Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice
Open accessAbstract
Ubiquitous blood pressure (BP) monitoring is needed to improve hypertension detection and control and is becoming feasible due to recent technological advances such as in wearable sensing. Pulse transit time (PTT) represents a well-known potential approach for ubiquitous BP monitoring. The goal of this review is to facilitate the achievement of reliable ubiquitous BP monitoring via PTT. We explain the conventional BP measurement methods and their limitations; present models to summarize the theory of the PTT-BP relationship; outline the approach while pinpointing the key challenges; overview the previous work toward putting the theory to practice; make suggestions for best practice and future research; and discuss realistic expectations for the approach.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Models that require a single calibration cuff measurement are convenient but potentially less accurate than those that use multiple measurements to form a patient-specific calibration curve”
From Wearable Photoplethysmography for Cardiovascular Monitoring (2022) · cites this paper as [20]
In the citing paper
“Pulse wave propagation delay between the genders indicates baseline differences in arterial stiffness with women displaying lower degrees of pulse arrival latency delay indicating higher vascular stiffness [106,107].”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [106]
In the citing paper
“New devices that promise to change the current paradigm of hypertension monitoring use either inflating wristband cuffs (e.g., Omron HeartGuide, currently undergoing clinical testing) or optical pulse wave-based metrics of BP [75,76].”
From Wearables and the Medical Revolution (2018) · cites this paper as [76]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · cited 1×
“A number of effective R-wave detection methods have been described, with the most popular being the Pan and Tomp-kins method [156].”
MethodExpert consensus document on arterial stiffness: methodological issues and clinical applications
Laurent, S., Cockcroft, J., Van Bortel, L., Boutouyrie, P., Giannattasio, C., Hayoz, D., Pannier, B., Vlachopoulos, C., Wilkinson, I., Struijker-Boudier, H., et al. (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl254 · Cited by 6,077 · cited 2×
“…1) reliably estimate aortic PWV in order to assess cardiovascular risk independent of BP more conveniently than traditionally employed tonometry [126]; 2) track BP changes (rather than BP levels) to help control BP in patients with known hypertension; and 3) indicate whether BP levels are not…”
BackgroundMethodPhotoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · cited 3×
“The advantage of reflectance-mode PPG is that it is less restrictive in terms of measurement locations [66], [67], [69].”
BackgroundMethodBurt, V. L., Whelton, P., Roccella, E. J., Brown, C., Cutler, J. A., Higgins, M., Horan, M. J., & Labarthe, D. (1995) · Hypertension · doi:10.1161/01.hyp.25.3.305 · Cited by 2,998 · cited 1×
“Hypertension is a major cardiovascular risk factor that is treatable, yet high BP detection and control rates are abysmally low, especially in low resource settings [1], [2].”
BackgroundAnderson, R. R., & Parrish, J. A. (1981) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12479191 · Cited by 2,456 · cited 1×
“Infrared (IR) and red optical wavelengths should be used for transmission-mode PPG due to the “optical water window” [66] and higher tissue penetration depth [66], [67], [74].”
MethodLakatta, E. G., & Levy, D. (2003) · Circulation · doi:10.1161/01.cir.0000048892.83521.58 · Cited by 2,244 · cited 1×
“We similarly hypothesized the model from reviews [24], [25], in vitro animal studies [32], [34], an in vivo animal study [35], an in vitro human study [36], and an in vivo human study [33].”
The Reference Values for Arterial Stiffness' Collaboration (2010) · European Heart Journal · doi:10.1093/eurheartj/ehq165 · Cited by 2,085
Schwartz, G., Tee, B. C.-K., Mei, J., Appleton, A. L., Kim, D. H., Wang, H., & Bao, Z. (2013) · Nature Communications · doi:10.1038/ncomms2832 · Cited by 2,042 · cited 1×
“Flexible strain or pressure sensors placed over a superficial artery can measure waveforms indicative of BP via the tonometric principle [109]–[112].”
BackgroundHuman blood pressure determination by sphygmomanometry.
Perloff, D., Grim, C. M., Flack, J. M., Frohlich, E. D., Hill, M. D., McDonald, M., & Morgenstern, B. Z. (1993) · Circulation · doi:10.1161/01.cir.88.5.2460 · Cited by 1,592 · cited 1×
“Auscultation is the standard clinical method [7].”
Advancements in Noncontact, Multiparameter Physiological Measurements Using a Webcam
Poh, M., McDuff, D., & Picard, R. W. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2010.2086456 · Cited by 1,514
Flagged first, then by citations
3 of the 212 references couldn’t be matched to a known paper.
Research works printed but not linked: 9
- 10.1016/0161-7346(79)90028-2 · DOI
- costanzo, 2003, Physiology
- froelicher, 2006, Exercise and the Heart
- katz, 2010, Physiology of the Heart
- patterson, 0, Development of an electrical impedance plethysmography system to monitor cardiac output, Proc 1st Annu Rocky Mountain Bioeng Symp, 56
- o’brien, 1993, The British hypertension society protocol for the evaluation of blood pressure measuring devices, J Hypertension, 11, 43s
- 2014, Standard for Wearable Cuffless Blood Pressure Measuring Devices
- 1993, Electronic or Automated Sphygmomanometers
- eckerle, 1988, Tonometry arterial, Encyclopedia of Medical Devices and Instrumentation
Cited by
Sorted by citations
Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0287-x · Cited by 895
Wearables and the Medical Revolution
Dunn, J., Runge, R., & Snyder, M. (2018) · Personalized Medicine · doi:10.2217/pme-2018-0044 · Cited by 654
“, Omron HeartGuide, currently undergoing clinical testing) or optical pulse wave-based metrics of BP [75,76].”
Cuffless Blood Pressure Estimation Algorithms for Continuous Health-Care Monitoring
Kachuee, M., Kiani, M. M., Mohammadzade, H., & Shabany, M. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2016.2580904 · Cited by 621
“where P0 , P1 , and Am are varying among individuals [7].”
BackgroundMethodInfluentialBlood Pressure Estimation from Photoplethysmogram Using a Spectro-Temporal Deep Neural Network
Slapničar, G., Mlakar, N., & Luštrek, M. (2019) · Sensors · doi:10.3390/s19153420 · Cited by 387
“The PTT approach was thoroughly tested in the past 10 years [15,16] and a number of slight variations were proposed, however, the underlying principle remains unchanged.”
BackgroundSelf-powered cardiovascular electronic devices and systems
Zheng, Q., Tang, Q., Wang, Z. L., & Li, Z. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0426-4 · Cited by 348
Metadata
- Authors
- Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S.
- Affiliations
- Michigan State University (US)
- University of Maryland, College Park (US)
- Georgia Institute of Technology (US)
- Palo Alto Research Center (US)
- Corresponding author
- Mukkamala, Ramakrishna
- Venue
- IEEE Transactions on Biomedical Engineering
- Published
- 2015-08
- Volume
- 62
- Issue
- 8
- Pages
- 1879-1901
- ISSN
- 0018-9294 (ISSN-L)
- 1558-2531
- PMC
- PMC4515215
- OpenAlex
- W2294418644
- Cited by
- 935
- References
- 212
- Citations by year
- 2026: 92 · 2025: 93 · 2024: 96 · 2023: 89 · 2022: 93 · 2021: 87 · 2020: 84 · 2019: 92 · 2018: 83 · 2017: 71 · 2016: 50 · 2015: 4
- FWCI
- 34.44
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- IEEE Transactions on Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Biomedical Imaging and Bioengineering
- National Institutes of Health (U01 EB018818)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Keywords
- blood pressure monitoring, pulse transit time, ubiquitous health monitoring, wearable sensors, hypertension detection, non-invasive blood pressure estimation
- MeSH terms
- Blood Pressure, Blood Pressure Determination, Clothing, Equipment Design, Humans, Signal Processing, Computer-Assisted, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review