerve Regrowth Control and Neural Interface in Non-Human Primate Model
Blasiak, A., Ng, K. A., Wong, M. D. S., Tsai, C.-W., Rusly, A., Gammad, G. G. L., Voges, K., Libedinsky, C., Yen, S., Thakor, N. V., et al. (2021) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2021.3113653 · Cited by 4 · see citing works
Dong, J. C., Jensen, W., Geng, B., Kamavuako, E. N., & Dosen, S. S. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2021.580385 · Cited by 4 · see citing works
“In addition, the subdermal interface can be designed and manufactured as an implantable electrode with stretchable materials (Musick et al., 2015; Thompson et al., 2016), which could endure the local tissue stretching during body movements and long-term implantation.”
BackgroundLiao, Y., Dai, Y., Tang, Z., Ma, H., Tu, Z., Yuan, T., Li, Y., & Wang, B. (2025) · Materials & Design · doi:10.1016/j.matdes.2025.115142 · Cited by 4 · see citing works
Wu, Y., & Guo, L. (2020) · Neural Interface Engineering · doi:10.1007/978-3-030-41854-0_5 · Cited by 3 · see citing works
Vecchi, J. T., Rhomberg, M., Guymon, C. A., et al. (2024) · PLoS ONE · doi:10.1371/journal.pone.0308389 · Cited by 2 · see citing works
“In this pursuit, they seek to engineer biomaterial surfaces or coatings of the electrode to guide neurite outgrowth into close proximity to or contact with the electrode array [5, 41–43].”
High-Porosity Sieve-Type Neural Electrodes for Motor Function Recovery and Nerve Signal Acquisition
Choi, W., Park, H., Oh, S., Seok, S., Yoon, D. S., & Kim, J. (2024) · Micromachines · doi:10.3390/mi15070862 · Cited by 2 · see citing works
Sensitivity of CNN image analysis to multifaceted measurements of neurite growth
Vecchi, J. T., Mullan, S. M., Lopez, J. A., Rhomberg, M., Yamamoto, A., Hallam, A., Lee, A., Milan Sonka, A., & Hansen, M. R. (2023) · BMC Bioinformatics · doi:10.1186/s12859-023-05444-4 · Cited by 2 · see citing works
“…deciphering the mechanisms that regulate neurite growth is essential to understand neural development, treat or prevent peripheral neuropathies [6], direct neurite growth towards targets such as a neural electrode [7, 8], and understand pathophysiological underpinnings of neurological diseases [9].”
BackgroundNeural electrodes for long-term tissue interfaces
del Valle, J., Rodríguez‐Meana, B., & Navarro, X. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822828-9.00009-5 · Cited by 2 · see citing works
Implantation of Elongated Porous Silicon Neural Probe Array in Rat CortexPreprint
Hajj-Hassan, M., Fayad, R., Berro, S., Chodavarapu, V. P., & Musallam, S. (2018) · Preprints.org · doi:10.20944/preprints201801.0080.v1 · Cited by 1 · see citing works
Neuronal excitability and network formation on optically transparent electrode materials
Thompson, C. H., Khan, S., Khan, W., Li, W., & Purcell, E. K. (2017) · International IEEE/EMBS Conference on Neural Engineering · doi:10.1109/ner.2017.8008315 · Cited by 2 · see citing works
Self-citationInduced Pluripotent Stem Cell-Derived Autologous Cell Therapy for Age-Related Macular Degeneration
Khristov, V., Jha, B. S., Rising, A., Li, Y.-C., Qian, H., Maminishkis, A., Amaral, J., Campos, M., & Bharti, K. (2017) · Springer eBooks · doi:10.1007/978-3-319-49479-1_3 · Cited by 2 · see citing works
Advances in neuroprostheses: interfaces, materials, and applications
He, E., Chen, K., Liu, S., Chen, H., Xiao, Y., Chen, R., Tu, P., Pan, G., & Lin, P. (2026) · Nano Convergence · doi:10.1186/s40580-026-00552-2 · Cited by 1 · see citing works
“…signals decoding [13], and neural interfaces [14, 15]), or emerging directions (such as personalized neuroprosthetics [16], regenerative electrodes [17], and neuromorphic hardware [18]), there is a lack of reviews addressing system-level designs to generalize key functional modules across…”
Current Developments and Challenges in the Field of Biohybrid Neural Interfaces—A Scoping Review
Yu, Y., Muzaffar, J., Phillips, V., Yang, Y., Barone, D. G., Malliaras, G. G., & Bance, M. (2025) · Advanced Electronic Materials · doi:10.1002/aelm.202500259 · Cited by 1 · see citing works
Realizing Upper Extremity Bionic Limbs: Leveraging Neuroprosthetic Control Strategies
Kumar, N. G., Chestek, C. A., Cederna, P. S., & Kung, T. A. (2023) · Plastic & Reconstructive Surgery · doi:10.1097/prs.0000000000011183
“12,45 In contrast, sieve electrodes were designed as an alternative to these methods by enabling transected axons to attempt regeneration through microchannels surrounded by electrodes.”
Long Porous Silicon Probe Array for Neural Recordings
Hajj-Hassan, M. (2020) · Archives in Neurology & Neuroscience · doi:10.33552/ann.2020.08.000678
“Neurotrophic probes have had much success, particularly in humans, where they have been able to isolate single units for over 4 years [29-31].”
BackgroundSergi, P. N., Jensen, W., Yoshida, K., & Micera, S. (2018) · Biosystems & biorobotics · doi:10.1007/978-3-030-01845-0_25
- Authors
- Thompson, C. H., Zoratti, M. J., Langhals, N. B., & Purcell, E. K.
- Affiliations
- Michigan State University (US)
- Corresponding author
- Purcell, Erin K.
- Venue
- Tissue Engineering Part B: Reviews
- Publisher
- SAGE Publications
- Published
- 2015-11-23
- Volume
- 22
- Issue
- 2
- Pages
- 125-135
- ISSN
- 1937-3368 (ISSN-L)
- 1937-3376
- OpenAlex
- W2175643771
- Cited by
- 67
- References
- 93
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 4 · 2023: 12 · 2022: 1 · 2021: 11 · 2020: 8 · 2019: 7 · 2018: 5 · 2017: 11 · 2016: 2
- FWCI
- 0.72
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Tissue Engineering Part B Reviews (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Michigan State University
- College of Engineering, Michigan State University
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Health Sciences › Medicine › Surgery › Tissue Engineering and Regenerative Medicine
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Nerve injury and regeneration
- Keywords
- neural prostheses, electrode-tissue interface, nerve regeneration, electrode arrays, biohybrid devices, neuronal growth, topographical cues, bioactive coatings, peripheral nervous system, central nervous system
- MeSH terms
- Electrodes, Implanted, Humans, Nerve Regeneration, Peripheral Nerves, Neural Prostheses
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-11.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-11.
Cited sources are checked by gpt-6-sol.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review