Characterization of signals and noise rejection with bipolar longitudinal intrafascicular electrodes
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
Longitudinal intrafascicular electrodes (LIFE's) are fine electrodes threaded into the extracellular space between axons in peripheral nerves or spinal roots. We are developing these electrodes for application in functional electrical stimulation and in basic physiology. An area of concern in chronic recording application of LIFE's is the possibility of electromyogram and other external noise sources masking the recorded neural signals. We characterized neural signals recorded by LIFE's and confirmed by three independent methods that increasing interelectrode spacing for bipolar LIFE's increases signal amplitude. The spectrum of neural signal from bipolar and monopolar LIFE lies between 300 Hz and 10 kHz. The amplitude of the spectrum increases with increasing interelectrode spacing, although the distribution is not affected. Single unit analysis of LIFE recordings show that they record selectively from units closest to the electrode active site. Units with conduction velocities ranging from 50-120 m/s were identified. Extraneural noise, as stimulus artifact or electromyogram, is much reduced with bipolar LIFE recording, as compared to monopolar recordings. Relative improvement in neural signal to extraneural noise increases with interelectrode spacing up to about 2 mm. Since there is no further improvement beyond 2 mm, we conclude that the preferred interelectrode spacing for bipolar LIFE's is 2 mm.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Selective stimulation of peripheral nerve fibers using dual intrafascicular electrodes
Yoshida, K., & Horch, K. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.243412 · Cited by 208 · cited 1×
“Acute animal studies have shown that LIFE’s can provide topologically selective stimulation [1] as well as recordings from a limited number of nerve fibers, some of which can be resolved individually [2]–[6].”
BackgroundSelf-citationChronically implanted intrafascicular recording electrodes
Lefurge, T., Goodall, E., Horch, K., Stensaas, L., & Schoenberg, A. (1991) · Annals of Biomedical Engineering · doi:10.1007/bf02368469 · Cited by 172
Cutaneous whole nerve recordings used for correction of footdrop in hemiplegic man
Haugland, M. K., & Sinkjaer, T. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481970 · Cited by 168
Sensory nerve recording for closed‐loop control to restore motor functions
Popovic, D. B., Stein, R. B., Jovanovic, K. L., Dai, R., Kostov, A., & Armstrong, W. W. (1993) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.247801 · Cited by 159
Self-citationAn intrafascicular electrode for recording of action potentials in peripheral nerves
Malagodi, M. S., Horch, K. W., & Schoenberg, A. A. (1989) · Annals of Biomedical Engineering · doi:10.1007/bf02368058 · Cited by 107
Long‐term effects of axotomy on neural activity during cat locomotion.
Gordon, T., Hoffer, J. A., Jhamandas, J., & Stein, R. B. (1980) · The Journal of Physiology · doi:10.1113/jphysiol.1980.sp013283 · Cited by 65 · cited 1×
“In fact, the signal amplitude in LG/S and MG cuffs is generally similar to the sciatic cuffs [9], [12].”
BackgroundSelf-citationInformation contained in sensory nerve recordings made with intrafascicular electrodes
Goodall, E. V., Lefurge, T. M., & Horch, K. W. (1991) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.83604 · Cited by 63 · cited 1×
“Unit shape recognition and tracking may be used as a means to get multiple channels of information out of a single electrode [2]–[4], [6].”
Method
Research works printed but not linked: 15
- 10.1385/0-89603-185-3:65 · DOI · Paper page
- stein, 1975, principles underlying new methods for chronic neural recording, Can J Neurological Sci, 2, 235, 10.1017/S0317167100020333 · DOI · Paper page
- clark, 1968, the extracellular potential field of the single active nerve fiber in a volume conductor, Biophys J, 8, 842, 10.1016/S0006-3495(68)86524-5 · DOI · Paper page
- 10.1109/TBME.1974.324406 · DOI · Paper page
- stein, 1983, Nerve and Muscle
- 10.1016/S0006-3495(76)85719-0 · DOI
- 10.1111/j.1525-1594.1997.tb04661.x · DOI · Paper page
- 10.1007/BF00233991 · DOI
- 10.1007/BF02458045 · DOI
- 10.1109/10.301727 · DOI · Paper page
- 10.1109/10.481986 · DOI · Paper page
- 10.1109/10.125014 · DOI · Paper page
- 10.1055/s-2007-1006661 · doi:10.1055/s-2007-1006661
- 10.1109/10.277276 · DOI
- 10.1007/BF00698791 · DOI
Cited by
Navarro, X., Krueger, T. B., Lago, N., Micera, S., Stieglitz, T., & Dario, P. (2005) · Journal of the Peripheral Nervous System · doi:10.1111/j.1085-9489.2005.10303.x · Cited by 845
Control of Hand Prostheses Using Peripheral Information
Micera, S., Carpaneto, J., & Raspopovic, S. (2010) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2010.2085429 · Cited by 403
Micera, S., Navarro, X., Carpaneto, J., Citi, L., Tonet, O., Rossini, P. M., Carrozza, M. C., Hoffmann, K. P., Vivo, M., Yoshida, K., et al. (2008) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2008.2006207 · Cited by 120
Self-citationLawrence, S. M., Dhillon, G. S., Jensen, W., Yoshida, K., & Horch, K. W. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2004.831491 · Cited by 106
Self-citationInterfaces with the Peripheral Nerve for the Control of Neuroprostheses
del Valle, J., & Navarro, X. (2013) · International Review of Neurobiology · doi:10.1016/b978-0-12-420045-6.00002-x · Cited by 105
Metadata
- Authors
- Yoshida, K., & Stein, R. B.
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 1999
- Volume
- 46
- Issue
- 2
- Pages
- 226-234
- ISSN
- 0018-9294 (ISSN-L)
- OpenAlex
- W2098338390
- Cited by
- 76
- References
- 7
- FWCI
- 0.34
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Characterization of signals and noise rejection with bipolar longitudinal intrafascicular electrodes. https://app.cochat.ai/papers/10.1109/10.740885
- The paper
- Yoshida, K., & Stein, R. B. (1999). Characterization of signals and noise rejection with bipolar longitudinal intrafascicular electrodes. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/10.740885
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.