Impedance and Noise of Passive and Active Dry EEG Electrodes: A Review
Open accessThe DOI matches its Crossref record; 80 of the 80 references it cites were found in a registry or an index.
Abstract
Dry electrodes are a promising solution for prolonged EEG signal acquisition, whereas wet electrodes may lose their signal quality in the same situation and require skin preparation for set-up. Here, we review the impedance and noise of passive and active dry EEG electrodes. In addition, we compare noise and input impedance of the EEG amplifiers. As there are multiple definitions of impedance in each EEG system, they are all first defined. Electrodes must be compatible with amplifiers to accurately record EEG signals. This implies that their impedance plays a significant role in amplifier compatibility and affects total input-referred noise. Therefore, we review the impedance and noise of state-of-the-art amplifiers and electrodes. Furthermore, we compare the various structures and materials used and their final impedance to that of wet electrodes. Finally, we compare state-of-the-art electrodes and amplifiers to the standards of the IFCN and IEC80601-2-26. We investigate bottlenecks and propose a guideline for future work on passive and active dry electrodes, as well as EEG amplifiers.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Computed Tomography — An Increasing Source of Radiation Exposure
Brenner, D. J., & Hall, E. J. (2007) · New England Journal of Medicine · doi:10.1056/nejmra072149 · Cited by 8,748 · cited 1×
“C URRENTLY, there are numerous techniques for monitoring brain activity, such as computer tomography (CT) [1], magnetic resonance imaging (MRI) [2], functional magnetic resonance imaging (fMRI) [3], positron emission tomography (PET) [4], magnetoencephalography (MEG) [5], and electroencephalogram…”
MethodDesign of Analog CMOS Integrated Circuits
Razavi, B. (1999) · Cited by 6,609
Electroencephalography—Basic principles, clinical applications and related fields
(1982) · Journal of the Autonomic Nervous System · doi:10.1016/0165-1838(82)90067-4 · Cited by 3,427
Nunez, P. L., & Srinivasan, R. (2006) · doi:10.1093/acprof:oso/9780195050387.001.0001 · Cited by 2,531
Electric Fields of the Brain: The Neurophysics of EEG
Nunez, P. L., & Harth, E. (1982) · Physics Today · doi:10.1063/1.2915137 · Cited by 2,232 · cited 1×
“INTRODUCTION CURRENTLY, there are numerous techniques for monitoring brain activity, such as computer tomography (CT) [1], magnetic resonance imaging (MRI) [2], functional magnetic resonance imaging (fMRI) [3], positron emission tomography (PET) [4], magnetoencephalography (MEG) [5], and electroencephalogram (EEG) [6].”
MethodElectroencephalography: Basic Principles, Clinical Applications and Related Fields
Torres, F. (1983) · Archives of Neurology · doi:10.1001/archneur.1983.04050030085025 · Cited by 1,812
Lam, C.-W., James, J. T., McCluskey, R., Arepalli, S., & Hunter, R. L. (2006) · Critical Reviews in Toxicology · doi:10.1080/10408440600570233 · Cited by 1,237 · cited 1×
“Given the high potential risk of CNTs, but their good properties, there is an extensive literature in this field that has focused on various parameters that affect the toxicity of CNTs, including size, length, agglomeration, and impurities [74]–[78].”
BackgroundDry-Contact and Noncontact Biopotential Electrodes: Methodological Review
Chi, Y. M., Jung, T.-P., & Cauwenberghs, G. (2010) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2010.2084078 · Cited by 1,061 · cited 6×
“Although this is the flicker noise of a single MOSFET, an increase in noise density can be observed at lower frequencies for a single small metal electrode or one of another material, proving the existence of flicker noise in other materials [29].”
BackgroundInfluentialCrispin, X., Jakobsson, F. L. E., Crispin, A., Grim, P. C. M., Ersman, P. A., Volodin, A., Haesendonck, C. V., der Auweraer, M. V., Salaneck, W. R., & Berggren, M. (2006) · Chemistry of Materials · doi:10.1021/cm061032+ · Cited by 991 · cited 1×
“Various materials have been investigated to identify the best quality of the recorded signal, including platinum, PEDOT/PSS [72], sintered Ag/AgCl, disposable Ag/AgCl, silver, gold-plated silver, stainless steel, tin [57], [66], and iridium oxide [34].”
BackgroundA direct comparison of wet, dry and insulating bioelectric recording electrodes
Searle, A., & Kirkup, L. (2000) · Physiological Measurement · doi:10.1088/0967-3334/21/2/307 · Cited by 725 · cited 1×
“The electrical power line around electrodes is capacitively coupled to them [53].”
BackgroundIFCN standards for digital recording of clinical EEG
Nuwer, M. R., Comi, G., Emerson, R., Fuglsang-Frederiksen, A., Guérit, J.-M., Hinrichs, H., Ikeda, A., Jose C. Luccas, F., & Rappelsburger, P. (1998) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/s0013-4694(97)00106-5 · Cited by 671
Carbon nanotubes for biological and biomedical applications
Yang, W., Thordarson, P., Gooding, J. J., Ringer, S. P., & Braet, F. (2007) · Nanotechnology · doi:10.1088/0957-4484/18/41/412001 · Cited by 648
Enhancing the quality of life through wearable technology
Park, S., & Jayaraman, S. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213625 · Cited by 614
Toxicity issues in the application of carbon nanotubes to biological systems
Firme, C. P., & Bandaru, P. R. (2010) · Nanomedicine: Nanotechnology, Biology and Medicine · doi:10.1016/j.nano.2009.07.003 · Cited by 545
Verma, N., Shoeb, A., Bohorquez, J., Dawson, J., Guttag, J., & Chandrakasan, A. P. (2010) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2010.2042245 · Cited by 526 · cited 1×
“It has been productively used in epilepsy, Parkinson’s disease, narcolepsy, depression, motor impairment, sports, entertainment, and computer-related functions [8]–[16].”
Method(2006) · The International Series in Engineering and Computer Science · doi:10.1007/b135984 · Cited by 511
Fan, Q., Sebastiano, F., Huijsing, J. H., & Makinwa, K. A. A. (2011) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2143610 · Cited by 473
Driven-right-leg circuit design
Winter, B. B., & Webster, J. G. (1983) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1983.325168 · Cited by 443 · cited 1×
“4 which is based on the structure in [31].”
BackgroundLopez-Gordo, M., Sanchez-Morillo, D., & Valle, F. (2014) · Sensors · doi:10.3390/s140712847 · Cited by 410 · cited 1×
“tem [28]–[30].”
MEG: An Introduction to Methods
(2010) · Oxford University Press eBooks · doi:10.1093/acprof:oso/9780195307238.001.0001 · Cited by 408 · cited 1×
“…there are numerous techniques for monitoring brain activity, such as computer tomography (CT) [1], magnetic resonance imaging (MRI) [2], functional magnetic resonance imaging (fMRI) [3], positron emission tomography (PET) [4], magnetoencephalography (MEG) [5], and electroencephalogram (EEG) [6].”
MethodWearable Electroencephalography
Casson, A., Yates, D., Smith, S., Duncan, J., & Rodriguez-Villegas, E. (2010) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2010.936545 · Cited by 404
Alshehri, R., Ilyas, A. M., Hasan, A., Arnaout, A., Ahmed, F., & Memic, A. (2016) · Journal of Medicinal Chemistry · doi:10.1021/acs.jmedchem.5b01770 · Cited by 381
Is Resting Anterior EEG Alpha Asymmetry a Trait Marker for Depression?
Debener, S., Beauducel, A., Nessler, D., Brocke, B., Heilemann, H., & Kayser, J. (2000) · Neuropsychobiology · doi:10.1159/000026630 · Cited by 358
Evaluation of commercially available electrodes and gels for recording of slow EEG potentials
Tallgren, P., Vanhatalo, S., Kaila, K., & Voipio, J. (2005) · Clinical Neurophysiology · doi:10.1016/j.clinph.2004.10.001 · Cited by 344 · cited 2×
“Various materials have been investigated to identify the best quality of the recorded signal, including platinum, PEDOT/PSS [72], sintered Ag/AgCl, disposable Ag/AgCl, silver, gold-plated silver, stainless steel, tin [57], [66], and iridium oxide [34].”
BackgroundSkin impedance from 1 Hz to 1 MHz
Rosell, J., Colominas, J., Riu, P., Pallas-Areny, R., & Webster, J. G. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.4599 · Cited by 343 · cited 1×
“The range of impedance of this skin layer varies from person to person, and is from approximately 10 k ! to 1 M ! per square centimeter at 1 Hz [32], [33].”
BackgroundNovel Dry Polymer Foam Electrodes for Long-Term EEG Measurement
Lin, C.-T., Liao, L.-D., Liu, Y.-H., Wang, I.-J., Lin, B.-S., & Chang, J.-Y. (2011) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2102353 · Cited by 336 · cited 1×
“In [65], a dry polymer foam-based electrode for EEG measurements was introduced.”
MethodDry and Noncontact EEG Sensors for Mobile Brain–Computer Interfaces
Chi, Y. M., Wang, Y.-T., Wang, Y., Maier, C., Jung, T.-P., & Cauwenberghs, G. (2012) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2011.2174652 · Cited by 297 · cited 2×
“Different types of dry electrodes: A) insulated electrode [22], B) non-contact electrode [23], and C) dry contact electrode [24]. in some cases.”
BackgroundAn Implantable 455-Active-Electrode 52-Channel CMOS Neural Probe
Lopez, C. M., Andrei, A., Mitra, S., Welkenhuysen, M., Eberle, W., Bartic, C., Puers, R., Yazicioglu, R. F., & Gielen, G. G. E. (2014) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2013.2284347 · Cited by 264 · cited 1×
“Active electrodes offer the best solution to reduce most noise sources, e.g., crosstalk, movement artifacts, noise pick-up, other forms of interference, and especially 50/60 Hz noise, due to its impedance conversion [54].”
BackgroundBristle-sensors—low-cost flexible passive dry EEG electrodes for neurofeedback and BCI applications
Grozea, C., Voinescu, C. D., & Fazli, S. (2011) · Journal of Neural Engineering · doi:10.1088/1741-2560/8/2/025008 · Cited by 256 · cited 2×
“Another type of dry electrode is the bristle dry electrode, proposed in [59].”
BackgroundSafe Clinical Use of Carbon Nanotubes as Innovative Biomaterials
Saito, N., Haniu, H., Usui, Y., Aoki, K., Hara, K., Takanashi, S., Shimizu, M., Narita, N., Okamoto, M., Kobayashi, S., et al. (2014) · Chemical Reviews · doi:10.1021/cr400341h · Cited by 243 · cited 1×
“Given the high potential risk of CNTs, but their good properties, there is an extensive literature in this field that has focused on various parameters that affect the toxicity of CNTs, including size, length, agglomeration, and impurities [74]–[78].”
BackgroundSoft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG Recording
Chen, Y.-H., De Beeck, M., Vanderheyden, L., Carrette, E., Mihajlović, V., Vanstreels, K., Grundlehner, B., Gadeyne, S., Boon, P., & Van Hoof, C. (2014) · Sensors · doi:10.3390/s141223758 · Cited by 238
A $160~\mu {\rm W}$ 8-Channel Active Electrode System for EEG Monitoring
Xu, J., Yazicioglu, R. F., Grundlehner, B., Harpe, P., Makinwa, K. A. A., & Van Hoof, C. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2170985 · Cited by 221 · cited 1×
“The first low-power active electrode application-specific integrated circuit (ASIC) system for EEG monitoring was designed in 2011 [44].”
Hinrichs, H., Scholz, M., Baum, A. K., Kam, J. W. Y., Knight, R. T., & Heinze, H.-J. (2020) · Scientific Reports · doi:10.1038/s41598-020-62154-0 · Cited by 219 · cited 1×
“Due to huge progress in dry electrodes technology, they comply with the needs of clinical EEG applications and higher user comfort is achieved for dry electrodes in comparison with wet electrodes [95]–[97].”
BackgroundPrutchi, D., & Norris, M. (2004) · Wiley eBooks · doi:10.1002/0471681849 · Cited by 215
A Study of Evoked Potentials From Ear-EEG
Kidmose, P., Looney, D., Ungstrup, M., Rank, M. L., & Mandic, D. P. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2264956 · Cited by 201 · cited 1×
“For example, several studies have performed EEG on the ear [17], [18].”
BackgroundIntroduction to Functional Magnetic Resonance Imaging: Principles and Techniques
Buxton, R. B. (2009) · Cambridge University Press eBooks · doi:10.1017/cbo9780511605505 · Cited by 200
In-Ear EEG From Viscoelastic Generic Earpieces: Robust and Unobtrusive 24/7 Monitoring
Goverdovsky, V., Looney, D., Kidmose, P., & Mandic, D. P. (2016) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2471183 · Cited by 192 · cited 1×
“For example, several studies have performed EEG on the ear [17], [18].”
BackgroundKam, J. W. Y., Griffin, S., Shen, A., Patel, S., Hinrichs, H., Heinze, H.-J., Deouell, L. Y., & Knight, R. T. (2019) · NeuroImage · doi:10.1016/j.neuroimage.2018.09.012 · Cited by 188
Chandrakumar, H., & Markovic, D. (2017) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2017.2753824 · Cited by 187 · cited 2×
“The proposed chopper amplifier in [100] has an input impedance of 1.”
BackgroundA dry electrode for EEG recording
Taheri, B. A., Knight, R. T., & Smith, R. L. (1994) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(94)90053-1 · Cited by 176 · cited 1×
“, which is considerably lower than that of the epidermis layer [36].”
Backgroundvon Lühmann, A., Wabnitz, H., Sander, T. H., & Müller, K. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2594127 · Cited by 166 · cited 1×
“ADS1299 is one of the conventional amplifiers used to implement an EEG ampli-fier and can also be used to measure the impedance of an electrode [46], [47].”
MethodMethodological aspects of EEG and body dynamics measurements during motion
Reis, P. M. R., Hebenstreit, F., Gabsteiger, F., von Tscharner, V., & Lochmann, M. (2014) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2014.00156 · Cited by 159
A review of electrodes for the electrical brain signal recording
Im, C., & Seo, J.-M. (2016) · Biomedical Engineering Letters · doi:10.1007/s13534-016-0235-1 · Cited by 154 · cited 1×
“Although the quality of the final signal is associated with the compatibility between the amplifier and electrodes, most of the exisiting literature has focused on electrodes instead of the entire sys-tem [28]–[30].”
BackgroundA 5 $\mu$W/Channel Spectral Analysis IC for Chronic Bidirectional Brain–Machine Interfaces
Avestruz, A.-T., Santa, W., Carlson, D., Jensen, R., Stanslaski, S., Helfenstine, A., & Denison, T. (2008) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2008.2006460 · Cited by 143
Ground-Free ECG Recording with Two Electrodes
Thakor, N. V., & Webster, J. G. (1980) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1980.326595 · Cited by 143
Kamousi, B., Liu, Z., & He, B. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847386 · Cited by 140
Lin, C.-T., Ko, L.-W., Chang, M.-H., Duann, J.-R., Chen, J.-Y., Su, T.-P., & Jung, T.-P. (2010) · Gerontology · doi:10.1159/000230807 · Cited by 133 · cited 1×
“It has been productively used in epilepsy, Parkinson’s disease, narcolepsy, depression, motor impairment, sports, entertainment, and computer-related functions [8]–[16].”
MethodA Wearable 8-Channel Active-Electrode EEG/ETI Acquisition System for Body Area Networks
Xu, J., Mitra, S., Matsumoto, A., Patki, S., Hoof, C. A. V., Makinwa, K. A. A., & Yazıcıoğlu, R. F. (2014) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2014.2325557 · Cited by 133
New horizons in ambulatory electroencephalography
Waterhouse, E. (2003) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2003.1213629 · Cited by 129
Ultra-High Input Impedance, Low Noise Integrated Amplifier for Noncontact Biopotential Sensing
Yu, M., Maier, C., & Cauwenberghs, G. (2011) · IEEE Journal on Emerging and Selected Topics in Circuits and Systems · doi:10.1109/jetcas.2011.2179419 · Cited by 129 · cited 2×
“Numerous techniques have been proposed to overcome this challenge, such as an auxiliary path [80], a positive feedback loop [81], bootstrap resistor [82], [83], or negative capacitor [84].”
MethodNovel Active Comb-Shaped Dry Electrode for EEG Measurement in Hairy Site
Huang, Y.-J., Wu, C., Wong, M., & Lin, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2347318 · Cited by 125 · cited 1×
“Furthermore, electrodes are not always metal and flat [19], [20], and a wide-variety of structures and techniques are used to capture EEG signals.”
BackgroundA 15-Channel Digital Active Electrode System for Multi-Parameter Biopotential Measurement
Xu, J., Busze, B., Hoof, C. A. V., Makinwa, K. A. A., & Yazıcıoğlu, R. F. (2015) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2015.2422798 · Cited by 121
Analysis for the change of skin impedance
Yamamoto, T., & Yamamoto, Y. (1977) · Medical & Biological Engineering & Computing · doi:10.1007/bf02441041 · Cited by 121 · cited 1×
“As the skin under the electrode starts to sweat, the contact between electrode and skin will be improved [98].”
BackgroundKrachunov, S., & Casson, A. J. (2016) · Sensors · doi:10.3390/s16101635 · Cited by 116 · cited 5×
“In [57], a 3D-printed dry electrode was proposed with different length pins, to accommodate people with different hair densities.”
BackgroundInfluentialThe Thickness of Human Scalp: Normal and Bald
Hori, H., Moretti, G., Rebora, A., & Crovato, F. (1972) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12540633 · Cited by 115 · cited 2×
“5) [35].”
BackgroundZou, Y., Nathan, V., & Jafari, R. (2014) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2370646 · Cited by 112 · cited 1×
“fier and can also be used to measure the impedance of an electrode [46], [47].”
BackgroundAn EEG-NIRS Multimodal SoC for Accurate Anesthesia Depth Monitoring
Ha, U., Lee, J., Kim, M., Roh, T., Choi, S., & Yoo, H. (2018) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2018.2810213 · Cited by 102 · cited 1×
“For example, in [93], the authors achieved 0.44 µ V rms from 0.5 to 250 Hz, while their input impedance was 1 G ! at 60 Hz.”
BackgroundInfluentialMicro-spike EEG electrode and the vacuum-casting technology for mass production
Ng, W. C., Seet, H. L., Lee, K. S., et al. (2008) · Journal of Materials Processing Technology · doi:10.1016/j.jmatprotec.2008.10.051 · Cited by 100 · cited 2×
“The micro-pillar was used to penetrate the hair, while the micro-tips were designed to penetrate the stratum corneum [79].”
BackgroundMethodDias, N. S., Carmo, J. P., da Silva, A. F., Mendes, P. M., & Correia, J. H. (2010) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2010.09.016 · Cited by 90 · cited 3×
“Various materials have been investigated to identify the best quality of the recorded signal, including platinum, PEDOT/PSS [72], sintered Ag/AgCl, disposable Ag/AgCl, silver, gold-plated silver, stainless steel, tin [57], [66], and iridium oxide [34].”
BackgroundMethodInfluentialLow-cost active electrode improves the resolution in biopotential recordings
MettingVanRijn, A. C., Kuiper, A. P., Dankers, T. E., & Grimbergen, C. A. (2002) · Proceedings of 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.1996.656866 · Cited by 88 · cited 1×
“Furthermore, a low output impedance amplifier helps to suppress cable artifacts and eliminate the need of using a shield for wires [45].”
BackgroundClinical application of an active electrode using an operational amplifier
Nishimura, S., Tomita, Y., & Horiuchi, T. (1992) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.161342 · Cited by 69
ENOBIO dry electrophysiology electrode; first human trial plus wireless electrode system
Ruffini, G., Dunne, S., Farrés, E., Lozano, I., Watts, P. C. P., Silva, S. R. P., Grau, C., Fuentemilla, L., Marco‐Pallarés, J., & Vandecasteele, B. (2007) · Conference proceedings · doi:10.1109/iembs.2007.4353895 · Cited by 67
Flexible PDMS electrode for one-point wearable wireless bio-potential acquisition
Chen, C.-Y., Chang, C.-L., Chien, T., & Luo, C. (2013) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2013.08.010 · Cited by 54 · cited 1×
“The second sub-category consists of dry electrodes with flexible polymer micro-needles and substrate, such as polydimethylsiloxane (PDMS) [70] or SU-8 [71] to increase user comfort.”
BackgroundA Motion Interference-Insensitive Flexible Dry Electrode
Zhang, H., Pei, W., Chen, Y., Guo, X., Wu, X. M., Yang, X., & Chen, H. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2485269 · Cited by 51 · cited 2×
“A motion interference-insensitive flexible dry electrode was produced in [55].”
Microelectronic skin electrode
Padmadinata, F. Z., Veerhoek, J. J., van Dijk, G. J. A., & Huijsing, J. H. (1990) · Sensors and Actuators B Chemical · doi:10.1016/0925-4005(90)80257-z · Cited by 48 · cited 1×
“In other words, they consume power solely to convert impedance [40]–[43].”
BackgroundExtraction of SSVEP signals of a capacitive EEG helmet for Human Machine Interface
Oehler, M., Neumann, P., Becker, M., Curio, G., & Schilling, M. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4650211 · Cited by 43 · cited 1×
“Different types of dry electrodes: A) insulated electrode [22], B) non-contact electrode [23], and C) dry contact electrode [24]. in some cases.”
BackgroundDynamics of slow wave activity in narcoleptic patients under bed rest conditions
Nobili, L., Besset, A., Ferrillo, F., Rosadini, G., Schiavi, G., & Billiard, M. M. (1995) · Electroencephalography and Clinical Neurophysiology · doi:10.1016/0013-4694(95)00138-7 · Cited by 43
Nathan, V., & Jafari, R. (2015) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2015.2471080 · Cited by 34 · cited 1×
“In [89], configurable dry contact electrodes were studied, in which noisy pins were found and the corresponding signal is eliminated.”
BackgroundStudy of impedance spectra for dry and wet EarEEG electrodes
Kappel, S. L., & Kidmose, P. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7319063 · Cited by 33 · cited 1×
“[56] at higher frequencies.”
Capacitively-Coupled Chopper Amplifiers
Fan, Q., Makinwa, K. A. A., & Huijsing, J. H. (2016) · Analog circuits and signal processing/Analog circuits and signal processing series · doi:10.1007/978-3-319-47391-8 · Cited by 30
Lee, J. S., Han, C. M., Kim, J. H., & Park, K. S. (2015) · Electronics Letters · doi:10.1049/el.2015.1873 · Cited by 29 · cited 2×
“SAHARASYS [61], E) reverse-curve-arch-shaped dry electrode [62], and F) micro-needle arrays dry electrode [63].”
BackgroundYu, Y.-H., Lu, S.-W., Liao, L., & Lin, C. (2014) · IEEE Journal of Translational Engineering in Health and Medicine · doi:10.1109/jtehm.2014.2367518 · Cited by 27 · cited 1×
“comparison with wet electrodes [95]–[97].”
Patterned Vertical Carbon Nanotube Dry Electrodes for Impedimetric Sensing and Stimulation
Abu-Saude, M., & Morshed, B. I. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2449301 · Cited by 23 · cited 2×
“A vertical patterned carbon nanotube was proposed in [73].”
BackgroundLow Power Active Electrode ICs for Wearable EEG Acquisition
Xu, J., Yazıcıoğlu, R. F., Hoof, C. A. V., & Makinwa, K. A. A. (2018) · Analog circuits and signal processing/Analog circuits and signal processing series · doi:10.1007/978-3-319-74863-4 · Cited by 18
A Self-Wetting Paper Electrode for Ubiquitous Bio-Potential Monitoring
Guo, X., Pei, W., Wang, Y., Gong, Q., Zhang, H., Xing, X. F., Xie, Y., Gui, Q., & Chen, H. (2017) · IEEE Sensors Journal · doi:10.1109/jsen.2017.2684825 · Cited by 18 · cited 1×
“epidermis, a self-wetting paper electrode was proposed in [64].”
BackgroundZhao, Z., Ivanov, K., Lubich, L., Omisore, O. M., Mei, Z., Fu, N., Chen, J., & Wang, L. (2018) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2018.8513457 · Cited by 17 · cited 2×
“[58].”
BackgroundA feasibility study on a single-unit wireless EEG sensor
Luan, B., Jia, W., Thirumala, P. D., Balzer, J. R., Gao, D., & Sun, M. (2014) · International Conference on Signal Processing · doi:10.1109/icosp.2014.7015401 · Cited by 15 · cited 1×
“A low-impedance, skin-grabbing dry electrode was proposed in [90], [91] to solve this problem.”
BackgroundReducing the noise level of EEG signal acquisition through reconfiguration of dry contact electrodes
Nathan, V., & Jafari, R. (2014) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2014.6981790 · Cited by 14 · cited 1×
“variations in scalp surface due to sweat, hair density, and the shape of the skull [88].”
BackgroundDry electrode design and performance evaluation for EEG based BCI systems
Lee, S., Shin, Y., Woo, S., Kim, K., & Lee, H.-N. (2013) · International Winter Conference on Brain-Computer Interface (BCI) · doi:10.1109/iww-bci.2013.6506627 · Cited by 11 · cited 1×
“Furthermore, electrodes are not always metal and flat [19], [20], and a wide-variety of structures and techniques are used to capture EEG signals.”
BackgroundFabrication of polymeric dry microneedle electrodes coated with nanoporous parylene
Nishinaka, Y., Jun, R., Prihandana, G. S., & Miki, N. (2013) · International Conference on Solid-State Sensors, Actuators and Microsystems · doi:10.1109/transducers.2013.6627021 · Cited by 9 · cited 1×
“The second sub-category consists of dry electrodes with flexible polymer micro-needles and substrate, such as polydimethylsiloxane (PDMS) [70] or SU-8 [71] to increase user comfort.”
Background
Cited by
Soft wearable devices for deep-tissue sensing
Lin, M., Hu, H., Zhou, S., & Xu, S. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00427-y · Cited by 334
Metadata
- Authors
- Habibzadeh Tonekabony Shad, E., Molinas, M., & Ytterdal, T.
- Venue
- IEEE Sensors Journal
- Published
- 2020-12-15
- Volume
- 20
- Issue
- 24
- Pages
- 14565-14577
- ISSN
- 1530-437X (ISSN-L)
- 1558-1748
- 2379-9153
- OpenAlex
- W3046015843
- Cited by
- 173
- References
- 80
- FWCI
- 10.14
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.