Electric crosstalk impairs spatial resolution of multi-electrode arrays in retinal implants
Open accessAbstract
Active multi-electrode arrays are used in vision prostheses, including optic nerve cuffs and cortical and retinal implants for stimulation of neural tissue. For retinal implants, arrays with up to 1500 electrodes are used in clinical trials. The ability to convey information with high spatial resolution is critical for these applications. To assess the extent to which spatial resolution is impaired by electric crosstalk, finite-element simulation of electric field distribution in a simplified passive tissue model of the retina is performed. The effects of electrode size, electrode spacing, distance to target cells, and electrode return configuration (monopolar, tripolar, hexagonal) on spatial resolution is investigated in the form of a mathematical model of electric field distribution. Results show that spatial resolution is impaired with increased distance from the electrode array to the target cells. This effect can be partly compensated by non-monopolar electrode configurations and larger electrode diameters, albeit at the expense of lower pixel densities due to larger covering areas by each stimulation electrode. In applications where multi-electrode arrays can be brought into close proximity to target cells, as presumably with epiretinal implants, smaller electrodes in monopolar configuration can provide the highest spatial resolution. However, if the implantation site is further from the target cells, as is the case in suprachoroidal approaches, hexagonally guarded electrode return configurations can convey higher spatial resolution.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“In contrast, current prosthesis users report seeing highly distorted phosphenes, which vary in shape across subjects as well as electrodes and often fail to assemble into more complex percepts (Wilke et al., 2011; Beyeler et al., 2019; Beauchamp et al., 2020; Erickson-Davis and Korzybska, 2021; Fernández et al., 2021).”
From Towards a Smart Bionic Eye : AI-powered artificial vision for the treatment of incurable blindness (2022) · cites this paper as Wilke et al., 2011
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Gabriel, S., Lau, R.-W.-M., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/002 · Cited by 4,487
Electrical stimulation of excitable tissue: design of efficacious and safe protocols
Merrill, D. R., Bikson, M., & Jefferys, J. G. R. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2004.10.020 · Cited by 2,078
Subretinal electronic chips allow blind patients to read letters and combine them to words
Zrenner, E., Bartz-Schmidt, K. U., Benav, H., Besch, D., Bruckmann, A., Gabel, V.-P., Gekeler, F., Greppmaier, U., Harscher, A., Kibbel, S., et al. (2011) · Proceedings of the Royal Society B: Biological Sciences · doi:10.1098/rspb.2010.1747 · Cited by 819
Self-citationWill Retinal Implants Restore Vision?
Zrenner, E. (2002) · Science · doi:10.1126/science.1067996 · Cited by 736
Retinal remodeling during retinal degeneration
Jones, B. W., & Marc, R. E. (2005) · Experimental Eye Research · doi:10.1016/j.exer.2005.03.006 · Cited by 430
Electrical Stimulation of Mammalian Retinal Ganglion Cells With Multielectrode Arrays
Sekirnjak, C., Hottowy, P., Sher, A., Dąbrowski, W., Litke, A. M., & Chichilnisky, E. J. (2006) · Journal of Neurophysiology · doi:10.1152/jn.01168.2005 · Cited by 342
Design of a high-resolution optoelectronic retinal prosthesis
Palanker, D., Vankov, A., Huie, P., & Baccus, S. A. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/1/012 · Cited by 330
Non-linear projection of the retinal image in a wide-angle schematic eye.
Drasdo, N., & Fowler, C. W. (1974) · British Journal of Ophthalmology · doi:10.1136/bjo.58.8.709 · Cited by 312
Axonal Sodium-Channel Bands Shape the Response to Electric Stimulation in Retinal Ganglion Cells
Fried, S. I., Lasker, A., Desai, N. J., Eddington, D. K., & Rizzo, J. F. (2009) · Journal of Neurophysiology · doi:10.1152/jn.91081.2008 · Cited by 206
Current focusing and steering: Modeling, physiology, and psychophysics
Bonham, B. H., & Litvak, L. M. (2008) · Hearing Research · doi:10.1016/j.heares.2008.03.006 · Cited by 198
High-Resolution Electrical Stimulation of Primate Retina for Epiretinal Implant Design
Sekirnjak, C., Hottowy, P., Sher, A., Dąbrowski, W., Litke, A. M., & Chichilnisky, E. J. (2008) · Journal of Neuroscience · doi:10.1523/jneurosci.5138-07.2008 · Cited by 195
Stimulation with a Wireless Intraocular Epiretinal Implant Elicits Visual Percepts in Blind Humans
Klauke, S., Goertz, M., Rein, S., Hoehl, D., Thomas, U., Eckhorn, R., Bremmer, F., & Wächtler, T. (2010) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.09-4410 · Cited by 162
Development and Implantation of a Minimally Invasive Wireless Subretinal Neurostimulator
Shire, D. B., Kelly, S. K., Chen, J., Doyle, P., Gingerich, M. D., Cogan, S. F., Drohan, W. A., Mendoza, O., Theogarajan, L. S., Wyatt, J. L., et al. (2009) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2021401 · Cited by 156
Wilke, R., Gabel, V.-P., Sachs, H., Bartz Schmidt, K.-U., Gekeler, F., Besch, D., Szurman, P., Stett, A., Wilhelm, B., Peters, T., et al. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-6946 · Cited by 155
Self-citationFeasibility Study of a Retinal Prosthesis
Caspi, A. (2009) · Archives of Ophthalmology · doi:10.1001/archophthalmol.2009.20 · Cited by 149
Focal activation of the feline retina via a suprachoroidal electrode array
Wong, Y. T., Chen, S. C., Seo, J. M., Morley, J. W., Lovell, N. H., & Suaning, G. J. (2009) · Vision Research · doi:10.1016/j.visres.2009.02.018 · Cited by 109
Self-citationSpatiotemporal Interactions in Retinal Prosthesis Subjects
Horsager, A., Greenberg, R. J., & Fine, I. (2010) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.09-3746 · Cited by 95
Brindley, G. S. (1956) · The Journal of Physiology · doi:10.1113/jphysiol.1956.sp005647 · Cited by 92
Laminar profile of resistivity in frog retina
Karwoski, C. J., Frambach, D. A., & Proenza, L. M. (1985) · Journal of Neurophysiology · doi:10.1152/jn.1985.54.6.1607 · Cited by 84
Retinal Neurostimulator for a Multifocal Vision Prosthesis
Wong, Y. T., Dommel, N., Preston, P., Hallum, L. E., Lehmann, T., Lovell, N. H., & Suaning, G. J. (2007) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2007.903958 · Cited by 79
Self-citationStett, A., Mai, A., & Herrmann, T. (2007) · Journal of Neural Engineering · doi:10.1088/1741-2560/4/1/s02 · Cited by 59
A CMOS retinal neurostimulator capable of focussed, simultaneous stimulation
Dommel, N., Wong, Y. T., Lehmann, T., Dodds, C. W. D., Lovell, N. H., & Suaning, G. J. (2009) · Journal of Neural Engineering · doi:10.1088/1741-2560/6/3/035006 · Cited by 54
Self-citationFornos, A. P., Sommerhalder, J., Pittard, A., Safran, A. B., & Pelizzone, M. (2008) · Vision Research · doi:10.1016/j.visres.2008.04.027 · Cited by 50
Temporal interactions during paired-electrode stimulation in two retinal prosthesis subjects
Horsager, A., Boynton, G. M., Greenberg, R. J., & Fine, I. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-5282 · Cited by 48
Lyle, W. M. (1989) · Optometry and Vision Science · doi:10.1002/j.1538-9235.1989.tb02323.x · Cited by 31
Electric Field Stimulation of Bipolar Cells in a Degenerated Retina—A Theoretical Study
Gerhardt, M., Alderman, J., & Stett, A. (2010) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2037323 · Cited by 21
In-Vitro Testing of Simultaneous Charge Injection and Recovery in a Retinal Neuroprosthesis
Dommel, N., Suaning, G. J., Preston, P., Lehmann, T., & Lovell, N. H. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1616274 · Cited by 17
Self-citationZiv, O. R., Rizzo, J. F., & Jensen, R. J. (2005) · Journal of Neural Engineering · doi:10.1088/1741-2560/2/1/002 · Cited by 15
Current Distribution During Parallel Stimulation: Implications for an Epiretinal Neuroprosthesis
Lovell, N. H., Dokos, S., Cloherty, S. L., Preston, P., & Suaning, G. J. (2005) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2005.1615661 · Cited by 14
Self-citationWilke, R. G. H., Greppmaier, U., Štingl, K., & Zrenner, E. (2011) · Investigative Ophthalmology & Visual Science · Cited by 12
Self-citationFactors Affecting Perceptual Thersholds of Subretinal Electric Stimulation in Blind Volunteers
Wilke, R. G. H., Greppmaier, U., Harscher, A., Benav, H., & Zrenner, E. (2010) · Investigative Ophthalmology & Visual Science · Cited by 7
Self-citationModeling Advantages of Subretinal Microphotodiode-Arrays Utilizing Sequential Electrode Activation
Benav, H., Wilke, R. G. H., Stett, A., & Zrenner, E. (2009) · Investigative Ophthalmology & Visual Science · Cited by 3
Self-citationElectric Stimulation of the RetinaPreprint
Schmid, E. W., & Wilke, R. G. H. (2010) · arXiv (Cornell University) · doi:10.48550/arxiv.1012.5958 · Cited by 2
Self-citation
Research works printed but not linked: 7
- Benav H Wilke R Stett A Zrenner E, 2009, ARVO Meeting Abstracts, 50, 4593
- Dommel N, 2005, Conf. Proc. IEEE Eng. Med. Biol. Soc., 7, 7612
- Lovell N H, 2005, Conf. Proc. IEEE Eng. Med. Biol. Soc., 5, 5242
- Schmid E W, 2008, U S Patent Specification 7840273
- Wilke R, 2011, Invest. Ophthalmol. Vis. Sci.
- Wilke R Greppmaier U Harscher A Benav H Zrenner E, 2010, ARVO Meeting Abstracts, 51, 2026
- Wilke R G, 2011, ARVO Meeting Abstracts, 52, 458
Cited by
Sorted by citations
A Materials Roadmap to Functional Neural Interface Design
Wellman, S. M., Eles, J. R., Ludwig, K. A., Seymour, J. P., Michelson, N. J., McFadden, W. E., Vazquez, A. L., & Kozai, T. D. Y. (2018) · Advanced Functional Materials · doi:10.1002/adfm.201701269 · Cited by 389
Retinal stimulation strategies to restore vision: fundamentals and systems
Yue, L., Weiland, J. D., Roska, B., & Humayun, M. S. (2016) · Progress in Retinal and Eye Research · doi:10.1016/j.preteyeres.2016.05.002 · Cited by 272
A model of ganglion axon pathways accounts for percepts elicited by retinal implants
Beyeler, M., Nanduri, D., Weiland, J. D., Rokem, A., Boynton, G. M., & Fine, I. (2019) · Scientific Reports · doi:10.1038/s41598-019-45416-4 · Cited by 143
Wilke, R., Gabel, V.-P., Sachs, H., Bartz Schmidt, K.-U., Gekeler, F., Besch, D., Szurman, P., Stett, A., Wilhelm, B., Peters, T., et al. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-6946 · Cited by 155
Self-citationGold nanostructures: synthesis, properties, and neurological applications
Zare, I., Yaraki, M. T., Speranza, G., Najafabadi, A. H., Shourangiz-Haghighi, A., Nik, A. B., Manshian, B. B., Saraiva, C., Soenen, S. J., Kogan, M. J., et al. (2022) · Chemical Society Reviews · doi:10.1039/d1cs01111a · Cited by 133
Metadata
- Authors
- Wilke, R. G. H., Moghadam, G. K., Lovell, N. H., Suaning, G. J., & Dokos, S.
- Affiliations
- UNSW Sydney (AU)
- Corresponding author
- Lovell, N H
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2011-06-15
- Volume
- 8
- Issue
- 4
- Pages
- 046016
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2084193319
- Cited by
- 110
- References
- 34
- Citations by year
- 2026: 3 · 2025: 9 · 2024: 8 · 2023: 5 · 2022: 15 · 2021: 8 · 2020: 8 · 2019: 4 · 2018: 6 · 2017: 10 · 2016: 9 · 2015: 3 · 2014: 6 · 2013: 6 · 2012: 8
- FWCI
- 2.92
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Journal of Neural Engineering (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Deutsche Forschungsgemeinschaft
- University of New South Wales
- Eberhard Karls Universität Tübingen
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Analog and Mixed-Signal Circuit Design
- Physical Sciences › Engineering › Electrical and Electronic Engineering › CCD and CMOS Imaging Sensors
- Keywords
- retinal prosthesis, microelectrode arrays, spatial resolution, electric field distribution, electrode configuration, monopolar electrodes, electrode diameter, electrode spacing, epiretinal implants, finite element analysis
- MeSH terms
- Algorithms, Electrodes, Implanted, Electromagnetic Fields, Electronics, Prosthesis Design, Retina, Models, Statistical, Prostheses and Implants, Finite Element Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review