A Compact, Low Input Capacitance Neural Recording Amplifier
Abstract
Conventional capacitively coupled neural recording amplifiers often present a large input load capacitance to the neural signal source and hence take up large circuit area. They suffer due to the unavoidable trade-off between the input capacitance and chip area versus the amplifier gain. In this work, this trade-off is relaxed by replacing the single feedback capacitor with a clamped T-capacitor network. With this simple modification, the proposed amplifier can achieve the same mid-band gain with less input capacitance, resulting in a higher input impedance and a smaller silicon area. Prototype neural recording amplifiers based on this proposal were fabricated in 0.35 μm CMOS, and their performance is reported. The amplifiers occupy smaller area and have lower input loading capacitance compared to conventional neural amplifiers. One of the proposed amplifiers occupies merely 0.056 mm(2). It achieves 38.1-dB mid-band gain with 1.6 pF input capacitance, and hence has an effective feedback capacitance of 20 fF. Consuming 6 μW, it has an input referred noise of 13.3 μVrms over 8.5 kHz bandwidth and NEF of 7.87. In-vivo recordings from animal experiments are also demonstrated.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Modern circuit designs and IC fabrication technologies have produced innovative designs that allow designers to relax the size-functionality compromises.
“However, modern circuit designs [77, 92] and IC fabrication technologies [18, 86] have resulted in innovative designs, enabling designers to relax such compromises.”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [92]
Although input transistor gate areas are usually large to reduce flicker noise, that reduction is counteracted by an increased capacitance multiplying effect from the parasitic gate area.
“Although the gate areas of the input transistors are usually large to reduce the flicker noise component, the flicker noise reduction is contradicted by the increased capacitance multiplying effect contributed by the parasitic gate area [92].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [92]
The gate area optimization technique proposed by Ng and Xu must be used to reduce input-referred noise from flicker noise and the capacitor transformer effect for complementary input pairs.
“Thus, the gate area optimization technique proposed by Ng and Xu [92] must be used to reduce the input-referred noise due to flicker noise and capacitor transformer effect for the complimentary input pairs.”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [92]
The gate areas of the input differential pair must be appropriately sized according to the method described by Ng and Xu.
“Hence, the gate areas of the input differential pair must be appropriately sized according to the method described by Ng and Xu [92].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [92]
Ng and Xu proposed relaxing the area-gain trade-off by introducing additional shunt elements onto the negative feedback network.
“However, Ng and Xu proposed relaxing this area–gain trade-off by introducing additional shunt elements onto the negative feedback network [92].”
From Implantable neurotechnologies: a review of integrated circuit neural amplifiers (2016) · cites this paper as [92]
Cites
Response of brain tissue to chronically implanted neural electrodes
Polikov, V. S., Tresco, P. A., & Reichert, W. M. (2005) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2005.08.015 · Cited by 1,896 · see citing works
Analog integrated circuit design
Hurst, S. L. (1998) · Microelectronics Journal · doi:10.1016/s0026-2692(97)00051-7 · Cited by 1,734 · see citing works
A low-power low-noise cmos for amplifier neural recording applications
Harrison, R. R., & Charles, C. (2003) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2003.811979 · Cited by 1,695 · see citing works
Design with Operational Amplifiers and Analog Integrated Circuits
Franco, S. (1988) · Cited by 898 · see citing works
A micropower low-noise monolithic instrumentation amplifier for medical purposes
Steyaert, M. S. J., & Sansen, W. M. C. (1987) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.1987.1052869 · Cited by 577 · see citing works
An Energy-Efficient Micropower Neural Recording Amplifier
Wattanapanitch, W., Fee, M., & Sarpeshkar, R. (2007) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2007.907868 · Cited by 441 · see citing works
The Design of Integrated Circuits to Observe Brain Activity
Harrison, R. R. (2008) · Proceedings of the IEEE · doi:10.1109/jproc.2008.922581 · Cited by 330 · see citing works
The 128-Channel Fully Differential Digital Integrated Neural Recording and Stimulation Interface
Shahrokhi, F., Abdelhalim, K., Serletis, D., Carlen, P. L., & Genov, R. (2010) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2041350 · Cited by 263 · see citing works
Design of Ultra-Low Power Biopotential Amplifiers for Biosignal Acquisition Applications
Zhang, F., Holleman, J. H., & Otis, B. P. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2011.2177089 · Cited by 237 · see citing works
256-Channel Neural Recording and Delta Compression Microsystem With 3D Electrodes
Aziz, J. N. Y., Abdelhalim, K., Shulyzki, R., Genov, R., Bardakjian, B. L., Derchansky, M., Serletis, D., & Carlen, P. L. (2009) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2008.2010997 · Cited by 184 · see citing works
Review of signal distortion through metal microelectrode recording circuits and filters
Nelson, M. J., Pouget, P., Nilsen, E. A., Patten, C. D., & Schall, J. D. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2007.12.010 · Cited by 167 · see citing works
Energy Efficient Low-Noise Neural Recording Amplifier With Enhanced Noise Efficiency Factor
Majidzadeh, V., Schmid, A., & Leblebici, Y. (2011) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2010.2078815 · Cited by 165 · see citing works
A micropower dry-electrode ECG preamplifier
Burke, M. J., & Gleeson, D. T. (2000) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.821734 · Cited by 164 · see citing works
A Versatile Integrated Circuit for the Acquisition of Biopotentials
Harrison, R. R. (2007) · 2007 IEEE Custom Integrated Circuits Conference · doi:10.1109/cicc.2007.4405694 · Cited by 156 · see citing works
Lee, J., Rhew, H.-G., Kipke, D. R., & Flynn, M. P. (2010) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2010.2052403 · Cited by 140 · see citing works
Analysis and Design of Tunable Amplifiers for Implantable Neural Recording Applications
Rezaee-Dehsorkh, H., Ravanshad, N., Lotfi, R., Mafinezhad, K., & Sodagar, A. M. (2011) · IEEE Journal on Emerging and Selected Topics in Circuits and Systems · doi:10.1109/jetcas.2011.2174492 · Cited by 103 · see citing works
Rodriguez-Perez, A., Ruiz-Amaya, J., Delgado-Restituto, M., & Rodriguez-Vazquez, Á. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2187352 · Cited by 103 · see citing works
Electrical Performance of Penetrating Microelectrodes Chronically Implanted in Cat Cortex
Kane, S. R., Cogan, S. F., Ehrlich, J., Plante, T. D., McCreery, D. B., & Troyk, P. R. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2248152 · Cited by 95 · see citing works
An integrated system for simultaneous, multichannel neuronal stimulation and recording.
Blum, R. A., Ross, J. D., Brown, E. A., & DeWeerth, S. P. (2007) · IEEE Transactions on Circuits and Systems I Regular Papers · doi:10.1109/tcsi.2007.906071 · Cited by 74 · see citing works
A 0.09 $\mu$ W Low Power Front-End Biopotential Amplifier for Biosignal Recording
Tseng, Y., Ho, Y., Kao, S.-T., & Su, C. (2012) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2012.2188029 · Cited by 65 · see citing works
Bazley, F. A., Hu, C., Maybhate, A., Pourmorteza, A., Pashai, N., Thakor, N. V., Kerr, C. L., & All, A. H. (2012) · Journal of Neurosurgery Spine · doi:10.3171/2012.1.spine11684 · Cited by 50 · see citing works
Effects of Glial Cells on Electrode Impedance Recorded from Neural Prosthetic Devices In Vitro
Frampton, J. P., Hynd, M. R., Shuler, M. L., & Shain, W. G. (2010) · Annals of Biomedical Engineering · doi:10.1007/s10439-010-9911-y · Cited by 37 · see citing works
Agrawal, G., Sherman, D. L., Maybhate, A., et al. (2010) · Journal of Clinical Neuroscience · doi:10.1016/j.jocn.2010.02.005 · Cited by 36 · see citing works
Perlin, G. E., Sodagar, A. M., & Wise, K. D. (2006) · 2006 International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/iembs.2006.260193 · Cited by 30 · see citing works
Chen, W., Yang, W.-C., Tsai, T.-Y., Chiueh, H., & Wu, C. (2011) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2011.6091185 · Cited by 20 · see citing works
A compact, low input capacitance neural recording amplifier with Cin/Gain of 20fF.V/V
Ng, K. A., & Xu, Y. P. (2012) · IEEE Biomedical Circuits and Systems Conference (BioCAS) · doi:10.1109/biocas.2012.6418430 · Cited by 12 · see citing works
Self-citationA Bionic Neural Link for peripheral nerve repair
Xu, Y. P., Yen, S.-C., Ng, K. A., Liu, X., & Tan, T. C. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346184 · Cited by 11 · see citing works
Self-citationImportant opamp configurations
Sansen, W. (2007) · Springer eBooks · doi:10.1007/0-387-25747-0_7 · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Implantable neurotechnologies: a review of integrated circuit neural amplifiers
Ng, K. A., Greenwald, E., Xu, Y. P., & Thakor, N. V. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-015-1431-3 · Cited by 83 · see citing works
Self-citationMulti-Channel Neural Recording Implants: A Review
Hashemi Noshahr, F., Nabavi, M., & Sawan, M. (2020) · Sensors · doi:10.3390/s20030904 · Cited by 66 · see citing works
- Authors
- Ng, K. A., & Xu, Y. P.
- Affiliations
- National University of Singapore (SG)
- Venue
- IEEE Transactions on Biomedical Circuits and Systems
- Published
- 2013-10
- Volume
- 7
- Issue
- 5
- Pages
- 610-620
- ISSN
- 1932-4545 (ISSN-L)
- 1940-9990
- OpenAlex
- W2047700412
- Cited by
- 81
- References
- 28
- Citations by year
- 2026: 3 · 2025: 3 · 2024: 5 · 2023: 4 · 2022: 5 · 2021: 13 · 2020: 9 · 2019: 6 · 2018: 8 · 2017: 6 · 2016: 4 · 2015: 8 · 2014: 7
- FWCI
- 3.49
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- IEEE Transactions on Biomedical Circuits and Systems (publishedVersion)
- PubMed (publishedVersion)
- National University of Singapore (submittedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Radio Frequency Integrated Circuit Design
- Keywords
- input capacitance, feedback capacitor, input impedance, chip area reduction, CMOS integrated circuit, low-power amplifier, input-referred noise, in vivo neural recording, compact amplifier design
- MeSH terms
- Amplifiers, Electronic, Animals, Equipment Design, Neurons, Signal Processing, Computer-Assisted, Rats, Wistar, Equipment Failure Analysis, Electric Capacitance, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review