Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care/hospital/ambulatory patients in the United States and Canada
The DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
OBJECTIVES: Existing handheld glucose meters are glucose oxidase (GO)-based. Oxygen side reactions can introduce oxygen dependency, increase potential error, and limit clinical use. Our primary objectives were to: a) introduce a new glucose dehydrogenase (GD)-based electrochemical biosensor for point-of-care testing; b) determine the oxygen-sensitivity of GO- and GD-based electrochemical biosensor test strips; and c) evaluate the clinical performance of the new GD-based glucose meter system in critical care/hospital/ambulatory patients. DESIGN: Multicenter study sites compared glucose levels determined with GD-based biosensors to glucose levels determined in whole blood with a perchloric acid deproteinization hexokinase reference method. One site also studied GO-based biosensors and venous plasma glucose measured with a chemistry analyzer. Biosensor test strips were used with a handheld glucose monitoring system. Bench and clinical oxygen sensitivity, hematocrit effect, and precision were evaluated. SETTING: The study was performed at eight U.S. medical centers and one Canadian medical center. PATIENTS: There were 1,248 patients. RESULTS: The GO-based biosensor was oxygen-sensitive. The new GD-based biosensor was oxygen-insensitive. GD-based biosensor performance was acceptable: 2,104 (96.1%) of 2,189 glucose meter measurements were within +/-15 mg/dL (+/-0.83 mmol/L) for glucose levels of < or = 100 mg/dL (< or = 5.55 mmol/L) or within +/-15% for glucose levels of > 100 mg/dL, compared with the whole-blood reference method results. With the GD-based biosensor, the percentages of glucose measurements that were not within the error tolerance were comparable for different specimen types and clinical groups. Bracket predictive values were acceptable for glucose levels used in therapeutic management. CONCLUSIONS: The performance of GD-based, oxygen-insensitive, handheld glucose testing was technically suitable for arterial specimens in critical care patients, cord blood and heelstick specimens in neonates, and capillary and venous specimens in other patients. Multicenter findings benchmark the performance of bedside glucose testing devices. With the new +/-15 mg/dL --> 100 mg/dL --> +/-15% accuracy criterion, point-of-care systems for handheld glucose testing should score 95% (or better), as compared with the recommended reference method. Physiologic changes, preanalytical factors, confounding variables, and treatment goals must be taken into consideration when interpreting glucose results, especially in critically ill patients, for whom arterial blood glucose measurements will reflect systemic glucose levels.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Reichard, P., Nilsson, B.-Y., & Rosenqvist, U. (1993) · New England Journal of Medicine · doi:10.1056/nejm199307293290502 · Cited by 1,202
Design and in vitro studies of a needle-type glucose sensor for subcutaneous monitoring
Bindra, D. S., Zhang, Y., Wilson, G. S., Sternberg, R., Thevenot, D. R., Moatti, D., & Reach, G. (1991) · Analytical Chemistry · doi:10.1021/ac00017a008 · Cited by 309
Zhang, Y., & Wilson, G. S. (1993) · Analytica Chimica Acta · doi:10.1016/0003-2670(93)85009-9 · Cited by 92
Cited by
Emerging Technologies for Next-Generation Point-of-Care Testing
Vashist, S. K., Luppa, P. B., Yeo, L. Y., Ozcan, A., & Luong, J. H. T. (2015) · Trends in Biotechnology · doi:10.1016/j.tibtech.2015.09.001 · Cited by 705
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-1135-eodhlo · Cited by 293
Self-citationContinuous glucose monitoring: an Endocrine Society Clinical Practice Guideline
Klonoff, D. C., Buckingham, B., Christiansen, J. S., Montori, V. M., Tamborlane, W. V., Vigersky, R. A., & Wolpert, H. (2011) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2010-2756 · Cited by 211
Approved IFCC recommendation on reporting results for blood glucose (abbreviated)
D'Orazio, P. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.051979 · Cited by 208
Tang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · American Journal of Clinical Pathology · doi:10.1309/qaw1-x5xw-bvrq-5lkq · Cited by 207
Self-citationTang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00038 · Cited by 188
“Table 1 summarizes the characteristics of the six glucose meter systems evaluated: (1) Accu-Chek Advantage H and (2) Accu-Chek Comfort Curve (Roche Diagnostics, Indianapolis, IN), (3) Precision PCx and (4) Precision QID (Abbott Laboratories, Bedford, MA), (5) SureStep (LifeScan, Milpitas, CA), and (6) Glucometer Elite…”
BackgroundMethodSelf-citationLouie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Self-citationCorstjens, A. M., Ligtenberg, J. J., van der Horst, I. C., Spanjersberg, R., Lind, J. S., Tulleken, J. E., Meertens, J. H., & Zijlstra, J. G. (2006) · Critical Care · doi:10.1186/cc5048 · Cited by 163
“[21] are (a) within ± 0.8 mmol/l (15 mg/dl) of the reference measurement for glucose levels ≤ 5.6 mmol/l (100 mg/dl) and (b) within approximately 15% of the reference measurement for glucose levels greater than 5.6 mmol/l (100 mg/dl).”
BackgroundInfluentialOverview on self-monitoring of blood glucose
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.01.002 · Cited by 116
“0825 mmol/L, when compared with the reference method [44].”
MethodInterferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816641433 · Cited by 115
IFCC recommendation on reporting results for blood glucose
(2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00431-4 · Cited by 110
Balion, C., Grey, V., Ismaila, A., Blatz, S., & Seidlitz, W. (2006) · BMC Pediatrics · doi:10.1186/1471-2431-6-28 · Cited by 46
Performance of a new interference-resistant glucose meter
Vanavanan, S., Santanirand, P., Chaichanajarernkul, U., Chittamma, A., DuBois, J. A., Shirey, T., & Heinz, M. (2010) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.09.010 · Cited by 39
“[11] have proposed a modification of the previous CLSI guideline that the discrepancy between meters and routine”
MethodTendl, K. A., Christoph, J., Bohn, A., Herkner, K. R., Pollak, A., & Prusa, A.-R. (2013) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm-2012-0864 · Cited by 18
Metadata
- Authors
- Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., & Whitlow, K. J.
- Venue
- Critical Care Medicine
- Published
- 1998-03
- Volume
- 26
- Issue
- 3
- Pages
- 581-590
- ISSN
- 0090-3493 (ISSN-L)
- OpenAlex
- W2055790308
- Cited by
- 120
- References
- 4
- FWCI
- 5.02
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.